humimeter.com P2, LM5, P4 User Manual

Usermanual
Paper resp. Leather
Moisture Meter
P2 P4
LM5
Version 2.2_en
© Schaller GmbH
2005
Page 2 Schaller GmbH Version 2.2_en
Quality information:
Congratulations on purchasing this product at Messtechnik Schaller GmbH (Ltd.).
We would like to inform you that your measuring device corresponds to the CE norms EN50081 and EN50082. The function of your instrument has been examined and calibrated before delivery.
Application
P2: Mainly for very strong paper, packing paper,
cardboard
P4: Mainly for paper used for offset printing LM5: mainly for leather
First setting into operation:
Your measuring device should be supplied with a battery by the manufacturer. If this is not the case, open the battery compartment and stick a 9 V block battery or a charged accumulator onto the connecting contacts. Then put the battery in the battery compartment. Make sure that the power cable does not get caught in the lid. Press the power switch (ON) at the front of the instrument and the display will go on. If this is not the case, check if the batteries are charged. If you hold your measuring instrument in the air you will see a certain value on the display depending on the switch position. Now you can compare this value to the rated values according to tables on pages 9 and 10. Furthermore it is advisable to check the measuring device at the proof plane which is the bottom of the wooden case.
Version 2.2_en Schaller GmbH Page 3
Measuring procedure
Exact measuring results can best be achieved by adjusting the measuring instrument to the respective product temperature. So you should store the device near the object you wish to measure. A temperature difference of more than 5 °C results in a falsified measured value. Adjust the switch by shifting it to the correct position (refer to type of product). Turn on your measuring device and put it on the product with suitable pressure or move it with constant pressure lengthwise along the product roll to detect the wet streak. Take care that you hold your measuring instrument with a moderate but sufficient pressure on the product because the compressed density influences the measuring result. Make sure that you always exert the same amount of pressure since the density of the material increases when it is pressed tighter or softer; so you could get different measuring results every time you measure the material. The absolute density can immediately be seen on the LC-Display. Please note that the bigger the supporting surface of the sensor, the more exact is the measurement (refer to sketch). The device turns off automatically after 90 seconds or when the measuring range has been exceeded.
Firmly press the instrument on the pile in the
middle
Paper piles must be at least 2 cm thick, leather piles
at least 1 cm
incorrect
Press the instrument only lengthwise on the
paper roll
Page 4 Schaller GmbH Version 2.2_en
How to choose type of product
Since there is no unified manufacturing process of the measuring products the adjustment of the switch can not be standardized. But there is always the possibility of having your device calibrated by Messtechnik Schaller GmbH (Ltd.). In this case we need a pile of your product (2 cm thick) size A4 in a plastic wrapping that is as tight as possible. You can classify your products by means of a comparison measurement applying methods appropriate for verification according to DIN20287. Take a 2 cm thick pile with an average quantity of moisture size A5, weigh some of the sheets and dry them in a drying stove. Store the rest of the paper in a plastic wrapping to preserve the moisture. The dried sheets of paper are weighed again. By using the following formula the absolute moisture is calculated. Caution! The dried sheets of paper must not be used for measurement again.
Mn: mass of sheets with an average moisture. Mt : mass of dried sheets %F: calculated absolute moisture.
100% ×
=
Mn
MtMn
F
Now take the rest of the pile out of the plastic wrapping and measure it with your measuring device. Adjust the switch until the most exact value can be seen on the display and note down the position of the switch.
Version 2.2_en Schaller GmbH Page 5
Verification and correction of switch position
Now weigh the rest of the material sample, store it in more humid or drier ambient conditions than where your calibration has been carried out. Do not forget to adapt the measuring device to the corresponding temperature. Then weigh your pile once more, calculate the moisture value by means of weight applying the formula of control value calculation, measure this moisture value and compare the result with the display
Control value calculation:
M
ter
: Calculated dry mass of your control sample
Mnk : mass of the remaining material
(
)
100
%100 FMnk
Mter
×
=
M
nneu
: new mass by drying or moistening
%F
neu
: New moisture value for comparison
measurement
Now insert the switch position in the table.
Please note: Messtechnik Schaller GmbH (Ltd.) provides special versions of P2, P4 and LM5 adapted to certain products. Refer to additional information on leather, page 11.
Page 6 Schaller GmbH Version 2.2_en
Measuring method
Even if the kiln drying method is the only measuring procedure appropriate for verification according to DIN 20287 it is time consuming, in-situ application can very seldom be carried out; the method also has the disadvantage that the samples are destroyed. Schaller GmbH (Ltd.) has developed the P2, P4 and LM5 generation – a measuring method which provides exact measuring results within seconds. Yet it is not possible to develop a measuring device that is ideal under every single condition; even P2, P4 and LM5 have a certain limited range to provide optimum results. In order to minimize incorrect applications resulting from that fact we have provided you with a list.
