![](/html/f2/f268/f268bc3611fcf2fae1cf6645e4b008dfc7525e930337ca345b96fdb9f434c939/bg1.png)
Straight Blade Devices
Technical Information
HBL® Extra Heavy Duty, Straight Blade
Hospital Grade Duplex Receptacles
Specifications
Receptacles Part Description
Typical Specification – Catalog No. HBL8200 Receptacle 15A
Manufacturer’s Identification – Hubbell HBL8200 Top PBT*
Description – Hospital Grade, Straight Blade Duplex Receptacle Base RTP**
Type – 2 Pole, 3 Wire, Grounding By-Pass Power Contacts .037 in. (.9) Brass – Nickel Finish
Rating – 15A, 125V Clamping Plate .031 in. (.8) Brass – Nickel Finish
Certification: UL Listed to UL 498 File E2186, Fed. Spec. WC596, Mounting Strap .050 in. (1.3) Brass – Nickel Finish
Hospital Grade, NEMA
Certified to CSA, C22.2, No. 42 File 285 Hex Head Grounding Screw #10-32 (Green) Silicon Bronze, Multiple Drive
®
WD-6 Compliant Line Terminal Screws #10-32 Silicone Bronze, Multiple Drive
Center Assembly Rivet Silicon Bronze
Automatic Self-Grounding Clip Stainless Steel
Flat Head Mounting Screws Stainless Steel
HBL820 0GY
Performance
Electrical
Dielectric Voltage Withstands 2,000V minimum.
Max. Working Voltage 125V or 250V as rated.
Current Interrupting Certified for current interrupting at full rated current.
Temperature Rise Max 30°C temperature rise at full rated current after 50 cycles of overload at
Mechanical
Terminal Identification Terminals identified in accordance with UL 498 and CSA (Brass, White, Green).
Terminal Accommodation #14-10 AWG stranded or solid copper conductor only.
Product Identification Ratings are a permanent part of the device.
Environmental – Material
Flammability UL 94V-0 and UL 94-5 VA.
Operating Temperature Maximum continuous 75°C; minimum -40°C (w/o impact).
Note: *Polybutylene Terephthalate.
**Reinforced Thermoplastic Polyester.
150% of rated current at a power factor of 75%.
NEMA® is a registere d trademark of the National Electrical Manufacturers Associ ation.
A-50
®
Wiring Device-Kellems