HP 83751A, 83752A,83752B, 83751A,83751B User Manual

Page 1
User’s Guide
HP 83751A/B and HP 83752AlB Synthesized Sweepers
Page 2
HP part number: Printed in USA
Serial Numbers.
83750-90004
March 1996
This manual applies directly to instruments with serial prefix
3610A
and
below.
This manual also applies to Ermware revision 2.0 and above. For Ermware revisions below 2.0 contact your nearest Hewlett-Packard service center for a Ermware upgrade.
Notice.
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and Etness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
@Copyright Hewlett-Packard Company 1993, 1996 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403, USA
Page 3
The HP 8375lA/B and HP Synthesized Sweepers
83752A/B
The HP 83751AA and HP
“sweepers” throughout this manual) provide continuous analog or digital stepped sweep capability. The HP 2 to 20 to 20 high power output (approximately + 17 sweepers are SCPI and for drop-in replacement of an HP 8350 sweep oscillator. The sweepers are designed for optimum use with HP 8757 scalar analyzers. For specillcation and option information, refer to Chapter 17, “Specilkations and Options,” in this manual.
This User’s Guide is written to provide operating information to the user who is comfortable with the front panel layout and basic operation of the sweeper. For installation and basic sweeper operation, refer to the HP
HP
provided with your shipment.
GHz,
while the HP
GHz.
The “B” versions of both the HP 83751 and HP 83752 provide
83752A/B
S~thesized Sweepers Installation and Quick Start Guide,
83752A/B
83752A/B
HP-II3
programmable, with HP 8350 HP-IB mnemonics
synthesized sweepers (referred to as
83751A/B
provides a frequency range of 10 MHz
provides a frequency range of
dBm
maximum leveled power). The
83751A/B
and
. . .
111
Page 4
The Synthesized Sweeper at a Glance
The following Egure and accompanying text explain some features of the sweeper.
iv
pg425ab
The Synthesized Sweeper
Page 5
1. The
(m)
are used to save and subsequently recall sweeper operating parameters from one of nine nonvolatile register locations.
2. The line POWER switch turns the sweeper to either on or standby.
3. The Marker keys allow selection of up to ten markers to be set anywhere within the current frequency range. Marker A measurements are made using these keys as well.
4. The Modulation keys allow selection of pulse, AM, or FM modulation.
6. The SWEEP OUT BNC connector provides voltage proportional to the sweep ranging from 0 V et the start of a sweep at the end of the sweep, regardless of sweep width.
6. The Automatic
[AK INI
connector is used as the feedback path to the sweeper when its RF output power level is being leveled externally.
7. The Power keys allow selection of the RF output power level of the sweeper, as well as other power related functions, such as Automatic Level Control
[ALC
MODE 0) key is used to select
the method used to regulate the sweeper output power level. Either internal leveling or external leveling [with a diode, power meter, or source module] can be selected. Additional equipment is required when external leveling is used.
and
(m)
level
Control voltage input
IALCI.
The
keys
to’+10
V
6. The date entry keys are used to enter and/or modify various sweeper parameters. The @ key cancels all or part of an erroneous parameter entry before it has been terminated. The terminator keys rightmost column of keys) are used to choose the units for the entered parameter as well as to terminate the oarameter entry. The
a, 0,
used to increase or decrease a parameter in predetermined steps.
9. The toggle the RF output power on and off.
and
(-SIZE)
(RF
ON/OFF) key is used to
[the
keys are
10. The RF OUTPUT connector mates with
a female standard instruments. The connector metes with a type-N male connector on option instruments. The RF OUTPUT connector will be found on the rear panel of sweepers with option
APC-3.5mm
lE4.
precision connector on
1ED
11. The front panel knob is used to increase
or decrease active parameters under the pointers, and is used in manual frequency and power sweeps.
12. The displays show the current values of sweeper parameters as well as the status of many of the sweeper functions. The left-hand display shows the current frequency status, whether it be swept or CW. The right-hand display shows current marker, modulation end sweep status parameters when they are selected, as well as the current output power level. The annunciators that appear below the parameters are only visible when their associated function is active. For example, the
STEP annunciator will only be visible when
operating in stepped sweep mode.
13. The Frequency keys are used to set the
various frequency parameters for the sweeper. Swept frequency selections include Start/Stop,
CWSpan
and Marker 1 + Marker 2 functions. Continuous Wave also be selected for- outputting single frequencies.
14. The
(w)
the instrument into a known state. There are two preset modes: the factory mode, and a user-defined mode.
15. The
(m)
of some of the kevs. When vou
(m)
the sweeper performs the function printed in
blue above the key.
key changes the function
key and then press
ICWl
mode may
key is used to put
mess
the
enother
key,
Sweeper rear panel features are depicted and described in detail in Chapter 7,
“Front/Rear Panel” in this manual.
V
Page 6
In This Book
This book is divided into two sections: the task reference and the dictionary reference. These sections are subdivided into chapters. The task reference
section (identified by light blue tabs) provides step-by-step instructions for many of the tasks that you perform with your sweeper. The chapters in the task reference section are as follows:
Chapter 1
Chapter 2
Chapter 3
“Performing the Operator’s Check, )) contains a procedure for
assuring you that your instrument is operating correctly.
“Externally Leveling the Sweeper, )) provides the steps necessary to externally level your sweeper with detectors/couplers/splitters, power meters, and source modules.
“Generating a Stepped Sweep,” explains how to set up the instrument to generate a stepped sweep.
Chapter 4 “Generating a Millimeter Signal,” illustrates the setups for
using a millimeter head with your option
Chapter 5
Chapter 6
The dictionary reference section (identified by dark blue tabs) provides information about instrument features and functions. Information is divided into chapters as follows:
“Creating User Flatness Arrays,” explains how to create user flatness correction arrays, which calibrate the power level at a remote test port.
“Operating a Master/Slave Setup, ’ shows the steps necessary to configure two sweepers for two-tone measurement capabilities.
1EE
sweeper.
Chapter 7
Chapter 8
Chapter 9 “Marker Keys,
vi
“Front/Rear Panel, ’ contains entries that explain different aspects of the sweeper front and rear panel. (For example, you turn to this chapter for information on the sweepers various connectors).
“Instrument State Keys,” explains the functions of the keys in the Instrument State group.
explains the functions of the keys in the
Markers group.
Page 7
Chapter 10
“Modulation Keys,’ explains the functions of the keys in the
Mod group.
Chapter 11
Chapter 12
Chapter 13
Chapter 14
Chapter 15
Chapter 16
Chapter 17
Chapter 18
“Frequency Keys,” explains the functions of the keys in the
Frequency group.
“Sweep Keys,” explains the functions of the keys in the
Sweep group.
“Power Keys, ” explains the functions of the keys in the
Power group.
“Entry Keys,” explains the functions of the keys in the Entry
group.
“Special Functions,
explains the various selections in the
special functions menu, which is accessed by selecting
m
SPECIAL.
“Error Messages,” contains lists of the error messages that
might be generated during use of the instrument.
“Specifications and Options,” contains a list of the sweeper’s warranted performance specifications and typical operating parameters, as well as the various mechanical, electrical, warranty, and documentation options that are available.
“Safety and Regulatory information, ’ contains required regulatory and safety information that is not included elsewhere in the instrument documentation.
Vii
Page 8
Certification
Hewlett-Packard Company certihes that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Regulatory
Information.
The “Safety and Regulatory Information” chapter contains regulatory information.
. . .
Vlll
Page 9
Warranty
This Hewlett-Packard instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Hewlett-Packard Company will, at its option,
either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service
facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and Hewlett-Packard for use with an instrument will execute its programming instructions when properly installed on that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modihcation or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
fumware
designated by
ix
Page 10
Assistance
Product maintenance agreements and other customer assistance agreements
are available
Fbr
any assistance, contact your nearest
Ome.
Refer
for
Hewlett-Packard products.
Haolett-Packard
to the list of Sales and Service 0me.s
Sales and Service
on
the following page.
X
Page 11
Hewlett-Packard Sales and Service Offices
IN THE UNITED STATES California
Hewlett-Packard Co. 1421 South Manhattan Ave
PO.
Box 4230 Fullerton, CA 92631 17141 999-6700
Hewlett-Packard Co. 301 E. Evelyn Mountain View, CA 94039 14151
694-2000
Colorado
iewlett-Packard
24 Inverness Place, East Englewood, CO 80112
3031
649-5000
Co.
Georgia
iewlett-Packard 2000 South Park Place ?O.
Box 105005
Atlanta, GA 30339
Co.
4041 955-1500
:llinois
iewlett-Packard Co.
j201 Tollview IoIling
3121
Drive Meadows, IL 60008
255-9800
Yew Jersey
iewlett-Packard
120 W. Century Road
laramus,
201 I
265-5000
Co.
NJ 07653
rexas
hewlettPackard
130 E. Campbell Rd.
hchardson,
2141 231-6101
Co.
TX 75081
IN AUSTRALIA
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 895-2895
IN CANADA
Hewlett-Packard 17500 South Service Road
Trans.Canada
Kirkland, 15141 697-4232
[Canada]
Highway
Quebec H9J 2X8
Ltd.
IN FRANCE
Hewlett-Packard France
F-91947 Les Orsey
161 907-78-25
Ulis Cedex
IN GERMAN FEDERAL REPUBLIC
Hewlett-Packard Vertriebszentrale Frankfurt Berner Strasse 117 Postfach 560 140
D-6000
Frankfurt 56
1061 II
50-04-I
GmbH
IN GREAT BRITAIN
Hewlett-Packard Ltd. King Street Lane
Winnersh, Wokingham
Berkshire
0734 784774
RGll 5AR
IN OTHER EUROPEAN COUNTRIES
Hewlett-Packard
Allmend
2
CH-8967
Widen [Zurich]
100411 57 31 21 11
[Schweizl
AG
IN JAPAN
Yokogawa-Hewlett-Packard Ltd. 29-21 Takaido-Higashi, 3 Chome Suginami-ku Tokyo 168
1031 331-6111
IN PEOPLE’S REPUBLIC OF CHINA
China Hewlett-Packard, Ltd.
PO. Box 9610, Beijing
4th Floor,
Main Bldg. Shuang Yu Shu,
Beijing, PRC
256-6888
2nd
Watch Factory
Bei
San Huan Rd.
