Hukseflux SR05, SR05-D1A3-PV User Manual

Copyright by Hukseflux | manual v1801 | www.hukseflux.com | info@hukseflux.com
USER MANUAL
SR05-D1A3-PV
Digital second class pyranometer – alternative for PV reference cell
Hukseflux
Thermal Sensors
SR05-D1A3-PV manual v1801 2/83
Warning statements
Putting more than 30 Volt across the sensor wiring of the main power supply can lead t o permanent damage to the sensor.
Keep the voltage on the RS-485 data wiring of SR05-D1A3-PV between -7 and +12 V to avoid permanent damage.
For proper instrument grounding: use SR05 with its original factory-made SR05 cable.
Using the same Modbus address for more than one device will lead to irregular behaviour of the entire network.
Disconnect power whi le performing service or maintenance.
Locally connect the cable shield to ground if SR05-D1A3-PV is not connected to ground throu gh the installation platform.
Modbus
®
is a registered trademark of Schneider Electric, licens e d to the Modb us Organization, Inc.
SR05-D1A3-PV manual v1801 3/83
Contents
Warning statements 2 Contents 3 List of symbols 5 Introduction 6 1 Ordering and checking at delivery 11
1.1 Ordering SR05-D1A3-PV 11
1.2 Included items 13
1.3 Quick instrument check 14
2 Instrument principle and theory 15 3 Specifications of SR05-D1A3-PV 18
3.1 Specifications of SR05-D1A3-PV 18
3.2 Dimensions of SR05 22
4 Standards and recommended practices for use 23
4.1 Classification standard 23
4.2 General use for solar radiation measurement 23
4.3 General use for sunshine duration measurement 23
4.4 Specific use for outdoor PV system perfo rmance testing 24
4.5 Specific use in meteorology and climatology 25
5 Installation of SR05 26
5.1 Site selection and installation 26
5.2 Mounting and levelling SR05 27
5.3 Installing SR05 27
5.4 Installing SR05 with its ball levelling and tube mount 28
5.5 Placing and removing SR05’s ball levelling shim 30
5.6 Electrical connection of SR05-D1A3-PV: wiring diagram 32
5.7 Grounding and use of t he shield 32
5.8 Using SR05-D1A3-PV’s digital output 33
5.9 Using SR05-D1A3-PV’s analogue 0 to 1 V output 36
6 Communication with SR05 37
6.1 PC communication: Sensor Manager software 37
6.2 Network communication: function codes, registers, coils 43
6.3 Silicon Reference Cell compatible Modbus output 45
6.4 Standard Hukseflux Modbus Output 48
6.5 Network communication: getting started 53
6.6 Network communication: example master reques t to SR05 54
7 Making a dependable measurement 56
7.1 The concept of dependability 56
7.2 Reliability of the measurement 57
7.3 Speed of repair and maintenance 58
7.4 Uncertainty evaluation 58
8 Maintenance and trouble shooting 61
8.1 Recommended maintenance and quality assurance 61
8.2 Trouble shooting 62
8.3 Calibration and che cks in the field 63
8.4 Data quality assurance 64
9 Appendices 66
9.1 Appendix on cable extension / replacement 66
9.2 Appendix on tools for SR05 68
9.3 Appendix on spare parts for SR05 68
SR05-D1A3-PV manual v1801 4/83
9.4 Appendix on standards for classification and calibration 69
9.5 Appendix on calibration hierarchy 70
9.6 Appendix on meteorological radiation quantities 71
9.7 Appendix on ISO and WMO classification tables 72
9.8 Appendix on definition of pyranometer specifications 73
9.9 Appendix on terminology / glossary 74
9.10 Appendix on floating point format conversion 75
9.11 Appendix on function codes, register and co il overview 76
9.12 Appendix on the sensor model name in the register 80
9.13 EU declaration of conformity 81
SR05-D1A3-PV manual v1801 5/83
List of sy m bols
Quantities Symbol Unit Voltage output U V
Sensitivity S V/(W/m
2
)
Solar irradiance E W/m
2
Output of 0-1 V U V Transmitted range of 0-1 V r W/m
2
(see also appendix 9.6 on meteorological qua ntities)
Subscripts
Not applicable
Notation Example
Decimal numbers are indicated without prefix 30 Hexadecimal numbers are indicated with a 0x prefix 0x1E
SR05-D1A3-PV manual v1801 6/83
Introduction
SR05 series is the most affordable range of pyranometers meeting ISO 9060 requirements. These sensors are ideal for general solar radiation measurements and popular for monitoring photovoltaic (PV) systems. Model SR05-D1A3–PV is made as a perfect alternative to PV reference cells. It offers the same Modbus interface as the most common PV reference cell model for easy compatibility. Relative to PV reference cells, SR05-D1A3-PV has the advantage of higher stability, independence of the PV cell type or anti-reflection coating, and better availability and price of recalibration.
