7.2. Service ................................................................................................................ 22
EN - 1
HT32
1. SAFETY PRECAUTIONS AND PROCEDURES
This instrument conforms with safety Standard EN 61010-1 related to electronic
measuring instruments.
For your own safety and to avoid damaging the instrument follow the procedures
described in this instruction manual and read carefully all notes preceded by this symbol
.
When taking measurements:
Avoid doing that in humid or wet places.
Avoid doing that in rooms where explosive gas, combustible gas, steam or excessive
dust is present.
Keep you insulated from the object under test.
Do not touch exposed metal parts such as test lead ends, sockets, fixing objects,
circuits etc.
Avoid doing that if you notice anomalous conditions such as breakages, deformations,
fractures, leakages of battery liquid, blind display etc.
Be particularly careful when measuring voltages exceeding 20V to avoid risks of
electrical shocks.
The following symbols are used:
CAUTION - refer to the instruction manual - an improper use may damage the
instrument or its components
Danger high voltage: risk of electric shock
Double insulated meter
AC voltage or current
DC voltage or current
1.1. PRELIMINARY
This instrument has been designed for use in environments of pollution degree 2.
It can be used for VOLTAGE and CURRENT measurements on installations of overvoltage
category III 1000V and IV 600V.
When using the instrument always respect the usual safety regulations aimed at:
Protecting you against the dangerous electric currents.
Protecting the instrument against incorrect operations.
Only the leads supplied with the instrument guarantee compliance with the safety
standards in force. They must be in good conditions and, if necessary, replaced with
identical ones.
Do not test or connect to any circuit exceeding the specified overload protection.
Do not effect measurements under environmental conditions exceeding the limits
indicated in paragraph 6.2.1.
Make sure that batteries are properly installed.
EN - 2
HT32
Before connecting the test probes to the installation, make sure that the function
selector is positioned on the required measurement.
Make sure that LCD and function selector indicate the same function.
1.2. DURING USE
Read the recommendations which follow and the instructions in this manual:
CAUTION
An improper use may damage the instrument and/or its components or
injure the operator.
When changing the range, first disconnect the test leads from the circuit under test in
order to avoid any accident.
When the instrument is connected to measuring circuits never touch any unused
terminal.
When measuring resistors do not add any voltage. Although there is a protection
circuit, excessive voltage could cause malfunctioning.
If during measurement the displayed values remain constant check whether the HOLD
function is active.
1.3. AFTER USE
After using the instrument turn it off.
If you expect not to use the instrument for a long period remove the battery to avoid
leakages of battery liquids which may damage its inner components.
EN 61010-1: Safety requirements for electrical equipment for measurement, control and
laboratory use, Part 1: General requirements, gives a definition of measuring category,
usually called overvoltage category. Paragraph 6.7.4: Measuring circuits:
(OMISSIS)
circuits are divided into the following measurement categories:
Measurement category IV is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent
protection devices and ripple control units.
Measurement category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit breakers, wiring,
including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, and equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to fixed installation.
Measurement category II is for measurements performed on circuits directly
connected to the low voltage installation..
Examples are measurements on household appliances, portable tools and similar
equipment..
Measurement category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses
are variable; for that reason, the norm requires that the transient withstand
capability of the equipment is made known to the user.
EN - 4
HT32
2. GENERAL DESCRIPTION
This instrument performs the following measurements:
DC and AC TRMS Voltage
DC and AC TRMS Current
Resistance and Continuity test
Diode test
Frequency
Capacitance
All selectable by means of a 10-position function selector (including OFF position). 4
FUNCTION keys are also available to hold the displayed value and select the desired
function. The selected quantity is displayed with indication of measuring unit and active
functions.
The instrument disposes of an Auto Power Off function consisting in an automatic
switching off 10 minutes after last pressure on keys or rotation of selector. To resume
normal operation turn the selector on OFF and switch it on again.
2.1. MEAN VALUE AND TRMS : DEFINITION
Safety testers for AC quantities are divided in two big families:
MEAN VALUE instruments, measuring only the value of the wave at the fundamental
frequency (50 or 60 Hz)
TRUE ROOT MEAN SQUARE (or “TRMS”) instruments, measuring the true root mean
square value of the quantity under test.
