Hukseflux LP02 User Manual

Copyright by Hukseflux | manual v1606 | www.hukseflux.com | info@hukseflux.com
USER MANUAL LP02
Second class pyranometer
Thermal Sensors
LP02 manual v1606 2/47
Warning statements
Putting more than 12 Volt across the se nsor wiring can lead to permanent damage to t h e sensor.
Do not use “open circuit detection” when measuring the sensor output.
LP02 manual v1606 3/47
Contents
Warning stat em e nts 2 Contents 3 List of symbols 4 Introduction 5 1 Ordering and checking at delivery 7
1.1 Ordering LP02 7
1.2 Included items 7
1.3 Quick instrument che ck 8
2 Instrument principle and theory 9 3 Specifications of LP02 12
3.1 Specifications of LP02 12
3.2 Dimensions of LP02 15
4 Standards and recommended practices for use 16
4.1 Classification standard 16
4.2 General use for solar radiation measurement 16
4.3 General use for sunshine duration measurement 16
4.4 Specific use in meteorology and climatology 17
5 Installation of LP02 18
5.1 Site selection and installation 18
5.2 Installation of the optional sun screen 19
5.3 Electrical connection 19
5.4 Requirements for data acquisition / amplification 20
6 Making a dependable measureme nt 21
6.1 The concept of dependability 21
6.2 Reliability of the measurement 22
6.3 Speed of repair and maintenance 23
6.4 Uncertainty evaluation 24
7 Maintenance and trouble shooting 26
7.1 Recommended maintenance and quality assurance 26
7.2 Trouble shooting 27
7.3 Calibration and checks in the field 28
7.4 Data quality assurance 29
8 LP02-TR 30
8.1 Introduction LP02-TR 30
8.2 Dimensions of LP02-TR 32
8.3 Electrical connection of LP02-TR 34
9 Appendices 37
9.1 Appendix on cable extension / replacement 37
9.2 Appendix on tools for LP02 38
9.3 Appendix on spare parts for LP02 38
9.4 Appendix on standards for classification and calibration 39
9.5 Appendix on calibration hierarchy 40
9.6 Appendix on meteorological radiation quantities 41
9.7 Appendix on ISO and WMO classification tables 42
9.8 Appendix on definition of pyranometer specifications 43
9.9 Appendix on terminology / glossary 44
9.10 EU declaration of conformity 45
LP02 manual v1606 4/47
List of sy m bols
Quantities Symbol Unit Voltage output U V
Sensitivity S V/(W/m
2
) Temperature T °C Electrical resistance R
e
Solar irradiance E W/m
2
Solar radiant exposure H W∙h/m
2
Time in hours h h (see also appendix 9.6 on meteorological quantities)
Subscripts
Not applicable
LP02 manual v1606 5/47
Introduction
LP02 is a solar ra diation se nsor that is applied in most common s olar radiation observations. It complies with the seco nd class specifications of the ISO 9060 standard and the WMO Guide. LP02 pyranometer is widely used in (agro-)meteorological applications and for PV system performance monitoring.
LP02 measures the solar radiation received by a plane surface from a 180
o
field of view
angle. This quantity, expressed in W/m
2
, is called “hemispherical” solar radia tion. LP02 pyranometer can be employed outdoors under the sun, as well as indoors with lamp­based solar simulators. Its orientation depends on the application and may be horizontal, tilted (for plane of array radiation) or inverted (for reflected radiation). LP02 pyranometer is a very good alternative to silicon cell (photodiode-based) pyranometers, which do not comply to the ISO 9060 standard. In combination with the right software, also sunshine duration may be measured.
Using LP02 is easy. The pyranometer can be connected directly to commonly used data logging systems. The irradiance in W/m
2
is calculated by dividing the LP02 output, a small voltage, by the sensitivity. This sensitivity is provided with LP02 on its calibration certificate.
The central equation governing LP02 is: E = U/S (Formula 0.1)
The instrument should be used in accordance with the recommended practices of ISO, WMO and ASTM.
Suggested use for LP02:
general meteorlogical observations
agricultural networks
PV system performance monitoring
Figure 0.1 LP02 second class pyranometer
LP02 manual v1606 6/47
Figure 0.2 LP02 second class pyranometer with LI19 read-out unit / datalogger
WMO has approved the “pyranometric method” to calculate sunshine duration from pyranometer measurements in WMO-No. 8, Guide to Meteorological Instruments and Methods of Observation. This implies that LP02 may be used, in combination with appropriate software, to estimate sunshine duration. This is much more cost-effective than using a dedicated sunshine duration sensor. Ask for our application note.
Model LP02-TR houses a 4-20 mA transmitter for easy read-out by dataloggers commonly used in the industry. For more information see the chapter on LP02-TR.
Model LP02-LI19 offers LP02 with handheld read-out unit / datalogger LI19 in a practical transport case. The LP02-LI19 combination is well suited for mobile measurements and short term datalogging. For more information see the LP02-LI19 brochure and ask for the LI19 manual.
LP02 manual v1606 7/47
1 Ordering and checking at delivery
1.1 Ordering LP02
The standard configuration of LP02 is with 5 metres cable.
Common options are:
L onger cable (in multiples of 5 m). Specify total cable length.
LP02-TR second class pyranometer with 4-20 mA transmitter. Standard setting is 4
mA at 0 W/m
2
and 20 mA at 1600 W/m2. Specify setting and total cable length.
Adap ted sensitivity range. Specify the lower en higher end of the allowed range in x
10
-6
V/(W/m2).
Sun screen. Specify order number SCR01.
For a fast response second class pyranometer see model SR03.
1.2 Included items
Arriving at the customer, the delivery should include:
pyranometer LP02
cable of the length as ordered
product certificate matching the instrument serial number
any other options as ordered
Please store the certificate in a safe place.
LP02 manual v1606 8/47
1.3 Quick instrument check
A quick test of the instrument can be done by using a simple hand held multimeter and a lamp.
1. Check the electrical resistance of the sensor between the green (-) and white (+) wire. Use a multimeter at the 200 Ω range. Measure the sensor resistance first with one polarity, than reverse the polarity. Take the average value. The typical resistance of the wiring is 0.1 Ω/m. Typical resistance should be the typical sensor resistance of 40 to 60
Ω plus 1.5 Ω for the total resistance of two wires (back and forth) of each 5 m. Infinite
resistance indicates a broken circuit; zero or a low resistance indicates a short circuit.
2. Check if the sensor reacts to light: put the multimeter at its most sensitive range of DC voltage measurement, typically the 100 x 10
-3
VDC range or lowe r. Expose the sensor to a strong light source, for instance a 100 W light bulb at 0.1 m distance. The signal should read > 2 x 10
-3
V now. Darken the sensor either by putting something over it or switching off the light. The instrument voltage output should go down and within one minute approach 0 V.
3. If applicable, remove the optional sun screen, using the hex key (see chapter on installation of the s un screen). Inspect the bubble level.
4. Inspect the instrument for any damage.
LP02 manual v1606 9/47
2 Instrument principle and theory
Figure 2.1 Overview of LP02: (1) cable, (standard length 5 metres, optional longer cable)
(2) cable gland (3) thermal sensor with black coating (4) glass dome (5) sensor body (6) levelling feet (7) mounting hole (8) bubble level
1
2
3
4
5
6
7
8
LP02 manual v1606 10/47
LP02’s scientific name is pyranometer. A pyranometer measures the solar radiation received by a plane surface from a 180 ° field of view angle. This quantity, expressed in W/m
2
, is called “hemisph erical” solar radiation. The solar radiation spectrum extends
roughly from 285 to 3000 x 10
-9
m. By definition a pyranometer should cover that
spectral range with a spectral selectivity that is as “flat” as possible.
In an irradiance measurement by definition the response to “beam” radiation varies with the cosine of the angle of incidence; i.e. it should have full response when the solar radiation hits the sensor perpendicularly (normal to the surface, sun at zenith, 0° angle of incidence), zero response when the sun is at the horizon (90 ° angle of incidence, 90 ° zenith angle), and 50 % of full response at 60 ° angle of incidence. A pyranometer should have a so-called “directional response” (older documents mention “cosine response”) that is as close as possible to the ideal cosine characteristic.
