HT instruments HT204 User Manual

ENGLISH
User’s manual
© Copyright HT ITALIA 2012 Release EN 1.01 - 09/02/2012
HT204
1. SAFETY PRECAUTIONS AND PROCEDURES ............................................................. 2
1.1. Preliminary ............................................................................................................ 2
1.2. During use ............................................................................................................. 2
1.3. After use ................................................................................................................ 2
2. GENERAL DESCRIPTION .............................................................................................. 3
2.1. Features ................................................................................................................ 3
2.2. Instrument description ........................................................................................... 3
2.3. Description of function keys ................................................................................... 3
2.3.1. ON/OFF key .................................................................................................................. 3
2.3.2. M-H key ......................................................................................................................... 3
2.3.3. W/B key ......................................................................................................................... 4
2.3.4. D-H key ......................................................................................................................... 4
2.3.5. R key ............................................................................................................................. 4
3. PREPARATION FOR USE .............................................................................................. 4
3.1. Initial ...................................................................................................................... 4
3.2. Power supply ......................................................................................................... 4
3.3. Calibration ............................................................................................................. 4
3.4. Storage .................................................................................................................. 4
4. OPERATING INSTRUCTIONS ........................................................................................ 5
4.1. Measurement description ...................................................................................... 5
5. PREVENTIVE MAINTENANCE ....................................................................................... 5
5.1. General information ............................................................................................... 5
5.2. Battery replacement .............................................................................................. 5
5.3. Cleaning ................................................................................................................ 5
5.4. End of life .............................................................................................................. 5
6. TECHNICAL SPECIFICATIONS ..................................................................................... 6
6.1. Characteristics ....................................................................................................... 6
6.1.1. General data .................................................................................................................. 6
6.2. Environment .......................................................................................................... 6
6.2.1. Environmental conditions .............................................................................................. 6
6.3. Accessories ........................................................................................................... 6
6.3.1. Standard accessories .................................................................................................... 6
7. SERVICE ......................................................................................................................... 7
7.1. Warranty conditions ............................................................................................... 7
7.2. Service .................................................................................................................. 7
EN - 1
HT204
1. SAFETY PRECAUTIONS AND PROCEDURES
This meter is in compliance with safety standard EN 61010-1 related to electronic measuring instruments. For your own safety and to avoid damaging the instrument follow the procedures described in this instruction manual and read carefully all notes preceded
by this symbol . When taking measurements:
Avoid doing that in humid or wet places
Avoid doing that in rooms where explosive gas, combustible gas, steam or excessive
dust is present
Avoid doing that if you notice anomalous conditions such as breakages, deformations, fractures, leakages of battery liquid, blind display etc
The following symbol is used:
Caution: refer to the instruction manual. An incorrect use may damage the luxmeter or its components.
1.1. PRELIMINARY
Before turning on the instrument make sure that the battery is correctly installed.
Don’t touch the sensor during measurements to prevent any damage due to
1.2. DURING USE
Always keep to the instructions contained in this manual.
If the reading value or the sign indication remains unchanged during the
1.3. AFTER USE
Replace the battery as soon as the low battery indication (“ ”) is displayed contamination or static electricity
CAUTION
An improper use may damage the instrument and/or its components or injure the operator.
measurement, check if the DATA HOLD function is active
After using the instrument turn it off
If you expect not to use the instrument for a long period remove the battery to avoid
leakages of battery liquids which may damage its inner components
EN - 2
HT204
2. GENERAL DESCRIPTION
The digital solar power meter HT204 is used to measure the sunlight. Possible applications are the evaluation of the solar power in industrial installations (photovoltaic plants) or its danger level for people’s skin. The sunlight is usually expressed in W/m2 or BTU/(ft2*h).