IN SINGAPORE
Hewlett-Packard Singapore
Pm.
Ltd. 1150 Depot Road Singapore 0410 273 7388
Telex HPSGSO Fax
1651
AS34209
2788990
IN TAIWAN
Hewlett-Packard Taiwan
8th Floor, Hewlett-Packard
Building 337 Fu Hsing North Road Taipei
1021 712-0404
IN ALL OTHER LOCATIONS
Hewlett-Packard Inter-Americas 3495 Deer Creek Rd. Palo Alto, California 94304
xi
Page 12
Safety Notes
The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
CAUTION
WARNING
The caution note denotes a hazard. It calls attention to a procedure
tihich,
if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution note until the indicated conditions are fully understood and met.
The warning note denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in
iqiury
or
loss of life. Do not proceed beyond a warning note until the indicked conditions are fully understood and met.
Instrument
Instruction
Manual
!
A
The
CE93
Markings.
The instruction manual symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the manual.
mark shows compliance with European Community 1993 standards.
The CSA mark is the Canadian Standards Association safety mark.
The
&Ml-A
mark stands for Industrial Scientific and Medical Group 1. Class A.
xii
Page 13
General Safety Considerations
WARNING
WARNING
WARNING
CAUTION
Before this
instrummt
is
switched on, make sure it has been properly grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact.
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
No operator serviceable parts inside the instrument. Refer servicing to
qualified personnel. To prevent electrical shock, do not remove covers. Any adjustments or service procedures that require operation of the instrument with protective covers removed should be performed only by trained service personnel.
For continued protection against fire hazard, replace line fuse only with the same type and rating (F
6.3A/250V).
The use of other fuses or
material is prohibited.
If this instrument is used in a manner not specihed by Hewlett-Packard Co.,
the protection provided by the instrument may be unpaired.
CAUTION
CAUTION
Always use the three-pronged ac power cord supplied with this instrument. Failure to ensure adequate earth grounding by using this cord may cause instrument damage.
This instrument has automatic selection input. Be sure the supply voltage is
within the speciEed range.
Xlll
. . .
Page 14
How to Use This Guide
This guide uses the following conventions.
(FRONT-PANEL
KEY_)
SHIFT FUNCTION
ANNUNCIATOR
This represents a key physically located on the instrument.
This represents a shift function (blue text above front panel keys).
Text in this font represents FREQUENCY, MARKER/SWEEP/STATUS, and POWER displays.
Text in this font represents the annunciators that are displayed in the lower portion of the sweeper display.
Page 15
Contents
1.
Performing
To run the To run peak power-tracking To
check
To
check
If
you
If the self-test If the maximum leveled power
2. Externally Leveling the Sweeper
Leveling
External leveling with the option 1El See
also
Leveling
See
also
Leveling with millimeter-wave source modules
(option
See
also
the Operator’s Check
full
self-test
the maximum leveled power the output
have
a
problem
fails
with detectors/couplers/splitters
.....................
with
a
power meter
.....................
1EE
only)
.....................
................
power
.................
.................
................
..............
..............
check fails
..............
.........
step
.......
.......
attenuator
l-3
1-5 1-6
l-8 l-10 l-10 l-11
2-3
.
2-6 2-6 2-7 2-8
2-9
2-11
3.
Generating a Generating a
4.
Using Using
Creating
5.
Creating
Creating a user flatness array for use in a scalar analysis
millimeter heads with “B” model
(high power) sweepers ................
millimeter heads with “A” model
(standard power) sweepers
See also
To
set up the sweeper ...............
To
set up the power meter
To
start the user flatness cal
measurement
To
set up
Stepped Sweep Millimeter Signal
..............
.....................
User Flatness Arrays
a
user flatness array
..................
the sweeper
.............
.............
............
...............
4-3 4-5
4-6
5-3 5-4 5-5 5-5
5-6 5-7
Contents-l
Page 16
To
To set
To start the user flatness cal To reactivate the
6.
Operating a
To
set up
To
set up
See
also
set up
up
the master sweeper the
......................
the analyzer
the power meter
HP
8757 system interface
Master/Slave
slave
sweeper
...............
...............
Setup
.............
..............
..........
......
5-7
5-8
5-8
5-9
6-4 6-6 6-7
Front/Rear
7.
Connectors
BNC
Panel
.....................
Connectors
Multi-pin Connectors
AUXILIARY HP-IB
.....................
SOURCE MODULE INTERFACE (Option
RF Output Connector
Display
......................
Frequency Display Marker/Sweep/Status Annunciators
HP-IB Lang/Address Switch
See
Knob
Also
.....................
.......................
Equivalent SCPI Commands
Line
POWER Switch
Power
Cables
Instrument
....................
State
.....................
.....................
.....................
.....................
.....................
.....................
.................
...............
INTERFACE
...............
.................
and Power
...................
................
Keys
............
Display
.............
............
.......
1EE
Only)
7-3 7-5 7-8
7-8 7-10 7-13 7-15 7-16 7-17 7-18
7-19
7-23 7-25 7-26 7-26 7-27
7-29
8-3
8-4
8-6
8-8
B-10
8-11
Contents-2
Page 17
9.
Marker
10.
Modulation Keys
11.
12.
Sweep Keys
Keys
(iziG)
........................
m (MKRa_l
(PULSE MODE
(AMMODEU
(FM MODE
ITIME) . . .
(TRIG MODE [SINGLE/TRIG)
........................
.......................
8)
...................
....................
$1.
....................
....
.
....................
....................
Q)
....................
...............
...................
...................
...................
...................
...................
9-4 9-7 9-8
10-4
10-6
10-7
11-4 11-6 11-8
11-9 11-10 11-12
12-4
12-7
12-10
13.
Power Keys
14.
Entry
(POWER LEVEL]
@OWER/SWEEP) [ALC
MODE 3) . . . . . . . . . . . . . . . . . . . .
(FLTNEss ON/OFF_) . . . . . . . . . . . . . . . . . . .
....................
...................
Keys
Q)a @EEiz)
......................
Number Pad Keys Unit
(j-ON/OFF)
......................
*+. ENTRY OFF
PEgK
.......................
Keys
.....................
......................
.....................
.......................
..................
Contents-3
13-3 13-5 13-7
13-10
14-3 14-4 14-5 14-6 14-7
14-8 14-9
14-10
Page 18
15.
Special
l- CWCFAUTO
2 - CW PEAKNG . . . . . . . .
3 - SWPTIME AUTO . . . . .
4 - STEP SWPTIME . . . . ,
5 - SWPTIME
6 - ATT COUPLING . . . . .
7 - ATT SETTING . . . . . .
8 ­9 -
10 - V/GHz
11- V/GHz OFFSET . . . . .
12-PWRMETERTY
13-PWRMETERAD
14 -
15 - LANGUAGE . . . . . . .
16 - FW REVISION . . . . . .
17 -
18 - DP DEFIN : : : ‘. ‘. ‘. ‘. ’ .
19 - CONTROL MODE . . . .
21-
Functions
ROSC
ROSC
FM
AUTO
SOURCE : : : : : :
SENSITIVITY . . . .
*SECURITY
FULL
. . . . . .
LLIM . . . .
SCALE . . . . .
. . . . . .
SELFTST . . . . .
...........
.
...........
...........
...........
.
...........
...........
...........
...........
...........
.
...........
...........
.
...........
.
...........
...........
...........
...........
...........
............
...........
...........
15-6
15-7
15-8
15-9 15-10
15-11 15-12 15-13 15-14 15-15 15-16 15-17 15-18 15-19 15-20 15-21
15-22
15-26 15-27 15-28
16. Error
The Error/Event Queue Error numbers
Error SCPI Error
Contents-4
Messages
................
....................
Messages List
Messages
Command Error Execution Error Device-specific Error Query Error
Instrument Block Bus
Control Errors Parsing and Compatibility Errors Diagnostics and Internal Hardware Errors
Hardware Configuration Errors
Calibration Routine Errors Loops Miscellaneous Hardware Dependent Errors
Specific
Transfer Errors
Unlocked Errors
.................
................
...................
..................
................
....................
Error
................
Self-test
Messages
...............
Errors
..............
.............
...............
.........
..........
...........
...........
......
16-3
16-4
16-5 16-7
16-7 16-13 16-19 16-21
16-23 16-23 16-24 16-25 16-28 16-32 16-32
16-33 16-35 16-36
Page 19
17.
Specifications and
Specifications Frequency
Range Timebase Stability
CW
Mode
Stepped Ramp (Analog) Bandswitch Points
RF Output
Maximum Minimum Settable Resolution Accuracy and Flatness
Power Sweep
Power Slope
External User Source Match
Spectral Purity
Harmonics (at max leveled power) Subharmonics Non-harmonic Spurious Residual
Modulation
Pulse AM FM
General
Compatibility : : : : Programming Master/Slave (two-tone) Measurements Environmental Warmup Time Power Dimensions Weight
Options
Electrical Options
Option 1El - Add Output Option lE4 - Rear Panel RF Output
Options
....................
.....................
......................
.................
....................
Sweep Mode
Sweep
.................
.....................
Leveled
Power (25 f5
Power
....................
...................
...................
Leveling
Flatness
(Level)
.................
Correction
...................
...................
...............
FM (0.05 to 15 kHz
.....................
......................
.......................
.......................
................
...................
..................
..................
Requirements
................
....................
......................
......................
.................
...............
Mode
..............
...............
..............
...............
Step
............
“C)
.........
..........
..........
bandwidth)
Attenuator
.......
.......
........
17-3 17-4 17-4 17-4 17-4 17-5 17-5 17-5 17-7 17-7 17-8 17-8 17-8 17-9 17-9
17-9
17-9 17-10 17-11 17-11
....
17-11 17-11 17-12 17-13 17-13 17-13 17-14 17-15 17-15
17-15 17-15
17-16
17-16 17-17 17-17 17-18 17-19 17-19
....
.
17-19 17-19
Contents-5
Page 20
Option lE5 - Add
Option
Option
Cable
.
1ED
1EE - Source
....................
Mechanical Options
Option 1CM -
Option
Option
1CP - Rack AX2 -
Warranty Options
Option
W30 - Two
Service Option Option
W50 ­ W52 -
Service Option
W54 -
Documentation Options
High
Type-N RF
Stability Timebase
Output
Connector
Module Connector and Extension
................