In order to attain the proper directional and spectral characteristics, a pyranometer’s main components are:
a thermal sensor with black coating. It has a flat spectrum covering the 200 to 50000
x 10
-9
m range, and has a near-perfect directional response. The coating absorbs all solar radiation and, at the moment of absorption, converts it to heat. The heat flows through the sensor to the sensor body. The thermopile sensor generates a voltage output signal that is proportional to the solar irradiance.
a glass dome. This dome limits the spectral range from 285 to 3000 x 10
-9
m (cutting
off the part above 3000 x 10
-9
m), while preserving the 180 ° field of view angle. Another function of the dome is that it shields the thermopile sensor from the environment (conve ction, rain).
a second (inner) glass dome: For a first class pyranometer, two domes are used, and not one single dome. This construction pro vides an additional “radiation shield”, result in g in a better thermal equilibrium between the sensor and inner dome, compared to using a single dome. The effect of having a second dome is a strong reduction of instrument offsets.
Pyranometers can be manufactured to different specifications and with different levels of verification and characterisation during production. The ISO 9060 - 1990 standard, “Solar energy - specification and classification of instruments for measuring hemispherical solar and direct solar radiation”, distinguishes b etween 3 classes; secondary standard (highest accuracy), first class (second highest accuracy) and second class (third highest accuracy).
From second class to first class and from first class to secondary standard, the achievable accuracy improves by a factor 2.
LP02 manual v1606 11/47
Figure 2.2 Spectral response of the pyranometer compared to the solar spectrum. The pyranometer only cuts off a negligible part of the total solar spectrum.
0
0,2
0,4
0,6
0,8
1
1,2
100 1000 10000
relative spectral conten t /
response [arbitrary units]
wavelength [x 10
-9
m]
solar radiation
pyranometer response
LP02 manual v1606 12/47
3 Specifications of LP02
3.1 Specifications of LP02
LP02 measures the solar radiation received by a plane surface from a 180o field of view angle. This quantity, expressed in W/m
2
, is called “hemisph erical” solar radiation. Working completely passive, using a thermopile sensor, LP02 generates a small output voltage proportional to this flux. It can only be used in combination with a suitable measurement system. The instrument is classified according to ISO 9060 and should be used in accordance with the recommended practices of ISO, IEC, WMO and ASTM.
Table 3.1.1 Specifi ca t i on s of LP02 (continued on next pages)
LP02 MEASUREMENT SPECIFICATIO NS: LIST OF CLASSIFICATION CRITERIA OF ISO 9060*
ISO classification (ISO 9060: 1990)
second class pyranometer
WMO performance level (WM O -No-8, seventh edition 2008)
moderate quality pyranometer Response time (95 %)
18 s
Zero offset a (response to 200 W/m2
net thermal radiat ion)
< ± 15 W/m
2
unventilated
Zero offset b (response to 5 K/h
change in ambient temperature)
< ± 4 W/m2 Non-stability
< ± 1 % change per year
Non-linearity
< ± 1 % (100 to 1000 W/m2)
Directional resp onse
< ± 25 W/m2
Spectral selectivity
< ± 5 % (0.35 to 1.5 x 10
-6
m)
Temperature response
< ± 3 % (-10 to +40 °C)
Tilt response
< ± 2 % (0 to 90 ° at 1000 W/m2)
*For the exact definition of pyranometer ISO 9060 specifications see the appendix.
LP02 manual v1606 13/47
Table 3.1.1 Specifications of LP02 (continued)
LP02 ADDITIONAL SPECIFICATION S
Measurand
hemispherical solar radiation
Measurand in SI r a diometry units
irradiance in W/m2
Optional measurand
sunshine duration
Field of view angle
180 °
Measurement range
0 to 2000 W/m2
Sensitivity range
7 to 25 x 10-6 V/(W/m2)
Sensitivity (nominal)
15 x 10-6 V/(W/m2)
Expected voltage output application under natural solar radiation: -0.1 to + 50
x 10-3 V
Measurement function / required programming
E = U/S
Measurement function / optional
programming for sunshine duration
programming according to W M O guide paragraph
8.2.2
Required readout 1 differential voltage channel or 1 single ended
voltage channel, input resistance > 106
Optional readout
1 temperature channel in cas e opti onal temperature sensor is ordered
Rated operating temperatu r e r a nge
-40 to +80 °C
Sensor resistance range
40 to 60
Required sensor power
zero (passive sensor)
Spectral range (20 % transmission
points)
285 to 3000 x 10-9 m
Standard governing use of the
instrument
ISO/TR 9901:1990 Solar en er gy -- Field pyranometers
-- Recommended practice for u s e ASTM G183 - 05 Standa r d P r a c tice for Field Use of Pyranometers, Pyrheliometers and UV Radiom e ter s
Standard cable length (see options)
5 m
Cable diameter
4 x 10-3 m
Cable gland: cable diameter ra nge
accepts cable diameters from 4 to 6 x 10-3 m
Cable replacement cable can be removed and installed by the user
provided that the cable is sealed a t th e sensor side against humidity ingress. Consult Hukseflux for
instructions or u s e Hukseflux-supplied parts
Mounting 2 x M5 bolt at 65 mm centre-to-centre distance on
north-south ax is
Levelling
bubble level and adjustable lev elling feet are included
Levelling accuracy
< 0.4 ° bubble entirely in ring
IP protection cla s s
IP67
Gross weight including 5 m cable
0.5 kg
Net weight including 5 m cable
0.3 kg
Packaging
box of 170 x 90 x 230 x mm
LP02 manual v1606 14/47
Table 3.1.1 Specifications of LP02 (started on previous pages)
CALIBRATION
Calibration trace a b ility
to WRR
Calibration hierarchy from WRR through ISO 9846 and ISO 9847, applying
a correction to ref er e nce conditions
Calibration method
indoor calibration according to ISO 9847, Type IIc
Calibration un certainty
< 1.8 % (k = 2)
Recommended recalibrati on interval
2 years
Reference conditions 20 °C, normal incidence solar radiation, horizontal
mounting, irradiance level 1000 W/m2
Validity of calibra tion
based on experience the instrument sensitivity will not change during storage. Durin g use under exposure to solar radiation th e instrument “non-stability”
specification is applicable.
MEASUREMENT ACCURACY
Uncertainty of the measurement statements about the overall measurement
uncertainty can only be made on an individual basis.
See the chapter on uncertainty evaluation
VERSIONS / OPTIONS
Adapted sensitivity range
the rated sensitivity range c a n be adapted to
customer requirements (to a lower range only) option code = lower end - higher end of the allowed
range x 10-6 V/(W/m2)
4-20 mA transmitter creating a 4-20 mA output signal,
option code = TR, with adapted housing standard setting is 4 x 10
-3
A at 0 W/m2 and
20 x 10
-3
A at 1600 W/m
2
for specifications see the chapter on LP02-TR
Longer cable, in multiples of 5 m
option code = total cable length
ACCESSORIES
Sun screen for use on LP02
SCR01
Separate amplifiers
AC100 and AC420
Handheld read-out unit LI19
LP02-LI19 consists of one LP02 pyranometer, one programmed LI19 handheld read-out unit / datalogger, two spare batter ies, one USB cable,
software and a transport case
LP02 manual v1606 15/47
3.2 Dimensions of LP02
Figure 3.2.1 Dimensions of LP02 in 10
-3
m.
59
Ø 5
22
65
Ø 78
LP02 manual v1606 16/47
4 Standards and recommended practices
for use
Pyranometers are classified according to the ISO 9060 standard and the WMO-No. 8 Guide. In any application the instrument should be used in accordance with the recommended practices of ISO, IEC, WMO and / or ASTM.
4.1 Classification standard
Table 4.1.1 Standards for pyranometer classification. See the appendix for definitions of
pyranometer specifications, and a table listing the specification limits.
