Hottinger Baldwin Messtechnik C16A User Manual

C16A../20t ... 100t
Self--restoring rockerpin load cell
Max. capacities: 20t ... 100t
Special features
Self--restoring function
Simple to install
laser welded, IP68
Legal for trade verification
d
d up to 5000d (OIML R60)
dd
d
d up to 10 000d (NTEP class III LM)
dd
Optimized for parallel connection by corner pre-adjustment
Meets EMC / ESD requirements according to EN 45 501
Available Options:
d
d Explosion proof versions
dd
G EEx ib IIC T4 G EEx d IIC T6
d
d Lightning protection
dd
Wiring code
Wiring code (6--wire circuit):
D.C16A.1e -- 284.00--1050.4 -- Version d -- 19.03.2001
(grey) (black)
(white)
(blue) (green)
(red)
(--)
Sense (--) Excitation (--)
Signal (+)
Excitation (+) Sense (+)
Signal (--)
Shield / wire strand connected with housing
Data sheet D.C16A.1e
Technical Data
%of
C
n
%
Type C16A D1 C16A C3
Maximum capactity (E
Accuracy class accord. to OIML R60 Max. numbers of load cell verification intervals (n
LC
)
Min. load cell verification interval (v
Min. scale verification interval (e accordingtoEN45501
[... LC = max. Number of load cells]
Sensitivity (Cn) mV/V 2
Sensitivity tolerance
Temperature effect on sensitivity (TK
)
C
Temperature effect on zero signal (TK
)
0
Hysteresis error (dhy)
Non-Linearity (d
2)
)
lin
Creep (dcr), 30 min.
Input resistance (RLC) (black--blue)
Output resistance (R0)1)(red--white)
Reference excitation voltage(U
Nominal range of excitation voltage (B
U
Insulation resistance (Ris) G >5
Nominal range of ambient temperature (B
)
T
Service temperature range (Btu)
Storage temperature range (Btl) --50 ... +85 [-60...185]
Limit load (EL) 150
Breaking load (Ed)
Permissible dynamic load (F (vibration amplitude according to DIN 50100)
Deflection at E
max(snom
Weight (G) with cable, approx. kg 2.1 2.3 2.9 3.7 8 2.1 2.3 2.9 3.7 8
Protection class according to EN60529 (IEC529)
Material: Meas. body + housing
Cable fitting Sealing Cable-sheath
1)
Smaller tolerances on Sensitivity and Output resistanc e are not necessary. Throughout corner pre-adjustment the Sensitivity and Output resistance are coordinated, so that the indicated value of the scale is within permissible limits when off--center load is applied.
2)
The data for Non--Linearity (d sum of these data meets the requirements for p
) 20t 30t 40t 60t 100t 20t 30t 40t 60t 100t
max
D1 (0.0330%)
1000 (10000 NTEP III LM)
%
)
min
of E
max
)
min
kg -- -- -- -- --
1)
2)
% ≤±0.5000
%ofC
n
/ 10K
2)
%ofC
n
0.0200
(0.0068 NTEP)
5
[6 LC]
10
[10 LC]
≤±0.0250
2)
≤±0.0285 ≤±0.0140
≤±0.0330
≤±0.0300
2)
2)
C3 (0.0170%)
3000
0.0100 0.0083 0.0167
10
10
20
[4 LC]
20
[10 LC]
2)
1)
10
[10 LC]
≤±0.0080
[6 LC]
[10 LC]
≤±
0.0116≤±0.0234
≤±0.0170
≤±0.0180
2)
2)
≤±0.0330 ≤±0.0167
) 5
ref
V
)
700 ±20
706 ±3.5
0.5 ... 12
1)
--10 ... +40 [15...105]
°C[°F]
--30 ... +70 [-20...160]
>350
srel
)
of E
max
70
), approx. mm 0.65 0.75 0.85 1.22 1.57 0.65 0.75 0.85 1.22 1.57
IP68 (test conditions 100h at 1m water column)
IP69K (water at high pressure, steam jet cleaning)
stainless steel
stainless steel (Maximum capacity 100t: nickel--plated brass)
neoprene
thermoplastic elastomer
), Hysteresis error (dhy) and Temperature effect on sensitivity (TKC) are typical values. The
lin
= 0.7 according to OIML R60 respectively NTEP.
LC
50
[8 LC]
D.C16A.1e -- 284.00--1050.4 -- Version d -- 19.03.2001
Loading...