Rack
Mount Kit without Handles
Mount Kit with Handles
Portable Handle and Front-panel
.................
Additional Rears Return-to-HP
...................
Five Five
Year Year
Return to
HP
Return-to-HP Calibration
...................
Five
Year MIL-STD
Calibration
..............
Option OB2 - Extra Operating Documentation Option
Certification Options
Option 1BN - Certification Option
OB3 - Service
................
1BP -
Certification with Data
Documentation
............
......
Repair
.......
.......
.....
....
Cover
Service
Service
...
. .
17-19
17-19
17-20 17-20 17-20 17-20 17-20 17-21
17-21 17-21
17-21
17-21
17-22 17-22
17-22
17-22 17-22
17-22
18.
Safety and
Regulations and Licensing
Safety
Index
Contents-6
Notice
for
Information
Statement
Regulatory
Germany:
of
Compliance
Information
Noise
..................
Environmental Conditions Ventilation Requirements
Cleaning Instructions
..............
Declaration
..............
.............
..............
...............
........
18-3
18-3 18-8 18-8 18-8 18-9 18-9
Page 21
Figures
l-l. The UNLEV Annunciator Location ............. l-7
l-2. Connections for Checking Output Power ..........
2-l.
ALC Circuit Externally Leveled ..............
2-2. Typical Diode Detector Response at 25
2-3. Leveling with a Power Meter ...............
2-4. Millimeter-wave Source Module Leveling ..........
2-5. Millimeter-wave Source Module Leveling Using a Microwave
Ampliher ......................
4-1. Millimeter-wave Source Module Leveling ..........
4-2. Millimeter-wave Source Module Leveling Using a Microwave
Amplifier ......................
5-1. Creating a User Flatness Array ......... .....
5-2. Scalar Measurement System Setup ............
6-l.
Master/Slave Setup ...................
7-l.
Sweeper Connectors - Front Panel .............
7-2. Sweeper Connectors - Rear Panel .............
7-3. Auxiliary Interface Connector ...............
7-4. HP-B Connector and Cable ................
7-5. Interface Signals of the Source Module Connector ......
7-6. Sweeper Display ............... .....
7-7. Sweeper Display and LED Annunciators ..........
7-8. HP-B Address Switch Settings ..............
7-9. Instrument Language Switch Settings ...........
7-10. Clear Register Contents Settings .............
7-l 1. The Sweeper Line POWER Switch .............
7-12. Power Cable and Line (Mains) Plug Part Numbers ......
8-l.
Instrument State Group .................
9-l.
The Markers Group ...................
10-l. The Modulation Group ...................
11-l. The Frequency Group ..................
12-1. The Sweep Group ....................
13-1. The Power Group ....................
14-1. The Entry Group ....................
15-1. The Instrument Group ..................
17-l. Typical Swept Frequency Accuracy (100 ms sweep, ramp
mode) .......................
17-2. Typical Maximum Available Power ............
“C
.........
l-8
2-3 2-5 2-7 2-9
2-10
4-4
4-6
5-3 5-6
6-3
7-3 7-4 7-8
7-10
7-13
7-16
7-19
7-23
7-24 7-24
7-27
7-30
8-2
9-2 10-2 11-2 12-2 13-2 14-2 15-2
17-6
17-7
Contents-7
Page 22
Contents
17-3. Typical Phase Noise (10 17-4. Dimensions
.......................
GHz
Carrier)
...........
17-12 17-17
Contents-8
Page 23
l%bles
7-l.
Pin Description of the Auxiliary Interface
8-l.
Factory Preset Conditions 14-1. Step Sizes 15-1. Special Functions for the HP 83750 Series Sweepers 15-2. FM Sensitivity When Using Source Modules
.......................
................
.........
........
....
7-9
8-4
14-3
15-4
15-19
Contents-9
Page 24
1
Performing the Operator’s
Check
Page 25
Performing the Operator’s Check
The operator’s check consists of a series of tasks that, when completed, will either assure you that your instrument is operating correctly, or will help to point to problem areas if it is not. The operator’s check performance to specifications.
The operator’s check should be performed on a weekly basis, or whenever the integrity of the sweeper is in question.
To perform the operator’s check, the following tasks should be performed, in order:
1. Run the full self-test.
2. Run peak power-tracking.
3. Check the maximum leveled power.
4. Check the output power.
does
not ensure
l-2
Page 26
To run the full self-test
Attention!
All external cables [including HP-IB and BNC cables) must be disconnected from the sweeper prior to running the full self-test. Failure to do so may cause self-test failures or lock-ups.
1. Disconnect all external cables prior to running the self-test.
2. Activate the FULL SELFTST special function by pressing the following keys on the sweeper:
The message that appears in the MARKER/SWEEP/STATUS display is one of the following:
Full self-test has not been performed since the last time the line power was turned on.
Full self-test has been performed and passed all tests.
Full self-test has been performed, and test XXXXXX was either the only test that failed, or was the most significant failure.
3. Press the
While the self-test
(jj)
key again to initiate the self-test routine.
routine is running, a
M++I,,I.~ i t. ++s+
appear in the MARKER/SWEEP/STATUS display.
message
wfll
l-3
Page 27
Performing the Operator’s Check
To run the full self-test
4. When the self-test routine is
mushed,
a message will appear in the
MARKER/SWEEP/STATUS display. If the message does not say
“Ful 1
Ted PHASED,”
refer to the section entitled “If You Have a
Problem” at the end of this chapter.
The full self-test is actually a series of tests performed to check different instrument functions. If the sweeper fails just one test, the display shows the title of the failed test. If the instrument fails more than one test, the test title shown is the most signilicant failure.
1-4
Page 28
lb run peak power-tracking
1. Before running peak power-tracking, make sure the RF OUTPUT is either connected in a 50 0 system, or has a load on it, such as a power sensor or attenuator.
CAUTION
For optimum power at all frequencies, do not run peak power-tracking with a millimeter head connected to the sweeper.
2.
Press (PRESET)
Cm)
P&W .
The MARKER/SWEEP/STATUS display shows the progression of the sweep as the instrument adjusts the power-tracking.
When you run peak power-tracking (or autotracking), the instrument
optimizes its output power over the sweeper’s full frequency range by tracking the output
NOTE
Peak power tracking takes approximately 1 to 3 minutes to complete, and can be aborted by pressing
(EEFi],
if necessary.
titer
with the RF source output.
1-5
Page 29
lb check the maximum leveled power
1. Before checking maximum leveled power, make sure the RF OUTPUT is either connected in a 50 61 system, or has a load on it, such as a power sensor or attenuator.
2. Press [PRESET).
3. Press [POWER LEVEL) and enter the specified maximum leveled power for your instrument. (For an HP < 2
GHz.)
83752B,
set the power to the
specikation
for
HP
83751N33752A
HP 837518” HP
837528”
< 2
>
2
GHz
*
Option
1El
GHz
[step attenuatorl, reduce by 1
+lO dBm +17 dBm
+16 dBm +17 dBm
dB.
4. Make sure the UNLEV annunciator is not lit at any time. The UNLEV annunciator is located in the POWER display area. (See Figure l-l.)
5. Press
m
@
cm)
to change the sweep time to 2 seconds.
6. Press [SINGLE TRIG] and make sure the UNLEV annunciator does not come on at any point during the sweep. (If the annunciator comes on, it indicates the instrument is unleveled.) The UNLEV annunciator is located in the POWER display area. (See Figure l-l.)
1-6
Page 30
Performing the Operator’s Check
To check the maximum leveled power
\
000 0 000 0 00000
00
00
:a::
For HP
The HP
7. Press the following keys to set a sweep range of 2 power level of 17
0 0 000 0 00000 0 0 000 0 00000
-*
0
0
0’0
Figure l-l. The UNLEV Annunciator location
83752B
83752B
Instruments Only.
sweepers have a power specification that is split at 2
dBm
(or 16
000000
dBm
for instruments with option
0
0
0
GHz
to 20
UNLEV ANNUNCLATOR
tiEARS
pg426ob
GHz.
GHz
and a
1El):
(START_) 0 @iqGqLq)
[POWER
8. Press @NGLE on at any point during the sweep. (If the annunciator comes on, it
indicates the instrument is unleveled.) The UNLEV annunciator is located in the POWER display area. (See Figure l-l.)
mm) 117)
TRIG_)
(or
[16)) [GHz/dB(mL)
and make sure the UNLEV annunciator does not come
l-7
Page 31
To check the output power
1. Connect the equipment as shown.
POWER METER
SENSOR
SYNTHESIZED SWEEPER
RF OUTPUl
I
POWER SENSOR ADAPTER
Figure 1-2. Connections for Checking Output Power
2.
Press (PRESET_).
3. Press ccw)
4. Press
you set. Refer to the following table. Terminate the power level entry by pressing
(ZiJ CGHz/dB(mL).
POWER
LEVEL) and enter the specified power for the CW frequency
(GHz/dB(mZ).
pg427ob
l-8
Instrument
HP
83751Ml3752A”
HP 837518”
HP 837528”
< 2
1
2
1 *
Option
Maximum leveled
Power Specification
+I0 dBm +I7 dBm
GHz
GHz
IEI
lstep attenuetorl, reduce by 1
+I6 dBm +I7 dBm
dB.1
Page 32
Performing the Operator’s Check
To check the output power
5. Set the power meter calibration factor to the value listed on the power sensor that corresponds to the frequency you set.
6. Verify that the measured output power meets the instrument specification.
l If the measured value is less than the specihed power, turn the front
panel knob until the value measured is at least the speciEed power level.
l Make sure that the UNLEV annunciator doesn’t light. (If the annunciator
does light, this indicates an unleveled condition.)
7. Repeat steps 3 through 6 for the following setting(s):
2.5
GHz
1.5
GHz -
for HP
83752AA
instruments only
50 MHz - for HP
83752A/B
instruments only
l-9
Page 33
If you have a problem
If you have a problem while performing the operator’s check, check the following list of commonly encountered problems. If the problem you have encountered is not here, contact the nearest Hewlett-Packard office for assistance.
If the self-test fails
The self-test has failed if the message that appears when it is done, is anything except
Full Test.
