Humboldt HM-4187A User Manual

08.11
product manual
HM-4187A
De-Airing Tank
1
Introduction:
The Humboldt HM-4187A De-Airing Tank is designed to remove dissolved gasses from ordinary tap water. The De-Airing tank requires a source for vacuum, tap water, and a 110 VAC electrical supply. The apparatus consists of a closed cylinder with a motor driven impeller. The rotation of the impeller causes the gasses to separate from the water while under a constant vacuum. The gasses are evacuated from the tank at the top. The HM-4187A can be used in conjunction with any of the distribution panels, as well as the Flex panels.
Installation:
Using 1/4 inch O/D tubing, attach a vacuum line to the connector at the top of the tank by pushing the line into the connector until it is fully seated. Using the same size tubing, connect a line from the “Tap Water Out” connector at the back of the distribution panel to one of the connectors at the back of the tank base. Connect another line from the “De-Aired Water In” line at the back of the distribution panel to the other tubing connector at the back of the tank base.
The HM-4187A De-Airing Tank requires an electrical connection to operate the motorized impeller. The tank is supplied with an AC to DC transformer which plugs into a standard 110 VAC outlet. The transformer also has a plug which connects to the jack at the back of the tank base under the power switch.
Operation:
After all of the connections are made, fill the tank to approximately ¾ full with tap water. Turn on the impeller using the switch at the back of the tank base. Turn on the vacuum by rotating the knob on the front of the distribution panel or Flex panel. On the distribution panel or the Flex panel, there are two valves for use with the de-airing tank. One valve controls the filling and draining of tap water into and out of the tank and the other controls the vacuum or pressure to the tank. The draining and pressurizing operation is for the purpose of cleaning the tank and not for test purposes. During a test, the de-aerated water from the tank would be controlled by another valve on the distribution panel where the water connection to the base of the triaxial cell is located. Do not fill the tank with tap water while de-aerated water is being used as this would allow tap water to mix with the de-aerated water in the sample. Also, do not fill the tank while using the vacuum as accidental overfilling of the tank would allow water to be drawn into the vacuum lines and possibly into the vacuum pump.
2
VACUUM LINE
TAP WATER LINE
DE-AIRED WATER LINE
HM-4187A AIR, WATER, AND ELECTRIC CONNECTIONS
POWER SWITCH
AC TO DC TRANSFORMER JACK
FIGURE 1
TURN KNOB LEFT TO FILL DE-AIRING TANK WITH TAP WATER OR TURN KNOB RIGHT TO DRAIN DE-AIRING TANK
3
TURN KNOB LEFT TO APPLY VACUUM TO DE-AIRING TANK OR TURN KNOB RIGHT TO APPLY PRESSURE TO ASSIST DRAINING OF DE-AIRING TANK
DE-AIRED WATER
FILL DRAIN
OFF
VACUUM
FIGURE 2
DE-AIRING TANK CONTROL VALVES
PRESSURE
AUXILIARY AIR PRESSURE
ROTATE THE KNOB
CLOCKWISE TO
INCREASE PRESSURE
TO THE DE-AIRING
TANK & TO THE
AIR
AUXILIARY AIR
4
CONNECTOR
AUXILIARY VACUUM
FIGURE 3
DRAIN
OFF
FILL
DE-AIRED WATER
PRESSURE
VACUUM
POINT THE KNOB
TO THE RIGHT
TO APPLY PRESSURE
TO THE DE-AIRING TANK
REGULATING AUXILIARY AIR PRESSURE
Warranty
Humboldt Mfg. Co. warrants its products to be free from defects in material or workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co., factory replacement of any part or parts of such product, for the warranty of this product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions of Sale. The purchaser is responsible for the transportation charges. Humboldt Mfg. Co. shall not be responsible under this warranty if the goods have been improperly maintained, installed, operated or the goods have been altered or modified so as to adversely affect the operation, use performance or durability or so as to change their intended use. The Humboldt Mfg. Co. liability under the warranty contained in this clause is limited to the repair or replacement of defective goods and making good, defective workmanship.
Humboldt Mfg. Co.
875 Tollgate Road Elgin, Illinois 60123 U.S.A.
Testing Equipment for
HUMBOLDT
www.humboldtmfg.com
U.S.A. Toll Free: 1.800.544.7220
Voice: 1.708.456.6300
Fax: 1.708.456.0137
Email: hmc@humboldtmfg.com
Construction Materials