1.2. During use........................................................................................................................................................ 3
1.3. After use........................................................................................................................................................... 3
2. GENERAL DESCRIPTION ........................................................................................................................................4
2.2. TRMS value and mean value - definitions........................................................................................................ 4
2.3. TRUE ROOT MEAN SQUARE VALUE and CREST FACTOR - definitions.....................................................4
3. PREPARATION FOR USE ........................................................................................................................................5
3.2. Power supply.................................................................................................................................................... 5
4.1.1. Turning on ................................................................................................................................................... 7
4.1.2. Auto Power Off ............................................................................................................................................ 7
4.1.3. Current measurement through external transducer..................................................................................... 7
4.1.4. Phase sequence and phase conformity measurement................................................................................ 7
4.2.1. HOLD .......................................................................................................................................................... 8
4.3.1. Frequency measurement .......................................................................................................................... 10
4.3.2. Anomalous cases which may occur during V
4.4. A
4.4.1. Frequency measurement .......................................................................................................................... 13
5.1. General .......................................................................................................................................................... 39
5.4. End of life ....................................................................................................................................................... 39
6.1.3. General specifications ............................................................................................................................... 42
7.2. After-sale service ...........................................................................................................................................43
Hz: DC/AC voltage and frequency measurement (M73 and M75).......................................................... 9
This instrument complies with safety Standards EN61557 and EN61010-1 related to
electronic measuring instruments.
CAUTION
For your own safety and to avoid damaging the instrument follow the
procedures described in this instruction manual and read carefully all notes
preceded by this symbol .
When taking measurements:
) avoid doing that in humid or wet places - make sure that humidity is within the limits
indicated in section “environmental conditions”.
) avoid doing that in rooms where explosive gas, combustible gas, steam or excessive
dust is present.
) keep you insulated from the object under test.
) do not touch exposed metal parts such as test lead ends, sockets, fixing objects,
circuits etc.
) avoid doing that if you notice anomalous conditions such as breakages, deformations,
fractures, leakages of battery liquid, blind display etc.
) be particularly careful when measuring voltages exceeding 25V in particular places
(building yards, swimming pools, etc.) and 50V in ordinary places to avoid risks of
electrical shocks.
The following symbols are used:
CAUTION - refer to the instruction manual - an improper use may damage the
instrument or its components
DC or AC voltage or current
CAUTION for dangerious voltage. Risk of electric shock
Meter with double insulation
1.1. PRELIMINARY INSTRUCTIONS
) This instrument has been designed for use in environments of pollution degree 2.
) It can be used for tests on electrical installations of overvoltage category III 265V and
550V maximum rated interlinked voltage (and to earth).
) You are recommended to respect the usual safety regulations aimed at protecting you
against dangerous currents and protecting the instrument against improper use.
) Only the original test leads supplied along with the instrument guarantee compliance
with the safety Standards in force. They must be in a good conditions and, if
necessary, replaced with identical ones.
) Do not test nor connect to any circuit exceeding the specified overload protection.
) Do not take measurements under environmental conditions exceeding the limits
indicated in this manual.
) Make sure that batteries are correctly installed.
) Before connecting the test probes to the installation make sure that the right function is
chosen.
EN - 2
Page 4
MULTITEST M72 - M73 - M74 - M75
1.2. DURING USE
CAUTION
An improper use may damage the instrument and/or its components or injure
the operator.
) Before selecting any function, first disconnect the test leads from the circuit under test.
) When the instrument is connected to circuits never touch any unused terminal.
) Do not measure resistance in presence of external voltages; although the instrument is
protected, an excessive voltage may cause malfunctioning.
CAUTION
If the “low battery” symbol is displayed during use interrupt testing and replace
batteries following the procedure described in paragraph 5.2.
1.3. AFTER USE
) Disconnect the test leads from the circuit under test and switch off the instrument.
) If you expect not to use the instrument for a long period remove batteries.
1.4. OVERVOLTAGE CATEGORIES - DEFINITIONS
Standard EN61010-1 (Safety requirements for electrical equipment for measurement, control
and laboratory use, Part 1: General requirements) defines what a measurement category
(usually called “overvoltage category”) is. At paragraph 6.7.4: Measuring circuits it says:
(OMISSIS)
Circuits are divided into the following measurement categories:
•Measurement category IV is for measurements performed at the source of the
low-voltage installation.
Examples are electricity meters and measurements on primary overcurrent
protection devices and ripple control units.
•Measurement category III is for measurements performed in the building
installation.
Examples are measurements on distribution boards, circuit breakers, wiring,
including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, and equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to fixed installation.
•Measurement category II is for measurements performed on circuits directly
connected to the low voltage installation.
Examples are measurements on household appliances, portable tools and similar
equipment.
•Measurement category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of
the equipment is made known to the user.
EN - 3
Page 5
MULTITEST M72 - M73 - M74 - M75
2. GENERAL DESCRIPTION
Dear Customer, the instrument you have purchased, whether used according to the
instructions given in this manual, will grant you accurate and reliable measurements.
Thanks to a development of newest conception assuring double insulation and overvoltage
category III you will enjoy the highest safety. This manual refers to the entire range of
MULTITEST family including M72, M73, M74 e M75 models. Except where otherwise
expressly indicated, this manual refers to all these models.
2.1. WORKING
) VHz: DC and AC TRMS voltage measurement, frequency measurement.
) A
) Ω: measurement of resistance / continuity with sound signal
Hz: DC and AC TRMS current measurement, frequency measurement by means
of current transducer (clamp) with max. full scale of 1V.
)
: phase sequence detection at one or two terminals
) LAN: wire mapping for cables UTP, STP, FTP (SCTP), SFTP, SSTP in any
category (Cat. 3, 5, 5E, 6, etc.) with connector RJ45 capable of measuring
through connection to remote unit (M75 only).
) Ω 0.2A: continuity test on earth, protective and potential equalising conductors with a test
current higher than 200mA and open voltage ranging from 4V to 24V (M72, M74,
M75 only)
) MΩ: insulation resistance measurement with test DC voltage of 250, 500VDC (M72,
M74, M75 only).
) RCD: measurement on RCDs type AC (
)(M73, M74, M75 only)
) Ra measurement of total earth resistance (M73, M74, M75 only)
) AUTOperformance of measurements in Ra , RCD and MΩ with automatic sequence (
M74, M75 only)
2.2. TRMS VALUE AND MEAN VALUE - DEFINITIONS
Safety testers for alternate quantities are divided into two categories:
•MEAN VALUE instruments: instruments measuring only the value of the wave at the
fundamental frequency (50 or 60Hz);
•True Root Mean Square (TRMS) instruments: instruments measuring the true root
mean square value of the quantity under test.
Mean value instruments provide only the value of the fundamental wave while TRMS
instruments provide the value of the entire wave, including harmonics (within the passband
of the instrument). Accordingly, the measured values are identical only if the wave is
purely sinusoidal.
2.3. TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR - DEFINITIONS
The current effective value is defined as follows: “In an interval of time equivalent to a
period, an alternate current with effective value having an intensity of 1A, by passing on a
resistor, disperses the same energy which would be dispersed in the same period of time
by a direct current having an intensity of 1A”. From this definition we get the numerical
expression: G=
0
1
T
The effective value is indicated as RMS (root mean square).
2
dttg
)(
∫
t
0
+Tt
The crest factor is defined as the ratio between the peak value of a signal and its effective
G
value: CF (G)=
p
. This value varies according to the waveform of the signal, for a purely
G
RMS
sinusoidal wave it’s worth 2 =1.41. In presence of distortions, the higher the wave
distortion is, the higher the crest factor values get.
