Humboldt HM-2850.3F Product Manual

Page 1
product manual
03.18
HM-2850.3F
Humboldt
Multi-Speed Load Frame
Page 2
Application
The HM-2850 features a quiet, direct drive stepper motor that provides a range of loading speeds from 0.00001 to 2.00000 in/ min (0.00001 to 50.80000mm/min.). This speed range is more than adequate for the majority of standard soil tests.
The HM-2850 also incorporates a separate, dedicated control to accommodate 2.00000 in/min (50.80000mm). for use in Marshall and TSR Testing, as well as a rapid travel speed of 2.25 in/min for moving the platen into position quickly. Speeds are controlled through the use of edit keys and the digital display.
Product Use
This product is intended for use only in accordance with the directions and specifications contained in this User Guide. The HM-2800 Multi-speed Load Frame has been designed for testing that complies with ASTM D1883, D2850, D2166, D4767, D5581 and D6927 and AASHTO T193, T296, T297, T208, T245, T246, as well as other applicable International standards.
Features include:
• 8" plated platen provides roomy, stable base for test equipment
• 7" color, touchscreen
• Test speeds adjustable from 0.00001 to 2.00000 in/min (0.00001 to 50.80000mm/min.) via touchscreen
• User selectable unit change from keyboard between U.S. Standard and Metric units.
1
Page 3
Machine Specifications
Covers: CBR, UU, CU, CD, UC, Marshall and Hveem Tests. ASTM: D1883, D2850, D2166, D4767, D5581 and D6927 AASHTO: T193, T296, T297, T208, T245, and T246; BS 1377: Part 4: 1990, BS 1377: Part 7: 1990, BS 1377: Part 8: 1990, BS 598: Part 107
General Warnings
Safety Warnings
Operators should take care to operate this machine under the maximum load restrictions. The machine is programmed at the factory to provide safety shutdown if the upper or lower maximum travel is exceeded as well as if the upper instrument calibration is exceeded.
Electrical Warnings
Typically, there is no reason for the operator to open the machine. However, if the customer’s engineers attempt to change settings to the circuit board connected to the back panel, the machine must always be unplugged before this operation. Unplugging the internal connection to the back panel circuit board while the machine is under power will result in permanent damage to the circuit board.
2
Specifications
Load capacity 11000 lbf (50kN)
Speed range
0 - 2.00000 in/min. 0 - 50.80000 mm/min.
Platen Size 8" (203mm)
Platen Travel 4" (100mm)
Clearance, vertical 40" (1000mm) Max
Clearance, horiz. 11" (279mm)
Dimensions (l x w x h).
17 x 22 x 51 inch (432 x 559 x 1295mm)
Voltage 110/220V 50/60Hz. 5.0 amps
Page 4
IMPORTANT NOTICE
The information contained herein is supplied without representation or warranty of any kind. Humboldt Mfg. Co. therefore assumes no responsibility and shall have no liability, consequential or otherwise, of any kind arising from the use of the described equipment contained in this manual.
Updated products
The manufacturer reserves the right to change or modify product design or construction without prior notice and without incurring any obligation to make such changes and modifications on products previously or subsequently sold.
Fitness for application
The manufacturer makes no recommendations or claims regarding fitness for applications other than the specific tests as defined in this User Guide.
Installation and Set-up
Unpacking
During your inspection, make certain that the contents of your shipment match the documentation provided by your packing list. Accessories that are part of your order may have been packed with or installed onto the HM-2850, so make sure they are accounted for and match the packing list.
Listed below are those items included with an HM-2850 Multi­Speed Load Frame:
Multi-Speed Load Frame— HM-2850.3F
Includes: SHH04—3/4-16 Bolt x 3-1/2 inch long WF04—Washer for SHH04 Bolt HM-000352—AC Power Cord
3
Page 5
Electrical Connections
Place unit on a flat smooth surface and use leveling feet (supplied) and a bubble level to ensure that the unit is level side-to-side and back-to­front.
