Hewlett-Packard
Company 1989,1990, 1991,
1993, 1994
All Rights Reserved.
Reproduction, adaptation,
or translation without
permission is prohibited,
except as allowed under
the copyright laws.
HP part number
05890-90271
First edition—Jun 1989
Printed In U.S.A.
Second edition—Oct 1989
Printed In U.S.A.
Third edition—Jan 1990
Printed In U.S.A.
Fourth edition—Oct 1990
Printed In U.S.A.
Fifth edition—Oct 1991
Printed In U.S.A.
Fifth edition—Mar 1993
Printed In U.S.A.
Sixth edition—Jul 1994
Printed In U.S.A.
Printed in USA
Warranty
The information contained
in this document is subject
to change without notice.
Hewlett-Packard makes no
warranty of any kind with
regard to this material,
including, but not limited
to, the implied warranties
of merchantability and
fitness for a particular
purpose. Hewlett-Packard
shall not be liable for errors
contained herein or for
incidental or consequential
damages in connection with
the furnishing,
performance, or use of this
material.
Safety Information
The HP 5890 Series II and
HP 5890 Series II Plus are
IEC (International
Electrotechnical
Commission) Safety Class 1
instruments. This unit has
been designed and tested in
accordance with recognized
safety standards. Whenever
the safety protection of the
HP 5890 Series II has been
compromised, disconnect
the unit from all power
sources and secure the unit
against unintended
operation.
Safety Symbols
This manual contains
safety information that
should be followed by the
user to ensure safe
operation.
WARNING
A warning calls attention
to a condition or possible
situation that could cause
injury to the user.
CAUTION
A caution calls attention to
a condition or possible
situation that could
damage or destroy the
product or the user’s work.
Important User
Information for In Vitro
Diagnostic Applications
This is a multipurpose
product that may be used
for qualitative or
quantitative analyses in
many applications. If used
in conjunction with proven
procedures (methodology)
by qualifiedoperator, one
of these applications may
be In Vitro Diagnostic
Procedures.
Generalized instrument
performance characteristics
and instructions are
included in this manual.
Specific In Vitro Diagnostic
procedures and
methodology remain the
choice and the
responsibility of the user,
and are not included.
Sound Emission
Certification for Federal
Republic of Germany
If Test and Measurement
Equipment is operated with
unscreened cables and/or
used for measurements in
open set-ups, users have to
assure that under these
operating conditions the
Radio Interference Limits
are still met at the border
of their premises.
The following information
is provided to comply with
the requirements of the
German Sound Emission
Directive dated January 18,
1991
Sound pressure Lp <
70db(A)
During normal operation
At the operator position
According to ISO 7779
(Type Test)
When operating the HP
5890 Series II with cryo
valve option, the sound
pressure
cryo valve operation for
short burst pulses.
78 db(A) during
Schallemission
Werden Meß- und
Testgeräte mit
ungeschirmten Kabeln
und/oder in offenen
Meßaufbauten verwendet,
so ist vom Betreiber
sicherzustellen, daß die
Funk-Entströbedingungen
unter Betriebsbedingungen
an seiner
Grundstücksgrenze
eingehalten werden.
Diese Information steht im
Zusammenhang mit den
Anforderungen der
Maschinenl
sverordnung vom 18
Januar 1991.
Schalldruckpegel LP < 70
dB(A)
Am Arbeitsplatz
Normaler Betrieb
Nach DIN 45635 T. 19
(Typpr
Bei Betrieb des HP 5890
Serie II mit Cryo Ventil
Option treten beim Oeffnen
des Ventils impulsfoermig
Schalldrucke Lp bis ca. 78
dB(A) auf.
ärminformation
üfung)
Little Falls Site
Contents
Chapter 1 — Columns and Fittings9.................
The HP 5890 SERIES II (hereafter referred to as HP 5890) provides
flexibility in choices among inlets, columns, and detectors through use of
liners and adapters, allowing any standard column to be used without
sacrificing performance. Additionalflexibility is gained through positions
of inlets and detectors relative to each other and through the large
internal volume of the oven.
This section provides information for the following:
C
The column oven.
C
Fittings.
C
Liners and inserts.
C
ECD and TCD capillary makeup gas adapters.
The first three items must be considered before a column may be
installed properly at either an inlet or a detector. In addition, for an FID
or NPD, and depending upon the column to be installed (packed versus
capillary), the correct jet must be installed before installing the column.
