ENGLISH
User manual
HT63
EN - 1
Table of contents:
1.PRECAUTIONS AND SAFETY MEASURES ............................................................... 2
1.1. Preliminary instructions ..................................................................................................... 2
1.2. During use ......................................................................................................................... 3
1.3. After use ............................................................................................................................ 3
1.4. Definition of Measurement (Overvoltage) category ........................................................... 3
2.GENERAL DESCRIPTION ........................................................................................... 4
2.1. Measuring average values andTRMS values .................................................................... 4
2.2. Definition of true root mean square value and Crest factor ............................................... 4
3.PREPARATION FOR USE ........................................................................................... 5
3.1. Initial checks ...................................................................................................................... 5
3.2. Instrument power supply ................................................................................................... 5
3.3. Calibration ......................................................................................................................... 5
3.4. Storage .............................................................................................................................. 5
4.OPERATING INSTRUCTIONS ..................................................................................... 6
4.1. Description of the instrument ............................................................................................. 6
4.1.1. Description of the controls .......................................................................................................... 6
4.2. Description of function keys .............................................................................................. 7
4.2.1. A HOLD key ................................................................................................................................ 7
4.2.2. RANGE key ................................................................................................................................ 7
4.2.3. MAX MIN/PEAK key ................................................................................................................... 7
4.2.4. Hz%/REL key ........................................................................................................................... 7
4.2.5. MODE key .................................................................................................................................. 8
4.2.6. HIRES key .................................................................................................................................. 8
4.2.7. Disabling the Auto Power Off function ........................................................................................ 8
4.3. Description of rotary switch functions ................................................................................ 9
4.3.1. DC Voltage measurement .......................................................................................................... 9
4.3.2. AC Voltage measurement ........................................................................................................ 10
4.3.3. Frequency and Duty Cycle measurement ................................................................................ 11
4.3.4. Resistance measurement and Continuity test .......................................................................... 12
4.3.5. Diode test .................................................................................................................................. 13
4.3.6. Capacitance measurement ....................................................................................................... 14
4.3.7. Temperature measurement with K-type probe ......................................................................... 15
4.3.8. DC Current measurement and e 4-20mA% reading ................................................................ 16
4.3.9. AC Current measurement ......................................................................................................... 17
5.MAINTENANCE ......................................................................................................... 18
5.1. Replacing the batteries and the internal fuses ................................................................ 18
5.2. Cleaning the instrument .................................................................................................. 18
5.3. End of life ........................................................................................................................ 18
6.TECHNICAL SPECIFICATIONS ................................................................................ 19
6.1. Technical characteristics ................................................................................................. 19
6.1.1. Reference standards ................................................................................................................ 21
6.1.2. General characteristics ............................................................................................................. 21
6.2. Environment .................................................................................................................... 21
6.2.1. Environmental conditions for use ............................................................................................. 21
6.3. Accessories ..................................................................................................................... 21
6.3.1. Accessories provided ............................................................................................................... 21
6.3.2. Optional accessories ................................................................................................................ 21
7.ASSISTANCE ............................................................................................................. 22
7.1. Warranty conditions ......................................................................................................... 22
7.2. Assistance ....................................................................................................................... 22
HT63
EN - 2
1. PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1 relevant to electronic measuring instruments. For your safety and in order to prevent damaging the instrument, please carefully follow the procedures described in this manual and read all
notes preceded by symbol with the utmost attention. Before and after carrying out measurements, carefully observe the following instructions:
Do not carry out any measurement in humid environments. Do not carry out any measurements in case gas, explosive materials or flammables are
present, or in dusty environments.
Avoid any contact with the circuit being measured if no measurements are being
carried out.
Avoid any contact with exposed metal parts, with unused measuring probes, circuits,
etc.
Do not carry out any measurement in case you find anomalies in the instrument such
as deformation, breaks, substance leaks, absence of display on the screen, etc.
Pay special attention when measuring voltages higher than 20V, since a risk of
electrical shock exists.
In this manual, and on the instrument, the following symbols are used:
Warning: observe the instructions given in this manual; improper use could damage the instrument or its components.
