READ THE HYBRESIS CONTROLLER AND PATCH INSTRUCTIONS FOR USE FOR
ADDITIONAL IMPORTANT INFORMATION.
REF: 199586
Charging Station - Instructions For Use
![](/html/f1/f1ad/f1adda00c95252b47ab7bc47144b4d0649e36c47566bb7969eda7b0c9a1e430a/bg2.png)
GLOSSARY OF SYMBOLS
This device may contain one or all of the following symbols:
Refer to Instruction Manual/Booklet
Council Directive 2002/96/EC concerning Waste Electrical and Electronic Equipment (WEEE). Indicates a
requirement not to dispose of WEEE as municipal waste. Contact your local distributor for information
regarding disposal of the unit and accessories.
Dangerous Voltage
Underwriters Laboratories, Inc., indicates product meets US and Canadian product safety standards.
This device complies with UL 60601-1 and CSA C22.2 No. 601-1-M90
Precautionary Instructions
Keep Dry
Keep Away from Sunlight
Class II Equipment
For Prescription Only
11N1
![](/html/f1/f1ad/f1adda00c95252b47ab7bc47144b4d0649e36c47566bb7969eda7b0c9a1e430a/bg3.png)
Warnings
• Keep out of the reach of children.
• Electronic monitoring equipment (such as ECG monitors and alarms) may not operate properly when this equipment
is in use.
• Electrodes should be removed before using equipment that generates high frequency or high energy electromagnetic
radiation such as welding equipment, surgical electrocautery units, shortwave and microwave therapy. Operation in
close proximity (< 3m) to this equipment may startle the user by causing output instability or improper operation of
the stimulator. Simultaneous connection to the patient may result in burns and possible damage to the stimulator.
• Care must be taken when operating this equipment adjacent to or stacked with other equipment. Potential
electromagnetic or other interference could occur to this or to the other equipment. Try to minimize this interference
by not using other equipment in conjunction with it.
• Do not operate these units in an environment where other devices are being used that intentionally radiate
electromagnetic energy in an unshielded manner. Portable and mobile RF communications equipment can affect
Medical Electrical Equipment.
• This unit generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee
that interference will not occur in a particular installation. Harmful interference to other devices can be determined
by turning this equipment on and off. Try to correct the interference using one or more of the following: reorient or
relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet
on a different circuit from that to which the other device(s) are connected or consult the Empi Repair Department for help.
• The system is not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide.
• Use only accessories that are specially designed for this unit and recommended by Empi. The use of other accessories
or cables may result in increased emissions or decreased immunity of this unit.
• Avoid exposure to water. Do not allow the charging station to be immersed in fluids.
Precautions
1
![](/html/f1/f1ad/f1adda00c95252b47ab7bc47144b4d0649e36c47566bb7969eda7b0c9a1e430a/bg4.png)
DESCRIPTION
The Hybresis Charging Station is designed to charge and store Hybresis
Controllers.
OPERATION
The Hybresis Charging Station charges up to four Hybresis Controllers,
maintains them in a fully charged state and provides a convenient place
for storage.
General -
When a Controller is inserted into one of the charging bays, the Charging
Station will automatically detect the battery level of the Controller and
will begin to charge it.
2
![](/html/f1/f1ad/f1adda00c95252b47ab7bc47144b4d0649e36c47566bb7969eda7b0c9a1e430a/bg5.png)
Light Status Corrective Action
Green Charge Complete;
The Controller is fully charged
and ready for use.
None Required
Yellow Charging;
The Controller is charging
normally.
None Required
Red-
Continuous
or Flashing
Charging Fault;
Controller battery not charging
properly.
Return Controller to
Empi for service
Charging Station Lights -
The two windows on each of the charging bays indicate the following:
On rare occasions, the problem may be due to a defective Charger. If you have
another Controller, please charge it to determine if the Charger is functioning
correctly or not. If the same problem occurs with multiple Controllers, then
the Charger may be defective.
3