Homedics BPA-150 User Manual

Deluxe Automatic
Blood Pressure Monitor
El manual en español empieza en la página 34
lifetime
limited warranty
BPA-150
TABLE OF CONTENTS
INTRODUCTORY INFORMATION: Important Product Notices and Safety
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
About Blood Pressure . . . . . . . . . . . . . . . . . . . .5
What is Blood Pressure? Why Measure Your Blood Pressure?
Blood Pressure Standard . . . . . . . . . . . . . . . . . .5
How This Blood Pressure Monitor Works . . . . .7
USAGE INFORMATION:
Name/Function of Each Part . . . . . . . . . . . . . . . .8
Display Explanations . . . . . . . . . . . . . . . . . . . . .9
Installing Batteries . . . . . . . . . . . . . . . . . . . . . .11
Using the AC Adapter . . . . . . . . . . . . . . . . . . . .13
Date & Time Set Procedure . . . . . . . . . . . . . . .14
Applying the Cuff . . . . . . . . . . . . . . . . . . . . . . .15
Measurement Procedure . . . . . . . . . . . . . . . . .18
Risk Category Indicator . . . . . . . . . . . . . . . . . .21
Recalling Values from Memory . . . . . . . . . . . .23
Clearing Values from Memory . . . . . . . . . . . . .24
Important Notes Regarding Your
Blood Pressure Measurement . . . . . . . . . . . . .25
Care, Maintenance & Cleaning . . . . . . . . . . . . .27
Potential for Radio/Television Interference . . .29
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . .30
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .31
Lifetime Limited Warranty . . . . . . . . . . . . . . . .32
IMPORTANT PRODUCT NOTICES AND SAFETY INSTRUCTIONS
When using your blood pressure monitor, basic pre­cautions should always be followed. Please read and follow all instructions and warnings before using this product. Save these instructions for future reference.
• Please note that this is a home healthcare product only and it is not intended to serve as a substitute for the advice of a physician or medical professional.
• Do not use this device for diagnosis or treatment of any health problem or disease. Measurement results are for reference only. Consult a healthcare professional for interpretation of pressure measurements. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or healthcare professional.
• Proper cuff size is critical for accurate measurements. If the arrow on the cuff falls outside of the solid color line when applied to the user’s arm, contact HoMedics Consumer Relations regarding additional cuff sizes.
• This product is not suitable for people with arrythmias. This device may have difficulty determining the proper blood pressure for individuals with diabetes, poor circulation of blood, kidney problems or for users who have suffered from a stroke.
• The pulse display is not suitable for checking the frequency of heart pacemakers.
• Electromagnetic interference: The device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g. mobile telephones, microwave ovens). These may lead to temporary impairment of measurement accuracy.
• Use blood pressure monitor only for its intended use.
• Do not wrap the cuff around body parts other than your arm.
• This unit is intended for use in measuring the blood pressure and pulse rate of adults. Do not use this device on infants or those who cannot express their own consent. This unit should not be operated by children.
• Do not plug or unplug the adapter power cord with wet hands.
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope ausculatory method, within the limits prescribed by the American National Standard, Manual, electronic, or automated sphygmomanometers.
ABOUT BLOOD PRESSURE
What is Blood Pressure?
Blood pressure is the pressure exerted on the artery walls while blood flows through the arteries. The pressure measured when the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the heart dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure.
Why Measure Your Blood Pressure?
Among today’s various health problems, those associated with high blood pressure are very common. High blood pressure dangerously correlates with cardiovascular diseases. Therefore, blood pressure monitoring is important for identifying those at risk.
BLOOD PRESSURE STANDARD
The U.S. Department of Health and Human Services and the National Institutes of Health have developed a blood pressure standard that identifies areas of low­and high-risk blood pressure. This standard, however, is a general guideline as an individual’s blood pressure varies among different people, age groups, etc.
Categories for Blood Pressure Levels in Adults
Systolic (mmHg) Diastolic (mmHg)
Stage 2 Hypertension >160 >100 Stage 1 Hypertension 140~159 90~99 Prehypertension 120~139 80~89 Normal <120 <80
__
For adults 18 and older who are not on medicine for high blood pressure, are not having a short-term serious illness, and do not have other conditions, such as diabetes and kidney disease.
To determine category of risk when systolic and diastolic readings fall into two areas, use the higher of the two numbers for classification.
There is an exception to the above definition of high blood pressure for people with diabetes and chronic kidney disease. A blood pressure of 130/80 mmHg or higher is considered high blood pressure for those individuals.
