The information contained in this document is subject to change
without notice.
Agilent T echnologies makes no war ranty of any kind with r egard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a partic ular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Safety Information
The following safety symbols are used throughout this manual.
Familiarize yourself with the symbols and their meaning before
operating this instrument.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in dam a g e to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
NOTENote calls out special information for the user’s attention. It provides
operational information or additional instructions of which the user
should be aware.
The instruction documentation symbol. The product is
marked with this symbol when it is necessary for the
user to refe r to th e instructions in the docu mentation .
This symbol is used to mark the on position of the
power line switch.
This symbol is used to mark the s tandby p osit ion o f the
power line switch.
This symbol indicates that the input power required is
AC.
2
WARNINGThis is a Safety Class 1 Product (provided with a protective
earth ground incorporated in the power cord). The mains plug
shall be inserted only in a socket outlet provided with a
protected earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption i s prohib ited.
WARNINGNo operator serviceable parts inside. Refer servicing to
qualified personnel. To prevent electrical shock do not remove
covers.
WARNINGIf this product is not used as specified, the protection provided
by the equipment could be impaired. This product must be used
in a normal condition (in which all means for protection are
intact) only.
CAUTIONAlways use the three-prong AC power cord supplied with this product.
Failure to ensure adequate grounding may cause product damage.
Where to Find the Latest Information
Documentation is updated periodically. Fo r the latest information about
Agilent Technologies PSA and ESA spectrum analyzers, including
firmware upgrades and application information, please visit the
following Internet URL:
Using C to Make a Power Suite ACPR Measurement on a cdmaOne Signal . . . . . . . . . . 176
Using C to Serial Poll the Analyzer to Determine when an Auto-alignment is Complete . 179
Using C and Service Request (SRQ) to Determine When a Measurement is Complete . . 182
Using Agilent VEE to Capture the Equivalent SCPI Learn String . . . . . . . . . . . . . . . . . . .342
8
Recommended Test Equipment
1Recommended Test Equipment
9
Recommended Test Equipment
NOTETo find descriptions of specific analyzer functions, for the ESA, refer to
the Agilent Technologies ESA Series Spectrum Analyzers
User’s/Programmer’s Reference Guide and for the PSA, refer to the
Agilent Technologies PSA Series Spectrum Analyzers User’s and
Programmer’s Reference Guide.
Test EquipmentSpecificationsRecommended Model
Signal Sources
Signal Generator (2)0.25 MHz to 4.0 GHz
Ext Ref Input
Adapters
Type-N (m) to BNC (f) (3)1250-0780
Termination, 50
Comparing Signals on the Same Screen Using Marker Delta
Comparing Signals on the Same Screen Using
Marker Delta
Using the analyzer, you can easily compare freque ncy and amplitude
differences between signals, such as radio or television signal spectra.
The analyzer delta marker function lets you compare two signals when
both appear on the screen at one time .
In this procedure, the analyzer 10 MHz signal is used to measure
frequency and amplitude differences between two signals on the same
screen. Delta marker is used to demonstrate this comparison.
Figure 2-1An Example of Comparing Signals on the Same Screen
Step 1. Preset the an alyzer:
Press
Step 2. (PSA)
Preset, Factory Preset (if present).
a. Enable the rear panel 10 MHz output.
Measuring Multiple Signals
Press
System, Reference, 10 MHz Out (On).
b. Connect the 10 MHz OUT (SWITCHED) from the rear panel to t he front
panel RF input.
(ESA)
Connect the rear panel 10 MHz REF OUT to the front panel RF input.
Step 3. Set the analyzer center frequency, span and reference level to view the
10 MHz signal and its harmonics up to 50 MHz:
Press
Press
Press
12Chapter 2
FREQUENCY Channel, Center Freq, 30, MHz.
SPAN X Scale, Span, 50, MHz.
AMPLITUDE Y Scale, Ref Level, 10, dBm.
Measuring Multiple Signals
Comparing Signals on the Same Screen Using Marker Delta
Step 4. Place a marker at the highest peak on the display (10 MHz):
Press
The
Peak Search.
Next Pk Right and Next Pk Left softkeys are available to move the
marker from peak to peak. The marker should be on the 10 MHz
reference si g n a l :
Step 5. Anchor the first marker and activate a second marker:
Press
Marker, Delta.
