IRROLIS SENSE Tx
SOIL AND SUBSTRATE TENSION MEASUREMENT
derived from a classical tensiometer technology that allows readings of soil and substrate tension for optimized plant growing, experiment monitoring and environmental follow-up.
LCD Display
The tension measurement is displayed on the LCD Display. The LCD Display is always ON when the device is powered. If the device is equipped with a temperature sensor (optional), both tension and temperature measurements are displayed alternatively.
Low Battery Indicator
When the battery level is getting low, an ar row is displayed next to the Low Battery pictogram.
Units
The IRROLIS SENSE Tx displays soil or substrate tension in kilopascal (kPa) or centibar (cbar), two equivalent measurement units. The temperature (optional) is displayed in ºC (Fahrenheit in USA).
Analog Output Connector
The IRROLIS SENSE Tx is equipped with an Analog Output Connector. This feature enables connection with most analog data loggers and control systems.
Dust Cap
The Analog Output Connector and Battery Pack Connector are equipped with Dust Caps to protect against dust and debris accumulation. Leave them in place when connectors are not in use.
INSERTING AND CHANGING BATTERIES
1. Separate the cover from the housing of the Battery Pack.
2. Remove the used batteries (if any).
3. Insert two (2) new AA batteries.
4. Verify that the gasket is well positioned.
5. Put in place the cover on top of the IRROLIS SENSE Tx.
6. Screw back the four (4) screws.
Before using or preparing your IRROLIS SENSE Tx, check that you
have all parts at hand and in good working order.
Handle the IRROLIS SENSE Tx with care at all times.
The Cer amic Tip MUST be in contact with soil or substrate in order to generate an accurate reading.
Do NOT put the Ceramic Tip in contact with oil, fats or any contaminant that could affect its wetability.
Never totally immerse the IRROLIS SENSE Tx.
The IRROLIS SENSE Tx is sealed and cannot be opened except for the Water Inlet and the Ceramic Tip which can be replaced if broken.
Protect the Ceramic Tip at all times against shocks.
The IRROLIS SENSE Tx should be protected at all times against frost. Under no circumstances should the storage temperature be bellow 5 ºC (40 Fahrenheit).
Avoid exposure to magnetic fields.
Do not expose to extreme temperatures.
When powered-on, the IRROLIS SENSE Tx goes through an initialization routine :
1. Battery voltage displayed.
2. Tension displayed (the IRROLIS SENSE Tx is now ready).
INSTALLING THE ANTENNA
FILLING THE IRROLIS SENSE Tx
1. Unscrew the cap of the Filling Bottle
2. Pour drinking water in the Filling Bottle and screw the cap back.
3. Using the Hexagonal Key, remove completely the Water Set Screw.
4. Squeeze the Filling Bottle to push water in the IRROLIS SENSE Tx directly.
IMPORTANT
1. Screw the Antenna on the Antenna Connector. Make sure the Antenna is tightly screwed. This will ensure good transmission.
The user manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Operation is subject to the following two conditions: (1) This device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. To reduce potential radio interference to other users, the antenna type and its gain should be so cho sen that the equivalent isotropically radiated power (EIRP) is not more than that required for successful communication. This equipment has been tested and found to comply with the limits for Class A digital device, pursuant to Part 15 of FCC Rules. These limits are designed to provide re asonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio freq uency and, if not installed and used in accordance with the instruction manual, m ay cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful inte rference in which case the user will be required to correct the interference at its own expense. This device has been designed to operate with an antenna having a maximum gain of 2.2 dB. Antenna having a higher gain is strictly prohibited per regulations of Industry Canada. The required antenna impedance is 50 ohms. The ¼ wave antenna provided with this device has a gain of 2.2 dB and is manufactured by Antenna Factor (P/N ANT-418-CW-QW).
5. Using the Hexagonal Key, screw the Water Set Screw back, making sure it is firmly tightened.
6. Your IRROLIS SENSE Tx is now ready to use.
STORING THE IRROLIS SENSE Tx WHILE FILLED
If you do not install the IRROLIS SENSE Tx after filling it, it is recommended to leave it with the Ceramic Tip in contact with free water. A convenient tip is to place the IRROLIS SENSE Tx in a bucket containing two (2) inches of water until it is installed. Otherwise, the Ceramic Tip will dry and the IRROLIS SENSE Tx will eventually empty.