EN - 4
Page 6
MULTITEST M72 - M73 - M74 - M75
3. PREPARATION FOR USE
3.1. PRELIMINARY CHECKS
This instrument was checked both mechanically and electrically prior to shipment. All
possible cares and precautions were taken to let you receive the instrument in perfect
conditions. Notwithstanding we suggest you to check it rapidly (eventual damages may
have occurred during transport – if so please contact the local distributor from whom you
bought the item).
Make sure that all standard accessories mentioned in paragraph 6.3 are included.
Should you have to return back the instrument for any reason please follow the
instructions mentioned in paragraph 7.
3.2. POWER SUPPLY
The instrument is powered by batteries (refer to paragraph 6.1.3 for details on model, no.
and battery life). When batteries are low, a low battery indication is displayed.
To replace/insert batteries follow the instructions indicated in paragraph 5.2.
3.3. CALIBRATION
The instrument complies with the technical specifications contained in this manual and
such compliance is guaranteed for 1 year. Annual recalibration is recommended.
3.4. STORAGE
After a period of storage in extreme environmental conditions exceeding the limits
mentioned in paragraph 6.2.1 let the instrument resume normal measuring conditions
before using it.
EN - 5
Page 7
MULTITEST M72 - M73 - M74 - M75
4. OPERATING INSTRUCTIONS
4.1. INSTRUMENT - DESCRIPTION
LEGEND:
1. Inputs
2. Display
3. ON/OFF key
4. MODE PEAK key
5. Arrows key
6. FUNC HOLD key
7. GO key
8. Remote units for LAN tests
(M75 only)
MODE
PEAK
FUNC
HOLD
GO
Fig. 1: Instrument description
to switch on/off the instrument
MODE key to select working mode
PEAK key to select peak measurement
FUNC key to select working mode
HOLD key to stop display updating
to start a measurement
Arrow keys to select measurements
EN - 6
Page 8
MULTITEST M72 - M73 - M74 - M75
4.1.1. Turning on
When the instrument is turned on, it emits a brief sound and all
display segments are lit for just one second. Then the model
number and the firmware release appear on the display (see
picture referred to M75). Finally the instrument is ready for
operation.
4.1.2. Auto Power Off
The instrument automatically turns off 10 minutes after last pressure of keys. To resume
operation turn on the instrument again. When the instrument must be used for long
periods, the operator may need to disable the Auto Power Off function. In order to do so:
keep pressed the FUNC HOLD key and turn on the instrument; this way the Auto Power
Off function is deactivated. On next switch on autopower off will be automatically restored.
On the instrument’s LCD the symbol is displayed only when auto power off is active.
4.1.3. Current measurement through external transducer
The instrument measures current through a clamp transducer to be connected to the input
terminals. Unlike traditional multimeters it’s therefore not necessary to interrupt the current
circuit to insert the measuring device. Besides it’s possible to use more clamps having
different full scales according to the current to be measured from time to time. To set the
full scale of the clamp being used, turn on the instrument by keeping pressed MODE
PEAK: the instrument displays the value of the set full scale. To modify such value press
the arrow keys to select the desided full scale, then press MODE PEAK to validate the
modification.
The instrument is supplied with full scale already pre-set for the provided clamp.
4.1.4. Phase sequence and phase conformity measurement
Additional functions (patented) distinguishing this instrument from other multimeters
available on the market are phase sequence and phase conformity measurements
(indispensable to interconnect two three-phase triads) with one test probe.
The instrument performs tests by simply touching the cable under test with the built in test
probe. Note! The instrument performs test by only touching the conductors under voltage
with the single test probe, therefore without having to connect both test probes. The
traditional 2-terminal measurement is however available.
EN - 7
Page 9
MULTITEST M72 - M73 - M74 - M75
4.2. HOLD, MAX/MIN/AVG, PEAK±
The following functions are available for measurements of AC and DC voltage, AC current,
frequency and resistance.
4.2.1. HOLD
The HOLD function permits to block on the display the detected value during
measurements of AC and DC voltage, AC current, frequency and resistance. Just press
FUNC HOLD for at least one second. The symbol HOLD is displayed. To escape this
function press again FUNC HOLD or the arrow keys.
This function is not available when MAX/MIN/AVG or PEAK± functions are active.
4.2.2. MAX/MIN/AVG
During measurements of AC and DC voltage, AC current, frequency and resistance it’s
possibile to measure and display the maximum (MAX), minimum (MIN) and average
(AVG) valus of the quantity under test. Press FUNC HOLD for more than one second to
enter this function and press it repeatidly for less than one second to run through MAX,
MIN or AVG. The corresponding symbol is displayed.
Maximum, minimum and average values are detected since this function is activated and
are continuously updated even if not displayed. For example, while the AC current average
value is displayed, the maximum and minimum values of the same quantity are
continuously updated.
To escape the MAX/MIN/AVG function press again FUNC HOLD for more than one
second or the arrow keys.
The MAX/MIN/AVG function is not available when HOLD or PEAK± functions are active.
4.2.3. PEAK±
During measurements of AC and DC voltage and AC current it’s possibile to measure and
display the maximum (PEAK+) and minimum (PEAK-) peak values of the quantity under
test with a resolution of 1ms. Press MODE PEAK for more than one second to enter this
function and press it repeatidly for less than one second to run through PEAK+ or PEAK-.
The corresponding symbol is displayed.
Maximum and minimum peak values are detected since this function is activated and are
continuously updated even if not displayed. For example, while the maximum peak values
of AC current is displayed, the minimum peak value of the same quantity is continuously
updated.
When displaying maximum and minimum peaks it’s not mentioned whether the
corresponding quantity is AC or DC: a peak value is absolutely a peak value, regardless of
the quantity at which it’s detected.
To escape this function press again MODE PEAK for more than one second or the arrow
keys.
The HOLD and MAX/MIN/AVG functions are not available when when PEAK± is active.
EN - 8
Page 10
MULTITEST M72 - M73 - M74 - M75
4.3. V Hz: DC/AC VOLTAGE AND FREQUENCY MEASUREMENT (M73 AND M75)
CAUTION
The maximum input voltage is 550+10%V. Don’t try to measure higher
voltages to avoid risks of electrical shocks or serious damages to the
instrument.
Fig. 2: Connection of the instrument’s terminals during V Hz test
1.
Turn on the instrument
2.
Insert the black and green cables in the corresponding input terminals of the
3.
instrument
Press the arrow keys to select V Hz
4. If necessary, insert the croco clips on the test probes
5. Connect the cables to the desired points of the circuit under test as shown in Fig. 2. The
voltage and frequency values will be displayed with automatic range selection.
6. The instrument automatically switches from AC to DC voltage basing on the signal applied
to terminals.
7.
Example of display of AC voltage and
AC voltage value
frequency values. The minimum reading
limit of AC voltage is 0.5V. Lower input
values are displayed as 0.0V.
EN - 9
Frequency value
Page 11
MULTITEST M72 - M73 - M74 - M75
8.
Example of display of DC voltage. The
minimum reading limit of DC voltage is
1.2V. Lower input values are displayed
as 0.0V.
DC voltage value
MODE
9.
10.
11.
12.
PEAK
MODE
PEAK
FUNC
HOLD
FUNC
HOLD
Press MODE PEAK for less than 1 second to pass to frequency
measurement (only during AC measurements, see paragraph 4.3.1)
Press MODE PEAK for more than 1 second to detect the voltage peak value
(see paragraph 4.2.3)
Press FUNC HOLD for less than 1 second to block the detected values on the
display (see paragraph 4.2.1)
Press FUNC HOLD for more than 1 second to detect maximum, minimum
and average voltage values (see paragraph 4.2.2)
4.3.1. Frequency measurement
1. In order to detect minimum, average, maximum and peak values of frequency it's
necessary to pass to this measuring parameter.
MODE
2.
PEAK
3.
Example of display of frequency value.
