Hudson Robotics Dried Blood Spot Processor User Manual

DBS USER MANUAL
10 Stern Avenue, Springfield, NJ 07081
Tel: 973-376-7400 Fax: 973-376-8265
SECTION I ......................................................................................................................1
Introduction ..........................................................................................................1
SECTION II .....................................................................................................................2
Processor Overview .............................................................................................2
A. Hardware ............................................................................................2
1. Main Menu ..............................................................................3
2. Run Screen ..............................................................................4
3. Calibration Screen ...................................................................5
4. Data Handling .........................................................................6
SECTION III ....................................................................................................................7
Machine Operation ...............................................................................................7
A. Start-Up ..............................................................................................7
1. Cable Connections ..................................................................7
2. Power-Up The Processor ........................................................7
B. Calibration ..........................................................................................8
1. Factory Calibration Method ....................................................9
SECTION IV ....................................................................................................................11
Troubleshooting ...................................................................................................11
A. Problem/Cause/Solution ....................................................................11
1. Processor .................................................................................11
2. XY Stage .................................................................................11
3. Punch Mechanism ...................................................................11
4. Barcode ...................................................................................12
DBS Processor - Table of Contents Page i
SECTION I
INTRODUCTION
The Hudson Dried Blood Spot Processor (DBS) links the barcode identification of a sample (a dried blood spot or other sample impregnated on paper) to a location in a micro-titer plate. After the micro-titer plate is processed, the assay results can easily be linked eliminating the chance of sample mis-identification.
The operator loads the micro-titer plate into the plate holder and the operator then instructs the processor to move to the first well. The operator picks up a sample and positions the barcode under the scanner. When the barcode is read and recognized, the operator then positions the sample under the punch selecting the best part of the sample to be inserted into the plate well. When the spot is properly positioned, the operator will press the foot switch, activating the punch. The spot is automatically inserted into the plate well.
The DBS automatically indexes the micro-titer plate to the next well location. The operator selects the next sample and follows the procedure described above. The operator repeats this cycle until the last well of the plate is completed, at which time, the result data is processed.
At completion of the plate, the result data is written to a data file and archived to the computer’s hard drive. The result data includes the micro-titer plate id, date/time, and specimen bar code ids.
The processor utilizes several means to avoid error:
1. The plate holder is cornered allowing the plate to be inserted in only one orientation.
2. The computer, communicating to the processor via RS-232, confirms a plate is loaded into
the holder by monitoring a sensor located internal to the plate holder.
3. The computer monitors a sensor (located on the punch mechanism) to insure the punch has
returned to its up position.
DBS Processor Page 1
SECTION II
PROCESSOR OVERVIEW
A. Hardware
Part Name Part #
Peripheral Equipment
footswitch (1) ................................842-6091
computer cables (1) ......................83F888
AC line cord (1) ............................ELC-1009
Barcode
Hand held scanner (1) ....................LS1006-I000-1900Z
Punch and Die
die plate (1) ...................................A1578
punch (1) .......................................A2431
punch holder (1)…………………..A2430
X-Y Stage
lead screw shaft (2) .......................SPM4-3708-10.4375
lead screw nut (2) .........................TAB4-3708M
brg support unit (2) .......................B6-01
linear guide (4) .............................SR20V2UU+280L-II
Electrical
photosensor (6) .............................EE-SX671
limit switch (1) .............................ELC-1025
indexer/driver (2) ..........................IM483I
driver power supply (1) ................IP404
I/O power supply (1) .....................800-9756
rocker switch (1) ...........................ELC-1019
fuse (1) ..........................................10 amp
fuse (1) ...........................................2 amp
fuse (1) ...........................................3 amp
motor……………………………..M2-2220-S interface board……………………OTP4-232
DBS Processor Page 2
B. Software
This button begins a run using a new destination plate. See below for more information on processing plates.
This button will display the calibration dialog where a user can adjust the positioning of the DBS. See below for more information on calibrating.
This button displays a history of all plates stored in the system and allows the user to select plates to view or delete.
This button displays a history of all plates stored in the system and allows the user to select plates to export to a text file.
This button allows the user to change which RS232 ports are used for communication with the DBS and the barcode scanner.
1. Main Screen
The Dried Blood Spot Testing System boots up to display its main screen as show below.
The main screen provides the following available selections:
DBS Processor Page 3