STANDARDS FOR INSTRUMENT CLASSI F I CAT I O N
ISO STANDARD EQUIVALENT
ASTM STANDARD
WMO
ISO 9060:1990 Solar energy -- specification and classification of instruments for measuring hemispherical solar and direct solar radiation
Not available
WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, chapter 7, measurement of radiation, 7.3 measurement of global and diff use solar radiation
4.2 General use for solar radiation measurement
Table 4.2.1 Standards with recommendations for instrument use in solar radiation
measurement
STANDARDS FOR INSTRUMENT USE FOR HEM ISP H ERICAL SOLAR RADIATION
ISO STANDARD EQUIVALENT
ASTM STANDARD
WMO
ISO/TR 9901:1990 Solar energy -- Field pyranometers -- Recommended practice for use
ASTM G183 - 05 Standard Practice for Field Use of Pyranometers, Pyrheliometers and UV Radiometers
WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, chapter 7, measurement of radiation, 7.3 measurement of global and diff use solar radiation
4.3 General use for sunshine duration measurement
According to the World Meteorological Organization (WMO, 2003), sunshine duration during a given period is defined as the sum of that sub-period for which the direct solar irradiance exceeds 120 W/m
2
.
LP02 manual v1606 17/47
WMO has approved the “pyranometric method” to estimate sunshine duration from pyranometer measurements (Chapter 8 of the WMO Guide to Instruments and Observation, 2008). This implies that a pyranometer may be used, in combination with appropriate software, to estimate sunshine duration. Ask for our application note.
Table 4.3.1 Standards with recommendations for instrument use in sunshine duration measurement
STANDARDS FOR INSTRUMENT USE F O R SUNSHINE DURATION
WMO
WMO-No. 8; Guide to Meteorological Instruments and Method s of Observation, chapter 8, measurement of s unshine duration, 8.2.2 Pyranometric Method
4.4 Specific use in meteorology and climatology
The World Meteorological Organization (WMO) is a specialised agency of the United Nations. It is the UN system's authoritative voice on the state and behaviour of the earth's atmosphere and climate. WMO publishes WMO-No. 8; Guide to Meteorological Instruments and Met hod s of Observation, in which a table is includ ed on “level of performance” of pyranometers. Nowadays WMO conforms itself to the ISO classification system.
LP02 manual v1606 18/47
5 Installation of LP02
5.1 Site selection and installation
Table 5.1.1 Recommendations for installation of pyranometers
Location
The situation tha t s hadows are cast on the
instruments is u sually not desirable. The horizon should be as free from obstacles as possible. Ideally there should be no objects between th e c ourse of the sun and the instrument.
Mechanical mounting / thermal insulation preferably use connection by bolts to the bottom plate
of the instrument. A pyranometer is sen s itive to thermal shocks. Do not mount the instrument with the body in direct thermal contact to the mounting plate (so always use the levelling feet also if the mountin g is not horizontal), do not mount the instrumen t on objects that become very hot (black coated metal plates).
Instrument moun ting with 2 bolts
2 x M5 bolt at 65 x 10-3 m centre to centre distance
on north-south axis, connection through the pyranometer flange.
Instrument moun ting with one bolt not applicable
Performing a representativ e measurement
the pyranometer measures the solar radiation in the plane of the sensor . This may require installation in a tilted or inverted position. The black sensor su rface (sensor bottom plate) should be mounted parallel to the plane of interest. In case a pyranometer is not mounted horizontally or in case the horizon is obstructed, the representativeness of th e loca tion becomes an important element of the meas urement. See the chapter on uncertainty evaluation.
Levelling in case of horizontal m ounting only use the bubble
level and levelling feet. The optional sun screen must be removed for inspection of the bubble level.
Instrument orientation
by convention w ith the cable exit pointing to the
nearest pole (so the cable exit should point north in the northern hemisphere, south in the southern hemisphere).
Installation height
in case of inverte d installation, WMO recommend s a
distance of 1.5 m between soil surface and sensor (reducing the effect of sh adows a nd in order to obtain good spatial averaging).
LP02 manual v1606 19/47
5.2 Installation of the optional sun screen
The optional SCR01 sun screen can be installed and removed using a hex key (size 2 x 10
-6
m). See the drawing below.
Figure 5.2.1 Installation and removal of sun screen: Turn the set screw using the hex key and lift of the sun screen. (1) hex key, (2) sun screen, (3) set scre w
5.3 Electrical connection
In order to operate, a pyranometer should be connected to a measurement system, typically a so-called datalogger. LP02 is a passive sensor that does not need any power. Cables generally act as a source of distortion, by picking up capacitive noise. We recommend keeping the distance between a datalogger or amplifier and the sensor as short as possible. For cable extension, see the appendix on this subject.
Table 5.3.1 The electrical connection of LP02
WIRE
COLOUR
MEASUREMENT SYSTEM
Sensor output + White
Voltage input +
Sensor output - Green
Voltage input - or ground
Shield
Black
Analogue ground
Figure 5.3.1 Electrical diagram of LP02. The shield is connected to the sensor body.
1 2 3
LP02 manual v1606 20/47
Table 5.3.2 Standard internal connection of LP02 at the internal printed circuit board
5.4 Requirements for data acquisition / amplification
The selection and programming of dataloggers is the responsibility of the user. Please contact the supplier of the data acquisition and a mp lification equipment to see if directions for use with the LP02 are available. In case programming for similar instruments is available, this can typically also be used. LP02 can usually be treated in the same way as other thermopile pyranometers. Pyranometers usually have the same programming as heat flux sensors.
In case of the LP02-TR version, the output is 4 to 20 x 10
-3
A. See the chapter on the LP02-TR. When using LP02 combined with read-out unit / datalogger LI19, please consult the LI19 manual as well.
Table 5.4.1 Requirements for data acquisition and amplification equipment for LP02 in the standard configuration
Capability to measure small voltage signals
preferably: 5 x 10
-6
V uncertainty
Minimum requirement: 20 x 10
-6
V uncertainty (valid for the entire expected temperature range of the acquisition / am plification equipment)
Capability for th e da ta logger or the software
to store data, and to per form division by the sensitiv ity to calculate the solar irradiance. E = U/S (Formula 0.1)
Data acquisition input resistance
> 1 x 106
Open circuit detec tion (WARNING)
open-circuit detec tion should not be used, unless th is is don e separately from the norma l mea s urement by more than 5 times the sensor resp onse time and with a small current only. Thermopile s ensors are sensitive to the curr ent that is used during open c ir c uit detection. The current will generate heat, which is measured and will appear as an offset.
SENSOR
PRINTED CIRCUIT COLOUR CODE WIRE
Plus (+)
+
White
Minus (-) - Green
Shield
SH
Bare metal
TR Not connected unless the sensor
has a trimmed sens itivity or limited sensitivity range
LP02 manual v1606 21/47
6 Making a dependable measurement
6.1 The concept of dependability
A measurement with a pyranometer is called “dependable” if it is reliable, i.e. measuring within required uncertainty limits, for most of the time and if problems, once they occur, can be solved quickly.
The requiremen t s for a measurement with a pyranometer may be expressed by the user as:
required uncertainty of the measurement (see following paragraphs)
requirements for maintenance and repairs (possibilities for maintenance and repair
including effort to be made and processing time)
a requirement to the expected instrument lifetime (until it is no longer feasible to repair)
It is important to realise that the uncertainty of the measurement is not only determined by the instrument but also by the way it is used.
See also ISO 9060 note 5. In case of pyranometers, the measurement uncertainty as obtained during outdoor measurements is a function o f:
the instrument class
the calibration procedure / uncertainty
the duration of instrument employment under natural sunlight (involving the
instrument stability specification)
the measurement conditions (such as tilting, ventilation, shading, instrument temperature)
maintenance (mainly fouling)
the environmental conditions*
Therefore, ISO 9060 says, “statements about the overall measurement uncertainty under outdoor conditions can only be made on an individual basis, taking all these factors into account”.