In presence of a perfectly sinusoidal wave, both families provide identical results. While in
presence of distorted waves, readings are different. Mean value instruments provide only
the value of the fundamental wave while TRMS instruments provide the value of the entire
wave, including harmonics (within the passband of the instrument). Accordingly, if the
same quantity is measured with both kinds of instruments, the measured values are
identical only if the wave is purely sinusoidal. Should it be distorted, TRMS instruments
provide higher values than MEAN VALUE instruments.
2.2. TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR : DEFINITION
The effective current value is defined as follows: “In an interval of time equivalent to a
period, an alternate current with effective value having an intensity of 1A, by passing on a
resistor, disperses the same energy which would be dispersed in the same period of time
by a direct current having an intensity of 1A”. From this definition comes the numerical
Tt
0
1
expression: G=
T
The crest factor is defined as the ratio between the peak value of a signal and its effective
G
value: CF (G)=
G
purely sinusoidal wave it’s worth
assumes higher values as long as the wave distortion is higher.
2
dttg
)(
t
0
p
. This value varies according to the waveform of the signal, for a
RMS
The effective value is indicated as RMS (root mean square).
=1.41. In presence of distortions the crest factor
2
EN - 5
HT32
3. PREPARATION FOR USE
3.1. INITIAL
This instrument was checked both mechanically and electrically prior to shipment.
All possible cares and precautions were taken to let you receive the instrument in perfect
conditions.
Notwithstanding we suggest you to check it rapidly (eventual damages may have occurred
during transport – if so please contact the local distributor from whom you bought the
item).
Make sure that all standard accessories mentioned in paragraph 6.3. are included.
Should you have to return back the instrument for any reason please follow the
instructions mentioned in paragraph 7.
3.2. SUPPLY VOLTAGE
The instrument is powered by batteries mod. 9V NEDA1604, JIS006P, IEC6F22 included
in the packaging. Battery life is approx. 300 hours. When batteries are low the symbol
"" is displayed.
To replace/insert batteries follow the instructions indicated in paragraph 5.2.
3.3. CALIBRATION
The instrument complies with the technical specifications contained in this manual and
such compliance is guaranteed for 1 year.
3.4. STORAGE
After a period of storage in extreme environmental conditions exceeding the limits
mentioned in paragraph 6.2.1 let the instrument resume normal operating conditions
before using it.
EN - 6
HT32
4. OPERATING INSTRUCTIONS
4.1. INSTRUMENT DESCRIPTION
4.1.1. Front panel
LEGEND:
1. LCD display.
2. Functions keys.
3. Rotary selector.
4. A input for AC/DC
current measurements.
5. HzVA input for
other measurements.
6. COM input.
Fig. 1: Instrument description
EN - 7
HT32
4.2. FUNCTION KEYS
When pressing a key, the corresponding symbol is displayed with a beep. To resume
default state turn the selector on another function.
4.2.1. MIN MAX
By pressing MIN MAX key, maximum and minimum values are measured. Both values are
stored and can be recalled by pressing the same key. The symbol corresponding to the
desired function is displayed: “MAX” for maximum value, “MIN” for minimum value. MIN MAX key is disabled when HOLD function is active. To exit this function keep MIN MAX
key pressed for at least 1 second or rotate the selector to another position.
4.2.2. RANGE
By pressing RANGE key, the manual mode is activated and the “AUTO” symbol
disappears from the display. Press RANGE cyclically to change the measuring range and
fix the decimal point on the display.
To exit this function keep RANGE key pressed for at least 1 second or rotate the selector
to another position.
4.2.3. HOLD
By pressing HOLD the measured value is frozen on the display where the symbol “HOLD”
appears. To exit this function press HOLD again or rotate the selector to another position.
4.2.4. Disable Auto Power OFF
When the meter is to be used for long periods of time, the operator might want to disable
the Auto Power OFF function. Once the Auto Power OFF function is disabled the meter
stays on continuously. To disable the Auto Power OFF function:
Switch OFF the meter.
Turn ON the meter keeping pressed MIN MAX and RANGE keys.
The Auto Power OFF function is automatically activated when turning ON again the meter.
EN - 8
HT32
4.3. MEASUREMENTS
4.3.1. DC Voltage measurement
Maximum input for DC voltage is 1000V. Do not attempt to take any voltage
measurement that exceeds the limits. Exceeding the limits could cause
electrical shock and damage the multimeter.
CAUTION
Fig. 2: DC Voltage measurement
1. Turn the selector on V.
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 2).