The plants (growth stage, size, species, etc.) and the ambient conditions, as well as the characteristics of the soil or substrate, are variable parameters. Hence, soil water tension (moisture) varies from one location to the next. Here is how to choose a location for the IRROLIS SENSE Tx and determine the number of IRROLIS SENSE Tx needed in a given situation.
DEFINING IRRIGATION ZONES
If your irrigation system is installed, this exercise must have been completed during the design stage of the system. Normally, an irrigation valve controls the watering for a specific zone. Watering in this zone is relatively uniform; henc e the plants it contains must normally have similar water needs. If you water manually, define the zones where plants have similar water needs. Install at least one (1) IRROLIS SENSE Tx per irrigation zone. Ideally, five (5) IRROLIS SENSE Tx are needed to manage irrigation and obtain an average representation of the water needs of plants within a given irrigation zone. However, three (3) IRROLIS SENSE Tx are often en ough to ensure proper management within a zone.
Autonomous Mode
The tension measurement can be read directly on the LCD Display. The display is refreshed on a random average of five (5) minutes for the Tx-80 model, and every two (2) minutes for the Tx-07 model.
Analog Mode
The IRROLIS SENSE Tx can be connected to an irrigation control system or a data logging system by a wire linked to the Analog Connector. It is necessary to supply power to the IRROLIS SENSE Tx through the analog connector to get a voltage output. To transform the output voltage into tension, use the following equations where “y” is tension in kPa or cbar and “x” is voltage in Volt: To-07 and To-07-WL : ( y) = 2,4069 (x) - 0,2537 To-35 and To-35-WL : (y) = 11,9750 (x) - 1,1032
Wireless Mode
The tension and temperature (optional) measurements, as well as the battery level and the wireless signal quality, are transmitted to the IRROLIS COM WR (Hortau’s wireless receiver, P/N 6010). The Irricom WR translates the wireless signal into a RS-232 signal that can be used by your PC computer using the Irrolis Software by Hortau (P/N 5001). For installation and use, please refer to the User Guide of the IRROLIS COM WR provided with it.
SOIL TENSION IS A VARIABLE PARAMETER BASED ON, AMONG
OTHERS, CLIMATIC CONDITIONS, AND SOIL CHARACTERISTICS.
THE IRROLIS SENSE Tx PERFORMS A MEASUREMENT FROM THE
LOCATION WHERE IT HAS BEEN INSTALLED. THIS IN NO WAY
GUARANTEES THAT THE TENSION READING WOULD BE THE
SAME AT ANOTHER LOCATION, EVEN IF IT WERE CLOSE TO THE
INSTRUMENT. THE IRROLIS SENSE TX CANNOT IN ANY CASE
INSERTING THE IRROLIS SENSE Tx INTO THE SOIL OR SUBSTRATE
1. Determine depth of penetration
For plants growing in a pot or in a grow bag, the penetration depth must be equal to half the height of the pot or bag. As for fie ld-grown plants, the penetration depth must be equal to the average rooting depth of the plants.
When not in use, the IRROLIS SENSE Tx should be stored properly. This will prevent some damage and extend the shelf life of the device.
STORING THE IRROLIS SENSE Tx
1. Remove the device from the soil or substrate.
2. Wash the aluminium casing and the Ceramic Tip with water.
3. Remove the batteries from the IRROLIS SENSE Tx following the
4. Release the pr essure from the sensor by removing the Water Set
5. Store the IRROLIS SENSE Tx in a d ry place. Storage
THE IRROLIS SENSE TX SHOULD BE PROTECTED AGAINST FROST
REPLACE PERSONAL JUDGEMENT.
