HTA 77596, 77597 Operation Manual

OPERATION MANUAL
PORTABLE IAQ METER
Model:
77597, 4 in 1 77596, 3 in 1
INTRODUCTION
Thank you for purchasing this portable IAQ meter. The meter measures CO
2
level,CO level, air temp., dew point, wet bulb temp. and humidity(DP, WB, RH are for models 77597 only) and is an ideal instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, especially on CO
2
level and air ventilation become widely applied in public areas such as offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of industrial hygiene in some countries. (Appendix)
The portable CO meter uses NDIR (non-
2
dispersive infrared) technology to ensure the reliability and long term stability. It’s useful in verifying HVAC system perform­ance and air ventilation control.
CO is a colorless, odorless, and tasteless gas which is slightly lighter than air. It is highly toxic to humans and animals.
Maximum indoor air quality level per ASHRAE Residential standards 62-1989 for living area is under 9ppm. Besides, it is also necessary to remove employees from enclosed space if the CO concentration exceeds100ppm per OSHA exposure limit. Please refer to Appendix for the details.
Features:
Big LCD display with blue backlight to use in dark area One touch to display CO2 /Dew point temperature/Wet bulb temperature/ Air temperature/Humidity/CO in turns. Designed with NDIR (non-dispersive infrared) Waveguide technology CO2 sensor Programmable warning CO level Long time drift compensation on CO2 sensor Audible alarm (~80db)threshold setting Max, Min, TWA, STEL functions included Mini USB to PC download function for analysis 99 points manually recording 32000/24000 points auto recording Review 99 points manual record Hold function freezes current readings Housing design surrounds with rails to help Air ventilation for quick & accurate response Easy to manually calibrate CO2 at fresh air around 380-420ppm
MATERIAL SUPPLIED
Meter 4pcs AA batteries Mini USB cable and software CD Operation manual Hard carrying case
This package contains:
Optional accessory:
33% calibration salt (VZ0033AZ1) 75% calibration salt (VZ0075AZ1) Adaptor (9V/100~240Vac)
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(9V/1A output).
Install the batteries into the battery compartment on the rear and make sure they are in correct polarity and good contact.
will
appear on the LCD .
atter
When an adaptor is used, it will cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low,
(Fig.1)
The CO sensor can’t work normally
2
under low voltage, please replace with fresh b ies or connect with an adaptor.
B
CO2
PM
B
Fig.1
HOLDHOLD
CO2
%rhrh
TA
DP
WBT
OO
FF
OO
CC
REC
Recall
AM
PM
CO
CAL
B
MINMIN MAXMAX AVGAVG
LCD DISPLAY
Upper LCD
MIN/MAX Minimum/Maximun readings STEL Short-term exposure limit (15 minutes weighted average) TWA Time weighted average(8 hours) Low battery indicator DP Dew point TA Air temperature WBT Wet bulb %rh Unit of relative humidity (77597)
o
C/ F Celsius/Fahrenheit of temp. CAL In calibration status REC In manual/automatic logging RECALL In manual records recall mode
CO2/CO/Temp....reading
Lower LCD Real time display
CO2 Carbon dioxide reading CO Carbon monoxide reading
HOLD Readings are freezed unchanged
temperature(77597)
temperature(77597)
o
Symbols
Turns on and off the meter. Enters setup mode while meter is off. Sets as non-sleep mode with .
SET
START ESC
Exits setup/recall page. Start automatic logging.
KEYPAD
ON/OFF
MODE RECALL
HOLD
Press to switch displayed mode. Long press to enter memory recall mode.
TWATWA STELSTEL
B
Freezes the current readings. Selects unit or increases value in setup. Cancels data hold function.
Press to manually record the reading. Selects unit or decreases value in setup.
Activates MIN, MAX, TWA, STEL function. Saves and finishes settings.
OPERATION
Press to turn the meter on and off. At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO and real time
displayed (Fig.3). The real time display date and time in turns.
Suggest to wait for 30 seconds warm up time. For quick start, you may press for 2 seconds to end the warm up and entering normal display.
POWER ON/OFF
SET
ON/OFF
Fig.2
Fig.3
CO2
AM
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
2
NOTE: Do not hold the meter close to faces in case exhalation affects CO
2
levels.
N X
M /AVG
HOLD
MEM
N X
M /AVG
Press to switch to temperatures display.(Fig.6). In 77597 model, dew point temp. and wet bulb temp. are also available.(Fig.7). The lower display shows the real time clock.
AIR(all models),DP,WBT(77597)
Press to switch the mode to CO2. User can get the CO2 reading in ppm unit on main display.(Fig.4). The lower display shows the real time clock.
CO2 (Carbon Dioxide)
MODE RECALL
CO2
AM
Fig.4
Press to switch the mode to CO. User can get the CO reading in ppm unit on main display.(Fig.5). The lower display shows the real time clock.
CO (Carbon Monoxide)
MODE RECALL
CO
AM
Fig.5
MODE RECALL
TA
OO
CC
PM
WBT
OO
CC
Fig.6
Fig.7
Press to switch to humidity display. (Fig.8).The lower display shows the real time clock.
Humidity (77597)
MODE RECALL
%rhrh
Fig.8
DATA HOLD In normal display mode, press to freeze the readings, "HOLD” icon is flashed on the left top of LCD(Fig.9). All
current readings are kept unchanged. Press again to cancel the hold function.
Fig.9
BACKLIGHT
The backlight will be activated for 10
seconds by pressing any key.
MIN,MAX,STEL,TWA, CURRENT REVIEW
This meter allows you to check the
minimun, maximun, STEL,TWA and
current value from the moment you press
the key. Under normal mode, press to see the minimum, maximum, STEL, TWA and current value in turns. Each press of ,it displays MIN, MAX, STEL,TWA, current value in sequence. (Fig.10~12). To return to normal mode, press key to escape(Fig.13)
If the meter is turned on for shorter than 15 mins, the STEL value will be the weighted average of readings take since power on. Same for TWA values appear before 8 hours.
It takes at least 5 mins to calculate STEL and TWA. The display shows “----” during the first 5 mins from power on.
HOLD
HOLD
HOLDHOLD
%rhrh
AM
NOTE: THE HOLD FUNCTION IS DISABLE IN MIN/MAX/ AVG MODE.
N X
M /AVG
N X
M /AVG
N X
M /AVG
START ESC
In this max/min/STEL/TWA/current modes, it shows the corresponding readings on main display and the accumulated time of how long you are in this mode.(Fig10~12).
NOTE: “HOLD” & “MANUALLY RECORD” & “RECALL” FUNCTION ARE DISABLE IN MIN/MAX/AVG MODE.
MODE RECALL
CO
MAXMAX
Fig.14
START ESC
While in “HOLD” mode, STEL and TWA will remain updating every 5 mins.
CO2
MINMIN
CO2
CO2
CO2
AM
Fig.10
Fig.11
Fig.12
Fig.13
TWATWA
In this mode, press to switch the displayed parameters. Fig 14.
The dispalyed sequence in this mode is min max STEL TWA current min
max.....Press key can immediately
leave this mode and return to normal. Distinguishing the normal display and
current review by checking the time display format. In normal display, the date & time displayed in turns. In current review, the time display is the accumu- lated time of how long you are in max/min mode.
NOTE: TWA &STEL ARE FOR CO2 ONLY. MIN/MAX ARE FOR ALL PARAMETERS.