Honeywell EI2 Operating Instructions For Operators And Installers

Page 1
GB
F
NL
I
SK
1000106001-002-06
03250952
D GB
I
www.docuthek.com

Safety

Safety
Please read and keep in a safe place
Operating instructions for operators and installers
Electronic index EI
© 2017 Elster GmbH · Edition 08.17
alpha
Translation from the German

Contents

Contents
Electronic index EI .....................
Contents ..............................
Safety.................................
Checking the usage .....................
Installation ............................
Operating the electronic index ............
Navigating within the menu ..............
Service mode ..........................5
Establishing an optical communications link 7
Changing the battery ....................7
Setting the index parameters .............8
Releasing the valve .....................8
Retrofitting/replacing the communication
module ...............................9
Check test.............................9
Assistance in the event of malfunction ....4
Spare parts ...........................4
Technical data ........................5
Logistics .............................5
Contact ..............................6
Please read through these instructions
installation, pass the instructions on to the opera­tor. This unit must be installed and commissioned in accordance with the regulations in force. These instructions can also be found at www.docuthek.com.
Explanation of symbols
• ,  ,  ,  ... = Action
= Instruction
Liability
We will not be held liable for damage resulting from non-observance of the instructions and non­compliant use.
Safety instructions
Information that is relevant for safety is indicated in the instructions as follows:
DANGER
Indicates potentially fatal situations.
WARNING
Indicates possible danger to life and limb.
CAUTION
Indicates possible material damage.
All interventions may only be carried out by qualified gas technicians. Electrical interventions may only be carried out by qualified electricians.
Conversion, spare parts
All technical changes are prohibited. Only use OEM spare parts.
Changes to edition .
The following chapters have been changed: – Checking the usage – Logistics
GB-1
Page 2
GB
F
NL
I
SK

Checking the usage

*
Current meter reading
OFF
Electronic index EI.00, EI.05, EI.06, EI. for diaphragm gas meters BK-G...E, BK-G...ET or BK-G...ETe
Electronic index for reading out absolute meter read­ings and for retrieving consumption data, current tariffs, messages and the valve position.
This function is only guaranteed when used within the
specified limits – see page 15 (Technical data). Any other use is considered as non-compliant.
Type code
Code Description
EI Electronic index based on EI2
.00 .05 .06 .
Zigbee with valve Vs
Zigbee with valve Ve Zigbee, valve Vs, top reading Zigbee, valve Ve, top reading
Communication module
.00 .0 .0
The index version is shown on the index plate,
ECM.10 (Zigbee SMETS 1)
ECM.02 (Zigbee)
see page 2 (Type label/Index plate). When retrofitting or replacing the communication
module, it may be the case that the last two char­acters in the type code are no longer applicable.
Part designations
Variant
None
ATEX
The electronic index is suitable for use in potentially
explosive atmospheres. For the exact use (zone), see ATEX sticker on the diaphragm gas meter or see the operating instructions for diaphragm gas meters BK-G1.6 to BK-G25 http://docuthek. kromschroeder.com/doclib/main.php?language
=1&folderid=400041&by_class=2&by_lang=-1.

Installation

Installing the gas meter
For installing the gas meter in the pipework, refer to the operating instructions for diaphragm gas meters BK-G1.6 to BK-G25 http://docuthek. kromschroeder.com/doclib/main.php?language=1 &folderid=400041&by_class=2&by_lang=-1.
Gas meter with integrated valve
▷ If the integrated shut-off valve in the gas meter
is closed, it must be released, see page 8 (Releasing the valve).

Operating the electronic index

The display on the index is switched off.
Briefly press any key.
123 45678
BK-Gx
DE-07-MI002-PTB001
max
Q
min
Q
M07 0102
002545632734
5
6
Electronic index EI2
Display User keys 4 Opto-adapter interface 5 Service cap 6 Installation seal/screw locking cap
Type label/Index plate
Please quote for all enquiries:
Manufacturer’s serial number S/N (at the bottom left)
▷ Index version EI (next to the serial number)
V p
S/N: ............ EI2. ...
BK-G..E..
Q
max
Q
min
max
M ... 0102
12345678
class 1,5 t
m
t
g
class 1,5 Valve Ve
1
2
3
A beep sounds and the main screen appears.
An internal buzzer gives audible feedback, e.g. a short beep indicates a valve is open and a long beep indicates a valve is closed. A short beep sounds each time a key is pressed or if the unit automatically changes back to the main screen.
4
This function can be switched off, see page 8
(Setting the index parameters).
Main screen
1
3
00000.
ON
004
m
2
3
Menu area Information area Status line (symbols)
The ON/
symbols are only displayed when
a valve is integrated in the gas meter.
GB-2
Page 3
GB
F
NL
I
SK
OFF
ON
User keys, selection key and symbols
You can navigate through the menu using the
user keys , and the selection key .
Symbol Meaning
Navigate to the left or the right on each
,
level using the user keys. Briefly pressing the selection key selects
a sub-menu. Holding the selection key pressed down switches the display back to the previ­ous menu.
Briefly pressing the selection key selects a sub-menu.
Holding the selection key pressed down switches the display back to the previ­ous menu. Keys inactive
, ,
Valve/gas flow closed. This symbol is only displayed when a valve is integrated in the gas meter.
Valve/gas flow released. This symbol is only displayed when a valve is integrated in the gas meter.
New messages
Invalid data
Error message Low battery. This symbol is only dis-
played when battery power is low. Marking for metrology-relevant data Substitute for undisplayable character
in sentence
In the “Icon definitions” menu, the most important symbols are described briefly.
Icon denitions
OFF
Valve closed Valve open
ON
Low battery
Press for more ...
OFF

