HP app an57 3 figure measurement 2 5980 1916e schematic

Page 1
10 Hints for Making Successful Noise Figure Measurements
Application Note 57-3
Page 2
2
Table of Contents
Introduction 3
HINT 1: Select the appropriate noise source 4
HINT 3: Minimize mismatch uncertainties 6
HINT 4: Use averaging to minimize display jitter 7
HINT 5: Avoid non-linearities 8
HINT 6: Account for mixer characteristics 9
HINT 7: Use proper measurement correction 11
HINT 8: Choose the optimal measurement
bandwidth 12
HINT 9: Account for path losses 13
HINT 10: Account for the temperature of
the measurement components 14 Appendix A: Checklist 15 Appendix B: Total uncertainty calculations 16 Appendix C: References 18 Appendix D: Abbreviations 18 Appendix E: Glossary and definitions 19 Key web resources 20
Page 3
3
Introduction
To achieve accurate and repeatable results at RF or microwave frequencies, measurement uncertainties and barriers to measurement repeatability must be minimized. The performance of a device can be obscured if errors are allowed to accumulate. For the most accurate measurement, it is important to understand the nature of the error contributors and identify which of these can be influenced or changed to improve the quality of the results. This application note provides useful tips that will assist in making accurate noise figure measurements.
The checklist in Appendix A is a helpful tool to verify that all hints have been considered for a particular measurement.
A detailed explanation of the uncertainties in noise figure measurements is provided in Agilent Technologies’ Application Note 57-2, Noise Figure Measurement Accuracy (see reference 5). For a general understanding of noise figure and its variety of measurement techniques, refer to Agilent Technologies Application Note 57-1, Fundamentals of Noise Figure Measurement (see reference 2).
Page 4
4
Frequency range
Commercial noise sources cover frequencies up to 50 GHz with choices of co-axial or waveguide connectors. The frequency range of the noise source must include the input frequency range of the DUT, of course. If the DUT is a mixer or frequency translation device, the output frequency range of the DUT must also be addressed. If one source does not include both frequency ranges, a second source will be required. A second noise source may also be necessary when measuring a non-frequency translating device with low noise and high gain. Low ENR is best for the measurement, however, high ENR is necessary to calibrate the full dynamic range of the instrument. In either case, a full-featured noise figure analyzer can account for the different ENR tables required for calibration and measurement.
Match
If possible, use a noise source with the lowest change in output impedance between its ON and OFF states. The noise source’s output impedance
changes between its ON and OFF states, which varies the match between the noise source and the DUT. This variation changes the gain and noise figure of the DUT, especially for active devices like GaAs FET amplifiers. To minimize this effect, 6 dB ENR noise sources are commercially available that limit their changes in reflection coefficient between ON and OFF states to better than 0.01 at frequencies to 18 GHz.
Adapters
Use a noise source with the correct connector for the DUT rather than use an adapter, particularly for devices
with gain. The ENR values for a noise source apply only at its connector. An adapter adds losses to these ENR values. The uncertainty of these losses increases the overall uncertainty of the measurement. If an adapter must be used, account for the adapter losses.
HINT 1:
Select the appropriate noise source
ENR
The output of a noise source is defined in terms of its frequency range and excess noise ratio (ENR). Nominal ENR values of 15 dB and 6 dB are commonly available. ENR values are calibrated at specific spot frequencies. The uncertainties of these calibrations vary over the frequency range of the noise source and add to the uncertainty of the measurement. This uncertainty is typically limited to approximately 0.1 dB using the root-sum-of-squares method (RSS).
Use a 15 dB ENR noise source for:
• general-purpose applications to measure noise figure up to 30 dB.
• user-calibrating the fullest dynamic range of an instrument (before measuring high-gain devices)
Use a 6 dB ENR noise source when:
• measuring a device with gain that is especially sensitive to changes in the source impedance
• the device under test (DUT) has a very low noise figure
• the device’s noise figure does not exceed 15 dB
A low ENR noise source will minimize error due to noise detector non-linearity. This error will be smaller if the measurement is made over a smaller, and therefore more linear, range of the instrument’s detector. A 6 dB noise source uses a smaller detector range than a 15 dB source.