During AC measurements it’s possible to pass to frequency measurement by
pressing MODE PEAK for less than 1 second
Frequency value
The minimum reading limit of frequency
is 30.0Hz. Lower input values are
displayed as <30.0Hz
4.
5.
6.
7.
MODE
PEAK
MODE
PEAK
FUNC
HOLD
FUNC
HOLD
To resume voltage measurement press MODE PEAK for less than 1 second
To detect the frequency peak value press MODE PEAK for more than 1
second (see paragraph 4.2.3)
To block the detected frequency value on the display press FUNC HOLD for
less than 1 second (see paragraph 4.2.1)
To detect maximum, minimum and average frequency values press FUNC HOLD for more than 1 second (see paragraph 4.2.2)
EN - 10
Page 12
MULTITEST M72 - M73 - M74 - M75
4.3.2. Anomalous cases which may occur during V Hz measurements
1.
The maximum input voltage is
550+10%V.
If the detected voltage value exceeds
605V TRMS the instrument displays the
screen beside.
Disconnect immediately the instrument
from the circuit under test to avoid
electrical shocks and damages to the
instrument.
2.
If during a voltage measurement the
detected frequency value exceeds 400Hz
the instrument displays the screen
beside.
3.
If during a frequency measurement the
detected value exceeds 400Hz the
instrument displays the screen beside.
4.
If during a frequency measurement the
detected value does not reach 30.0Hz
the instrument displays the screen
beside.
EN - 11
Page 13
MULTITEST M72 - M73 - M74 - M75
4.4. A Hz: DC/AC CURRENT AND FREQ UENCY MEASUREMENT
1.
Turn on the instrument
Fig. 3: Connection of the instrument’s terminals during A
Hz test
2.
Press the arrow keys to select A Hz
3. Insert the banana connectors of the clamp transducer in the corresponding input
terminals of the instrument (black with black, green or red with green)
4. Make sure that the clamp full scale and the instrument full scale do correspond. If they
do not, the measured value will be wrong. To set the clamp full scale refer to
paragraph 4.1.3
5. Open the jaws and insert the cable in the very middle as shown in Fig. 3. Current and
frequency values will be displayed.
6. The instrument automatically switches from AC to DC current basing on the signal applied
to terminals.
7.
Example of DC current displaying. The
DC current value
minimum limit of DC current is:
1.0mV x transduction ratio of the clamp
lower values are nullified.
EN - 12
Page 14
MULTITEST M72 - M73 - M74 - M75
8.
Example of AC current displaying. The
minimum limit of AC current is:
1.0mV x transduction ratio of the clamp
lower values are nullified.
AC current value
Frequency value
The minimum reading value of AC and DC current is given by the herewith:
1mV x transduction ratio of the clamp
Therefore, with a clamp 400A/400mV, the minimum measurable current is 1.0A.
Lower input values are displayed as 0.0A
MODE
9.
10.
11.
12.
4.4.1. Frequency measurement
PEAK
MODE
PEAK
FUNC
HOLD
FUNC
HOLD
To pass to frequency measurement press MODE PEAK for less than 1
second (only during AC current measurements, see paragraph 4.4.1)
To detect the current peak value press MODE PEAK for more than 1 second
(see paragraph 4.2.3)
To block the detected values on the display press FUNC HOLD for less than
1 second (see paragraph 4.2.1)
To detect maximum, minimum and average current values press FUNC HOLD for more than 1 second (see paragraph 4.2.2)
1. In order to detect minimum, average, maximum and peak values of frequency it's
necessary to pass to this measuring parameter.
MODE
2.
PEAK
3.
Example of display of frequency value.
It’s possible to pass to frequency measurement by pressing MODE PEAK for
less than 1 second
Frequency value
The minimum reading limit of frequency
is 30.0Hz. Lower input values are
displayed as <30.0Hz.
MODE
4.
PEAK
MODE
5.
PEAK
FUNC
6.
HOLD
FUNC
7.
HOLD
To resume voltage measurement press MODE PEAK for less than 1 second
To detect the frequency peak value press MODE PEAK for more than 1
second (see paragraph 4.2.3)
To block the detected frequency value on the display press FUNC HOLD for
less than 1 second (see paragraph 4.2.1)
To detect maximum, minimum and average frequency values press FUNC HOLD for more than 1 second (see paragraph 4.2.2)
EN - 13
Page 15
MULTITEST M72 - M73 - M74 - M75
4.4.2. Anomalous cases which may occur during A Hz measurements
1.
If the detected current value exceeds the
clamp full scale the instrument displays
Example of clamp full
scale set at 400A AC
the screen beside.
Disconnect immediately the clamp from
the circuit under test to avoid electrical
shocks and damages to the instrument
The instrument is 20% overchargeable
than the clamp full scale
2.
If during a current measurement the
detected frequency value exceeds 400Hz
the instrument displays the screen beside
3.
If during a frequency measurement the
detected value exceeds 400Hz the
instrument displays the screen beside
4.
If during a frequency measurement the
detected value does not reach 30.0Hz
the instrument displays the screen beside
EN - 14
Page 16
MULTITEST M72 - M73 - M74 - M75
4.5. Ω : RESISTANCE MEASUREMENT AND CONTINUITY TEST
CAUTION
Before taking resistance measurements make sure that the circuit under test is
not powered and that eventual condensers are discharged.
Fig. 4: Connection of the instrument’s terminals during Ω test
1.
2.
Turn on the instrument
Press the arrow keys to select Ω
3. If the measuring cables being used have not been calibrated, first calibrate them as
described in paragraph 4.5.1
4. Insert the black and green cables in the
corresponding input terminals of the
instrument
5. Position the test probes on the desired
points of the circuit under test (see Fig. 4)
Resistance value
6. Example of display of resistance value. If
such value is lower than 40Ωthe
instrument emits an acoustic signal
FUNC
7.
HOLD
FUNC
8.
HOLD
To block the detected value on the display press FUNC HOLD for less than 1
second (see paragraph 4.2.1)
To detect maximum, minimum and average values press FUNC HOLD for
more than 1 second (see paragraph 4.2.2)
9. The measured value is out of accuracy if an input voltage is present
4.5.1. "CAL" mode
1. Any addition or replacement of cables, extensions and croco clips nullify the previous
calibration and make necessary a new calibration before performing further
measurements. Therefore the instrument must be calibrated in the same conditions at
which it will operate during measurements
2. Short-circuit the cable ends with each other as shown in Fig. 5 making sure that the
metallic parts of test probes and crocodiles are in good touch
EN - 15
Page 17
MULTITEST M72 - M73 - M74 - M75
Fig. 5: Connection of the instrument’s terminals during calibration procedure
MODE
3.
PEAK
CAUTION
Press MODE PEAK for more than 1 second. The instrument resets the
resistance of the cables, the symbol "CAL" is displayed
While MODE PEAK is pressed the instrument is measuring. During this phase
never disconnect test leads.
4.
The instrument performs the calibration of cables with resistance lower than 5Ω.
5. At the end of the test the measured value is stored by the instrument and used as
OFFSET, which means it’s deducted from all continuity tests performed) for all further
measurements until a new calibration is made
6.
If the value measured during the calibration phase is higher than 5Ω the instrument
interruptus the calibration, removes the offset value previously stored and does not
display the CAL symbol until the next positive calibration.
This method can be used to nullify the last calibration performed
7. Each time the instrument is switched off the calibrated value is lost
4.5.2. Anomalous cases which may occur during Ω measurement
1. The full scale of the instrument is
39.99kΩ.
If the resistance value is higher than this
value, or in case of open or interrupted
probes, the instrument displays the screen
beside.
EN - 16
Page 18
MULTITEST M72 - M73 - M74 - M75
4.6. : PHASE SEQUENCE AND CONFORMITY MEASUREMENT
CAUTION
The maximum input voltage is 550+10%V. Don’t try to measure higher
voltages to avoid risks of electrical shocks or serious damages to the
instrument. Do not use the instrument on plants whose interlinked rated
voltage is higher than 550V.
