HTA 77596, 77597 Operation Manual

OPERATION MANUAL
PORTABLE IAQ METER
Model:
77597, 4 in 1 77596, 3 in 1
INTRODUCTION
Thank you for purchasing this portable IAQ meter. The meter measures CO
2
level,CO level, air temp., dew point, wet bulb temp. and humidity(DP, WB, RH are for models 77597 only) and is an ideal instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, especially on CO
2
level and air ventilation become widely applied in public areas such as offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of industrial hygiene in some countries. (Appendix)
The portable CO meter uses NDIR (non-
2
dispersive infrared) technology to ensure the reliability and long term stability. It’s useful in verifying HVAC system perform­ance and air ventilation control.
CO is a colorless, odorless, and tasteless gas which is slightly lighter than air. It is highly toxic to humans and animals.
Maximum indoor air quality level per ASHRAE Residential standards 62-1989 for living area is under 9ppm. Besides, it is also necessary to remove employees from enclosed space if the CO concentration exceeds100ppm per OSHA exposure limit. Please refer to Appendix for the details.
Features:
Big LCD display with blue backlight to use in dark area One touch to display CO2 /Dew point temperature/Wet bulb temperature/ Air temperature/Humidity/CO in turns. Designed with NDIR (non-dispersive infrared) Waveguide technology CO2 sensor Programmable warning CO level Long time drift compensation on CO2 sensor Audible alarm (~80db)threshold setting Max, Min, TWA, STEL functions included Mini USB to PC download function for analysis 99 points manually recording 32000/24000 points auto recording Review 99 points manual record Hold function freezes current readings Housing design surrounds with rails to help Air ventilation for quick & accurate response Easy to manually calibrate CO2 at fresh air around 380-420ppm
MATERIAL SUPPLIED
Meter 4pcs AA batteries Mini USB cable and software CD Operation manual Hard carrying case
This package contains:
Optional accessory:
33% calibration salt (VZ0033AZ1) 75% calibration salt (VZ0075AZ1) Adaptor (9V/100~240Vac)
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(9V/1A output).
Install the batteries into the battery compartment on the rear and make sure they are in correct polarity and good contact.
will
appear on the LCD .
atter
When an adaptor is used, it will cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low,
(Fig.1)
The CO sensor can’t work normally
2
under low voltage, please replace with fresh b ies or connect with an adaptor.
B
CO2
PM
B
Fig.1
HOLDHOLD
CO2
%rhrh
TA
DP
WBT
OO
FF
OO
CC
REC
Recall
AM
PM
CO
CAL
B
MINMIN MAXMAX AVGAVG
LCD DISPLAY
Upper LCD
MIN/MAX Minimum/Maximun readings STEL Short-term exposure limit (15 minutes weighted average) TWA Time weighted average(8 hours) Low battery indicator DP Dew point TA Air temperature WBT Wet bulb %rh Unit of relative humidity (77597)
o
C/ F Celsius/Fahrenheit of temp. CAL In calibration status REC In manual/automatic logging RECALL In manual records recall mode
CO2/CO/Temp....reading
Lower LCD Real time display
CO2 Carbon dioxide reading CO Carbon monoxide reading
HOLD Readings are freezed unchanged
temperature(77597)
temperature(77597)
o
Symbols
Turns on and off the meter. Enters setup mode while meter is off. Sets as non-sleep mode with .
SET
START ESC
Exits setup/recall page. Start automatic logging.
KEYPAD
ON/OFF
MODE RECALL
HOLD
Press to switch displayed mode. Long press to enter memory recall mode.
TWATWA STELSTEL
B
Freezes the current readings. Selects unit or increases value in setup. Cancels data hold function.
Press to manually record the reading. Selects unit or decreases value in setup.
Activates MIN, MAX, TWA, STEL function. Saves and finishes settings.
OPERATION
Press to turn the meter on and off. At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO and real time
displayed (Fig.3). The real time display date and time in turns.
Suggest to wait for 30 seconds warm up time. For quick start, you may press for 2 seconds to end the warm up and entering normal display.
POWER ON/OFF
SET
ON/OFF
Fig.2
Fig.3
CO2
AM
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
2
NOTE: Do not hold the meter close to faces in case exhalation affects CO
2
levels.