PHSSED.
q Make sure all external cables are disconnected from the sweeper and run
the self-test again.
q If the test still fails, send the instrument to an HP service center for repair,
including a description of the failed test and any other error messages.
NOTE If you need to ship your sweeper, remove the front handles (if so equipped) and use the original
packaging (or comparable).
l-10
Page 34
Performing the Operator’s Check
If the maximum leveled power check fails
q Make sure that the RF OUTPUT connector was terminated before the peak
power-tracking sequence was run, and before the maximum leveled power check.
q Make sure the power you entered in steps 2 and 5 of the maximum leveled
power check procedure are correct for your particular instrument.
q
If
the measured value of power in step 6 of “To check the output power” is
mo?-e
than the specified power, turn the front panel knob counterclockwise until the value measured equals the specified power. Note the sweeper’s front panel reading and use this value to repeat the maximum leveled power check.
q If the check still fails, contact the nearest Hewlett-Packard office for
assistance.
NOTE
If you need to ship your sweeper, remove the front handles packaging (or comparablel.
(if
so equipped) and use the original
l-11
Page 35
I
2
Externally Leveling the Sweeper
Page 36
Externally Leveling the Sweeper
In externally leveled operations, the output power from the sweeper is detected by an external sensor. The output of this detector is returned to the leveling circuitry, and the output power is automatically adjusted to keep power constant at the point of detection.
2-2
Page 37
Leveling with detectors/couplers/splitters
1. Connect the equipment as shown in Figure 2-1.
SYNTHESIZED SWEEPER
DIRECTIONAL COUPLER
OR POWER
SPLITTER)
pg417ab
Figure 2-1. ALC Circuit Externally leveled
2. Press
[ALC
shows
MODE
HLC=
8)
repeatedly until the MARKER/SWEEP/STATUS display
Diode. Note that the EXT ALC annunciator is lit.
3. Enter the coupling factor by selecting the desired number.
m
LEVELED
OUTPUT
EXT CAL and then entering
2-3
Page 38
Externally Leveling the Sweeper
leveling with detectorslcouplerslsplitters
NOTE
The coupling factor (in
leveled power (P
locations of P lev and P det. After the coupling factor has been entered, the front panel knob may be used to fine tune the displayed power to equal the actual leveled output power (P levl.
Figure
2-l
illustrates a typical setup for external leveling. When externally
dB)
is defined by the equation
lev) -
detected power (P
det).
Refer to Figure ‘2-1 for
leveled, the power level feedback is taken from the external negative detector
input rather than the internal detector. This feedback voltage controls the ALC system to set the desired RF output.
Figure 2-2 shows the input power versus output voltage characteristics for
typical HP diode detectors. From the chart, the leveled power at the diode detector input resulting from any external level voltage setting may be determined. The ALC feedback voltage present at the output of the detector must be between -0.2 mV and -0.5 V. For a typical
corresponds to a detector input of approximately -35
HF’
diode detector, this
dE%m
to +5
dBm.
(See
Figure 2-2.)
2-4
Page 39
Externally Leveling the Sweeper
leveling with detectors/couplers/splitters
10 v
1.0
100 mv
10
mv
1 mv
+20
dBV
+lO
dBV
+6
dBV
v
-10
-20
-30
-40
-50
-60
0 dBV
dBV
dBV
dBV
dBV
dBV
dBV
-66
dBV
-70 di3V
.I mv
-40 -30
-
DETECTOR INPUT POWER,
0
+10 +20 f30
Figure 2-2. Typical Diode Detector Response at 25
dBM
-80
dBV
pg431ob
‘C
2-5
Page 40
Externally leveling the Sweeper
leveling with detectors/couplers/splitters
External leveling with the option 1El step attenuator
Some external leveling applications require low output power from the sweeper. The sweeper automatically uncouples the attenuator from the ALC system for the UNCPLD annunciator is lit.
For example, leveling the output of a 30 dB gain amplifier to a level of
-
10 leveled. At some frequencies this level is beyond the range of the ALC modulator alone. If so, the UNLEV warning message is displayed. Inserting 40
dB
of attenuation results in an ALC level (power level + attenuator value) of
0
dBm,
AM or other functions that vary the power level.
all
external leveling points. Note that in external leveling modes,
dBm
requires the output of the sweeper to be around -40
which is well within the range of the ALC. This gives a margin for
dBm
when
The ALC level should be greater than or equal to -10 sweepers (standard power), and -5 power). Adjust the attenuator so that the ALC level is within the specified power range of your sweeper. For an “A” model sweeper, this is achieved by using attenuation equal to the tens digit of output power. Example: for a desired sweeper output power of -43
1.
Press
2. Set the attenuator to 40
CSHIFT)
SPECIAL
(YJ
dB:
dBm
press
for “B” model sweepers (high
dBm;
do the following:
c40) [GHz/dB(mL)
dBm
for “A” model
See also
To obtain flatness corrected power refer to the chapter entitled “Creating User
Flatness Correction Arrays, n later in this manual.
2-6
Page 41
Leveling with a power meter
Leveling with a power meter is similar to leveling with a diode detector.
1. Set up the equipment as shown in Figure 2-3. Be sure to set the power meter to the correct manual range mode for the output power setting at which you are leveling.
POWER METER
=j-$g;pLER
POWER SENSOR
I
Figure 2-3. leveling with a Power Meter
2. Press
3. Select
shows
(ALC
MODE $J repeatedly until the MARKER/SWEEP/STATUS display
ALC= Pcllwr Met.w-.
m
EXT CAL @
Note that the EXT ALC annunciator is lit.
CGHz/dB(mL).
enter the coupling factor of the coupler.)
LEVELED
OUTPUT
pg418ab
(If a directional coupler is used,
2-7
Page 42
Externally leveling the Sweeper
leveling with a power meter
NOTE
The coupling factor is defined by the equation
leveled power (P
locations of P lev and P det. After the coupling factor has been entered, the front panel knob may be used to fine tune the displayed power to equal the actual leveled output power fP lev).
lev) -
detected power (P
det).
Refer to Figure 2-1 for
4. If the power meter and the sweeper power setting don’t agree, set the coupling factor step size to 10 dR and then use the a Q) keys to adjust the coupling factor up or down until the power meter and the sweeper power setting agree.
5. Select the sweep tune by pressing
I‘TIME)
and then entering the desired
sweep rate.
NOTE
Due to the settling time required by power meters, it is recommended to use a 40 second sweep rate.
Unlike detector leveling, power meter leveling provides calibrated power out of the leveled RF port.
See also
To obtain flatness corrected power refer to “Creating User Flatness Correction Arrays, n later in this manual.
2-8
Page 43
Leveling with millimeter-wave source modules (option 1EE only)
Millimeter-wave source module leveling is similar to power meter leveling, except that a slow sweep time is not required. Figure 2-4 and Figure 2-5 illustrate the setups for leveling with a mm-wave source module.
Figure 2-4 illustrates the setup that is used with the HP HP
83752B
(high power models). No external amplifier is required to obtain
maximum specified power. The interface extender cable
5062-7202) allows the source module to be positioned in front of the sweeper.
SOURCE MODULE I NTERFACE
RF
OUTPUT
ADAPTER
( IF
REQUI
RED)
83751B
(HP
and
part number
MM-WAVE SOURCE
MODULE
RF IN
b LEVELED
OUTPUT
pg420ob
Figure 2-4. Millimeter-wave Source Module leveling
2-9
Page 44
Externally leveling the Sweeper
Figure 2-5 illustrates the setup that is used with the HP HP
83752A
(standard power models). An external amplifier is required to
obtain maximum specified power.
SOURCE MODULE I NSERFACE
RF OUTPUT
I
INTERFACE EXTENDER CABLE
I
ADAPTER
(IF REQUIRED)
MICROWAVE AMPLIFIER
\
-cxJ@
0
0
RF IN
C
RF OUT
W-WAVE SOURCE
MODULE
83751A
and
LEVELED OUTPUT
Figure 2-5. Millimeter-wave Source Module leveling Using a Microwave Amplifier
1.
Turn the sweeper line power off.
2. Set up the equipment as shown in Figure 2-4 or Figure 2-5.
3. Turn the sweeper line power on and press
4. When the
c-j
key is pressed, the sweeper configures itself for source
C-1.
module operation, and all of the ALC data necessary to communicate properly with the sweeper is exchanged via the rear panel SOURCE MODULE INTERFACE. Note that the EXT ALC annunciator is lit, and that the POWER display shows the millimeter head output power.
2-10
Page 45
Externally leveling the Sweeper
See also
To obtain flatness corrected power refer to “Creating User Flatness Correction Arrays, )) later in this manual.
2-11
Page 46
3
Generating a Stepped
Sweep
Page 47
Generating a Stepped Sweep
To generate a stepped sweep, perform the following steps:
1. Press (PRESET).
2.
Select the desired
3.
Select the desired number of points by pressing
entering the desired number.
4.
Press
m
SWEEP MODE $ .
5. Press the @J key. The MARKER/SWEEP/STATUS display should read
Sweep=
St>eF’
6. The instrument is now running in stepped sweep mode.
Iknt
(START_)
and the STEP annunciator should be lit.
and IsTopl frequencies.
(=I
POINTS and
NOTE
The stepped sweep time is calculated by the following formula:
SteppedSweep
stepped sweep time can be changed in one of two ways:
l By changing the dwell time. (Press
l By changing the number of points. (Press
Time = Dwell Time *Number of Points. Therefore the
CTIME)
while in stepped sweep
(GiiFF~ PDINTS .)
mode.1
3-2
Page 48
4
Generating a Millimeter Signal
Page 49
Generating a Millimeter Signal
If your sweeper was ordered with Option lEE, you have the capability of using HP 83550 series millimeter head source modules with the sweeper. The interface extender cable (HP part number 5062-7202) that is provided with all option sweeper.
After a source module is connected to the sweeper via the SOURCE MODULE INTERFACE connector, the sweep will automatically configure itself for source
module operation when the line power is cycled, when the sweeper is preset,
or when a register is recalled.
1EE
sweepers allows the source module to be positioned in front of the
4-2
Page 50
Using millimeter heads with “B” model
(high power) sweepers
Figure HP
1.