Fig. 6: Connection of the instrument’s terminals during test
1.
2.
MODE
3.
PEAK
CAUTION
Turn on the instrument
Press the arrow keys to select
Press MODE PEAK to select function 1W (1-terminal measurement) or 2W (2-
terminal measurement)
Mode 1W requires the operator to touch the measuring key (without gloves)
and be at the earth potential, and the star centre of the system under test to be
at the earth potential. Only if these conditions are met mode 1W provides
correct results. In absence of just one of the above mentioned conditions
(operator wearing protective gloves or mounting a ladder, IT systems etc.)
select mode 2W.
4. Insert the black wire in the corresponding input terminal of the instrument. If necessary
install the croco-clip on the test probe
5. If mode 2W has been selected, insert the green wire in the corresponding input
terminal of the instrument and connect the test probe to the neutral wire or to the
neutral wire of the plant under test. If necessary install the croco-clip on the test probe
6. Following messages are displayed:
EASURING…" the instrument is ready to measure the first phase voltage
"M
"PH1" (secondary display): the operator is invited to connect the measuring cable to the
cable of the first phase voltage
EN - 17
Page 19
MULTITEST M72 - M73 - M74 - M75
CAUTION
For a correct functioning of mode 1W it’s necessary that the star centre of the
three-phase triad under test is at the earth potential.
In plants with insulated neutral wire, like IT systems (often present in hospitals,
airports etc.) it’s necessary to select mode 2W and connect the green probe to
the neutral conductor (not to the protective conductor). In this kind of plants
mode 1W may not provide correct results.
Only for mode 1W press and keep pressed GO, or simply touch the surface of
7.
8.
GO
the key for the entire duration of the measurement. Connect the test probe to
the first wire of the three-phase triad to be tested
When a higher voltage than 110V is detected, the symbol "
PH" is displayed and the
buzzer emits a prolonged sound
CAUTION
During measurement:
•GO must be always kept pressed or at least its surface must be always
touched (only for mode 1W)
• the test probe, except for the phase cable under test, must not be in touch
or close to any voltage source which may block the measurement due to
the instrument’s sensitivity
•the test probe must be kept in touch with the phase cable
9. At the end of the measurement the wordings "MEASURING…" and "PH1" disappear.
The buzzer emits and intermitting sound until the test probe is disconnected drom the
phase cable
10.
Disconnect the test probe from the cable of the first phase voltage. The wording "PH"
(present only when the input voltage is detected) disappears from the display
Only for mode 1W keep pressed GO, or simply touch its surface, for the entire
duration of the measurement. An eventual release and new pressure on the
11.
GO
key cancels all performed measurements. In this case repeat all previous
passages starting from point 6
12. Following messages are displayed:
"MEASURING…" the instrument is ready to measure the second phase voltage
"PH2" (secondary display): the operator is invited to connect the measuring cable to
the cable of the second phase voltage
CAUTION
If more than 10 seconds pass between the first and the second measurement,
a message "t.out" is displayed. In this case it’s necessary to repeat the entire
procedure. Press GO and re-start from point 6.
Only for mode 1W keep pressed GO, or simply touch its surface, for the entire
13.
14.
GO
duration of the measurement. Connect the test probe to the second cable of
the three-phase triad to be tested
When a higher voltage than 110V is detected, the symbol "PH" is displayed and the
buzzer emits a prolonged sound
EN - 18
Page 20
MULTITEST M72 - M73 - M74 - M75
A
A
A
A
CAUTION
During measurement:
•GO must be always kept pressed or its surface must be always touched
(only for mode 1W)
• the test probe, except for the phase cable under test, must not be in touch
or close to any voltage source which may block the measurement due to
the instrument’s sensitivity
•the test probe must be kept in touch with the phase cable
15.
t the end of the test, if two tested
cables are in a correct phase sequenze,
the instrument emits a double sound to
signal the positive outcome of the test
and displays a screen like this
Correct phase
sequence
16.
t the end of the test, if two cables
belong to the same phase, the
instrument emits a double sound to
sigla the positive outcome of the test
and displays a screen like this
17.
t the end of the test, if two tested
cables are not in the correct phase
sequenze, the instrument emits a
prolonged sound to signal the negative
outcome of the test and displays a
screen like this
18. To perform a new measurement press GO, then re-start from point 6
CAUTION
lthough two cables are in sequence it doesn’t mean that the third cable is in
sequence too. It’s not excluded that the cabling was made by mistake with a
double phase cable. To clear any possible doubt always perform at least two
measurements by testing the cables two by two.
Phase rotation
Cables belonging to
the same phase
Compliance between
one cable and
another
Wrong phase
sequence
Phase rotation
EN - 19
Page 21
MULTITEST M72 - M73 - M74 - M75
4.6.1. Anomalous cases which may occur during tests
1.
If you wait more than 10 seconds
between the first measurement and the
second one, the instrument emits a
prolonged sound to signal the negative
outcome of the test and displays a screen
like this. It’s necessary to repeat the
entire procedure. Press GO and re-start
from point 6
2.
If mode 1W is selected and the
instrument detects the connection of the
second probe like in mode 2W, a screen
like this is displayed to signal the error. A
prolonged sound is emitted until the error
condition is removed
3.
If mode 2W is selected and the
instrument detects an input voltage
(between the two bushes) higher than
605V, a screen like this is displayed and
a prolonged sound is emitted until the
error condition is removed. Disconnect
the instrument promptly
EN - 20
Page 22
MULTITEST M72 - M73 - M74 - M75
4.7. LAN: CABLING TEST (M75)
CAUTION
Before taking any measurement make sure that the circuit under test is not
powered. Connections to phone lines or active networks could damage the
instrument.
Fig. 7: Connections of the instrument’s terminals during LAN tests
1.
2.
MODE
3.
PEAK
Turn on the instrument
Press the arrow keys to select LAN
Select the type of cable under test by pressing MODE PEAK: set STP whether
shielded, UTP whether unshielded
Note: STP must be chosen for all shielded cables such as:
4. Connect the cable under test to MULTITEST 75 and to the remote unit if necessary
through patch cables (see Fig. 7)
CAUTION
The remote unit must be necessarily connected to the other end of the cable
being tested, otherwise no measurement is performed.
GO
Press GO to perform all tests related to the selected type of cable
Identification number
of the remote unit
5.
6.
If cabling is correct, a screen like this is
displayed. The identification number (02)
refers to the remote unit connected to the
other end of the cable being tested
EN - 21
Page 23
MULTITEST M72 - M73 - M74 - M75
7.
If cabling is not correct, a screen like this
is displayed (NOT OK). Referring to this
example, “FAULT 1/4” means that the
Identification number
of the remote unit (if
possible to find it)
detected errors are 4, of which the first
one is currently displayed. Details on the
detected error are given on the right side:
the couple 1-2 is open. By pressing
FUNC HOLD key it’s possible to run over
the remaining screens and display other
cabling errors (“FAULT 2/4”, “FAULT
3/4”, “FAULT 4/4”). The number of the
Number of the
displayed error /
number of the
detected errors
remote unit can be not displayed
CAUTION
It’s indispensable to select the right type of cable. If UTP is selected although
a STP cable is tested, test results may be not reliable due to the shield
affecting the measurement.
4.7.1. Anomalous cases which may occur during LAN tests
If the voltage present at the terminal is
higher 0.2V the instrument does not
perform the test and emits a sound to
signal the anomalous situation. The
screen beside is displayed.
CAUTION
Before taking any measurement make sure that the circuit under test is not
powered. Connections to phone lines or active networks could damage the
instrument.