THE NUMBER OF IRROLIS SENSE Tx PER IRRIGATION ZONE IS GIVEN FOR INDICATION ONLY. EVERY SITUATION INVOLVES
PARTICULAR CIRCUMSTANCES THAT CANNOT BE ESTABLISHED
IN THIS MANUAL. THESE INDICATIONS CANNOT IN ANY WAY
REPLACE PERSONAL JUDGEMENT.
INSERTING AND CHANGING THE BATTERIES instructions of section 5 - GETTING STARTED of this manual.
Screw using the Hexagonal Key.
temperature must stay between 5 and 30 ºC (40 and 85 Fahrenheit)
AT ALL TIMES. UNDER NO CIRCUMSTANCES SHOULD THE
STORAGE TEMPERATURE BE BELOW 5 ºC (40 FAHRENHEIT).
2. Preparing the insertion hole
Using a 25 mm ( 1 in.) diameter rod, make an insertion hole in the soil or substrate where the IRROLIS SE NSE Tx will be installed. Once inserted in the soil, the entire surface of the Ceramic Tip must remain in contact with the soil. Therefore, do not insert the rod at a depth exceeding the penetration depth of the IRROLIS SE NSE Tx to allow the Ceramic Tip to remain in contact with the soil at the bottom of the insertion hole. For deep insertion it might be difficult to insert a rod in the soil. To make the insertion easier, pour water at the surface prior making the hole, and pour some more water in the ho le as you go deeper. This will make the soil easier to dig. A 25 m m (1 in.) power­driven drill will make deep installation faster and easier. The blue line on the metal tube of the IRROLIS SENSE Tx indicates the maximum insertion depth, this in order to leave the Air Set Screw above soil.
Displaying
1. Normal while the Ceramic Tip is in the water for a
H20 while the Ceramic Tip is in the water
No Display 1. Verify that the batteries are not dead by replacing
Worn or broken Ceramic Tip
Displaying H2O constantly
long time.
2. Insert the Ceramic Tip into the soil and wait 15 minutes.
them with new ones.
1. The material us ed for the end piece is brittle. Minute splinters and scratches may appear at its surface with extended use. These do not hinder the good operation of the instrument.
2. If the end piece is broken, order a replacement from your local distributor
1. The IRROLIS SENSE Tx is empty and needs to be refilled in order to work properly. Just refill it with water following the FILLING THE IRROLIS SENSE Tx instructions of section 5 - GETTING STARTED of this manual
Or
2. The IRROLIS SENSE Tx is in contact with the water table. Wait for the soil to dry.
3. Inserting the IRROLIS SENSE Tx Insert the IRROLIS SENSE Tx, the Ceramic Tip first, by pushing lightly on the Hortimetre Head. Do not exert undue pressure on the Hortimetre Head. If it does not glide in easily, widen the inser tion hole by using a 30 mm (1 1/4 in.) diameter metal rod. Try once more. If it is still not possible to push the IRROLIS SENSE Tx into the insertion hole, dig another hole at a different place. While inserting the IRROLIS SENSE Tx, be careful not to unscrew the Ceramic Tip by rotating the IRROLIS SENSE Tx counter clockwise. Once the installation is done, replace soil o r subs trate around the Stainless Steel Tubing to avoid water infiltration along the tube.
4. Wetting the soil
Once installed, it is recommended to pour water at the soil or substrate surface, around the IRROLIS SENSE Tx, to ensure adequate contact between the Ceramic Tip and the soil or substrate.
5. Operation / Maintenance
The IRROLIS SENSE Tx is designed to redu ce the maintenance to a minimum. However, if the soil tension g ets high low or after long use periods, the IRROLIS SENSE Tx might need to be refilled. In such cases, the IRROLIS SENSE Tx will display the value H2O. This display indicates that a refilling is required.
Power Supply
RF Frequency
Tension Range
Analog Data Output
Analog wires color code
2.2 to 3.3 Volt DC
418 MHz
0.0 – 7.0 kPa (cbar): Models Tx-07
0.0 – 80.0 kPa (cbar): Models Tx-80
0.0 – 3.3 Volt DC
Red wire: Supply (+) Black wire (twisted with red wire): Ground (-) White wire: Signal (+) Black wire (twisted with white wire): Signal (-)