3. Connect the red and black test leads to the positive and negative poles of the circuit
under test respectively. The voltage value will be displayed with automatic detection of
the range.
4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
5. The message "OL" means that the voltage exceeds the measuring limits. In this case
disconnect the test leads from the circuit under test to avoid damaging the instrument
or endanger your own safety.
6. The symbol "-" on the display means that voltage has opposite direction than shown in
Fig. 2.
7. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 9
HT32
4.3.2. AC Voltage measurement
Maximum input for AC voltage is 750V rms. Do not attempt to take any
voltage measurement exceeding such limit to avoid the risk of electrical
shock and damages to the instrument.
CAUTION
Fig. 3: AC Voltage measurement
1. Turn the selector on V.
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 3).
3. Connect the red and black test leads to the positive and negative poles of the circuit
under test respectively. The voltage value will be displayed with automatic detection of
the range.
4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
5. The message "OL" means that the voltage exceeds the measuring limits. In this case
disconnect the test leads from the circuit under test to avoid damaging the instrument
or endanger your own safety.
6. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 10
HT32
4.3.3. DC Current measurement
Maximum input for DC current is 10A. Do not attempt to take any current
measurement exceeding such limit to avoid the risk of electrical shock and
damages to the instrument.
CAUTION
Fig. 4: DC Current measurement
1. Switch OFF the circuit under test.
2. Turn the selector on A.
3. Insert the red test lead into A jack and the black plug into COM jack (Fig. 4).
4. Connect red and black plugs in series with the circuit under test respecting the polarity
and current flow indicated in Fig. 4.
5. Power the circuit under test. The current value is displayed.
6. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
7. The message "OL" means that the current exceeds the measuring limits. In this case
disconnect the test leads from the circuit under test to avoid damaging the instrument
or endanger your own safety.
8. If the measured value is lower than 6mA, to get a better resolution:
Switch off the circuit under test.
Turn the selector on A.
Remove the red test lead from A jack, and insert it into HzVA jack.
Power the circuit under test.
9. The symbol "-" on the display means that current has opposite direction than shown in
Fig. 4.
10. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 11
HT32
4.3.4. AC Current measurement
Maximum input for AC current is 10A. Do not attempt to take any current
measurement exceeding such limit to avoid the risk of electrical shock and
damages to the instrument.
CAUTION
Fig. 5: AC Current measurement
1. Switch OFF the circuit under test.
2. Turn the selector on A.
3. Insert the test leads into the jacks, the red plug into A jack and the black plug into
COM jack (Fig. 5).
4. Connect red and black plugs in series with the circuit under test as indicated in Fig. 4.
5. Power the circuit under test. The current value is displayed.
6. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
7. The message "OL" means that the current exceeds the measuring limits. In this case
disconnect the test leads from the circuit under test to avoid damaging the instrument
or endanger your own safety.
8. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 12
HT32
4.3.5. Frequency measurement
Maximum input for AC voltage is 750V rms. Do not attempt to take any
voltage measurement exceeding such limit to avoid the risk of electrical
shock and damages to the instrument.
CAUTION
Fig. 6: Frequency measurement
1. Turn the selector on Hz.
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 6).
3. Connect the red and black test leads to the positive and negative poles of the circuit
under test respectively. The voltage value will be displayed with automatic detection of
the range.
4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
5. The message "OL" means that the frequency exceeds the measuring limits. In this
case disconnect the test leads from the circuit under test to avoid damaging the
instrument or endanger your own safety.
6. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 13
HT32
4.3.6. Resistance measurement
Before taking resistance measurements in circuit remove power from the
circuit being tested and discharge all capacitors.
CAUTION
Fig. 7: Resistance measurement
1. Turn the selector on .
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 7).
3. Connect the red and black test leads to the circuit under test, the resistance value will
be displayed with automatic detection of the range.
4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
5. The message "OL" means that the resistance exceeds the measuring limits. In this
case disconnect the test leads from the circuit under test to avoid damaging the
instrument or endanger your own safety.
6. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 14
HT32
4.3.7. Diode test and continuity test
Before taking resistance measurements in circuit remove power from the
circuit being tested and discharge all capacitors.
CAUTION
Fig. 8: Diode test and continuity test
1. Turn the selector on /.
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 8).
