Hukseflux IR20 User Manual

Page 1
Copyright by Hukseflux | manual v1604 | www.huksefluxusa.com | info@huksefluxusa.com
USER MANUAL IR20
Research grade pyrgeometer
Page 2
IR2 0 manual v 1604 2/39
Warning statements
Putting more than 12 Volt across the sensor wiring can lead to permanent damage to the sensor.
Do not use open circuit detection when measuring the sensor output.
Page 3
IR2 0 manual v 1604 3/39
Contents
Warning statements 2 Contents 3 List of symbols 4 Introduction 5 1 Ordering and checking at delivery 8
1.1 Ordering IR20 8
1.2 Included items 8
1.3 Quick instrument check 9
2 Instrument principle and theory 10
2.1 Pyrgeometer functionality 10
2.2 Solar and longwave radiation 10
2.3 IR20 pyrgeometer design 12
2.4 Typical measurement results 14
2.5 Optional heating 14
2.6 Optional shading 14
2.7 Use as a net radiation sensor 14
3 Specifications of IR20 and IR20WS 15
3.1 Specifications of IR20 and IR20WS 15
3.2 Dimensions of IR20 18
4 Standards and recommended practices for use 19
4.1 Site selection and installation 20
4.2 Installation of the sun screen 21
4.3 Electrical connection 22
4.4 Requirements for data acquisition / amplification 23
5 Making a dependable measurement 24
5.1 The concept of dependability 24
5.2 Reliability of the measurement 25
5.3 Speed of repair and maintenance 26
5.4 Uncertainty evaluation 26
6 Maintenance and trouble shooting 28
6.1 Recommended maintenance and quality assurance 28
6.2 Trouble shooting 29
6.3 Calibration and checks in the field 30
6.4 Data quality assurance 30
7 Appendices 31
7.1 Appendix on cable extension / replacement 31
7.2 Appendix on tools for IR20 32
7.3 Appendix on spare parts for IR20 32
7.4 Appendix on standards for classification and calibration 32
7.5 Appendix on calibration hierarchy 33
7.6 Appendix on meteorological radiation quantities 35
7.7 Appendix on terminology / glossary 36
7.8 EU declaration of conformity 38
Page 4
IR2 0 manual v 1604 4/39
List of symbols
Quantities Symbol Unit
Voltage output U V Sensitivity S V/(W/m2) Sensitivity at reference conditions S0 V/(W/m2) Temperature T °C Equivalent blackbody radiative temperature T °C Electrical resistance Re Longwave irradiance E W/m2 Stefan–Boltzmann constant (5.67 x 10-8) σ W/(m2∙K4)
temperature coefficient a 1/°C2 temperature coefficient b 1/°C temperature coefficient c -
(see also appendix 7.6 on meteorological quantities)
Subscripts
sky relating to the atmosphere surface relating to the ground surface ambient relating to ambient air body relating to the instrument body sensor relating to the sensor
Page 5
IR2 0 manual v 1604 5/39
Introduction
IR20 is a research grade pyrgeometer suitable for high-accuracy longwave irradiance measurement in meteorological applications. IR20 is capable of measuring during both day and night. In absence of solar radiation, model IR20WS offers even better accuracy because of its wider spectral range.
IR20 measures the longwave or far-infra-red radiation received by a plane surface, in W/m2, from a 180 ° field of view angle. In meteorological terms pyrgeometers are used to measure “downward and upward longwave irradiance” (WMO definition). Longwave radiation is the part of radiation that is not emitted by the sun. The spectral range of longwave radiation is not standardised. A practical cut-on is in the range of 4 to 5 x 10-6 m. IR20 has a dome with a solar blind filter with a cut-on at 4.5 x 10-6 m, making it suitable for day- and night observations.
Model IR20WS has a wide spectral range with a cut-on at 1.0 x 10-6 m. It offers a superior accuracy during night-time, when solar radiation is absent.
The main purpose of a pyrgeometer is to measure longwave radiation. As secondary measurands, the sky temperature T
sky
, and the equivalent surface temperature T
surface
can be measured. Both are so-called equivalent blackbody temperatures, i.e. temperatures calculated from pyrgeometer data, assuming the source behaves as a blackbody with an emission coefficient of 1.
Using IR20 is easy. It can be connected directly to commonly used data logging systems. The irradiance in W/m2 is calculated by dividing the IR20 output, a small voltage, by the
sensitivity and taking in account the irradiated heat by the sensor itself (Planck’s law).
The sensitivity is provided with IR20 on its calibration certificate. Please note that the IR20 sensitivity is corrected for temperature dependence in the measurement equation by using 3 additional constants. These coefficients are provided as well.
The central measurement equation governing IR20 is:
E = U/(S0·(a·T² + b·T + c)) + σ·(T + 273.15)4 (Formula 0.1)
S = S0·(a·T² + b·T + c) (Formula 0.2)
The instrument should be used in accordance with the recommended practices of WMO.
Suggested use for IR20 and IR20WS:
climatological networks extreme climates (polar / tropical) moving platforms (aircraft, buoys) uncertainty assessment (IR20 + IR20WS) calibration reference (IR20WS)
Page 6
IR2 0 manual v 1604 6/39
Distinguishing features and benefits of IR20 are:
correction of temperature dependence by use of the measurement function. This is
far more accurate than temperature compensation in the instrument, especially at very low and high temperatures. Every pyrgeometer is supplied with temperature coefficients to enter into the equation.
high sensitivity. With sufficient input signal a typical datalogger no longer significantly
contributes to the uncertainty of the measurement.
low thermal-resistance of the sensor. Competing designs need a significant correction
for the difference in temperature between pyrgeometer body and sensor surface. For IR20 this is not needed.
fast response time (3 s). A low response time is a benefit for measurements on
moving platforms such as aircraft and buoys.
on-board heater. Heating prevents condensation of water on the pyrgeometer dome
which, when occurring, leads to very large measurement errors.
instrument cut-on wavelength (5 %) and the two 50 % transmission points are
displayed on the product certificate for individual sensors.
Figure 0.1 IR20 research grade pyrgeometer with its sun screen removed
More about the instrument principle, theory and specifications can be found in the following chapters.
Page 7
IR2 0 manual v 1604 7/39
Pyrgeometers are not subject to a classification standard.
Calibration of pyrgeometers is usually traceable to the World Infrared Standard Group (WISG). This calibration takes into account the spectral properties of typical downward longwave radiation. As an option, calibration can be made traceable to a blackbody and the International Temperature Scale of 1990 (ITS-90). This alternative calibration is appropriate for measurements of upward longwave radiation (IR20 facing down).
See the specific paragraph in this manual about calibration and uncertainty assessment for more information.
This manual is intended for users of both IR20 and IR20WS. The specifications of IR20WS are identical to IR20’s except for its spectral range.
Figure 0.1 IR20WS research grade pyrgeometer
Page 8
IR2 0 manual v 1604 8/39
1 Ordering and checking at delivery
1.1 Ordering IR20
The standard configuration of IR20 is with 5 metres cable and a connector.
Common options are:
Longer cable (in multiples of 5 m). Specify total cable length. Five silica gel bags in an air-thight bag for IR20 desiccant holder. Specify order
number DC01.
Optional calibration to blackbody (ITS-90). IR20WS for the special wide spectrum model of IR20.
