HT Neptune, F3000U User Manual

ENGLISH
User manual
 Copyright HT ITALIA 2018 Release EN 1.00 - 31/10/2018
NEPTUNE
TABLE OF CONTENTS
1.PRECAUTIONS AND SAFETY MEASURES ............................................................... 2
1.1. Preliminary instructions ..................................................................................................... 2
1.2. During use ......................................................................................................................... 3
1.3. After use ............................................................................................................................ 3
1.4. Definition of Measurement (Overvoltage) category ........................................................... 3
2.GENERAL DESCRIPTION ........................................................................................... 4
2.1. Measuring average values andTRMS values .................................................................... 4
2.2. Definition of true root mean square value and Crest factor ............................................... 4
3.PREPARATION FOR USE ........................................................................................... 5
3.1. Initial checks ...................................................................................................................... 5
3.2. Instrument power supply ................................................................................................... 5
3.3. Storage .............................................................................................................................. 5
4.NOMENCLATURE ........................................................................................................ 6
4.1. Description of the instrument ............................................................................................. 6
4.1.1. Instrument’s initial screen ........................................................................................................... 6
4.2. Description of function keys .............................................................................................. 7
4.2.1. Key GO/HOLD ............................................................................................................................ 7
4.2.2. Key H/H%/H ............................................................................................................................ 7
4.2.3. Key MODE/MXMNPK ................................................................................................................. 8
4.2.4. Keys / e / ZERO .............................................................................................................. 8
4.2.5. Key VTEST/LIM .......................................................................................................................... 8
4.2.6. LoZ function ................................................................................................................................ 9
4.2.7. AC+DC function .......................................................................................................................... 9
4.2.8. Inrush current function (INRUSH) .............................................................................................. 9
4.2.9. Disabling the Auto Power Off function ........................................................................................ 9
4.2.10.Setting the full scale for flexible clamp transducer ................................................................... 10
5.OPERATING INSTRUCTIONS ................................................................................... 11
5.1. DC voltage measurement ................................................................................................ 11
5.2. AC, AC+DC voltage measurement ................................................................................. 12
5.3. AC, DC, AC+DC voltage with low impedance (LoZ) ....................................................... 13
5.4. Resistance measurement and continuity test .................................................................. 14
5.5. Phase sequence and phase concordance with 1 terminal .............................................. 15
5.6. Measurement of insulation resistance ............................................................................. 17
5.7. Continuity of protective conductors with 200mA ............................................................. 23
5.7.1. Function ZERO – Zeroing of test cable resistance ................................................................... 27
5.8. Measurement of DC, AC, AC+DC, INRUSH current with clamp transducers ................. 29
6.MAINTENANCE ......................................................................................................... 33
6.1. Battery replacement ........................................................................................................ 33
6.2. Cleaning the instrument .................................................................................................. 33
6.3. End of life ........................................................................................................................ 33
7.TECHNICAL SPECIFICATIONS ................................................................................ 34
7.1. Technical characteristics ................................................................................................. 34
7.1.1. General characteristics ............................................................................................................. 36
7.2. Environmental conditions for use .................................................................................... 36
7.3. Accessories ..................................................................................................................... 36
8.ASSISTANCE ............................................................................................................. 37
8.1. Warranty conditions ......................................................................................................... 37
8.2. Assistance ....................................................................................................................... 37
9.THEORETICAL APPENDIXES ................................................................................... 38
9.1. Continuity of protective conductors ................................................................................. 38
9.2. Measurement of insulation resistance ............................................................................. 39
9.2.1. Polarization Index (PI) .............................................................................................................. 42
9.2.2. Dielectric Absorption Ratio (DAR) ............................................................................................ 42
9.3. Voltage and current harmonics ....................................................................................... 43
NEPTUNE
1. PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1 relevant to electronic measuring instruments. For your safety and in order to prevent damaging the instrument, please carefully follow the procedures described in this manual and read all
notes preceded by symbol with the utmost attention. Before and after carrying out measurements, carefully observe the following instructions:
Do not carry out any measurement in humid environments. Do not carry out any measurements in case gas, explosive materials or flammables are
present, or in dusty environments.