A low ENR noise source will require the instrument to use the least internal attenuation to cover the dynamic range of the measurement, unless the gain of the DUT is very high. Using less attenuators will lower the noise figure of the measurement instrument, which will lower the uncertainty of the measurement.
If ENR values must be entered into the instrument manually, double-check them to ensure that the table in the instrument is correct.
Page 5
5
Follow these guidelines:
(a) Ensure that mating connectors are clean and not worn or damaged (see Reference 8 for further
information). If the measurement becomes unstable when the cables and connectors are shaken lightly by hand, try other cables or connectors.
(b) Use threaded connectors in the signal path whenever possible. (For example BNC connectors are
very susceptible to stray signals.)
(c) Use double shielded cables (common flexible braided coaxial cable is too porous to RF).
(d) Use shielded GPIB cables.
(e) Move the measurement setup to a screened room.
If a transmitter that has any frequency content within the measurement bandwidth is nearby and any covers are off of the DUT, move the measurement setup to a screened room. Test for such signals with a spectrum analyzer with a simple wire antenna on the input. Attenuate these stray signals by 70 to 80 dB.
(f) Use shielding. This is especially important for making measurements on an open PC breadboard. (See Reference 5 for further information.)
(g) Use an analyzer with minimal electromagnetic emissions. Devices may be susceptible to stray emissions
from some measurement instruments. Some modern noise figure analyzers have electromagnetic emission characteristics low enough to have negligible impact on the measurement.
HINT 2:
Minimize extraneous signals
A noise figure analyzer measures the noise power from the noise source as affected by the DUT. It uses the power ratio at two detected noise levels to measure the noise figure of whatever is between the noise source and the instrument’s detector. Any interference, airborne or otherwise, is measured as noise power from the DUT and can cause an error of any magnitude.
Figure 2-1 demonstrates the types of stray signals that can get coupled into the signal path and affect the measurement. Fluorescent lights, adjacent instruments, computers, local TV and Radio stations, pocket pagers, mobile phones and base stations are notorious for their adverse effects on noise measurements. Random stray signals can cause several tenths of a dB difference between individual readings, and result in unstable measurements (i.e. jitter that will not average to a stable mean).
Figure 2-1
Interference
Noise
Source
Lights ( esp. fluorescent)
FM TV Computers
Double-shielded cables for IF (ordinary braid is
too porous)
Shielded GPIB Cables Enclose all circuits
NF
Analyzer
RF comms basestation
Test connectors by shaking leads
DUT
Page 6
6
Alternately, insert a well matched attenuator (pad) between the noise source and the DUT to
attenuate multiple reflections. As an example, with a 10 dB attenuator, the re-reflections are attenuated by 20 dB. The advantage of an attenuator vs. an isolator is broadband response. The disadvantage is that the noise source’s ENR values will be reduced by the attenuator’s insertion loss (10 dB in this example).
If the DUT has a high output reflection coefficient (S22) and low or no insertion gain (S21), then, in some cases, place a low noise pre-amp between
the DUT and the measurement instrument to reduce the total measurement uncertainty. The
pre-amp should have an input reflection coefficient (S11) as low as possible. The effective bandwidth of a pre-amp with very low S11 is usually narrow. More than one may be required for the entire frequency range of interest. To identify an appropriate pre-amp, see Agilent Technologies Application Note 57-2 (Reference 5).
HINT 3:
Minimize mismatch uncertainties
Mismatch at connection planes will create multiple reflections of the noise signal in the measurement and calibration paths (as shown in Figure 3-1). Mismatch uncertainties at these planes will combine vectorially and will contribute to the total measurement uncertainty.
One method to reduce the mismatch uncertainty is to
place an isolator in the RF path between the noise source and the DUT. This isolator can prevent multiple
re-reflections from reaching the DUT and can suppress the build-up of error vectors. Isolators, however, operate over restricted frequency ranges. Several may be needed for the frequency range of interest. Isolators also add to path losses. As a result compensation is required. Full­featured noise figure analyzers have a loss compensation feature to account for the insertion losses of any isolators.
