![](/html/c4/c42a/c42a20dba754f145422b5221154fd5eff12ff521c79d7093e416ab95b8b6477b/bg1.png)
ENG
HT7003
USER MANUAL
Copyright HT ITALIA 2015 Release EN 1.00 - 26/01/2015
6. FUNCTION KEYS DESCRIPTION
6.1 ON/OFF key
Press ON/OFF key to switch on and off the instrument
6.2 DH key
By pressing DH key the measured value is hold on the display
where the symbol "DH" appear. Pressing again DH to disable this
function and resume normal operation
6.3 Auto Power OFF feature
In order to preserve internal battery, the instrument switches
automatically off approximately 12 minutes after it was last used.
The symbol “APO” appears on the display when this function is
active
Press twice the ON/OFF key to switch on the
instrument after an Auto Power OFF operation
7. PRELIMINARY INSTRUCTIONS
This instrument has been designed for use in
environments with pollution degree 2
It can be used for AC CURRENT
measurements on installations with
measurement category CAT III 300V
Do not test or connect to any circuit exceeding
the specified overload protection
Do not effect measurements under
environmental conditions exceeding the limits
indicated in § 12.2
Make sure that battery is properly installed
When the instrument is connected to
measuring circuits never touch any unused
terminal
If during measurement the displayed values
remain constant check whether the HOLD
function is active
CAUTION
CAUTION
1. PRECAUTIONS AND SAFETY MEASUREMENTS
This instrument complies with safety Standard IEC/EN61010-1
related to electronic measuring instruments. For your own safety and
to avoid damaging the instrument follow the procedures described in
this instruction manual and read carefully all notes preceded by this
symbol
CAUTION
Avoid measuring in humid or wet places
Avoid measuring in rooms where explosive
gas, combustible gas, steam or excessive dust
is present
Keep you insulated from the object under test
Do not touch exposed metal parts such as test
lead ends, sockets, fixing objects, circuits etc.
Avoid doing that if you notice anomalous
conditions such as breakages, deformations,
fractures, leakages of battery fluid, blind display
etc.
Be careful when measuring on circuits with
voltages exceeding 20V to avoid risks of
The following symbols are used in user manual and on the meter:
electrical shock
CAUTION - refer to the instruction manual - an
improper use may damage the instrument or its
components
Double insulated meter
Use appropriate security measures to apply/remove
the instrument on/from live conductors
AC current
Ground reference
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
8. HOW TO PERFORM THE MEASUREMENTS
8.1 AC Current measurement
CAUTION
Put the conductor as close as possible to the
middle of the jaws in order to meet the meter
accuracy specifications
When apply/remove the instrument on/from live
conductors, if is not possible to de-energize the
circuit on test, use appropriate security
measures (e.g. protective gloves) to avoid
dangerous electric shock for the operator and a
possible damage of the instrument
Disconnect the main voltage or wear protective
gloves before performing measurement
Put the hand below the safety line during the
1. Switch on the instrument by pressing ON/OFF key
2. Open the clamp jaw and insert the cable in the middle of it (see
figure – left part)
3. Do not perform measurements with more cables in cluded in the
clamp jaw (see figure – right part)
4. The AC current value will be shown on the display
5. If the message "OL" is displayed the maximum current value was
exceeded
6. For Data HOLD feature see § 6.2
7. At the end of the measurements remove the instrument from the
cable and switch off by pressing ON/OFF key
measurement as indicated in § 5
1.1 Preliminary instructions
CAUTION
This instrument has been designed for use in
environments with pollution degree 2
It can be used for AC CURRENT
measurements on installations with
measurement category CAT III 300V
Do not test or connect to any circuit exceeding
the specified overload protection
Do not effect measurements under
environmental conditions exceeding the limits
indicated in § 12.2
Make sure that battery is properly installed
When the instrument is connected to
measuring circuits never touch any unused
terminal
If during measurement the displayed values
remain constant check whether the HOLD
1.2 Measuring overvoltage category definitions
function is active
According to the IEC/EN61010-1 guidelines the circuits are divided
into the following measurement categories:
CAT IV is for measurements performed at the source of the low-
voltage installation
Examples are electricity meters and measurements on primary
overcurrent protection devices and ripple control units
CAT III is for measurements performed in the building installation
Examples are measurements on distribution boards, circuit
breakers, wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in the fixed installation, and equipment
for industrial use and some other equipment, for example,
stationary motors with permanent connection to fixed installation
CAT II is for measurements performed on circuits directly
connected to the low voltage installation
Examples are measurements on household appliances, portable
tools and similar equipment
CAT I is for measurements performed on circuits not directly
connected to MAINS
Examples are measurements on circuits not derived from
MAINS, and specially protected (internal) MAINS-derived circuits.
In the latter case, transient stresses are variable; for that reason,
the standard requires that the transient withstand capability of the
equipment is made known to the user
9. BATTERY REPLACEMENT
When the "BAT" low battery indication is displayed the battery must
be replaced.
CAUTION
Only expert technicians should perform this operation.