Double-insulated meter AC voltage or current
DC voltage or current Connection to earth
1.1. PRELIMINARY INSTRUCTIONS
This instrument has been designed for use in environments of pollution degree 2. It can be used for VOLTAGE and CURRENT measurements on installations with CAT
IV 600V and CAT III 1000V.
We recommend following the normal safety rules devised by the procedures for
carrying out operations on live systems and using the prescribed PPE to protect the user against dangerous currents and the instrument against incorrect use.
In case the lack of indication of the presence of voltage may represent a danger for the
operator, always carry out a continuity measurement before carrying out the measurement on the live system, in order to confirm the correct connection and condition of the leads.
Only the leads supplied with the instrument guarantee compliance with the safety
standards. They must be in good conditions and be replaced with identical models, when necessary.
Do not test circuits exceeding the specified voltage limits. Do not perform any test under environmental conditions exceeding the limits indicated
in § 6.2.1.
Check that the battery is correctly inserted. Make sure that the LCD display and the rotary switch indicate the same function.
HT63
EN - 3
1.2. DURING USE
Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the caution notes and/or instructions may damage the instrument and/or its components or be a source of danger for the operator.
Before activating the rotary switch, disconnect the test leads from the circuit being
measured.
When the instrument is connected to the circuit being measured, do not touch any
unused terminal.
Do not measure resistance in case external voltages are present; even if the
instrument is protected, an excessive voltage may cause malfunction.
While measuring, if the value or the sign of the quantity being measured remain
unchanged, check if the HOLD function is enabled.
1.3. AFTER USE
When measurement is complete, set the rotary switch to OFF to turn off the instrument. If the instrument is not to be used for a long time, remove the batteries.
1.4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements, defines what measurement category, commonly called overvoltage category, is. § 6.7.4: Measured circuits, reads:
(OMISSIS) Circuits are divided into the following measurement categories: Measurement category IV is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Measurement category III is for measurements performed on installations inside
buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.
Measurement category II is for measurements performed on circuits directly
connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user.
HT63
EN - 4
2. GENERAL DESCRIPTION
The instrument carries out the following measurements:
DC Voltage AC TRMS Voltage DC Current 4-20mA% display AC TRMS Current Resistance and Continuity test Diode test Capacity Current and voltage frequency Duty Cycle Temperature with K-type probe
Each of these functions can be selected by means of the appropriate switch. The instrument is also equipped with function keys (see § 4.2), an analogue bargraph and backlight. The instrument is also equipped with an Auto Power OFF function (which can be disabled), which automatically switches off the instrument 15 minutes after the last time a function key was pressed or the rotary switch was turned. To switch on the instrument again, turn the rotary switch.
2.1. MEASURING AVERAGE VALUES ANDTRMS VALUES
Measuring instruments of alternating quantities are divided into two big families: AVERAGE-VALUE meters: instruments measuring the value of the sole wave at
fundamental frequency (50 or 60 Hz).
TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMS
value of the quantity being tested. With a perfectly sinusoidal wave, the two families of instruments provide identical results. With distorted waves, instead, the readings shall differ. Average-value meters provide the RMS value of the sole fundamental wave; TRSM meters, instead, provide the RMS value of the whole wave, including harmonics (within the instruments bandwidth). Therefore, by measuring the same quantity with instruments from both families, the values obtained are identical only if the wave is perfectly sinusoidal. In case it is distorted, TRMS meters shall provide higher values than the values read by average-value meters.
2.2. DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR
The root mean square value of current is defined as follows: In a time equal to a period, an alternating current with a root mean square value of 1A intensity, circulating on a resistor, dissipates the same energy that, during the same time, would be dissipated by a direct current with an intensity of 1A". This definition results in the numeric expression:
G=
Tt
t
dttg
T
0
0
)(
1
2
The root mean square value is indicated with the acronym RMS.
The Crest Factor is defined as the relationship between the Peak Value of a signal and its RMS value: CF (G)=
RMS
p
G
G
This value changes with the signal waveform, for a purely
sinusoidal wave it is
2
=1.41. In case of distortion, the Crest Factor takes higher values
as wave distortion increases.
HT63
EN - 5
3. PREPARATION FOR USE
3.1. INITIAL CHECKS
Before shipping, the instrument has been checked from an electric as well as mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged.