SR05 series is an economical range of ISO 9060 second class pyranometers for measurement of solar radiation received by a plane surface, in W/m
2
, from a 180 ° field of view angle. SR05 is perfect for predicting generated power and monitoring the overall efficiency of PV power plants. Different mounting options are available, allowing SR05 to be mounted in virtually any situation. The combination of easy installation and its low cost makes SR05 the preferred solution for commercial scale PV systems.
There are several versions of SR05 series:
Version SR05-D1A3: digit al s e nsor wit h Mo db us ove r RS-485 and analogue 0-1 V output
Version SR05-D2A2: digit al s e nsor wit h Mo db us ove r TTL and ana log ue 4–20 mA output
Version SR05-A1: analogue senso r w it h an a log u e m ill iv olt output
and:
Version SR05-D1A3-PV: d igit al se nso r w ith M odb us o ver RS-485, alternative for PV
reference cell
This user manual covers use of SR05-D1A3-PV. Specifications of this version differ from those of the other digital and analogue sensors in the SR05 series range. For use of SR05-D1A3 or SR05-D2A2, consult the separate SR05-D1A3 & SR05-D2A2 user manual. For use of SR05-A1, offering analogue millivolt output, consult the separate SR05-A1 user manual.
Figure 0.1 SR05-D1A3-PV digital second class pyranometer seen from above
SR05-D1A3-PV manual v1801 7/83
Model SR05-D1A3-PV has a digital output that is identical to the most commonly used photovoltaic reference cell with Modbus over RS-485 output. This allows for easy installation in existing PV monitoring systems, without the need to make major modifications to data logging software, instrument libraries and infrastructure.
Compared to silicon reference cells, pyranometers offer several advantages such as a perfect (cosine) directional response and a flat spectral response over a wide range. Pyranometers therefore meas ure the maximum available resource and are suitable to act as a reference for all types (for example amorphous, crystalline or thin-film) of photovoltaic cells both with and without anti-reflection coating. See also our Application note pyranometers versus PV reference cells. Moreover, since the working principle of a pyranometer is different from a solar cell, the pyranometer offers a truly independent measurement of the irradiance.
PV system performance monitoring: compliant with IEC 61724-1, Class C
IEC 61724-1: Photovoltaic System Performance Monitoring - Guidelines for Measurement, Data Exchange and Analysis – suggests to use pyranometers for PV monitoring; SR05 complies with IEC 61724-1 class C system requirements.
Features and benefits of SR05-D1A3-PV
higher stability than PV reference cells
indepen den t of P V ce ll typ e
affordable calibration
register structure and content identical to most common reference cells for easy
exchangeability
easy imple me nta tio n an d ser vic ing
easy mou n tin g a n d le v e llin g
pricing: affordable second class pyranometers
Figure 0.2 Two SR05-D1A3-PV digital second class pyranometers, of which one measuring in Plane of Array, replacing PV reference cells
SR05-D1A3-PV manual v1801 8/83
SR05-D1A3-PV design
SR05 pyranometers employ a thermopile sensor with black coated surface, one dome and an anodised aluminium body with visible bubble level. Optionally the sensor can be delivered with a unique ball levelling mechanism and tube mount or dedicated mounting fixture, for easy installation. SR05-D1A3-PV has an industry standard digital output: Modbus RTU over half-duplex RS-485, that allows multiple sensors to be installed on a single network. In addition, SR05-D1A3-PV has analogue 0 to 1 V output.
