Pump Assembly Parts List ..................................1
Mattress Assembly Parts List .................................7
CHAPTER 9
Service Contact Details .............................1
Issue 1
Jun 2001SER0001Page (v)
ContentsNimbus / DFS Service Manual
Issue 1
Page (vi)SER0001Jun 2001
Nimbus / DFS Service ManualIntroduction
CHAPTER 1
INTRODUCTION
1About This Manual
Huntleigh Healthcare strongly recommend that their equipment is only serviced by trained
personnel and provide courses for customers who wish to become licensed to service their
own equipment. In no event will Huntleigh Healthcare be responsible for any service
performed by customers or third parties.
This manual contains information on maintenance, servicing, repair, troubleshooting and testing for
the Nimbus
and understand this manual before attempting to service or repair the equipment.
Warnings, Cautions and Notes
WARNINGS given in this manual identify possible hazards in procedures or conditions which, if not
correctly followed, could result in death, injury or other serious adverse reactions.
Cautions given in this manual identify possible hazards in procedures or conditions which, if not
correctly followed, could result in equipment failure or damage.
Notes given in this manual are used to explain or amplify a procedure or condition.
®
2, Nimbus 3, DFS 2 and DFS 3 systems which comprise a pump and a mattress. Read
WARNINGS:
BEFORE PERFORMING ANY SERVICE OR MAINTENANCE PROCEDURES,
ENSURE THAT THE EQUIPMENT HAS BEEN ADEQUATELY
DECONTAMINATED.
VOLTAGES IN EXCESS OF 30 VOLTS RMS OR 50 VOLTS DC CAN, IN
CERTAIN CIRCUMSTANCES, BE LETHAL. WHEN WORKING ON
EQUIPMENT REQUIRING EXPOSURE TO LIVE, UNPROTECTED
CONDUCTORS WHERE SUCH VOLTAGES ARE PRESENT, EXTREME CARE
MUST BE EXERCISED.
2Technical Description
The Nimbus 2, Nimbus 3, DFS 2 and DFS 3 are Dynamic Flotation Systems for use in aiding the
prevention, treatment and management of pressure sores.
It is a powered, air-filled mattress replacement system with microprocessor-aided control and alarm
functions which have LED indication on the front panel of the pump.
Operating Principle
The patient is supported by twenty cells which are enclosed by a polyurethane coated fabric cover.
This cover has waterproof, vapour permeable, and two-way stretch properties.
The system incorporates a sensor that automatically adjusts the support pressure within the system to
accommodate different weight patients and their weight density or distribution. This sensor pad,
known as the Auto-Matt
minimum to support that patient in a given position.
Issue 1Chap 1
Jun 2001SER0001Page 1
®
sensor, optimises the pressure to which the cells are inflated to be the
IntroductionNimbus / DFS Service Manual
A
c
Small adjustments can be made to the mattress pressure, using the comfort control knob on the
pump. This is a patient comfort feature to make small adjustments to the softness or firmness of the
mattress.
®
The Nimbus 3 and DFS 3 mattress also includes the HeelGuard
which is designed to maximise pressure relief under the heels.
Dynamic Mode
In dynamic mode the support surface beneath the patient is cycled every 10 minutes providing
periods of pressure relief for the whole body.
Figures 1 and 2 show how the system operates in dynamic and static modes. Fig 2 shows the
pneumatic circuit in principle. A dual compressor system is used to provide a source of air supply.
This is fed to a high pressure switch, the Auto-Matt sensor and a rotary valve. The high pressure
switch is used to detect system over-pressures through blockages. The low pressure switch, which is
located in the pressure control assembly, is used to detect if the air pressure in the system falls below
minimum. The rotary valve channels the air supply as shown below:
C to outlet A with B at atmosphere (D)- A & C Cells Inflated
•
C to outlet A and B- All Cells Inflated (A,B&C)
•
C to outlet B with A at atmosphere (D)- B & C Cells Inflated
•
C to outlet A and B- All Cells Inflated (A,B&C)
•
special powered-down feature
This then allows cells A and B in the mattress to be inflated and deflated in a cycle period of ten
minutes.