The label on the first marker now reads 1R , indicating that it is the
reference point.
Step 6. Move the second marker to another signal peak using the front-panel
knob or by using the
Press
Press
Peak Search, Next Peak or
Peak Search, Next Pk Right or Next Pk Left.
Peak Search key:
The amplitude and frequency difference between the markers is
displayed in the active function block. For ESA see the lef t side of
Figure 2-2 and the right side for PSA.
Figure 2-2Using the Delta Marker Function (ESA left, PSA right)
NOTEThe resolution of the marker readings can be increased by turning on
the frequency count function.
Measuring Multiple Signals
Chapter 213
Measuring Multiple Signals
Comparing Signals on the Same Screen Using Marker Delta Pair
Comparing Signals on the Same Screen Using
Marker Delta Pair
In this procedure, the analyzer 10 MHz signal is used to measure
frequency and amplitude differences between two signals on the same
screen using the delta pair marker function.
Step 1. Refer to the previous procedure “Comparing Signals on the Same
Screen Using Marker Delta” on page 12 and follow steps 1, 2 and 3.
Step 2. Turn on Delta Pair reference marker to compare the 10 MHz signal and
the 30 M Hz signal:
Press
Note that the
Step 3. Use the knob or Peak Search to move the second marker (labeled 1) to
Peak Search, Marker, Delta Pair (ref).
Delta Pair marker does not anchor the first marker.
the 30 MHz peak:
Press
Step 4. Use the front panel knob to move the ref marker to the 20 MHz peak:
Peak Search, Next Peak or Next Pk Right.
The active function displays the amplit ude and frequency difference
between the 20 MHz and 30 MHz peaks as shown in Figure 2-3.
Figure 2-3Using the Delta Pair Marker Function (ESA left, PSA right)
Measuring Multiple Signals
NOTEIn Figure 2-3 notice that the active function reado ut has moved to the
top left of the analyzer display. The active function position has three
positions: top, center and bott om. To modify the active function pos ition:
Press
Display, Active Fctn Position, Top (Center, or Bottom).
Center position is the factory default setting.
14Chapter 2
Measuring Multiple Signals
Comparing Signals not on the Same Screen Using Marker Delta
Comparing Signals not on the Same Screen
Using Marker Delta
Measure the frequency and amplitude difference between two signals
that do not appear on the screen at one time. (This technique is useful
for harmonic distortion tes ts when narr ow span and narrow b andwidth
are necessary to measure the low level harmonics.)
In this procedure, the analyzer 10 MHz signal is used to measure
frequency and amplitude differences bet we en one signal on screen and
one signal off screen. Delta marker is used to demonstrate this
comparison.
Figure 2-4Comparing One Signal on Screen with One Signal Off Screen
Step 1. Preset the an alyzer:
Press
Step 2. (PSA)
a. Enable the rear panel 10 MHz output:
Press
b. Connect the 10 MHz OUT (SWITCHED) from the rear panel to t he front
panel RF input:
(ESA)
Connect the rear panel 10 MHz REF OUT to the front panel RF input.
Preset, Factory Preset (if present).
System, Reference, 10 MHz Out (On).
Measuring Multiple Signals
Chapter 215
Measuring Multiple Signals
Comparing Signals not on the Same Screen Using Marker Delta
Step 3. Set the center frequency, span and reference level to view only the
10 MHz signal:
Press
Press
Press
Step 4. Place a marker on the 10 MHz peak and then set the center frequency
FREQUENCY Channel, Center Freq, 10, MHz.
SPAN X Scale, Span, 5, MHz.
AMPLITUDE Y Scale, Ref Level, 10, dBm.
step size equal to the marker frequency (10 MHz):
Press
Press
Step 5. Activate the marker delta function:
Press
Step 6. Increase the center frequency by 10 MHz:
Press
Peak Search.
Marker →, Mkr → CF Step.
Marker, Delta.
FREQUENCY Channel, Center Freq, ↑.
The first marker moves to the left edge of the screen, at the amplitude
of the first sign a l peak.