3. Connect the red lead to the positive side (anode) of the diode and the black lead to the
negative side (cathode). The meter displays the diode voltage to approximately 0.4 ~
0.9V for good junction.
4. Reverse the connections and measure the voltage across the diode again. The
message “OL" on the display corresponds to a correct junction.
5. The continuity test is always active and the test is performed using the test leads in the
same way of resistance measurement (refer to paragraph 4.3.6). The buzzer is on for
resistance values <500.
EN - 15
HT32
4.3.8. Capacitance measurement
Before taking any in circuit or capacitance measurement, remove power
from the circuit being tested and discharge all capacitors. Use the short test
lead pair for measurement in order to reduce the stray capacitance. Before
connecting the test capacitor, note the display, which may show a reading
other than zero whenever the range is changed. Subtract this offset reading
from the test result of a capacitor to obtain the true value. Connect the test
capacitor to the input clamps noting the polarity connections when required.
CAUTION
Fig. 9: Capacitance measurement
1. Turn the selector on .
2. Insert the test leads into the jacks, the red plug into HzVA jack and the black
plug into COM jack (Fig. 9).
3. Connect the test leads to the capacitor terminals taking care of the shown polarity. The
capacitance value will be displayed with automatic detection of the range.
4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from
display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.
5. The message "OL" means that the capacitance exceeds the measuring limits. In this
case disconnect the test leads from the circuit under test to avoid damaging the
instrument or endanger your own safety.
6. For minimum and maximum value measurement and HOLD function please refer to
paragraph 4.2.
EN - 16
HT32
5. MAINTENANCE
5.1. GENERAL INFORMATION
1. This is a precision instrument. To guarantee its performances be sure to use it according
to these instructions and keep it stored on suitable environmental conditions.
2. Do not expose it to high temperatures or humidity or direct sunlight.
3. Be sure to turn it off after use. If you expect not to use the instrument for a long period
remove batteries to avoid leakages of battery liquid which could damage the its inner
components.
5.2. BATTERY REPLACEMENT
When the symbol "" is displayed, batteries need replacement.
Before replacing batteries disconnect the test leads from any energized
circuits to avoid electrical shocks.
CAUTION
LEGEND:
1. Turn OFF the meter and
disconnect the test leads from the
input terminals.
2. Remove the protective holster from
the meter.
3. Unscrew the battery cover and
remove the battery.
4. Insert a new battery of the same
type (9V NEDA1604, JIS006P,
IEC6F22) observing the proper
polarity, re-screw the battery cover
and reposition the protective
holster.
Fig. 10: Battery replacement
EN - 17
HT32
5.3. FUSE REPLACEMENT
Before replacing fuses, disconnect test leads from any energized circuit to
avoid electrical shock.
CAUTION
LEGEND:
1. Turn OFF the meter and
disconnect the test leads from the
input terminals.
2. Remove the protective holster from
the meter.
3. Unscrew the battery cover and
remove the battery.
4. Remove the defective fuse and
install a new fuse of the same size
and rating (fast type 10A/1000V).
Make sure the new fuse is
centered in the fuse holder. Rescrew the battery cover and
reposition the protective holster.
Fig. 11: Fuse replacement
5.4. CLEANING
To clean the instrument use a soft dry cloth. Never use a wet cloth, solvents or water.
5.5. END OF LIFE
CAUTION: this symbol indicates that equipment and its accessories shall be
subject to a separate collection and correct disposal.
EN - 18
HT32
6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL FEATURES
The accuracy is indicated as [% of reading + number of digits]. It is referred to the
following environmental conditions: temperature 23°C ± 5°C, relative humidity < 80%.