1.2 Included items
Arriving at the customer, the delivery should include:
pyrgeometer IR20 or IR20WS cable of the length as ordered with connector sun screen product certificate matching the instrument serial number calibration certificate matching the instrument serial number temperature dependence report any other options as ordered
Please store the certificates in a safe place.
Page 9
IR2 0 manual v 1604 9/39
1.3 Quick instrument check
A quick test of the instrument can be done by using a simple hand held multimeter and a thermal source.
1. Check the electrical resistance of the sensor between the green (-) and white (+) wire.
Use a multimeter at the 1000 Ω range. Measure the sensor resistance first with one
polarity, than reverse the polarity. Take the average value. The typical resistance of the wiring is 0.1 Ω/m. Typical resistance should be the typical sensor resistance of 300 to 500 Ω plus 1.5 Ω for the total resistance of two wires (back and forth) of each 5 m. Infinite resistance indicates a broken circuit; zero or a low resistance indicates a short circuit.
2. Check if the sensor reacts to heat: put the multimeter at its most sensitive range of DC voltage measurement, typically the 100 x 10-3 VDC range or lower. Make sure that the sensor is at 25 °C or lower. Expose the sensor to a heat source at a short distance from the window of more than 50 °C, for instance a heavy (> 5 kg) painted block of metal, or a painted metal container holding hot water. Face the side of the container to avoid condensation of water on the pyrgeometer window. Stir the water to attain homogeneity. A painted surface will act as a blackbody in the far-infra-red (FIR), irrespective of the visible colour. The signal should read positive and > 1 x 10-3 V now. In case of using your hand as a heat source, the signal should be significantly lower.
3. Remove the sun screen (see chapter on installation of the sun screen). Inspect the bubble level.
4. Check the electrical resistance of the thermistor. This should be in the 104 Ω range.
5. Check the electrical resistance of the heater. This should be in the 100 Ω range.
6. Inspect the instrument for any damage.
7. Inspect if the humidity indicator is blue. Blue indicates dryness. The colour pink indicates it is humid: in the latter case replace the desiccant (see chapter on maintenance).
Page 10
IR2 0 manual v 1604 10/39
2 Instrument principle and theory
2.1 Pyrgeometer functionality
IR20’s scientific name is pyrgeometer. IR20 measures the longwave or far-infra-red (FIR) radiation received by a plane surface, in W/m2, from a 180 ° field of view angle. In
meteorological terms pyrgeometers are used to measure “downward and upward longwave irradiance” (WMO definition).
As secondary measurands, the sky temperature T
sky
, and the equivalent surface (ground)
temperature T
surface
can be measured. Both are so-called equivalent blackbody radiative temperatures, i.e. temperatures calculated from the pyrgeometer measurement assuming these are uniform-temperature blackbodies with an emission coefficient of 1.
2.2 Solar and longwave radiation
Longwave radiation is the part of the radiation budget that is not emitted by the sun. The spectral range of the longwave radiation is not standardised. A practical cut-on is in the range of 4 to 5 x 10-6 m (see figure 2.2.1). In meteorology, solar- and longwave radiation are typically measured as separate parameters. The instrument to measure solar radiation is called pyranometer.
Figure 2.2.1 Atmospheric radiation as a function of wavelength plotted along two logarithmic axes to highlight the longwave radiation. Longwave radiation is mainly present in the 4 to 50 x 10-6 m range, whereas solar radiation is mainly present in the
0.3 to 3 x 10-6 m range. In practice, the two are measured separately using pyrgeometers and pyranometers
0.001
0.010
0.100
1.000
1 10 100
spectral irradiance x 10
-9
W/(m
2
/m)
wavelength x 10-6m
downwelling longwave solar
Page 11
IR2 0 manual v 1604 11/39
In the longwave spectrum, the sky can be seen as a temperature source; colder than ground level ambient air temperature, with its lowest temperatures at zenith, getting warmer (closer to ambient air temperature) at the horizon. The uniformity of this longwave source is much better than that in the range of the solar spectrum, where the sun is a dominant contributor. The “equivalent blackbody” temperature, as a function of zenith angle, roughly follows the same pattern independent of the exact sky condition (cloudy or clear). This explains why for pyrgeometers the directional response is not very critical.
The downwelling longwave radiation essentially consists of several components:
1. low temperature radiation from the universe, filtered by the atmosphere. The atmosphere is transparent for this radiation in the so-called atmospheric window (roughly the 10 to 15 x 10-6 m wavelength range).
2. higher temperature radiation emitted by atmospheric gasses and aerosols.
3. in presence of clouds or mist, the low temperature radiation from the universe is almost completely blocked by the water droplets. The pyrgeometer then receives the radiation emitted by the water droplets.
The spectral distribution of longwave irradiance varies significantly as a function of the source composition. A pyrgeometer is relatively insensitive to these variations, but all the same blackbody calibration tends to differ from WISG calibration by up to 5 %. In addition there may be effects that are uncompensated for in the calibration (for instance related to atmospheric water vapour content) in the 5 to 10 W/m2 range. These effects are still under investigation by the international scientific community. Comparison between IR20 and IR20WS may serve to investigate this effect.
Upwelling longwave irradiance is measured with downfacing instruments. These are presumably looking directly at the surface (absorption and emission of the atmosphere is low over a short distance of around 2 m), which behaves like a normal blackbody. Hukseflux suggests calibrating downfacing instruments against a blackbody rather than having WISG as a reference.
Page 12
IR2 0 manual v 1604 12/39
2.3 IR20 pyrgeometer design
Figure 2.3.1 Overview of IR20 pyrgeometer:
(1) cable (standard length 5 metres, optional longer cable) (2) fixation of sun screen (3) dome with solar blind filter (4) sun screen (5) humidity indicator (6) desiccant holder (7) levelling feet (8) bubble level (9) connector
By definition a pyrgeometer should not measure solar radiation, and in the longwave
have a spectral selectivity that is as “flat” as possible.
In an irradiance measurement by definition the response to “beam” radiation varies with
the cosine of the angle of incidence; i.e. it should have full response when the radiation hits the sensor perpendicularly (normal to the surface, 0 ° angle of incidence), zero response when the source is at the horizon (90 ° angle of incidence, 90 ° zenith angle), and 50 % of full response at 60 ° angle of incidence.
A pyrgeometer should have a so-called “directional response” (older documents mention “cosine response”) that is as close as possible to the ideal cosine characteristic.
1
2
3
4
5
6
7
8
9
Page 13
IR2 0 manual v 1604 13/39
In order to attain the proper directional and spectral characteristics, a pyrgeometer’s main components are:
a thermal sensor with black coating. It has a flat spectrum covering the
0.3 to 50 x 10-6 m range, and has a near-perfect directional response. The coating absorbs all longwave radiation and, at the moment of absorption, converts it to heat. The heat flows through the sensor to the sensor body. The thermopile sensor generates a voltage output signal that is proportional to the irradiance exchange between sensor and source. The sensor not only absorbs, but also irradiates heat as a blackbody.
a silicon dome. This dome limits the spectral range from 1.0 to 40 x 10
-6
m (cutting off the part below 1.0 x 10-6 m), while preserving the 180 ° field of view angle. Another function of the dome is that it shields the thermopile sensor from the environment (convection, rain).
a solar blind interference coating deposited on the dome (not for model IR20WS):
this coating limits the spectral range. It now becomes 4.5 to 40 x 10-6 m (cutting off the part below 4.5 x 10-6 m).