Jet installation is described in Chapter 8, Preventive Maintenance.
10
For specific information on ordering fittings, liners, and inserts, see
Hewlett•Packard's analytical supplies catalog.
Columns and Fittings
Column oven
Column oven
Figure 1-1
Inlet Ftg
Nut Plate
The Column Oven
The oven door latch, located beneath the lower right corner of the door, is
pressed upward to open the door.
Motor•drivenflapsat the rear of the oven admit room air for cool down or
near•ambientoperation, so the door is kept closed except for access to
columns (the oven cools most efficiently with its door closed).
Det Ftg
The oven can maintain temperature down to about 7
^
C above ambient
without auxiliary cooling. If lower temperatures are required, a
cryogenic valve (for either liquid CO
1
or liquid N1) is needed. Liquid CO
permits reliable temperature control down to -50^C; liquid N1provides
reliable control down to -80
450
^
C.
^
C. The maximum oven temperature is
1
11
Columns and Fittings
Column oven
Column placement
Generally, a column may be installed between any inlet and detector.
A rigid1/4•inchpacked glass column, however, if installed in the B
(rear•most)inlet, can only be installed in the B (rear•most)detector.
Distance relationships among inlets and detectors are shown in
Figure 1•2.
Figure 1-2
Top View
(showing relationshipof inlets to detectors)
B
A
Front
Installation Restrictions, Rigid Columns
1mm
228 +
B
A
12
Packed column
Packed columns require no physical support other than that provided by
proper installation at inlet and detector fittings.
Columns and Fittings
Column oven
Hewlett-Packard capillary columns
Hewlett•Packardcapillary columns are wound on wire frames which
mount on a pair of brackets which slip into slots at the top of the oven
interior.
Figure 1-3
Typical Hewlett-Packard Capillary Columns
13
Columns and Fittings
Fittings
Figure 1-4.
Column Hanger
Part No. 1460-1914
Installed Bracket for Hewlett-Packard Capillary Columns
Column Installed
The bracket has two positions from which to hang the column wire frame.
Depending upon frame diameter, use the position which best centers the
column in the oven. Column ends should come off the bottom of the
frame, making smooth curves to inlet and detector fittings. Avoid
allowing any section of the column itself to come in contact with oven
interior surfaces.
Fittings
The following is a brief comparison of common types of fittings (nuts,
ferrules, O•rings)used to install columns, and to install inlet and detector
liners and/or inserts. Each type has its own set of advantages and
disadvantages:
14
Columns and Fittings
Fittings
C
C
C
Graphite O•ringsor ferrules have excellent sealing quality and long
service life, can be used continuously to 400
^
C, and are generally
recommended for most applications, particularly capillary and glass
columns. They are also recommended for inlet and detector liners, and
for split/splitlesscapillary inlet inserts.
Since they do not adhere permanently to glass or metal, they can be
removed easily without damage to the column, tubing, liner, or insert.
A rear metal ferrule may be needed if recommended by the
manufacturer.
Commonly used with metal columns and tubing, brass nuts and
ferrules on the column prevent damage to inlet and detector liners,
but may develop leaks above 250
^
C or with temperature
programming.
Also commonly used with metal columns and tubing, stainless steel
fittings minimize possibility of leakage at high temperature, but
require care in installing columns; overtightening may damage the
column end or inlet/detector fitting.
C
Teflon ferrules may be used to 250^C but are recommended only for
isothermal work, because they develop leaks when
temperature•programmed.
C
Commonly used with glass columns, Vespel (or graphite•filledVespel)
ferrules are reusable and work well to 350
^
C. These ferrules may
leak or crack if tightened when cold.
C
Glass columns can be installed using silicone O•rings. For O•ring
installation, a back metal ferrule, reversed, is necessary to provide a
flat surface to seal against.
Silicone O•ringsare useful to about 250
^
C, but, due to bleed, interfere
in high•sensitivitywork. They also gradually lose elasticity and crack,
so they must be replaced fairly often.
15
Columns and Fittings
Fittings
Table 1-1.Typical Fittings for Columns and Inlet/Detector Liners, Adapters, and Inserts
TypeDescriptionTypical UsePart No.