Important! Here are the most common reasons for measuring errors.
Product temperature beyond ambient conditions: The temperature of the measuring device and of the product should be nearly equal.
Wrong switch position
Uneven pressing
Wrong application on rolls
Press your measuring device lengthwise to the longitudinal axis of the product roll, otherwise the bearing area of the sensor is too small.
Product pile not thick enough
Electrically conductive material
All metallic objects as well as electrically conductive packaging film or paper coloured with carbon black influence the measuring result in a negative way and must no be positioned within a range of 500 mm below the sensor area.
Version 2.2_en Schaller GmbH Page 7
Exceeding of measuring range: This can occur depending on the type of product and the position of the selector switch. It is indicated on the display by blinking and means that the measuring accuracy is less precise.
Measuring principle
The material in the measuring field is penetrated by an electromagnetic field whose features change due to the material's quantity of moisture. Because of the characteristic polarity of the water molecule and the thus resulting high relative permittivity of water (about 83) the capacity of the measuring field changes together with the changing moisture of the measuring sample.
Measuring plate at the
back of the instrument
8 x 5 cm 7 x 4,5 cm
Measuring instrument
P2, P4 Measuring plate 8cm long x 5cm broad x 1,3cm depth LM5 Measuring plate 7cm long x 4,5 cm broad x 0,5cm depth
Caution! When the material thickness is below a certain limit, the measuring result is falsified.
Page 8 Schaller GmbH Version 2.2_en
Exclusion of liability
The producer can not be made liable for any measuring errors and the damage that might result from it. This measuring principle is a quick measuring procedure which can be influenced by product and user-specific margin conditions. Therefore it is recommended to carry out a plausibility check of the measured values. In every measuring device there is a serial number and a seal of warranty. If this seal is broken no warranty claims will be accepted. In case of defect send back your measuring device to Messtechnik Schaller or your dealer. The device must be sufficiently packaged for protection reasons. (refer to technical support)
Care regulations
In order to guarantee that your measuring device is kept in good condition, please take care that it is not imposed to too much mechanical strain (it should not be dropped or exposed to excessive temperatures). Always use a dry cloth when cleaning your measuring device because it can be destroyed by incoming water or other cleaning agents. Do not leave your device unattended – put it back into the delivered wooden case after the measuring procedure. We recommend an ISO-approved examination of the device in regular intervals by applying the kiln method or by using the enclosed proof plane PP2 (base plate of the wooden case)! If needed, Messtechnik Schaller GmbH (Ltd.) will also carry out a calibration at your expense. In this case a calibration certificate will be issued and delivered.
Version 2.2_en Schaller GmbH Page 9
ISO-approved examination of the instrument
The surface of the proof plane PP2 must not be scratched and must be free of dust, dirt, fats and moisture. Field of application: temperatures ranging from 10°C to 30°C, relative humidity 30-80%
Procedure of examination Torn on the device, place it in the middle of the proof plane PP2 and press firmly onto it. Now check the values on the display against the set measuring values in the table and fill them in the provided table. We have defined the ideal temperature for your device – 20°C, but it can range between 17°C and 23°C. If an adjustmen t is not possible, a calculated correction of the temperature of the displayed value is necessary. The corrected value and the set value must be within the range of tolerance.
Set measuring values for PP2 should be at a temperature of 20°C.
Switch
set measuring
value adjust readjust
position
P2
P4
tolerance
tolerance
S1 6.4 4.3 0.6 0.9 S2 5.8 4 0.6 0.9 S3 5.1 3.7 0.5 0.8 S4 4.5 3.3 0.5 0.7 S5 3.8 3 0.4 0.6 S6 3.1 2.6 0.4 0.6
Value when held up in the air at switch position 1 is: P2= 1.3% und P4= 2.5% range of adjustment tolerance by the plant 0.6%
Page 10 Schaller GmbH Version 2.2_en
switch
set measuring
value adjust readjust
position
LM5 tolerance tolerance
S1 11,9 0,9 1,5 S2 11 0,8 1,3 S3 10,1 0,8 1,1 S4 9,2 0,7 1,0 S5 8,2 0,6 0,9 S6 7,2 0,5 0,8
Value when held up in the air at switch position 1 is: LM5= 3,7% range of adjustment tolerance by the plant 0,8%
Example of temperature correction This example has only been calculated for P2, but the temperature correction can be calculated for both P4 and LM5. Your measuring device is adjusted to a temperature of 28°C. The display of P2 with switch position S1 on proof plane PP2 is 6.1%. Tcf: Temperature correction factor corresponds to 0.04% (S1) moisture of paper per 1°C. Tcv: Temperature compensated value
Tcv = displayed value + (ambient temperature - 20°C) x x Tcf P2: Tcv= 6,1 + (28 – 20) x 0,04 results 6,4%
If your value is beyond the set range of adjustment tolerance a new adjustment by the factory is necessary. If there is a linear deviation from the set values, you can adjust the device to the zero point adjustment yourself. This is done as follows:
Version 2.2_en Schaller GmbH Page 11
Zero point adjustment
Shift the selector switch to S1, put off the lid of the battery compartment, take out the battery (which must still be stuck to the connecting contacts). Then put off the serial number tag carefully. Now hold the device into the air with one hand and adjust the inner potentiometer as long as the following values appear on the display: P2:1.3%, P4: 2.5%, LM5: 3.7%. Again the temperature correction value must be taken into consideration. Put on the serial number tag again, insert the battery and close the lid.