Pyrgeometers can be manufactured to different specifications and with different levels of verification and characterisation during production. Hukseflux also manufactures lower accuracy pyrgeometers; see our pyrgeometer model IR02.
Model IR20 has a dome with a solar blind filter with a cut-on at 4.5 x 10-6 m, making it suitable for day- and night observations. Model IR20WS has a wide spectral range with a cut-on at 1.0 x 10-6 m. It offers a superior accuracy under night-time conditions, when solar radiation is absent. See also the appendix on uncertainty evaluation.
Page 14
IR2 0 manual v 1604 14/39
2.4 Typical measurement results
Please note that the signal generated by an upfacing pyrgeometer usually has a negative sign. The most important factors determining downward longwave irradiance are:
ambient air temperature sky condition / cloud cover atmospheric moisture content
Table 2.4.1 Expected pyrgeometer output U/S at different ambient air temperatures, T
ambient
, and at different cloud conditions. Under clear sky conditions the U/S is around
-100 W/m2 while under cloudy conditions it will be close to 0 W/m2. Also calculated: the sky temperature, T
sky
, and the longwave downward irradiance E.
EXPECTED PYRGEOMETER OUTPUT CONDITIONS
T
ambient
Sky condition
U/(S0·(a·T² + b·T + c))
T
sky
E
[°C]
[cloudy], [clear]
[W/m2]
[°C]
[W/m2]
-20
cloudy
0
-20
232
-20
clear sky
-100
-53
132 0 cloudy
0
+0
314 0 clear sky
-100
-24
214
+30
cloudy
0
+30
477
+30
clear sky
-100
+12
377
2.5 Optional heating
A low-power heater is located in the body of the pyrgeometer. The heater is not necessarily switched on; recommended operation is to activate the heater when there is a risk of dew deposition.
2.6 Optional shading
One of the larger errors in the daytime measurement of downwelling longwave irradiance is the offset caused by solar radiation; the “solar offset”. Errors due to solar offset, are of the order of + 10 W/m2 at 1000 W/m2 global horizontal irradiance. For ultra high accuracy measurements this offset can be reduced by around 60 % by shading, which means preventing the direct radiation to reach the instrument. Shading is typically done by using a shading disk on a solar tracker.
2.7 Use as a net radiation sensor
Two pyrgeometers mounted back to back may be used to measure net longwave radiation. Net longwave radiation is defined as downwelling minus upwelling longwave irradiance. In case the two instruments are thermally coupled, the body temperatures of the instruments are identical. In that case the body temperature cancels from the equation for the net radiation. However for calculation of sky temperature and surface temperature the instrument temperature still needs to be measured.
Page 15
IR2 0 manual v 1604 15/39
3 Specifications of IR20 and IR20WS
3.1 Specifications of IR20 and IR20WS
IR20 research grade pyrgeometer measures the longwave irradiance received by a plane surface, in W/m2, from a 180 ° field of view angle. In meteorological terms IR20 measures downward and upward longwave irradiance. Working completely passive, using a thermopile sensor, IR20 generates a small output voltage proportional to the radiation balance between the instrument and the source it faces. It can only be used in combination with a suitable measurement system. The instrument is not subject to classification. It should be used in accordance with the recommended practices of WMO. IR20 measures during both day and night. In the absence of solar radiation IR20WS offers a higher accuracy because of its wider spectral range. For ultra-high accuracy measurements the user should consider to use the incorporated heater and should consider shading the instrument during daytime.
Table 3.1.1 Specifications of IR20 and IR20WS (continued on next pages)
IR20 & IR20WS SPECIFICATIONS
MEASURANDS
Measurand
longwave radiation
Measurand in SI radiometry units
longwave irradiance in W/m2
Optional measurand
sky temperature
Optional measurand
surface temperature
IR20 VERSUS IR20WS: SPECTRAL RANGE & USE
Spectral range IR20
4.5 to 40 x 10-6 m (nominal, see product certificate for individual value)
Spectral range IR20WS
1.0 to 50 x 10-6 m (based on typical material properties only)
IR20WS restrictions for use
only in the absence of solar radiation
Solar offset (IR20 only, not specified for IR20WS)
< 10 W/m2 (at 1000 W/m2 global horizontal irradiance on the dome)
MAIN SPECIFICATIONS Field of view angle
180 °
Response time (95 %)
3 s
Sensitivity (nominal)
17 x 10-6 V/(W/m2)
Sensitivity range
10 to 25 x 10-6 V/(W/m2)
Rated operating temperature range
-40 to +80 °C
Temperature dependence
< ± 0.4 % (-30 to +50 °C) using the measurement function
Temperature sensor
10 kΩ thermistor
Required sensor power
zero (passive sensor)
Heater
12 VDC, 1.5 W (see below for details)
Standard cable length
5 m
Page 16
IR2 0 manual v 1604 16/39
Table 3.1.1 Specifications of IR20 and IR20WS (continued)
ADDITIONAL SPECIFICATIONS
Zero offset b (response to 5 K/h change in ambient temperature)
< ± 2 W/m2 Non-stability
< ± 1 % change per year
Non-linearity
< ± 0.5 % (100 to 300 W/m2, relative to 200 W/m2 sensor to source exchange)
Measurement range
-1000 to +1000 W/m2 (sensor to source exchange: U/(S0·(a·T² + b·T + c)) )
Tilt dependence
< ± 0.5 % (0 to 90 ° at 300 W/m2)
Sensor resistance range
300 to 500 Ω
Expected voltage output
Application for outdoor measurement of downward longwave irradiance: -7.5 to 7.5 x 10-3 V
Measurement function / required programming
E = U/(S0·(a·T² + b·T + c)) + σ·(T + 273.15)4
Measurement function / optional programming for sky temperature
T
sky
= (El↓/σ)
1/4
- 273.15
Measurement function / optional programming for surface temperature
T
surface
= (El↑/σ)
1/4
- 273.15
Required readout
1 differential voltage channel or 1 single ended voltage channel, input resistance > 106 1 temperature channel
STANDARDS
Standard governing use of the instrument
WMO-No. 8, Guide to Meteorological Instruments and Methods of Observation, seventh edition 2008, paragraph 7.4 "measurement of total and long-wave radiation"
MOUNTING, CABLING, TRANSPORT
Chassis connector
M16 panel connector, male thread, 10-pole
Chassis connector type
HUMMEL AG 7.840.200.000 panel connector, front mounting, short version
Cable connector
M16 straight connector, female thread, 10-pole
Cable connector type
HUMMEL AG 7.810.300.00M straight connector, female thread, for cable diameter 3 to 6 x 10-3 m, special version
Connector protection class
IP67 / IP69 K per EN 60 529 (connected)
Cable diameter
5.3 x 10-3 m
Cable replacement
replacement cables with connector can be ordered separately from Hukseflux
Instrument mounting
2 x M5 bolt at 65 x 10-3 m centre-to-centre distance on north-south axis, or 1 x M6 bolt at the centre of the instrument, connection from below under the bottom plate of the instrument
Levelling
bubble level and adjustable levelling feet are included
Levelling accuracy
< 0.1 ° bubble entirely in ring
Desiccant
two bags of silica gel, 0.5 g, 35 x 20 mm
Humidity indicator
blue when dry, pink when humid
IP protection class
IP67
Gross weight including 5 m cable
1.2 kg
Net weight including 5 m cable
0.85 kg
Packaging
box of 200 x 135 x 225 mm
Page 17
IR2 0 manual v 1604 17/39
Table 3.1.1 Specifications of IR20 and IR20WS (started on previous pages)
HEATING
Heater operation
the heater is not necessarily switched on; recommended operation is to activate the heater when there is a risk of dew deposition
Required heater power
1.5 W at 12 VDC
Heater resistance
95 Ω
Steady state zero offset caused by heating
0 W/m2
CALIBRATION Calibration traceability
to WISG
Optional traceability
to blackbody (ITS-90 )
Calibration hierarchy
from WISG through Hukseflux internal calibration procedure involving outdoor comparision to a reference pyrgeometer
Calibration method
outdoor comparison to a reference pyrgeometer
Calibration uncertainty
< 6 % (k = 2)
Recommended recalibration interval
2 years
Reference conditions
horizontal mounting, atmospheric longwave irradiance, clear sky nights, 20 °C
Validity of calibration
based on experience the instrument sensitivity will not change during storage. During use under exposure to solar radiation the instrument “non-stability” specification is applicable. Hukseflux recommends ITS-90 traceable calibration for upward longwave irradiance measurement.