4.7.2. SPLIT PAIRS - explication note
A LAN cable contains 8 conductors, twisted two by two thus forming 4 pairs: 1-2, 3-6,
4-5, 7-8. The error “SPLIT PAIRS” consists in the exchange of two conductors
belonging to different pairs. The pin to pin correspondence seems intact, but physically
the conductors of two couples are split. Such interaction hardly affects (or even makes
impossible) the exchange of data at high frequency/speed
CAUTION
The error condition “SPLIT PAIRS” is verified only when the cable mapping is
fully correct. For a correct detection of such error condition it’s necessary that
the cable under test is at least 1m long.
EN - 22
Page 24
MULTITEST M72 - M73 - M74 - M75
4.7.3. Cabling errors
Cabling errors Description Visualization Mapping
OPEN PAIR
REVERSED PAIR
SHORTED CABLES
TRANSPOSED
(CROSSED) PAIRS
MISWIRE
SPLIT PAIRS
One or both conductors
of the pair are interrupted
(open)
The conductors of the
same pair are reversed
Two conductors are in
short circuit between
each other
Two pairs are crossed
Generic cabling error,
such as for example two
conductors belonging to
different pairs are
exchanged
The pin to pin
correspondence is hold,
but physically the
conductors of two pairs
are crossed
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
1
2
3
4
5
6
7
8
S
EN - 23
Page 25
MULTITEST M72 - M73 - M74 - M75
4.8. Ω 0.2A: CONTINUITY TEST ON EARTH, PROTECTIVE AND EQUALIZING
POTENTIAL CONDUCTORS (M72, M74, M75)
The measurement is performed with a test current higher than 200 mA (R<5Ω) and open
circuit voltage ranging from 4 to 24V DC according to EN 61557-2 and VDE 0413 part 4.
CAUTION
Before performing the continuity test make sure that no voltage is present at the
ends of the conductor under test.
Fig. 8: Connection of the instrument’s terminals during Ω 0.2A test
1.
2.
Turn on the instrument
Press the arrow keys to select Ω 0.2A
3. Insert the black and green cables in the corresponding input terminals of the instrument
4. If the cable length is not sufficient to perform the test, extend the black one
5. If necessary insert the croco clips on the test probes
6. If the measuring cables being used have not been calibrated, first calibrate them as
described in paragraph 4.8.1
7. Connect the instrument’s terminals to the ends of the conductor on which the continuità test
must be performed (see Fig. 8)
8.
GO
Press GO to perform the measurement
EN - 24
Page 26
MULTITEST M72 - M73 - M74 - M75
A
CAUTION
The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads. Connect the
instrument just BEFORE measuring and do not change connections while the
message “Measuring” is present on the display.
9. The continuity test is performed by supplying a current higher than 200mA if the
resistance value is lower than 5Ω (including the cable resistance stored as offset after
calibration). For higher resistance values the instrument performs the test with
decreasing current
10.
t the end of the test, if it has been
Resistance value
possible to generate at least 200mA
(not particularly high resistance value),
the instrument emits a double sound to
signal the positive outcome of the test.
The screen beside is displayed
11.
At the end of the test, if it has not been
Test current value
Resistance value
possible to generate 200mA due to the
high resistance value, the instrument
emits a prolonged sound to signal the
negative outcome of the test. The
screen beside is displayed
Test current value
4.8.1. "CAL" mode
Fig. 9: Connection of the instrument’s terminals during calibration procedure
EN - 25
Page 27
MULTITEST M72 - M73 - M74 - M75
MODE
1.
PEAK
By pressing MODE select CAL
2. Any addition or replacement of cables, extensions and croco clips nullify the previous
calibration and make necessary a new calibration before performing further
measurements. Therefore the instrument must be calibrated in the same conditions at
which it will operate during measurements
3. Short-circuit the cable ends with each other as shown in Fig. 9 making sure that the
metallic parts of test probes and crocodiles are in good touch
4.
GO
CAUTION
Press GO to start the calibration procedure
The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads.
5.
The instrument performs the calibration of cables with resistance lower than 5Ω
6.
At the end of the test the measured
value is stored by the instrument and
used as OFFSET, which means it’s
deducted from all continuity tests
performed) for all further measurements
until a new calibration is made.
The instrument emits a double sound to
signal the positive outcome of the
calibration procedure and displays a
screen similar to this for 2 seconds.
Then, the default screen corresponding
to the Ω 0.2A test is displayed
7.
If the value measured during the
CAL message:
it means that the
instrument has been
calibrated. This
symbol remains
displayed during any
further measurement
even in case the
instrument is turned
off and on.
Current supplied by
the instrument during
calibration procedure
calibration phase is higher than 5Ω the
instrument interruptus the calibration,
removes the offset value previously
stored and does not display the CAL
symbol until the next positive calibration.
The instrument emits a prolonged sound
to signal the negative outcome of the
calibration and displays a screen similar
to this for 2 seconds. Then, the default
screen related to the Ω 0.2A test is
displayed. This method can be used to
nullify the last calibration performed.
EN - 26
Page 28
MULTITEST M72 - M73 - M74 - M75
4.8.2. Anomalous cases which may occur during Ω 0.2A tests
1.
If the following condition occur:
R
MEASURED
- R
CALIBRATION
< -0.02Ω
the instrument displays the screen beside
and emits a prolonged sound to signal
the anomalous situation
2.
If the voltage present at the terminals is
higher than 10V the instrument does not
perform the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds, after which the instrument
displays the default value related to the Ω
0.2A test
3.
If the resistance value is higher than the
full scale the instrument emits a
Input voltage
prolonged sound to signal the anomalous
situation. A screen similar to this is
displayed. The same message may also
mean that mesuring cables are
disconnected or open
EN - 27
Page 29
MULTITEST M72 - M73 - M74 - M75
4.9. MΩ: INSULATION RESISTANCE MEASUREMENT TEST VOLTAGE 500V
The measurement is performed according to EN 61557-2 and VDE 0413 part 1.
CAUTION
• before performing the insulation test make sure that the circuit under test is
not energized and all relative loads are disconnected
• the insulation measurement requires particolar care and attention to avoid
providing wrong test results and causing damages to third parties
• before the insulation test prepare the plant adequately by disconnecting
everything must not be tested. During the insulation test continuously make
sure that the applied voltage is not accessible to third parties
• a measurement with a cable disconnected by mistake may provide a good
result also in presence of a faulty insulation. It’s necessary to take all
possibile cares to avoid that. Once prepared the plant and connected the
measuring cables, make sure that they are correctly connected. In case of
doubt, bifore performing an insulation test, perform a Ω 0.2A measurement
by short-circuiting the cables under test at a point of the plant which is as
far as possibile from the measuring clips. Remove the short circuit before
performing the insulation test
Fig. 10: Connection of the instrument’s terminals during MΩ test
1.
2.
Turn on the instrument
Press the arrow keys to select MΩ. Select test voltage pressing MODE PEAK
key between 250 and 500VDC
3. Insert the black and green cables in the corresponding input terminals of the instrument
4. If the cable length is not sufficient for the measurement extend the black one with an
adequately insulated cable, as its insulation is in parallel to the resistance to be measured. It
must be suspended and not laid to earth and all supports must be of insulated material
5. If necessary insert the croco clips on the test probes
6. Disconnect the circuit or the part of plant under test from power and all eventual loads
7. Connect the instrument’s terminals to the end of the conductors on which the insulation test
must be performed (see Fig. 10)
GO
8.
Press GO to start the measurement
EN - 28
Page 30
MULTITEST M72 - M73 - M74 - M75
A
A
A
CAUTION
The message “Measuring” on the display means that the instrument is
measuring or discharging eventual capacitors. During this phase never
disconnect nor touch test leads.