Electrical Connections
The HM-2850 arrives ready for operation. Attach the supplied IEC electrical cord to the machine and plug into a standard wall receptacle for use in the United States.
Power Cord:
Your Multi-use Test Machine is fitted with a molded vinyl grounding plug. The cord is made of three conductors either
(American color coding). Green Ground (Earth) White Cold (Neutral) Black Hot (Live)
or
(International color coding). Green/Yellow Ground (Earth) Blue Cold (Neutral)
Brown Hot (Live)
Fuses:
The HM-2850 uses (2) 5x20mm, slo-blow fuses in the IEC power entry module at the right rear of the machine. The replacement part number for these fuses is H-1322.13.
4
Page 6
Operation
Control Panel
An On-Off rocker switch on the back right side of the machine is the power switch. The control panel is situated on the front left hand side of the machine base. A 7" touch-screen displays the speed and units in either in/min or mm/min. A touch-screen en­ables setting the speeds and units and control of machine travel.
Speed Selection
On the control panel, pressing Motor Speed causes the display to display the motor speed keypad where you can enter the desired speed.
Speed Units Selection
Press the unit to select in/min or mm/min.
Running a Test
After selecting the desired speed and positioning the sample, ini­tiate a test by pressing the Up Arrow. Upon conclusion of the test press the Stop Test key. See Page 7 for a chart of common tests and test speeds.
Rapid Travel
Pressing and holding down either the up or down arrows allows travel at slightly over 2.25 in/min in the direction chosen and releasing either key will stop the travel.
1
5
Page 7
Maximum Travel
A redundant limit switch prevents over travel in the up or down directions by stopping the machine automatically. The maximum travel of the platen is 4.00 inches (100mm).
Load Ring
When a load ring is used it can be attached directly to the cross-beam using the 3/4-16 x 3” long bolt.
Maximum Load
This Compression and Testing Machine is rated a 11,000 lbf (50kN).
Accuracy of Platen Speeds
The indicated speeds are based on zero load conditions and will be maintained within +/- 5%. The actual motor speed will be maintained to +/- 1%. These are not closing gap speeds.
The speeds used in the various test are, in fact, the closing gap speed, i.e., the distance between the cross-beam and the platen. These closing gap speeds will be dependent on the stiffness of the sample being tested and the compressibility of any load measuring device (such as the load ring). At high loads, some machine stretch may occur. The test speeds used should, therefore, take into account these factors.
Maintenance
Both the jack unit and the gearbox are packed with grease and do not normally require further attention. The machine can generally be maintained by keeping it clean, and if adjusting the crosshead beam, lightly oiling the exposed threads.
6
Page 8
Test
Description
Standards
Speed
in/min.
Tolerance Comments
CBR, 1mm BS 1377, Clause 7.4.2.2 .039
(1mm/min.)
± 20% ie.
.2 mm/min.
UU Triaxial
Unconsolidated
Undrained, 1.4"
ASTM D2850
Para. 5.1 & 7.5
AASHTO T208
1% min. plastic,
.3% min. brittle
—5% constant rate,
strain control
Cohesive soils;
no ripples in a
glass of water
UU Triaxial
Unconsolidated
Undrained, 1"
BS 1377, Part 7 .3% min.
Unconfined
Cohesive Soil
ASTM D2166
Para. 7.1
1/2 to 2% min. Strain Controlled Clay or
Cemented Soils
UC Consolidated
Undrained, 1.4"
ASTM D4767
Para. 5.2
.028 or formula
in Para. 8.4.2
±1% Cohesive
Plastic Soils
Marshall/TSR ASTM D6927 & 5581;
AASHTO T245
2" ±.15
or 7.5%
Mold Soil Cement ASTM D1633 Approx. .05 Continuous w/o shock
CBR/LBR
Hveem &
Soil Cement
ASTM D1883, D4429
(field) AASHTO T193
Approx. .05
no pulses
Continuous
run at idle
w/o shock
UU Unconsolidated
Undrained quick triaxial, 1.4"
ASTM D2850 .008 ±5% Brittle material
UU Unconsolidated
Undrained quick triaxial, 2.8"
ASTM D2850 .017 ±5% Brittle material
UC Unconsolidated
Undrained triaxial for
cohesive soil
ASTM D4767
Para. 5.2
.056 or formula
in Para. 8.4.2
±1% Cohesive soils
w/plasticity
Common Speeds for Specimen Testing
7
Page 9
Typical Test Setups for the HM-2850
The charts below list the parts necessary to perform these standard tests.