Figure 0.3 On the left SR05-D1A3-PV pyranometer with bubble level and M12-A cable connector in its standard configuration (3 metre ca ble standard included); on the right SR05 with optional ball levelling, for easy mounting and levelling on (non-)horizontal surfaces (included mounting bolts not displayed)
Figure 0.4 SR05-D1A3-PV digital second class pyranometer with optional ball levelling and tube mount for easy mounting and levelling on a tube (tube not included)
SR05-D1A3-PV manual v1801 9/83
Communication with a PC: Hukseflux Sensor Manager Software
For communication between a PC and SR05-D1A3-PV(‘s), the Hukseflux Sensor Manager software can be used. It is available for download on our website. The software allows the user to quickly configure SR05-D1A3-PV Modbus address and serial communication settings (baud rate, parity and stopbits) and to plot and export data. Also, the digital outputs may be viewed for sensor diagnostics.
Figure 0.5 U
ser
interface of the Hukseflux Sensor Manager
Compatibility
SR05-D1A3-PV’s Modbus interface is exchangeable with IMT-Solar Si-RS485TC-T-MB PV reference cell’s interface. Other reference cells upon request.
Suggested use for SR05-D1A3-PV
replacement of PV reference cells
measuring global tilted irradiance (GTI) in the Plane of Array (PoA) of solar panels
measuring global horizontal irradiance (GHI)
SR05-D1A3-PV is suited for use in SCADA (Supervisory Control And Data Acquisition) systems, supporting Modbus RTU (Remote Terminal Unit) protocol over RS-485. In these networks the sensor operates as a slave. Using SR05-D1A3-PV in a network is easy. Once it has the correct Modbus address and communication settings and is connected to a power supply, the instrument can be used in RS-485 networks. A typical network wil l request the irradiance (either register 0x0000 or registers 0x1002 + 0x1003) and temperature data (either register 0x0007 or register 0x1006) every 1 second, and eventually store the averages every 60 seconds. How to issue a request, process the register content and convert it to useful data is described in the paragraphs about network communication. The user should have sound knowledge of the Mo db us communication protocol when installing sensors in a network. When using the analogue 0 to 1 V output provided by SR05-D1A3-PV, the instrument can be connected directly to commonly used datalogging systems capable of handling a 0 to 1 V signal.
SR05-D1A3-PV manual v1801 10/83
The recommended calibration interval of pyranometers is 2 years. The registers containing the applied sensitivity and the ca libra tion history of the digital versions of SR05 are accessible for users with a password. This allows the user to choose his own local calibration service. The same register access may also be used for remotely controlled re-calibration of pyranometers in the field. Ask Hukseflux for information on this feature and on ISO and ASTM standardised procedures for field calibration.
The ASTM E2848 “Standard Test Method for Reporting Photo voltaic Non-Concentrator System Performance” (issued end 2011) confirms that a pyr anometer is the preferred instrument for PV system performance monitoring. SR05 pyranometer complies with the requirements of this standard. For more information, see our pyranometer selection
guide.
WMO has approved the “pyranometric method” to calculate sunshine duration from pyranometer measurements in WMO-No. 8, Guide to Meteorological Instruments and Methods of Observation. This implies that SR05 may be use d, in combination with appropriate software, to estimate sunshine duration. This is much more cost-effective than using a dedicated sunshine duration sensor. Ask for our application note.
All SR05 versions should be used in accordance with the recommended practices of ISO, WMO and ASTM.
See also
view our comp lete range of sensors
PMF01 pyranometer mounting fixture, compatible with SR05 ball levelling
SR05-D1A3-PV manual v1801 11/83
1 Ordering and checking at delivery
1.1 Ordering SR05-D1A3-PV
SR05-D1A3-PV second class pyranometer has a digital output that is identical to the most commonly used photovoltaic reference cell with Modbus over RS-485 output.
Besides SR05-D1A3-PV, SR05 series offers several other versions with industry standard outputs, both digital and analogue. Each version offers multiple mounting options and various cable lengths:
SR05-D1A3 digital second class sensor, with Modbus over RS-485 and 0-1 V output
1
SR05-D2A2 digital second class sensor, with Modbus over TTL and 4-20 mA output
SR05-A1 analogue second class pyranometer with millivolt output
1
This is our standard Modbus model which is not directly exchangeable with the commonly used PV reference cells.
For an overview of all versions and options, and how to order, please take a look at Table
1.1.1 on the next page.