1/4-inch swage,nut1/4-inch packed metal columns5080-8753
stainless steel,front ferrule
pkg, 20 of eachback ferrule
1/8-inch swage,nut1/8-inch packed metal columns5080-8751
stainless steel,front ferrule
pkg, 20 of eachback ferrule
1/4-inch swage,nut1/4-inch packed metal columns5080-8752
brass, pkg,front ferrule
20 of eachback ferrule
1/8-inch swage,nut1/8-inch packed metal columns5080-8750
brass, pkg,front ferrule
20 of eachback ferrule
Vespel,1/4-inch ferruleinlet/detector liners,5080-8774
pkg of 101/4-inch glass packed columns
Vespel,1/8-inch ferrulemetal columns0100-1107
pkg of 10
graphite,1.0-mm ferrulecapillary columns5080-8773
pkg of 10
graphite,0.5-mm ferrulecapillary columns5080-8853
pkg of 10
split/splitless capillaryinlet inserts
silicone1.0 mm O-ringcapillary columns0905-0759
Note: Dimensions given are id’s of the O-ring or ferrule.
16
Columns and Fittings
Liners/adapters and inserts, general
Liners/adapters and inserts, general
A liner/adapter is installed from below, inside the oven; it serves both as
an adapter to mate the particular column to the inlet or detector and to
provide correct internal volume for proper operation.
Inserts are used with inlets only, and, when required, are installed from
above, at the top of the inlet; these are discussed specificallylater in this
section (see Inlet inserts).
In general, the analysis to be performed determines the column to be
used. The column then dictates hardware required for the inlet and
detector (liner, insert, adapter, jet (FID or NPD)).
Note:
C
A correctly designed 1/4•inchpacked glass column requires no liners
since the column ends themselves serve this purpose.
C
The appropriate liner/adapter, and insert if required, must be
installed prior to installing a column.
Tables 1•2and 1•3summarize hardware required for various
combinations of inlets, columns, and detectors.
17
Columns and Fittings
Liners/adapters and inserts, general
Table 1-2.Hardware and Recommended Fittings for Packed Column Installation
Packed Columns
1/8-inch Metal1/4-inch Metal1/4-inch Glass
Recommended1/8-inch1/4-inch1/4-inch swage-
Column Fittingsswage-type nutswage-type nuttype nut and
and ferrules
3
and ferrules
3
graphic ferrule or
silicone O-ring(s)
Packed Column19243-80510
1
19243-80520
1
Inlet Linersoror
19243-80530
1
19243-80540
1
(requires glass insert)(requires glass insert)
FID/NPD
2
19231-80521
1
19231-80530
1
Liners/Adapters
TCD Liners/AdaptersNone19302-80020
ECD Liner/Adapters19301-80530
1
Use 1/4-inch swage-typenut and Vespel or graphite ferrule to install liner/adapter.
2
See Chapter8 for details regardingjet exchange(if necessary).
3
See information later in this chapter regarding proper installation of swage-type fittingson packed
metal columns.
1
NoneNone
1
None
None
19302-80020
(may require
altering the
column)
1
18
Columns and Fittings
Liners/adapters and inserts, general
Table 1-3.Hardware and Recommended Fittings for Capillary Column Installation
Capillary Columns
HP Series 530
¿
320¿m ID200¿m IDMetal/
Glass
RecommendedCapillary columnCapillary columnSame as 320¿mSame as
Column Fittingsnut and 1.0-mmnut and 0.5-mmHP Series
Split/Splitless18740-80220SameSameSame
Capillary Inletwith graphite or
Insert: Splitlesssilicone O-ringSampling
Programmable19245-20580
3
19245-2052019245-2051019245-
On-Column20550
Capillary Inlet
Insert
FID/NPD
2
19244-80550
1
SameSameSame
Liners/Adapters
TCD18740-2095019232-80550
1
SameSame
Liners/Adaptersand
18740-20960
ECD19244-805501,
3
3
19233-80530SameSame
Liners/Adapters
1
Use 1/4-inch swage-typenut (if a nut is notsuppliedas part ofthe adapter) and graphite or Vespel ferrule to
install liner/adapter.
2
0.11-inch jet must beused; see Chapter8 for informationregarding jet exchange (if necessary).
3
Use only if detector is not configured with capillary makeupgas adapter. If makeup adapter is provided,it is
used instead (usually with makeup gas turnedoff).