Avoid any contact with the circuit being measured if no measurements are being
carried out.
Avoid contact with exposed metal parts, with unused measuring probes, etc. Do not carry out any measurement in case you find anomalies in the instrument such
as deformation, breaks, substance leaks, absence of display on the screen, etc.
Pay special attention when measuring voltages higher than 50V, since a risk of
electrical shock exists.
In this manual, and on the instrument, the following symbols are used:
Warning: observe the instructions given in this manual; improper use could damage the instrument or its components.
High voltage danger: electrical shock hazard.
Double-insulated meter
AC voltage or current
DC voltage or current
Connection to earth
1.1. PRELIMINARY INSTRUCTIONS
This instrument has been designed for use in environments of pollution degree 2. It can be used for VOLTAGE and CURRENT measurements on installations with CAT
IV 600V, CAT III 1000V to earth and between inputs.
We recommend following the normal safety rules devised by the procedures for
carrying out operations on live systems and using the prescribed PPE to protect the user against dangerous currents and the instrument against incorrect use.
In case the lack of indication of the presence of voltage may represent a danger for the
operator, always carry out a continuity measurement before carrying out the measurement on the live system, in order to confirm the correct connection and condition of the leads.
Only the leads supplied with the instrument guarantee compliance with the safety
standards. They must be in good conditions. If necessary, only replace them with original HT accessories.
Do not test circuits exceeding the specified voltage limits. Do not perform any test under environmental conditions exceeding the limits indicated
in § 6.2.1
Check that the battery is correctly inserted. Make sure that the LCD display and the rotary switch indicate the same function.
NEPTUNE
1.2. DURING USE
Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the caution notes and/or instructions may damage the instrument and/or its components or be a source of danger for the operator.
Before activating the rotary switch, disconnect the test leads from the circuit being
measured.
When the instrument is connected to the circuit being measured, do not touch any
unused terminal.
During current measurement, any other current near the clamps may affect
measurement precision.
When measuring current, always put the conductor as near as possible to the middle of
the clamp jaw, to obtain the most accurate reading.
Do not measure resistance in case external voltages are present; even if the
instrument is protected, an excessive voltage may cause malfunction.
Before attempting any resistance measurement, cut off power supply from the circuit to
be measured and make sure that all capacitors are discharged, if present.
While measuring, if the value or the sign of the quantity being measured remain
unchanged, check if the HOLD function is enabled.
1.3. AFTER USE
When measurement is complete, set the rotary switch to OFF If the instrument is not to be used for a long time, remove the batteries
1.4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements, defines what measurement category, commonly called overvoltage category, is. § 6.7.4 reads: Circuits are divided into the following measurement categories:
Measurement category IV is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Measurement category III is for measurements performed on installations inside
buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.
Measurement category II is for measurements performed on circuits directly
connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user.
NEPTUNE
2. GENERAL DESCRIPTION
The instrument carries out the following measurements:
DC / AC, AC+DC TRMS voltage DC / AC / AC+DC TRMS voltage with low impedance (LoZ) DC / AC / AC+DC TRMS current with standard clamp transducer AC TRMS current with flexible clamp transducers Automatic recognition of AC and DC quantities Inrush current (Dynamic INRUSH - DIRC) Current/voltage harmonics up to the 25th and THD% calculation Resistance and Continuity test Current and voltage frequency Insulation resistance with test voltage 50,100,250,500,1000VDC Measurement of Polarization Index (PI) and Dielectric Absorption Ratio (DAR) Continuity of protective conductor with 200mA Phase sequence with 1 terminal
Each of these functions can be selected by means of the appropriate switch. The instrument is also equipped with function keys (see § 4.2), an analogue bargraph and backlight. The instrument is also equipped with an Auto Power OFF function (which can be disabled), which automatically switches off the instrument 15 minutes after the last time a function key was pressed or the rotary switch was turned. To switch on the instrument again, turn the rotary switch.
