4
Frequency range
Commercial noise sources cover frequencies up to
50 GHz with choices of co-axial or waveguide connectors.
The frequency range of the noise source must include the
input frequency range of the DUT, of course. If the DUT
is a mixer or frequency translation device, the output
frequency range of the DUT must also be addressed.
If one source does not include both frequency ranges,
a second source will be required. A second noise source
may also be necessary when measuring a non-frequency
translating device with low noise and high gain. Low
ENR is best for the measurement, however, high ENR
is necessary to calibrate the full dynamic range of the
instrument. In either case, a full-featured noise figure
analyzer can account for the different ENR tables
required for calibration and measurement.
Match
If possible, use a noise source with the lowest
change in output impedance between its ON and
OFF states. The noise source’s output impedance
changes between its ON and OFF states, which varies
the match between the noise source and the DUT. This
variation changes the gain and noise figure of the DUT,
especially for active devices like GaAs FET amplifiers.
To minimize this effect, 6 dB ENR noise sources are
commercially available that limit their changes in
reflection coefficient between ON and OFF states to
better than 0.01 at frequencies to 18 GHz.
Adapters
Use a noise source with the correct connector for the
DUT rather than use an adapter, particularly for devices
with gain. The ENR values for a noise source apply only at
its connector. An adapter adds losses to these ENR values.
The uncertainty of these losses increases the overall
uncertainty of the measurement. If an adapter must be
used, account for the adapter losses.
● HINT 1:
Select the appropriate noise source
ENR
The output of a noise source is defined in terms of its
frequency range and excess noise ratio (ENR). Nominal
ENR values of 15 dB and 6 dB are commonly available.
ENR values are calibrated at specific spot frequencies.
The uncertainties of these calibrations vary over the
frequency range of the noise source and add to the
uncertainty of the measurement. This uncertainty is
typically limited to approximately 0.1 dB using the
root-sum-of-squares method (RSS).
Use a 15 dB ENR noise source for:
• general-purpose applications to measure noise
figure up to 30 dB.
• user-calibrating the fullest dynamic range of an
instrument (before measuring high-gain devices)
Use a 6 dB ENR noise source when:
• measuring a device with gain that is especially
sensitive to changes in the source impedance
• the device under test (DUT) has a very low
noise figure
• the device’s noise figure does not exceed 15 dB
A low ENR noise source will minimize error due to
noise detector non-linearity. This error will be smaller
if the measurement is made over a smaller, and therefore
more linear, range of the instrument’s detector. A 6 dB
noise source uses a smaller detector range than a
15 dB source.
A low ENR noise source will require the instrument to
use the least internal attenuation to cover the dynamic
range of the measurement, unless the gain of the DUT
is very high. Using less attenuators will lower the noise
figure of the measurement instrument, which will lower
the uncertainty of the measurement.
If ENR values must be entered into the instrument
manually, double-check them to ensure that the table in
the instrument is correct.