19
Columns and Fittings
Inlet/detector liners/adapters
Inlet/detector liners/adapters
Interchangeable stainless steel liners/adapters, installed from inside the
oven, are used with the packed column inlet, and with all detectors,
depending upon the column to be installed.
Packed column inlet liners
Figure 1-5
20
Liner
Installed Liner, Packed Column Inlet
Liners for the packed column inlet are available in three sizes: one for
1/8•inchcolumns, one for 1/4•inchcolumns, and one for HP Series 530
capillary columns.
¿
Columns and Fittings
Inlet/detector liners/adapters
In addition, liners for the packed column inlet are available to accept
glass inserts (discussed later) for reduced reactivity, to trap nonvolatile
residues, or for use with an HP Series 530
C
No liner is used with 1/4•inchpacked glass columns. The long leg of
the column fits into the inlet body, replacing the liner. Packing and
glass wool plug must be below the tip of the needle for best results.
C
Metal columns are installed with a liner appropriate for the column
diameter.
C
If necessary, glass columns can be installed using a metal liner
(preferably those accepting a glass insert), but this is not
recommended. There may be problems with dead volume in
connections, and preventing contact of sample with metal surfaces.
¿
capillary column.
21
Columns and Fittings
Inlet/detector liners/adapters
Detector liners/adapters
Figure 1-6
Liner/Adapter
22
Typical Installed Detector Liner/Adapter
Detectors require a liner/adapter to be installed when used with packed
metal columns (either 1/8•or 1/4•inch),and with any type of capillary
column. Normally, no liner is required with 1/4•inchpacked glass
columns, since the leg of the column itself serves as the liner.
For the FID or NPD, the correct detector jet must be installed prior to
installation of the liner. (If jets must be exchanged, see Chapter 8,
Preventive Maintenance.)
Columns and Fittings
Inlet/detector liners/adapters
ECD and TCD adapters
A makeup gas adapter must be installed in the ECD or TCD base to
install a capillary column, and to augment carrier flow through the
column with additional gas flow needed for optimal detector operation.
The adapter must be removed for packed column applications.
In addition, to install an HP Series 530
¿
capillary column in an ECD or
TCD having no capillary makeup gas adapter, the following adapters are
used: Part No. 19244•80550for the ECD, and Part No. 18740•20950and
18740•20960for the TCD.
Finally, to use a 1/4•inchcolumn with the TCD (having a base designed
for 1/8•inchcolumns), a 1/8•to 1/4•inchadapter is required (Part No.
19302•80020). For the ECD (having a base designed for 1/4•inch
columns), to use a 1/8•inchcolumn, a 1/4•to 1/8•inchadapter is required
(Part No. 19301•80530).
23
Columns and Fittings
Inlet/detector liners/adapters
Liner/adapter installation
Figure 1-7
Liner
1/4-inch Ferrule
Liner Retainer Nut
Capillary Column Nut
WARNING
Packed Column Inlet Liner for HP Series 530
Capillary Column Use
Nut and Ferrule Installed on a Liner/Adapter
¿
1-mm Graphite Ferrule
With one exception, liners/adapters are installed in the same manner;
if the liner/adapter has not been used before, a new ferrule must be
installed.
¿
The single exception is the adapter to install an HP Series 530
capillary
column in a TCD without provision for capillary makeup gas (Part No.
18740•20950and 18740•20960). In this case, no ferrule is required to
form a seal with the detector base.
Note: A graphite ferrule is strongly recommended; since metal ferrules
tend to lock permanently onto the liner/adapter, their use may require
replacing the entire liner/adapter, should a permanent leak develop.
Exercise care! The oven, and/or inlet or detector fittings may be hot
enough to cause burns.
24
Note: The liner/adapter must be kept as clean as possible to prevent
introducing contamination into the inlet or detector. Use a clean,
lint•freecloth to remove fingerprints, etc., from the end of the
liner/adapter to be inserted into the inlet or detector base. CH
2
OH
(methanol) may be used as a solvent.
Columns and Fittings
Inlet inserts
1. Assemble a brass nut and graphite ferrule onto the liner/adapter.
2. Insert the liner/adapter straight into the detector base as far as
3. Holding the liner/adapter in this position, tighten the nut finger•tight.
4. Use a wrench to tighten the nut an additional 1/4 turn.
5. Install the column; then heat the oven, inlet, and detector to desired
Inlet inserts
Inserts are used in inlets, and can be installed from the top of the
particular inlet.