The standard configuration of SR05-D1A3-PV is with 3 metres cable length. Common options for SR05-D1A3-PV are:
longer cable (10, 20 metres). Specify total cable length
extension cable with connector pair (10, 20 metres). Specify total cable length
with ball levelling (-BL)
with ball levelling and tube mount (for tube diameters 25 to 40 mm, -TMBL)
Ball levelling and tube m ount are suited for retrofitting.
SR05-D1A3-PV’s Modbus interface is exchangeable with IMT-Solar Si-RS485TC-T-MB PV reference cell’s interface. Other reference cells upon request.
SR05-D1A3-PV manual v1801 12/83
Table 1.1.1 Ordering codes for the versions of model SR05
VERSIONS OF SR05 (part numbers), without cable
SR05-D1A3-PV
digital second class pyranometer, with Modbus ove r
RS-485 output, alternative for PV reference cell
SR05-D1A3-PV-BL
digital second class pyranometer, with Modbus ove r RS-485 output, alternative for PV reference cell, with ball
levelling
SR05-D1A3-PV-TMBL
digital second class pyranometer, with Modbus ove r RS-485 output, alternative for PV reference cell, with tube
mount on ball levelling
SR05-D1A3
digital second class pyranometer, with Modbus ove r
RS-485 and 0-1 V output
SR05-D1A3-BL
digital second class pyranometer, with Modbus ove r RS-485 and 0-1 V output, with ball levelling
SR05-D1A3-TMBL
digital second class pyranometer, with Modbus ove r
RS-485 and 0-1 V output, with tube mount on ball
levelling
SR05-D2A2
digital second class pyranometer, with Modbus ove r TTL and 4-20 mA output
SR05-D2A2-BL
digital second class pyranometer, with Modbus over TTL
and 4-20 mA output, with ball levelling
SR05-D2A2-TMBL
digital second class pyranometer, with Modbus ove r TTL
and 4-20 mA output, with tube mount on ball levelling
SR05-A1
analogue second class pyranometer, with millivolt output
SR05-A1-BL
analogue second class pyranometer, with millivolt output,
with bal l le v elling
SR05-A1-TMBL
analogue second class pyranometer, with millivolt output, with tube mount on ball levelling
CABLE FOR SR05, with female M12-A connector at sensor end, non-stripped on other end
‘-03’ after SR05 part number
standard cable length: 3 m
‘-10’ after SR05 part number
cable length: 10 m
‘-20’ after SR05 part number
cable length: 20 m
CABLE EXTENSION FOR SR05, with male and female M12-A connectors
C06E-10
cable length: 10 m
C06E-20
cable length: 20 m
An extension cable (wit h connector pair) can be used in combination w ith a regular cable (with one connector at sensor end) to make alternative SR0 5 cable lengths possible.
Example: Cable length needed: 15 m. In this case, it is eas iest to buy SR05 with a 20 m cable and to cut it to desired length.
Example: Cable length needed: 30 m. In this case, it is eas iest to buy SR05 with 10 m cable and a cable extension of 20 m.
SR05-D1A3-PV manual v1801 13/83
1.2 Included items
Arriving at the customer, the delivery should include:
pyranometer SR05-D1A3-PV
cable of the length as ordered
product certificate matching the instrument serial number
For SR05-D1A3-PV-BL, also
ball levelling
4 mm hex key
1 x shim
2 x M5x20 bolts
2 x M5 nuts
For SR05-D1A3-PV-TMBL, also
ball levelling
4 mm hex key
1 x shim
2 x M5x20 bolts
2 x M5 nuts
tube mount
2 x M5x30 bolts
2 x M5x40 bolts
Please store the certificate in a safe place. The Hukseflux Sensor Manager can be downloaded via www.hukseflux.com/downloads
SR05-D1A3-PV SR05-D1A3-PV-BL SR05-D1A3-PV-TMBL
Figure 1.2.1 From left to right: SR05-D1A3-PV, SR05-D1A3-PV-BL and SR05-D1A3-PV-
TMBL (nuts and bolts, tools and certificates are not shown, tube itself is not included)
SR05-D1A3-PV manual v1801 14/83
1.3 Quick instrument check
A quick test of the instrument can be done by connecting it to a PC and installing the Sensor Manager software. See the chapters on installation and PC communication for directions.