* defined at Hukseflux as all factors outside the instrument that are relevant to the measurement such as the cloud cover (presence or absence of direct radiation), sun position, the local horizon (which may be obstructed) or condition of the ground (when tilted). The environmental conditions also involve the question whether or not the measurement at the location of measurement is representative of the quantity that should be measured.
LP02 manual v1606 22/47
6.2 Reliability of the measurement
A measurement is reliable if it measures within required uncertainty limits for most of the time. We distinguish between two causes of unreliability of the measurement:
related to the reliability of the pyranometer and its design, manufacturing, calibration (hardware reliability).
re lated to the reliability of the measurement uncertainty (measurement reliability), which involves hardware reliability as well as co ndition of use.
Most of the hardware reliability is the responsibility of the instrument manufacturer. The reliability of the measurement however is a joint responsibility of instrument manufacturer and user. As a function of user requirements, taking into account measurement conditions and environmental conditions, the user will select an instrument of a certain class, and define maintenance support procedures.
In many situations the re is a limit to a realistically attainable accuracy level. This is due to conditions that are beyond control once the measurement system is in place. Typical limiting conditions are:
the measurement conditions, for instance whe n working at extreme temperatures when the instrument temperature is at the extre m e lim its of the rated temperature range.
the environmental conditions, for instance when installed at a sub-optimal measurement location with obstacles in the path of the sun.
other environmental conditions, for instance when assessing PV system performance and the system contains panels at different tilt angles, the pyranometer measurement may not be representative of irradiance received by the entire PV system.
The measurement reliability can be improved by maintenance support. Important aspects are:
dom e fouling by deposition of dust, dew, rain or snow. Fouling results in undefined measurement uncertainty (sensitivity and directional error are no longer defined). This should be solve d by regular inspection and cleaning.
sensor instability. Maximum expected sensor aging is specified per instrument as its non-stability in [% change / year]. In case the sensor is not recalibrated, the uncertainty of the sensitiv ity gradually will increase. This is solved by regular recalibration.
moisture condensing under pyranometer domes resulting in a slow change of sensitivity (within specifications). This is solved by regular replacement of desiccant or by maintenance (drying the entire sensor) in case the sensor allows this. For non­serviceable sensors like Hukseflux second cl ass pyranometers (for exampl e model LP02), this may slowly develop into a def ect. For first class and secondary standard models (for instance model SR11 first class pyranometer) extra desiccant (in a set of 5 bags in an air-tight bag) is available.
LP02 manual v1606 23/47
Another way to improve measurement reliability is to introduce redundant sensors.
The use of redundant instruments allow s remote checks of one instrument using the other as a reference, which leads to a higher measurement reliability.
In PV system performance monitoring, in addition to instruments measuring in the plane of array, horizontally placed instruments are used for the measurement of global radiation. Global irradiance data enable the user to compare the local climate and system efficiency between different sites. These data can also be compared to measurements by local meteorological stations.
6.3 Speed of repair and maintenance
Dependability is not only a matter of reliability but also involves the reaction to problems; if the processing time of service and repairs is short, this contributes to the dependability.
Hukseflux pyranometers are designed to allow easy maintenance a nd repair. The main maintenance actions are:
replacement of desiccant (not applicable for LP02)
replacement of cabling
For optimisation of dependability a user should:
design a schedule of regular maintenance
design a schedule of repair or replacement in case of defects
When operating multiple instruments in a network Hukseflux recommends keeping procedures simple and having a few spare instruments to act as replacements during service, recalibrations and repair.
LP02 manual v1606 24/47
6.4 Uncertainty evaluation
The uncertainty of a mea surement under outdoor or indoor co nditions depends on many factors, see paragraph 1 of this chapter. It is not possible to give one figure for pyranometer measurement uncertainty. The work on uncertainty evaluation is “in progress”. There are several groups around the world participating in standardisation of the method of calculation. The effort aims to work according to the guidelines for uncertainty evaluation (according to the “G uide to Expression of Uncertainty in Measurement” or GUM).
6.4.1 Evaluation of measurement uncertainty under outdoor conditions
Hukseflux actively participates in the discussions about pyranometer measurement uncertainty; we also provide spreadsheets, reflecting the latest state of the art, to assist our users in making their own evaluation. The input to the assessment is summarised:
1) The formal evaluation of uncertainty should be performed in accordance with ISO 98-3
Guide to the Expression of Uncertainty in Measurement, GUM.
2) The spe cifications of the instrument according to the list of ISO 9060 classification of
pyranometers and pyrheliometers are entered as limiting values of possible errors, to be analysed as type B evaluation of standar d uncerta inty per paragraph 4.3.7. of GUM. A priori distributions are chosen as rectangular.
3) A separate estimate has to be entered to allow for estimated uncertainty due to the
instrument maintenance level.
4) The calibration uncertainty has to be entered. Please note that Hukseflux calibration
uncertainties are lower than those of alternative equipment. These uncertainties are entered in measurement equation (equation is usually Formula 0.1: E = U/S), either as an uncertainty in E (zero offsets, direc t ional response) in U (voltage readout errors) or in S (tilt error, temperature dependence, calibration uncertainty).
5) In uncertainty analysis for pyranometers, the location and date of interest is entered.
The course of the sun is then calculated, and the direct and diffuse components are estimated, based on a model; the angle of incidenc e of direct radiation is a major factor in the uncertainty.
6) In uncertainty analysis for modern pyrheliometers: tilt dependence often is so low that
one single typical observat ion may be sufficient.
7) In cas e of specia l measurement conditions, typical specification values are chosen.
These should for instance account for the measurement conditions (shaded / unshaded, ventilated/ unventilated, horizontal / tilted) and environmental conditions (clear sky / cloudy, working temperature range).
8) Among the various sources of uncertainty, some are “correlated”; i.e. present during
the entire measurement process, and not cancelling or converging to zero when averaged over time; the off-diagonal elements of the covariance matrix are not zero. Paragraph 5.2 of GUM.
9) Among the various sources of uncertainty, some are “uncorrelated”; cancelling or
converging to zero when averaged over time; the off-diagonal elements of the covariance matrix are zero. Paragraph 5.1 of GUM.
LP02 manual v1606 25/47
10) Among the various sources of uncertainty, some are “not included i n analysis”; this
applies for instance to non-linearity for pyranometers, because it is already included in the directional error, and the spectral response for pyranometers and pyrheliometers because it is already taken into account in the calibration process.
Table 6.4.1.1 Preliminary estimates of achievable uncertainties of measurements with Hukseflux pyranometers. The estimates are based on typical pyranometer properties and calibration uncertainty, for sunny, clear sky days and well maintained stations, without uncertainty loss due to lack of maintenance and due to instrument fouling. The table specifies expanded uncertainties with a coverage factor of 2 and confidence level of 95 %. Estimates are based on 1 s sampling. IMPORTANT NOTE: there is no international consensus on uncer t ainty evaluation of pyranometer measurements, so this table should not be used as a formal reference.
Pyranometer class (ISO 9060)
season latitude uncertainty
minute totals at solar noon
uncertainty hourly totals at solar noon
uncertainty daily totals
secondary standard
summer
mid-latitude
2.7 %
2.0 %
1.9 %
equator 2.6 % 1.9 % 1.7 %
pole
7.9 %
5.6 %
4.5 %
winter
mid-latitude
3.4 %
2.5 %
2.7 %
first class
summer
mid-latitude
4.7 %
3.3 %
3.4 %
equator
4.4 %
3.1 %
2.9 %
pole
16.1%
11.4 %
9.2 %
winter
mid-latitude
6.5 %
4.5 %
5.2 %
second class
summer
mid-latitude
8.4 %
5.9 %
6.2 %
(LP02)
equator
7.8 %
5.5 %
5.3 %
pole
29.5 %
21.6 %
18.0 %
winter
mid-latitude
11.4 %
8.1 %
9.9 %
6.4.2 Calibration uncertainty
From 2011 to 2012, calibration of LP02 has been improved. New procedures were developed in close cooperation with PMOD World Radiation Center in Davos, Switzerland. Our latest calibration method results in an uncertainty of the sensitivity of less than
1.8 %, compared to typical uncertainties of higher than 3.5 % for this pyranometer
class. See the appendix for detailed information on calibration hierarchy.