Navigating within the menu

The menu is constructed hierarchically.
Depending on the configuration, some menus may be missing.
The “Current meter reading” main screen appears
when switching on the index.
If you are in a different menu, the display will automatically change back to the main screen when no user key has been pressed for 30s, and switches off after a further 30s.
You can navigate from the main screen to the
various menus, such as “Meter information” us­ing the user keys
, .
Menu overview
The display can differ depending on the parameteriza-
tion or communication module.
Main menu –
Optional contents from communication module
Current meter reading Icon definitions
Valve release procedure
Date and time
Historic meter readings
Meter information
Meter identification
Calibration information
Temperature conversion (optional)
Optional contents from communication module
For further information, refer to the operating instructions for communication modules type ECM (for electronic index).
Current meter reading
The absolute meter reading and optionally the
current tariff are indicated in the main screen.
This appears when switching on the index.
You can receive information about the symbols by
pressing the selection key and the user keys
, , or see page 3 (User keys, selection
key and symbols).
Valve release procedure
▷ If the valve was released while the display was
switched off, the release note will appear the next time the index is switched on.
Valve open safety check
CHECK IF
APPLIANCES OFF
HOLD FOR GAS
▷ The note remains active until the valve has been
released, see page8 (Releasing the valve).
If the selection key is not pressed, the display will switch back to the main screen after 30s.
GB-3
Page 4
GB
F
NL
I
SK
Date and time
*
Date and time
*
4 hour summary: Sat
Information on the date and time display.
10012011 10
:02:06
The operator can transfer the switchover be-
tween winter and summer time to the commu­nication module, provided that it supports this.
The date is given in the format day - month - year.
The date format can differ depending on the
market.
This display is only visible if access to the historic
meter readings has been activated.
Historic meter readings
Consumption data dating as far back as 60weeks can be called up.
Interval data
Historic
meter readings
ON
The historic meter readings can be viewed as an
option. Depending on the parameterization, ac-
cess to the historic meter readings can either be – fully activated or – protected by entering a password or – deactivated.
▷ By pressing the selection key , consumption
data are displayed, which are given by month,
week or day, or 4-hour or ½-hour intervals.
“4 hour summary” example:
Electronic index with communication module:
▷ The
Meter information
symbol is displayed if, for example, the tolerance between the internal time recording and the actual time is too large. This can lead to invalid consumption data. After the next time synchronization, the consumption data are re­corded again correctly and disappears.
Meter-specific technical data are displayed in sub-menus by pressing the selection key several times.
Meter information
Identification & calibration info
ON
Meter identification: – No.: (Owner’s meter number) – EN 1359 Reg. No.: NG-4701BM0443 (exam-
ple) – Firmware version – CRC (checksum) – Details (firmware details)
For further information, see page 15 (Techni-
cal data).
Calibration information:
Meter calibration parameters Q1 to Q3 (adjust-
ment values Q1 to Q3 for three-point calibration) – Cyclic meter volume
Temperature conversion (optional): – Type of conversion (mechanical or electronic) – Base temperature tb (in accordance with
EN1359)
Specified centre temperature tsp (in accordance
with EN1359)
01-02-2011 04:00 → 08:00
00000.351 → 00000.353m
0.002 m
The timeframe is displayed with date and time
for the start and end of the period.
▷ The meter reading is displayed for the start and
end of the period in m
The consumption for this period is indicated in m3.
The tariff band may be displayed.
3
.
3
3
GB-4
Page 5
GB
F
NL
I
SK