inner Potentiometer
Additional information ( for LM 5 only)
Switch positions 1 to 6 serve as an approach to the real quantity of moisture of the material. The density of the material has the biggest influence on the measuring result. For leather, the relevant range of density corresponds to 0.45 gram/cm³ for switch position 1 and
0.85 gram/cm³ for switch position 6. Various manufacturing procedures can result in a change of up to two switch positions. A higher moisture value is indicated on the display when measuring values lie below or above the set measuring range. These values can be used as reproducible comparative values and with the help of a conversion table you get values corresponding to absolute moisture values. When measuring leather it is especially important that the pressure against the material is the same pressure when determining the switch position.
Page 12 Schaller GmbH Version 2.2_en
P2 Examination table
Copy this table and fill in the displayed values – without temperature correction – and the ambient temperature. Enclose a copy of the data if the instrument should be adjusted again by the factory.
serial number measuring instrument:_____________
serial number proof plate:______________
temperature :______°C date :_____________
set
displayed
temperature- drift
switch- measuring
value compensation in readjust
position
value P2 value % tolerance
S1 6.4
0.6
S2 5.7
0.6
S3 5.1
0.5
S4 4.5
0.5
S5 3.8
0.4
S6 3.1
0.4
Version 2.2_en Schaller GmbH Page 13
P4 Examination table
Copy this table and fill in the displayed values – without temperature correction – and the ambient temperature. Enclose a copy of the data if the instrument should be adjusted again by the factory.
serial number measuring instrument:____________
serial number proof plate:_____________
temperature :______°C date :_____________
set displayed temperature- drift
switch- measuring
value compensation in readjust
position
value P4 value % tolerance
S1 4,3
0,6
S2 4
0,6
S3 3,7
0,5
S4 3,3
0,5
S5 3
0,4
S6 2,6
0,4
Page 14 Schaller GmbH Version 2.2_en
LM5 Examination table
Copy this table and fill in the displayed values – without temperature correction – and the ambient temperature. Enclose a copy of the data if the instrument should be adjusted again by the factory.
serial number measuring instrument:____________
serial number proof plate:_____________
temperature :______°C date :_____________
set displayed temperature- drift
switch- measuring
value compensation in readjust
position
value LM5 value % tolerance
S1 11,9
1,5
S2 11
1,3
S3 10,1
1,1
S4 9,2
1
S5 8,2
0,9
S6 7,2
0,8
Version 2.2_en Schaller GmbH Page 15
Technical data:
dissolution 0.1% moisture of material depth of measurement P2, P4:13mm
LM5: 5mm measuring range P2: 3-10% moisture of paper
P4: 3-8% moisture of paper
LM5: 8-20% moisture of leather indication of % refers to wet weight P2, P4: switch position 4, LM5: switch position 2
working temperature 5°C to 35°C temperature compensation P2:0,04%/°C ,P4:0,03%/°C
LM5: 0,05%/C° power supply 9V alkaline batteries or
accumulator
drawing of current 5mA, sufficient for about
4000 measuring procedures, when BAT appears on the display, still about 200 measurements are possible
display three-digit LC-Display size 60 x 120 x 26 mm weight without battery about 140 gramms
delivery wooden case, proof plate, 9V
alkaline compound battery, case protection made of rubber
Page 16 Schaller GmbH Version 2.2_en
Technical Support
Max
-
Schaller
-
Straße 99
A –
8181 St.Ruprecht an der Raab
Tel +43 (0)31
78 28899
Fax +43 (0)31
78 28899
-
902
E-Mail support@humimeter.com
Web http://support.humimeter.com
or
your local retailer
date / stamp