However, we recommend generally checking the instrument in order to detect possible damage suffered during transport. In case anomalies are found, immediately contact the forwarding agent.
We also recommend checking that the packaging contains all components indicated in §
6.3.1. In case of discrepancy, please contact the Dealer. In case the instrument should be returned, please follow the instructions given in § 7.
3.2. INSTRUMENT POWER SUPPLY
The instrument is supplied with 4x1.5V alkaline batteries type AAA IEC LR03, included in the package. When batteries are flat, the symbol
appears on the display. To replace
the batteries, see § 5.1.
3.3. CALIBRATION
The instrument has the technical specifications described in this manual. The instrument's performance is guaranteed for one year.
3.4. STORAGE
In order to guarantee precise measurement, after a long storage time under extreme environmental conditions, wait for the instrument to come back to normal condition (see §
6.2.1).
HT63
EN - 6
4. OPERATING INSTRUCTIONS
4.1. DESCRIPTION OF THE INSTRUMENT
4.1.1. Description of the controls
CAPTION:
1. LCD display
2. HIRES key
3. Hz% / REL key
4. RANGE key
5. A HOLD key
6. MODE key
7. MAXMIN/PEAK key
8. Rotary selector switch
9. Input terminal 10A
10. Input terminal VHz%CAPTemp
11. Input terminal mAA
12. Input terminal COM
Fig. 1: Description of the instrument
HT63
EN - 7
4.2. DESCRIPTION OF FUNCTION KEYS
4.2.1. A HOLD key
Pressing the A HOLD key freezes the value of the measured quantity on the display. After pressing this key, the message HOLDappears on the display. Press the HOLD key again to exit the function. Press and hold the A HOLD key for a long time in order to activate/deactivate the
AutoHOLDfunction. The message A HOLDappears on the display. With this function
enabled, the instrument freezes the value of the quantity on the display, automatically updating it when a new stable value is detected; a buzzer sounds when this happens. This function is useful in case of input signals with quick oscillations.
4.2.2. RANGE key
Press the RANGE key to activate the manual mode and to disable the Autorange function. The symbol AUTOdisappears from the upper left part of the display. In manual mode, press the RANGE key to change measuring range: the relevant decimal point will change its position. The RANGE key is not active in positions
, , CAP, Hz%, Temp°C°F and 4-20mA%. In Autorange mode, the instrument selects the most appropriate ratio for carrying out measurement. If a reading is higher than the maximum measurable value, the indication
O.Lappears on the display. Press and hold the RANGE key for more than 1
second to exit the manual mode and restore the Autorange mode.
4.2.3. MAX MIN/PEAK key
Pressing the MAX MIN/PEAK key once activates the detection of maximum (MAX), minimum (MIN) and average (AVG) value of the quantity being measured. The values are constantly updated and are displayed cyclically every time the same key is pressed. The
MAX MIN/PEAK key is not active when the HOLD function is activated. The MAX MIN/PEAK key is not active in positions , , CAP, Hz%, Temp°C°F and 4-20mA%.
Press and hold the MAX MIN/PEAK key for more than 1 second or turn the selector to exit the function. Press and hold the MAX MIN/PEAK key for more than two seconds to activate the detection of Maximum and Minimum peak values o AC Voltage and Current with a response time of 1ms. Both values are constantly updated and are displayed cyclically every time the same key is pressed again. The display shows the symbol associated with the selected function: PMAXfor maximum peak value, PMINfor minimum peak value. The Auto Power OFF function is automatically disabled. Press and hold the MAX MIN/PEAK key for more than 2 seconds or turn the selector to exit the function.
4.2.4. Hz%/REL key
Press the Hz%/REL key to select frequency measurement and duty cycle test in positions VHz%, 10AHz%, mA (AC), A and Hz% of the rotary selector. The frequency range is different in the different positions. Long pressing the Hz%/REL key activates relative measurement. The instrument zeroes the display and saves the displayed value as a reference value which subsequent measurements will be referred to. The symbol appears on the display. This function is not active in positions
, , Hz%, Temp°C°F and 4-20mA% . Press and hold the
Hz%/REL key again to exit the function.