Figure 3-1
Measuring
System
Noise
Source
DUT
Calibration
Measurement
Mismatch
Uncertainty
ρ1ρ
2
ρ
4
ρ
3
ρ
= reflection coefficient at a reference plane
Page 7
7
If time constraints limit the number of averages during DUT measurement (e.g. in manufacturing), use more averages during calibration. This will make the correction for all subsequent DUT measurements more accurate.
Modern noise figure analyzers have the ability to do point averaging or trace averaging. Point averaging makes all measurements for the first frequency point, calculates and displays their average, then moves to the next frequency point. The process is repeated until each frequency has been measured and averaged the number of times specified in the measurement setup. Trace averaging measures only once at each frequency point through one entire sweep of the frequency range. It then begins the second sweep, averaging each individual measurement at each frequency with the previous average for that frequency, as it sweeps. It repeats this process until the number of sweeps or “traces” averaged together equals the number of averages specified in the measurement setup.
Both types of averaging give the same answer. Point averaging is faster overall since the analyzer’s tuner has to retune fewer times for all measurements to complete. Trace averaging displays a rough measurement over the entire frequency range faster. This enables a user to see any obvious problems with the measurement (e.g. an extraneous signal) sooner.
Use trace averaging first. Watch a few sweeps across the display and look for indications of RF interference such as a spike in the response at a single frequency, or even a small step in the response.
HINT 4:
Use averaging to minimize display jitter
Noise measurement inherently displays variability or jitter because of the random nature of the noise being measured. Averaging many readings can minimize displayed jitter and bring the measurement closer to the true mean of the noise’s gaussian distribution.
Modern noise figure analyzers enable the user to select the number of readings that will be averaged for each measurement. This will reduce jitter in the measurement by the square root of N, where N is the number of measurements in the average. The table below shows some examples of the effect of averaging on jitter. For example, jitter may be reduced by almost 70% by averaging approximately 10 readings.
Increasing the averaging will add time to make the measurement. This would affect the device test time and total test throughput in manufacturing. There is a trade-off between the speed of the measurement and the level of jitter.
Decreasing the bandwidth of the measurement increases proportionately the number of readings necessary to obtain the same level of jitter reduction. For example, half the bandwidth requires twice as many readings of noise figure averaged together to obtain the same jitter reduction; one fourth requires four time as many readings, and so on. These additional readings do not necessarily extend the time for the measurement proportionately.
N SQRT % jitter
(N) reduction
110
4250 16 4 75 64 8 87.5
256 16 93.75
Page 8
8
A Y-Factor noise figure analyzer assumes a linear change in the detected noise power as the noise source is switched between T
hot
and T
cold
. Any variations from linearity in either the DUT or in the detector directly produce an error in the Y value and hence in the noise figure that is displayed. The instrumentation uncertainty specification accounts for the linearity of the analyzer’s detector. The linearity of the DUT, however, should be carefully considered when making the measurement.
The measured noise figure is determined by the power in the instrument’s resolution bandwidth. The instrument’s attenuation settings, however, may be determined by the power in the instrument’s overall frequency range that reaches the range detector. The analyzer is therefore susceptible to being overdriven by noise outside the bandwidth of any one individual measurement, and therefore vulnerable to non-linearity errors. In such cases, attenuate any broadband power outside the analyzer’s resolution bandwidth. Use a filter wider than the measurement frequency range and before the DUT.
Consider measuring a familiar “reference” or “gold standard” device at the beginning of each day to assure that the same result is obtained as prior days for the same device, to add assurance that the measurement instrument is warmed up sufficiently.
HINT 5:
Avoid non-linearities
Avoid all predictable sources of nonlinearities:
• circuits with phase lock loops (and any circuit that relies on signal presence to set its operating condition)
• circuits that oscillate (even if at a far-removed frequency)
• amplifiers or mixers that are operating near saturation
• AGC circuits or limiters (AGC circuits have been
known to contribute additional noise power at the power levels near its operational point, even when disabled.)