N X
M /AVG
HOLD
MEM
N X
M /AVG
Press to switch to temperatures display.(Fig.6). In 77597 model, dew point temp. and wet bulb temp. are also available.(Fig.7). The lower display shows the real time clock.
AIR(all models),DP,WBT(77597)
Press to switch the mode to CO2. User can get the CO2 reading in ppm unit on main display.(Fig.4). The lower display shows the real time clock.
CO2 (Carbon Dioxide)
MODE RECALL
CO2
AM
Fig.4
Press to switch the mode to CO. User can get the CO reading in ppm unit on main display.(Fig.5). The lower display shows the real time clock.
CO (Carbon Monoxide)
MODE RECALL
CO
AM
Fig.5
MODE RECALL
TA
OO
CC
PM
WBT
OO
CC
Fig.6
Fig.7
Press to switch to humidity display. (Fig.8).The lower display shows the real time clock.
Humidity (77597)
MODE RECALL
%rhrh
Fig.8
DATA HOLD In normal display mode, press to freeze the readings, "HOLD” icon is flashed on the left top of LCD(Fig.9). All
current readings are kept unchanged. Press again to cancel the hold function.
Fig.9
BACKLIGHT
The backlight will be activated for 10
seconds by pressing any key.
MIN,MAX,STEL,TWA, CURRENT REVIEW
This meter allows you to check the
minimun, maximun, STEL,TWA and
current value from the moment you press
the key. Under normal mode, press to see the minimum, maximum, STEL, TWA and current value in turns. Each press of ,it displays MIN, MAX, STEL,TWA, current value in sequence. (Fig.10~12). To return to normal mode, press key to escape(Fig.13)
If the meter is turned on for shorter than 15 mins, the STEL value will be the weighted average of readings take since power on. Same for TWA values appear before 8 hours.
It takes at least 5 mins to calculate STEL and TWA. The display shows “----” during the first 5 mins from power on.
HOLD
HOLD
HOLDHOLD
%rhrh
AM
NOTE: THE HOLD FUNCTION IS DISABLE IN MIN/MAX/ AVG MODE.
N X
M /AVG
N X
M /AVG
N X
M /AVG
START ESC
In this max/min/STEL/TWA/current modes, it shows the corresponding readings on main display and the accumulated time of how long you are in this mode.(Fig10~12).
NOTE: “HOLD” & “MANUALLY RECORD” & “RECALL” FUNCTION ARE DISABLE IN MIN/MAX/AVG MODE.
MODE RECALL
CO
MAXMAX
Fig.14
START ESC
While in “HOLD” mode, STEL and TWA will remain updating every 5 mins.
CO2
MINMIN
CO2
CO2
CO2
AM
Fig.10
Fig.11
Fig.12
Fig.13
TWATWA
In this mode, press to switch the displayed parameters. Fig 14.
The dispalyed sequence in this mode is min max STEL TWA current min
max.....Press key can immediately
leave this mode and return to normal. Distinguishing the normal display and
current review by checking the time display format. In normal display, the date & time displayed in turns. In current review, the time display is the accumu- lated time of how long you are in max/min mode.
NOTE: TWA &STEL ARE FOR CO2 ONLY. MIN/MAX ARE FOR ALL PARAMETERS.
parameter you choose to see on main display.
TA
REC
AM
Fig.15
If the reading changes quickly, you can press to freeze the reading before manually recording the data.
HOLD
99 MEMORIES RECALL
The meter features 99 point memories
review function.
In normal or hold mode, press key
for more than 2 seconds until the
icon flashes.
Press or to scroll the memories.
The memory serial number displays on
main LCD fist and reading comes after.
(Fig.16&17)
MODE RECALL
Recall
Recall
CO2
Recall
PM
HOLD
MEM
Fig.16 Fig.17
MANUALLY RECORDING
The meter features 99 point memories. In normal or hold mode, Press key
to record, icon and main display flash for about 3 seconds. The main display shows the memory serial number , at most 99 points.(Fig. 15)
Each memory contains all parameters
(CO2, CO, TA.....%rh), not only the
MEM
REC
The manually recording function is disabled in min/max mode.