2. Turn the sweeper on, and press
3. The sweeper automatically configures itself for source module operation
4-l
shows the equipment setup for using a millimeter head with your
83751B
Turn the sweeper line power off, and connect the equipment as shown in
Figure 4-
when the
or HP
1.
@EE]
83752B.
fjPRESET).
key is pressed.
4-3
Page 51
Generating a Millimeter Signal
Using millimeter heads with “B” model
SOURCE MODULE
I NTERFACE
I NTERFACE EXTENDER CABLE
RF
OUTPUT
ADAPTER
(IF REQUIRED)
(high power) sweepers
Figure
4-l. Millimeter.wave
MM-WAVE SOURCE
MODULE
LEVELED
OUTPUT
pg420ab
Source Module leveling
4-4
Page 52
Using millimeter heads with “A” model
(standard power) sweepers
Figure 4-2 shows the equipment setup for using a millimeter head with your HP
maximum specihed power.
1. Turn the sweeper line power off, connect the equipment as shown in Figure 4-1.
83751A
or HP
83752A.
An external amplifier is required to obtain
2. Turn the sweeper on, and press
3. The sweeper automatically configures itself for source module operation when the (PRESET_) key is pressed.
(EE’EQ.
4-5
Page 53
Generating a Millimeter Signal
SOURCE MODULE INTERFACE
RF OUTPUT
I
ADAPTER
(IF REQUIRED)
NTERFACE
:XTENDER :ABLE
MICROWAVE AMPLIFIER
RF IN
Wd-WAVE
SOURCE
MODULE
b
LEVELED
OUTPUT
Figure 4-2. Millimeter-wave Source Module leveling Using a Microwave Amplifier
See also
To obtain flatness corrected power refer to “Creating User Flatness Correction Arrays,” later in this section.
4-6
Page 54
5
Creating User Flatness Arrays
I-
Page 55
Creating User Flatness Arrays
This chapter explains how to create user flatness correction arrays, which calibrate the power level at a remote test port. Two examples are provided:
l The first example shows the basic setup and steps to create a user flatness
array.
l The second example shows how to set up a scalar analysis measurement
using a user flatness correction array.
5-2
Page 56
Creating a user flatness array
In this example an HP the interface bus
(HP-IB)
437B
power meter controlled by the sweeper through
is used to enter the correction data into a flatness
array. Figure 5-l shows a typical system setup. The setup shown assumes that if
the setup has an external leveling configuration, that the steps necessary to correctly level have been taken. Refer to Chapter 2, “Externally Leveling the Sweeper, )) for information on external leveling.
;ALC
IN
I
1
--------.
I
I
;
I
I
----d
AND OTHER
i
*- - - -,-- --’
CAmEs
DO/ICES
I
1
RF
INPUT PORT
I
,
OUTPUT
I
;
-0-uq-J-N
“l%tE”E::
TEST
----------
Figure
5.1.
IFLATNESS CORRECTED ,OUTPUT
i .-----,--,
TEST PORT
POWER SENSOR
Creating a User Flatness Array
5-3
Page 57
Creating User Flatness Arrays
Creating a user flatness array
To set up the sweeper
Connect the equipment as shown in Figure 5-l. Do not connect the power
1. sensor to the system yet. Press
If a frequency range other than the full range of the instrument is desired,
2. use the
If external cables and/or devices are used between the sweeper leveling
3.
CSTART)
and
m
point (the RF OUTPUT if internally leveled, or the coupler/splitter output if externally leveled) and the remote test port, the nominal (average) loss of these components should be entered as an offset. To enter the power
offset, press
nominal loss from the leveling point of the sweeper to the test port. (For
(SHIFT)
OFFSET (in the POWER key group) and then enter the
example, if there is a 6 dB loss from the leveling point to the remote test port, enter a power offset of +6
(PRESET)
on the sweeper.
keys to input the desired frequency range.
dB.)
Set the power level to the level desired at the test
4.
[POWER LEVEL) and entering the desired number.
5.
Select
[ml
SPECIAL
112) (jj).
Use the @j @J keys to select
the type of power meter you will be using.
6.
Select (SHIFT) SPECIAL
113) (w’.
Enter the HP-IB address of the
(437B
purt
by pressing,
for this example.)
power meter you will be using for the calibration. (Thirteen is the default address for power meters.)
7.
Press
[%iW)
a. Select whether to calibrate over
FLTNESS
CAL
S?.at-:~t,op
(correction points will be
linearly spaced over the selected Start/Stop frequency range, or Fu 11
Bard (correction points will be linearly spaced over the full frequency
range of the instrument). Use the m Q) keys to make your selection, then press [Hz/s/ENTER).
b. Select the number of correction points, using the keypad for your entry
(valid entries range from 2 to Sol), then press
I-1.
c. The MARKER/SWEEP/STATUS display should now read:
Conned.
437B PM
--
EHTER.
5-4
Page 58
Creating User Flatness Arrays
Creating a user flatness array
To set up the power meter
l Zero and calibrate the power meter/sensor. l Enter the appropriate power sensor calibration factors into the power
meter.
l Enable the power meter/sensor
on the HP
l Connect the power sensor to the point where corrected
437B
power meter refer to its operating and service manual.
cal
factor array. For operating information
power
is desired.
(See Figure 5- 1.)
To start the user flatness cal
l
The MARKER/SWEEP/STATUS display should still read:
C:onnect.
l Press
correction point’s frequency and power as it is measured.
l When the calibration is finished, the flatness correction is automatically
turned on, and the FLTNESS ON LED annunciator is lit on the front panel
of the sweeper. Power correction will be linearly interpolated between the measured correction points. The POWER display will port power.
437B
PM -- ENTER.
(Hz/s/ENTEji)
to start the calibration. The display will show each
now.show
the test
Attention!
Before doing anything else, save this calibration in one of the instrument’s registers. If the calibration
has not been saved, and the instrument is preset, a register is recalled, or the power offset is changed, the calibration will be lost. To save the calibration, press of the instrument register you wish to save it in.
ISAVE)
and then enter the number
5-5
Page 59
Page 60
Creating User Flatness Arrays
Creating a user flatness array for use in a scalar analysis measurement
To set up the sweeper
1. Connect equipment as shown in Figure 5-2. Do not connect the power sensor to the system yet.
2. Press (PRESET) on the scalar analyzer (HP 8757).
3. If a frequency range other than the full range of the instrument is desired, use the
frequency range.
4. If external cables and/or devices are used between the sweeper leveling point (the RF OUTPUT if interrmlly leveled, or the coupler/splitter output if externally leveled) and the remote test port, the nominal (average) loss
of these components should be entered as an offset. To enter the power
offset, press nominal loss from the leveling point of the sweeper to the test port. (For
example, if there is a 6 dB loss from the leveling point to the remote test
port, enter a power offset of +6
m
and
@iYWj
CsTopl
keys on the sweeper to input the desired
OFFSET (in the POWER key group) and then enter the
dB.)
5. Set the power level to the level desired at the tat port by pressing
(POWER LEVEL] and entering the desired number.
6.
Select
the type of power meter you will be using.
7.
Select (SHIFT)
@iiF]
SPECIAL (12)
SPECfAL @
(HZ/S/ENTER_).
Use the
(437B
Q) @Jj
for this example.)
(Hz/s/ENTER_). Enter the
keys to select
HP-R3
address of the
power meter you will be using for the calibration. (Thirteen is the default address for power meters.)
To set up the analyzer
1. On the analyzer, set up the appropriate measurement (i.e. gain for an amplifier). Calibrate the measurement (thru and short/open calibration). Press sweeper parameters in storage register 1.
m
(iJ on the analyzer to store the analyzer configuration and
5-7
Page 61
Creating User Flatness Arrays
Creating a user flatness array for use in a scalar analysis measurement
Attention!
Do not proceed until the HP 8757 system interface is turned off by pressing pressing softkeys, MORE,
SWEZP
MODE, and
SYSIElTF
ON
@FiiZ].
OFF
on the analyzer.
Then by
To set up the power meter
1. Zero and calibrate the power meter/sensor.
2. Enter the appropriate power sensor calibration factors into the power meter.
3. Enable the power meter/sensor cal factor array. For operating information
on the HP
437B
power meter refer to its operating and service manual.
I
4. Connect the power sensor to the point where corrected power is desired. (See Figure 5-2.)
To start the user flatness cal
1. On the sweeper, press a. Select whether to calibrate over St>arSt.op (correction points will be
linearly spaced over the selected Start/Stop frequency range, or
Band
(correction points will be linearly spaced over the full frequency
range of the instrument). Use the @)
then press (Hz/s/ENTER).
5-8
m
FLTMESS
CAL
(JJ
keys to make your selection,
FIJ
11
Page 62
Creating User Flatness Arrays
b. Select the number of correction points, using the keypad for your entry.
Valid entries range from 2 to 801, then press
CH+./ENTER).
c. The MARKER/SWEEP/STATUS display should now read:
Connel:t.
437E:
PM --
Et,ITER.
2. Press
CHZ/S/ENTER)
to start the calibration. The display will show each
correction point’s frequency and power as it is measured.
3. When the calibration is finished, the flatness correction is automatically turned on, and the FLTNESS ON LED annunciator is lit on the front panel of the sweeper. Power correction will be linearly interpolated between the
measured correction points.
4. Save the calibration by pressing
Attention!
If the calibration is not saved in an instrument register, and the sweeper or analyzer is preset, a
register is recalled, or the power offset is changed, the calibration will be lost. Reactivating the SYSINTF on the analyzer will also cause a preset and loss of the calibration. Be sure to save the calibration as described in the previous step before continuing.
CsAVEJ
(iJ on the sweeper.
To reactivate the HP 8757 system interface
1.
Press the softkey SYSINTF OH OFF on the analyzer and the sweeper will
preset.
2. Press
3. Devices can now be tested.
(RECALL_)
(iJ to pull up the calibration.
5-9
Page 63
I
I
6
Operating a Master/Slave Setup
Page 64
Operating a Master/Slave Setup
Two HP 83750 series sweepers can be configured for two-tone measurement capabilities. Two synchronously tracking sweepers can be configured as a
“master/slave” pair for mixer characterization or inter-modulation distortion
measurements. The swept frequency accuracy allows control of the sweepers
at Exed- or swept-frequency offsets.