• High Gain DUT’s without in-line attenuation
(Attenuate the output of the DUT if necessary. See Hint 7 for details)
• power supply drifts
• DUTs or measurement systems that have not warmed up
• logarithmic amplifiers (The standard Y-Factor
measurement is invalid for amplifiers in a logarithmic mode.)
Page 9
9
Figure 6-1
DSB measurements are easier to perform since they don’t require the additional burden of image rejection filter design and matching. When a mixer with a DSB specified noise figure is going to be used in an SSB application, careful correction is needed. (See reference 5 for more information)
b) For double sideband measurements, select
a LO frequency as close as possible to the RF band of interest.
The choice of LO frequency, and the resulting IF, can make a dramatic difference in the results of DSB measurements. Instruments typically display the average of the LSB and USB noise figures. They measure the power in the IF band (which is the sum of both sidebands after conversion), diminish that power by half (3 dB), and display the result. The closer the USB and LSB are together, the more likely they are to be equal (as shown in Figure 6-1) and the more likely the default 3 dB correction will be accurate. Since noise power versus frequency for a mixer is rarely flat, if too wide an IF is used the error and its correction will be unknown. To minimize this error, choose the LO frequency as close as possible to the RF band of interest, within the limitation that the resulting IF cannot be below the lower frequency limit of the instrument being used (often 10 MHz).
HINT 6:
Account for mixer characteristics
If the device under test is a mixer:
• Measure the same sideband(s) that will be used in the application of the mixer.
• For double sideband measurements, select a LO frequency close to the RF band of interest.
• For single sideband measurements, select a LO far from the RF band of interest, if possible.
• Choose the LO to suit the mixer.
• Filter the RF signal (i.e. the noise source) if necessary to remove unwanted signals that would mix with the LO’s harmonics or spurious signals.
Always document a frequency plan to identify which of the above precautions are necessary.
a) Select a double sideband or a single
sideband measurement.
A mixer will translate input signals and noise from the upper sideband (USB) and lower sideband (LSB) as Figure 6-1 shows. (Note that the LSB and USB are separated by twice the IF.) A double sideband (DSB) measurement, shown in Figure 6-2, measures the noise powers for both the USB and LSB. Some receiver systems, like those in radio astronomy, intentionally use both sidebands. DSB noise figure measurement is appropriate in these cases. In many applications the desired signal will be seen in only one sideband. A single sideband (SSB) measurement is appropriate in these cases. In SSB measurement setups, the noise power in the unwanted sideband is suppressed by appropriate “image rejection” filtering at the input of the mixer.
Noise Power
Freq
LSB
USB
F
LO
F
IF
average
Page 10
10
Balanced and double-balanced mixers have more than one diode to perform the frequency conversion. This improves the LO to IF isolation (20 dB and better), but usually requires more LO power or biasing. This higher power may raise the LO’s noise output, produce spurs away from the set frequency, or produce LO harmonics inside the mixer. Any of these can mix additional, yet unwanted, RF signal into the IF, or leak through to the IF. This may raise the measured noise and possibly overdrive the analyzer’s first stage to a non-linear state. Experiment with the LO to see what improves performance and produces the lowest noise figure.
e) Filter the RF signal (i.e. the noise source)
if necessary.
It may be necessary to filter the RF noise source to remove frequencies that are outside the band of interest as shown in Figure 6-4. This can remove unwanted signals that would mix with the LO or its harmonics or spurious signals and raise the measured noise figure.
To determine if sideband-averaging error is a problem, set up the noise figure measurement for a mixer with swept LO and fixed IF (modern noise figure analyzers allow this). Monitor the noise figure reading. If the noise figure values change dramatically as the LO sweeps, then SSB measurement is recommended (See reference 5 for further information.)
Figure 6-2
c) For single sideband measurements, select
a LO far from the RF band of interest, if possible.
The farther the LO is from the RF, the higher will be the IF, and the less stringent will be the rolloff requirement for the filter that suppresses the unwanted sideband as shown in Figure 6-3.
d) Choose the LO to suit the mixer.