ALARM
The meter features audible alarm to
give warnings when CO concentration
exceeds the limit. (See P20 in setup
for setting alarm threshold). It emits
beeps(Abt.80dB)when CO level goes 
over the set value and stops only when
the readings fall below the set value. It
beeps again when value goes over the
limit.
To escape the memory recall mode, press key to leave and return to normal display.
START ESC
DATA LOGGING
The meter can automatically record readings of CO2/CO/TEMP./RH for long time environment monitoring. The memory capacity is 32000 (#77597) or 24000 (#77596) points. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds (See SETUP P40).
After sampling rate is selected, press for 2 seconds under normal mode to start logging. The icon flashes to indicate the logging status and LCD main display shows the real time CO2 value. Lower displays are the real time clock.(Fig.19)
START ESC
REC
10
MODE RECALL
CO
Recall
PM
Fig.18
The time displayed in memory recall mode is the recording time of this memory.
Press key to switch the display parameter.(Fig.18)
CO2
REC
Fig.19
To terminate data logging, press for 2 seconds, icon stops flashing
While press to start the logging again, the previous logging records are overwrite.
Press key to switch to the mode you want to see. Except the mode switch function, min/max, manual record, hold and recall functions are all disabled during logging.
START ESC
REC
START ESC
MODE RECALL
The meter turns off automatically after 20 minutes of inactivity. To override the function, hold down and for 2 seconds to turn on the meter until “n” appears.(Fig.20)
NOTE: AUTO SLEEP FUNCTION WILL BE DISABLED DURING CALIBRATION MODE.
AUTO POWER OFF
SET
ON/OFF
HOLD
When the meter is off, hold down for more than 1 sec to enter setup mode. To exit setup, press in P10 ~ P60 and it returns to normal mode. (Fig.21)
SETUP
Fig.20 Fig.21
SET
ON/OFF
START ESC
11
P20: CO ALARM
12
P10: 99 MEMORIES CLEAR
When entering setup mode, P10 and “CLr” (Fig.21) are displayed on the LCD. Press to go into P11 and decide to clear or keep all the manual records. The current set will be blinking on LCD (Fig.22).
Press or to choose NO or YES and press to confirm.
N X
M /AVG
HOLD
MEM
N X
M /AVG
Fig.22
Press to escape and return to P10.
START ESC
When entering setup mode, P20 and “ALAr” (Fig.23) are displayed on LCD. Press to go into P21 for setting CO alarm threshold. The current set value will be blinking on LCD(Fig.24). Press to increase the value or to decrease. The selectable alarm limits are 15~200ppm,each press tunes 5ppm. When the preferred alarm value is set, press to save the setting or without saving and return to P20.
N X
M /AVG
HOLD
MEM
N X
M /AVG
START ESC
CO
CO
Fig.23 Fig.24
P40:LOGGING SAMPLING RATE
P30 TEMPERATURE UNIT
When entering setup mode, P30 and “unit” (Fig.25) are displayed on the LCD. Press to go into P31 and decide the temperature unit.
The current set will display on LCD (Fig.26).
o o
Press or to choose C or F and press to confirm.
N X
M /AVG
HOLD
MEM
N X
M /AVG
Fig.25
Press to escape and return to P30.
START ESC
Fig.27 Fig.28
Fig.26
TA
OO
CC
TA
Go into P40 for setting sampling rate of data logging(Fig.27). The range is from 1 second to 4 hours 59 minutes and 59 seconds.
Press and it goes into setting with blinking Hour digits on the lower display. To change the digit, press to increase and to decrease. Press to confirm and enter Minute setting . Press again to confirm and enter Second setting (Fig.28). Press to confirm the rate setting or without saving and return to P40.
N X
M /AVG
HOLD
N X
M /AVG
N X
M /AVG
START ESC
13
MEM
N X
M /AVG
P60:REAL TIME CLOCK
P50 PRESSURE COMPENSATION
When entering setup mode, P50 and “PrES” (Fig.29) are displayed on the LCD. Press to set pressure compensation value for CO2 measurement.
The current set will flash on LCD (Fig.30).The barometric pressure unit is kpa.