Characterisation of the dependence of sensitivity to temperature
temperature coefficients a, b and c of the measurement equation are determined in an independent experiment, and reported on the product certificate
MEASUREMENT ACCURACY
Uncertainty of the measurement
statements about the overall measurement uncertainty can only be made on an individual basis. See the chapter on uncertainty evaluation
Achievable uncertainty (95 % confidence level) daily totals
± 8 % (Hukseflux own estimate)
VERSIONS / OPTIONS
Longer cable, in multiples of 5 m
option code = total cable length
Calibration
optional to blackbody (ITS-90 )
ACCESSORIES
Bags of silica gel for desiccant
set of 5 bags in an air tight bag option code = DC01
Page 18
IR2 0 manual v 1604 18/39
3.2 Dimensions of IR20
Figure 3.2.1 Dimensions of IR20 and IR20WS in 10-3 m
68
Ø 150
M6M5 (2x)
65
Page 19
IR2 0 manual v 1604 19/39
4 Standards and recommended practices
for use
Pyrgeometers are not subject to standardisation.
The World Meteorological Organization (WMO) is a specialised agency of the United Nations. It is the UN system's authoritative voice on the state and behaviour of the earth's atmosphere and climate. WMO publishes WMO-No. 8; Guide to Meteorological Instruments and Methods of Observation, in which paragraph 7.4 covers "measurement of total and long-wave radiation".
For ultra high accuracy measurements, the following manual may serve as a reference: Baseline Surface Radiation Network (BSRN) Operations Manual, Version 2.1, L. J. B. McArthur, April 2005, WCRP-121, WMO/TD-No. 1274. This manual also includes chapters on installation and calibration.
Page 20
IR2 0 manual v 1604 20/39
4.1 Site selection and installation
Table 4.1.1 Recommendations for installation of pyrgeometers
Location
the horizon should be as free from obstacles as possible. Ideally there should be sources of longwave irradiance between the course of the sun and the instrument, only free sky.
Mechanical mounting / thermal insulation
preferably use connection by bolts to the bottom plate of the instrument. A pyrgeometer is sensitive to thermal shocks. Do not mount the instrument with the body in direct thermal contact to the mounting plate (so always use the levelling feet also if the mounting is not horizontal), do not mount the instrument on objects that become very hot (black coated metal plates).
Instrument mounting with 2 bolts
2 x M5 bolt at 65 x 10-3 m centre to centre distance on north-south axis, connection from below under the bottom plate of the instrument.
Instrument mounting with one bolt
1 x M6 bolt at the centre of the instrument, connection from below under the bottom plate of the instrument.
Performing a representative measurement
the pyrgeometer measures the solar radiation in the plane of the sensor. This may require installation in a tilted or inverted position. The sensor surface (sensor bottom plate) should be mounted parallel to the plane of interest. In case a pyrgeometer is not mounted horizontally or in case the horizon is obstructed, the representativeness of the location becomes an important element of the measurement. See the chapter on uncertainty evaluation.
Levelling
in case of horizontal mounting only use the bubble level and levelling feet. For inspection of the bubble level the sun screen must be removed.
Instrument orientation
by convention with the cable exit pointing to the nearest pole (so the cable exit should point north in the northern hemisphere, south in the southern hemisphere).
Installation height
in case of inverted installation, WMO recommends a distance of 1.5 m between soil surface and sensor (reducing the effect of shadows and in order to obtain good spatial averaging).
Optional shading
for ultra-high accuracy measurements the solar offset can be reduced by around 60 % by “shading”, which means preventing the direct radiation to reach the instrument. Shading is typically done by using a shading disk on a solar tracker.
Page 21
IR2 0 manual v 1604 21/39
4.2 Installation of the sun screen
IR20’s sun screen can be installed and removed by using the dedicated thumb screw. See item 2 of the drawing below. The thumb screw can be turned without tools for fixation or loosening of the sun screen, as visualised below. Once the thumb screw has turned the sun screen loose, the screen can be lifted off manually. After removal the user may inspect the bubble level, item 8 of the drawing, and remove the cable / connector, item 9.
Figure 4.2.1 Installation and removal of IR20’s sun screen
1
2
3
5
6
7
8
9
4
Page 22
IR2 0 manual v 1604 22/39
4.3 Electrical connection
In order to operate, a pyrgeometer should be connected to a measurement system, typically a so-called datalogger. IR20 is a passive sensor that does not need any power. Cables generally act as a source of distortion, by picking up capacitive noise. We recommend keeping the distance between a datalogger or amplifier and the sensor as short as possible. For cable extension, see the appendix on this subject.
Table 4.3.1 The electrical connection of IR20. The heater is not necessarily used. The temperature sensor must be used.
PIN
WIRE
IR20
2
Red
10 kΩ thermistor [+]
3
Pink
10 kΩ thermistor [+]
6
Blue
10 kΩ thermistor [−]
8
Grey
10 kΩ thermistor [−]
1
Brown
heater
4
Yellow
heater
9
Black
ground
7
White
signal [+]
5
Green
signal [−]
Note 1: 10 kΩ thermistors are usually connected in a 2-wire configuration.
Note 2: the heater is not necessarily connected. In case it is connected, the polarity of the connection is not important. Note 3: signal wires are insulated from ground wire and from the sensor body. Insulation resistance is tested during production and larger than 1 x 106 Ω. Note 4: ground is connected to the connector, the sensor body and the shield of the wire.
Figure 4.3.1 Electrical diagram of the internal wiring of IR20. The shield is connected to the sensor body.
Page 23
IR2 0 manual v 1604 23/39
4.4 Requirements for data acquisition / amplification
The selection and programming of dataloggers is the responsibility of the user. Please contact the supplier of the data acquisition and amplification equipment to see if directions for use with the IR20 are available. In case programming for similar instruments is available, this can typically also be used. IR20 can usually be treated in the same way as other thermopile pyrgeometers.