Choose the LO to avoid spurious output where the mixer is sensitive and to avoid high broadband noise floor. Filter the LO if necessary to diminish spurious signals and broadband noise, since the IF pass band will include noise power one IF away from the LO frequency and one IF away from any spurious signals or harmonics of the LO.
Noise Figure
Analyzer
Noise source
IF Amplifier
Mixer
f
LO
IF
f
RF
LO
Measurement System
Noise Source Drive
(DUT)
f
Page 11
11
FIG 6-3
FIG 6-4
f) Filter the IF signal (ie, the DUT output)
if necessary.
Keep the LO outside of the frequency range of the instrument if possible. LO power will almost certainly
leak through to the IF port of the mixer. Assume that LO to IF isolation will be insufficient. If this leakage is within the band of the measurement, it will add to the noise figure measured. If it is outside the band of the measurement but within (or close to) the overall frequency range of the instrument, it could cause the instrument to autorange itself to use more attenuation, which would increase the uncertainty of the measurement. Filter the output of the mixer to remove this LO leakage and any LO harmonics created in the mixer without appreciably attenuating the wanted mixer output signal.
g) Document a frequency plan to evaluate
which of the above precautions are necessary.
A frequency plan will help identify which of the precautions above are the most important before the measurement is made.
Noise Figure
Analyzer
Noise source
IF Amplifier
Mixer
f
f
LO
IF
f
MEAS
LO
Measurement System
Noise Source Drive
(DUT)
Noise Power
Freq
LSB
USB
F
LO
F
IF
filter response
Page 12
12
The cascade equation shows how F12 is very sensitive to uncertainty margins in the second stage term [(F2 - 1) / G1]. (To see how F12 would vary with marginal changes in F2 or G1, see reference 6.) If the DUT has
insertion loss (e.g. a mixer, attenuator, etc.), use a low noise pre-amplifier before the instrument to reduce the uncertainty margin. Choose a pre-amp with the
lowest noise figure and a gain of more than 100 (20 dB +). (See Agilent Technologies Application Note 57-2 for more information on selecting the pre-amp.) Make the noise figure of the second stage as low as practical and the uncertainty of F12 (and hence F1), as low as possible. Adding a pre-amp also gives the measurement some resilience against noise figure variations versus frequency of the second stage.
In the case of a high-gain DUT, there may not be a need for a pre-amp. In order to make that decision, place the DUT’s linear gain and the measurement system’s linear noise figure (i.e. noise factor) into the Cascade Equation. Notice the noise figure of the cascade will converge to the noise figure of the DUT.
Figure 7-1.
HINT 7:
Use proper measurement correction
Take the following steps to ensure the measurement system itself does not add error to the measurement.
• Remove the noise figure of the measurement system with regular user calibration.
• Avoid exceeding the maximum input power of the measurement instrument. Modern instruments
can handle around 65 dB of device gain for narrow band devices. For wider band devices with high gains it is likely that an attenuator will be required after the device to keep the overall power within the instrument’s range. Use the analyzer’s compensation feature to account for the losses of the attenuator. Use appropriate filters/isolators/circulator to suppress out-of-band responses that would otherwise contribute noise power at the high gain level and overpower the instrument’s input.
A Y-Factor noise figure analyzer measures the noise figure of the measurement system and the DUT combined (see Figure 7-1). Below is F12 in the cascade equation:
F12 = F1 + [(F2 - 1 ) / G1] (7-1)
F1 and F2 are linear noise figure values for the DUT and the measurement system, respectively, and G1 is the gain of the DUT. User calibration (termed “second stage correction”) determines F2; measurement determines F12 and G1. The analyzer calculates F1 from the cascade equation.
Perform user-calibration prior to the measurement to remove the second stage contribution. Calibrate out the second stage contribution at regular intervals depending on how sensitive the noise figure and gain are to temperature drifts.
Noise source
Measuring
system
DUT Gain, G
Calibration path
Measurement
path
F2 = noise figure of the measurement system F1 = noise figure of the device under test (DUT)
Page 13
13
Figure 8-1
HINT 8:
Choose the optimal measurement bandwidth
Select a measurement bandwidth no larger than the pass band of the DUT. Modern noise figure analyzers
provide a selection of various measurement bandwidths to enable measurements that are more relevant to current practical applications (e.g. individual wireless GSM channels). 4 MHz bandwidths were common in past generations of noise figure instruments; modern analyzers can measure down to at least 100 kHz bandwidth.
