BEFORE USING SCALE
Thank you for purchasing the Hornady® GS-Series digital
scale. Please read all operating instructions carefully before
use. This electronic scale is a precision instrument. With
normal care and proper treatment, it will provide years of
reliable service.
Never load the scale with more than the maximum capacity.
Overloading will permanently damage it! Avoid any exposure
to extreme heat or cold. Your scale works better when
operated at normal room temperature. Always allow the scale
to acclimate to normal ro om temperature for at least one
hour before use.
Give your scale suffi cient warm up time; usually 30-60
seconds before calibration to give the internal components a
chance to stabilize.
Keep your scale in a clean environment. Dust, dirt, moisture,
vibration, air currents and a close distance to other
electronic equipment can all cause an adverse effect on the
reliability and accuracy of your scale.
Handle with care. Gently apply all items to be weighed onto
tray top. Although this scale is designed to be quite durable,
try to avoid rough treatment as this may permanently
damage the internal sensor.
Avoid shaking, dropping or otherwise shocking the scale.
This is a precision instrument and must be handled with
extreme care.
Only operate the scale on a stable, vibration free surface.

OPERATION
Place Scale on a horizontal fl at surface.
Press [ON/ OFF] key.1.
Wait until "0.0" is displayed.2.
Put the object(s) on the weighing platform.3.
Using the [MODE] key, you can switch bet ween4.
the weighing units.
TARE WEIGHING
Turn on scale as described above.1.
Place the “tare item” on the platform.2.
Press [ TARE] and wait until "0.0" is displayed.3.
Add the “net-weight-item.”4.

CALIBRATION
Calibration is NOT r equired before fi rst use. Only calibrate if
scale is not weighing properly. Make sure you use the correct
calibration weight for your scale or the calibration will not be
accurate - GS-1500: 100g calibration weight
Calibration Procedures for GS-1500
Turn on the scale by pressing [ON /OFF] key.1.
Wait until the LCD displays "0.00" and then pr ess 2.
and HOLD the [ MODE ] key for 3 seconds, the LCD will
display "CAL" then release the [MODE] key.
Press the [ MODE] key again, the LCD will display 3.
"CAL" and then "100.00"
Add a 100g calibration weight, the display will show 4.
"PASS" and return to normal weighing mode.
Put away the calibration weight and press [ON /OFF] 5.
to turn of the scale and the calibration is fi nished!

TROUBLESHOOTING
The primary r easons for inaccuracy or malfunction are low
batteries, incorrect calibration, overload or operating on an
unstable surface. Please ke ep this in mind and maintain
and operate your scale properly. This scale is a pre cise
instrument and must be handled with the utmost care
and caution.
[Display] Elimination of errors
Nothing, Change the batteries.
Battery-Symbol,
[lo] o r [88 888]
[OUE2] Recalibrate the scale
[EEEEE] You are overloading
[LLLLL] The delicate weighing load cells
[UNST] Use the scale in a more
as shown above.
the scale. Remove the
excessive load immediately!
have been damaged by misusing
the scale (i.e. it was shocked,
dropped or overloaded) .
You can try recalibrating the
scale. If the load cell has not
been damaged too much it will
work again after recalibration.
stable position.

SPECIFICATIONS
Model GS -1500
Capacity 100g
3.528oz
500ct
1543 gn
Resolution 0.01g
0.001oz
0.05ct
0.1gn
Calibration Weight 100g
Plat form Dimensions 2.5 x 3"
(inches)
Scale Dimensions 3 x 5 x 0.75"
(inches)
Power 2 x AAA (included)
P.O. Box 1848 • Grand Island, NE 68802-1848
(308) 382-1390 • www.hornady.com