Before carrying out this operation, make sure you
have removed the cable being tested from inside the
1. Switch off the instrument by pressing ON/OFF key
2. Remove the fixing screw and the battery compartment cover in the
3. Use the tip of a screwdriver or a similar tool in the internal hole
4. Replace the battery with new one of the same type (see § 12.2) by
5. Replace battery compartment cover
6. Do not scatter old batteries into the environment. Use the relevant
10. CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths,
solvents, water, etc.
11. END OF LIFE
clamp jaw
direction of the arrow on it
(see figure) to remove the battery
respecting the correct polarity
containers for disposal
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
2. GENERAL DESCRIPTION
HT7003 have the following features:
AC current measurement up to 300A in Autorange
Data HOLD
Auto Power OFF
Function keys DH and ON/OFF are available (see § 6.1 and §
6.2). The selected quantity is displayed with indication of
measuring unit and active functions. The instrument is provided
with an Auto Power Off function consisting in an automatic
switching off 12 minutes after last pressure on keys.
3. PREPARATION FOR USE
3.1 Initial checks
This instrument was checked both mechanically and electrically
prior to shipment. All possible cares and precautions were taken
to let you receive the instrument under perfect conditions.
Notwithstanding we suggest you to check it rapidly (any damage
may have occurred during transport – if so please contact the
local distributor from whom you bought the item).
Make sure that all standard accessories mentioned in § 12.2 are
included. Should you have to return back the instrument for any
reason please follow the instructions mentioned in § 13
3.2 Supply voltage
The instrument is powered by 1x3V button battery type CR2032
included on meter. When battery is low the symbol "BAT"is
displayed. To replace/insert the battery follow the instructions
indicated in § 9
3.3 Calibration
The instrument complies with the technical specifications
contained in this manual and such compliance is guaranteed for
12 months
3.4 Storage
After a period of storage under extreme environmental conditions
exceeding the limits mentioned in § 12.2 let the instrument
resume normal operating conditions before using it
12. TECHNICAL SPECIFICATIONS
12.1 Technical characteristics
The accuracy is referred to 23°C± 5°C, <80%RH
AC Current
Range Resolution Frequency Accuracy (*)
40.00A 0.01A
300.0A 0.1A
Overload protection: 400A AC (300V max) for 1 minute
(*) Frequency response (0 200A / 40Hz 50Hz and 60Hz 500Hz): add
(1.5%rdg+3dgt) to the accuracy
NOTE: for measurement 0A 0.5A add 3 dgt to the accuracy
12.2 General specifications
Reference guidelines
Safety: IEC/EN61010-1, IEC/EN61010-2-32
EMC: IEC/EN61326-1
Measurement category: CAT III 300V
Pollution degree: 2
Insulation: double insulation
Max height of use: 2000m (6562ft)
Compliance with 2004/108/CE (EMC) and 2006/95/CE (LVD)
Display
Characteristics: LCD, 3 ¾dgt, 4000 points + decimal point
Conversion type: mean value
Sampling rate: 2.5 times/s
Over range indication: “OL” symbol at display
Power supply
Battery type: 1x3V button battery type CR2032
Low battery indication: symbol “BAT” at display
Battery life: approx. 70 hours
Auto Power OFF: after 12 minutes of idleness
Environmental conditions for use
Reference temperature: 23°C±5°C (73°F±41°F), <80%RH
Working temp.: 040°C (32°F 104°F), <80%RH
Storage temperature: -10 50°C (14°F 122°F), <70%RH
Mechanical characteristics
Dimensions (LxWxH): 150x52x24mm (6x2x1in)
Weight (with battery): 110g (4 ounces)
Max diameter cable: 27mm (1in)
Standard accessories
Battery
Carrying bag
User manual
50Hz 60Hz (1.8%rdg+5dgt)
4. INSTRUMENT AND DISPLAY DESCRIPTION
5. SAFETY LINE
Put the cable as close as possible to the middle of the jaws in order
to meet the meter accuracy specifications and take the hand under
the safety line (see below figure)
13. SERVICE
13.1 Warranty conditions
This instrument is guaranteed for one year against material or
production defects, in accordance with our general sales conditions.
During the warranty period the manufacturer reserves the right to
decide either to repair or replace the product.
The warranty shall not apply in the following cases:
Repair and/or replacement of accessories and battery (not
covered by warranty).
Repairs that may become necessary as a consequence of an
incorrect use of the instrument or due to its use together with
non-compatible appliances.
Repairs that may become necessary as a consequence of
improper packaging.
Repairs which may become necessary as a consequence of
interventions performed by unauthorized personnel.
Modifications to the instrument performed without the
manufacturer’s explicit authorization
Use not provided for in the instrument’s specifications or in the
instruction manual
13.2 Service
Should the instrument not work properly, make sure that battery is
correctly installed and working and replace if necessary before
contacting your distributor. Should you need for any reason to return
back the instrument for repair or replacement take prior agreements
with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault).
Use only original packaging. Any damage occurred in transit due to
not original packaging will be charged anyhow to the customer. The
manufacturer will not be responsible for any damage to persons or
things
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – ITALY
Tel: +39-0546-621002
Fax: +39-0546-621144
Web: www.ht-instruments.com
Email: ht@htitalia.it