To implement master/slave operation, configure the sweepers as shown in Figure 6-l. The master sweeper generates the 10 MHz frequency reference for both sources. The source synchronization cable (HP part no. 83750-60059) must be connected between the auxiliary interface connectors on the rear panels of the sweepers.
When conEguring a two-tone measurement system, the slave does not need to interface with any instruments other than the master. The master should interface with the display device (scalar analyzer or oscilloscope, for example) as if it were a stand-alone sweeper.
6-2
Page 65
SOURCE SYNCHRO­NIZATION CABLE
AUXILIARY
INTERFACE
:
I I
Operating a Master/Slave Setup
-----------.e---
10 MHz REF OUT
pg424ab
Figure
“SLAVE”
SYNTHESIZED
SWEEPER
6-l.
Master/Slave Setup
AUXILIARY
INTERFACE
10 MHz REF IN
6-3
Page 66
Tb set up the master sweeper
1. Press
2. Activate the master mode by pressing
If the display does not read
(jPRESETJ.
[M)
13-A rlMsde=MHSTER,
(SPECIAL)
then press the
(19) C-1.
once to set the control mode to master.
3. Set the desired frequencies by using the
(CJV(SPAN)
NOTE
It is recommended that the master and slave sweepers frequencies not differ more than 5 point in the sweep. If this requirement is not met, it is possible for unlocks to occur. If frequency offsets greater than 5
unlocks.
keys, or the
GHz
are required, it may be necessary to reduce the sweep speeds to eliminate
Icw]
key.
CSTARTY(STOP_)
keys, the
4. Set the desired power level using the (POWER LEVEL] key.
m
GHz
key
at any
For swept measurements.
5. Set the desired sweep time using the (TIME) key. (The recommended minimum sweep time for master/slave mode is 100 ms.)
NOTE
The master and slave sweepers must have identical sweep times. Therefore, the value entered here must also be entered into the slave sweeper.
6-4
Page 67
Operating a Master/Slave Setup
To set up the master sweeper
For stepped measurements.
6.
Set the sweep mode to stepped by pressing
then pressing the @) key until the display reads:
7.
Set the number of points by pressing
@iE)
(GiF]
SWEEP MODE $ , and
Si>Jeep=
St. ep lzt:tr,t.
POINTS and entering the
desired number of points.
NOTE
The master and slave sweepers must have identical sweep mode, number of points, and sweep time settings. Therefore, the values entered here must also be entered into the slave sweeper.
.
6-5
Page 68
31 set up the slave sweeper
1. Press (PRESET).
2. Activate the slave mode by pressing (SHIFT) Press the @) key until the display reads
3. Set the desired frequencies by using the
(EEiK] 119) (-1.
Cnt.
t- 1
Mode=SLHVE.
(START)/0
keys, the
@J(SPAN) keys, or the ccw) key.
NOTE
It is recommended that the master and slave sweepers frequencies not differ more than 5 point in the sweep. If this requirement is not met, it is possible for unlocks to occur. If frequency offsets greater than 5
unlocks.
4. Set the desired power level using the
For
swept measurements.
GHz
are required, it may be necessary to reduce the sweep speeds to eliminate
POWER
LEVEL] key.
GHz
at any
5. Set the sweep time to the same value as the master sweeper using the
m
key. (The recommended minimum sweep time for master/slave
mode is 100 ms.)
For stepped measurements.
6.
Set the sweep mode to stepped by pressing (SHIFT)
then pressing the @) key until the display reads:
7. Set the number of points to the same value as the master sweeper by pressing
6-6
(SHIFTI)
SWEEP
Sweep= St. ep
MODE $ , and
POINTS and entering the number of points.
C:ont..
Page 69
See also
For a
“Multi-pin Connectors” section of Chapter 7, “Front/Rear Panel.”
pinout
description of the source synchronization cable, see the
6-7
Page 70
7
I
-
-
Front/Rear Panel
Page 71
Front/Rear Panel
This chapter contains detailed information on various aspects of the sweeper front and rear panel. Information on the following can be found in this chapter:
0
Connectors
l Display l HP-B Lang/Address Switch l Knob l Line Power Switch l Power Cables
7-2
Page 72
Connectors
pg412ab
00 00 00
.-I 0 ::
0 0
a
0 0
-
LJLJ 00 00 00 0
@
I \
SWEEP OUT
Figure 7-1. Sweeper Connectors - Front Panel
0
@@”
0
ALC IN
‘El0000
00000
moooo
l30000
Cl
0
0
la
Q @I
L
RF
,
\
\/
OUTPUT
7-3
Page 73
(OPT
RF
OUTPUT
(OPT lE4 ONLY)
Front/Rear Panel
Connectors
SOURCE MODULE
I NTERFACE
1EE
ONLY)
I
SWEEP
OUTPUT
Z AXIS
BLANK/MKRS
I
STOP
SWEEP
PULSE
I N/OUT
FM INPUT
AUX
I NTE
I
LIARY
DIARY
:RFACE:RFACE
HP-IB
HPllB
VOLTi/GHz
VOLTS/GHz
OUTPUTOUTPUT
A’M
AM
INPUTINPUT
TRIGGER TRIGGERTRIGGER TRIGGER
OUTPUTOUTPUT
INPUTINPUT
Figure 7-2. Sweeper Connectors - Rear Panel
1OilHz
1OMHz
REF
REF
OUT
OUT
1OivlHz
1OMHz
REF
REF
IN
IN
pg413ab
7-4
Page 74
BNC Connectors
Front/Rear Panel
Connectors
10 MHz REF IN
10 MHz REF
OUT
ALC IN
AM INPUT
FM
INPUT
This rear panel female BNC connector accepts a -5 to + 10
dBm
signal from an external time base reference which is within MHz. The nominal input impedance is 50
510
ppm of 10 MHz or any sub-multiple down to 1
61.
This connector detects when a valid reference signal is connected to it and automatically switches from internal to external reference operation.
This rear panel female BNC connector provides a nominal
signal level of 0
dBm,
and output impedance of 50 0. The
accuracy is determined by the tune base used. This front panel female BNC connector is used for external
power meter leveling or external negative detector leveling. The damage level is
100
kQ.
f15 V.
The nominal input impedance is
This rear panel female BNC connector accepts the amplitude modulating signal input when External AM is enabled. The nominal input impedance is 3.5
f15
v.
kQ.
The damage level is
This rear panel female BNC connector accepts the frequency
modulating signal input when DC or AC FM is enabled. The
nominal input impedance is 1
f15
v.
k62.
The damage level is
PULSE IN/OUT
This rear panel female BNC connector can be used as either an external pulse input or output for internal pulse
modulation. In either case, it is TTL level compatible and
has a nominal impedance of 5
khl.
The damage level is
2 +lOVor< -5V.
A TTL high level (> +2 V) enables the selected power level to be at the RF OUTPUT connector, while a TTL low level turns the RF off.
7-5
Page 75
Front/Rear Panel
Connectors
STOP SWEEP
SWEEP OUT
This rear panel female BNC connector causes a sweep to be stopped when this input is pulled low. Retrace does not occur, and the sweep resumes when this input is pulled high. The open circuit voltage is TTL high and is internally pulled low when the instrument stops its sweep. The damage level is 2 + 10 V or 5 -4
V.
These front and rear panel female BNC connectors provide a voltage range of 0 to + 10 V. When the instrument is sweeping, the SWEEP OUT ranges from 0 V at the beginning of the sweep and + 10 V at the end of the sweep regardless of the sweep width. In CW mode, the SWEEP OUT ranges from 0 V at the sweeper minimum frequency to + 10 V at the specifled maximum frequency, with a proportional voltage for frequencies between the specified minimum and
maximum. When the sweeper is in manual sweep operation,
the SWEEP OUT voltage is a percentage of the span. The nominal output impedance is 100
f0.05%, f5 mV
When used with the HP
into a high impedance load.
8757D
61.
The typical accuracy is
scalar analyzer in
ramp-sweep mode, the rear panel output is a series of 0 to
+ 10 V pulses similar to the Trigger Output.
TRIGGER INPUT
TRIGGER OUTPUT
VOLTS/GHz
OUTPUT
7-6
This rear panel female BNC input is activated on a TTL
rising edge. It is used to externally initiate an analog sweep
or to advance to the next point in stepped sweep mode. The
damage level is 2 + 10 V or 2 -4
This rear panel female BNC connector outputs a 1
V.
,ns
wide
negative-going TTL pulse at 1601 points evenly spaced
across an analog sweep, or at each point in stepped sweep
mode. When used with the HP
8757D
scalar analyzers, the number of pulses per sweep (in analog sweep mode) is determined by the number of HP 8757 trace points.
This rear panel female BNC connector supplies a voltage that is proportional to the RF output frequency, which can be configured to any desired sensitivity and offset within
a
f12
V range. The default setting is 0.5
minimum load impedance is 2
3~0.1% f10
mV
kbl.
The typical accuracy is
V/GHz.
The
Page 76
Z
AXIS
BLANK/MKRS
Front/Rear Panel
Connectors
This rear panel female BNC connector supplies a positive rectangular pulse (approximately +5 V into 2
kfl)
during the retrace and band switch of the RF output when the instrument is sweeping. This output also supplies a -5 V pulse when the RF output is coincident with a marker frequency.
7-7
Page 77
Front/Rear Panel
Connectors
Multi-pin Connectors
AUXILIARY INTERFACE This connector provides special control signals used in master/slave interface
operation with another source, and in other special-purpose applications. This connector is a
25-pin
AUXILIARY INTERFACE
D-subminiature receptacle located on the rear panel.
SOURCE SYNCHRONIZATION
(HP P.N. 8375040059)
Figure 7-3. Auxiliary Interface Connector
CABLE
7-8
Page 78
Table 7-1. Pin Description of the Auxiliary Interface
Pin
#
1
2 Z-Axis Blanking/Markers out
3 4 5 6
1
8 9 10 Rear Panel Control 1 11 12 13 14 Low Marker out 15 16 17 Rear Panel Control 2 18 19 Digital Ground 20 No Connection 21 22 No Connection 23 low Source Settled 24 25
*
Or 0 to 10 V pulses when used with a HP 87570 scalar analyzer.