Service mode

Test instructions
*
Part designations
Service mode menu overview
8
1234 5678
BK-Gx
DE-07-MI002-PTB001
max
Q
min
Q
M07 0102
002545632734
5
1
2 3
4
7
6
Electronic index EI2
Display User keys 4 Opto-adapter interface 5 Service cover 6 Installation seal/screw locking cap 7 Battery 8 Communication module (optional)
▷ In Service mode, meter-specific operating data
can be called up.
Activating Service mode  Hold the selection key
pressed down.
▷ A pixel will appear in each corner of the display.
Observe one pixel: while the pixel is visible, hold
the selection key pressed down. Release the key as soon as the pixel has disappeared.
Repeat the process, until all the pixels are off and
“Test instructions” appears in the menu area.
▷ Service mode is activated.
Main menu
Test instructions
Temperature (optional)
Battery diagnosis
Date & time
LCD pixel test
Cyclic test
UMI peripherals UMI-1 UMI-2 UMI-3
Optional contents from communication module
Meter information
Test instructions
Automatic return to main menu after 5 minutes of inactivity Hold on any screen to return immediately
ON
Temperature (optional)
This menu is available if temperature conversion
has been activated.
Temperature
tg 25.04°C TC : electronic tg : [-10, 40]°C tsp : 20°C tb : 0°C
OFF
tg currently measured gas temperature t – TC: type of temperature conversion
Electronic: mathematical conversion to tb
in index
Mechanical: mechanical conversion to t
in measuring unit
GB-5
g
b
Page 6
GB
F
NL
I
SK
tg: [ ] max. allowable gas temperature range
OFF
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : OK
OFF
OFF
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : OK
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : Change required
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : Test in progress Time : 05:00 h:m
*
Date and time
*
Cyclic test: Starting
– tsp: specified centre temperature tsp (in ac-
tb: base temperature t
Battery diagnosis
▷ When the battery is connected, the status “OK”
t
[min. value, max. value]
g
cordance with EN 1359)
EN 1359), see page 15 (Technical data) Check test for temperature test, see page 9 (Check test).
The measured value is updated once per minute.
is shown in the display.
(in accordance with
b
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : OK
OFF
– The U(hlc) value indicates the current voltage
measured on the battery or HLC.
– The U(min) value signals the minimum voltage
measured on the battery or HLC.
▷ If battery power is low, the display changes to
the status “Change required”. The battery must be changed within a short time.
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : Change required
Having disconnected the battery, the display switches to the status “Removed”.
Battery diagnosis
U(hlc) : 3.60 V U(min) : 3.00 V Status : Removed
OFF
OFF
Date and time
Information on the date and time display.
10012011 10
:02:06
The operator can transfer the switchover be-
tween winter and summer time to the commu­nication module, provided that it supports this.
The date is given in the format day - month - year.
The date format can differ depending on the
market.
This display is only visible if access to the historic
meter readings has been activated.
LCD pixel test
A display test can be carried out in this menu.
Follow the displayed instructions.
A test pattern is shown in the display.
Briefly press the selection key
A further test pattern appears in the display.
Hold the selection key pressed down. The display
switches to the previous menu.
Cyclic test
▷ The accuracy of the meter can be checked us-
ing a cyclic test.
C 00.000000 U 00.000000
m m
tg 25.04°C N 00000-0 t 00000-0 s
Press to abandon test
– C = converted volume (temperature only)
for BK-G...E: no function (C = U) for BK-G...ET: C = V for BK-G...ETe: C = Vb, conversion to tb (see
page 15 (Technical data))
– U = non-converted volume V
for BK-G...ET: U = converted volume V
The conversion takes place in the measuring unit. – tg = measured gas temperature in °C – N = number of complete measuring cycles
(measuring unit revolutions) – number of inter­mediate sampling points in the measuring cycle (max. 8)
– t = total testing time in seconds
▷ for BK-G...ETe, the following applies: The relationship between Vb and VP is as follows:
Vb = VP x Tb / T
GB-6
, conversion to tb = 15°C
b15
g
.
3
3
P
b15
Page 7
GB
F
NL
I
SK
where Tb = (273.15 + {tb}) K
ON
Meter information
Identification & calibration info
Calibration information
Q1: 0.000m3/h CF1:16384 Q2: 6.000m
3
/h CF2:16384
Q3: 12.000m
3
/h CF3:16384
5
m
m
T
= (273.15 + {tg}) K
g
tb is specified in the technical data, see page
15 (Technical data).
▷ The curly brackets mean “numerical value of”.
For detailed information on the check test op-
erating sequence, see page 9 (Check test).
UMI peripherals
The status of the available interfaces for optional
communication modules is displayed.
Meter information
Meter-specific technical data are displayed in sub-menus by pressing the selection key and the user keys , .
Meter information
Version :x.xx.xx CRC :0x0000 Details :REL YYMMDD Year of man. :YYYY
Meter information: – Software version – CRC: software checksum – Software details – Year of manufacture
▷ Other screen descriptions (not illustrated):
Calibration information: – Meter calibration parameters Q1 to Q3 (adjust-
ment values Q1 to Q3 for three-point calibration)
Meter characteristics: – Cyclic meter volume – Transitional flow rate Q – EN 1359 Reg. No.: NG-4701BM0443 (example)
Ambient conditions: – Electromagnetic – Mechanical

Establishing an optical communications link

▷ Depending on customer requirements, the opti-
cal interface can be locked.
▷ In order to configure the electronic index for the
respective application, the optical communica­tions link must be activated.
Position the opto-adapter head on the interface
provided.
1234 5678
Press any user key.
Optical communication is enabled for 1minute.
If the optical communications link is not used during this time, the interface will be deactivated.
Initiate communication.
The procedure depends on your user software.
For further information, refer to the operating instructions for communication modules type ECM (for electronic index).