Anti-sink
Torso Section
Upper cells alternate,
lower cells semi-dynamic
C
CCCC
C
CCCC
C
CCCC
Head Section
Semi-dynami
AAAA
CCC
AAAA
CCC
C
AAAA
CCCCCC
CCC
C
C
lternating
Static
Alternating
1
2
3
Lower Body Section
Upper and lower cells
alternate
BBBB
AAAA
AAABBBBA
All cells constantly inflated
AAA
BBBB
A
BBBB
A
Semi-dynamic cells
BBBB
A
BBBB
A
Static
Chap 1Issue 1
Page 2SER0001Jun 2001
4
AAABBBBA
All cells constantly inflated
Figure 1 - Mattress Cell Inflation Cycle
BBBB
A
C
CCCC
AAAA
CCC
CCC
Nimbus / DFS Service ManualIntroduction
In the mattress there is a third section of cells labelled as C. These are fed through partial non-return
valves from cells A and B and are therefore never at zero pressure. These are positioned as shown in
Fig 1 to prevent the torso section from ‘bottoming’ during dynamic mode and when a patient is sitting
upright. Semi-dynamic pressure is also applied to the head cells (C) and the heel protector (above
CPR unit) for added patient comfort.
On Nimbus 3 and DFS 3, loop sheets, the heel cells have a special powered-down feature to
facilitate fast deflation and thereby improve pressure relief under the heels.
Static Mode
In static mode the support surface is kept constant and provides pressure reduction without cycling.
The rotary valve is held in a constant position so that cells A, B, C and the heel protector are equally
inflated.
Compressor 1
High Pressure
Switch
Pressure Control
Adjuster
Compressor 2
Restrictor
Atmosphere
Low Pressure
Switch
(Atmosphere)
D
C
B
Rotary
Valve
52314
Not
Used
A
Pump
Tubeset
Cells A
Cells B
Partial Non
Return Valves
Sensor Pad (Auto Matt)
Cells C
Mattress
Figure 2 - Pneumatic System Diagram
Issue 1Chap 1
Jun 2001SER0001Page 3
IntroductionNimbus / DFS Service Manual
Auto-Matt
The Auto-Matt sensor consists of a flexible air sensor pad which is located underneath the mattress
cells. This acts as a pressure regulator and accommodates changes in patient weight and position in
the following manner:
It is pneumatically connected in parallel to the mattress cells and when it is not occluded it channels
the air supply away from the cells. Occlusion of the Auto-Matt sensor is by patient weight and position.
In this manner it allows internal mattress air pressures to be compensated to that required to support
the patient.
With no patient on the mattress the Auto-Matt sensor is not occluded and pressure in the mattress is
governed by the pressure control shown in Fig 2. This is in turn set to a pressure range which can be
slightly adjusted to change mattress pressures for patient comfort.
With a patient on the mattress, pressure is exerted onto the cells and this can partially or totally
occlude the Auto-Matt sensor. This results in more air being supplied to the cells increasing their
internal pressure until full patient support is achieved. Should the patient change to a position which
necessitates a lower pressure, air is released via the Auto-Matt sensor through the pressure control
valve until the correct lower pressure is achieved in the cells.
System Start Up Sequence
After the pump is switched on the start up sequence is as follows:
All display indicators illuminate for a few seconds.
•
Low Pressure and Wait indicators remain illuminated while the mattress inflates (approx
•
20 Min). All the other indicators go out.
• The pump runs one compressor until the rotor turns to the cross over position (within 5 Min).
•
The rotor then depresses the microswitch which stops the rotor and activates the second
compressor.
•
Both compressors run for 20 minutes or when the mattress reaches the minimum pressure
setting.
•
When the minimum pressure setting is reached the Low Pressure and Wait indicators go out.
•
The pump cuts down to one compressor and the rotor starts to rotate.
•
In Dynamic Mode the compressors will alternate every 4 hours.
•
In Static Mode the Coils (4 off) will alternate every 4 hours (one at a time). The compressors
alternate depending on which coil is working (there are two coils in each compressor).
Alternating Head Cells Modification
The Nimbus and DFS beds are normally supplied with the top three (head) cells configured so that
they are inflated to a specific pressure. Although the pressure in the head cells normally remains
constant, the pressure in the rest of the cells slowly alternates.
The Alternating Head Cells Modification converts the static head cells into alternating cells. The
modification can be applied in two configurations that follow:
Configuration 1 converts all three head cells from static pressure to alternating pressure.
Configuration 2 converts the second and third cells from static pressure to alternating pressure. The
top head cell remains as a static pressure cell.
Chap 1Issue 1
Page 4SER0001Jun 2001
Nimbus / DFS Service ManualMaintenance
CHAPTER 2
MAINTENANCE
1Pump Maintenance Checks
If any parts are found to be damaged they must be replaced in accordance with Chapter 3, Pump
Repair.
