MULTIGUARD CALL/RESET
POINT S1663P
This call point (S1663P) is designed for use with the
MULTIGUARD 2000 patient nurse call system (MG311000 and
MG311100). In addition to being able to generate a call by
pressing the large red button on the facia it is also possible to
reset any calls by simultaneously pressing the blue button and
the central green button/LED. An emergency recall can also be
generated at any time by simultaneously pressing the yellow
button and the central green button/LED.
The call point has an integral electronic timer which is started
whenever a call is generated. If the call has not been reset within
a predetermined period then a late attendance signal is sent to
the control equipment to re-alert staff at the nurse station. The
integral timer is adjustable over the range 30 seconds to about 2
minutes and will continue to send the signal at the periodic
intervals until the call point has been reset.
Facilities are provided for the connection of auxiliary devices
such as:-
A. an external ceiling mounted Pull-Cord with re-assurance
LED S/1600.
B. a hand held pillow push. S/1631 or other open circuit device
5
87
15
The call point is
designed to fit into
a standard single
gang pattress box
or flush mount
accessory box.
A box of 25mm or
greater depth is
recommended.
Eg. S/1680 or
TE55
Dimensions
(mm)
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
RESET RE-CALL
Late Attend
Timer
87
CALL
Call In
Reset In
LED
+ve
-ve
+ve
C. an overdoor light unit. S/1636
Details of these devices and their connection to the call point can be found overleaf.
Note: The MG311XXX controllers is a non-latching device. A pull cord, pillow push or other non-
latching call devices cannot be connected directly to the MULTIGUARD MG311XXX controllers
without an S/1663 call point. The call point provides an electronic latching circuit.
To additional
call points
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend
Timer
Pillow push or Pull Cord
Input. Normally Open, Close
to Call
Output for Re-assurance LED
on Pull Cord or Overdoor Light
See overleaf for connection
details
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend
Timer
A 10 core cable can be used for
each six zones. It is possible to
connect a number of call points
into one zone of the annunciator
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend
Timer
+ve
-ve
Reset In
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
To additional
call points
Call In
+ve
-ve
LED
+ve
Late Attend
Timer
MG311000
& MG311100
Dwg No:60010:7 Iss 2 Page 1 Jul 09
1 2 3 4 5 6 + + 1 2 3 4 + -
INPUTS
FUNCTION
INPUTS
Link F2 to +ve to disable silence function
S/1663
Complies with Council
Directive 89/336/EEC
Hoyles Electronic Developments Ltd
T. 01744 886600 E. sales@hoyles.com
F. 01744 886607 W. www.hoyles.com

Call Point with S/1600
Ceiling Pull Cord
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
LED
+ve
-ve
S/1600
N.O.
COM
N.C.
To further pull
cords if required
S/1663
Call Point
Late Attend
Timer
Up to three ceiling pull cords with re-assurance
LEDs can be connected to an S/1663 call
point/reset unit. Each re-assurance LED draws
approximately 15ma. The LED on the pull cord will
flash in sympathy with the green LED on the S/1663 call point. If more than
one is fitted then both the switch contacts and the LEDs should be connected
in parallel.
The +ve feed for the LED and the +ve feed for the switch COM are common
therefore a three core cable could be used if this would be more convenient.
Call Point with S/1631 Pillow Push
2000
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
If a hand held pillow push device is required the call point/reset unit should be
S/1663
Call Point
Late Attend
Timer
installed within 1 metre of the bedhead and an S/1631 Pillow push unit connected
to the Call In terminals as shown. The pillow push unit is normally supplied with 2
metres of two core cable, this can be extended on site using a wall mounted
junction box. It is best to avoid excessively long trailing cable. The edge of the call
point plate is notched out as a cable entry for the pillow push.
Dimensions
(mm)
90
33
2000
90
45
Call Point with S/1636 Overdoor Lights
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663
Call Point
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend
Timer
Call In
+ve
Reset In
-ve
LED
+ve
SW
+ve
O/P
S/1636
S/1636s used as
overdoor lights
SW
+ve
O/P
S/1636
Overdoor or end of corridor indicators can be incorporated on the system if
required. A blocking diode is included in each overdoor light to enable a
S/1663
Call Point
Late Attend
Timer
group of overdoor lights to signal onto to an end of corridor light. If a voltage
is applied to unit A then both A & C will illuminate, if applied to unit B, then B
& C will illuminate. In this way the system can be configured as a follow the
lights type of system to enable patrolling staff to respond without returning
to the panel to verify if any alarm conditions have occurred. Overdoor lights
Dwg No:60010:7 Iss2 Page 2 Jul 09
are fabricated on single gang plastic plates for flush or surface mounting.
A
B
SW
+ve
O/P
S/1636
C
S/1636 used as
end of corridor
lights
87
Dimensions (mm)
87
25