Changing the battery

WARNING
Risk of explosion in explosion-hazard areas!
– As a general rule, maintenance and repair work
should be avoided in explosive atmospheres.
– Check that the electrical system complies with
the special electrical explosion protection re­quirements.
– When working on electrical equipment in an
explosion-hazard area, only design-approved electrical operating equipment may be used.
– Use original spare parts supplied by Elster
GmbH, see page 14 (Spare parts).
▷ The battery is available as a spare part.
Prepare the index for changing the battery.
The battery can only be replaced when the optical
communications link has been established, see page 7 (Establishing an optical communica-
t
GB-7
tions link). Otherwise, data can be lost.
Start the battery change procedure.
The procedure depends on your user software.
Important! The battery may only be removed once a release message has been issued by the index. Otherwise, an integrated valve will be closed and/or an error message will be trans­ferred via the communication module.
▷ The index confirms disconnection of the battery
with a sound sequence.
Replace the battery as quickly as possible.
Q Q
00254
The index confirms connection of the battery
max
Q
min
Q
M07 0102
002545632734
with a sound sequence.
The main screen appears when any key is
pressed.
▷ Optical communication is enabled for 1minute.
If the optical communications link is not used during this time, the interface will be deactivated.
Page 8
GB
F
NL
I
SK
4 Reprogram the battery parameters.
ON
Valve open safety check
CHECK IF
APPLIANCES OFF
HOLD FOR GAS
Gas ow check complete
Gas flow check
successful
ON
OFF
Valve open safety check
CHECK IF
APPLIANCES OFF
HOLD FOR GAS
Opening valve
Please wait
Gas ow check complete
Gas flow check
successful
Gas flow check
failed
Gas ow check complete
▷ The procedure depends on your user software.
Electronic index without communication module:
▷ After the battery has been replaced, check the
time and update it if necessary. Note the global UTC format.
5 Replace the service cover. 6 Push in a new screw locking cap, see page 14
(Spare parts). The body carrying out this task should apply its own seal.
▷ After a short time, the unit switches to initializa-
tion mode.
Opening valve
Please wait

Setting the index parameters

The following index parameters can be adjusted using
the user software: – properties of the historic values screen, – time display, – audible feedback when a key is pressed, – parameters for valve release procedure.

Releasing the valve

If a valve is integrated in the diaphragm gas meter BK, this must be released/opened for commissioning.
CAUTION
To avoid damage: – Ensure the customer’s consumers are closed.
▷ The valve can only be released when the optical
communications link has been established or via a communication module.
▷ Unless otherwise agreed, the valve is open on
delivery as standard.
▷ Once the valve has been released via the com-
munication module, this must be confirmed at the meter on site by the consumer or another person if applicable.
Establish the optical communications link, see
page 7 (Establishing an optical communica­tions link).
Release the valve. The procedure depends on
your user software.
Valve open safety check
After successful initialization, the gas flow check is started. The test duration is shown in the dis­play.
Gas ow check in progress
Max time: Min time
Once the release criteria have been checked, the results are shown in the display.
Gas ow check complete
Gas flow check
successful
ON
00:29:56
:
00:06:08
CHECK IF APPLIANCES OFF HOLD FOR GAS
▷ The test procedure may differ from the descrip-
tion.
▷ Press the selection key and hold down.
GB-8
Page 9
GB
F
NL
I
SK
Retrofitting/replacing the
31 2
G
2
31 2
3
4
5
communication module
WARNING
Risk of explosion in explosion-hazard areas!
– As a general rule, maintenance and repair work
should be avoided in explosive atmospheres.
– Check that the electrical system complies with
the special electrical explosion protection re-
Replacing the communication module  Establish the optical communications link, see
page 7 (Establishing an optical communica­tions link).
Deactivate the respective communication mod-
ule.
BK-
DE-07-MI00
max
Q
min
Q
M07 0102
002545632734
quirements.
– When working on electrical equipment in an
explosion-hazard area, only design-approved electrical operating equipment may be used.
6
– The communication module may only be pulled
out if it has first been deactivated.
– When installing a communication module, en-
sure that the communication interface has been deactivated.
– Further information can be found in the operat-
ing instructions for communication modules type ECM (for electronic index).
– Use original spare parts supplied by Elster
GmbH, see page 14 (Spare parts).
The communication module is available as a
spare part.
▷ The communication module can only be retrofit-
ted or replaced when the optical communica­tions link has been established.
The procedure differs depending on whether
the communication module is to be retrofitted or replaced. See “Retrofitting the communica­tion module” or “Replacing the communication module”.
Retrofitting the communication module
Check the status of the interface and deactivate if necessary.
7 The communication module and index must be
paired.
The procedure depends on your user software. Set the parameters of the communication mod-
ule according to the instructions or the servicing steps provided by the relevant operator.
8 Replace the service cover. 9 Push in a new screw locking cap, see page 14
(Spare parts). The body carrying out this task should apply its own seal.