CU/UU Triaxial Setup with HM-2850
Components Item
Load
50kN (11240 lbf) capacity HM-2850.3F
Strain
Load ring 2,200 lbf (10 kN)
H-4454.020
Dial gauge 2.0" travel,
0.001" divisions)
H-4463
Pore pressure transducer HM-4170
Ball seat adapter HM-200387
Single channel readout (choose)
HM-2350
HM-2350.4F
NEXT software basic, user defined
included
Pressure
Pressure distribution panel for CU triaxial (choose)
HM-4150.3F
HM-4150M.3F
Pressure distribution panel for UU triaxial (choose)
HM-4140.3F
HM-4140M.3F
De-airing system HM-4187A.3F
Vacuum pump (choose)
H-1763A
H-1763A.4F
Triaxial Cell (choose 1 below)
3"/ 75mm dia. capacity HM-4199B
4"/ 100mm dia. capacity HM-4199B-4
Top Cap/ Base Pedestal Set (specify specimen size)
HM-4199.XX
Page 10
9
Part Numbers ending in .XX require a size code to be entered referring to
the sample size to be tested. For Triaxial samples, sizes are: .14 = 1.4"; .15 = 1.5"; .20 = 2.0"; .25 = 2.5"; .28 = 28"; .30 = 3.0".40 = 4.0"; .35 = 35mm; .38 = 38mm; .50 = 50mm; .70 = 70mm, and .100 = 100mm.
Typical Test Setups for the HM-2850 (cont.)
Typical Soil Cement Setup
Components
Load
50kN (11240 lbf) capacity HM-2850.3F
Upper swivel platen HM-2003E
Strain
Load ring 5,000 lbf (25 kN)
H-4454.050
Unconfined Compression Setup with HM-2850
Components
Load
50kN (11240 lbf) capacity HM-2850.3F
Upper unconfined platen HM-2002
Displacement indicator platform
HM-3000.10.2
Displacement indicator rod HM-3000.10.1
Load ring 500 lbf (2.5 kN)
H-4454.005
Dial gauge 2.0" travel
0.001" divisions
H-4463
Typical Triaxial Sample Prep Items
Components
Part # Qty. Description
HM-4180.XX
1
Membranes, 12/pk
HM-4181.XX
1
Membrane Stretcher
HM-4182.XX
1
O-Rings, 12/pk
HM-4184.XX
2
Porous Stone
HM-4187E
1
De-Airing Tank, 6 Liter, 110V
HM-3847.XX
1
Split Miter Box
HM-4189.XX
1
Filter Paper, 100/pk
HM-4189FS
1
Filter Strips 5 x 150mm 100/pk
Page 11
10
Page 12
www.humboldtmfg.com
HUMBOLDT
Testing Equipment for
Construction Materials
Humboldt Mfg. Co.
3801 North 25th Avenue Schiller Park, Illinois 60176 U.S.A.
U.S.A. Toll Free: 1.800.544.7220
Voice: 1.708.456.6300
Fax: 1.708.456.0137
Email: hmc@humboldtmfg.com
Warranty
Humboldt Mfg. Co. warrants its products to be free from defects in material or workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co., factory replacement of any part or parts of such product, for the warranty of this product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions of Sale. The purchaser is responsible for the transportation charges. Humboldt Mfg. Co. shall not be responsible under this warranty if the goods have been improperly maintained, installed, operated or the goods have been altered or modified so as to adversely affect the operation, use performance or durability or so as to change their intended use. The Humboldt Mfg. Co. liability under the warranty contained in this clause is limited to the repair or replacement of defective goods and making good, defective workmanship.
Loading...