This monitor comes equipped with a Risk Category Indicator to visually indicate the assumed risk level of each reading based upon these standards. See page 21 for more information on this feature.
It is important that you consult with your physician regularly. Your physician will tell you your normal blood pressure range as well as the point at which you will be considered at risk.
For reliable monitoring and reference of blood pressure, keeping long-term records is recommended
HOW THIS BLOOD PRESSURE MONITOR WORKS
This monitor uses Smart Sense™ Technology to detect your blood pressure. With one touch of a but­ton, the cuff will automatically inflate to block the blood flow through your artery. Next, the deflation process starts. Smart Sense™ Technology enables the monitor to automatically inflate and deflate to the appropriate level for each individual. Please note that any muscle movement during inflation or deflation will cause measurement error. When measurement is complete, the monitor will display your systolic pres­sure, diastolic pressure, and pulse readings.
Based upon your measurement results, the monitor's Risk Category Indicator will alert you if your reading falls into the pre-hypertension, stage 1 hypertension, or stage 2 hypertension categories. See page 21 for more information on the Risk Category Indicator.
This monitor also features an irregular heartbeat (IHB) detector that will alert the user to the presence of an irregu- lar heartbeat. It is strongly recommended that you consult your physician if the IHB symbol appears often (five times or more in a row). See page 22 for more information on the IHB Detector.
NAME/FUNCTION OF EACH PART
Smart Fit™ Arm Cuff
Air Tube and Connector
Memory Recall
Button
Other Accessories Included :
Date/Time Set Buttons
(Located on
back of unit)
START/STOP Button
1 12V DC Output AC Adapter and 4 “AAA” size, 1.5V alkaline batteries Note: Please unload the batteries from the battery pack when operating with the AC adapter for an extended period of time.
On/Off Switch
(Located on back of unit.)
LCD Display
Removable battery pack
(Located
on back of unit)
User-Select Button
DISPLAY EXPLANATIONS
Date/Time Indicator
Systolic Pressure
Diastolic Pressure
Pulse Rate
Display Symbols:
User 1: Appears when the monitor is operated by User 1. User 2: Appears when the monitor is operated by User 2.
Weak Battery Symbol: Appears when batteries should be
replaced.
Pulse Symbol: Shows the pulse rate per minute.
Important Health Bulletin. See page 22 for more information.
Memory Average: Displays average of last 3 readings.
Risk Category Indicator: Compares readings against blood
pressure guidelines. See page 21 for more information.
If any of the following letters and numbers appear in the area that diastolic pressure should be displayed, an error has occurred with your reading. See Troubleshooting section of this manual for more information.
Measurement Error: Measure again. Wrap the cuff correctly and keep arm steady during measurement.
Measurement Error: Measurement result exceeds measurement range.
Air Circuit Abnormality: Check cuff connection. Measure again.
Pressure Exceeding 300 mmHg: Switch the unit off to
clear, then measure again.
Data Error: Remove and reload the batteries.
10
INSTALLING BATTERIES
Note: It is recommended that you use batteries only
for travelling or when it is not possible to use the AC adapter.
1. Remove the battery pack by pressing down on
latch and pulling away from the monitor.
2. Install or replace 4 “AAA” sized batteries in the
battery pack according to the indications inside the compartment.
3. Replace the battery pack by clicking in the
bottom hook first, then push in the top end of the battery door.
Battery Cover
11
Replace the batteries if:
1. The weak battery symbol appears on the display.
2. Nothing appears on the display when the power is
switched on. As the supplied batteries are for test only, they may be discharged earlier than batteries you buy in stores. Replace the batteries in pairs. Remove batteries when unit is not in use for extended periods of time.
Note: Batteries are hazardous waste. Do not dispose
of them together with the household garbage.
12
USING THE AC ADAPTER
Note: • Please unload the batteries from the battery
pack when operating with the AC adapter for an extended period of time. Leaving the batter­ies in the battery pack for a long time may cause leakage, which may lead to damage of the unit.
• No batteries are needed when operating with the AC adapter.
1. Remove the battery pack by pressing down on the latch and pulling away from the monitor.
2. Connect the AC adapter with the AC adapter jack in the back of the unit as shown below.
3. Please use only the AC adapter included with this monitor.
AC Adapter
Input: 120V 60Hz Output: 12V DC 500mA
Connect AC adapter here
13
DATE & TIME SET PROCEDURE
1. Slide the “On/Off” Switch located on the back of the monitor into the “On” position.
2. To adjust the date/time, press the Set button , located on the back of the monitor.
3. The display will show a blinking number showing the month. Change the month by pressing the button. Each press will increase the number by one in a cycling manner. Press the Set button again to confirm the entry, and the screen will show a blinking number representing the day of the month.