HT63
EN - 8
4.2.5. MODE key
Pressing the MODE key allows selecting a double function on the rotary switch. In particular, it is active in position CAP to select diode test, continuity test, capacitance measurement and resistance measurement, in position Temp°C°F to select temperature measurement in °C o °F and mA , A to select AC or DC measurements. In position VHz%, pressing the MODE key activates/deactivates the function, whic h allows reading AC voltage with a low-pass filter, capable of reducing the effects of disturbance on the signal. In this condition, the instrument switches to Manual mode.
4.2.6. HIRES key
Press the HIRES key activates the instruments high-resolution operating mode, and the message HiResappears on the display. In this condition, the instrument switches to 4½ digits (60000 dots) for a better detail on data reading. This function is not active in positions Temp°C°F and in the capacitance measurement. Press again the HIRES key for a long time in order to go back to 3½ digits (6000 dots) standard display.
4.2.7. Disabling the Auto Power Off function
In order to preserve internal batteries, the instrument switches off automatically approximately 15 minutes after it was last used. To disable the Auto Power Off function, proceed as follows:
Switch off the instrument (OFF) Press and hold the MODE key and switch on the instrument by turning the rotary
switch. The message 1OFFappears on the display.
Switch off and then on again the instrument to enable the function.
HT63
EN - 9
4.3. DESCRIPTION OF ROTARY SWITCH FUNCTIONS
4.3.1. DC Voltage measurement
CAUTION
The maximum input DC voltage is 1000V. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 2: Use of the instrument for DC Voltage measurement
1. Select positions V or mV.
2. Insert the red cable into input terminal VHz%CAPTemp and the black cable into
input terminal COM.
3. Position the red lead and the black lead respectively in the spots with positive and
negative potential of the circuit to be measured (see Fig. 2). The display shows the value of voltage.
4. If the display shows the message "O.L", select a higher range.
5. When symbol "-" appears on the instruments display, it means that voltage has the
opposite direction with respect to the connection in Fig. 2.
6. To use the HOLD, RANGE, MAX MIN, REL and HIRES functions, see § 4.2
HT63
EN - 10
4.3.2. AC Voltage measurement
CAUTION
The maximum input AC voltage is 1000V. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 3: Use of the instrument for AC voltage measurement
1. Select positions VHz% or mV
2. In position mV , press the MODE key to display the ACsymbol on the display
3. Insert the red cable into input terminal VHz%CAPTemp and the black cable into
input terminal COM.
4. Position the red lead and the black lead respectively in the spots of the circuit to be
measured (see Fig. 3). The display shows the value of voltage.
5. If the display shows the message "O.L", select a higher range.
6. Press the Hz%/REL key to select measurements
Hzo %in order to display the
values of frequency and duty cycle of input voltage. The bargraph is not active in these functions. Press the MODE key to go back to voltage measurement and activate the function, if necessary.
7. To use the HOLD, RANGE, MAX MIN, PEAK, HIRES and REL functions, see § 4.2
HT63
EN - 11
4.3.3. Frequency and Duty Cycle measurement
CAUTION
The maximum input AC voltage is 1000V. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 4: Use of the instrument for frequency measurement and duty cycle test.
1. Select position Hz%.
2. Press the Hz% key to select measurements
Hzor %in order to display the values of
frequency and duty cycle of input voltage.
3. Insert the red cable into input terminal VHz%CAPTemp (HT62) and the black
cable into input terminal COM.
4. Position the red lead and the black lead respectively in the spots of the circuit to be
measured (see Fig. 4). The value of frequency (Hz) or of duty cycle (%) is shown on the display. The bargraph is not active in these functions.
5. If the display shows the message "O.L", select a higher range.
6. To use the HOLD and HIRES function, see § 4.2.
HT63
EN - 12
4.3.4. Resistance measurement and Continuity test
CAUTION
Before attempting any resistance measurement, cut off power supply from the circuit to be measured and make sure that all capacitors are discharged, if present.
Fig. 5: Use of the instrument for Resistance measurement and Continuity test
1. Select position CAP
2. Insert the red cable into input terminal VHz%CAPTemp and the black cable into
input terminal COM.