Check test

Directive 2014/32/EU (MID) prescribes that it must be possible to check the meter.
The requirements and test methods must comply
with national laws and regulations.
The following tests describe the check tests
which are carried out by accredited testing agencies.
Always conduct a pressure and temperature correction in accordance with established pro­cedures (unit under test against master meter).
Measurement accuracy class, see page 15 (Technical data).
4 Establish the optical communications link, see
page 7 (Establishing an optical communica­tions link).
▷ The intended interface for the optional commu-
nication module is UMI-1.
5 The communication module and index must be
paired.
The procedure depends on your user software. Set the parameters of the communication mod-
ule according to the instructions or the servicing steps provided by the relevant operator.
6 Replace the service cover. 7 Push in a new screw locking cap, see page 14
should apply its own seal.
(Spare parts). The body carrying out this task
The unit under test must be acclimatized and
installed on the test rig.
Maintain the climatic conditions constant dur­ing the entire test duration. Otherwise, the test results will be inaccurate.
Immediately before the beginning of the test, the quantity of test air, which corresponds to at least 50x the cyclic volume of the meter to be tested, is fed through the meter at a flow rate of Q (maximum flow rate of a gas meter).
During an active cyclic test, the display dis­appears after 5 minutes but lights up every min­ute for 10 seconds. This function is available for max. 5 hours.
max
GB-9
Page 10
GB
F
NL
I
SK
To conduct the tests, the thermowell and the
pressure test point (if available) can be used as a reference for the temperature and pressure measured by the meter.
Cyclic test
The cyclic test is designed for checking the meter
with a master meter. The recorded volume of the unit under test during the testing period can be read off directly from the index once the test has been completed and can be compared with the master meter. Testing at a constant flow rate thus ensures the lowest possible level of measurement uncertainty for the unit under test.
Legend ΔtN = total master meter testing time in s ΔtP = testing time of the unit under test in s Q
= maximum flow rate of a gas meter
max
Q
= minimum flow rate of a gas meter
min
QN = flow on master meter in m³/h based on
the displayed volume V
Q
= actual flow rate on master meter in m³/h
act,N
QP = flow determined on unit under test based
N
on VP in m³/h VN = displayed volume on master meter in m³ V
= actual volume having flowed through the
act,N
master meter in m³ VP = displayed volume on unit under test in m³ Value after C or U in display, depending
on device configuration and test method.
See test procedure below for further
details. FN = error of the master meter in % FP = error of the unit under test in % pN = absolute pressure on the master meter in
mbar pP = absolute pressure on the unit under test in
mbar
TN = absolute temperature on the master meter
in K
TP = absolute temperature on the unit under
test in K tb = base temperature in °C V
= converted volume with reference to
b15
Vb = converted volume with reference to t
There are two different options for carrying out the
tb=15°C
b
cyclic test:
Option : cyclic test at a constant flow rate
▷ The test rig is in pre-trial operation, i.e. start of
measurement on the unit under test will be de­layed.
▷ Maintain the flow rate constant.
Test load and minimum test volumes for the test with
index readout:
Type
BK-G1.6 2.5 1.2 1.2 12 60 BK-G2.5 4.0 1.2 1.2 12 60 BK-G4 6.0 1.2 1.2 12 60 BK-G2.5 4.0 2 2 20 100 BK-G4 6.0 2 2 20 100 BK-G6 10 2 2 20 100 BK-G6 10 4 4 40 200 BK-G6 10 6 6 60 300 BK-G10 16 6 6 60 300 BK-G16 25 6 6 60 300 BK-G25 40 12 12 120 600
Q
The minimum test volumes are recommended
max
in
m3/h
Cyclic volume in dm
Test volume in dm3 at
Q
0.2 Q
min
3
maxQmax
guide values. The measurement uncertainty of the complete system (test rig plus unit under test) must not exceed 1/3 of the maximum permis­sible error (MPE).
The testing time must be at least 10s due to
the system design.
In the test procedure described below, it is guar­anteed that the unit under test always performs full measuring unit rotations.
Master meter test procedure
Q
N
∆t
N
61
∆t
P
2 3 4 5
Set the test flow rate. Start measuring the reference timeΔtN at mark-
er1.
Immediately afterwards, briefly press the selec-
tion key on the index to start the cyclic test on the unit under test– marker2. The index will thus be “armed” for measurement.
As soon as one of the significant sensor posi­tions has been detected, the unit changes to measuring mode– marker3. A beep confirms the start of the measurement.
GB-10
Page 11
GB
F
NL
I
SK
Once the required minimum testing time has been reached, the measurement can be termi­nated– marker4.
4 Briefly press the selection key in order to end
the measurement. Measurement on the unit under test stops au-
tomatically once the full number of measuring unit revolutions has been completed– marker5.
▷ A beep acknowledges the end of the measure-
ment.
Measurement is terminated automatically after 5 hours.
5 Stop the test on the master meter– marker6.
▷ The measurements are then available.
6 Read off the flow rate on the master meter or
Option : cyclic test with a given volume
Test load and minimum test volumes for the test with
index readout:
Cyclic
Q
max
Type
BK-G1.6 2.5 1.2 36 72 72 BK-G2.5 4.0 1.2 36 72 72 BK-G4 6.0 1.2 36 72 72 BK-G2.5 4.0 2 60 120 120 BK-G4 6.0 2 60 120 120 BK-G6 10 2 60 120 120 BK-G6 10 4 120 240 240 BK-G6 10 6 180 360 360 BK-G10 16 6 180 360 360 BK-G16 25 6 180 360 360 BK-G25 40 12 360 720 720