2.1. FEATURES
Sunlight measurement up to 1999W/m2 / 634BTU/(ft2*h)
High accuracy and rapid response
DATA HOLD function to hold measuring values
Unit and sign display for easy reading
Measuring unit selection among W/m2 and BTU/(ft2*h)
Manual scale selection
Direct reading with no adjustments needed
Maximum and minimum values
Low battery indication
2.2. INSTRUMENT DESCRIPTION
LEGEND:
1. DATA HOLD active
2. MAX active
3. MIN active
4. Decimal point
5. Measured value
6. Measuring unit W/m2
7. Measuring unit BTU/(ft2*h)
8. Low battery
LEGEND:
1. ON/OFF key
2. M-H key to select the maximum, minimum or current reading
3. W/B key to select the measuring unit
2
W/m
4. D-H key to hold displayed data
or BTU/(ft2*h)
5. R key to select the full scale manually
Fig. 1: Instrument description
2.3. DESCRIPTION OF FUNCTION KEYS
2.3.1. ON/OFF key
Press the key to switch on and off the instrument.
2.3.2. M-H key
By pressing key, maximum and minimum values are measured. The symbol corresponding to the desired function is displayed: “MAX” for maximum value, “MIN” for minimum value. To exit this function keep M-H key pressed for at least 1 second.
EN - 3
HT204
2.3.3. W/B key
By pressing key the manual selection of measuring unit is possible. The cyclically pressure of the key changes the measuring unit among W/m2 and BTU/(ft2*h).
2.3.4. D-H key
By pressing key the measured value is frozen on the display where the symbol "HOLD" appears. Press again D-H to disable this function and resume normal operation.
2.3.5. R key
By pressing key the manual selection of measured range cyclically pressure of the key changes the measuring range and fix the decimal point on the display.
3. PREPARATION FOR USE
3.1. INITIAL
This instrument was checked both mechanically and electrically prior to shipment. All possible cares and precautions were taken to let you receive the instrument in perfect conditions. Notwithstanding we suggest you to check it rapidly (eventual damages may have occurred during transport – if so please contact the local distributor from whom you bought the item).
Make sure that all standard accessories mentioned in paragraph 6.3.1 are included. Should you have to return back the instrument for any reason please follow the instructions mentioned in paragraph 7.
3.2. POWER SUPPLY
The instrument is powered by batteries (refer to paragraph 6.1.1 for further details on model, number and battery life). The battery charge is displayed on the right top side. Replace it immediately, following the instructions given in paragraph 5.2.
3.3. CALIBRATION
The instrument complies with the technical specifications contained in this manual and such compliance is guaranteed for 1 year.
The calibration interval for the photo detector varies according to the operating conditions, but generally the sensitivity decreases in direct proportion to the product of luminous intensity and operating time. In order to maintain the basic accuracy of the instrument, a periodic recalibration is recommended.
3.4. STORAGE
After a period of storage in extreme environmental conditions exceeding the limits mentioned in paragraph 6.2.1 let the instrument resume normal measuring conditions before using it.
EN - 4
HT204
4. OPERATING INSTRUCTIONS
4.1. MEASUREMENT DESCRIPTION
1. Press the power key to turn on the meter
2. Press the key to select W/m2 or BTU/(ft2*h) measurement
3. Remove the protection cap of the photo detector and expose it to the light source in horizontal position. Read the sunlight value on the LCD display
4. Wait for values to be stable on the display. Press “D-H” key to activate the DATA HOLD function blocking the result on the display
CAUTION
If the instrument displays "OL", the input signal is too strong. A higher range must be selected.
5. When the measurement is completed, fit the photo detector cap and check that the indication value should be “000” regardless of the range. If no, adjust the “ADJ” trimmer on “000” before pressing the power key to turn off the meter
5. PREVENTIVE MAINTENANCE
5.1. GENERAL INFORMATION
This is a precision ins trum ent. To guarantee its performances be sure to use it according to these instructions and keep it stored on suitable environmental conditions Do n ot e xpo se i t to high temperatures or humidity. Be sure to turn it off after use. If you expect not to use the instrument for a long period remove batteries to avoid leakages of battery liquid which could damage the its inner components.