1. At power–up the signal may have a temporary output level different from zero; an offset. Let this offset settle down.
2. Check if the sensor reacts to light: expose the se nso r to a strong light source, for instance a 100 W light bulb at 0.1 m distance. The signal should read > 100 W/m
2
now. Darken the sensor either by putting something over it or switching off the light. The instrument irradiance output should go down and within one minute approach 0 W/m
2
.
3. Inspect the bubble level.
4. Inspect the instrument for any damage.
5. Check the instrument serial number as indicated by the software against the label on the instrument and against the certificates provided with the instrument.
SR05-D1A3-PV manual v1801 15/83
2 Instrument principle and theory
Figure 2.1 Overview of SR05: shaded areas in exploded view show ball levelling mount and shim
(1) cable (standard length 3 metres, optional longer cable) (2) connector (3) bubble level (4) thermal sensor with black coating (5) glass dome (6) sensor body (7) tube mount (optional) (8) mounting screw (included with ball levelling and tube mount; requires 4 mm hex key) (9) shim (included with and needed for ball levelling mount) (10) ball levelling mount (optional) (11) countersunk set screw for levelling adjustment (included with ball levelling mount;
requires 4 mm hex key)
(12) opening for Ø 25 to Ø 40 mm tube when using ball levelling and tube mount
1
2
3
4
5
6
7
8
10
11
9
12
SR05-D1A3-PV manual v1801 16/83
SR05’s scientific name is pyranometer. A pyranometer measures the solar radiation received by a plane surface from a 180 ° field of view angle. This quantity, expressed in W/m
2
, is called “hemispherical” solar radiation. The solar radiation spectrum extends
roughly from 285 to 3000 x 10
-9
m. By definition a pyranometer should cover that
spectral range with a spectral selectivity that is as “flat” as possible.
In an irradiance measurement by definition the response to “beam” radiation varies with the cosine of the angle of incidence; i.e. it should have full response when the solar radiation hits the sensor perpendicularly (normal to the surface, sun at zenith, 0 ° angle of incidence), zero response when the sun is at the horizon (90 ° angle of incidence, 90 ° zenith angle), and 50 % of full response at 60 ° angle of incidence. A pyranometer should have a so-called “directional response” (older documents mention “cosine response”) that is as close as possible to the ideal cosine characteristic.
In order to attain the proper directional and spectral characteristics, a pyranometer’s main components are:
a thermal sensor with black coating. It has a flat sp ectrum covering the 200 to 50000
x 10
-9
m range, and has a near-perfect directional response. The coating absorbs all solar radiation and, at the moment of absorption, converts it to heat. The heat flows through the sensor to the sensor body. The thermopile sensor generates a voltage output signal that is proportional to the solar irradiance.
a glass dome. This dome limits the spectral range from 285 to 3000 x 10
-9
m (cutting
off the part above 3000 x 10
-9
m), while preserving the 180 ° field of v iew angle. Another function of the dome is that it shields the thermopile sensor from the environment (conve ction, rain).
The digital versions of model SR05 have a high-end 24-bit A/D converter, which is used by SR05 to convert the analogue thermopile voltage to a digital signal. SR05­D1A3-PV has a digital output that is identical to the most commonly used photovoltaic reference cell with Modbus over RS-485 output for easy exchangeability.
Pyranometers can be manufactured to different specifications and with different levels of verification and characterisation during production. T he ISO 9060 - 1990 standard, “Solar energy - specification and classification of instruments for measuring hemispherical solar and direct solar radiation”, distinguishes b etween 3 classes; secondary standard (highest accuracy), first class (second highest accuracy) and second class (third highest accuracy).
From second class to first class and from first class to secondary standard, the achievable accuracy improves by a factor 2.
SR05-D1A3-PV manual v1801 17/83
Figure 2.2 Spectral response of the pyranometer compared to the solar spectrum. The pyranometer only cuts off a negligible part of the total solar spectrum.
0
0,2
0,4
0,6
0,8
1
1,2
100 1000 10000
relative spectral conten t /
response [arbitrary units]
wavelength [x 10
-9
m]
solar radiation
pyranometer
response
SR05-D1A3-PV manual v1801 18/83
3 Specifications of SR05-D1A3-PV
3.1 Specifications of SR05-D1A3-PV
SR05 pyranometers measure the solar radiation received by a plane surface from a 180 ° field o f view angle . This quantity, expressed in W/m
2
, is called “hemispherical” solar
radiation.