volume
in
3
/h
in dm
m
Test volume in dm3 at
Q
min
3
0.2 Q
maxQmax
calculate if necessary:
a) taking into account the inherent error of the
Master meter test procedure
master meter: Q b) If the inherent error of the master meter has
= VN x 3600 s/h / ((1+FN/100) x ΔtN)
act,N
V
N
already been taken into account in the dis-
played volume (VN = V
= V
Q
act,N
7 Calculate the flow rate on the unit under test:
x 3600 s/h / Δt
act,N
act,N
):
N
8
10
QP = VP/ΔtP for BK-G...ET, the following applies: VP = V
8 The accuracy is checked by comparing the flow
rates. The pressure and temperature values of the unit under test corrected with reference to
b15
97
11
the master meter have already been taken into
account here:
FP = 100% x (((QP x pP x TN) / (Q
The following applies:
x pN x TP)) - 1)
act,N
pP = relevant absolute pressure on the unit under
test The following also applies: for BK-G...E: T where t
under test in °C (display)
= (273.15 + {tg}) K
P
= relevant gas temperature on the unit
g
for BK-G...ET: TP = (273.15 + {tb}) K where tb = 15°C (see page 15 (Technical data)):
T
= 288.15 K
P
for BK-G...ETe:
in step 7 is determined from the converted
If Q
P
volume Vb, the following applies: TP = (273.15 + {tb}) K (tb see page 15 (Techni-
cal data))
in step 7 is determined from the non-con-
If Q
P
verted volume VP, the following applies: TP = (273.15 + {tg}) K where tg = relevant gas temperature on the unit
under test in °C (display)
The curly brackets mean “numerical value of”.
On a nozzle test rig with a known flow rate,
steps  and 6 can be omitted.
The error calculation is based on PTB Testing
Instructions, Volume 29: “Messgeräte für Gas –
meters), Edition 2003.
Gaszähler” (Measuring instruments for gas – gas
To activate the cyclic test on the unit under test,
briefly press the selection key on the index– marker7. The index will thus be “armed” for measurement.
Start the test on the master meter– marker8.
As soon as one of the significant sensor posi­tions has been detected, the unit changes to measuring mode– marker9.
Test is ended– marker10. 4 Read off the test results on the unit under test.
The measured values are updated with each 1/8
revolution of the measuring unit.
5 Compare the measurement results with the mas
ter meter and determine the measuring deviation on the unit under test:
a) taking into account the inherent error of the
master meter:
FP = 100% x (((VP x (1+FN/100) x pP x TN) /
(VN x pN x TP)) - 1)
b) If the inherent error of the master meter has al-
ready been taken into account in the displayed volume (VN = V
FP = 100% x (((VP x pP x TN) / (V
TP)) - 1)
The following applies: pP = relevant absolute
), the following applies:
act,N
act,N
x pN x
pressure on the unit under test
-
GB-11
Page 12
GB
F
NL
I
SK
▷ The following also applies: for BK-G…E: T where t
g
under test in °C (display)
= (273.15 + {tg}) K
P
= relevant gas temperature on the unit
for BK-G...ET: TP = (273.15 + {tb}) K where tb = 15°C (see page 15 (Technical data)):
T
= 288.15 K
P
for BK-G…ETe: If the converted volume V TP = (273.15 + {tb}) K (tb see page 15 (Techni-
is taken for VP:
b
cal data))
If the non-converted volume V
= (273.15 + {tg}) K
T
P
where tg = relevant gas temperature on the unit
is taken for VP:
P
under test in °C (display)
▷ The curly brackets mean “numerical value of”. 6 Stop execution of the cyclic test– marker 11.
Briefly press the selection key twice in order to stop the measurement.
A beep confirms interruption of the measurement.
Measurement is terminated automatically after 5 hours.
The error calculation is based on PTB Testing
Instructions, Volume 29: “Messgeräte für Gas – Gaszähler” (Measuring instruments for gas – gas meters), Edition 2003.
Pulse test
The pulse test is designed for checking the meter with
a master meter. The recorded volume is provided by pulses which are emitted via the optical interface.
Test load and minimum test volumes for the test with
index readout:
Q
Cyclic
Type
BK-G1.6 2.5 1.2 36 72 72 BK-G2.5 4.0 1.2 36 72 72 BK-G4 6.0 1.2 36 72 72 BK-G2.5 4.0 2 60 120 120 BK-G4 6.0 2 60 120 120 BK-G6 10 2 60 120 120 BK-G6 10 4 120 240 240 BK-G6 10 6 180 360 360 BK-G10 16 6 180 360 360 BK-G16 25 6 180 360 360 BK-G25 40 12 360 720 720
Pulse value V
This test can only be carried out when the opti-
max
in
m3/h
volume
in dm
Imp
Test volume in dm3 at
Q
0.2 Q
min
3
maxQmax
, see page 15 (Technical data).
cal communications link has been established.
The procedure depends on your user software.
Establish the optical communications link, see
page7 (Establishing an optical communica­tions link).
Interrupt communication to the installed commu-
nication modules before starting the test so that the measurement accuracy will not be adversely
affected. As of step 3, continue as described in one of the two following sections.
Pulse test at a constant flow rate (optical interface)
Master meter test procedure
Q
N
∆t
N
12
∆t
P
13 14
17
15 16
Set the test flow rate. 4 Start measuring the reference time ∆tN at mark-
er12.
▷ Marker12 indicates the release of the test flow
rate on the master meter.
5 Immediately afterwards, start the pulse test on
the unit under test– marker13.
▷ Then the unit under test generates the volume
pulses on the optical interface each time the lowest-order decimal place on the meter display is incremented– marker14. The test begins.
6 As soon as the required minimum test volume
has been reached on the unit under test, the time measurement on the unit under test can be stopped– marker15.
7 End the pulse test using any command– mark-
er16.
Measurement is terminated automatically after 90minutes.