2.1. MEASURING AVERAGE VALUES ANDTRMS VALUES
Measuring instruments of alternating quantities are divided into two big families: AVERAGE-VALUE meters: instruments measuring the value of the sole wave at
fundamental frequency (50 or 60 Hz).
TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMS
value of the quantity being tested. With a perfectly sinusoidal wave, the two families of instruments provide identical results. With distorted waves, instead, the readings shall differ. Average-value meters provide the RMS value of the sole fundamental wave; TRMS meters, instead, provide the RMS value of the whole wave, including harmonics (within the instruments bandwidth). Therefore, by measuring the same quantity with instruments from both families, the values obtained are identical only if the wave is perfectly sinusoidal. In case it is distorted, TRMS meters shall provide higher values than the values read by average-value meters.
2.2. DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR
The root mean square value of current is defined as follows: In a time equal to a period, an alternating current with a root mean square value of 1A intensity, circulating on a resistor, dissipates the same energy that, during the same time, would be dissipated by a direct current with an intensity of 1A". This definition results in the numeric expression:
Tt
0
1
2
dttg
G=
T
The Crest Factor is defined as the relationship between the Peak Value of a signal and its
RMS value: CF (G)=
sinusoidal wave it is 2=1.41. In case of distortion, the Crest Factor takes higher values as wave distortion increases.
)(
t
0
The root mean square value is indicated with the acronym RMS.
G
p
This value changes with the signal waveform, for a purely
G
RMS
NEPTUNE
3. PREPARATION FOR USE
3.1. INITIAL CHECKS
Before shipping, the instrument has been checked from an electric as well as mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged.
However, we recommend generally checking the instrument in order to detect possible damage suffered during transport. In case anomalies are found, immediately contact the forwarding agent.
We also recommend checking that the packaging contains all components indicated in §
6.3.1. In case of discrepancy, please contact the Dealer.
In case the instrument should be returned, please follow the instructions given in § 7.
3.2. INSTRUMENT POWER SUPPLY
The instrument is supplied with 4x1.5V alkaline batteries type AAA IEC LR03, included in the package. When batteries are flat, the symbol the batteries, see § 6.1.
3.3. STORAGE
In order to guarantee precise measurement, after a long storage time, wait for the instrument to come back to normal condition (see § 6.2.1).
is shown on the display. To replace
NEPTUNE
4. NOMENCLATURE
4.1. DESCRIPTION OF THE INSTRUMENT
CAPTION:
1. LCD display
2. Key / ZERO
3. Key/
4. Key MODE/MXMNPK
5. Key H/H%/H
6. Key V
TEST/LIM
7. Key GO/HOLD
8. Rotary selector switch
9. Input terminal COM/
10. Input terminal V /+
Fig. 1: Description of the instrument
4.1.1. Instruments initial screen
1. Turn the rotary switch to any position to switch on the instrument. The following initial
screen is shown on the display for a few seconds to identify the internal Hardware and
Firmware version.
Fig. 2: Instruments initial screen
2. Turn the rotary switch to OFF to switch off the instrument.
NEPTUNE
4.2. DESCRIPTION OF FUNCTION KEYS
4.2.1. Key GO/HOLD
Pressing the key GO/HOLD (for functions V , LoZV , and ) makes the instrument hold the value of the quantity shown on the display. The message HOLDappears on the display. Press the key again to exit the function. Pressing key GO/HOLD (for functions
M, Lo, , IRC) activates the corresponding measurement.
4.2.2. Key H/H%/H
Key H/H%/H (active in positions V , LoZV and ) allows the following operations:
Simple pressing of the key to display the amplitude of voltage and current harmonics
up to the 25th (Hdc, H01H25) in absolute or percentage format in relationship with
the fundamentals of input signals (for voltage values VAC >0.5V and current values AC
> 0.5A and frequency in the range 42.5Hz ÷ 69Hz) and the percentage value of
parameter THD% (see § 9.3) as shown in Fig. 3. Use keys /ZERO and /
to
increase/decrease the order of the harmonic.