Alternate Sweep Out out
Spare Rear Panel Control 3 low Stop Sweep
+5.2
v out Digital Ground Low Divider-Sync External Trigger
Rear Panel Control 0
low
Retrace
Alternate Sweep In
Low
Clualified
Stop Sweep
Rear Panel Control 4
Sweep Output
No Connection
low Source Settled
Spare
Function
In/Out Signal
In
I/O I/O
TTL
I/O TTL
In
I/O I/O TTL I/O
TTL
In
out out TTL
I/O
TTL
out
In TTL out out
Front/Rear Panel
level
TTL
Blank- +5 V Marker- -5 V
f12
v
TTL
+5.2
v
Gnd
TTL TTL
TTL TTL rr!
0
to 10 V ramp”
Gnd
TTL f12
v
Connectors
7-9
Page 79
Front/Rear Panel
Connectors
HP-II
This connector allows the sweeper to be connected to other instruments or devices on the interface bus. Details of this cable are shown in Figure 7-4.
HP
table following the figure.
HP-IE
SHl AH1
TE
LELo4 SRl
RL 1
PPO
DC1
COl-3, 28
DC1
El
part numbers for various HPIB cables that are available are shown in the
-07
24 :: 12
:: :: :: ::
13 ::
1
0
4
pg415ob
Figure 7-4. HP-IB Connector and Cable
7-10
HP-IB Interface Cables Available
Page 80
Front/Rear Panel
Connectors
As many as 14 HP-B instruments can be connected to the sweeper (15 total instruments in the system). The cables can be interconnected in a star pattern (one central instrument with the
HP-lB
cables emanating from that instrument like spokes on a wheel), or in a linear pattern (like boxcars on a train), or a combination of the two. There are certain restrictions that must be followed when interconnecting instruments:
l Each instrument must have a unique HP-B address, ranging from 0 to 30
(decimal). Refer to
“BDDRESS ’
in the “Instrument State Keys” chapter for information on
“HP-IB
Lang/Address Switch” later in this chapter, or
setting the sweeper HP-B address.
l In a two-instrument system that uses just one HP-B cable, the cable length
must not exceed 4 meters (13.2
l When more than two instruments are connected on the bus, the cable
ft).
length to each instrument must not exceed two meters (6.6 ft).
l The total cable length between all instruments must not exceed 20 meters
(65 ft).
Hewlett-Packard manufactures HPIB extender instruments (Models HP
37201A
and HP
37204AiB)
that overcome the range limitations unposed by the cabling rules. These extenders allow twin pair cable operation up to 1 km (3,280 service offices can provide additional information on the
ft),
and telephone modem operation over any distance. HP sales and
HP-IB
extenders.
7-11
Page 81
Front/Rear Panel
Connectors
The codes next to the
HP-II3
connector, illustrated in Figure 7-4, describe the HP-IB electrical capabilities of the synthesizer, using IEEE Std. 488-1978 mnemonics (HP-IB, GP-IB,
IEEE-488,
and RX-625 are all electrically
equivalent). Briefly, the mnemonics translate as follows:
SHl Source Handshake, complete capability. AH1 T6:
Acceptor Handshake, complete capability. Talker; capable of basic talker, serial poll, and unaddress if
MLA. TEO
L4
LEO
l%Iker,
Extended address; no capability. Listener, capable of basic listener, and unaddress if MTA. Listener, Extended address; no capability.
SRl Service Request, complete capability. RLl PPO DC1 DTl
Remote Local, complete capability. Parallel Poll, no capability. Device Clear, complete capability. Device Trigger, complete capability.
CO, 1, 2, 3, 28
El
Controller capability options; CO, no capabilities; Cl, system controller;
C28,
send I. F. messages.
C2,
send IFC and take charge;
C3,
send REN;
Electrical specihcation indicating open collector outputs.
These codes are described completely in the IEEE Standard 488 (1978)
IEEE Standard Digital Interface
for
Programmable Instrumentation
identical ANSI Standard MCl. 1.
7-12
or the
Page 82
Front/Rear Panel
Connectors
SOURCE MOOULE This connector sends and receives digital and analog signals to and from an INTERFACE Only)
(Option 1EE
HP 83550Series millimeter-wave source module. With the source module connected, the sweeper assumes the characteristics of the source module.
MOD
MOD
MOD Cl
Dl
DIG GND
f5V
EXT LVL (COAX)
RESERVED5
-15v
V/GHz
Figure 7-5. Interface Signals of the Source Module Connector
The codes indicated in Figure 7-5 translate as follows: MOD DO
Source module data line zero. Signals MOD DO through MOD D3 are the millimeter source module data bus lines (bidirectional).
RTN
SENSE
RESERVED
pg416ot
MOD MOD D2 MOD D3 MOD CO
MOD Cl
Dl
Data line one. Data line two. Data line three. Source module control line zero. Signals MOD CO and MOD
Cl are the control lines for the read/write to and from the millimeter source module.
Control line one.
7-13
Page 83
Front/Rear Panel
Connectors
MOD SENSE
L MOD RF OFF EXT
LVL
RET
EXT LVL
0.5V/GHz
-15v
+15v
+8V
+5v
NOTE
Source module sense. A 1 mA current is injected on this line by the millimeter source module to indicate its presence. This signal always equals 0 V.
Low = RF off. Source module RF is turned off. Source module external leveling return.
Source module external leveling input, from the millimeter
source module. Internal
0.5V/GHz
to the millimeter source module.
Power supply. Range is -14.25 to -15.45 V. (See note below.)
Power supply. Range is + 14.25 to + 15.45 V. (See note below.)
Power supply. Range is +7.50 to +8.45 V. (See note below.) Power supply. Range is
+4.75
to +5.40 V. (See note below.)
The values for the four power supplies above are valid when a millimeter head is connected. With no
load on the connector, the values would be:
-
15 v supply: + 15 v supply: + 8 v supply: + 5 v supply:
DIG GND
.5
V/GHz
RTN 0.5
ANLG GND
-14.4 to -15.45 v
+14.4
v to
+15.45
v
+7.75
v to
+a.45
v
+4.75
v to
+5.40
v
Digital ground.
V/GHz
return.
Analog ground return.
RET
7-14
Page 84
Front/Rear Panel
Connectors
RF Output Connector
The standard front panel RF OUTPUT connector is a precision 3.5 mm male connector. When Option ruggedized Type-N connector. The nominal source impedance is 50
When making connections to this connector, carefully align the center conductor elements, then rotate the knurled barrel while mating components remain still. Tighten the barrel until Iirm contact is made.
lkke
care when working with this connector. If the connector is mechanically
degraded in any way, high frequency losses can occur. Refer to application note 326, Coaxial number
5954-
Sz/stems -
1566) for more information.
1ED
is installed, this front panel connector is a
Principles of Microwave Connector Care
61.
(HP
part
7-15
Page 85
Display
The sweeper front panel display contains two areas (Frequency, and
Marker/Sweep/Status and Power) for displaying the current operating
parameters of the sweeper. Front panel annunciators show the status of several of the sweeper functions and settings.
FREQUENCY DISPLAY
MARKER/SWEEP/STATUS
DISPLAY
Figure 7-6. Sweeper Display
POWER DISPLAY
pg410pb
7-16
Page 86
Front/Rear Panel
Display
Frequency Display
The left-hand display is labelled FREQUENCY and is used exclusively to show information about frequencies being generated. (Information about
markers is not pertinent to generated frequencies and is not displayed here.)
In Start/Stop and CF/Span sweep modes, the left display has two separate
fields of 11 characters each, whose contents depend on the instrument mode.
Two blank spaces separate the two fields. Frequencies are displayed using eight signihcant digits, followed by a three-character frequency unit
MHz, or
GHz).
The decimal multiplier is changed automatically to keep the number being displayed in the range of 1 to 999.99999, and digits below 1
kHz
resolution are not shown. Leading zeros are suppressed.
(kHz,
In CW mode, there is one 24-character field. Frequencies are displayed using up to 11 significant digits, followed by a three-character frequency unit MHz, or
GHz).
The decimal multiplier is changed automatically to keep the
@Hz,
number being displayed in the range of 1 to 999.99999, and digits below 1 Hz resolution are not shown. Leading zeros are suppressed.
NOTE
The decimal separator (either . to the “Special Functions” chapter for more information.
or , I can be changed by selecting Special function number 18. Refer
7-17
Page 87
Front/Rear Panel
Display
Marker/Sweep/Status and Power Display
The right-hand display is divided into two fields, although there is no physical separation between them. (Normally, one blank space is inserted between the left and right Eelds. When some features are accessed, however, all 24
characters of the right display can be used.)
The left Eeld is labelled MARKER/SWEEP/STATUS, and is used to show a variety of information including marker frequencies, sweep time, and general instrument status. Sixteen characters are reserved for this Eeld, including any units displayed.
The right Eeld is labelled POWER, and displays the power level in
dBm.
seven character cells wide. Exed format.
NOTE
When operating the sweeper in a secure environment, the display can be blanked so that the sweeper
parameters and status cannot be viewed. For more information, refer to the “Special Functions” chapter of this manual.
It is
I
7-18
Page 88
Annunciators
Front/Rear Panel
Display
DISPLAY
ANNUNCIATORS
LED
ANNUNCIATORS
Figure 7-7. Sweeper Display and LED Annunciators
The front panel annunciators show the status of several of the sweeper
functions and settings. Refer to Figure 7-7. Notice that there are two types of
annunciators: display and LED.
A display annunciator is not visible if its associated function is not active or selected. For example, if external pulse modulation is currently selected, the
EXTMOD annunciator will be lit, otherwise, it will not be visible.
The display annunciators that will appear in the left-hand (frequency) display are SHIFT, START, CF, CW, MAN, SWEEP, EXT REF, ALT, STOP,
SPAN, MULT, and OFFSET. The display annunciators that will appear in
the right-hand (marker/sweep/status and power) display are MSG, MKRS,
EXTMOD, INTMOD, UNLOCK, STEP, UNCPLD, EXT ALC, PWR SWP,
SLOPE, UNLEV, and RF OFF.