8 Shut off the gas flow to the master meter–
marker17.
9 Determine the volume on the unit under test VP: VP = N x V N = number of pulses during Δt V
= pulse value V
Imp
Imp
cal data)
, see page 15 (Techni-
Imp
P
0 Calculate the flow rate on the unit under test:
QP = VP/Δt
 Read off the flow rate on the master meter or
P
calculate if necessary:
a) taking into account the inherent error of the
master meter: Q b) If the inherent error of the master meter has
= VN x 3600 s/h / ((1+FN/100) x ΔtN)
act,N
already been taken into account in the dis-
played volume (VN = V Q
act,N
= V
x 3600 s/h / Δt
act,N
act,N
):
N
GB-12
Page 13
GB
F
NL
I
SK
 The accuracy is checked by comparing the flow
rates. The pressure and temperature values of the unit under test corrected with reference to the master meter have already been taken into account here:
FP = 100% x (((QP x pP x TN) / (Q
TP)) - 1)
act,N
x pN x
The following applies:
pP = relevant absolute pressure on the unit under test The following also applies: for BK-G...E: TP = (273.15 + {tg}) K where t
= relevant gas temperature on the unit
g
under test in °C (display) for BK-G...ET and BK-G...ETe: TP = (273.15 + {tb}) K (tb see page 15 (Techni-
cal data))
The curly brackets mean “numerical value of”.
On a nozzle test rig with a known flow rate,
step 4 can be omitted.
The error calculation is based on PTB Testing
Instructions, Volume 29: “Messgeräte für Gas –
Gaszähler” (Measuring instruments for gas – gas
meters), Edition 2003.  Recommission the previously switched off com-
munication modules.
Pulse test with a given volume
Master meter test procedure
V
N
19
18 21
20
Start the pulse test on the unit under test–
marker18.
4 Start the test on the master meter– marker19. 5 Record the pulses from the unit under test.
The unit under test generates the volume pulses
on the optical interface each time the lowest-
order decimal place on the index display is in-
cremented. 6 The test volume is achieved and the test ends–
marker20. 7 End the pulse test on the unit under test using
any command– marker21.
Measurement is terminated automatically after
90minutes.
8 Determine the volume on the unit under test VP: V
= N x V
P
N = number of pulses V
= pulse value in m³
Imp
9 Read off the volume on the unit under test.
Imp
The inherent error of the master meter is to be
taken into account if applicable:
V 0 Determine the measuring deviation on the unit
= VN / (1 + FN/100)
act,N
under test:
FP = 100% x (((VP x pP x TN) / (V
The pressure and temperature values of the unit
x pN x TP)) - 1)
act,N
under test corrected with reference to the master
meter have already been taken into account here. The following applies: pP = relevant absolute pressure on the unit under
test The following also applies: for BK-G…E: T where tg = relevant gas temperature on the unit
= (273.15 + {tg}) K
P
under test in °C (display) for BK-G…ET and BK-G…ETe: TP = (273.15 + {tb}) K (tb see page15 (Techni-
cal data))
The curly brackets mean “numerical value of”.
The error calculation is based on PTB Testing
Instructions, Volume 29: “Messgeräte für Gas –
Gaszähler” (Measuring instruments for gas – gas
meters), Edition 2003.  Recommission the previously switched off com-
munication modules.)
RTC test
The climatic conditions must be maintained con-
stant at 22± 5°C during the entire test duration.
Temperature changes in 24hours ≤2K.
Ensure that conditions remain sufficiently stable
during the measurement.
▷ The accuracy of the time count can be verified
with this test. Acclimatize the unit under test and place next
to the time reference unit. If necessary, activate the time display on both
units.
Ensure synchronous reading by taking a photo. 4 Observe a min. testing time of 24hours. 5 Repeat steps  and  . 6 The clock deviation of the unit under test must
not exceed the maximum admissible deviation.
Maximum admissible deviation= 100ppm in
24hours.
GB-13
Page 14
GB
F
NL
I
SK
Temperature test
A temperature test is required on diaphragm gas meters with temperature conversionBK..Te only.
▷ The accuracy of the temperature measurement
can be verified with this test.
▷ The temperature test can only be carried out in
Service mode.
CAUTION
To avoid damage to the unit: – Comply with ambient temperature, see
page 15 (Technical data). Deviations from the permitted ambient temperature will be re­corded in the error memory.
Temperature measurement accuracy, see
page15 (Technical data).
Install the diaphragm gas meter in a climatic
chamber.
Activate Service mode– see page 5 (Service
mode).
Change to the “Temperature” menu.
▷ The current gas temperature is displayed.
4 Close the climatic chamber. 5 Select an ambient temperature as a reference
value and bring the climatic chamber to this tem­perature.
To ensure there is also a uniform temperature
in the meter, we recommend starting the meter air/gas flow during the temperature adjustment phase.
Ensure that temperature distribution remains uniform and stable during the temperature measurement.
6 Compare the measured value to the temperature
reference value.
If required, several reference values can be checked. In this case, repeat the test as of point 5 .
Leave the index unused for an extended period,
e.g. 24hours. After this, the user interface will once again be available.
? When pressing the user keys, the display
remains switched off and no beep can be heard.
! The index is defective.
Contact the manufacturer.
? The
symbol is displayed.
! Low battery. This symbol is only displayed when
battery power is low.
Replace the battery.
In the case of faults which are not described here, contact the manufacturer immediately.