Table 4.4.1 Requirements for data acquisition and amplification equipment for IR20 in the standard configuration
Capability to measure small voltage signals
preferably: better than 5 x 10-6 V uncertainty Minimum requirement: 20 x 10-6 V uncertainty (valid for the entire expected temperature range of the acquisition / amplification equipment)
Capability for the data logger or the software
to store data, and to perform division by the sensitivity to calculate the longwave irradiance. E = U/(S0·(a·T² + b·T + c)) + σ·(T + 273.15)4 (Formula 0.1) (see also optional measurands)
Data acquisition input resistance
> 1 x 106
Open circuit detection (WARNING)
open-circuit detection should not be used, unless this is done separately from the normal measurement by more than 5 times the sensor response time and with a small current only. Thermopile sensors are sensitive to the current that is used during open circuit detection. The current will generate heat, which is measured and will appear as an offset.
Capability to measure temperature
a 10 kΩ thermistor must be read-out. Required accuracy of the readout is ± 0.2 °C, which results in around 1 W/m2 uncertainty of the irradiance measurement.
Capability to power the heater (OPTIONAL)
IR20 has a 12 VDC, 1.5 W heater on board, which may optionally be activated to keep the instrument above dew point. Some users prefer to have the heater on full time, others prefer to switch it on during nighttime only.
Page 24
IR2 0 manual v 1604 24/39
5 Making a dependable measurement
5.1 The concept of dependability
A measurement with a pyrgeometer is called “dependable” if it is reliable, i.e. measuring within required uncertainty limits, for most of the time and if problems, once they occur, can be solved quickly.
The requirements for a measurement with a pyrgeometer may be expressed by the user as:
required uncertainty of the measurement (see following paragraphs) requirements for maintenance and repairs (possibilities for maintenance and repair
including effort to be made and processing time)
a requirement to the expected instrument lifetime (until it is no longer feasible to
repair)
It is important to realise that the uncertainty of the measurement is not only determined by the instrument but also by the way it is used.
In case of pyrgeometers, the measurement uncertainty as obtained during outdoor measurements is a function of:
• the instrument properties
• the calibration procedure / uncertainty
• the presence of natural sunlight (involving the instrument specification of solar offset)
the measurement conditions (such as tilting, ventilation, shading, heating, instrument
temperature)
• maintenance (mainly fouling and deposition of water)
• the environmental conditions* (such as temperature, position of the sun, presence of
clouds, horizon, representativeness of the location)
Therefore statements about the overall measurement uncertainty under outdoor conditions can only be made on an individual basis, taking all these factors into account.
* defined at Hukseflux as all factors outside the instrument that are relevant to the measurement such as the cloud cover (presence or absence of direct radiation), sun position, the local horizon (which may be obstructed) or condition of the ground (when inverted). The environmental conditions also involve the question whether or not the measurement at the location of measurement is representative of the quantity that should be measured.
Page 25
IR2 0 manual v 1604 25/39
5.2 Reliability of the measurement
A measurement is reliable if it measures within required uncertainty limits for most of the time. We distinguish between two causes of unreliability of the measurement:
related to the reliability of the pyrgeometer and its design, manufacturing, calibration
(hardware reliability).
related to the reliability of the measurement uncertainty (measurement reliability),
which involves hardware reliability as well as condition of use.
Most of the hardware reliability is the responsibility of the instrument manufacturer. The reliability of the measurement however is a joint responsibility of instrument manufacturer and user. As a function of user requirements, taking into account measurement conditions and environmental conditions, the user will select an instrument of a certain class, and define maintenance support procedures.
In many situations there is a limit to a realistically attainable accuracy level. This is due to conditions that are beyond control once the measurement system is in place. Typical limiting conditions are:
the measurement conditions, for instance when working at extreme temperatures
when the instrument temperature is at the extreme limits of the rated temperature range.
the environmental conditions, for instance when installed at a sub-optimal
measurement location with obstacles in the field of view.
the environmental conditions, for instance when assessing net radiation, the
downfacing pyrgeometer measurement may not be representative of irradiance received in that particular area.
The measurement reliability can be improved by maintenance support. Important aspects are:
dome fouling by deposition of dust, dew, rain or snow. With pyrgeometers the most
important source of unreliability is deposition of water on the dome. Water completely blocks the longwave radiation flux between sensor and sky. In particular at clear nights this causes very large errors. Water deposition under clear-sky nighttime conditions can largely be prevented by using the instrument heater. Fouling results in undefined measurement uncertainty (sensitivity and directional error are no longer defined). This should be solved by regular inspection and cleaning.
sensor instability. Maximum expected sensor aging is specified per instrument as its
non-stability in [% change / year]. In case the sensor is not recalibrated, the uncertainty of the sensitivity gradually will increase. This is solved by regular recalibration.
moisture condensing under pyrgeometer domes resulting in a slow change of
sensitivity (within specifications). This is solved by regular replacement of desiccant or by maintenance (drying the entire sensor) in case the sensor allows this. For non­serviceable sensors like Hukseflux flat window pyrgeometers (for example model
Page 26
IR2 0 manual v 1604 26/39
IR02), this may slowly develop into a defect. For model IR20 extra desiccant (in a set of 5 bags in an air tight bag) is available.
One of the larger errors in the daytime measurement of downwelling longwave
irradiance is the offset caused by solar radiation; the “solar offset”. Errors due to solar offset, are of the order of + 10 W/m2 at 1000 W/m2 global horizontal irradiance. For ultra-high accuracy measurements this offset can be reduced by around 60% by
“shading”, which means preventing the direct radiation to reach the instrument.
Shading is typically done by using a shading disk on a solar tracker.
Another way to improve measurement reliability is to introduce redundant sensors.
The use of redundant instruments allows remote checks of one instrument using the
other as a reference, which leads to a higher measurement reliability.
5.3 Speed of repair and maintenance
Dependability is not only a matter of reliability but also involves the reaction to problems; if the processing time of service and repairs is short, this contributes to the dependability.
Hukseflux pyrgeometers are designed to allow easy maintenance and repair. The main maintenance actions are:
replacement of desiccant replacement of cabling
For optimisation of dependability a user should:
estimate the expected lifetime of the instrument design a schedule of regular maintenance design a schedule of repair or replacement in case of defects
When operating multiple instruments in a network Hukseflux recommends keeping procedures simple and having a few spare instruments to act as replacements during service, recalibrations and repair.
5.4 Uncertainty evaluation
The uncertainty of a measurement under outdoor or indoor conditions depends on many factors, see paragraph 1 of this chapter. It is not possible to give one figure for pyrgeometer measurement uncertainty. The work on uncertainty evaluation is “in progress”. There are several groups around the world participating in standardisation of the method of calculation. The effort aims to work according to the guidelines for
uncertainty evaluation (according to the “Guide to Expression of Uncertainty in Measurement” or GUM).
Page 27
IR2 0 manual v 1604 27/39
The main ingredients of the uncertainty evaluation for pyrgeometers are:
Calibration uncertainty, which is in the order of ± 6 % (k = 2) for upfacing
instruments measuring downward longwave irradiance
Calibration uncertainty, which is larger for other than upfacing instruments; for
downfacing instruments a blackbody calibration seems preferable. Blackbody calibration will result in a lower sensitivity (S0) than WISG traceable calibraton.
Errors due to water deposition at clear nights; these completely block the longwave
irradiance exchange between pyrgeometer and may cause the signal U/(S0·(a·T² + b·T + c)) to change from a large negative value (-100 W/m2) to around 0 W/m2 . Water deposition at clear nights may largely be avoided by using the on-board heater of IR20.
