Hottinger Baldwin Messtechnik RTN User Manual

RTN...
Load cells
Special features
-- Low construction
-- Legal for trade according to OIML
-- Stainless steel
-- Protection class IP 68
-- High output signal of 2.85 mV/V
-- Input resistance 4350
-- Explosion proof versions
according to ATEX (Option)
Data Sheet
Dimensions (in mm; 1 mm= 0.03937 inches)
RTN... / 4,7 t
AB
K
(1)
L
WÄGEZELLE LOAD CELL
X
~15
RTN... 1t 2.2 t 4.7 t 10 t 15 t 22 t 33 t 47 t 68 t 100 t 150 t 220 t 330 t 470 t
A 49 49 49 74 75 75 95 130 130 150 150 225 225 270B 20 20 20 30 30 30 40 60 60 70 70 100 100 120C 60 60 60 75 75 75 95 130 130 150 150 225 225 270
H 43 43 43 50 50 50 65 75 85 90 100 130 145 170
J -- -- -- 7 7 7 7 7 7 7 7 10 10 10
K 7.5 7.5 7.5 6.5 6.5 6.5 10 14 14 16 16 24 24 28 L
C
5m 5m 5m 5m 5m 12 m 15 m 12 m 12 m 12 m 5m 5m 5m 5m
H
View X
depth of thread 7 mm
53
RTN... / 10 t ... 470 t
WÄGEZELLE
(1)
L
~15
LOAD CELL
AB
25
C
F8
K
H
J
L = Cable length
)
1
(
B0990-5.0 en
Specifications
n
%of
C
n
V
Typ e
Nominal (rated) load (E Accuracy class 0.05 C3
Maximum number of load cell intervals (nLC) -- 3000
Minimum LC verification interval (v
Sensitivity (Cn) mV/V
Temperature effect on sensitivity (TKC)
Temperature effect on zero balance (TK0)
Hysteresis error (dhy) 0.05 0.02 Non-linearity (d
Creep (dcr) over 30 min.
Input resistance (RLC)
Output resistance (R0)
Reference excitation voltage (U Nominal range of excitation voltage (BU) Carrier frequency of the excitation voltage Hz < 600 Insulation resistance (Ris) G >20 Nominal temperature range (BT) --10 ... +40 [+14 ... +104] Service temperature range (Btu)
S Load cell RTN S Accessory: VEN
Storage temperature range (Btl) --50 ... +85 [--58 ... +185]
lin
VPN
)
) 1 t ... 470 t
max
) %ofE
min
)
ref
%ofC
/10K
%ofC
°C[°F]
max
n
n
--30 ... +80 [--22 ... +176] (Option: up to +110 °C / +230 °F)
RTN 0,05 RTN C3
(OIML R60)
--
2.85 0.1 %
0.05 0.0080.03 0.007
0.05 0.02
0.03 0.017
4450 100
4010 2 4010 0.5
5
5 ... 30 (max. 60)
--30 ... +80 [--22 ... +176]
--30 ... +110 [--22 ... +230]
(= E
0.005
max
/ 20000)
1)
1)
1)
Nominal (rated) load (E Weight (G), approx. kg 0.6 0.6 0.7 1.2 1.3 1.3 2.1 4.3 4.8 7.0 8.6 22 29 50
Safe load limit (EL)
Breaking load (Ed)
Permissiblestaticsideload(Lq) 0.5 (E
Permissible dynamic load (F (vibration amplitude according to DIN 50100)
Deflection at E Protection class to EN60529 (IEC529) IP 68 Material: Measuring element
1)
The data for Non--linearity (d meets the requirements according to OIML R60.
max(snom
Cable fitting Cable--sheath
Wiring code RTN...
) t 1 2.2 4.7 10 15 22 33 47 68 100 150 220 330 470
max
1.7 3.8 8 17 25 38 56 80 115 170 250 380 500 700
t
4 9 19 40 60 88 130 190 270 400 600 770 1100 1500
-- 0 . 8 Lz), but no higher than L
max
E
= Nominal (rated) load; Lz= load in measuring direction
)
srel
), approx. mm 0.13 0.12 0.12 0.17 0.18 0.21 0.25 0.33 0.35 0.45 0.57 0.67 0.85 1.00
), Hysteresis error (dhy) and Temperature effect on sensitivity (TKC) are typical values. The sum of these data
lin
% of
E
max
max
Brass, nickel plated / Sealing: CR/NBR
Special silikon, RAL 7000 (grey), 6.5 mm
(black)
(red)
(blue)
(white)
(green--yellow)
70
Stainless steel
Excitation (+)
Signal (+)
Excitation (--)
Signal (--)
Shield
qmax
=0.3E
max
Available Options:
Explosion--proof versions according to ATEX: II 2 G EEx ia IIC T4 resp. T6 (Zone 1) * II 3 G EEx nA II T6 (Zone 2) II 2 D IP68 T80 °C (Zone 21) *
)
II 3 D IP68 T80 °C (Zone 22 for non--conductive dust)
)
*
with EC--Type Examination Certificate
Service temperature range up to +110 °C (not possible in connection with ATEX and/ or Elastomer mounts VEN)
2
)
B0990-5.0 enHBM
Loading...
+ 2 hidden pages