Packed column inlet inserts
possible.
operating temperatures and, only if necessary to stop leaks, tighten
fittings further.
Figure 1-8
Flared End
Glass Insert for Packed Column Inlet Liner
Assuming the correct inlet liner is installed, a glass insert is installed as
described on the next page.
25
Columns and Fittings
Inlet inserts
WARNING
Figure 1-9
Exercise care! the oven, and/or inlet, or detector fittings may be hot
enough to cause burns.
Flared End
Insert
26
Installing a Glass Insert in a Packed Column Inlet
1. In handling the insert, avoid contaminating its surface (particularly
its interior).
2. Remove the septum retainer nut and septum.
3. Carefully remove the old insert (if present) by withdrawing it straight
up. A match stick or similar fibrous item may be used as an aid in
lifting the insert from the inlet.
4. Install the new insert by dropping it carefully, straight into the inlet
liner, flared end up.
Columns and Fittings
Inlet inserts
Note: For the liner and insert for an HP Series 530¿capillary column, if
the column is already installed, a new insert may not seat properly in the
liner; the column may prevent it from dropping completely into the liner.
If the insert does not drop completely into the liner, do not force it
(either the liner or the column may shatter); instead, remove the column,
seat the insert, and then replace the column.
5. Replace the septum and septum retainer nut.
Split/splitless or split-only capillary inlet inserts
Figure 1-10
Viton O-ring (Preferred)
Viton O-ring
Split Use Splitless Use
Split/Splitless and Split-Only Capillary Inlet and Inserts
A specific inlet insert is required, depending upon the particular
sampling mode. Specific sampling modes include:
C
Split, for major•componentanalyses
C
Purged splitless, for trace•componentanalyses
27
Columns and Fittings
Inlet inserts
The split insert contains packing material (10% OV•1on 80/100 High
Performance Chromosorb•W),held in place by silanized glass wool plugs,
located immediately above a mixing chamber. This ensures proper
volatilization and homogeneous mixing of the sample prior to its entry
into the column.
WARNING
Caution
Exercise care! The oven, and/or inlet, or detector fittings may be hot
enough to cause burns.
If operating in split mode, carrier gas pressure must be reduced before
opening the inlet. If not done, pressure may blow insert packing out of
the inlet, altering its characteristics. Pressure is reduced at the
backpressure regulator for the inlet.
1. In handling the insert, avoid contaminating its surface (particularly
its interior).
2. Remove the insert retainer nut. The septum retainer nut need not be
3. Using tweezers, forceps, or similar tool, remove any insert already in
place.
4. Inspect the new insert to be installed: For a split mode insert, the end
with the mixing chamber and packing is inserted first into the inlet.
5. Place a graphite or silicone O•ringon the insert, about 2 to 3 mm from
its top end.
6. Install the insert, pressing it straight down, as far as possible, into
the inlet.
Do not add any seal either at the bottom of the inlet or at the bottom of
the insert; to do so will damage the inlet and/or shatter the insert.
7. Replace the insert retainer nut, tightening it to firm finger•tightness
to form a leak•freeseal. Do not overtighten.
29
Columns and Fittings
Jet replacement, FIDs or NPDs
Jet replacement, FIDs or NPDs
Depending upon the column type (packed versus capillary) to be used,
and/or analyses to be performed, exchanging the jet in an FID or NPD
may be necessary. This must be done prior to column installation, and is
particularly important in optimizing FID performance.
Exchanging the jet in either an FID or an NPD is described in Chapter 8,
Preventive Maintenance.
Metal capillary columns
Most metal capillary columns (0.6 to 1.0 mm od) can be connected
directly. Some metal capillaries have a large•diametersleeve soldered on
each end; this must be removed. Use a small triangular file to score the
tubing behind the sleeve; then bend the sleeve back and forth until it
breaks.
30
It is important to have fresh ends of the column, free of burrs, jagged
edges, and/or loose particles of column, stationary phase, and/or material
from a sealing ferrule or O•ring.
Therefore, whenever the column must be cut to provide fresh ends, use
a suitable file to first score the column at the point at which it is to be
broken. This is done normally after installing on the column the column
nut and ferrule (or O•ring)required for installation.
Loading...
+ 214 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.