1.1Visually inspect the following for damage, wear and potential faults:
1.1.1Outer CasingCheck for damage and security.
1.1.2Switches, Control KnobsCheck function.
1.1.3Hanging BracketCheck for damage and security.
1.1.4Outlet portsCheck for damage and free from obstructions.
1.1.5Rubber feet (8 0ff)Check for security.
1.1.6FusesCheck condition.
1.1.7Front panelCheck for damage to LED’s and buttons.
1.1.8Mains/Power leadCheck for damage and security of plug.
1.1.9LabelsCheck all labels are present and legible.
1.2Disassemble the Rear Case (Refer to Page 5, Section 4) and visually inspect the following for
damage, wear, security and potential faults:
1.2.1Compressor
1.2.2Rotary Valve Assembly
1.2.3Printed Circuit Boards (PCB’s)
1.2.4Pressure Control Assembly
1.2.5High Pressure Switch
1.2.6Silencer Bag
1.2.7Bio Filter (Nimbus 3 and DFS 3 Only)
1.2.8Tubing and Wiring
1.2.9Screws, Nuts and bolts
1.3Remove the compressor lids and visually inspect the following for damage, wear and potential
faults:
1.3.1Coils
1.3.2Elbows and connectors
1.3.3Filters and gaskets
Issue 1Chap 2
Jun 2001SER0001Page 1
MaintenanceNimbus / DFS Service Manual
1.3.4Shuttle Valve Assemblies
2Pump Soak Times
2.1A pump must be soaked with its ports occluded for the following reasons:
To ensure that the pump is at normal running temperature prior to pneumatic performance
•
tests being carried out.
To stress test major/critical parts which have been replaced or repaired.
•
2.2All pumps should be soaked for a period of 30 to 60 minutes prior to any pneumatic
performance test being carried out.
2.3If any of the following major/critical components have been replaced or repaired, then it is
recommended that the pump be soaked for 12 hours prior to being tested:
Valve Shuttle Assemblies
•
Silencer Bags
•
Printed Circuit Boards (PCB’s)
•
Compressor
•
Rotary Valve Assembly
•
3Pump Service
A pump should be serviced every 12 months or when thesymbol is illuminated. To carry out
a service on the pump, do the following:
3.1Carry out the Pump Maintenance Checks (Refer to Page 1, Section 1).
3.2Replace the following components:
3.2.1Case Seal (Refer to Chapter 3, Page 5, Section 4 and 5).
5.3Carry out an inflation test on the mattress in accordance with Chapter 6, Page 15, Section 9.
Issue 1Chap 2
Jun 2001SER0001Page 3
MaintenanceNimbus / DFS Service Manual
Chap 2Issue 1
Page 4SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
CHAPTER 3
PUMP REPAIR
1General
This chapter details repair procedures for the Nimbus 2, Nimbus 3, DFS 2 and DFS 3 pumps. All
repairs should be carried out by Huntleigh approved service personnel.
After carrying out a service or any repairs, the pump must be soaked for the recommended time
(Refer to Chapter 2, Page 2, Section 2) and tested for serviceability. The table below defines the test
requirements which must be carried out following certain repairs:
To carry out a flow, pressure and function test on the pump refer to Chapter 6, Page 1, Section 1.
To carry out the electrical tests on the pump refer to Chapter 6, Page 13, Section 8.
Table 1- Repair to Testing Requirements
Components / Assemblies
CompressorYesYes
Pressure ControlYesYes
Rotors & StatorsYesYes
Silencer BagYesNo
High / Low Pressure SwitchesYesYes
Shuttle AssemblyYesYes
CoilsYesYes
Elbows & ConnectorsYesNo
AV MountsYesYes
PCBYesYes
Front CasingYesYes
Rear CasingNoNo
BatteryNoYes
Flow/Pressure & Function
Test
Electrical Safety
SwitchesNoYes
FusesNoYes
Electrical ConnectorsNoYes
Mains/Power LeadNoYes
Issue 1Chap 3
Jun 2001SER0001Page 1
Pump RepairNimbus / DFS Service Manual
20
40 50
30
Figure 3 - Removing and installing the Hanging Bracket
Table 2 - Hanging Bracket Part List
ItemPart NumberDescriptionQty
10151123Hanging Bracket Assembly1
20151427Bracket1
30151428Wire1
40FAS046Screw M3 x 10 CSK2
50FAS216Nut M3 x 0.5 Euro Series2
Chap 3Issue 1
Page 2SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
2Removing the Hanging Bracket
2.1Ease the bottom of the hanging bracket off of the base of the pump.