4. Change the date and time as described in Step 2 above, using the button to change the num­bers and the Set button to confirm the entries.
5. A “0” will reappear as the blood pressure monitor is ready for measurement again.
Press the button
blinking month
increasing month
confirm the entry
repeat process to set date and time
14
P o s it ion
cu f f e
dg e
0 .8
1. 2 in ch e s (2
3cm )
above elb
o w .
F it cu f
All ow
to f
an
Ce n
mi
APPLYING THE CUFF
1. If the cuff is not assembled, pass the end of the cuff furthest away from the tubing through the metal D-ring in order to form a loop.
End of cuff
D-ring
2. Plug the cuff tube into the left side of the unit.
Insert plug here
(
0.8” 2 ~3
~ 1
c
.2”)
m
15
3. Remove tight fitting clothing from around your upper left arm.
4. Position cuff edge 0.8-1.2 inches (2-3cm) above elbow.
5. Center tube over middle of arm.
6. Pull the end of the cuff so that it tightens evenly around your arm.
Press the hook and loop material together to secure.
Allow room for 2 fingers to fit between the cuff and your arm.
7. Lay your arm on a table (palm upward) so the cuff is at the same height as your heart. Make sure the tube is not kinked.
16
Note:
P 0 ab
If it is not possible to fit the cuff to your left arm, it
can also be placed on your right arm. However, all
measurements should be made using the same arm.
To use on the right arm, you must position the artery
symbol “ ” over the main artery. Locate the main
artery by pressing with two fingers approximately 1"
(2 cm) above the bend
Main Artery
of your elbow on the
inside of your right arm.
Identify where the pulse
can be felt the strongest.
This is your main artery.
Very Important: Proper cuff size is critical to accurate
measurement. This cuff is suitable for your use if
the arrow falls within the solid color line as shown
below. If the Arrow falls outside of the Solid Color Line,
you will need a cuff with other circumferences. Contact
HoMedics Consumer Relations regarding additional
cuff sizes.
2~3 c
m
(0.8”~1.2”)
AND
FALL
B E U
RROWMUST
WITHINBL
A
Arrow
Position cuff edge 0.8˝
1.2˝
(2
3cm) above elbow.
Center tube over
One Size
middle of arm.
Cu Fits:
8.7-18.1 Inches
22-46 Centimeters
Latex free
Fit
cuff secu
re
l
y.
Allo
w
ro
o
m
f
o
r
tw
o
f
i
ng
e
r
s
t
o
fit be
t
ween
the cuff
an
d
yo
ur arm
.
Model# PP-BPCUFF7
Center tube over middle of arm
Solid Color Line
17
MEASUREMENT PROCEDURE
Important Notes:
• Blood pressure changes with every heartbeat and is in constant fluctuation throughout the day.
• Blood pressure measurement can be affected by the position of the user, his or her physiologic condition and other factors. For greatest accuracy, wait 1 hour after exercising, bathing, eating, drinking beverages with alcohol or caffeine, or smoking to measure blood pressure.
• Before measurement, it is suggested that you sit quietly for 15 minutes as measurements taken during a relaxed state will have greater accuracy. You should not be physically tired or exhausted while taking a measurement.
• During the measurement, do not talk or move your arm or hand muscles.
• See page 25 for additional notes regarding your blood pressure measurement.
If you are using this blood pressure monitor for the first time, please remove the protective film from the screen.
18
1. Press the User Select Button to choose User 1
START
STOP
/
or User 2.
OR
After the user number is selected, press the START/STOP button to confirm the chosen user.
2. With the cuff wrapped around your arm, press the
START/STOP button. Do not inflate the cuff unless it is wrapped around your arm. All digits will light up to check the display functions. The checking procedure will be completed after about 2 seconds.
19
3. After all symbols disappear, the display will
show a blinking “0”. The monitor is “Ready to Measure” and will automatically inflate the cuff to start measurement.
4. Once your systolic pressure is determined, the monitor
will begin deflating to measure your diastolic pressure.
Note: This monitor will re-inflate automatically if the
system detects that your body requires more pressure for measurement.
5. When the measurement is completed, the cuff will
deflate. Systolic pressure, diastolic pressure and pulse will be shown simultaneously on the LCD screen. The measurement is then automatically
stored into memory.
20