Figure 2-5 shows the reference annotation for the delta marker (1R) at
the left side o f th e display, indicating that th e 10 MHz refere n ce si g n a l
is at a lower frequency than the frequency range currently displayed.
The delta marker appears on the peak of the 20 MHz component. The
delta marker annotation displays the amplitude and frequency
difference between the 10 and 20 MHz signal peaks.
Figure 2-5Delta Marker with Reference Signal Off-Screen (ESA)
Measuring Multiple Signals
Step 7. Turn the markers off:
Press
Marker, Off.
16Chapter 2
Resolving Signals of Equal Amplitude
In this procedure a decrease in resol ution bandwidth is used in
combination with a decrease in video bandwidth to resolve two signals
of equal amplitude with a freq uency s eparation of 100 kHz. Notice that
the final RBW selection to resolve the signals is the same width as the
signal separation while the VBW is slightly narrowe r than the RBW.
Step 1. Connect two sources to the analyzer input as shown in Figure 2-6.
Figure 2-6Setup for Obtaining Two Signals
Measuring Multiple Signals
Resolving Signals of Equal Amplitude
Step 2. Set one source to 300 MHz. Set the frequency of the other source to
300.1 MHz. Set both source amplitudes to
both signals should be approximately
−20 dBm. The amplitude of
−20 dBm at the output of the
bridge.
Step 3. Setup the analyzer to view the signals:
Press
Press
Preset, Factory Preset (if present).
FREQUENCY Channel, Center Freq, 300, MHz.
Press BW/Avg, Res BW, 300, kHz.
Press
SPAN X Scale, Span, 2, MHz.
A single signal peak is visible. See Figure 2-7 for an ESA example.
NOTEIf the signal peak is not present on the display, span out to 20 MHz,
turn signal tracking on, span back to 2 MHz and turn signal tracking
off.:
Figure 2-7Unresolved Signals of Equal Amplitude (ESA)
Step 4. Change the resolution bandwidth (RBW) to 100 kHz so that the RBW
setting is less than or equal to the frequency separation of the two
signals:
Press
BW/Avg, Res BW, 100, kHz.
Notice that the peak of the signal has become flattened indicating that
two signals may be present .
Step 5. Decrease the vi d e o b a n dw i dth to 10 kHz:
Press
Video BW, 10, kHz.
Two signals are now visible as shown with the ESA on the left side in
Figure 2-8 and the PSA on the right side. Use the front-panel knob or
step keys to further reduce the resolution bandwidth and better resolve
the signals.
Figure 2-8Resolving Signals of Equal Amplitude (ESA left, PSA right)
Measuring Multiple Signals
18Chapter 2
Measuring Multiple Signals
Resolving Signals of Equal Amplitude
As the resolution bandwidth is decreased, resolution of the individual
signals is improved and the sweep time is increased. For fastest
measurement times, use the widest possible resolution bandwidth.
Under factor y pr eset conditi o n s, the re so l u ti on bandwidt h is “coupled”
(or linked) to the span.
Since the resolution bandwidth has been changed from the coupled
value, a # mark appears next to Res BW in the lower-left corner of the
screen, indicating that the resolution bandwidth is uncoupled. (For
more infor m a ti on on coupli n g, ref e r to the
Auto Couple key descr i ption
in the Agilent Technologies ESA Spectrum Analyzers
User’s/Programmer’s Reference Guide and the PSA Spectrum
Analyzers User’s/Programmer’s Reference Guide.)
NOTETo res olve two signals of equal amplitude with a frequency separation
of 200 kHz, the resolution bandwidth must be less than the signal
separation so a resolution bandwidth of 100 kHz must be used. (For
analyzers that use a 1-3-10 RBW step sequence, a 100 kHz RBW is the
best choice for signal separation, but for high performance analyzers,
like the PSA, a 180 kHz RBW can be selected by fine tuning the RBW
filters at 10% increments.) Filter widths above 200 kHz exceed the
200 kHz signal separation and would not resolve the signals.
Measuring Multiple Signals
Chapter 219
Measuring Multiple Signals
Resolving Small Signals Hidden by Large Signals
Resolving Small Signals Hidden by Large
Signals
This procedure uses narrow resoluti on bandwidths to res olve two input
signals with a frequency separation of 155 kHz and an amplitude
difference of 60 dB .