Errors due to solar offset, which is of the order of + 10 W/m
2
at 1000 W/m2 global horizontal irradiance. This offset can be reduced by around 60% by shading of the instrument, typically by using a shading disk on a solar tracker. This error is partially caused by heating of the dome, partially by transmission of solar radiation by the dome / filter combination. This uncertainty is not taken into account in the WISG calibration of the reference instrument. With IR20WS, only measuring at night, this uncertainty does not play a significant role.
Errors due to the choice of the cut-on wavelength of the pyrgeometer. Depending on
the atmospheric water content, the pyrgeometer will block a variable percentage of the downward longwave irradiance. This causes an uncertainty of the sensitivity S0. With IR20, this uncertainty is already taken into account in the WISG calibration of the reference instrument. With IR20WS this uncertainty does not play a significant role.
Errors due to instrument non-stability. This is now estimated at < ± 1% change per
year. The main factor in instrument non-stability is the aging of the pyrgeometer solar blind filter. For IR20WS this filter is not present. For that reason we expect IR20WS to have a better non-stability.
Errors due to the temperature measurement T. For this a 10 kΩ thermistor must be
read-out. Required accuracy of the readout is ± 0.2 °C, which results in around 1 W/m2 uncertainty of the irradiance measurement. To this the uncertainty of the thermistor itself should be added. In measurement of net radiation, in case the upfacing and downfacing instruments are thermally coupled, the temperature measurement (and also its uncertainty) cancel from the equation.
Page 28
IR2 0 manual v 1604 28/39
6 Maintenance and trouble shooting
6.1 Recommended maintenance and quality assurance
IR20 can measure reliably at a low level of maintenance in most locations. Usually unreliable measurements will be detected as unreasonably large or small measured values. As a general rule this means that regular visual inspection combined with a critical review of the measured data, preferably checking against other measurements, is the preferred way to obtain a reliable measurement.
Table 6.1.1 Recommended maintenance of IR20. If possible the data analysis and cleaning (1 and 2) should be done on a daily basis.
MINIMUM RECOMMENDED PYRGEOMETER MAINTENANCE
INTERVAL
SUBJECT
ACTION
1
1 week
data analysis
compare measured data to maximum possible / maximum expected irradiance and to other measurements nearby (redundant instruments). Also historical seasonal records can be used as a source for expected values. Look for any patterns and events that deviate from what is normal or expected
2
2 weeks
cleaning
use a soft cloth to clean the dome of the instrument, persistent stains can be treated with soapy water or alcohol
3
6 months
inspection
inspect cable quality, inspect cable glands, inspect mounting position, inspect cable, clean instrument, clean cable, inspect levelling, change instrument tilt in case this is out of specification, inspect mounting connection
4 desiccant replacement
desiccant replacement (if applicable). Change in case the humidity indicator shows more than 50 %, then replace desiccant. Coat the rubber of the cartridge with silicone grease or vaseline. Desiccant regeneration: heating in an oven at 70 °C for 1 to 2 hours. Humidity indicator regeneration: heating until blue at 70 °C
5
2 years
recalibration
recalibration by side-by-side comparison to a higher standard instrument in the field
6 lifetime assessment
judge if the instrument should be reliable for another 2 years, or if it should be replaced
7
6 years
parts replacement
if applicable / necessary replace the parts that are most exposed to weathering; cable, cable gland, desiccant holder, sun screen. NOTE: use Hukseflux approved parts only
8 internal inspection
if applicable: open instrument and inspect / replace O-rings; dry internal cavity around the circuit board
9 recalibration
recalibration by side-by-side comparison to a higher standard instrument at the manufacturer or a reference institute. Also recalibrate the temperature sensor
Page 29
IR2 0 manual v 1604 29/39
6.2 Trouble shooting
Table 6.2.1 Trouble shooting for IR20
The sensor does not give any signal
Check the electrical resistance of the sensor between the green (-) and white
(+) wire. Use a multimeter at the 1000 Ω range. Measure the sensor
resistance first with one polarity, than reverse the polarity. Take the average
value. The typical resistance of the wiring is 0.1 Ω/m. Typical resistance should be the typical sensor resistance of 100 to 200 Ω plus 1.5 Ω for the
total resistance of two wires (back and forth) of each 5 m. Infinite resistance indicates a broken circuit; zero or a low resistance indicates a short circuit. Check if the sensor reacts to heat: put the multimeter at its most sensitive range of DC voltage measurement, typically the 100 x 10-3 VDC range or lower. Make sure that the sensor is at 25 °C or lower. Expose the sensor to a heat source at a short distance from the window of more than 50 °C, for instance a heavy (> 5 kg) painted block of metal, or a painted metal container holding hot water. Face the side of the container to avoid condensation of water on the pyrgeometer window. Stir the water to attain homogeneity. A painted surface will act as a blackbody in the far-infra-red (FIR), irrespective of the visible colour. The signal should read positive and > 1 x 10-3 V now. In case of using your hand as a heat source, the signal should be significantly lower.Check the data acquisition by applying a 1 x 10-6 V source to it in the 1 x 10-6 V range.
The sensor signal is unrealistically high or low.
Check if the measurement function, including the constant a, b and c has been implemented properly. Please note that each sensor has its own individual calibration factor and constants, as documented in its production certificate. Check if the pyrgeometer has a clean dome. Check the location of the pyrgeometer; are there any obstructions / sources that could explain the measurement result. Check the condition of the wiring at the logger. Check the cable condition looking for cable breaks. Check the range of the data logger; signal is usually negative (this could be out of range) or the amplitude could be out of range. Check the data acquisition by applying a 1 x 10-6 V source to it in the 1 x 10-6 V range. Look at the output. Check if the output is as expected. Check the data acquisition by short circuiting the data acquisition input with a 100 to 1000 Ω resistor. Look at the output. Check if it is close to 0 W/m2.
The sensor signal shows unexpected variations
Check the presence of strong sources of electromagnetic radiation (radar, radio etc.) Check the condition of the shielding. Check the condition of the sensor cable. Check if the cable is not moving during the measurement
The instrument shows internal condensation.
Replace the desiccant and wait a few days to see if the situation improves. In case of condensation of droplets: disassemble the instrument and dry out the parts.
The instrument shows persistent internal condensation
Arrange to send the sensor back to Hukseflux for diagnosis.
Page 30
IR2 0 manual v 1604 30/39
6.3 Calibration and checks in the field
Recalibration of field pyrgeometers is typically done by comparison in the field to a reference pyrgeometer. There is no standard for this procedure.
Hukseflux recommendation for re-calibration: if possible, perform calibration outdoor by comparison to an identical or a higher class reference instrument, under nighttime as well as daytime conditions. Use nighttime data only to determine S0. Do not change the constants a, b and c.
Hukseflux main recommendations for field intercomparisons are:
1) perform field calibration during several days; 2 to 3 days and if possible under cloudless conditions.
2) to take a reference of the same brand and type as the field pyrgeometer or a pyrgeometer of a higher class, and
3) to connect both to the same electronics, so that electronics errors (also offsets) are eliminated.
4) to mount all instruments on the same platform, so that they have the same body temperature.
5) to analyse downward irradiance values at nighttime only to determine S0.