DC Voltage
Range Resolution Accuracy Input impedance Overload protection
600.0mV 0.1mV
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
1000V 1V
(0.5rdg+2dgt)
10M //
less of 100pF
AC TRMS Voltage
Range Resolution
Accuracy (*)
(50 500Hz)
Input impedance
600.0mV 0.1mV
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
(0.9rdg+5dgt)
10M //
less then100pF
750V 1V
(*) NOTES
Sinusoidal waveform 50/60Hz only for 600.0mV range
Basic accuracy is specified for sinusoidal waveform up to 4000 counts. Over 4000 counts add 0.6%rdg to the
accuracy
For non-sinusoidal waveform below 2000 counts add 1.5%rdg for crest factor from 1.4 to 3
DC Current
Range Resolution Accuracy Output voltage Overload protection
1000VDC
750VACrms
Overload
protection
1000VDC
750VACrms
600.0A 0.1A
6000A 1A
6.000A 1mA
10A 10mA
(1.0rdg+2dgt)
<4mV/A
2V max
AC TRMS Current
Range Resolution
6.000A 1mA
10A 10mA
Accuracy
(50 500Hz)
(1.5rdg+5dgt)
Output voltage Overload protection
2V max
Frequency
Range Resolution Accuracy
6000Hz 1Hz
60.00kHz 0.01kHz
600.0kHz 0.1kHz
6.000MHz 0.001MHz 250mVrms
60.00MHz 0.01MHz
(*) for measurements < 20Hz sensitivity is 1.5Vrms
(**) Max sensitivity: <5VACrms
(0.1rdg+1dgt)
Sensitivity (**) Overload protection
100mVrms (*)
1Vrms
600Vrms
Fast fuse
10A / 1000V
Fast fuse 10A /
1000V
600Vrms
EN - 19
HT32
Resistance
Range Resolution Accuracy Open voltage Overload protection
600.00.1
6.000k0.001k
60.00k0.01k
600.0k0.1k
6.000M0.001M(1.0rdg+2dgt)
60.00M0.01M(1.5rdg+2dgt)
(0.7rdg+2dgt)
1.3V 600V rms
Diode test
Range Resolution
Accuracy
(0.4 0.8V)
Test
current
Open
voltage
Overload protection
10mV
(1.5rdg+5dgt)
1.5mA <3V 600V rms
Continuity test
Range Buzzer Overload protection
R<500
600V rms
Capacitance
Range Resolution Accuracy Overload protection
6.000nF 0.001nF
60.00nF 0.01nF
600.0nF 0.1nF
6.000F 0.001F
60.00F 0.01F
600.0F 0.1F
6.000mF 0.001mF
(1.9rdg+8dgt)
600V rms
6.1.1. Electrical specifications
Conversion: TRMS
Measuring rate: 1.5 times per second
Temperature coefficient: 0.15×(accuracy)/°C (<18°C and >28°C)
NMRR Normal Mode Rejection Ratio: > 50dB for DC parameters and 50/60Hz
CMRR Common Mode Rejection Ratio: >100dB from DC up to 60Hz on DCV
> 60dB from DC up to 60Hz on ACV
6.1.2. Safety
The instrument complies with: EN 61010-1
Insulation: class 2, double insulation
Pollution degree: 2
Overvoltage category: CAT IV 600V, CAT III 1000V (V//A)
CAT IV 500V (A)
For indoor use, max height: 2000m
Battery type: 9V NEDA1604, JIS006P, IEC6F22.
Low battery indication: "" is displayed when battery voltage is too low.
Battery life: about 300 hours.
AutoPowerOFF: 10 minutes after last pressure on keys or rotation
of selector.
Display
Specifications: 4 Digits with max. reading 6000 counts + symbol
This instrument is guaranteed for one year against material or production defects, in
accordance with our general sales conditions. During the warranty period the
manufacturer reserves the right to decide either to repair or replace the product.
Should you need for any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault). Use only original
packaging. Any damage occurred in transit due to non original packaging will be charged
anyhow to the customer.
The warranty doesn’t apply to:
Accessories and batteries (not covered by warranty).
Repairs made necessary by improper use (including adaptation to particular
applications not foreseen in the instructions manual) or improper combination with
incompatible accessories or equipment.
Repairs made necessary by improper shipping material causing damages in transit.
Repairs made necessary by previous attempts for repair carried out by non skilled or
unauthorized personnel.
Instruments for whatever reason modified by the customer himself without explicit
authorization of our Technical Dept.
Use not provided by the instrument's specifications or in the instruction manual.
The contents of this manual may not be reproduced in any form whatsoever without the
manufacturer’s authorization.
Our products are patented and our logotypes registered. We reserve the right to
modify specifications and prices in view of technological improvements or
developments which might be necessary.
7.2. SERVICE
Shouldn’t the instrument work properly, before contacting your distributor make sure that
batteries are correctly installed and working, check the test leads and replace them if
necessary.
Should you need for any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault). Use only original
packaging. Any damage occurred in transit due to non original packaging will be charged
anyhow to the customer.
The manufacturer will not be responsible for any damage to persons or things.
EN - 22
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.