If the measurement bandwidth is similar to the DUT’s operational pass band, then any instabilities between the center frequency of the analyzer’s final IF and the DUT’s operational bandwidth will lead to an error in the displayed gain. The analyzer may recognize the noise power outside of the DUT’s pass band during calibration and inside during measurement. The extent of this effect depends on the shape of the DUT’s pass band overlapping with the shape of the analyzer’s final IF pass band. This issue is less significant if the analyzer’s pass band is narrower than the DUT’s pass band.
Past generations of noise figure instruments required additional time to obtain the same level of jitter reduction in a narrower bandwidth. For the same jitter reduction, half the bandwidth required twice the time, one fourth required four times as long, etc. In modern full-featured noise figure analyzers, this time penalty is mostly eliminated due to digital signal processing (DSP) techniques that measure a group of adjacent frequencies simultaneously.
Frequency
Measurement Bandwidth Setting
Noise power detected during calibration but not during DUT measurement understates DUT's noise figure
Noise Power
Lowest possible actual measurement bandwidth due to downward frequency drift
Analyzer's frequency instability due to drift during time required for all measurements to be made and averaged
Highest possible actual measurement bandwidth due to upward frequency drift
Page 14
14
Figure 9-1
HINT 9:
Account for path losses
Adapters must be used if the connectors between the noise source, DUT, and measurement system do not mate, as in Figure 9-1. It is most important to avoid adapters where the signal is smallest in the measurement setup. For devices with gain, avoid adapters before the DUT. For devices with loss, avoid adapters after the DUT.
Keep track of the insertion losses of adapters and any additional components such as cables, baluns, filters, pads, and isolators that are not part of the DUT. These insertion losses have to be taken into account
by subtracting them from the measurement result. Any adapters used between the noise source and the DUT will introduce some loss, effectively reducing the ENR of the noise source.
Modern noise figure analyzers can subtract these losses from the measurement automatically. The analyzer will also need to know the temperature of these
components and whether they are before or after the DUT. The analyzer will correct the displayed noise
figure for all loss elements and their temperatures that are entered.
Losses in the measurement system are corrected by user calibration. Adapters and cables that are connected to the instrument during both user calibration and measurement are considered part of the measurement system and are accounted for during user calibration.
If a pre-amp is needed prior to the measurement system then choose a pre-amp with the correct connector for the noise source, and include the pre-amp as part of the calibration setup that measures the second stage noise figure.
Connection Diagram
Noise Source
DUT
Measurement system
G 1
G 2
Cable C1
Cable C 2
Coax/WG adapter
Coax/WG adapter
A
B
Page 15
15
If the actual local ambient temperature is not entered into the analyzer, it will make its calculations with the wrong temperature assumption. An extra error term in the over-all uncertainty margin will be present. This can be a significant value if the true noise figure of the DUT is low. Figure 10-2 shows typical differences between the displayed and actual noise figures for different ambient temperatures.
Figure 10-2
HINT 10:
Account for the temperature of the measurement components
Y-Factor noise figure analyzers assume that the surface temperatures of all components in the measurement (noise source, DUT, connectors, cables, etc.) are the default value for Tcold, 290K (16.8°C, 62.2°F). If this is not the case, enter the correct temperature of each
component into the analyzer and monitor them regularly. Full featured noise figure analyzers allow the
entry of the temperatures of any components added before and after the device.
Figure 10-1 shows the measurement response curve where the noise powers N
1
and N2are due to input noise temperatures Tcold and Thot, respectively. The solid line on this graph represents the response when T
0
is considered as the reference ambient temperature of 290K. If the T
cold
of the noise source (for a solid state noise
source, its surface temperature) is not equal to T
0
, then the dotted line may become the response curve. The noise power added by the DUT would then be N
a
´ and not Na.