LP02 manual v1606 26/47
7 Maintenance and trouble shooting
7.1 Recommended maintenance and quality assurance
LP02 can measure reliably at a low level of maint enance in most locations. Usually unreliable measurements will be detected as unreasonably large or small measured values. As a general rule this means that regular visual inspection combined with a critical review of the measured data, preferably checking against other measurements, is the preferred way to obtain a reliable measurement.
Table 7.1.1 Recommended maintenance of LP02. If possible the data analysis and cleaning (1 and 2) should be done on a daily basis.
MINIMUM RECOMMENDED PYRANOMETER MAI NTENANCE
INTERVAL
SUBJECT
ACTION
1 1 week data analysis compare measured data to maximum possible / maximum
expected irradiance and to other measurements nearby (redundant instruments). Also historical seasonal records can be used as a source for expected values. Analyse night time signals. These signals may be negative (down to -5 W/m
2
on clear windless nights), due to zero offset a. In case of use with PV systems, compare daytime measurements to PV system output. Look for a ny patterns and events that dev ia te from
what is normal or expected.
2
2 weeks
cleaning
use a soft cloth to clea n the dome of the instrument, persistent stains can be treated with soapy water or a lcohol
3 6 months inspection inspect cable quality, inspect cable glands, ins pec t mounting
position, inspect cable, clean instrument, clean cable, inspect levelling, change instrument tilt in case this is ou t of specification, inspect mounting connection, inspect interior of
dome for condensation
4 desiccant
replacement
desiccant replacement (not applicable for LP02)
5 2 years recalibration recalibration by side-by-side comparison to a higher s ta ndard
instrument in th e field according to ISO 9847
6 lifetime assessment
judge if the instrument should be reliable for another 2 years, or if it should be replaced
7
6 years
parts
replacement
if applicable / necessary replace the parts that are most
exposed to weathering; cable, c a ble gland, sun screen.
NOTE: use Hukseflux approved parts only.
8 internal
inspection
if applicable: open instrument and inspect / replace O-rings;
dry internal cav ity around the circuit board
9 recalibration recalibration by side-by-side comparison to a higher s ta ndard
instrument indoors according to ISO 9847 or outdo ors
according to ISO9846
LP02 manual v1606 27/47
7.2 Trouble shooting
Table 7.2.1 Trouble shooting for LP02
The sensor does not give any signal
Check the electrical resista nce of the sensor between the green (-) and white (+) wire. Use a multimeter at the 200 Ω range. Measure the sensor resistance first with one polarity, than rev er se the polarity. Take the average value. The typical resistance of the wiring is 0.1 Ω/m. Typical resistance should be the typical sensor resistance of 40 to 60 Ω plus 1.5 Ω for the total resistance of two wires (back and forth) of each 5 m. Infinite resistance indic a tes a broken circuit; zero or a low r e s istance indicates a short circuit. Check if the sens or reacts to light: put the multimeter at its most sensitive range of DC voltage measu r e m ent, typically the 100 x 10
-3
VDC range or lower. Expose
the sensor to strong light source, for instance a 1 00 W light bulb at 1 x 10
-1
m
distance. The signal should read > 2 x 10
-3
V now. Darken the sensor either by putting somethin g over it or switching off the light. The instrument voltage out put should go down an d within one minute approach 0 V. Check the data a c q uisition by applying a 1 x 10
-6
V source to it in the
1 x 10
-6
V range.
The sensor signal is unrealistically high or low.
Note that night-time signals may be negative (down to -5 W/m
2
on clear windless nights), due to zer o offset a. Check if the pyranometer ha s c lea n domes. Check the location of the pyranometer; are there any obstructions that could explain the measurement result. Check the orientation / levelling of the pyranometer. Check if the right c alibration factor is entered in to the algorithm. Please note that each sensor has its own individual calibration factor, as documented in its calibration cert ificate. Check if the voltage r e a d ing is divided by the calibrati on factor in review of the algorithm. Check the condition of the wiring at the logger. Check the cable condition looking for cable breaks. Check the range of the data logger; signal can be negative (this could be out of range) or the amplitude could be ou t of ra nge. Check the data acquisition by a pplying a 1 x 10
-6
V source to it in the
1 x 10
-6
V range. Look at the output. Check if the output is as e xpected. Check the data a c q uisition by short circuiting the da ta acquisition input with a 100 Ω resistor. Look at the output. Check if the output is close to 0 W/m2.
The sensor signal shows unexpected variations
Check the presence of strong sources of electromagnetic radiation (radar, radio etc.) Check the condition of the shielding. Check the condition of the sensor cable. Check if the cable is not moving du r in g the measurement
The dome shows internal condensation
Arrange to send the sensor back to Hukseflux for diagnosis.
LP02 manual v1606 28/47
7.3 Calibration and checks in the field
Recalibration of field pyranometers is typically done by comparison in the field to a reference pyranometer. The applicable standard is ISO 9847 “International Standard­Solar Energy- calibration of field pyranometers by comparison to a reference pyranometer”. At Hukseflux an indoor calibration according to the same standard is used.
Hukseflux recommendation for re-calibration: if possible, perform cal ibration indoor by comparison to an identical reference instrument, under normal incidence conditions.
In case of field comparison; ISO recommends field calibration to a higher class pyranometer. Hukseflux suggests also allowing use of sensors of the same model and class, because intercomparisons of similar instruments has the advantage that they suffer from the same offsets. It is therefore just as good to compare to pyranometers of the same brand and type as to compare to an instrument of a higher class. ISO recommends to perform field calibration during several days; 2 to 3 days under cloudless conditions, 10 days un d er cloudy conditions. In general th is is not achievable. In order to shorten the calibration process Hukseflux suggests to allow calibration at normal incidence, using hourly totals near solar noon.
Hukseflux main recommendations for field intercomparisons are:
1) to take normal incidence as a reference and not the entire day.
2) to take a reference of the same brand and type as the field pyranometer or a pyranometer of a higher class, and
3) to connect both to the same electronics, so that electronics errors (also offsets) are eliminated.
4) to mount all instruments on the same platform, so that they have the same body temperature.
5) assuming that the electronics are indepe ndently calibrated, to analyse radiation values at normal incidence radiation (possibly tilting the radiometers to approximately normal incidence); if this is not possible to compare 1 hour totals around solar noon for horizontally mounted in struments.
6) for second class radiometers, to co rr ect deviations of more than ± 10 %. Lower deviations should be interpreted as acceptable and should not lead to a revised sensitivity.
7) for first class pyranometers, to correct deviations of more than ± 5 %. Lower deviations should be interpreted as acceptable and should not lead to a revised sensitivity.
8) for secondary standard instruments, to correct deviations of more than ± 3 %. Lower deviations should be interpreted as acceptable and should not lead to a revised sensitivity.
LP02 manual v1606 29/47
7.4 Data quality assurance
Quality assurance can be done by:
analysing trends in solar irradiance signal
• plotting the measured irradiance against mathematically generated expected values
comparing irradiance measurements between sites
analysis of night time signals
The main idea is that one should look out for any unrealistic values. There are programs on the market that can automatically perform data screening. See for more information on such a program http://www.dqms.com.
LP02 manual v1606 30/47
8 LP02-TR
8.1 Introduction LP02-TR
As a special version of LP02, Hukseflux offers model LP02-TR: a second class pyranometer with 4-20 mA transmitter.
LP02-TR is a solar radiation sensor that is applied in most common solar radiation observations. It complies with the second class sp ecifications of the ISO 9060 standard and the WMO Guide. LP02-TR pyranometer is widely used in (agro-)meteorological applications and for PV system performance monitoring. LP02-TR houses a 4-20 mA transmitt er for ea s y rea d-out by dataloggers commonly used in the industry.