9. At the end of the test, before giving the result of the measurement, the instrument
automatically discharge eventual capacitors and parasite capacitances present among the
conductors involved in the measurement
10.
t the end of the test, if the resistance
Resistance value
value is higher than 0.5MΩ (or 0.25MΩ
for 250V), the instrument emits a double
sound to signal the positive outcome of
the test. A screen similar to this is
displayed
11.
t the end of the test, if the resistance
Test voltage value
Resistance value
value is higher than 999MΩ, so higher
than the full scale, the instrument emits
a double sound to signal the positive
outcome of the test. A screen similar to
this is displayed. Note! An insulation
value higher than 999MΩ is an excellent
insulation value, generally much higher
than the minimum requirements
prescribed by Standards
12.
t the end of the test, if the resistance
Test voltage value
Resistance value
value is lower than 0.5MΩ (or 0.25MΩ
for 250V) the instrument emits a
prolonged sound to signal the negative
outcome of the test. A screen similar to
this is displayed
Test voltage value
4.9.1. Anomalous cases which may occur during MΩ tests
If, during measurement, the voltage
present at terminals is higher than 10V
the instrument does not perform the test
and emits a prolonged sound to signal
the anomalous situation. The screen
beside is displayed for 5 seconds after
which the instrument displays the default
screen related to the MΩ test
Input voltage value
EN - 29
Page 31
MULTITEST M72 - M73 - M74 - M75
4.10. RCD: TESTS ON AC TYPE RCDS (M73, M74, M75)
The test is performed in compliance with CEI 64.8 612.9, CEI 64.8/6 appendix D,
EN61008, EN61009, EN 60947-2 part B 4.2.4.1 and VDE 0413 part 6.
CAUTION
Testing an RCD involves the tripping of the RCD itself. Therefore, before taking
this measurement, make sure that no loads are connected to the RCD under test to avoid damaging them. Disconnect all loads connected to the RCD as
they could add further leakage currents to those moved by the instrument, thus
nullifying the test results.
Fig. 11: Connection of the instrument’s terminals during RCD test
1.
2.
MODE
3.
PEAK
CAUTION
Turn on the instrument
Press the arrow keys to select RCD
By pressing MODE select the test current among the possible values 30mA,
30mA x5, 100mA, 300mA which turn cyclically at each key pressure
Pay attention when setting the test current of the RCD to make sure that the
correct one is selected. In case a higher current than the nominal one of the
device under test is selected, the RCD would be tested at a higher current
than the correct one, thus favouring a quicker tripping of the RCD itself.
As an alternative:
4. Insert the black and green cables in the corresponding input terminals of the
instrument. If necessary insert the croco clips on the test probes
5. Connect the green cable to the protective conductor (earth) and the black cable to
the phase conductor at the lower end of the RCD under test (Fig. 11)
Or:
4. Insert the Shuko cable in the input terminals of the instrument
5. Insert the Shuko cable in a socket at the lower end of the RCD under test (Fig. 11)
EN - 30
Page 32
MULTITEST M72 - M73 - M74 - M75
A
A
Keep GO pressed for at least one second to perform the leakage current
measurement in phase with the positive semiwave of the network voltage (0°),
6.
GO
or keep GO pressed for at least one second and, when the hyphens on the
display start disappearing, press GO again to perform the measurement with
the leakage current in phase with the negative semiwave of the network
voltage (180°)
CAUTION
The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads.
7.
tripping time is lower than 300ms (40ms
t the end of the test, if the detected
for I∆n=30mA x5), the instrument emits a
double sound to signal the positive
outcome of the test and displays a screen
like this
Tripping time
8.
t the end of the test, if the detected
tripping time is higher than 300ms (40ms
Test current value
Tripping time
exceeding the limit
for I∆n=30mA x5), or in case the RCD
does not trip, the instrument emits a
prolonged sound to signal the negative
outcome of the test and displays a screen
like this
Test current value
4.10.1. Anomalous cases which may occur during RCD tests
1.
If during measurement a higher input
voltage than 265V is detected (for
example, both cables connected to the
phase conductors of a 400V threephases plant) the instrument does not
perform the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds after which the instrument
displays the default screen of RCD test
EN - 31
Page 33
MULTITEST M72 - M73 - M74 - M75
2.
If during measurement a lower input
voltage than 110V is detected, the
instrument does not perform the test and
emits a prolonged sound to signal the
anomalous situation. The screen beside
is displayed for 5 seconds after which the
instrument displays the default screen of
RCD test
This can happen for example if the black cable is erroneously connected to the neutral
conductor instead of the phase one. If a Shuko cable is used, rotate the plug and
repeat the test
3.
If during measurement the green probe is
connected to the phase conductor and
the black probe is connected to the
protective conductor, the instrument does
not perform the test and emits a
prolonged sound to signal the anomalous
situation. The screen beside is displayed
for 5 seconds after which the instrument
displays the default screen of RCD test
4.
If during measurement an excessive
contact voltage is detected (higher than
50V) the instrument does not perform the
test and emits a prolonged sound to
signal the anomalous situation. The
screen beside is displayed for 5 seconds
after which the instrument displays the
default screen of RCD test
5.
If during measurement an excessive
earth resistance is detected, such to
prevent the instrument from generatine
the test current, the instrument does not
perform the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds after which the instrument
shows the default screen related to RCD
test
EN - 32
Page 34
MULTITEST M72 - M73 - M74 - M75
4.11. RA : MEASUREMENT OF GLOBAL EARTH RESISTANCE (M73, M74, M75)
CAUTION
Disconnect all loads connected to the lower end of the RCD as they could
introduce additional leakage currents, thus nullifying the test results. Is possible to
perform measurement on plants whose phase to earth rated voltage is up to
265V. Do not use the instrument on plants whose interlinked rated voltage is
higher than 550V.
Fig. 12: Connection of the instrument’s terminals during Ra test
1.
2.
MODE
3.
PEAK
CAUTION
Turn on the instrument
Press the arrow keys to select Ra
By pressing MODE select the test current among the possible values 15mA
and 100mA which turn cyclically at each key pressure
If a RCD is present, select a lower current value than the nominal current
value of the device. Otherwise the RCD could trip during the measurement
and therefore prevent it from being executed.
By selecting a test current of 100mA you will get also the value of the prospective short
4.
U
circuit current phase to earth, calculated according to the formula ICC=
N
Z
where:
PE
ZPE is the global earth resistance value
U
is the nominal phase to earth voltage whose value is: 127V if 100V ≤ V
N
230V if 150V ≤ V
measured
measured
< 150V
< 265V
As an alternative:
5. Insert the black and green cables in the corresponding input terminals of the
instrument. If necessary insert the croco climps on the test probes
6. Connect the green cable of the instrument to the protective conductor (earth) and
the black cable to the phase conductor (as shown in Fig. 12)
EN - 33
Page 35
MULTITEST M72 - M73 - M74 - M75
A
A
Or:
5. Insert the Shuko cable in the input terminals of the instruments
6. Insert the Shuko cable in a power socket (as shown in Fig. 12). The picture
represents the connection to power socket
7.
GO
CAUTION
Keep pressed GO for at least one second, the instrument performs the
measurement
The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads.
8.
t the end of the test, if the earth
resistance value is lower than 1999Ω,
the instrument emits a double sound
and displays a screen like this reporting
the value of the measured global earth
resistance and the current at which the
measurement has been performed
9.
If a test current of 100mA has been
selected and the earth resistance value
is lower than 1999Ω, by pressing FUNC
you will display alternatively the values
of the global earth resistance and the
prospective short circuit current phase
to earth, as well as the current at which
the measurement has been performed
10.
t the end of the test, if the global earth
resistance value is higher than 1999Ω,
the instrument emits a double sound
and displays a screen like this
Measured value of
global earth
resistance
The symbol HOLD
remains on until a
new measurement
can be taken
Current used during
the measurement
Measured value of
global earth
resistance
The symbol HOLD
remains on until a
new measurement
can be taken
Current used during
the measurement
Measured value of
global earth
resistance higher
than full scale
CAUTION
To guarantee the correctness of measurements, a certain interval of time is
necessary between a measurement and the following. During this period the
symbol HOLD is displayed and no measurement can be taken. When the
symbol HOLD disappears, the instrument is ready for a new measurement.