Spare parts

Battery
For meters without communication module: Elster Part No.: 32906520. For meters with communication module: Elster Part No.: 72910321.
Communication module
Use ATEX compliant communication modules only. A safety sticker with the following code must be attached to the communication module:
II 3G Ex ic IIA Gc (-25°C ≤ Ta ≤ +55°C). Electrical connection parameters, see page 15 (Technical data).
Screw locking cap

Assistance in the event of malfunction

? Fault ! Cause
• Remedy
Possible faults and suggested solutions ? The ! If the symbol appears next to a measured
After the next data synchronization, the data are
? When pressing the user keys, the backlight-
! Energy-saving mode is active. Due to excessive
symbol is displayed.
value, this means that the value is invalid.
recorded again correctly and disappears.
ing and/or display remain switched off. A beep can nevertheless be heard.
use of the index, the average energy consump­tion has been exceeded.
Elster Part No.: 32447510.
GB-14
Page 15
GB
F
NL
I
SK

Technical data

Application with diaphragm gas meters BK..E RoHS compliant Enclosure: IP 54. Battery life: approx. 10years. Maximum allowable ambient temperature range: see type label/index plate. Data logger for historic meter readings: up to 60weeks in 30-minute intervals. Optical interface: pursuant to EN62056-21, Mode(E), AnnexB.2. Accuracy of the clock: 0.4 s/day at 20°C on the day of manufacture. Diaphragm gas meter BK..ETe with temperature conversion:
Temperature measurement accuracy: ± 1°C on the
day of manufacture.
The base temperature tb is specified on the index
plate.
Pulse output of optical interface
Pulse value V
Gas meter
BK-G 1.6 – BK-G 6 3 1 BK-G 10 – BK-G 25 2 10
Pulse duration: 90 ms
For further technical data on diaphragm gas meters BK– see the operating instructions for diaphragm gas meters BK-G1.6 to BK-G25 http://docuthek. kromschroeder.com/doclib/main.php?language=1& folderid=400041&by_class=2&by_lang=-1.
Imp
:
Decimal
place
in display
Pulse
value V
in dm
Imp
3

Logistics

Transport
Diaphragm gas meters are always to be transported in the upright position. On receipt of the product, check that the delivery is complete, see page 2 (Part designations). Report any transport damage immediately.
Storage
Diaphragm gas meters are always to be stored in the upright position and in a dry place. Ambient tem­perature: see page 15 (Technical data).
Disposal
Meters with electronic components: Components, particularly batteries, are to be dis­posed of separately. On request, old units may be returned carriage paid to the manufacturer, see page 16 (Contact), in accordance with the relevant waste legislation re­quirements.
ATEX explosion protection
Battery
Only use approved batteries from Elster.
The “Battery pack cpl. EI2” is certified as part of the
electronic index.
Communication modules
Please contact the manufacturer for details on suit­able communication modules. Information on the ATEX suitability can be found in the operating in­structions of the respective communication module.
The UMI interface features the following parameters: – UO = 3.9 V peak, 3.6 V nominal, – IO = 7.5 A peak, 506 mA nominal, – PO = 1.55 W, – CO = 12 µF, – LO = 50 nH.
For more data on explosion protection of diaphragm gas meters BK– see the operating instructions for diaphragm gas meters BK-G1.6 to BK-G25 http://
uage=1&folderid=400041&by_class=2&by_lang=-1.
docuthek.kromschroeder.com/doclib/main.php?lang
GB-15
Page 16
GB
F
NL
I
SK

Contact

Contact
United Kingdom
Elster Metering Limited Paton Drive
Tollgate Business Park
Beaconside Stafford, ST16 3EF
Tel. +44 1785 275200
Fax +44 1785 275305 jeavons.info@gb.elster.com www.elstermetering.co.uk
Germany
Elster GmbH
Strotheweg 1 49504 Lotte Tel. +49 541 1214-0
Fax +49 541 1214-370
info@elster-instromet.com
www.elster-instromet.com
Ireland
Active Energy Control Ltd. Unit 4, Clare Marts Quin Road Ennis, Co. Clare
Tel. +353 65 6840600
Fax +353 65 6840610 info@aec.ie www.aec.ie
GB-16
in the interests of progress.
We reserve the right to make technical modifications
Loading...