Using LP02-TR is easy. The pyranometer can be connected directly to commonly used data logging systems. The irradiance in W/m
2
is calculated by using the transmitter’s output. In LP02-TR’s standard configuration, the 4 to 20 mA output corresponds to a transmitted range of 0 to 1600 W/m
2
. This range can be adjusted at the factory upon
request.
Figure 8.1.1 LP02-TR second class pyranometer with 4-20 mA transmitter
LP02 manual v1606 31/47
Table 8.1.1 Specifications of LP02-TR
LP02-TR SPECIFICATIONS
Description
second class pyranometer with 4-20 m A transmitter
Transmitted range
0 to 1600 W/m2
Output signal
4 to 20 x 10-3 A
Principle
2-wire current loop
Supply voltage
7.2 to 35 VDC
Options adapted transmitted range
longer cable, in multiples of 5 m
For de finition of pyranometer ISO 9060 specifications see the appendix.
Table 8.1.2 Requirements for data acquisition and amplification equipment with the LP02–TR configuration
Capability to
- measure 4-20 mA or
- measure currents or
- measure voltages
The LP02-TR has a 4-20 mA output. There are several possibilities to h a ndle this signal. It is important to r ealise that the signal wire s not only act to transmit the signal but also act as power supply. Some dataloggers have a 4-20 mA input. In that case the connection can be directly made. Some dataloggers have the capability to measure currents. In some cases the datalogger accepts a voltage input. Usually a 100 Ω precision resistor is used to convert the current to a voltage ( this will then be in the 0.4 – to 2 VDC range). This r esistor must be put in the + wire of the sensor. In the two latter ca s e s the user must check that the low side of the input channel is connected to ground, and the high side to a positive voltage in th e r equired range. See the chapter on LP02-TR’s electrical connections for wiring diagrams and electrical connections to am - and voltmeters.
LP02 manual v1606 32/47
8.2 Dimensions of LP02-TR
Figure 8.2.1 Overview of LP02-TR:
(1) cable, (standard length 5 metres, optional longer cable) (2) cable gland (3) thermal sensor with black coating (4) glass dome (5) sensor body (6) transmitter hous i ng (7) levelling feet (8) bubble level
1
2
3
4
5
6
7
8
LP02 manual v1606 33/47
Figure 8.2.2 Dimensions of LP02-TR in 10
-3
m
106
Ø 78
65
M5
M6
LP02 manual v1606 34/47
8.3 Electrical connection of LP02-TR
In order to operate, a pyranometer should be connected to a measurement system, typically a so-called datalogger. LP02-TR operates on a supply voltage of 7.2 to 35 VDC.
Table 8.3.1 The electrical connection of LP02-TR
WIRE
COLOUR MEASUREMENT SYSTEM
Sensor output + White
Voltage input +
Sensor output - Green
Voltage input - or ground
Shield
Black
Analogue ground
Figure 8.3.1 Electrical diagram of LP02-TR. The shield is connected to the sensor body.
Table 8.3.2 Standard internal connection of LP02 at the internal printed circuit board
SENSOR
PRINTED CIRCUIT COLOUR CODE WIRE
Plus (+) + White
Minus (-) - Green
Shield SH Bare metal
TR Not connected unless the sensor
has a trimmed sens itivity or limited sensitivity range
TR
LP02 manual v1606 35/47
Figure 8.3.2 Electrical diagram of the connection of LP02-TR to a typical ammeter or datalogger with capacity to measure current signals. LP02-TR operates on a supply voltage of 7.2 to 35 VDC.
Figure 8.3.3 Electrical diagram of the connection of LP02-TR to a typical voltmeter or datalogger with the capacity to measure voltage signals. Usually a 100 Ω shunt resistor (R) is used to convert the current to a voltage. LP02-TR operates on a supply voltage of
7.2 to 35 VDC.
TR sensor
white [+]
green [–]
A
ground
7.2 - 35 VDC
ammeter I = 4 to 20 mA
black
TR sensor
white [+]
green [–]
V
ground
7.2 - 35 VDC
I = 4 to 20 mA
voltmeter I = U/R
R
black
LP02 manual v1606 36/47
LP02 manual v1606 37/47
9 Appendices
9.1 Appendix on cable extension / replacement
The sensor cable can be removed and installed by the user prov ided that the cable is sealed at the sensor side against humidity ingress. Please consult Hukseflux for instructions on cable preparation or use Hukseflux-supplied parts.
LP02 is equipped with one cable. Keep the distance betwee n d a ta logger or amplifier and sensor as short as possible. Cables act as a source of distortion by picking up capacitive noise. In an electrically “quiet” environment the LP02 cable can however be extended without problem to 100 meters. If done properly, the sensor signal, although small, will not significantly degrade because the sensor resistance is very low (so good immunity to external sourc e s) and because there is no current flowing (so no resistive losses). Cable and connection specifications are summarised below.
NOTE: the body of LP02 contains connector blocks that can be used for the internal connection of a new cable. See the chapter on electrical connections. Usually it is e a s ie r to connect a new extended cable inside the pyranometer body cable than to make a good (weatherproof) connection to an existing cable.
Table 9.1.1 Preferred specifications for cabl e e xt en si on of LP02
General
please consult Hukseflux for instructions or use Hukseflux-supplied parts.
Cable
2-wire, shielded, with copper conductor (at Hukseflux 3-wire shielded cable is used, of which only 2 wires are used)
Sealing
sealed at the sensor side against humidity ingr es s
Core resistance
< 0.1 /m
Outer diameter
± 4 to 6 x 10
-3
m (to fit cable gland)
Length
cables should be kept as short as possible, in any case the total cable length should be less than 100 m
Outer sheet
with specificati ons for outdoor use (for good stability in outdoor applications)
Connection
either solder the new cable core and shield to the original sensor cable, and make a waterproof connection using cable s hrink, or use gold plated waterproof connectors. Always connect sh ie ld.
LP02 manual v1606 38/47
9.2 Appendix on tools for LP02
Table 9.2.1 Specifications of tools for LP02
tooling required for optional sun screen f ixation and removal hex key 2 mm
tooling required for cable gland fixation and r e m oval
spanner size mm
tooling required for wire fixation and removal (internal wiring inside LP02 body)
screwdriver blade width mm
9.3 Appendix on spare parts for LP02
Levelling feet (set of 3)
LP02 cable (specify length in multiples of 5 m), sealed at one end
Cable gland LP02
O-ring LP02
LP02 manual v1606 39/47
9.4 Appendix on standards for classification and calibration
Both ISO and ASTM have standards on instrument classification and methods of calibration. The World Meteorologica l Organisation (WMO) has largely adopted the ISO classification system.
Table 9.4.1 Pyranometer standardisation in ISO and ASTM.
STANDARDS ON INSTRUMENT CLASSIFICATI O N AND CALIBRATION
ISO STANDARD
EQUIVALENT ASTM STANDARD
ISO 9060:1990 Solar energy -- Specification and classification of instruments for measu r ing hemispherical so lar and direct solar radiation
not available Comment: work is in progress on a new ASTM equivalent standard
Comment: a standard “Solar energy --Methods for testing pyranometer an d pyrheliometer characteristics” has been announced in ISO 9060 but is not yet implemented.
not available
ISO 9846:1993 Solar energy -- Calibration of a pyranometer using a pyrheliometer
ASTM G167 - 05 Standa r d Test Method for Calibration of a Pyranometer Using a Pyrheliometer
ISO 9847:1992 Solar energy -- Calibration of field pyranometer s by comparison to a reference pyranometer
ASTM E 824 -10 Standard Test Method for Transfer of Calibration from Reference to Fie ld Radiometers
ASTM G207 - 11 Standa r d T es t M ethod for Indoor Transfer of Calibration from Referenc e to Field Pyranometers
ISO 9059:1990 Solar energy -- Calibration of field pyrheliome te r s b y comparison to a reference pyrheliometer
ASTM E 816 Standard Test Method for Calibration of Pyrheliometers by Comparis on to Reference Pyrheliometers
LP02 manual v1606 40/47
9.5 Appendix on calibration hierarchy
The World Radiometric Reference (WRR) is the measurement standard representing the SI unit of irradiance. It was introduced in order to ensure world-wide homogeneity of solar radiation measurements and is in use since 1980. The WRR was determined from the weighted mean of the measurements of a group of 15 absolute cavity radiometers which were fully characterised. It has an estimated accuracy of 0.3%. The WMO introduced its mandatory use in its status in 1979. The world-wide homogeneity of the meteorological radiation measurements is guaranteed by the World Radiation Center in Davos Switzerland, by ma intaining the World Standard Group (WSG) which materialises the World Radiometric Reference.