EN - 34
The symbol HOLD
remains on until a
new measurement
can be taken
Page 36
MULTITEST M72 - M73 - M74 - M75
4.11.1. Anomalous cases which may occur during Ra tests
1.
If during measurement the RCD
protecting the line trips, the instrument
interrupts the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds after which the instrument
displays the default screen of Ra test
2.
If during measurement a higher input
voltage than 265V is detected (for
example, if both cables connected to
phase conductors of a 400V three-phase
plant) the instrument does not perform
the test and emits a prolonged sound to
signal the anomalous situation. The
screen beside is displayed for 5 seconds
after which the instrument displays the
default screen of Ra
3.
If during measurement a lower input
test
voltage than 110V is detected, the
instrument does not perform the test and
emits a prolonged sound to signal the
anomalous situation. The screen beside
is displayed for 5 seconds after which the
instrument displays the default screen of
Ra test
This can happen for example if the black cable is erroneously connected to the neutral
conductor instead of the phase one. If a Shuko cable is used, rotate the plug and
repeat the test
4. If during measurement an excessive
contact voltage is detected (higher than
50V) the instrument does not perform the
test and emits a prolonged sound to
signal the anomalous situation. The
screen beside is displayed for 5 seconds
after which the instrument displays the
default screen of Ra
test
EN - 35
Page 37
MULTITEST M72 - M73 - M74 - M75
5.
If during measurement the green probe is
connected to the phase conductor and
the black probe is connected to the
protective conductor, the instrument does
not perform the test and emits a
prolonged sound to signal the anomalous
situation. The screen beside is displayed
for 5 seconds after which the instrument
displays the default screen of Ra
This can happen even if the wrong connection is performed on the back side of the
test
power socket
EN - 36
Page 38
MULTITEST M72 - M73 - M74 - M75
4.12. AUTO: AUTOMATIC CYCLE OF MEASUREMENTS TO TEST A PLANT (M74,
M75)
This function permits to test an electrical plant in a completely automatic way without any
intervention of the operator.
CAUTION
Testing an RCD involves the tripping of the RCD itself. Therefore, before taking
this measurement, make sure that no loads are connected to the RCD under test to avoid damaging them. Disconnect all loads connected to the RCD as
they could add further leakage currents to those moved by the instrument, thus
nullifying the test results.
Fig. 13: Connection of the instrument’s terminals during AUTO test
1.
2.
MODE
3.
PEAK
CAUTION
Turn on the instrument
Press the arrow keys to select AUTO
By pressing MODE select the test current among the possible values 30mA,
30mA x5, 100mA, 300mA which turn cyclically at each key pressure
Pay attention when setting the test current of the RCD to make sure that the
correct one is selected. In case a higher current than the nominal one of the
device under test is selected, the RCD would be tested at a higher current
than the correct one, thus favouring a quicker tripping of the RCD itself.
4. Insert the black and green cables in the corresponding input terminals of the
instrument. If necessary insert the alligator clips on the test probes or use the shuko
cable
5. Connect the green cable to the protective conductor (earth) and the black cable to the phase
conductor (Fig. 13) or insert the Shuko cable in a power socket
Keep GO pressed for at least one second, the instrument performs
6.
GO
measurements without any intervention of the operator in the following
sequenze: Ra (15mA), RCD, MΩ (phase to earth)
EN - 37
Page 39
MULTITEST M72 - M73 - M74 - M75
A
A
CAUTION
The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads.
7. During measurements, at the end of each test, the partial values are displayed for 5
seconds, then the instrument passes to the following test
8.
At the end of the Ra test, if the earth
resistance value is lower than 50V/I
the instrument displays for 5 seconds a
screen like this, then it passes to the
∆n
Measured value of
global earth
resistance
following measurement. Refer to
paragraph 4.11 for further details or
information regarding the negative
outcome of the test or anomalous
situations
9.
t the end of the test, if the detected
tripping time is lower than the maximum
admittable limit, the instrument displays
Measured value of
tripping time
for 5 seconds a screen like this, then it
passes to the following screen. For
further details or information regarding
the negative outcome of the test or
anomalous situations refer to paragraph
4.10
10.
At the end of the MΩ test, if the
detected resistance value is higher than
0.5MΩ, the instrument displays for 5
Value of test current
Measured value of
resistance
seconds a screen like this, then it
passes to the following screen. For
further details or information regarding
the negative outcome of the test or
anomalous situations refer to paragraph
4.9
11.
t the end of the AUTO test, if all tests
have a positive outcome, the instrument
emits a double sound to signal the
Value of test voltage
positive outcome and displays a screen
like this. To display partial results press
FUNC HOLD. The single test results
are dislayed cyclically by any new
pressure on this key
EN - 38
Page 40
MULTITEST M72 - M73 - M74 - M75
5. MAINTENANCE
5.1. GENERAL
This is a precision instrument. Strictly follow the instructions for use and storage reported
in this manual to avoid any possible damage or danger during use.
Do not use this tester under unfavorable conditions of high temperature or humidity. Do
not expose to direct sunlight.
Be sure to turn off the tester after use. If the instrument is not to be used for a long period
you are recommended to remove batteries to avoid leakages of battery liquid which may
damage its internal circuits.
5.2. BATTERY REPLACEMENT
When the low battery indication (refer to paragraph 6.1.3) is displayed the batteries are to
be replaced.
CAUTION
Only skilled technicians can open the instrument and replace batteries. Before
removing batteries disconnect the test leads from the input terminals to avoid
electrical shocks.
1. Switch off the instrument
2. Remove test leads from the input terminals
3. Remove the battery compartment cover by using a screwdriver
4. Remove all batteries replacing them with new ones all of the same type (refer to
paragraph 6.1.3) respecting the polarity signs
5. Re-allocate the battery pack taking care that the part from which the black and red
wires come out is positioned backwards
6. If the battery pack is re-allocated in a wrong way, the battery compartment can not be
closed. In this case do not force the plastic parts, but re-position the battery pack
correctly before closing
7. Replace the battery compartment cover making a pressure to close it
8. Use the appropriate battery disposal methods for your area
5.3. CLEANING
Use a soft dry cloth to clean the instrument. Never use wet cloths, solvents, water, etc.
5.4. END OF LIFE
Caution: this symbol indicates that equipment and its accessories shall be
subject to a separate collection and correct disposal
EN - 39
Page 41
MULTITEST M72 - M73 - M74 - M75
6. TECHNICAL SPECIFICATIONS
This product conforms to the prescriptions of the European directive on low voltage
73/23/EEC (LVD) and to EMC directive 89/336/EEC, amended by 93/68/EEC.
6.1. CHARACTERISTICS
Accuracy is indicated as [% of reading + digit number]. It is referred to the following
reference conditions: 23°C ± 5°C with RH <70%.