3. Position the test leads in the desired spots of the circuit to be measured (see Fig. 5).
The display shows the value of resistance.
4. If the display shows the message "O.L", select a higher range.
5. Press the MODE key to select “ ” measurement, relevant to the continuity test, and
position the test leads in the desired spots of the circuit to be measured.
6. The value of resistance (which is only indicative) is displayed in and the instrument
sounds if the value of resistance is <35.
7. To use the HOLD, RANGE, MAX MIN, REL and HIRES functions, see § 4.2
HT63
EN - 13
4.3.5. Diode test
CAUTION
Before attempting any resistance measurement, cut off power supply from the circuit to be measured and make sure that all capacitors are discharged, if present.
Fig. 6: Use of the instrument for diode test
1. Select position CAP
2. Press the MODE key to select measurement.
3. Insert the red cable into input terminal VHz%CAPTemp and the black cable into
input terminal COM.
4. Position the leads at the ends of the diode to be tested (see Fig. 6), respecting the
indicated polarity. The value of directly polarized threshold voltage is shown on the display.
5. If threshold value is equal to 0mV, the P-N junction of the diode is short-circuited.
6. If the display shows the message O.L, the terminals of the diode are reversed with
respect to the indication given in Fig. 6 or the P-N junction of the diode is damaged.
7. To use the HOLD and HIRES function, see § 4.2.
HT63
EN - 14
4.3.6. Capacitance measurement
CAUTION
Before carrying out capacitance measurements on circuits or capacitors, cut off power supply from the circuit being tested and let all capacitance in it be discharged. When connecting the multimeter and the capacitance to be measured, respect the correct polarity (when required).
Fig. 7: Use of the instrument for Capacitance measurement
1. Select position CAP
2. Press the MODE key until the symbol nFis displayed.
3. Insert the red cable into input terminal VHz%CAPTemp and the black cable
into input terminal COM.
4. Press the REL/ key before carrying out measurements.
5. Position the leads at the ends of the capacitor to be tested, respecting, if necessary ,
the positive (red cable) and negative (black cable) polarity (see Fig. 7). The display shows the value of capacitance. The bargraph is not active in this function.
6. The message O.L.indicates that the value of capacitance exceeds the maximum
measurable value.
7. To use the HOLD and REL functions, see § 4.2.
HT63
EN - 15
4.3.7. Temperature measurement with K-type probe
CAUTION
Before attempting any temperature measurement, cut off power supply from the circuit to be measured and make sure that all capacitors are discharged, if present.
Fig. 8: Use of the instrument for Temperature measurement
1. Select position Temp°C°F
2. Press the MODE key until the symbol °Cor °Fis displayed.
3. Insert the provided adapter into input terminals VHz%CAPTemp (polarity +) and
COM (polarity -) (see Fig. 8)
4. Connect the provided K-type wire probe or the optional K-type thermocouple (see § ) to
the instrument by means of the adapter, respecting the positive and negative polarity on it. The display shows the value of temperature. The bargraph is not active in this function.
5. The message O.L.indicates that the value of temperature exceeds the maximum
measurable value.
6. To use the HOLD function, see § 4.2.
HT63
EN - 16
4.3.8. DC Current measurement and e 4-20mA% reading
CAUTION
Maximum input DC current is 10A (input 10A) or 600mA (input mAA). Do not measure currents exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 9: Use of the instrument for DC current measurement and 4-20mA% reading.
1. Cut off power supply from the circuit to be measured.
2. Select position A , mA or 10A to measure DC current or the same position 4-
20mA% for 4-20mA% reading.
3. Insert the red cable into input terminal 10A or into input terminal mAA and the black
cable into input terminal COM.
4. Connect the red lead and the black lead in series to the circuit whose current you want
to measure, respecting polarity and current direction (see Fig. 9).
5. Supply the circuit to be measured.
6. The value of DC current (see Fig. 9 left side) appears on the display.
7. The value of reading 4-20mA% (0mA = -25%, 4mA = 0%, 20mA = 100% and 24mA =
125%) (see Fig. 9 right side) appears on the display. The bargraph is not active in this function
8. If the display shows the message O.L, the maximum measurable value has been
reached.