Press or to adjust the pressure value and press to confirm.
N X
M /AVG
HOLD
MEM
N X
M /AVG
Fig.29
Press to escape and return to P50.
START ESC
Fig.31
Fig.32
Fig.30
Go into P60 for setting the real time clock of this meter.
Press and it goes into P61to set the time format as 12hour or 24 hour. The current set will flash on LCD. Press or to change the format and press to confirm and entering real time clock setting.(Fig.31)
The blinking Year digits on the lower display. To change the digit, press or to adjust. Press to confirm and enter Month setting. Repeat above steps to complete the month/date/hour/ minute/second setting.(Fig.32)
N X
M /AVG
HOLD
MEM
N X
M /AVG
N X
M /AVG
CO2
CO2
HOLD
MEM
PM
14
15
CO CALIBRATION
2
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration.
2
Otherwise, it will be calibrated as 400ppm by default and leads to inaccurate measurements.
NOTE:
Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration.
The manual calibration is suggested to to be done in fresh outdoor air that is well ventilated and in sunny weather.
Place the meter in the calibration site. First press + , then also simultaneously for 3 seconds to turn on the meter and enter calibration mode (Fig.33). Three calibration menu is available: Humidity/CO/CO2.
Press key to select the CO2 calibration. 400ppm and “ “ icon are blinking on the LCD. Press to start calibration. (Fig.34)
Wait about 10 minutes until the blinking stops and the calibration is completed automatically. To abort the calibration, turn off the meter at any time.
The meter is calibrated at standard 400ppm CO concentration in factory.
2
It’s suggested to do manual calibration regularly to maintain good accuracy.
Note:
When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.
START ESC
MEM
SET
ON/OFF
N X
M /AVG
CAL
N X
M /AVG
%rh%rh
CAL
CO2
CAL
Fig.33
Fig.34
33% calibration
Plug the sensor probe into 33% salt bottle. First p
Hold down to enter 33% calibration. ¡§CAL¡¨ and calibrating value (32.8% if at are blinking on the LCD.
Meter is now calibrating, and will finish in about 60 minutes when ¡§CAL¡¨ and humidity stop blinking.
ress + , also simultaneously for 3 seconds to turn on the meter and enter calibration mode (Fig.33.).
o
25 C)
To abort the cali-
bration, turn off the meter at any time.
The meter defaults to be calibrated the humidity with 33% and 75% salt solution. The ambient condition is recommended
o
to be at 25 C and stable humidity(better to be close to the calibrating value). To abort calibration, just turn off the meter.
(77597)
RH CALIBRATION
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact the dealer for calibration salt or services.
START ESC
MEM
SET
ON/OFF
N X
M /AVG
75% calibration
Plug sensor probe into 75% salt bottle and entering calibration mode. In calibration mode, press to select 75% calibration(Fig.35), press to start.
HOLD
N X
M /AVG
16
CAL
17
“CAL” and calibrating value (75.3% if at
o
25 C) are blinking on the LCD. Meter is now calibrating, and will finish
in about 60 minutes when “CAL” and humidity stop blinking. To abort the cali­bration, turn off the meter at any time.
%rh%rh
Fig.35
CO CALIBRATION
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration. Otherwise, it will be calibrated as 0 or 400ppm by default and leads to inaccurate measurements.
When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration. It is necessary to have standard CO gas for calibration.
Place the meter in the calibration site. First press + , then also simultaneously for 3 seconds to turn on the meter and enter calibration mode (Fig.33).
Press or to select the CO 0ppm or 400ppm calibration. ppm and “ “ icon are blinking on the LCD. Press to start calibration. (Fig.36&37)
Wait about 10 minutes until the blinking stops and the calibration is completed automatically. To abort the calibration, turn off the meter at any time.
START ESC
MEM
SET
ON/OFF
HOLD
MEM
N X
M /AVG
CAL
? Error messages
E01/E33: CO sensor is out of order. Try
2
new battery first. E02: The value is under range. E03: The value is over range. E04: The original data error results in this error (DP, WB) E11: Retry humidity calibration. E16: Retry CO calibration. E31: Temperature sensor or AD damaged. E32: Memory IC damaged. E33: Humidity sensor or circuit damaged.
