Huaxin HS108 User Manual

User manual
HS108
Digital Pressure Gauge
The operating instructions are parts of the products and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The manufacturer's liability is void in the case of any damage caused by using the product contrary to its intended use, non-compliance with these operating instructions, assignment of insufficiently qualified
skilled personnel or unauthorized modications to the instrument.
HUAXIN reserve the rights to change the contents or form of these operating instructions at any time without prior notice having been given.
In order to protect your Huaxin products,your own and others safety,pls read this manual carefully before your operation,Otherwise it may cause trouble.
Claim
Disassembling the inner components may take toll or completely damage the equipment. Only the qualified technician could make a repairment. Any abnormal happened to the equipment pls call our service staff to make a repairment under their instruction.
To avoid over pressure damages,don't apply pressure that exceeds
the limits listed in the pressure specication table.
Safety information
Please don't disassemble & repair the equipment by yourself
Pay attention to the range
This action might injure the product and inuence the performance.
That will damage the items by using the wrong medium.
Don't beat and shock
Pay attention to the medium of the items
1
3
2
4
Warning
That will damage the items.and even may cause the re or an electric
shock.
Other power supply will damage to the product,and even may cause
the re or an electric shock.
The calibration should be carried out under a standard calibration condition, such as the standard gauge must be accuracy 3 times of HS108, ambient temperature (20±2)℃.
Don't immersed in water or exposed to water or getting wet
Only use the special charger or battery
The calibration should be carried out under a standard calibration condition
Symbols
Labelling,Safety Marks
Symbols used on the product and in the menu
Symbol Meaning Symbol Meaning
Indicates warning
Conforms to European Union directives
Indicates important information
Conforms to standard of PCEC
Digital Pressure Gauge Model:HS108 NO.1401011708DN Voltage:8.4DCV Range:(0-1)bar Date:2014.1 Accuracy:±0.5%F.S Huaxin Instrument(Beijing)Co.,Ltd
Product name Model Voltage Date of manufacture
Accuracy
Manufacturer
Pressure range
Serial number
5
6
7
Claim Safety information
1. Summary - 1 -
2. Technical Spec - 2 -
3. W orking Principle - 4 -
4. Install Operation - 4 -
4.1. Basic structure - 4 -
4.2. Button&terminal Introduction - 5 -
4.3. Display - 6 -
4.4. Installation method - 6 -
4.5. The gauge type - 6 -
4.6. The Max/Min valve display - 7 -
4.7. Changing the original pressure unit - 7 -
4.8. Temperature modication display - 8 -
4.9. COM Communication setting - 9 -
4.10. Digital Display setting - 10 -
4.11. Backlight time adjustment - 11 -
4.12. The Auto off setting - 11 -
4.13.Displaying speed set - 12-
4.14. Factory reset - 13 -
5. Calibration - 13 -
5.1. Pressure Low point, High Point Calibration - 13 -
5.2. Clear two point modication value - 15 -
5.3. Clearing clear zero value - 16 -
5.4. Clearing secondary modication sign - 16 -
5.5. Clear temperature modication value - 17 -
6. Maintenance - 17 -
6.1 Maintenance of instruments - 17 -
6.2 Attention items - 18 -
7. Accessories List - 18 -
8. Standard - 18 -
1.Summary
Characteristics
Pressure range to 2500bar High accuracy up to ±0.025%F.S Large,easy to read dual display with 6-digit resolution White backlight and RS232 communication port Various Pressure Units selectable Zero Point,Full Scale calibration function Max/Min values function %Pressure indication with bar graph
7.4Vdc rechargeable lithium battery or AC adapter Data logging fuction,store measuring value Auto Off can be set The backlight time is adjustable Temperature compensating A compact cost effective pressure gauge in various application regions Wetted parts:316 Stainless steel
Dimension:Φ95mmx49mm
Total height:166(mm) Weight:0.75kg
HS108 is a high-precision single-range digital pressure gauge, which is composed of high-precision measurement chips, temperature measuring circuit&battery&voltage measuring circuit and intelligent digital pressure module.