2.2Rotate the hanging bracket (Fig 3) and remove it from the handle recess in the back of the
pump.
3Installing the Hanging Bracket
3.1Lay the pump face down and insert the bridge of the hanging bracket (Fig 3) into the handle
recess of the pump.
3.2Pull the bottom of the hanging bracket over the bottom of the pump and push firmly downwards
so that the bracket clips into place around the base of the pump.
Issue 1Chap 3
Jun 2001SER0001Page 3
Pump RepairNimbus / DFS Service Manual
t
10
30
50
40
20
Ends of cord
meet at this poin
50
Figure 4 - Rear Case Assembly
Chap 3Issue 1
Page 4SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
4Disassembling the Rear Case
WARNING:
BEFORE DISMANTLING THE PUMP UNIT , ENSURE THE UNIT HAS BEEN
ISOLATED FROM THE POWER SUPPLY BY REMOVING THE MAINS/POWER
PLUG FROM THE WALL SOCKET.
4.1Remove the hanging bracket from the pump (Refer to Page 3, Section 2).
4.2Place the pump unit on a flat surface with the front case (Fig 4, Item 20) face down.
4.3Unscrew the 8 pan head screws (Fig 4, Item 30) and lift the rear case (Fig 4, Item 10) away
from the front case (Fig 4, Item 20). Keep the screws (Fig 4, Item 30) in a safe place.
4.4Carefully remove the neoprene sponge cord (Fig 4, Item 40) and inspect for damage.
5Assembling the Rear Case
5.1Assembly of the rear case is in the reverse order of disassembly.
5.2Install the hanging bracket to the pump (Refer to Page 3, Section 3).
Table 3 - Rear Case Parts List
ItemPart NumberDescriptionQty
10151052Rear Case Assembly1
20151307Front Case1
30FAS 188Pan Head Screws M4 x 30 mm8
40151105Neoprene Sponge Cord (Dia 2 mm)1
50500381Rubber foot8
Issue 1Chap 3
Jun 2001SER0001Page 5
Pump RepairNimbus / DFS Service Manual
40
20
Slots
10
30
Figure 5 - Pump Filter Assembly
Chap 3Issue 1
Page 6SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
6Disassembling the Pump Filter
6.1Disassemble the Rear Case (Refer to Page 5, Section 4).
6.2Lay the Rear Case on a flat surface and remove the six screws (Fig 5, Item 40) and keep them
in a safe place.
6.3Remove the Filter Housing Plates (Fig 5, Item 20) and the Filter (Fig 5, Item 30).
6.4Discard the old Filter.
7Assembling the Pump Filter
7.1Assembly of the new Pump Filter is in the reverse order of disassembly. Make sure the slots in
the Filter Housing Plates (Fig 5, Item 20) are uppermost in the positions shown in Fig 5.
7.2Assemble the Rear Case (Refer to Page 5, Section 5).
Table 4 - Pump Filter Parts List
ItemPart NumberDescriptionQty
10151308Rear Case1
20151311Filter Housing Plate2
30151312Filter1
40FAS179Pan Hd Self Tapping Screw (Dia 2.5mm x 10 mm)6
Issue 1Chap 3
Jun 2001SER0001Page 7
Pump RepairNimbus / DFS Service Manual
Compressor Earth Wire
140
40
50
20
C
30
A
160
B
Rotor Assembly
140
150
110
60
70
Compressor
Earth Wires
PCB Board
Front Panel
60
170
80
90
100
110
120
10
45
70
Power Supply
160
140
130
Figure 6 - Remove the Double Compressor Assembly and
Disassemble the AV mounts
Chap 3Issue 1
Page 8SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
8Removing the Double Compressor Assembly
8.1Disassemble the Rear Case (Refer to Page 5, Section 4).
8.2Remove the screws (Fig 6, Item 50) and the compressor lids (Fig 6, Item 20). Lift the four pairs
of power cables, together with the rubber grommets, out from the side of each compressor
casing. Make sure the blanking grommets (Fig 6, Item 40) on the inner side of each
compressor remain in the compressor casing.
8.3If necessary, replace the items that follow:
8.3.1The compressor Shuttles (Refer to Page 15, Section 15).
8.3.2The compressor Coils (Refer to Page 15, Section 16).
8.4Disconnect the high pressure switch hose and the compressor hose from the compressor
(Fig 6, A and B) and disconnect the silencer bag (Page 46, Fig 26, Item 10) from the
compressor hose.