SR05-D1A3-PV offers irradiance in W/m
2
as a digital output and as a 0-1 V output. It must be used in combination with suitable power supply and a data acquisition system which uses the Modbus communication protocol over RS-485 or one that is capable of handling a 0-1 V signal.
This user manual covers use of SR05-D1A3-PV. Specifications of this version differ from those of the other digital and analogue sensors in the SR05 series range. For use of SR05-D1A3 or SR05-D2A2, consult the separate SR05-D1A3 & SR05-D2A2 user manual. For use of SR05-A1, offering analogue millivolt output, consult the separate SR05-A1 user manual.
The instrument is classified according to ISO 9060 and should be used in accordance with the recommended practices of ISO, IEC, WMO and ASTM.
Table 3.1.1 Specificat ions of SR05-D1A3-PV (continued on next pages)
SR05-D1A3-PV MEASUREMENT SPECIFICATIONS: LIST OF CLASSIFICATION CRITERIA OF ISO 9060*
ISO classification (ISO 9060: 1990)
second class pyranometer
WMO performance level (WM O -No. 8,
seventh edition 2008)
moderate quality pyr a nometer
Response time (95 %)
18 s
Zero offset a (response to 200 W/m2
net thermal radiat ion)
< 15 W/m2 unventilated
Zero offset b (response to 5 K/h
change in ambient temperature)
< ± 4 W/m2
Non-stability
< ± 1 % change per year
Non-linearity
< ± 1 % (100 to 1000 W/m2)
Directional resp onse
< ± 25 W/m2
Spectral selectivity
< ± 5 % (0.35 to 1.5 x 10
-6
m)
Temperature response
< ± 3 % (-10 to +40 °C)
Tilt response
< ± 2 % (0 to 90 ° at 1000 W/m2)
*For the exact definition of pyranometer ISO 9060 specifications see the appendix.
SR05-D1A3-PV manual v1801 19/83
Table 3.1.1 Specifications of SR05-D1A3-PV (continued)
SR05-D1A3-PV
ADDITIONAL SPECIFICATIONS
Measurand global irradiance
(hemispherical solar radiation)
Measurand in SI r a diometry units
irradiance in W/m2
Optional measurand
sunshine duration
Field of view angle
180 °
Measurement range
0 to 2000 W/m2
Spectral range
(20 % transmission p oints)
285 to 3000 x 10-9 m
Rated operating temperatu r e r a nge
-40 to +80 °C
Output definition running average over 4 last meas urements,
measurement every 0.1 s
Recommended data reques t in terval
1 s, storing 60 s averages
Measurement function / optional
programming for sunshine duration
programming according to W M O guide paragraph
8.2.2
Measurand
instrument body temperat ure
Temperature sensor
Solid state System on Chip (SoC) silicon bandgap temperature sensor
measurand in SI units
temperature in °C
Standard governing use of the
instrument
ISO/TR 9901:1990 Solar en er gy -- Field pyranometers
-- Recommended practice for u s e ASTM G183 - 05 Standa r d P r a c tice for Field Use of
Pyranometers, Pyrheliometers and UV Radiometer s
Standard cable length (see options)
3 m
Cable diameter
4.8 x 10-3 m
Cable conductor cross-section
0.25 x 10-6 m2 (24 AWG)
Chassis connector
M12-A straight male connector, male thread, 5-pole
Cable connector
M12-A straight female connector, female thread, 5­pole
Connector protec tion class
IP67
Cable replacement
replacement and extension c a bles with connector(s) can be ordered separately from Hukseflux
Mounting (see opti ons)
2 x M5 bolt at 46 mm centre-to-centre distance on north-south axis, r equires 4 mm hex key
Levelling (see op tions)
bubble level is included
Levelling accuracy
< 0.6 ° bubble entirely in ring
Desiccant
silica gel, 1.0 g, in a HDPE bag, (25 x 45) mm
IP protection cla s s
IP67
Gross weight including 3 m cable
0.45 kg excluding optional a c c es s or ies
Net weight including 3 m cable
0.35 kg excluding optional a c c es s or ies
Packaging
box of (170 x 100 x 80) mm
CALIBRATION
Calibration trace a b ility
to WRR
Calibration hierarchy from WRR through ISO 9846 and ISO 9847, applying
a correction to ref er e nce conditions
Calibration method
indoor calibration according to ISO 9847, Type IIc
Calibration un certainty
< 1.8 % (k = 2)
Recommended recalibrati on interval
2 years
Reference conditions
20 °C, normal incide nce solar radiation, horizonta l mounting, irradiance level 1000 W/m2
Validity of calibra tion
based on experience the ins trument sensitivity will not change during storage. Durin g use under exposure to solar radiation th e instrument “non-stability” specification is applicable.