Step 1. Connect two sources to the analyzer input as shown in Figure 2-6.
Step 2. Set one source to 300 MHz at −10 dBm. Set the second source to
300.05 MHz, so that the signal is 50 kHz higher than the first signal.
Set the amplitude of the signal to
signal).
Step 3. Set the analyzer as follows:
−70 dBm (60 dB below the first
Press
Press
Press
Press
NOTEIf the signal peak is not present on the display, span out to 20 MHz,
Preset, Factory Preset (if present).
FREQUENCY Channel, Center Freq, 300, MHz.
BW/Avg, 30, kHz.
SPAN X Scale, Span, 500, kHz.
turn signal tracking on, span back to 2 MHz and turn signal tracking
off:
Step 4. Set the 300 MHz signal to the reference level:
Press
Measuring Multiple Signals
NOTEThe ESA 30 kHz filter shape factor of 15:1 (PSA is 4.1:1) has a
bandwidth of 450 kHz at the 60 dB point (PSA has a BW of 123 kHz).
Peak Search, Mkr →, Mkr → Ref Lvl.
The half-bandwidth (225 kHz for ESA and 61.5 kHz for PSA) is NOT
narrower than the frequency separation of 50 kHz, so the input signals
can not be resolved.
20Chapter 2
Measuring Multiple Signals
Resolving Small Signals Hidden by Large Signals
Figure 2-9Signal Resolution with a 30 kHz RBW (ESA left, PSA right)
Step 5. Reduce the resolution bandwidth filter to view the smaller hidden
signal. Place a delta marker on the smaller signal:
Press
Press
NOTEThe ESA 1 kHz filter shape factor of 15:1 (PSA is 4.1:1) has a
BW/Avg, 1, kHz.
Peak Search, Marker, 50, kHz.
bandwidth of 15 kHz at the 60 dB point (PSA has a BW of 4.1 kHz). The
half-bandwidth (7.5 kHz for ESA and 2.05 kHz for PSA) is narrower
than 50 kHz, so the input signals can be resolved.
Figure 2-10Signal Resolution with a 1 kHz RBW (ESA left, PSA right)
Measuring Multiple Signals
NOTETo determine the resolution capability for intermediate amplitude
differences, assume the filter skirts between the 3 dB and 60 dB points
are parabolic, like an ideal Gaussian filter. The resolution capability is
approximately:
2
∆f
12.04 dB
-------------
•
RBW
where ∆f is the separation between the signals.
Chapter 221
Measuring Multiple Signals
Decreasing the Frequency Span Around the Signal
Decreasing the Frequency Span Around the
Signal
Using the analyzer signal track func tion, you can quickly decrease the
span while keeping the signal at center frequency. This is a fast way to
take a closer look at the area around the signal to identify signals that
would otherwise not be resolved.
This procedure uses signal trac king and span zoom to view the analyzer
50 MHz reference signal in a 200 kHz span.
Step 1. Perform a factory preset:
Press
Step 2. Enable the internal 50 MHz amplitude reference signal of the analyzer
Preset, Factory Preset (if present).
as follows:
(PSA)
Press
Input/Output, Input Port, Amptd Ref.
(ESA E4401B and E4411B)
Press
Input/Output, Amptd Ref (On).
(ESA E4402B, E4403B, E4404B, E4405B, E4407B and E4408B)
Connect a cable from the front panel AMPTD REF OUT to the analyz er
RF input:
Press
Step 3. Set the start frequency to 20 MHz and the stop frequency to 1 GHz:
Press
Press
Step 4. Place a marker at the peak:
Press
Measuring Multiple Signals
Step 5. Turn on the signal tracking function to move the signal to the center of
Input/Output, Amptd Ref Out (On).
FREQUENCY Channel, Start Freq, 20, MHz.
FREQUENCY Channel, Stop Freq, 1, GHz.
Peak Search.
the screen (if it is not already positioned there):
Press
FREQUENCY Channel, Signal Track (On).
See the left-side of figure Figure 2-11. (Note that the marker must be on
the signal before turning signal track on.)