7-19
Page 89
Front/Rear Panel
Display
The front panel LED annunciators (R, L, T, S, MKR A, Ml-+M2 SWP, UNLOCKED FM, and FLTNESS ON) will be lit to show you that an
associated condition exists. An explanation of each annunciator follows. Display
Annunciators SHIFT
This annunciator appears when goes off when any other key is pressed (or when
(ml
is pressed, and
m
pressed again).
is
START
CF
cw
MAN
SWEEP
EXT REF
ALT
This annunciator appears when and goes off when Icw),
ICF),
This annunciator appears when and turns off when Icw),
This annunciator appears when off when
a, @Gi), (START),
(START)
or
m
(ZJ
(START),
Icw_l
or
or [STOP) is pressed,
is pressed.
or
or
m
ISTOP)
is pressed,
is pressed.
is pressed, and turns
ISTOP)
is pressed.
This annunciator indicates that manual sweep mode has
been selected.
This annunciator indicates that the instrument is performing a sweep. This annunciator appears only while the
instrument actually sweeps, not while it waits for a trigger
signal (even though it is in sweep mode). In continuous sweep mode, the annunciator will blink with slow sweeps,
but remain on with fast sweeps.
This annunciator indicates that an external frequency
reference is in use.
This annunciator indicates that the alternate sweep mode is active.
STOP
SPAN
MLJLT
7-20
This annunciator appears when and turns off when Icw),
a,
This annunciator appears when turns off when Icw], (START), or
(VZ??i]
or
ISPAN)
LCF)
m
or
is pressed.
or
m
is pressed.
~
is pressed
is pressed and
This annunciator indicates that the displayed frequency is a
multiple of the actual RF output frequency.
Page 90
Front/Rear Panel
Display
OFFSET
MSG
MKRS
EXTMOD
INTMOD
UNLOCK
STEP
This annunciator indicates that the displayed frequency is offset from the actual RF output frequency.
This annunciator indicates that the sweeper has a new message that has not been displayed, such as an error message or instrument failure message. Messages are viewed by pressing the front panel
CMsGl
key.
This annunciator appears when any of the markers (0 to 9) are on.
This annunciator appears under any of the following conditions:
l AM Mode = External l FM Mode = AC/Locked l Pulse Mode = External
or
DC/Unlocked
This annunciator appears if Pulse Mode is set to Internal, Scalar, or 1
kHz
square-wave.
This annunciator appears when any of the internal synthesizer loops are unlocked.
This annunciator indicates that the Sweep Mode is set to Stepped Sweep.
UNCPLD
EXT ALC
PWR SWP
SLOPE
UNLEV
RF OFF
This annunciator indicates that automatic coupling between the attenuator and the ALC reference level is turned off in the Special area. (For Option
1El
instruments only). This
mode is automatically selected whenever the ALC mode is
not in internal mode or unleveled mode. This annunciator indicates that the ALC mode is set to
either Diode or Power meter. This annunciator indicates that the Power Sweep function is
active. This annunciator indicates that the Power Slope function is
active. This annunciator indicates that the source is unable to
maintain the correct output power level for the current
output frequency, or that the ALC is turned off. This annunciator indicates that the RF output is turned off.
7-21
Page 91
Front/Rear Panel
Display
LED
Annunciators
R
This LED annunciator is lit when the sweeper is in remote
HP-II3
operation.
L
This LED annunciator is lit when the sweeper is receiving information or commands over HP-IB.
T
This LED annunciator is lit when the sweeper is transmitting information over HP-IB.
S
MKR A
Ml-M2 SWF’
This LED annunciator is lit when the sweeper is receiving
an
SRQ over HP-IB. This LED annunciator is lit if the marker delta mode is on. This LED annunciator is lit when
(MI+MP
SWEEP) is pressed.
UNLOCKED FM This LED annunciator is lit if the FM mode is set to
DC/Unlocked.
FLTNESS ON
This LED annunciator is lit if a power flatness correction array is enabled.
7-22
Page 92
HP-IB Lang/Address Switch
The HP-IR switch on the rear panel is used to select the sweeper language and
HP-II!
address when you apply AC power. These parameters can also be
selected via front panel operation. Refer to Figure 7-8 for a pictorial explanation of the various switch settings
for the HP-IB address of the sweeper. The selected address depicted in this figure represents an HP-IR address of 19.
When the HP-IR portion of the switch (switch positions 1 through 5) are all set to 1, any address changes made from the front panel are retained through a power cycle of the instrument. If you want your sweeper to default to a particular address at power up, set the switch to the desired. address; you can still make changes from the front panel, but they won’t be retained through a
power cycle.
Refer to Figure 7-9 for a pictorial explanation of the switch settings for instrument language.
Refer to Figure 7-10 for a pictorial explanation of how to set the switch for clearing the register contents at power on.
SELECTED
POSITION
ADDRESS
ADDRESS-
ADDS 1 TO ADDRESS IF SELECTED POSITION = ADDS 2 TO ADDRESS IF SELECTED POSITION = ADDS 4 TO ADDRESS IF SELECTED POSITION = ADDS 8 TO ADDRESS IF SELECTED POSITION = ADDS 16 TO ADDRESS IF SELECTED POSITION =
1 1 1 1
1
Figure 7-8. HP-IB Address Switch Settings
7-23
Page 93
I
Front/Rear Panel
HP-18
Lang/Address Switch
= SCPI
1
1
I I=!
0
0
Figure 7-9. Instrument language Switch Settings
=
Register Contents Are
Through An AC Power Cycle.
1
1
1 I=!
0
=
(111;
Figure
7-l
0. Clear Register Contents Settings
Register Contents Are Cleared
Through An AC Power Cycle.
Saved
7-24
Page 94
See Also
Front/Rear Panel
HP-IB Lang/Address Switch
“Instrument
For information on setting the
State Keys” panel of the sweeper.
Special
Functions”
For information on setting the language from the front panel
of the sweeper.
HP-IB
address from the front
7-25
Page 95
Knob
The front panel knob is used to increase or decrease parameter values. Pointers ( that the function is active and its value will be increased or decreased when
knob is rotated.
The front panel knob is used to give an analog feel to the setting of the
values. Any of the values that can be set through the numeric entry pad, or
the step keys, can also be set using the knob. However, the rate at which the
active parameter varies, for a given amount of knob rotation, is dependent on the parameter that is being controlled.
WV )
over a parameter or numeric value in the display indicate
If ENTRY OFF
(m
[H
Z/S/ENTER
)
has been enabled, the pointers will disappear from the display and rotating the knob will have no affect on the sweeper parameters until another function key is pressed.
Equivalent SCPI Commands
The equivalent SCPI commands for knob rotation are SYST : KEY 6 1 which simulates clockwise knob rotation, and SYST:KEY 62 which simulates counterclockwise knob rotation.
7-26
Page 96
Line POWER Switch
P
INSTRUMENT STATE
7m
SHIFT
PRESET MODE
1
/
PRESETI
II
I--
II I
c
c
STANDBY LINE POWER SWITCH
pg49ab
Figure
The line POWER switch turns power to the sweeper to either on or standby. The sweeper line POWER switch is located at the bottom left corner of the
front panel. Pressing the switch so that the button locks in the “in” position turns the sweeper on and pressing the switch again so that it is in the “out” position turns it to standby. When set to standby, most of the sweeper circuitry is powered off, however, power to the internal timebase (Option
lE5) is not disconnected.
When the sweeper is on, the green LED indicator will be illuminated. When the sweeper is in standby, the amber LED indicator will be illuminated.
7-l
1. The Sweeper line POWER Switch
7-27
Page 97
Front/Rear Panel
line POWER Switch
WARNING
Before turning the sweeper on, make sure that it is grounded through
the protective conductor of the power cable to a mains power receptacle provided with protective earth contact. Any interruption of the protective grounding conductor inside or outside of the sweeper or
disconnection of the protective earth terminal can result in personal injury.
7-28
Page 98
Power Cables
The line power cable is supplied in one of several configurations, depending on the destination of the original shipment.
Each instrument is equipped with a three-wire power cable. When connected
to an appropriate AC power receptacle, this cable grounds the instrument
chassis. The type of power cable shipped with each instrument depends on the country of destination. See Figure 7-12, “Power Cable and Line (Mains) Plug Part Numbers”, for the part numbers of these power cables. Cables are available in different lengths and some with right-angle plugs to the instrument. Check with your nearest Hewlett-Packard service center for descriptions and part numbers for these cables.
NOTE
There is no line voltage selector switch on the instrument. The line module will automatically switch between 110 and 220 Vat as needed.
7-29
Page 99
FrontlRear Panel
PLUG TYPE * *
250V
250V
25DV
125V
250V
$;AT
NUMBER
8120-1351
8120-1703 90’
8120-1369
8 120-0696
8120-1689 8120-1692
8120-1348 Straight*
8120-1538
8120-1378
8120-4753 Straight
8120-1521
8120-4754 90’
8120-5182
8120-5181
90’
DESCRIPTION LENGTH COLOR
Straight*
Straight*
90-
Straight*
90’
Straight”
90’
Straight*
90’
PLUG
BS1363A
NZSS196/ASC112
CEE7-Yll
NEMAS-15P
NEMAS-15P
NEMA5-15P
CABLE
ct.4 (INCHES)
229 (90) Mint Gray 229 (90) Mint Gray Cyprus, Nigeria
201 (79)
221 (87)
201 (79)
201 (79) Mint Gray
203 (80) Block 203
(80)
203 (80) 230 (90) 203 (80)
230 (90)
200 (78)
200 (78)
CABLE
Gray
Gray
Mint Gray
Black
Jade Gray 200 v), Bra2 II,
Jade Gray
Jade Gray Jade
Gray
Jade
Gray
Jade Gray
FOR USE
IN COUNTRY
Great Britain.
Singapore,
Zimbabwe
Argentina, Australia. New Zealand,
Mainland China
East and West
Europe’
African Republic
United Arab Republic
(unpclorized
many nat United States
Canada,
Japan (100 V or
Colombia,
Philippines,
Saudia Arabia.
Taiwan
Israel
Central
Ions)
Mexico
in
*
Part number for plug 1s industry identifier for plug only. Number shown for cable
HP Part Number for complete cable.
**
E = Earth Ground; L = Line; N = Neutral.
Figure 7-12. Power Cable and line (Mains) Plug Part Numbers
7-30
lncludlng
plug.
i
Page 100
I
8
Instrument State Keys
Loading...