8.5For Nimbus 3 and DFS 3 pumps only, remove the Bio Filter (Refer to Page 49, Section 41).
8.6Cut and remove any cable ties as necessary and disconnect the 2 compressor power cables
from the PCB.
8.7Remove the items that follow from the left compressor bracket:
8.7.1Screw (Fig 6, Item 80) and washers (Fig 6, Items 90 and 100).
8.7.2‘P’ Clip (Fig 6, Item 110).
8.8Remove the nuts, washers (Fig 6, Items 140 and 160) and the compressor earth wire from the
top of the compressor.
8.9Remove the nuts, washers and ‘P’ clips (Fig 6, Items 140, 150 and 110) together with the
compressor hose. The ‘P’ clips can remain on the hose.
8.10 Remove the double compressor assembly (Fig 6, Item 10).
9Installing the Double Compressor Assembly
9.1Carefully install the double compressor assembly into the pump onto the AV mounts on the
compressor brackets in the pump casing.
9.2Install the screw, the washers and the ‘P’ Clip (Fig 6, Items 80, 90, 100 and 110) to the left
compressor bracket. Make sure that the ‘P’ Clip is around the 2 of the compressor power cable
wires which form the longer loop from point C in Fig 6.
9.3Connect the 2 compressor power cables to the PCB (Refer to Fig 11 or Fig 15).
9.4Attach the high pressure switch hose to the compressor (Fig 6, A) and the silencer bag to the
compressor hose (Page 46, Fig 26, Item 10). Attach the compressor hose to the compressor
(Fig 6, B).
9.5If necessary, replace the compressor lid gaskets (Fig 6, Item 30) as follows:
9.5.1Remove the old gasket and clean the compressor lid.
9.5.2Peel the back off of the new gasket (Fig 6, Item 30) and stick in position on the lid.
Issue 1Chap 3
Jun 2001SER0001Page 9
Pump RepairNimbus / DFS Service Manual
9.6Make sure that the blanking grommets (Fig 6, Item 40) are installed in the compressor casings
and install the compressor lids (Fig 6, Item 20). Secure the lids with screws (Fig 6, Item 50).
9.7Install cable ties where they were removed.
9.8Assemble the rear case (Refer to Page 5, Section 5).
10Removing the AV Mounts
10.1 Remove the double compressor assembly (Refer to Page 9, Section 8).
10.2 Remove the items that follow from the earthing stud on the right compressor bracket:
10.2.1The nut and washer (Fig 6, Items 60 and 70).
10.2.2The 4 earth wire connectors (PCB, Compressor, Front Panel and Rotary).
Note : On Nimbus 3 and DFS 3 pumps there is only one earth wire for the PCB and Front
Panel.
10.2.3The nut, washer (Fig 6, Items 60 and 70) and the Power Supply earth wire connector.
10.3 Remove the screws (Fig 6, Item 120) and the left compressor bracket (Fig 6, Item 130) from the
pump casing.
10.4 Remove the screw and washers (Fig 6, Items 80, 90 and 100) from the right compressor
bracket
10.5 Remove the screws (Fig 6, Item 120) and the right compressor bracket from the pump casing
10.6 Remove the nuts, washers and AV mounts (Fig 6, Items 140, 160 and 170) from each
compressor bracket.
11Installing the AV Mounts
11.1 Install the AV mounts, the washers and the nuts (Fig 6, Items 170, 160 and 140) to each
compressor bracket (Fig 6, Item 130).
11.2 Install the right compressor bracket (with the earth label) into the pump casing. Secure the
bracket with a screw and washers (Fig 6, Items 80, 90 and 100) at the back, nearest the PCB.
11.3 Install screws (Fig 6, Item 120) to the remaining holes in the bracket.
11.4 Install the items that follow to the earthing stud on the right compressor bracket:
11.4.1The Power Supply earth wire connector.
11.4.2The washer and nut (Fig 6, Items 70 and 60).
11.4.3The 4 earth wire connectors (Rotary, Front Panel, Compressor and PCB).
Note : On Nimbus 3 and DFS 3 pumps there is only one earth wire for the PCB and Front
Panel.
11.4.4The washer and nut (Fig 6, Items 70 and 60).
11.5 Install the left compressor bracket and secure with the screws (Fig 6, Item 120).
Chap 3Issue 1
Page 10SER0001Jun 2001
Nimbus / DFS Service ManualPump Repair
Note : Do not install the screw, washers and ‘P’ clip to the left compressor bracket at this stage.
11.6 Install the double compressor assembly (Refer to Page 9, Section 9).