? Can’t power on
-Press for more than 0.3 seconds and try again. Check whether batteries are in good contact and correct polarity, or the adaptor is well plugged.
-You may also take off the batteries for 10 mins and then put back again.
-Open the battery cover and use a small metal tip to touch and reset the meter.
TROUBLESHOOTING
? Fixed readings
Check whether data hold function was activated. (HOLD icon at the left top)
? Slow response
Check whether the air flow channels on the rear were blocked.
CO
CAL
SET
ON/OFF
NOTE: It is allowed to adjust the CO calibration point from 385~415ppm (400+15ppm)
CO
CAL
Fig.36
Fig.37
18
Reset point
19
PC CONNECTION
Connect the logger to PC, users can do logging setting and data transmission with the included software.
Plug the USB cable into the socket at the right side of the meter and the other port to PC. Then install the software in your PC with the following procedure.
1.Insert the CD Rom and run installation. Select a preferred directory and click “Next” step by step and finish it.
2. After the software installation is completed, it will run USB driver installation next.
20
Software Operation
Start the software and it will detect logger connection automatically and indicate the COM Port information at the bottom of the main screen as well as the COMPort setting column.
Logger Settings
To set up logging plan, click “Setting” icon and select “Logger”. The setting page is opened.
Users can set up the logger identifier number. Click OK for setting and Exit the screen to confirm the settings.
AutoConnect
Data Transmission
To transmit auto recorded data from the meter, click “Download” icon on the left top side of the main screen. All auto­recorded data in the logger will be transmitted. The raw data with time stamp will display at the lower part of the screen and the graph in the middle.
In the Graph display, the Y-axis indicates Temp, RH,DP, WBT, CO2&CO level in different line colors. And the X-axis can be switched to show Time or Recorded No. Stamp.
Data Review Function
The following functions help to view the GRAPH data in more detailed way.
Under View function, there are 4 tools to enlarge the Graph data for detailed data review.
Zoom in: Hold down “Ctrl” key on the PC keyboard and click any target point on the graph. It enlarges the point in each click.
21
22
Zoom Xaxis: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the X-axis of the selected area will be enlarged.
Zoom Yaxis: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the Y-axis of the selected area will be enlarged.
Three icons on the right top side of the main screen:
Redraw: After any review and zooming on the Graph, click “Redraw” to reset it to original format.
Clear: Click “Clear” to delete all data and Graph on the window. A warning box shows up for confirmation. Press “Yes” and all data will be cleared.
Cursor: Switch Cursor ON and it shows
the raw data with the cursor. In any click on the graph, the digital data displayed. Switch OFF to disable the function.
Zoom window: Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the selected area will be enlarged .
Data Processing The downloaded data can be Saved or
Printed via “File” processing functions. Load: To retrieve saved files, just click
“Load” and select a desired file and it will be loaded in the main screen with the file details at the left top side.
SPECIFICATION
23
Measuring range
o o
-20~60 C (-5~140 F)
CO2
Temperature
Accuracy CO2
Temperature
o o
+0.6 C/+0.9 F
30 seconds
CO2 Warm-up time Response time CO2
Tair
LCD / Meter size (mm) Operating condition
(avoid condensation) Storage condition Power supply
Weight Standard package
<30 seconds(90% step change)
<2 mins (90% step change)
26(H)x44(L), 205(L)x70(W)x56(H)
-20 to 50 °C (CO2 sensor),0 to 50 °C (CO sensor)
-20 to 60 °C (for the rest parameters)
o
-20~ 60 C,10~90%RH(avoid condensation) AA x 4pcs or 9V adaptor
200g
Meter, manual ,AA bat.,USB cable &CD, carry case
77597
77596
0~9999 ppm, 0~9999 ppm,
(5001~9999 ppm out of scale range)
+30ppm+5% of reading( 0~5000ppm)
Other ranges are not specified
Battery life
> 24 hours (Alkaline battery)
Resolution
o o
1ppm , 0.1 C/ F, 0.1%RH
Relative Humidity DP(Dew point temp.) WB(Wet bulb temp.)