Fig. 3: Display of amplitude of harmonic analysis
Long pressing of the key (at least 2s) in order to activate the H2O (Higher Harmonic
Ordering) ordering function of the amplitude of harmonics. In these conditions, function
HOLDis automatically activated and symbol oappears next to the displayed
harmonic order to indicate that the Ordering function is activated. The bargraph is
disabled and the instrument shows the value of amplitude of all the harmonics between
the DC value and the 25th, fundamental harmonic excluded, in a decreasing order
starting from the harmonic with the highest amplitude, as shown in Fig. 4
Fig. 4: Display of ordering of amplitude of harmonic analysis
In the example in Fig. 4, the harmonic with the highest value corresponds to the 7th.
Press key to observe the amplitude of the remaining harmonics and press key
H/H%/H again to switch between display in absolute and percentage values. Turn
the rotary switch to exit the function.
NEPTUNE
4.2.3. Key MODE/MXMNPK A simple pressing of key MODE/MXMNPK allows for the following operations:
Selection of measuring modes AUTO, AC, DC, AC+DCand FREQin positions
V , LoZV
Selection of measuring modes AUTO, AC, DCand AC+DC, FREQand IRC
(see § 4.2.8) in position
Selection of clamp transducer type in current measurement between options “ ”
(optional standard clamp) and “ ” (optional flexible clamp) in position
Selection of measurements AUTO” “TMRand PIin position M (see § 5.6). Selection of measurements AUTOand TMRin position Lo (see § 5.7). Selection of measurement of Resistance or Continuity test in position
A long pressing (>2s) of key MODE/MXMNPK allows for the activation/deactivation of continuous detection of maximum value (MAX), minimum value (MIN), positive peak (Pk+), negative peak (Pk-) of the quantity (voltage or current) to be measured. The values are constantly updated and are displayed cyclically every time the same key is pressed. This
function is not active in position . Press and hold key MODE/MXMNPK (>2s) or use the selector to exit the function.
4.2.4. Keys / e / ZERO A simple pressing of keys / and / ZERO allows for the following operations::
Setting of the full scale value of flexible clamp transducer (optional accessory - option
“ ” ) in position among values: 30A, 300A, 3000A for AC current measurement.
Setting of the full scale value of standard clamp transducer (option “ ” ) in position
among values: 1A, 10A, 30A, 40A, 100A, 200A, 300A, 400A, 1000A, 2000A, 3000A
for AC and DC current measurement.
 Selection of harmonic order DC ÷ 25°in positions V , LoZV and  Selection of calculation time of RMS value in DIRC function (see § 4.2.8).  Setting of threshold values in positions M and Lo.  Once insulation measurement is complete, the results appear on the display (M,
Vgen, PI, DAR).
Long pressing (>2s) of key / allows activating/deactivating display backlight. This function is activated in any position of the rotary switch and is automatically deactivated after approx. 2 minutes following operations:
idling. Long pressing (>2s) of key / ZERO allows for the
Zeroing of cable resistance in position Lo (see § 5.7) Zeroing of cable resistance in position (see § 5.4)
4.2.5. Key VTEST/LIM A simple pressing of key VTEST/LIM allows for the following operations:
Selecting test voltage in insulation measurement among the following options: 50V,
100V, 250V, 500V, 1000VDC in position M
Long pressing (>2s) of key VTEST/LIM allows for the following operations:
Setting the minimum threshold in insulation measurement among the following options:
no (no threshold), 0.10M, 0.230M, 0.50M, 1.00M, 100M in position M. Setting the maximum threshold for continuity test in range: 0.05 ÷ 9.99 in position
Lo.
NEPTUNE
4.2.6. LoZ function
This mode allows carrying out AC/DC voltage measurement with a low input impedance, in order to eliminate wrong readings due to parasite voltages for capacitive couplings.
