HP app an57 3 figure measurement 2 5980 1916e schematic

10 Hints for Making Successful Noise Figure Measurements
Application Note 57-3
2
Table of Contents
Introduction 3
HINT 1: Select the appropriate noise source 4
HINT 3: Minimize mismatch uncertainties 6
HINT 4: Use averaging to minimize display jitter 7
HINT 5: Avoid non-linearities 8
HINT 6: Account for mixer characteristics 9
HINT 7: Use proper measurement correction 11
HINT 8: Choose the optimal measurement
bandwidth 12
HINT 9: Account for path losses 13
HINT 10: Account for the temperature of
the measurement components 14 Appendix A: Checklist 15 Appendix B: Total uncertainty calculations 16 Appendix C: References 18 Appendix D: Abbreviations 18 Appendix E: Glossary and definitions 19 Key web resources 20
3
Introduction
To achieve accurate and repeatable results at RF or microwave frequencies, measurement uncertainties and barriers to measurement repeatability must be minimized. The performance of a device can be obscured if errors are allowed to accumulate. For the most accurate measurement, it is important to understand the nature of the error contributors and identify which of these can be influenced or changed to improve the quality of the results. This application note provides useful tips that will assist in making accurate noise figure measurements.
The checklist in Appendix A is a helpful tool to verify that all hints have been considered for a particular measurement.
A detailed explanation of the uncertainties in noise figure measurements is provided in Agilent Technologies’ Application Note 57-2, Noise Figure Measurement Accuracy (see reference 5). For a general understanding of noise figure and its variety of measurement techniques, refer to Agilent Technologies Application Note 57-1, Fundamentals of Noise Figure Measurement (see reference 2).
4
Frequency range
Commercial noise sources cover frequencies up to 50 GHz with choices of co-axial or waveguide connectors. The frequency range of the noise source must include the input frequency range of the DUT, of course. If the DUT is a mixer or frequency translation device, the output frequency range of the DUT must also be addressed. If one source does not include both frequency ranges, a second source will be required. A second noise source may also be necessary when measuring a non-frequency translating device with low noise and high gain. Low ENR is best for the measurement, however, high ENR is necessary to calibrate the full dynamic range of the instrument. In either case, a full-featured noise figure analyzer can account for the different ENR tables required for calibration and measurement.
Match
If possible, use a noise source with the lowest change in output impedance between its ON and OFF states. The noise source’s output impedance
changes between its ON and OFF states, which varies the match between the noise source and the DUT. This variation changes the gain and noise figure of the DUT, especially for active devices like GaAs FET amplifiers. To minimize this effect, 6 dB ENR noise sources are commercially available that limit their changes in reflection coefficient between ON and OFF states to better than 0.01 at frequencies to 18 GHz.
Adapters
Use a noise source with the correct connector for the DUT rather than use an adapter, particularly for devices
with gain. The ENR values for a noise source apply only at its connector. An adapter adds losses to these ENR values. The uncertainty of these losses increases the overall uncertainty of the measurement. If an adapter must be used, account for the adapter losses.
HINT 1:
Select the appropriate noise source
ENR
The output of a noise source is defined in terms of its frequency range and excess noise ratio (ENR). Nominal ENR values of 15 dB and 6 dB are commonly available. ENR values are calibrated at specific spot frequencies. The uncertainties of these calibrations vary over the frequency range of the noise source and add to the uncertainty of the measurement. This uncertainty is typically limited to approximately 0.1 dB using the root-sum-of-squares method (RSS).
Use a 15 dB ENR noise source for:
• general-purpose applications to measure noise figure up to 30 dB.
• user-calibrating the fullest dynamic range of an instrument (before measuring high-gain devices)
Use a 6 dB ENR noise source when:
• measuring a device with gain that is especially sensitive to changes in the source impedance
• the device under test (DUT) has a very low noise figure
• the device’s noise figure does not exceed 15 dB
A low ENR noise source will minimize error due to noise detector non-linearity. This error will be smaller if the measurement is made over a smaller, and therefore more linear, range of the instrument’s detector. A 6 dB noise source uses a smaller detector range than a 15 dB source.