5.2. BATTERY REPLACEMENT
When the low battery indication “ ” is displayed, replace the battery.
1. Switch off the instrument
2. Press the battery cover and push in the direction of the arrow to open
3. Replace the battery with a new one of the same type (9V 6LR61)
4. Replace the battery cover
5.3. CLEANING
To clean the instrument use a soft dry cloth. Never use a wet cloth, solvents or water, etc. The white plastic lens of the detector should be cleaned with a damp cloth when necessary.
5.4. END OF LIFE
CAUTION: this symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal.
EN - 5
HT204
6. TECHNICAL SPECIFICATIONS
6.1. CHARACTERISTICS
Accuracy is referred to the following reference conditions: 23 ± 5°C with RH < 70%. Uncertainty: ± 10 W/m2 [ ± 3 BTU/(ft2*h) ] or ± 5%, whichever is greater
Temperature added uncertainty: ± 0.38 W/m2/°C [ ± 0.12 BTU/(ft2*h)/°C] from 25°C Resolution: 1 W/m2 [ 1 BTU/(ft2*h) ] Range: 1 ÷ 1999 W/m2 [ 1 ÷ 634 BTU/(ft2*h) ]
6.1.1. General data
Mechanical characteristics
Size: 172(L) x 60(W) x 38(H)mm; 6,8 x 2,4 x 1,5in Weight (including battery): About 150g; 5,3ounces
Supply
Battery type: 1 battery 9V type 6LR61 Low battery indication: " " is displayed when the battery is low Battery life: About 100 hours (carbon zinc.)
Display
Characteristics: 3-1/2 digit LCD, 2000 counts with “OL” indication Sample rate: 4 times/sec
6.2. ENVIRONMENT
6.2.1. Environmental conditions
Reference temperature: 23° ± 5°C; 73° ± 41°F Operating temperature: 5° ÷ 40°C; 41° ÷ 104°F Operating humidity: <80%RH Storage temperature: -10° ÷ 60°C; 14° ÷ 140°F Storage humidity: <70%RH Internal use: Max. 2000m (6561ft), pollution degree 2
This instrument complies with the requirements of the European Directive:
EN61326 (1997) + A1 (1998) + A2 (2001)
6.3. ACCESSORIES
6.3.1. Standard accessories
The packaging contains:
Instrument HT204
Carrying case
Instruction manual
EN - 6
HT204
7. SERVICE
7.1. WARRANTY CONDITIONS
This instrument is guaranteed against material or production defects, in accordance with our general sales conditions. During the warranty period the manufacturer reserves the right to decide either to repair or replace the product.
Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it. Do not forget to enclose a report describing the reasons for returning (detected fault). Use only original packaging. Any damage occurred in transit due to non original packaging will be charged anyhow to the customer. The manufacturer will not be responsible for any damage to persons or things.
The warranty doesn’t apply to:
Accessories and batteries (not covered by warranty)
Repairs made necessary by improper use (including adaptation to particular
applications not foreseen in the instructions manual) or improper combination with incompatible accessories or equipment
Repairs made necessary by improper shipping material causing damages in transit
Repairs made necessary by previous attempts for repair carried out by non skilled or
unauthorized personnel
Instruments for whatever reason modified by the customer himself without explicit
authorization of our Technical Dept
The contents of this manual may not be reproduced in any form whatsoever without the manufacturer’s authorization.
Our products are patented and our logotypes registered. We reserve the right to modify specifications and prices in view of technological improvements or developments which might be necessary.
7.2. SERVICE
Shouldn’t the instrument work properly, before contacting your distributor make sure that the battery is correctly installed and working, check the test leads and replace them if necessary. Make sure that your operating procedure corresponds to the one described in this manual.
Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it. Do not forget to enclose a report describing the reasons for returning (detected fault). Use only original packaging. Any damage occurred in transit due to non original packaging will be charged anyhow to the customer. The manufacturer will not be responsible for any damage to persons or things.