See http://www.pmodwrc.ch The Hukseflux standard is traceable to an outdoor WRR calibration. Some small
corrections are made to transfer this calibration to the Hukseflux standard conditions: sun at zenith and 1000 W/m
2
irradiance level. During the outdoor calibration the sun is
typically at 20 to 40 ° zenith angle, and the total irradiance at a 700 W/m
2
level.
Table 9.5.1 Calibration hierarchy for pyranometers
WORKING STANDARD CALIBRATION AT PMOD / WRC DAVOS
Calibration of working standard pyranometers: Method: ISO 9846, type 1 outdo or. This working standard has an uncertainty “uncertainty of standard”. The working standard has been calibrated under certain “test conditions of the standard”. The working standard has traceability to WRR world radiometr ic reference.
CORRECTION OF (WORKING) STANDARD CALIBRATION TO STANDARDISED
REFERENCE CONDITIONS
Correction from “test condition s of the standard” to “referen c e c onditions” i.e. to normal incidence and 20 °C: Using known (working) standard pyranometer properties: directional, non linearity, offsets, temperature dependence) . This correction has an uncertainty; “uncertainty of correction”. At Hukseflux we a ls o call the working standard pyr anometer “standard”.
INDOOR PRODUCT CALIBRATION
Calibration of products, i.e. pyranometers: Method: according to ISO 9847, Type IIc, which is an indoor calibration. This calibration has an uncertainty associated with the m ethod. (In some cases like the BSRN network the product calibration is with a different method; for example again type 1 ou tdoor)
CALIBRATION UNCERTAINTY CALCULAT ION
ISO 98-3 Guide to the Expression of Uncertainty in Measurement, GU M Determination of combined expanded uncert a in ty of calibration of the product, including uncertainty of the working standard, uncertainty of corre c tion, uncertainty of the method (transfer error). The coverage factor must be determined; at Hukseflux we work with a coverage factor k = 2.
LP02 manual v1606 41/47
9.6 Appendix on meteorological radiation quantities
A pyranometer measures irradiance. The time integrated total is called radiant exposure. In solar energy radiant exposure is often given in W∙h/m
2
.
Table 9.6.1 Meteorological radiation quantities as recommended by WMO (additional symbols by Hukseflux Thermal Sensor). POA stands for Plane of Array irradi an ce. The term originates from ASTM and IEC standards.
SYMBOL DESCRIPTION CALCULATION UNITS ALTERNATIVE
EXPRESSION
E
downward irradiance
E = Eg ↓ + El↓
W/m2
H
downward radiant exposur e for a specified time interval
H = H
g
+ Hl
J/m
2
E↑
upward irradiance
E↑ = E
g
+ E
l
W/m2
H
upward radiant exposure for a specified time interval
H = H
g
+ Hl
J/m
2
Wh/m2 Change of
units
E direct solar irradiance
normal to the apparen t
solar zenith angle
W/m2 DNI Direct
Normal
Irradiance
E0 solar constant W/m2
E
g
h
global irradiance;
hemispherical irradiance on a specified, in this case
horizontal surface.*
E
g
= E cos θh + E
d
W/m2
GHI
Global
Horizontal Irradiance
Eg t
global irradiance; hemispherical irradiance on a specified, in this case
tilted surface.*
Eg ↓ = E∙cos θt + E
d
t + Er↑ t ***
W/m2 POA Plane of
Array
Ed
downward diffuse solar radiation
W/m
2
DHI Diffuse
Horizontal
Irradiance
E
l
, E
l
upward / downward long-
wave irradiance
W/m2 E
r
reflected solar irradiance
W/m2
E* net irradiance
E* = E↓ – E↑
W/m2
T↓
apparent surface temperature**
ºC or K
T
apparent sky
temperature**
ºC or K
SD
sunshine duration
h
θ is the apparent solar zenith angle θ
h
relative to horizontal, θt relative to a tilted surface g = global, l = long wave, t = tilted *, h = horizontal* * distinction horizontal and tilted from Hukseflux, ** T symbols introduced by Hukseflux, *** contributions of E
d
t and Er↑ t are Ed ↓ and E
r↑
both corrected for the tilt angle of the
surface
LP02 manual v1606 42/47
9.7 Appendix on ISO and WMO classification tables
Table 9.7.1 Classification table for pyranometers per ISO 9060 and WMO.
NOTE: WMO specification of spectral selectivity is different from that of ISO. Hukseflux conforms to the ISO limits. WMO also specifies expected accuracies. ISO finds this not to be a part of the classification system because it also involves calibration. Please note that WMO achievable accuracies are for clear days at mid latitudes and that the uncertainty estimate does not include uncertainty due to calibration*.
ISO CLASSIFICATION** T ABLE
ISO CLASS
SECONDARY STANDARD
FIRST CLASS
SECOND CLASS
Specification limit Response time (95 %) 15 s 30 s 60 s Zero offset a (respons e to 200 W/m2 net
thermal radiation)
+ 7 W/m2 + 15 W/m2 + 30 W/m2
Zero offset b (response to 5 K/h in a m b ient
temperature)
± 2 W/m2 ± 4 W/m2 ± 8 W/m2
Non-stability (change per year) ± 0.8 % ± 1.5 % ± 3 %
Non-linearity (100 to 1000 W/m2)
± 0.5 %
± 1 %
± 3 %
Directional response
± 10 W/m2
± 20 W/m2
± 30 W/m2
Spectral selectivity (350 to 1 500 x 10-9 m) (WMO 300 to 3 000 x 10
-9
m)
± 3 % ± 5 % ± 10 %
Temperature response (interval of 50 K)** 2 % 4 % 8 %
Tilt response
(0 to 90 ° at 1000 W/m2)
± 0.5 %
± 2 %
± 5 %
ADDITIONAL WMO SPECIFICATIONS
WMO CLASS HIGH QUALITY GOOD QUALITY MODERATE
QUALITY
WMO: achievable accuracy for daily sums*
2 %
5 %
10 %
WMO: achievable accuracy for hourly sums* 3 % 8 % 20 % WMO: achievable accuracy for minute sums* not specified not specified not specified
WMO: resolution
(smallest detectable change)
1 W/m2
5 W/m2
10 W/m2
CONFORMITY TESTING***
ISO 9060 individual
instrument only : all specs must
comply
group compliance
group compliance
* WMO 7.2.1: The estimated uncertainties are based on the following assumptions: (a) instruments are well-maintained, correctly aligned and clean; (b) 1 min and 1 h figures are for clear-sky irradiances at solar noon; (c) daily exposure values are for clear days at mid-latitudes. WMO 7.3.2.5: Table 7.5 lists the expected maximum deviation from the true value, excluding calibration errors. ** At Hukseflux we use the expression ± 1 % instead of a range of 2 %. *** an instrument is subject to conformity testing of its specifications. Depending on the classification, conformity compliance can be proven either by group- or individual compliance. A specificatio n is fulfilled if the mean value of the respective test result does not exceed the corresponding limiting value of the specification for the specific category of instrument.