Figure 10-1
+1K
T
amb
= 296.5K (=74 F)
T
amb
= 305K (=89 F)
0.24
0.20
0.16
0.12
0.08
0.04
0 0.4 0.8 1.2 1.4 1.8
NF(dB)
NF(dB) = F mea - Factual
P
out
N
2
N
1
N
a
´
N
a
T
o
T
c
ThT
s
Page 16
16
Account for mixer characteristics. See Hint # 6.
Measure the same sideband(s) that will be used in
the application.
For double-sideband measurements, select a LO
frequency close to the RF band of interest.
For single-sideband measurements, select a LO far
from the RF band of interest, if possible.
Choose the LO to suit the mixer. Filter the LO if necessary to diminish spurious signals
and broadband noise.
Keep the LO outside of the measurement bandwidth
if possible.
Filter the IF if necessary to remove LO harmonics
created inside the mixer.
Filter the RF to prevent unwanted mixing. Test for DSB error by changing IF. Experiment with different LOs to get the most
accurate (i.e. lowest) noise figure.
Document a frequency plan to evaluate which of the
above precautions are necessary.
Enter proper measurement correction.
See Hint # 7.
Calibrate regularly. Keep overall gain below the instrument’s spec. Filter out-of-band power. If 2nd stage effect is large, add a low-noise pre-amp
(with proper connectors).
Choose the optimal bandwidth. See Hint # 8.
Select a measurement bandwidth no larger than the
pass band of the DUT.
Account for path losses. See Hint # 9.
Avoid adapters as much as possible. If used, enter their losses into the instrument.
Account for the temperature of measurement
components. See Hint # 10.
Enter physical temperatures of the noise source and
the components of the measurement setup into the instrument.
Appendix A:
Checklist
Use this checklist to assist with locating the hint relating to specific issues or considerations. Reading the 10 Hints Application Note sequentially is not necessary.
Select the appropriate noise source. See Hint # 1.
Use a low ENR source whenever possible. Avoid adapters between the noise source and DUT. Double check manually entered ENR values.
Reduce EMI influence. See Hint # 2.
Use clean, undamaged connectors. Use threaded connectors. Use double shielded cables. Use shielded GPIB cables. Use a screened room. Use shielding. Use an analyzer with minimal electromagnetic
emissions.
Minimize mismatch uncertainties. See Hint # 3.
Use an attenuator or isolator if needed. Use a pre-amp if needed.
Use averaging to minimize display jitter.
See Hint # 4.
Select enough averages to stabilize the measurement. Use “trace averaging” first to spot measurement setup
problems soonest.
Look for spikes or even small steps in the display that
indicate RF interference.
If time constrained, use more averaging during
calibration than during DUT measurement.
Avoid non-linearities. See Hint # 5.
Avoid the following:
Circuits with phase lock loops Circuits that oscillate Amplifiers or mixers operating near saturation AGC circuits or limiters High-gain DUTs without in-line attenuation Power supply drifts DUTs or instruments that have not warmed up Logarithmic amplifiers
Page 17
17
Figure B-2 shows the typical display for the entry point to this interactive model. Figures B-3, B-4 and B-5 show simulated results for the uncertainty calculator. There are a number of graphs that can be plotted. These figures show examples of RSS uncertainty with respect to instrument match, DUT input match and instrument noise figure, respectively. This interactive uncertainty calculator and the spreadsheet version shown in Figure B-1 can be accessed via the Internet by using the URL on the back page of this application note.
Figure B-2
Figure B-3
Appendix B:
Total uncertainty calculations
The error model in the spreadsheet shown in Figure B-1 is obtained from the derivative of the Cascade equation (F1 = F12 - [(F2-1) / G1]). It takes into account the individual mismatch uncertainty calculations at each reference or incident plane of the DUT, noise source and measurement system. This example represents the total noise figure measurement uncertainty, RSS analysis, for a microwave transistor with an S11 of 0.5, S22 of 0.8 and S21 of 5 (14dB). Noise figure uncertainty, here is calculated as ±0.48 dB. The uncertainty dramatically improves to ±0.26dB, in this instance, if the gain is improved to 20dB. The calculation for this error model is derived in Reference 6.