.
SR05-D1A3-PV manual v1801 20/83
Table 3.1.1 Specifications of SR05-D1A3-PV (started on previous pages)
MEASUREMENT ACCURACY AND RESOLUTION
Uncertainty of the measurement statements about the overall measurement
uncertainty ca n only be made on an individual basis.
see the chapter on uncertainty evaluation
WMO estimate on achievable accuracy for daily sums (see appendix for a
definition of the measurement conditions)
10 %
WMO estimate on achievable accuracy
for hourly sums (see appendix for a definition of the measurement conditions)
20 %
Irradiance resolution 0.1 W/m2 (register address 0x0000)
0.01 W/m2 (register address 0x1002 + 0x1003)
Instrument body temperature resolution 0.1 °C (register address 0x0007)
0.01 °C (register address 0x 1006)
Instrument body temperat ure accuracy
± 0.5 °C
SR05-D1A3-PV
Rated operating voltage ra nge
5 to 30 VDC
Recommended operating voltage
12 VDC
Power consumption
< 48 x 10-3 W at 12 VDC
SR05-D1A3-PV: DIGITAL
Digital output irradiance in W/m2
instrument body temperat ure in °C
Communication pr otocol
Modbus over 2-wire RS-485
Transmission mode
RTU
RS-485 transceiver common mode range
-7 to +12 V
RS-485 transceiver type
2-wire, non-isolated
System requirements for us e with PC Windows 7 and later, USB or RS-232 (COM) port and
connector, RS-485 / USB converter or RS-485 / RS-
232 converter
Software requirements for use with PC
Java Runtime Environment 8 – latest version available
free of charge at http://www.java.com, Hukseflux Sensor Manager - software version v1817 or higher
User interface on PC Hukseflux Sensor Man a ger v1817 or higher software
downloadable: to download and for available software updates, see
http://www.hukseflux.com/page/downloads
SR05-D1A3-PV: ANALOGUE 0 TO 1 V
0 to 1 V output
irradiance in W/m2
Transmitted range
0 to 1600 W/m2
Output signal
0 to 1 V
Standard setting (see options) 0 V at 0 W/m2 and
1 V at 1600 W/m
2
OPTIONS
Longer cable: 10,20 m
Cable with M12-A female c onnector on
sensor end, non-stripped on other end
option code = total cable length
Extension cable with connector pair: 10, 20 m. Cable with male and fem a le M12-A connectors
option code = C06E-10 for 10 metres, C06E-20 for 20 metres
SR05-D1A3-PV manual v1801 21/83
Table 3.1.1 Specifications of SR05-D1A3-PV (started on previous pages)
Ball levelling mountab le on (non-)horizontal surfaces
with angle compens ation up to 10 °; retrofittable; one shim, two M5x20 mounting bolts and two M5 nuts included; requires 4 mm hex k ey for levelling and 4 mm hex key and 8 mm wrench for mounting
option code = BL
Tube mount with b a ll le velling
mountable on tubes Ø 25 to Ø 40 mm
with angle compen s ation up to 10 °; retrofittable; one shim, two M5x 3 0 a nd two M5x40 mounting bolts included; requires 4 m hex key for levelling and mounting
option code = TMB L
Alternative pyranometer mounting
fixture
for mounting an y H ukseflux pyranometer on
horizontal and v er tical tubes, on platforms, both horizontal and in Plane of Array order code = PMF01
PMF01 is compatible with SR05 ball levelling
Adapted transmitted range 0 to 1 V
can be adjusted at the factory upon request
SR05-D1A3-PV manual v1801 22/83
3.2 Dimensions of SR05
Figure 3.2.1 Dimensions of SR05 in x 10
-3
m. The bottom drawing shows the height of SR05 combined with its optional ball levelling mount and the tube diameter required for use with SR05’s optional tube mount. M5 mounting bolts and the countersu nk set screw require a 4 mm hex key for mounting and levelling.