9. When symbol "-" appears on the instrument's display, it means that current has the
opposite direction with respect to the connection in Fig. 9.
10. To use the HOLD, RANGE, MAX MIN, REL and HIRES functions, see § 4.2
HT63
EN - 17
4.3.9. AC Current measurement
CAUTION
Maximum input AC current is 10A (input 10A) or 600mA (input mAA). Do not measure currents exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 10: Use of the instrument for AC current measurement
1. Cut off power supply from the circuit to be measured.
2. Select positions A , mA or 10A.
3. Press the MODE key to select ACmeasurement.
4. Insert the red cable into input terminal 10A or into input terminal mAA and the black
cable into input terminal COM.
5. Connect the red lead and the black lead in series to the circuit whose current you want
to measure (see Fig. 10).
6. Supply the circuit to be measured. The display shows the value of current.
7. If the display shows the message O.L, the maximum measurable value has been
reached.
8. Press the Hz%/REL key to select measurements Hz” or “%” in order to display the
values of frequency and duty cycle of input voltage. The bargraph is not active in these functions.
9. To use the HOLD, RANGE, MAX MIN, PEAK, REL and HIRES functions, see § 4.2
HT63
EN - 18
5. MAINTENANCE
CAUTION
Only expert and trained technicians should perform maintenance
operations. Before carrying out maintenance operations, disconnect all cables from the input terminals.
Do not use the instrument in environments with high humidity levels or
high temperatures. Do not expose to direct sunlight.
Always switch off the instrument after use. In case the instrument is not to
be used for a long time, remove the battery to avoid liquid leaks that could damage the instruments internal circuits.
5.1. REPLACING THE BATTERIES AND THE INTERNAL FUSES
When the LCD display shows the symbol “” , it is necessary to replace the battery. Replacing the battery
1. Position the rotary switch to OFF and remove the cables from the input terminals.
2. Turn the fastening screw of the battery compartment cover from position to position
and remove it.
3. Remove the battery and insert a new battery of the same type (see § 6.1.2), respecting
the indicated polarity.
4. Restore the battery compartment cover into place and turn the fastening screw from
position to position .
5. Do not scatter old batteries into the environment. Use the relevant containers for
disposal.
Replacement of fuses
1. Position the rotary switch to OFF and remove the cables from the input terminals.
2. Turn the fastening screw of the battery compartment cover from position to position
and remove it.
3. Remove the damaged fuse and insert a new fuse of the same type (see § 6.1.2).
4. Restore the battery compartment cover into place and turn the fastening screw from
position to position .
5.2. CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths, solvents, water, etc.
5.3. END OF LIFE
WARNING: the symbol on the instrument indicates that the appliance and its
accessories must be collected separately and correctly disposed of.
HT63
EN - 19
6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL CHARACTERISTICS
Accuracy calculated as [%reading + (num. digits*resolution)] at 18°C 28°C <75%HR
DC Voltage
Range Resolution Accuracy Input impedance
Protection against
overcharge
600.0mV 0.1mV (0.9%reading + 5digits)
>10M
1000VDC/ACrms
6,000V 0,001V
60.00V 0.01V
600.0V 0.1V 1000V 1V
(1.2%reading + 5digits)
AC TRMS Voltage
Range Resolution
Accuracy (*)
Protection against
overcharge
(50Hz1kHz) (1kHz5kHz)
600.0mV 0.1mV
(1.0%reading + 5digits)
(3.0%rdg + 5digits)
1000VDC/ACrms
6,000V 0,001V
60.00V 0.01V
600.0V 0.1V 1000V 1V Not specified
(*) Accuracy specified from 5% to 100% of the measuring range, Input impedance: > 10M Accuracy PEAK function: 10%rdg, Response time PEAK function: 1ms
Accuracy
function: (1.5%rdg + 5dgt) (f  60Hz), (3.0%rdg + 5dgt) (60Hz f 1kHz), not specified (f: 1 5kHz)
