Hottinger Baldwin Messtechnik Z6FD1, Z6FC3, Z6FC4, Z6FC6 User Manual

Z6...
Load cell
Dimension (in mm; 1mm = 0,03937 inches)
Z6; Max. capacities 5kg...500kg
123
45
18
82
Fixing
10
Load introd.
Max. capacities 5kg...1t
Z6; Max. capacity 1t
40
60
Special features
Welded on metal bellow
Load cells and mounting aids
entirely made from stainless material
Six-wire circuit
Optimized for parallel
connection with corner-pre­adjustment
Meets today EMC/ESD requirements according to EN 45501
Available option: Explosion p roof version EEx ib IIC T4
Fixing
40
210
133±0,1
22±0,1
Load introd.
25
42
20
Cable: 3m, 5.4mm, 6 cores screened, screen connected to housing
31
9,9
AB
5...200kg 8.2 8.2 500kg 10.5 11.1
20
Cable: 3m, 5.4mm,
21
A
Metal bellow
B
6 cores, screened, screen connected to housing
34
15
13
70
Metal bellow
84
44
11, 1
+0,1
-- 0 , 1
38
Data sheet D.Z6.0e
Technical data
Typ e Z6FD1 Z6FC3 Z6FC4 Z6FC6
Accuracy class according to OIML R 60 Maximal numbers of load cell verif. intervals (n
Max. capacity (E
) kg 5; 10; 20; 50;
max
LC
)
D1
1000
100; 200;
C3
3000
10; 20; 50;
100; 200;
C4
4000
20; 50; 100;
200;
6000
50;100;200;
500 500 500
t 1 1
Minimum load cell verification interval (v
) %ofC
min
n
0.0360 0.0090 0.0066
Sensitivity (Cn) mV/V 2
Tolerance on sensitivity % <+1;--0.1 <+0.05
Temperature effect on sensitivity (TKC)
Temperature effect on zero balance (TK0) %of
Hysteresis error (dhy)
1)
(d
)
lin
1)
1)
%of
C
/10K
n
<+0.0500 <+0.0080 <+0.0070 <+0.0040
<+0.0500 <+0.0125 <+0.0093 <+0.0093
C
/10K
n
% <+0.0500 <+0.0170 <+0.0130 <+0.0080
% <+0.0500 <+0.0180 <+0.0150 <+0.0110
Creep (dDR)in30min. % <+0.0490 <+0.0166 <+0.0125 <+0.0083
Input resistance (RLC)(black--blue) 350 -- 480
Output resistance (R0)(red--white) 356 + 0.2 356 + 0.12
Reference excitation voltage (U
) V 5
ref
Nominal range of excitation voltage (BU) V 0.5...12
Insulation resistance (Ris) G >5
Nominal temperature range (BT)
Service temperature range (Btu)
Storage temperature range (Btl)
Safe load limit (EL) %ofE
Breaking load (Ed) %ofE
oC[o
F] --10...+40 [15...+105]
oC[o
F] --30...+70 [--20...+160]
oC[o
F] --50...+85 [--60...+185]
max
max
150
300
Max. capacity kg 5 10 20 50 100 200 500 1000
Permissible dynamic load (F
(vibration amplitude according to DIN 50100)
Deflection at max. load, (s
)
srel
) approx. (±15%) mm 0.24 0.3 0.29 0.27 0.31 0.39 0.6 0.55
nom
%ofE
100 100 100 100 100 100 70 100
max
Weight (G), approx. kg 0.5 0.5 0.5 0.5 0.5 0.5 0.5 2.3
Protection class (IP) acc. to EN60529 (IEC529) IP 67 (more rigorous test conditions: 100hat 1m Water column)
Material
Measuring body Metal below Cable entrance Cable sheath
stainless steel
stainless steel
stainless steel / Neoprene
PVC
Optionally Explosion proof version (EEx ib IIC T4) PTB-no. EX-90.C.2094
1)
The data for deviation of linearity, hysteresis and temperature effect on sensitivity are typical values. The sum of these data meets the requirementsaccordingtoOIMLR60.
Wiring code
C6
(gray) Sense (--)
(black) Excitation (--)
(white) Signal (+)
(blue) Excitation (+)
(green) Sense (+)
(red) Signal (--)
(yellow) Shield, connected with housing
Loading...