LP02 manual v1606 43/47
9.8 Appendix on definition of pyranometer specifications
Table 9.8.1 Definition of pyranometer specifications
SPECIFICATION DEFINITION SOURCE
Response time (95 %)
time for 95 % response. T he time interval between the instant when a stimulus is subjected to a spec ified abrupt change and the instant when the response rea c hes and remains within specified limits around its f inal steady value.The response tim e is a mea s ure of the thermal iner tia inherent in the stabilizat ion period for a final reading.
ISO 9060­1990 WMO
1.6.3
Zero offset a: (200 W/m
2
net thermal radiation )
response to 200 W/m
2
net thermal r a diation (ventilated). Hukseflux assu mes that unventilated instruments have to specify the zero-offset in unventilated – worst case – conditions. Zero offsets are a m easure of the stability of the z e r o-point. Zero offset a is vis ib le a t night as a negative offset, the instrument dome irradiates in the far infra red to the relatively cold sky. This causes the dome to cool down. The pyranometer sensor irradiates to the relatively cool dome, causing a negative offset. Zero offset
a is also assumed to be present during day time.
ISO 9060­1990
Zero offset b: (5 K/h in ambient
temperature)
response to 5 K/h change in a m bien t temper ature. Zero offsets are a m easure of the stability of the z e r o-point.
ISO 9060-
1990
Non-stability (change per
year)
percentage change in sensitivi ty per year. The dependence of sensitivity resulting from ageing effects which is a measure of the
long-term stability .
ISO 9060-
1990
Non-linearity
(100 to 1000 W/m
2
)
percentage deviation from the sensitivity at 500 W/m
2
due to the
change in irradiance with in th e r a nge of 100 W/m
2
to 1000 W/m2.
Non-linearity has an overlap with directional respons e, and
therefore should be handled with care in uncer ta inty evaluation.
ISO
9060­1990
Directional
response
the range of errors caused by a s su m in g that the normal incidence
sensitivity is valid for all di rections when measuring from any direction a beam ra dia tion whose normal incidence ir r adiance is 1000 W/m2 . Directional r esponse is a measure of the devia tions
from the ideal “cosine behaviour” and its az im uthal variation.
ISO
9060­1990
Spectral selectivity (350 to 1500 x 10
-9
m)
(WMO 300 to
3000 x 10-9 m)
percentage deviation of th e pr odu c t of s pec tr a l absorptance and spectral transmittance fr om the corresponding mean within 350 x 10
-9
m to 1500 x 10-9 m and the spectr al distribution of irradianc e.
Spectral selectivi ty is a measure of the spectral se lectivity of the
sensitivity.
ISO 9060­1990
Temperature response
(interval of 50 K)
percentage deviation of the sensitivity due to ch ange in ambient temperature within an in ter v a l of 50 K the temperature of the
pyranometer body.
ISO 9060-
1990
Tilt response
(0° to 90° at 1000 W/m
2
)
percentage deviation from the sensitivity at 0° t ilt (horizontal) due
to change in tilt from 0 ° to 90° at 1000 W/m
2
irradiance. Tilt
response describes changes of the sensitivity due to changes of
the tilt angle of the receiving s urface.
ISO
9060­1990
Sensitivity
the change in the response of a mea s uring instrument divided by the corresponding change in the stimulus.
WMO
1.6.3
Spectral range the spectral range of radiation to which the instrument is
sensitive. For a normal pyranometer this should be in the 0.3 to 3 x 10
-6
m range. Some pyranometers w ith c oloured glass domes
have a limited spectral range.
Hukseflux
LP02 manual v1606 44/47
9.9 Appendix on terminology / glossary
Table 9.9.1 Definitions and references of used terms
TERM DEFINITION (REFERENCE)
Solar energy or solar radiation
solar energy is the electroma gnetic energy emitted by the sun. Solar energy is also called solar radiation and shortwave radiation . The solar radiation inciden t on the top of the terrestrial atmosphere is called extra-terrestrial so lar radiation; 97 % of which is confined to the spectra l ra nge of 290 to 3 000 x 10
-9
m. Part of the extra-terrestrial solar r a diation penetrates the atmosphere and directly reaches the earth’s surf a ce, while part of it is scattered and / or absorbed by the gas molecules, aerosol particles, cloud dropl ets and cloud crystals in the atmosphere. The former is the direct c omponent, the latter is the diff use
component of the sola r r adiation. (ref: WMO, Hukseflux)
Hemispherical solar radiation
solar radiation received by a plane surface from a 180 ° field of view angle (solid angle of 2 π sr).(ref: ISO 9060)
Global solar radiation
the solar radiation received from a 180 ° field of view a ngle on a horizontal surface is refe r r e d to a s global radiation. Also called GHI. This includes radiation received directly from the s olid angle of the sun’s disc, as well as diffuse sky radiation that has been scattered in traversing the atmosphere. (ref: WMO) Hemispherical solar radiation received by a horizontal plane surface.
(ref: ISO 9060)
Plane-of-array irradiance
also POA: hemispherical solar irradiance in the plane of a PV array. (ref: ASTM E2848-11 / IEC 61724)
Direct solar radiation
radiation received from a small solid angle centre d on the sun’s disc, on a given plane. (ref: ISO 9060)
Terrestrial or Longwave radiation
radiation not of solar origin but of terrestrial and atmospheric origin and having longer wavelengths (3 000 to 100 00 0 x 10
-9
m). In case of down welling El ↓ also the background radiation from the universe is involved, passing through the ”atmospheric window”. In case of upwelling E
l
↑, composed of long -wave electromagnetic energy emitted by the earth’s surface and by the gases, aerosols and clouds of the a tm osphere; it is also partly absorbed within the atmosphere. For a temperature of 300 K, 99.99 % of the power of the terrestrial radia tion has a wavelength longer than 3 000 x 10
-9
m and about 99 per cent longer than
5 000 x 10
-9
m. For lower temperatures, th e s pec tr um shifts to longer
wavelengths. (ref: WMO)
World Radiometric Reference (WRR)
measurement standa r d r ep r esenting the Sl unit of irradian c e with an uncertain ty of less than ± 0.3 % (see the WMO Guide to Meteor ological Instruments and Methods of Observation, 1983, subclause 9.1.3). The reference wa s a dopted by the World Meteorological Organization (WMO ) and has been in effect since 1 July
1980. (ref: ISO 9060)
Albedo ratio of reflected and incoming solar radiation. Dimensionless nu m be r that varies
between 0 and 1. Typical albedo va lues are: < 0.1 for water, from 0.1 for wet
soils to 0.5 for dry sand, from 0.1 to 0.4 f or vegetation, up to 0.9 for fresh snow.
Angle of incidence
angle of radiation r elative to the sensor measured from normal incidence (varies from 0° to 90°).
Zenith angle angle of incidence of radiation, relative to ze nith. Equals angle of incidenc e for
horizontally mounted instruments
Azimuth angle
angle of incidence of radiation, projected in the plane of the sensor surface.
Varies from 0° to 360°. 0 is by definition the cable exit direction, also called north, east is + 90°. (ASTM G13-09)
Sunshine duration
sunshine duration during a given period is defined as the sum of that sub-period for which the direct solar irradia nce exceeds 120 W/m
2
. (ref: WMO)
LP02 manual v1606 45/47
9.10 EU declaration of conformity
We, Hukseflux Thermal Sensors B.V. Delftechpark 31 2628 XJ Delft The Netherlands
in accordance with the requirements of the following directive:
2014/30/EU The Electromagnetic Compatibility Directive
hereby declare under our sole responsibility that:
Product model: LP02 Product type: Pyranometer
has been designed to comply and is in conformity with the relevant sections and applicable requirements of the following standards:
Emission: EN 61326-1 (2006) Immunity: EN 61326-1 (2006) Emission: EN 61000-3-2 (2006) Emission: EN 61000-3-3 (1995) + A1 (2001) + A2 (2005). Report: 08C01340RPT01, 06 January 2009
Kees VAN DEN BOS Director Delft September 09, 2015
© 2016, Hukseflux Thermal Sensors B.V.
www.hukseflux.com
Hukseflux Thermal Sensors B.V. reserves the rig ht to change spec ifications without notice.
Loading...