NOTEBecause the signal track f unction automaticall y maintains the sig nal at
the center of the screen, you can reduce the span quic k l y for a closer
look. If the signal drifts off of the screen as you decrea se the span, use a
wider frequency span. (You can also use
as a quick way to perform the
SPAN key sequence.)
22Chapter 2
Peak Search, FREQUENCY, Signal Track,
Span Zoom, in the SPAN menu,
Measuring Multiple Signals
Decreasing the Frequency Span Around the Signal
Step 6. Reduce span and resolution bandwidth to zoom in on the marked
signal:
Press
NOTEIf the span change is large enough, the span decreases in steps as
SPAN X Scale, Span, 200, kHz.
automatic zoom is completed . See Figure 2-11 on the right side. You can
also use the front-panel knob or step k eys to decrease the span and
resolution bandwidth values.
Step 7. Turn off signal tracking:
Press
FREQUENCY Channel, Signal Track (Off).
Figure 2-11Signal Tracking
LEFT: Signal tracking on before span decrease
RIGHT: After zooming in on the signal
Measuring Multiple Signals
Chapter 223
Measuring Multiple Signals
Decreasing the Frequency Span Around the Signal
Measuring Multiple Signals
24Chapter 2
3Measuring a Low−Level Signal
25
Measuring a Low
−
Level Signal
Measuring a Low−Level Signal
Reducing Input Attenuation
Reducing Input Attenuation
The ability to measure a low-level signal is limited by internally
generated noise in the spectrum analyzer. The measurement setup can
be changed in several ways to improve the analyzer sensitivity.
The input attenuator affects the level of a signal pas sing through the
instrument. If a signal is very close to the noise floor, reducing input
attenuation can bring the signal out of the noise.
CAUTIONEnsure that the total power of all input signals at the analyzer RF
input does not exceed +30 dBm (1 watt).
Step 1. Preset the an a l y z er:
Press Preset, Factory Preset (if present).
Step 2. Set the freque n cy o f th e si g n a l so u r ce to 30 0 MHz. Set the source
amplitude to
RF INPUT.
−80 dBm. Connect the source RF OUTPUT to the analyzer
Step 3. Set the center frequency, span and reference level:
Press
Press
Press
FREQUENCY Channel, Center Freq, 300, MHz.
SPAN X Scale, Span, 5, MHz.
AMPLITUDE Y Scale, Ref Level, 40, −dBm.
Step 4. Move the desired peak (in this example, 300 MHz) to the center of the
display:
Press
Peak Search, Marker ➞, Mkr ➞ CF.
Step 5. Reduce the span to 1 MHz (as shown in Figure 3 -1) and if necessary
re-center the peak:
Press
Span, 1, MHz.
Figure 3-1 Measuring a Low-Level Signal (ESA Display)
Level Signal
−
Measuring a Low
26Chapter 3
Step 6. Set the attenuation to 20 dB:
Measuring a Low−Level Signal
Reducing Input Attenuation
Press
AMPLITUDE Y Scale, Attenuation, 20, dB.
Note that increasing the attenuation moves the noise floor closer to the
signal level.
A “#” mark appears next to the Atten annotation at the top of the
display, indicating that the attenuation is no longer coupled to other
analyzer settings.
Step 7. To see the signal more clearly, set the attenuation to 0 dB:
Press
AMPLITUDE, Attenuation, 0, dB.
See Figure 3-2 shows 0 dB input attenuation.
Figure 3-2 Measuring a Low-Level Signal Using 0 dB Attenuation (ESA)
CAUTIONWhen you finish this example, increase the attenuation to protect the
analyzer’s RF inpu t:
Press AMPLITUDE Y Scale, Attenuation (Auto) or press Auto Couple.
NOTEAll figures in this chapter are screen captures from an ESA. Display
and numerical results may be different for a PSA.
Chapter 327
Measuring a Low
−
Level Signal
Measuring a Low−Level Signal
Decreasing the Resolution Bandwidth
Decreasing the Resolution Bandwidth
Resolution bandwidth settings affect the level of internal noise without
affecting the level of continuous wave (CW) signals. Decreasing the
RBW by a decade reduces the noise floor by 10 dB.
Step 1. Refer to the first procedure “Reducing In put Attenuation” on page 26 of
this chapter and follow steps 1, 2 and 3.