DC Current
Range Resolution Accuracy Protection against overcharge
600.0A 0.1A
(1.5%reading + 5digits)
Quick fuse 800mA/1000V
6000A 1A
60.00mA 0.01mA
600.0mA 0.1mA
10.00A 0.01A Quick fuse 10A/1000V
AC TRMS Current
Range Resolution
Accuracy (*)
Protection against overcharge
(50Hz1kHz) (1kHz5kHz)
600.0A 0.1A
(2.5%rdg + 5digits) (3.5%rdg+5digits)
Quick fuse 800mA/1000V
6000A 1A
60.00mA 0.01mA
600.0mA 0.1mA
10.00A 0.01A Quick fuse 10A/1000V
(*) Accuracy specified from 5% to 100% of the measuring range Accuracy PEAK function: 10%rdg
4-20mA% reading
Range Resolution Accuracy Correspondence
-25%125%
0.1%
(50digits)
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Diode test
Function Test current Max voltage with open circuit
<1.5mA 2.8VDC
HT63
EN - 20
Resistance and Continuity test
Range Resolution Accuracy Buzzer
Protection against
overcharge
600.0 0.1 (2.0%rdg + 9digits)
<35
1000VDC/ACrms
6.000k 0.001k (1.2%rdg + 5digits)
60.00k 0.01k
600.0k 0.1k
6.000M 0.001M (2.0%rdg + 10digits)
60.00M 0.01M
Frequency (electronic circuits)
Range Resolution Accuracy
Protection against
overcharge
40.00Hz 10kHz
0.01Hz
(0.5%reading)
1000VDC/ACrms
Sensitivity: 15Vrms
Frequency (electronic circuits)
Range Resolution Accuracy
Protection against
overcharge
60.00Hz 0.01Hz
(1.0%rdg + 2digits)
1000VDC/ACrms
600.0Hz 0.1Hz
6.000kHz 0.001kHz
60.00kHz 0.01kHz
600.0kHz 0.1kHz
6.000MHz 0.001MHz
10.00MHz 0.01MHz
Sensitivity: >2.0Vrms (@ 20% 80% duty cycle) and f<100kHz; >5Vrms (@ 20% 80% duty cycle) and f>100kHz
Duty Cycle
Range Resolution Accuracy
0.1% 99.9%
0.1%
(1.2%reading + 2digits)
Pulse frequency range: 5Hz 10kHz, Pulse amplitude: ±5V (100s  100ms)
Capacity
Range Resolution Accuracy Protection against overcharge
60.00nF 0.01nF (3.5%rdg + 10digits)
1000VDC/ACrms
600.0nF 0.1nF
6,000F 0,001F
(2.5%rdg + 10digits)
60.00F 0.01F
600.0F 0.1F
6.000mF 0.001mF
(3.5%rdg + 10digits)
Temperature with K-type probe
Range Resolution Accuracy (*)
Protection against
overcharge
-40.0°C ÷ 600.0°C 0.1°C (2.0%rdg + 3°C)
1000VDC/ACrms
600°C ÷ 760°C 1°C
-58.0°F ÷ 600.0°F 0.1°F
(2.0%rdg+ 5.5°F)
600°F ÷ 1400°F 1°F
(*) Instrument accuracy with no probe
HT63
EN - 21
6.1.1. Reference standards
Safety: IEC/EN61010-1 EMC: IEC/EN 61326-1 Insulation: double insulation Pollution level: 2 Overvoltage category: CAT IV 600V, CAT III 1000V Max operating altitude: 2000m (6562ft)
6.1.2. General characteristics Mechanical characteristics
Size (L x W x H): 175 x 85 x 55mm (7 x 3 x 2in) Weight (batteries included): 360g (13 ounces)
Power supply
Battery type: 4x1.5V batteries type AAA IEC LR03 Low battery indication: symbol “” on the display Auto Power Off: after 15 minutesidling (can be disabled) Fuses: F10A/1000V, 10 x 38mm (input 10A) F800mA/1000V, 6 x 32mm (input mAA)
Display
Conversion: TRMS Characteristics: 3½ LCD, 6000 dots, decimal sign, point 4½ LCD, 60000 dots, decimal sign, point backlight and bargraph Sampling frequency: 3 times/s
6.2. ENVIRONMENT
6.2.1. Environmental conditions for use
Reference temperature: 18°C 28°C (64°F 82°F) Operating temperature: 5°C ÷ 40°C (41°F 104°F) Allowable relative humidity: <80%HR Storage temperature: -20° ÷ 60°C (-4°F  140°F) Storage humidity: <80%HR
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC
This instrument satisfies the requirements of European Directive 2011/65/EU (RoHS)
and 2012/19/EU (WEEE)
6.3. ACCESSORIES
6.3.1. Accessories provided
Pair of test leads Adapter + K-type wire probe Batteries Carrying bag User manual
6.3.2. Optional accessories
Pair of test leads Code KIT4000A K-type probe for air and gas temperature Code TK107 K-type probe for semisolid substance temperature Code TK108 K-type probe for liquid substance temperature Code TK109 K-type probe for surface temperature Code TK110 K-type probe for surface temperature with 90° tip Code TK111
HT63
EN - 22
7. ASSISTANCE
7.1. WARRANTY CONDITIONS
This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product.