DC and AC TRMS current measurement (with external jaws)
Range Resolution
1.0 ÷ 999.9mV
1.000 ÷ 1.200V
The mentioned accuracy does not consider the transducer's accuracy. Please refer to its user's manual
MAX, MIN, AVG, PEAK, resolution: ±(5.0% rdg + 10 dgt); responce time: 500ms (MAX, MIN, AVG),1ms (PEAK)
Minimum input current: 1mV x transduction ratio of the clamp
Max crest factor: 3.0: V<1.0V; 1.5: V≥1.0V
0.1mV
0.001V
Accuracy
DC
±(0.5% rdg
+ 2 dgt)
Accuracy
(30 ÷ 70Hz)
±(1.0% rdg +
2 dgt)
Accuracy
(70 ÷ 400Hz)
±(2.0% rdg +
2 dgt)
Frequency measurement with test leads
Range Resolution Accuracy Input impedance
30.0 ÷ 199.9Hz
200 ÷ 400Hz
Input voltage: 1mV ÷ 605.0V
0.1Hz
1Hz
±(0.5% rdg + 2 dgt) 1MΩ
Frequency measurement with jaws
Range Resolution Accuracy Overload protection
30.0 ÷ 199.9Hz
200 ÷ 400Hz
Input voltage: 1mV ÷ 1V
0.1Hz
1Hz
±(0.5% rdg + 2 dgt)
Resistance and continuity test
Range Resolution Accuracy Overload protection
0.00 ÷ 39.99 Ω 0.01Ω
40.0 ÷ 399.9 Ω 0.1Ω
400 ÷ 3999 Ω 1Ω
±(1.0% rdg + 5 dgt)
605V AC max RMS for 1 minute
4.00 ÷ 39.99 kΩ 10Ω
The buzzer sounds while measured resistance is lower than 40Ω
Accuracy
(70 ÷ 400Hz)
impedance
±(2.0% rdg +
2 dgt)
Input
impedance
1MΩ
Overload
protection
605V AC
max RMS
605V AC max RMS
Input
1MΩ
EN - 40
Page 42
MULTITEST M72 - M73 - M74 - M75
Phase sequence and phase conformity
Measuring method Wo r k i n g v o lta g e ( V ) Sy s t e m
1 test lead (1W)
2 test leads (2W)
Max crest factor: 1.5
Frequency range: 45 ÷ 65 Hz
(1) the two-wire measurement can be performed also phase to phase in plants without neutral, even with one phase to
earth, but always with phase to phase voltage up to 550V
90 ÷ 315 (Phase - Ground)
90 ÷ 315 (Phase - Neutral)
up to 315 V (Phase - Ground)
up to 550V (Phase - Phase)
up to 315 V (Phase - Neutral)
up to 550V (Phase - Phase) (1)
Ω 0.2A: Continuity (M72, M74, M75)
Range Resolution Accuracy Overload protection
0.00 ÷ 19.99Ω 0.01Ω
20.0 ÷ 99.9Ω
Test current: >200mA DC up to 5Ω (measuring cables resistance included)
Accuracy of current measurement: 1mA
Open circuit voltage: 4 < V0 < 24V
(*) For 500VDC test voltage. For 250VDC test voltage the range is: 200 ÷ 499MΩ
Autorange
Open circuit voltage: <1.3 x V
Accuracy of nominal voltage: –0% +10%
Short circuit current: <3.0mA
Nominal testing current: 1mA @ 1KΩ x V (1mA @ 500KΩ)
0
RC D : t e s t o n R CD dev i c e s AC type (M73, M74, M75)
Range Resolution Accuracy Overload protection
2 ÷ 400ms
Phase to ground voltage: 110 ÷ 265V
Test currents: 30mA, 30mA x 5, 100mA, 300mA
Frequency: 50Hz ± 0.5Hz / 60Hz ± 0.5Hz
1ms
±(2.0% rdg + 2 dgt)
605V max RMS
Ra : Measurement of global earth resistance (M73, M74, M75)
Test current Range Resolution Accuracy Overload protection
15mA
100mA
Phase to ground voltage: 110 ÷ 265V; Frequency: 50Hz ± 0.5Hz / 60Hz ± 0.5Hz
Nominal voltage used for the calculation of the prospective short circuit current: 127V if 100V ≤ V230V if 150V ≤ V
1 ÷ 1999Ω 1Ω ±(5% rdg + 2 dgt)
0.1 ÷ 199.9Ω 0.1Ω ±(5% rdg + 3 dgt)
605V max RMS
< 150V
measured
< 265V
measured
Wire mapping (M75)
Length of the cable: 1÷100m
Number of remote units: max 8 units
Detected errors: OPEN Pairs, REVERSED pairs, SHORT pairs, SPLIT pairs, CROSSED pairs, MISWIRING
According to the norm: TIA568B
CAUTION
In all measurements the instrument dispalys the CAUTION symbol when:
• the instrument is operating in a critical situation as for example in presence
of overvoltages
•the instrument cannot guarantee the measuring uncertainty lower than
30% of the reading, according to EN61557-1
EN - 41
Page 43
MULTITEST M72 - M73 - M74 - M75
6.1.1. Electrical
Conversion: ADC 16 bit, TRMS – True Root Mean Square
Measuring rate: 64 times per second
Display refreshing rate: 2 times per second
6.1.2. Safety standards
The instrument complies with: EN61010-1, EN61557
Insulation: Class 2, double insulation
Pollution level: 2
Max height: 2000m, indoor use
Overvoltage category: CAT III 550V (phase to earth)
CAT III 550V (phase to phase)
Indoor usage, maximum height: 2000m
6.1.3. General specifications
Mechanical features
Dimensions: 240(L) x 100(W) x 45 (H)mm
Weight (batteries included): approx. 630g
Power supply
Battery type: 4 batteries 1.5 V – LR6 – AA – AM3 – MN 1500
Low battery indication: the symbol "" is displayed when the battery
voltage is too low
Battery life: Multimeter: About 90 hours
: > 1000 tests
LAN: > 1000 tests
Ω 0.2A: > 1000 tests @ 1Ω
MΩ: > 1000 tests @ 480kΩ (500VDC) RCD: > 1000 tests
Ra
: > 1000 tests
AUTO: > 1000 tests
Display
Features: 4 LCD with max. reading 9999 counts + symbol
and decimal point
6.2. ENVIRONMENT
6.2.1. Environmental conditions
Reference temperature: 23° ± 5°C
Working temperature: 0°C ÷ 40 °C
Relative humidity allowed: <70%
Storage temperature: -10 ÷ 60 °C
Storage humidity: <70%
6.2.2. EMC
This product conforms to the prescriptions of the European directive on low voltage
73/23/EEC (LVD) and to EMC directive 89/336/EEC, amended by 93/68/EEC.
6.3. ACCESSORIES
Please see enclosed packing list.
EN - 42
Page 44
MULTITEST M72 - M73 - M74 - M75
7. SERVICE
7.1. WARRANTY CONDITIONS
This instrument is guaranteed against material or production defects, in accordance with
our general sales conditions. During the warranty period the manufacturer reserves the
right to decide either to repair or replace the product.
Should you need for any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault). Use only original
packaging. Any damage occurred in transit due to non original packaging will be charged
anyhow to the customer.
The manufacturer will not be responsible for any damage to persons or things.
The warranty doesn’t apply to:
• Accessories and batteries (not covered by warranty).
• Repairs made necessary by improper use (including adaptation to particular
applications not foreseen in the instructions manual) or improper combination with
incompatible accessories or equipment.
• Repairs made necessary by improper shipping material causing damages in transit.
• Repairs made necessary by previous attempts for repair carried out by non skilled or
unauthorized personnel.
• Instruments for whatever reason modified by the customer himself without explicit
authorization of our Technical Dept.
The contents of this manual may not be reproduced in any form whatsoever without the
manufacturer’s authorization.
Our products are patented and our logotypes registered. We reserve the right to
modify specifications and prices in view of technological improvements or
developments which might be necessary.
7.2. AFTER-SALE SERVICE
Shouldn’t the instrument work properly, before contacting your distributor make sure that
batteries are correctly installed and working, check the test leads and replace them if
necessary. Make sure that your operating procedure corresponds to the one described in
this manual.
Should you need for any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault). Use only original
packaging. Any damage occurred in transit due to non original packaging will be charged
anyhow to the customer.
The manufacturer will not be responsible for any damage to persons or things.
EN - 43
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.