6) to analyse the daytime data, independently, and look at the residuals between the calibration reference and calibrated instrument as a function of solar irradiance. The solar offset can serve as a quality indicator of the pyrgeometer filter condition.
6.4 Data quality assurance
Quality assurance can be done by:
analysing trends in longwave irradiance signal plotting the measured irradiance against mathematically generated expected values comparing irradiance measurements between sites analysis of daytime signals against solar irradiance
The main idea is that one should look out for any unrealistic values. There are programs on the market that can semi-automatically perform data screening:
See for more information on such a program: http://www.dqms.com
Page 31
IR2 0 manual v 1604 31/39
7 Appendices
7.1 Appendix on cable extension / replacement
The sensor cable of IR20 is equipped with a M16 straight connector. In case of cable replacement, it is recommended to purchase a new cable with connector at Hukseflux. An alternative is to choose for a Do-it-yourself (DIY) approach; please ask for the DIY connector assembly guide. In case of cable extension, the user may choose purchasing a new cable with connector at Hukseflux or extending the existing cable himself. Please note that Hukseflux does not provide support for DIY connector- and cable assembly. IR20 is equipped with one cable. Keep the distance between data logger or amplifier and sensor as short as possible. Cables act as a source of distortion by picking up capacitive noise. In an electrically “quiet” environment the IR20 cable can be extended without problem to 100 metres. If done properly, the sensor signal, although small, will not significantly degrade because the sensor resistance is very low (so good immunity to external sources) and because there is no current flowing (so no resistive losses). Connector, cable and cable connection specifications are summarised below.
Table 7.1.1 Preferred specifications for IR20 cable replacement and extension
General replacement
please order a new cable with connector at Hukseflux or choose for a DIY approach. In case of DIY replacement by the user see connector specifications below and ask for the DIY connector assembly guide
General cable extension
please order a new cable with connector at Hukseflux or solder the new cable conductors and shield to the original sensor cable and make a connection, using adhesive-lined heat shrink tubing, with specifications for outdoor use. Always connect shield
Connectors used
chassis: M16 panel connector, male thread, 10-pole, HUMMEL AG
7.840.200.000 panel connector, front mounting, short version. cable: M16 straight connector, female thread, 10-pole. HUMMEL AG
7.810.300.00M straight connector, female thread, for cable 3 to 6 x 10-3 m, special version
Cable
8-wire, shielded, with copper conductors (at Hukseflux 8-wire shielded cable is used, of which 2 wires are used for signal transmission, 2 for heating and 2 to 4 for the temperature sensor)
Conductor resistance
< 0.1 /m
Length
cables should be kept as short as possible, in any case the total cable length should be less than 100 m
Outer sheath
with specifications for outdoor use (for good stability in outdoor applications)
Page 32
IR2 0 manual v 1604 32/39
7.2 Appendix on tools for IR20
Table 7.2.1 Specifications of tools for IR20
tooling required for sun screen fixation and removal
by hand
tooling required for bottom plate fixation and removal
hex key 2.5 x 10-3 m
tooling required for desiccant holder fixation and removal
spanner size 20 x 10-3 m
tooling required for wire fixation and removal (internal wiring inside IR20 body)
screwdriver blade width 2 x 10-3 m
7.3 Appendix on spare parts for IR20
Desiccant holder (with glass window and rubber ring) Desiccant (set of 5 bags in an air tight bag) Humidity indicator Levelling feet (set of 2) Static foot Sun screen with metal ring and thumb screw IR20 cable with connector (specify length in multiples of 5 m) O-ring IR20
NOTE: The dome of IR20 and IR20WS cannot be supplied as spare part.
7.4 Appendix on standards for classification and calibration
Unlike pyranometers, pyrgeometers are not subject to a system of classification. At Hukseflux we distinguish between normal pyrgeometers, like our model IR02, and
“research grade” pyrgeometers, like IR20 and IR20WS. The term “research grade” is
used to indicate that this instrument has the highest attainable specifications.
Page 33
IR2 0 manual v 1604 33/39
7.5 Appendix on calibration hierarchy
Hukseflux pyrgeometers are traceable to the World Infrared Standard Group (WISG). WISG is composed of a group of pyrgeometers. The calibration hierarchy of Hukseflux IR20 is from WISG through Hukseflux internal calibration procedure involving outdoor comparison to a reference pyrgeometer.
The WISG group of instruments is maintained by World Radiation Center (WRC), in Davos Switzerland. An absolute sky-scanning radiometer provides the absolute longwave irradiance reference. Comparisons between the reference and the WISG are performed on a regular basis to maintain the WISG and supervise its long-term stability. It is essential that these intercomparisons take place under various sky conditions, but the predominant condition is a clear sky, which means that the validity of WISG calibration is a clear-sky spectrum. Typical exchange between pyrgeometer and sky is in the -70 to
-120 W/m2.
At Hukseflux in an independent lab experiment, the detector properties are determined as a function of temperature, resulting in temperature coefficients a, b, and c.
WRC works with a measurement equation involving additional constants k1, which corrects for thermal resistance of the thermopile sensor, k2 , which corrects for emissivity that does not equal 1.
E = (U / S) (1+ k
1
σ·(T + 273.15)
3
) + k2 σ·(T + 273.15)4 (Formula 7.5.1)
Pyrgeometers are calibrated by WRC in two ways:
- by nighttime comparison to WISG, to determine S. A typical uncertainty of S is 4.2%.
- by calibration under a blackbody to determine k1 and k2.
There also is a third constant, k3 , correcting for the difference between dome temperature and thermopile temperature.
The Hukseflux measurement equation for IR20 is:
E = U/(S0·(a·T² + b·T + c)) + σ·(T + 273.15)4 (Formula 0.1)
In other words, Hukseflux ignores k1 and assumes that k2 = 1. It is Hukseflux opinion that by the design of the instrument, the influence of these parameters is negligible.
Page 34
IR2 0 manual v 1604 34/39
Table 7.5.1 Calibration hierarchy for pyrgeometers
WORKING STANDARD CALIBRATION AT PMOD / WRC DAVOS
Calibration of working standard pyrgeometers traceable to WISG. A typical uncertainty of S0 is 4.2% (k = 2).
CORRECTION OF (WORKING) STANDARD CALIBRATION TO STANDARDISED REFERENCE CONDITIONS
Correction from “test conditions of the standard” to “reference conditions” : No corrections are applied. Reference conditions are: horizontal mounting, atmospheric longwave irradiance, clear sky nights, 20 °C.
OUTDOOR PRODUCT CALIBRATION AT HUKSEFLUX
Calibration of products, i.e. pyrgeometers: Outdoor side by side comparison to a reference pyrgeometer.
CALIBRATION UNCERTAINTY CALCULATION
ISO 98-3 Guide to the Expression of Uncertainty in Measurement, GUM Determination of combined expanded uncertainty of calibration of the product, including uncertainty of the working standard, uncertainty of correction, uncertainty of the method (transfer error). The coverage factor must be determined; at Hukseflux we work with a coverage factor k = 2. Hukseflux specifies a calibration uncertainty of < 6% (k = 2).
Page 35
IR2 0 manual v 1604 35/39
7.6 Appendix on meteorological radiation quantities
A pyrgeometer measures longwave irradiance. The time integrated total is called radiant exposure.
Table 7.6.1 Meteorological radiation quantities as recommended by WMO (additional symbols by Hukseflux Thermal Sensor). POA stands for Plane of Array irradiance. The term originates from ASTM and IEC standards.