Every HS108 pass the high and lower temperature aging,so it could assure the products more accurate and reliable.the product combines a variety of accuracy levels to meet the customer requirments.it provides an accurate,reliable and economic solution for pressure application.
This product can be used for real-time measurement of pressure value and as a pressure reference or in applications where high accuracy measurement is required.
Model
HS108 Digital Pressure Gauge
Accuracy
±0.025% F.S ±0.05% F.S ±0.1%F.S
Gauge types
Gauge pressure Compound pressure Absolute pressure Differential pressure
Work medium Oil, water&noncorrosive gases
Over pressure alarm
120%Full Scale
Display
Description: 6 digits big LCD with blacklight Display rate:Less than
0.025% of full scale is 150ms .More than 0.025% of full scale is 1s Numeral display height:16.5mm (0.65") %indication with bar graph scale for visual reference A battery icon indicates the power
Pressure unit
psi, bar, mbar, kgf/cm
2
,Pa, kPa,
MPa, mmH2O, mmHg
Temperature
Compensated temperature:
-10°C to 50°C Operating temperature:-10°C to 50°C Humidity: <95% Storing temperature:-20°C to 70°C
Pressure connection
1/4NPT male M20×1.5 male Other connections customized
Power
Battery:One piece lithium battery & charger Battery life:80 hours
Housing
Case material:Plastic Wetted parts:316 Stainless steel
Dimension:Φ95mmx49mm
Total height:166(mm) Weight:0.75kg
Data storing
Storage capacity : Calibration record of 30pcs gauge under test
Communication
RS232
Certicate
CE
.....................
EX
.....................
2.Technical Specication
Compound pressure
Gauge pressure
No.
Pressure Range
Accuracy Media
(psi)
HS108-1 (0-10) (0-0.7) 0.025,0.05,0.1 G HS108-2 (0-15) (0-1) 0.025,0.05,0.1 G HS108-3 (0-30) (0-2) 0.025,0.05,0.1 G,L HS108-4 (0-60) (0-4) 0.025,0.05,0.1 G,L HS108-5 (0-100) (0-7) 0.025,0.05,0.1 G,L HS108-6 (0-200) (0-14) 0.025,0.05,0.1 G,L HS108-7 (0-300) (0-20) 0.025,0.05,0.1 G,L HS108-8 (0-500) (0-35) 0.025,0.05,0.1 G,L
HS108-9 (0-600) (0-40) 0.025,0.05,0.1 G,L HS108-10 (0-1000) (0-70) 0.025,0.05,0.1 G,L HS108-11 (0-2000) (0-140) 0.025,0.05,0.1 G,L HS108-12 (0-3000) (0-200) 0.025,0.05,0.1 G,L HS108-13 (0-5000) (0-350) 0.025,0.05,0.1 G,L HS108-14 (0-6000) (0-400) 0.025,0.05,0.1 G,L HS108-15 (0-10000) (0-700) 0.025,0.05,0.1 G,L HS108-16 (0-15000) (0-1000) 0.025,0.05,0.1 G,L HS108-17 (0-24000) (0-1600) 0.05,0.1 G,L HS108-18 (0-37000) (0-2500) 0.05,0.1 G,L
No.
Pressure Range
Accuracy
Media
(psi)
HS108-19 (-0.4-0.4) (-0.025-0.025) 0.025,0.05,0.1 G,L HS108-20 (-0.8-0.8) (-0.05-0.05) 0.025,0.05,0.1 G,L HS108-21 (-1.5-1.5) (-0.1-0.1) 0.025,0.05,0.1 G,L HS108-22 (-6-6) (-0.4-0.4) 0.025,0.05,0.1 G,L HS108-23 (-15-15) (-1-1) 0.025,0.05,0.1 G,L
Remark:G=Gas,L=Liquid
Loading...
+ 16 hidden pages