0.1%~99.9%RH
o
-20.0~59.9 C
o
-5.0~59.9 C
Relative Humidity
o o
+3%RH (at 25 C, 10~90%RH) ; +5%RH ( at 25 C , other range )
RH
<10 minutes ( 90% step change)
(5001~9999 ppm out of scale range)
CO
0~1000 ppm, 0~1000 ppm,
N/A N/A N/A
CO
+/-10 ppm for less than 100 ppm +/-10% or reading for 101~500 ppm +/-20% of reading for 501ppm and above
CO
<60 seconds(90% step change)
24
CO2 LEVELS AND GUIDELINES
EPA Taiwan: 600ppm and 1000ppm Type 1 indoor areas such as department
stores, theaters, restaurants, libraries, the acceptable CO2 concentration of 8 hours avarge is 1000ppm. Type 2 indoor areas with special requirements of good air quality such as schools, hospitals, day care centers, the suggested CO2 level is 600ppm.
NIOSH recommendations 250-350 ppm: normal outdoor ambient
concentrations
600 ppm: minimal air quality complaints 600-1000 ppm: less clearly interpreted 1000 ppm: indicates inadequate ventilation;
complaints such as headaches, fatigue, and eye/throat irritation will be more widespread. 1000 ppm should be used as an upper limit for indoor levels.
Non-Enforced Reference levels
Regulatory exposure limit
ASHRAE Standard 62-1989: 1000ppm
CO2 concentration in occupied building should
not exceed 1000ppm. Building bulletin 101 (BB101):
UK standards for schools say that CO2 at
1500ppm
averaged over the whole day(i.e. 9am to 3.30 pm) should not exceed 1500ppm.
OSHA: 5000ppm Time weighted average over five 8-hour work days should not exceed 5000ppm.
Germany, Japan, Australia, UK...: 5000ppm 8 hours weighted average in occupational exposure limit is 5000ppm.
ppm
Symptoms and applicable standard
9
Maximum indoor air quality level: Maximum allowable concentration per ASHRAE Residential standards 62-1989 for living area.
25
Maximum limit 8 hrs of continuous exposure per California OSHA workplace standards.
35
Maximum 8 hrs average exposure level per US OSHA workplace standards.
50
Maximum concentration for continuous exposure in any 8 hrs average level per OSHA standards.
100
Remove employees from enclosed space if the CO concentration exceeds 100ppm per OSHA exposure limit.
400
Frontal headache, life threatening after 3hrs. Maximum concentration in flue gas per the US EPA and AGA standards.
1600
Nausea w/i 20 min.,death w/i 2-3 hrs
200
Mild headache, fatigue, nausea and dizziness w/i 2-3 hrs.
Dizziness, nausea, convulsions, death w/i 2-3 hrs.
0-1
Normal background levels
800
CO LEVELS AND GUIDELINES
25
26
WARRANTY
The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers normal operation and does not cover misuse, abuse, alteration, neglect, improper maintenance, or damage resulting from leaking batteries. Proof of purchase is required for warranty repairs. Warranty is void if the meter
has been opened.
RETURN AUTHORIZATION
OTHER RELATED PRODUCTS
Authorization must be obtained from the supplier before returning items for any reason. When requiring a RA (Return Authorization), please include data regarding the defective reason, the meters are to be returned along with good packing to prevent any damage in delivery and insured against
possible damage or loss.
Other related CO products:
2
a.Model 7721 Wall mount Temp./CO
2
meter, general purpose. b.Model 77231 Wall mount Temp./CO
2
meter, high performance. c.Model 7722 Wall mount Temperature/RH/
Co meter, general purpose.
2
d.Model 77232 Wall mount Temperature/RH/ Co meter, high performance.
2
Accuracy, the Zenith of Measuring / Testing Instruments !
More products available !
Hygrometer/Psychrometer Thermometer Anemometer Sound Level Meter Air Flow meter Infrared Thermometer K type Thermometer K.J.T. type Thermometer K.J.T.R.S.E. type Thermometer pH Meter Conductivity Meter T.D.S. Meter D.O. Meter Saccharimeter Manometer Tacho Meter Lux / Light Meter Moisture Meter Data logger Temp./RH transmitter