Figure B-1
Page 18
Figure B-4
Figure B-5
18
Appendix C:
References
1. Agilent 346A, 346B, and 346C Noise Sources (10 MHz to 26.5 GHz), Technical Specification Sheet,
literature number 5953-6452
2. Fundamentals of RF and Microwave Noise Figure Measurements, Application Note 57-1, literature number 5952-8255E
3. Noise Parameter Measurement Using the Agilent
8970B Noise Figure Meter and the ATN NP4 Noise Parameter Test Set, Product Note 8970B/S-3,
literature number 5952-6639
4. Noise Parameter Measurement Accuracy and Repeatability Considerations, ATN Application Note 001
5. Noise Figure Measurement Accuracy, Application Note 57-2, literature number 5952-3706
6. “Calculate the Uncertainty of NF Measurements” Microwaves and RF, October 1999. Also available at: http:\\www.mwrf.com (select Editorial Archives; then October 1999).
7. Agilent R/Q 347B Solid-State Noise Sources at Millimeter-Wave Frequencies, Technical Specifications, literature number 5954-8888
8. Principles of microwave connector care (for higher reliability and better measurements), Application Note 326, literature number 5954-1566
Page 19
19
Appendix E:
Glossary and definitions
1. Excess Noise Ratio (ENR):
ENR is the measure of how much more noise power is output from a noise source when “ON” (i.e. operating at virtual temperature T
hot
) than is output when “OFF” (i.e.
operating at ambient temperature T
cold
), normalized by its
output power at the standard temperature 290K.
ENR = 10 log (T
h
- Tc) / T
0
For example, an ENR of 15 dB means that the noise source output when ON is greater than when OFF by the antilog of 15/10 times the noise power ouput at 290K. This is equivalent to a resistor at 9171K, calculated as antilog (15/10) x 290.
2. Second Stage Contribution:
During the measurement process the noise source is connected to the input of the DUT and the DUT’s output is connected to the measurement system. In this cascade, the DUT is the first stage and the measurement system is the second stage. The measurement system will measure the noise figure of the cascade. Correction requires that the noise figure (F1) value of the DUT has to be de-embedded” from the cascade’s combined noise figure (F12) by removing the contribution of the second stage in the cascade or Friis equation:
F12 = F1 + [(F2 - 1) / G1]
The expression in the brackets is the second stage contribution. G1 is the gain of the DUT.
3. Y-Factor:
This is the linear ratio of the noise power seen by the measurement system when the noise source is turned ON, to that when the noise source is turned OFF. This is the basis of the calculation of noise figure. After the instrument recognizes the ENR, the noise figure is derived from the equation:
NF = ENR - 10 log (Y - 1)
Appendix D:
Abbreviations
AGC Automatic Gain Control DUT Device Under Test DSB Double Sideband ENR Excess Noise Ratio F Noise Factor (linear expression of NF) LSB Lower Sideband NF Noise Figure (noise factor [F] expressed in dB) RSS Root Sum of Squares SSB Single Sideband T
c
Cold Temperature
T
cold
Cold Temperature
T
h
Hot Temperature
T
hot
Hot Temperature USB Upper Sideband URL Universal Reference Locator WG Waveguide
Page 20
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain
dependable measurement accuracy for the life of those products.
For more assistance with your test & measurement needs go to
www.agilent.com/find/assist
Or contact the test and measurement experts at Agilent Technologies (During normal business hours)
United States:
(tel) 1 800 452 4844
Canada:
(tel) 1 877 894 4414 (fax) (905) 206 4120
Europe:
(tel) (31 20) 547 2000
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Latin America:
(tel) (305) 267 4245 (fax) (305) 267 4286
Australia:
(tel) 1 800 629 485 (fax) (61 3) 9272 0749
New Zealand:
(tel) 0 800 738 378 (fax) 64 4 495 8950
Asia Pacific:
(tel) (852) 3197 7777 (fax) (852) 2506 9284
Product specifications and descriptions in this document subject to change without notice. Copyright © 2000 Agilent Technologies Printed in U.S.A. 11/21/00 5980-0288EN
Loading...