SR05-D1A3-PV manual v1801 23/83
4 Standards and recommended practices
for use
Pyranometers are classified according to the ISO 9060 standard and the WMO-No. 8 Guide. In any application the instrument should be used in accordance with the recommended practices of ISO, IEC, WMO and / or ASTM.
4.1 Classification standard
Table 4.1.1 Standards for pyranometer classification. See the appendix for definitions of
pyranometer specifications, and a table listing the specification limits.
STANDARDS FOR INSTRUMENT CLASSIFICATION
ISO STANDARD EQUIVALENT
ASTM STANDARD
WMO
ISO 9060:1990 Solar energy -- specification and classification of instruments for measuring hemispherical solar and direct solar radiation
Not available
WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, chapter 7, measurement of radiation, 7.3 measurement of global and diff use solar radiation
4.2 General use for solar radiation measurement
Table 4.2.1 Standards with recommendations for instrument use in solar radiation
measurement
STANDARDS FOR INSTRUMENT USE FOR HEMISPHERICAL SOLAR RADIATION
ISO STANDARD
EQUIVALENT
ASTM STANDARD
WMO
ISO/TR 9901:1990 Solar energy -- Field pyranometers -- Recommended practice for use
ASTM G183 - 05 Standard Practice for Field Use of Pyranometers, Pyrheliometers and UV Radiometers
WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, chapter 7, measurement of radiation, 7.3 measurement of global and diff use solar radiation
4.3 General use for sunshine duration measurement
According to the World Meteorological Organization (WMO, 2003), sunshine duration during a given period is defined as the sum of that sub-period for which the direct solar irradiance exceeds 120 W/m
2
.
SR05-D1A3-PV manual v1801 24/83
WMO has approved the “pyranometric method” to estimate sunshine duration from pyranometer measurements (Chapter 8 of the WMO Guide to Instruments and Observation, 2008). This implies that a pyranometer may be used, in combination with appropriate software, to estimate sunshine duration. Ask for our application note.
Table 4.3.1 Standards with recommendations for instrument use in sunshine duration measurement
STANDARDS FOR INSTRUMENT USE FOR SUNSHINE DURATION
WMO
WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, chapter 8, measurement of s unshine duration, 8.2.2 Pyranometric Method
4.4 Specific use for outdoor PV system performance testing
Pyranometers are used for monitoring PV power plant efficiency, in order to measure incoming solar radiation independently from the PV system. Pyranometers can be placed in two positions:
plane of array (POA), parallel to the PV panels, for measurement of the in-plane
irradiance (also noted as G
i in IEC 61724-1)
horizontally, for measurement of the global horizontal irradiance (E, also noted as GHI in IEC 61724-1)
SR05 series is applicable in outdoor PV system performance testing. See also Huks eflux model SR15-D1 “digital first class pyranometer” and SR30-D1 “digital secondary standard pyranometer”.
Table 4.4.1 Standards with recommendations for i n strument use in PV system performance testing
STANDARDS ON PV SYSTEM PERFORMANCE TESTING
IEC / ISO STANDARD
EQUIVALENT ASTM STANDARD
IEC 61724-1; Photovoltaic system performance monitoring – guidelines for measurement, data exchange and analysis
COMMENT: Allows pyranometers or reference cells according to IEC 60904-2 and -6. Pyranometer reading requi r ed a c c uracy better than 5% of reading (Par 4.1)
COMMENT: equals JISC 8906 (Ja pa nese Industrial Standards Committee)
ASTM 2848-11; Standard Test M ethod for Reporting Photovoltaic Non-Concentrator System Performance
COMMENT: confirms that a pyranometer is the preferred instrument for outdoor PV testing. Specifically recommends a “first class” pyranometer (paragraph A 1. 2.1.)
SR05-D1A3-PV manual v1801 25/83
4.5 Specific use in meteorology and climatology
The World Meteoro logical Org anization (W MO) is a specialised agency of the United Nations. It is the UN system's authoritative voice on the state and behaviour of the earth's atmosphere and climate. WMO publishes WMO-No. 8; Guide to Meteorological Instruments and Met hod s of Observation, in which a table is included on “level of performance” of pyranometers. Nowadays WMO conforms itself to the ISO classification system.
Loading...
+ 58 hidden pages