Humboldt HP-1160 User Manual

02.11
product manual
HP-1160
Falling Sand Abrasion Tester
Loosen thumb screw #1. Move collar and sand container to the top of the post.
Loosen thumb screw #2. The box will stay because it's sitting on the bottom collar.
Loosen thumb screw #3. Slide box up pip until catch container fits under box. Tighten thumb screw #3.
With the brass gauge #4 on the bottom of the sand tube, loosen thumb screw #1 and lower until the brass gauge is somewhere close to the sample plate, approximately 1 inch. Remove plastic window from the bottom box. With screw #2 loose, the box will pivot around the post to allow the test sample to be put into place.
Pivot box back in to place, then raise and lower and pivot box until the brass gauge is centered and just touches the sample. Tighten thumb screw #2
A pivot and stop have been designed into the bottom collar of the sample box to allow for test repeatability, as long as the same sample thickness is used.
To set the stop, loosen thumb screw #3. Raise collar until it touches collar #3. Rotate bottom collar (thumb screw toward front of tester), make sure bottom collar contacts the stop. Tighten thumb screw #3.
Now loosen screw #2, pivot the box to allow removal of the brass gauge from the sand tube. Then, when the box is pivoted back into place, it will stop with the sample 1 inch from the bottom of the sand tube as per ASTM specification. Tighten screw #1, replace window. Conduct test.
To repeat test, remove window, loosen screw #2, pivot box to remove sample and put in a new one, If samples are the same thickness. Pivot box into place and conduct the next test. If sample thickness is different, replace the brass gauge and re-do the setup procedure for the new thickness.
Warranty
Humboldt Mfg. Co. warrants its products to be free from defects in material or workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co., factory replacement of any part or parts of such product, for the warranty of this product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions of Sale. The purchaser is responsible for the transportation charges. Humboldt Mfg. Co. shall not be responsible under this warranty if the goods have been improperly maintained, installed, operated or the goods have been altered or modified so as to adversely affect the operation, use performance or durability or so as to change their intended use. The Humboldt Mfg. Co. liability under the warranty contained in this clause is limited to the repair or replacement of defective goods and making good, defective workmanship.
Humboldt Mfg. Co.
875 Tollgate Road Elgin, Illinois 60123 U.S.A.
Testing Equipment for
HUMBOLDT
www.humboldtmfg.com
U.S.A. Toll Free: 1.800.544.7220
Voice: 1.708.456.6300
Fax: 1.708.456.0137
Email: hmc@humboldtmfg.com
Construction Materials
Loading...