SYMBOL
DESCRIPTION
CALCULATION
UNITS
ALTERNATIVE EXPRESSION
E
downward irradiance
E = E
g
↓ + E
l
W/m2
H
downward radiant exposure for a specified time interval
H = Hg ↓ + H
l
J/m2 E
upward irradiance
E = E
g
↑ + El
W/m2
H
upward radiant exposure for a specified time interval
H = H
g
↑ + Hl
J/m2
Wh/m2
Change of units
E
direct solar irradiance normal to the apparent solar zenith angle
W/m2
DNI
Direct Normal Irradiance
E0
solar constant
W/m2
E
g
h
global irradiance; hemispherical irradiance on a specified, in this case horizontal surface.*
Eg↓ = E cos θh + Ed↓
W/m2
GHI
Global Horizontal Irradiance
E
g
t
global irradiance; hemispherical irradiance on a specified, in this case tilted surface.*
E
g
= E∙cos θ
t
+
E
d
t + E
r ↑ t
***
W/m2
POA
Plane of Array
E
d
downward diffuse solar radiation
W/m2
DHI
Diffuse Horizontal Irradiance
E
l
↑, El
upward / downward longwave irradiance
W/m2 Er ↑
reflected solar irradiance
W/m2
E*
net irradiance
E* = E – E
W/m2
T
surface
equivalent blackbody radiative temperature of the surface**
ºC
T
sky
equivalent blackbody radiative temperature of the sky**
ºC
SD
sunshine duration
H
θ is the apparent solar zenith angle θh relative to horizontal, θt relative to a tilted surface
g = global, l = longwave, t = tilted *, h = horizontal* * distinction horizontal and tilted from Hukseflux, ** T symbols introduced by Hukseflux, *** contributions of E
d
t and E
r↑ t
are E
d
↓ and E
r↑
both corrected for the tilt angle of the
surface
Page 36
IR2 0 manual v 1604 36/39
7.7 Appendix on terminology / glossary
Table 7.7.1 Definitions and references of used terms
TERM
DEFINITION (REFERENCE)
Solar energy or solar radiation
solar energy is the electromagnetic energy emitted by the sun. Solar energy is also called solar radiation and shortwave radiation. The solar radiation incident on the top of the terrestrial atmosphere is called extra-terrestrial solar radiation; 97 % of which is confined to the spectral range of 290 to 3 000 x 10-9 m. Part of the extra-terrestrial solar radiation penetrates the atmosphere and directly reaches the earth’s surface, while part of it is scattered and / or absorbed by the gas molecules, aerosol particles, cloud droplets and cloud crystals in the atmosphere. The former is the direct component, the latter is the diffuse component of the solar radiation. (ref: WMO, Hukseflux)
Hemispherical solar radiation
solar radiation received by a plane surface from a 180 ° field of view angle (solid angle of 2 π sr).(ref: ISO 9060)
Global solar radiation
the solar radiation received from a 180 ° field of view angle on a horizontal surface is referred to as global radiation. Also called GHI. This includes radiation received directly from the solid angle of the sun’s disc, as well as diffuse sky radiation that has been scattered in traversing the atmosphere. (ref: WMO) Hemispherical solar radiation received by a horizontal plane surface. (ref: ISO 9060)
Plane-of-array irradiance
also POA: hemispherical solar irradiance in the plane of a PV array. (ref: ASTM E2848-11 / IEC 61724)
Direct solar radiation
radiation received from a small solid angle centred on the sun’s disc, on a given plane. (ref: ISO 9060)
Terrestrial or longwave radiation
radiation not of solar origin but of terrestrial and atmospheric origin and having longer wavelengths (3 000 to 100 000 x 10-9 m). In case of downwelling E
l
↓ also
the background radiation from the universe is involved, passing through the
”atmospheric window”. In case of upwelling E
l
↑, composed of longwave
electromagnetic energy emitted by the earth’s surface and by the gases, aerosols
and clouds of the atmosphere; it is also partly absorbed within the atmosphere. For a temperature of 300 K, 99.99 % of the power of the terrestrial radiation has a wavelength longer than 3 000 x 10-9 m and about 99 per cent longer than 5 000 x 10-9 m. For lower temperatures, the spectrum shifts to longer wavelengths. (ref: WMO)
World Radiometric Reference (WRR)
measurement standard representing the Sl unit of irradiance with an uncertainty of less than ± 0.3 % (see the WMO Guide to Meteorological Instruments and Methods of Observation, 1983, subclause 9.1.3). The reference was adopted by the World Meteorological Organization (WMO) and has been in effect since 1 July
1980. (ref: ISO 9060)
Albedo
ratio of reflected and incoming solar radiation. Dimensionless number that varies between 0 and 1. Typical albedo values are: < 0.1 for water, from 0.1 for wet soils to 0.5 for dry sand, from 0.1 to 0.4 for vegetation, up to 0.9 for fresh snow.
Angle of incidence
angle of radiation relative to the sensor measured from normal incidence (varies from 0° to 90°).
Zenith angle
angle of incidence of radiation, relative to zenith. Equals angle of incidence for horizontally mounted instruments
Azimuth angle
angle of incidence of radiation, projected in the plane of the sensor surface. Varies from 0° to 360°. 0 is by definition the cable exit direction, also called north, west is + 90°.
Sunshine duration
sunshine duration during a given period is defined as the sum of that sub-period for which the direct solar irradiance exceeds 120 W/m2. (ref: WMO)
Page 37
IR2 0 manual v 1604 37/39
WISG
World Infra Red Standard Group. Group of pyrgeometers, maintained by PMOD Davos Switzerland that forms the reference for calibration of pyrgeometers. WISG is traceable to international standards through an absolute sky scanning radiometer. WISG has been formally recognised by the World Meteorological Organisation WMO as “interim WMO Pyrgeometer Infrared Reference”.
Sky temperature
equivalent blackbody radiative temperature of the sky; i.e. the temperature calculated from pyrgeometer data measuring downwelling longwave radiation, assuming the sky behaves as a blackbody with an emission coefficient of 1.
Surface temperature
equivalent blackbody radiative temperature of the surface; i.e. the temperature calculated from pyrgeometer data measuring upwelling longwave radiation, assuming the ground behaves as a blackbody with an emission coefficient of 1.
Page 38
IR2 0 manual v 1604 38/39
7.8 EU declaration of conformity
We, Hukseflux Thermal Sensors B.V. Elektronicaweg 25 2628 XG Delft The Netherlands
in accordance with the requirements of the following directive:
2014/30/EU The Electromagnetic Compatibility Directive
hereby declare under our sole responsibility that:
Product model: IR20 and IR20WS Product type: Pyrgeometer
has been designed to comply and is in conformity with the relevant sections and applicable requirements of the following standards:
Emission: EN 61326-1 (2006) Immunity: EN 61326-1 (2006) Emission: EN 61000-3-2 (2006) Emission: EN 61000-3-3 (1995) + A1 (2001) + A2 (2005). Report: 08C01340RPT01, 06 January 2009
Kees VAN DEN BOS Director Delft 20 April, 2016
Page 39
© 2016, Hukseflux Thermal Sensors B.V.
www.hukseflux.com
Hukseflux Thermal Sensors B.V. reserves the right to change specifications without notice.
Loading...