CAUTION
By plugging the instrument between the phase and earth conductor,
because of the low impedance of the instrument during measurement, RCD protections may trip while carrying out the test. If this test is to be carried out, first carry out a measurement of at least 5s between phase and neutral in the presence of voltage
Do not leave the instrument connected for more than 1min
4.2.7. AC+DC function
The instrument is capable of measuring a possible presence of overlapping alternating components on a generic direct waveform (voltage or current). This can be useful when measuring typical impulsive signals of non-linear loads (e.g. welding machines, ovens, etc.).
4.2.8. Inrush current function (INRUSH)
Measurement of inrush current (see § 5.8) is intended as the recognition of an event detected upon exceeding of a trigger threshold. If the instant value exceeds this threshold (fixed, equal to 1%FS clamp), the instrument shows on the display the maximum Peak value (calculated in 1ms) and the maximum RMS value calculated with a time which can be selected among options: 16.7ms, 20ms, 50ms, 100ms (default), 150ms, 175ms and
200ms.
4.2.9. Disabling the Auto Power Off function
In order to preserve internal batteries, the instrument switches off automatically approximately 15 minutes after it was last used. Press key MODE/MXMNPK or turn the rotary switch from position OFF to switch on the instrument again. To disable the Auto Power Off function, proceed as follows:
Switch off the instrument (OFF) Press and hold key to switch on the instrument. The symbol
disappears from
the display Switch off and then on again the instrument to enable the function.
NEPTUNE
4.2.10. Setting the full scale for flexible clamp transducer
The instrument can be used with a flexible clamp transducer (optional accessory). For a correct current measurement, it is necessary to set the full scale of voltage for the clamp in use (please refer to the transducers user manual for the correct full scale value to be set). Proceed as follows:
1. Switch off the instrument (OFF)
2. Press and hold key MODE/MXMNPK and switch on the instrument by turning the
rotary switch. The following screen appears on the display:
Fig. 5: Setting the full scale for flexible clamp transducer
3. Press keys /
or to set the full scale value of the clamp in use among the options:
3VAC (model F3000U) or 1VAC (other models).
4. Press key GO/HOLD to confirm and go back to measuring screen.
5. Settings are maintained very time the instrument is switched on.
EN - 10
NEPTUNE
5. OPERATING INSTRUCTIONS
5.1. DC VOLTAGE MEASUREMENT
The maximum input DC voltage is 1000V. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
CAUTION
Fig. 6: Use of the instrument for DC voltage measurement
1. Select position V
2. Insert the red cable into input terminal V /+ and the black cable into input terminal
COM.
3. Position the red lead and the black lead respectively in the spots with positive and
negative potential of the circuit to be measured (see Fig. 6). The display shows the
value of voltage.
4. If the display shows the message >1000V(see Fig. 6), the maximum measurable
value has been reached.
5. When symbol "-" appears on the instruments display, it means that voltage has the
opposite direction with respect to the connection in Fig. 6.
6. To use functions HOLD, MAX/MIN/PK, see § 4.2
EN - 11
NEPTUNE
5.2. AC, AC+DC VOLTAGE MEASUREMENT
CAUTION
The maximum input AC voltage is 1000V to earth. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
Fig. 7: Use of the instrument for AC voltage measurement
1. Select position V
2. Press the MODE/MXMNPK key until symbol ACor AC+DCis displayed. The
instrument automatically recognizes AC or DC signals.
3. Insert the red cable into input terminal V /+ and the black cable into input terminal
COM/.
4. Position the red lead and the black lead respectively in the spots of the circuit to be
measured (see Fig. 7). The display shows the value of voltage. On the top of the
display on the right side, the value of voltage frequency is displayed. Press key
MODE/MXMNPK to display the value of frequency with a higher resolution.
5. If the display shows the message >1000V(see Fig. 7), the maximum measurable
value has been reached.
6. If the display shows the messages <32Hzor >1000Hz(see Fig. 7), the value of
frequency is out of the measuring range 32Hz ÷ 1000Hz.
7. To use functions HOLD, MAX/MIN/PK, H/H%/H, see § 4.2
EN - 12
Loading...
+ 32 hidden pages