Step 2. Decrease the resolution bandwidth:
Press
BW/Avg, ↓.
The low-level signal appears more clearly because the noise level is
reduced (see Figure 3-3).
Figure 3-3 Decreasing Resolution Bandwidth
A “#” mark appears next to the Res BW annotation in the lower left
corner of the screen, indicating that the resolution bandwidth is
uncoupled.
RBW SelectionsYou can use the step keys to change the RBW in a 1
−3−10 sequence.
For ESA, RBWs below 1 kHz are digital and have a selectivity ratio of
5:1 while RBWs at 1 kHz and higher have a 15:1 selectivity ratio. The
ESA’s maximum RBW is 5 MHz a nd the minimum is 1 Hz (optional).
All PSA RBWs are digital and have a selectivity ratio of 4.1:1. For PSA,
choosing the next lower RBW for better sensitivity increases the sweep
Level Signal
−
time by about 10:1 for swept measurements, and about 3:1 for FFT
measurements (within the limits of RBW). Using the knob or keypad,
you can select RBWs from 1 Hz to 3 MHz in approximately 10%
increments, plus 4, 5, 6 and 8 MHz. This enables you to make the trade
off between sweep tim e a n d se n s i t ivity with finer resolut i on.
Measuring a Low
28Chapter 3
Measuring a Low−Level Signal
Using the Average Detector and Increased Sweep Time
Using the Average Detector and
Increased Sweep Time
When the analyzer’s noise masks low-level signals, changing to the
average detector and increasing the sweep time smooths the noise and
improves the signal’s visibility. Slower sweeps are required to average
more noise variations.
Step 1. Refer to the first procedure “Reducing In put Attenuation” on page 26 of
this chapter and follow steps 1, 2 and 3.
Step 2. Select the average detector:
Press
Det/Demod, Detector, Average.
A “#” mark appears next to the Avg annotation, indicating that the
detector has been chosen manually (see Figure 3-4).
Step 3. Increase the sweep time to 100 ms:
Press
Sweep, Sweep Time, ↑.
Note how the noise smooths out, as there is more time to average the
values for each of the displ ayed data points .
Step 4. With the sweep time at 100 ms, change the average type to
log averaging:
(ESA) Press
(PSA) Press
BW/Avg, Avg Type, Video Avg.
BW/Avg, Avg/VBW Type, Log-Pwr.
Figure 3-4 Varying the Sweep Time with the Average Detector
Chapter 329
Measuring a Low
−
Level Signal
Measuring a Low−Level Signal
Trace Averaging
Trace Averaging
Averaging is a digital proces s in which each trace point is aver aged with
the previous average for the same trace point. Selecting averaging,
when the analyzer is autocoupled, changes the detection mode (from
peak in ESA and normal in PSA) to sample, smoothing the displa ye d
noise level. ESA sample mode displays the instantaneous value of the
signal at the end of the time or frequency interval represented by each
display point (for PSA it is t he ce nter of the t ime or fr equency interval),
rather than the value of the peak during the interval. Sample mode
may not measure a signal’s amplitude as accurately as normal mode,
because it may not find the true peak.
NOTEThis is a trace processing function and is not the same as using the
average detector (as described on page 29).
Step 1. Refer to the first procedure “Reducing In put Attenuation” on page 26 of
this chapter and follow steps 1, 2 and 3.
Step 2. Turn video averaging on:
Press
BW/Avg, Average (On).
As the averaging routine smooths the trace, low level signals become
more visible. Average 100 appears in the active function block.
Step 3. With average as the active function, set the number of averages to 25:
Press
25, Enter.
Annotation on the left side of the graticule shows the type of averaging
(the annotation for ESA is VAvg and is LgAv for PSA), and the number
of traces averaged.
Changing most active functions res tarts the averag ing , as do es toggling
Average key. Once the set number of sweeps com pletes , the analyzer
the
continues to provide a running average based on this se t nu mbe r.
NOTEIf you want the measurement to stop after the set number of sweeps,
use single sweep: Press
Sweep, Sweep (to select Single), or press Single
and then toggle the Average key.
Level Signal
−
Measuring a Low
30Chapter 3
Loading...
+ 316 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.