A low ENR noise source will require the instrument to use the least internal attenuation to cover the dynamic range of the measurement, unless the gain of the DUT is very high. Using less attenuators will lower the noise figure of the measurement instrument, which will lower the uncertainty of the measurement.
If ENR values must be entered into the instrument manually, double-check them to ensure that the table in the instrument is correct.
5
Follow these guidelines:
(a) Ensure that mating connectors are clean and not worn or damaged (see Reference 8 for further
information). If the measurement becomes unstable when the cables and connectors are shaken lightly by hand, try other cables or connectors.
(b) Use threaded connectors in the signal path whenever possible. (For example BNC connectors are
very susceptible to stray signals.)
(c) Use double shielded cables (common flexible braided coaxial cable is too porous to RF).
(d) Use shielded GPIB cables.
(e) Move the measurement setup to a screened room.
If a transmitter that has any frequency content within the measurement bandwidth is nearby and any covers are off of the DUT, move the measurement setup to a screened room. Test for such signals with a spectrum analyzer with a simple wire antenna on the input. Attenuate these stray signals by 70 to 80 dB.
(f) Use shielding. This is especially important for making measurements on an open PC breadboard. (See Reference 5 for further information.)
(g) Use an analyzer with minimal electromagnetic emissions. Devices may be susceptible to stray emissions
from some measurement instruments. Some modern noise figure analyzers have electromagnetic emission characteristics low enough to have negligible impact on the measurement.
HINT 2:
Minimize extraneous signals
A noise figure analyzer measures the noise power from the noise source as affected by the DUT. It uses the power ratio at two detected noise levels to measure the noise figure of whatever is between the noise source and the instrument’s detector. Any interference, airborne or otherwise, is measured as noise power from the DUT and can cause an error of any magnitude.
Figure 2-1 demonstrates the types of stray signals that can get coupled into the signal path and affect the measurement. Fluorescent lights, adjacent instruments, computers, local TV and Radio stations, pocket pagers, mobile phones and base stations are notorious for their adverse effects on noise measurements. Random stray signals can cause several tenths of a dB difference between individual readings, and result in unstable measurements (i.e. jitter that will not average to a stable mean).
Figure 2-1
Interference
Noise
Source
Lights ( esp. fluorescent)
FM TV Computers
Double-shielded cables for IF (ordinary braid is
too porous)
Shielded GPIB Cables Enclose all circuits
NF
Analyzer
RF comms basestation
Test connectors by shaking leads
DUT
6
Alternately, insert a well matched attenuator (pad) between the noise source and the DUT to
attenuate multiple reflections. As an example, with a 10 dB attenuator, the re-reflections are attenuated by 20 dB. The advantage of an attenuator vs. an isolator is broadband response. The disadvantage is that the noise source’s ENR values will be reduced by the attenuator’s insertion loss (10 dB in this example).
If the DUT has a high output reflection coefficient (S22) and low or no insertion gain (S21), then, in some cases, place a low noise pre-amp between
the DUT and the measurement instrument to reduce the total measurement uncertainty. The
pre-amp should have an input reflection coefficient (S11) as low as possible. The effective bandwidth of a pre-amp with very low S11 is usually narrow. More than one may be required for the entire frequency range of interest. To identify an appropriate pre-amp, see Agilent Technologies Application Note 57-2 (Reference 5).
HINT 3:
Minimize mismatch uncertainties
Mismatch at connection planes will create multiple reflections of the noise signal in the measurement and calibration paths (as shown in Figure 3-1). Mismatch uncertainties at these planes will combine vectorially and will contribute to the total measurement uncertainty.
One method to reduce the mismatch uncertainty is to
place an isolator in the RF path between the noise source and the DUT. This isolator can prevent multiple
re-reflections from reaching the DUT and can suppress the build-up of error vectors. Isolators, however, operate over restricted frequency ranges. Several may be needed for the frequency range of interest. Isolators also add to path losses. As a result compensation is required. Full­featured noise figure analyzers have a loss compensation feature to account for the insertion losses of any isolators.
Figure 3-1
Measuring
System
Noise
Source
DUT
Calibration
Measurement
Mismatch
Uncertainty
ρ1ρ
2
ρ
4
ρ
3
ρ
= reflection coefficient at a reference plane
Loading...
+ 14 hidden pages