Should the instrument be returned to the After-sales Service or to a Dealer, transport will be at the Customer's charge. However, shipment will be agreed in advance. A report will always be enclosed to a shipment, stating the reasons for the product's return. Only use original packaging for shipment. Any damage due to the use of non-original packaging material will be charged to the Customer. The manufacturer declines any responsibility for injury to people or damage to property.
The warranty shall not apply in the following cases:
Repair and/or replacement of accessories and battery (not covered by warranty). Repairs that may become necessary as a consequence of an incorrect use of the
instrument or due to its use together with non-compatible appliances.
Repairs that may become necessary as a consequence of improper packaging. Repairs which may become necessary as a consequence of interventions performed
by unauthorized personnel.
Modifications to the instrument performed without the manufacturer's explicit
authorization.
Use not provided for in the instrument's specifications or in the instruction manual. The content of this manual cannot be reproduced in any form without the manufacturer's
authorization.
Our products are patented and our trademarks are registered. The manufacturer reserves the right to make changes in the specifications and prices if this is due to improvements in technology.
7.2. ASSISTANCE
If the instrument does not operate properly, before contacting the After-sales Service, please check the conditions of battery and cables and replace them, if necessary. Should the instrument still operate improperly, check that the product is operated according to the instructions given in this manual. Should the instrument be returned to the After-sales Service or to a Dealer, transport will be at the Customer's charge. However, shipment will be agreed in advance. A report will always be enclosed to a shipment, stating the reasons for the product's return. Only use original packaging for shipment; any damage due to the use of non-original packaging material will be charged to the Customer.
HT INSTRUMENTS SA
C/ Legalitat, 89
08024 Barcelona - ESP
Tel.: +34 93 408 17 77, Fax: +34 93 408 36 30
eMail: info@htinstruments.com
eMail: info@htinstruments.es
Web: www.htinstruments.es
HT ITALIA SRL
Via della Boaria, 40
48018 Faenza (RA) - ITA
Tel: +39 0546 621002
Fax: +39 0546 621144
eMail: ht@htitalia.it
Web: www.ht-instruments.com
HT INSTRUMENTS BRASIL
Rua Aguaçu, 171, bl. Ipê, sala 108
13098321 Campinas SP - BRA
Tel: +55 19 3367.8775
Fax: +55 19 9979.11325
eMail: vendas@ht-instruments.com.br
Web: www.ht-instruments.com.br
HT INSTRUMENTS USA LLC 3145 Bordentown Avenue W3
08859 Parlin - NJ - USA
Tel: +1 719 421 9323
eMail: sales@ht-instruments.us
Web: www.ht-instruments.com
HT INSTRUMENTS GMBH
Am Waldfriedhof 1b
D-41352 Korschenbroich - GER
Tel: +49 (0) 2161 564 581
Fax: + 49 (0) 2161 564 583
eMail: info@ht-instruments.de
Web: www.ht-instruments.de
HT ITALIA CHINA OFFICE
意大利HT中国办事处
Room 3208, 490# Tianhe road, Guangzhou - CHN
地址:广州市天河路490号壬丰大厦3208
Tel.: +86 400-882-1983, Fax: +86 (0) 20-38023992
eMail: zenglx_73@hotmail.com
Web: www.guangzhouht.com
Loading...