Hoyles Electronic Call Point User Manual

Page 1
MULTIGUARD CALL/RESET POINT S1663P
This call point (S1663P) is designed for use with the MULTIGUARD 2000 patient nurse call system (MG311000 and MG311100). In addition to being able to generate a call by pressing the large red button on the facia it is also possible to reset any calls by simultaneously pressing the blue button and the central green button/LED. An emergency recall can also be generated at any time by simultaneously pressing the yellow button and the central green button/LED.
The call point has an integral electronic timer which is started whenever a call is generated. If the call has not been reset within a predetermined period then a late attendance signal is sent to the control equipment to re-alert staff at the nurse station. The integral timer is adjustable over the range 30 seconds to about 2 minutes and will continue to send the signal at the periodic intervals until the call point has been reset.
Facilities are provided for the connection of auxiliary devices such as:-
A. an external ceiling mounted Pull-Cord with re-assurance LED S/1600.
B. a hand held pillow push. S/1631 or other open circuit device
5
87
15
The call point is designed to fit into a standard single gang pattress box or flush mount accessory box. A box of 25mm or greater depth is recommended. Eg. S/1680 or TE55
Dimensions (mm)
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
RESET RE-CALL
Late Attend Timer
87
CALL
Call In
Reset In
LED
+ve
-ve
+ve
C. an overdoor light unit. S/1636 Details of these devices and their connection to the call point can be found overleaf. Note: The MG311XXX controllers is a non-latching device. A pull cord, pillow push or other non-
latching call devices cannot be connected directly to the MULTIGUARD MG311XXX controllers without an S/1663 call point. The call point provides an electronic latching circuit.
To additional
call points
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend Timer
Pillow push or Pull Cord Input. Normally Open, Close to Call
Output for Re-assurance LED on Pull Cord or Overdoor Light
See overleaf for connection details
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend Timer
A 10 core cable can be used for each six zones. It is possible to connect a number of call points into one zone of the annunciator
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend Timer
+ve
-ve
Reset In
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
To additional
call points
Call In
+ve
-ve
LED
+ve
Late Attend Timer
MG311000 & MG311100
Dwg No:60010:7 Iss 2 Page 1 Jul 09
1 2 3 4 5 6 + + 1 2 3 4 + -
INPUTS
FUNCTION
INPUTS
Link F2 to +ve to disable silence function
S/1663 Complies with Council Directive 89/336/EEC
Hoyles Electronic Developments Ltd
T. 01744 886600 E. sales@hoyles.com F. 01744 886607 W. www.hoyles.com
Page 2
Call Point with S/1600 Ceiling Pull Cord
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
LED
+ve
-ve
S/1600
N.O. COM N.C.
To further pull cords if required
S/1663 Call Point
Late Attend Timer
Up to three ceiling pull cords with re-assurance LEDs can be connected to an S/1663 call point/reset unit. Each re-assurance LED draws
approximately 15ma. The LED on the pull cord will flash in sympathy with the green LED on the S/1663 call point. If more than one is fitted then both the switch contacts and the LEDs should be connected in parallel.
The +ve feed for the LED and the +ve feed for the switch COM are common therefore a three core cable could be used if this would be more convenient.
Call Point with S/1631 Pillow Push
2000
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
If a hand held pillow push device is required the call point/reset unit should be
S/1663 Call Point
Late Attend Timer
installed within 1 metre of the bedhead and an S/1631 Pillow push unit connected to the Call In terminals as shown. The pillow push unit is normally supplied with 2 metres of two core cable, this can be extended on site using a wall mounted junction box. It is best to avoid excessively long trailing cable. The edge of the call point plate is notched out as a cable entry for the pillow push.
Dimensions (mm)
90
33
2000
90
45
Call Point with S/1636 Overdoor Lights
+ve
-ve
E/Recall
L/Attend
CALL OUT
S/1663 Call Point
+ve
-ve
E/Recall
L/Attend
CALL OUT
Call In
+ve
Reset In
-ve
LED
+ve
Late Attend Timer
Call In
+ve
Reset In
-ve
LED
+ve
SW +ve O/P
S/1636
S/1636s used as overdoor lights
SW +ve O/P
S/1636
Overdoor or end of corridor indicators can be incorporated on the system if required. A blocking diode is included in each overdoor light to enable a
S/1663 Call Point
Late Attend Timer
group of overdoor lights to signal onto to an end of corridor light. If a voltage is applied to unit A then both A & C will illuminate, if applied to unit B, then B & C will illuminate. In this way the system can be configured as a follow the lights type of system to enable patrolling staff to respond without returning to the panel to verify if any alarm conditions have occurred. Overdoor lights
Dwg No:60010:7 Iss2 Page 2 Jul 09
are fabricated on single gang plastic plates for flush or surface mounting.
A
B
SW +ve O/P
S/1636
C
S/1636 used as end of corridor lights
87
Dimensions (mm)