Homedics BPA-065 User Manual

WITH SMART MEASURE
TECHNOLOGY
Le manuel en français commence à la page 32
El manual en español empieza en la página 64
BPA-065
TABLE OF CONTENTS
INTRODUCTORY INFORMATION: Important Product Notices and Safety
Instructions ..............................................................3
About Blood Pressure ............................................5
Blood Pressure Standard ....................................... 6
How This Blood Pressure Monitor Works ............. 7
USAGE INFORMATION:
Name/Function of Each Part ..................................8
Display Explanations ..............................................9
Installing Batteries ................................................ 11
Using the AC Adapter (not included)...................12
Date & Time Set Procedure .................................. 13
Using the Arm Cuff ...............................................14
Measurement Procedure ......................................17
Risk Category Index ..............................................19
Irregular Heartbeat Detector (IHB).......................20
Recalling Values from Memory ............................ 21
Clearing Values from Memory ..............................22
Important Notes Regarding Your
Blood Pressure Measurement ..............................23
Care, Maintenance, & Cleaning ............................24
Potential for Electromagnetic Interference .........26
Electromagnetic Compatibility (EMC) .................27
Troubleshooting .................................................... 29
Specifications ........................................................ 30
5-Year Limited Warranty ....................................... 31
IMPORTANT PRODUCT NOTICES AND SAFETY INSTRUCTIONS
When using your blood pressure monitor, basic precautions should always be followed. Please read and follow all instructions and warnings before using this product. Save these instructions for future reference.
• Please note that this is a home health care product only and is not intended to serve as a substitute for the advice of a physician or medical professional.
• This device uses oscillometric method to measure systolic and diastolic blood pressure, as well as heart rate.
• DO NOT use this device for diagnosis or treatment of any health problem or disease. Measurement results are for reference only. Consult a health care professional for interpretation of pressure measurements. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or health care professional.
• Proper cuff size is critical for accurate measurements. Follow the instructions in this manual and printed on the arm cuff to ensure the appropriate size of cuff is being used.
• This device may have difficulty determining the proper blood pressure for pregnant women and for users with irregular heartbeat, diabetes, poor circulation, kidney problems, or for users who have suffered a stroke. This product is not suitable for people with arrhythmias
• Not suitable for people undergoing intravenous injection on any limb and for those with pre-eclampsia.
• Those who have had a mastectomy surgery (especially when lymph nodes have been removed), it is recommended to take measurements on the unaffected side.
• The pulse display is not suitable for checking the frequency of heart pacemakers.
• When used along with other electronic medical equipment on the same arm, pressurization of the cuff may cause the other devices to temporarily malfunction.
• Electromagnetic interference: This device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g. cell phones, microwave ovens). These may lead to temporary impairment of measurement accuracy.
• Consider the electromagnetic compatibility of the device (ex. power disturbance, radio frequency interference etc.) Please use this device in a home health care environment only.
• Use blood pressure monitor only for its intended use.
• DO NOT wrap the cuff around body parts other than your arm.
• Not for use by or on persons under the age of 18.
• DO NOT use this device on infants, children, or those who cannot express their own intention.
• DO NOT plug or unplug the adapter power cord with wet hands.
• Excessive use may result in blood flow interference, which is likely to cause uncomfortable sensations, such as partial subcutaneous hemorrhage, or temporary numbness to your arm. In general, these symptoms should not last long. However, if you continue to experience these sensations, please seek advice from a medical professional.
• Please use only a 6V AC adapter (as specified in this manual) or 1.5V AA alkaline batteries.
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/ stethoscope auscultatory method, within the limits prescribed by the American National Standard, manual, electronic, or automated sphygmomanometers.
ABOUT BLOOD PRESSURE
What is blood pressure?
Blood pressure is the pressure exerted on the artery walls while blood flows through the arteries. The pressure measured when the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the heart dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure.
Why measure your blood pressure?
Among today’s various health problems, those associated with high blood pressure are very common. High blood pressure dangerously correlates with cardiovascular diseases. Therefore, blood pressure monitoring is important for identifying those at risk.
Why do my readings vary?
Blood pressure is a body parameter that is subject to normal variations throughout the day. A single reading that is different from yours or your doctor's readings are not necessary inaccurate. The average of several readings, taken under similar conditions, using the same arm is preferred for accurate blood pressure readings.
Why are my readings different than those taken at my doctor's office?
Many experience a phenomenon called “White Coat Hypertension” when measured by a doctor. White Coat Hypertension refers to blood pressure that rises above its usual level when measured in a clinical setting, such as a doctor's office.
BLOOD PRESSURE STANDARD
The table below contains defined levels for hypertension that are publicly available from the American Heart Association® (AHA 2017).
Users can compare their own blood pressure readings against these defined levels to determine if they may be potentially at increased risk.
This table is applicable to most adults aged 18 and older.
Category Systolic (mmHg) Diastolic (mmHg)
Normal Less than 120 and Less than 80 Prehypertension 120-139 or 80-89
High blood pressure
Hypertension – Stage 1 140-159 or 90-99 Hypertension – Stage 2 160 or higher or 100 or higher Hypertension Crisis –
Consult your doctor immediately
Blood pressure tends to go up and down, even in people who normally don’t have high readings. If your numbers stay above the normal range most of the time, you may be at increased risk and should consult your physician.
Although one can easily find where their own blood pressure readings fall on this table, this monitor comes equipped with a Risk Category Indicator that automatically compares each reading to the defined levels and provides a helpful cue if your reading falls into one of the stages that could potentially indicate increased risk. See page 19 for more information on this feature.
Higher than 180 and/or Higher than 120
Please note that cues provided by this monitor are only intended to assist you in using this table. The table and cues are only provided for convenience in helping you to understand your noninvasive blood pressure reading as it relates to the American Heart Association (AHA 2017) information. They are not a substitute for a medical examination by your physician. It is important for you to consult with your physician regularly. Your physician will tell you your normal blood pressure range as well as the point at which you may actually be considered to be at risk.
HOW THIS BLOOD PRESSURE MONITOR WORKS
This monitor uses Smart Measure™ Technology to detect your blood pressure. This technology enables the monitor to automatically inflate and deflate at the appropriate level for each individual. With the touch of a button, the cuff will automatically inflate to block the blood flow through your artery. Next, the deflation process starts. Please note that any muscle movement during inflation or deflation will cause measurement error. When measurement is complete, the monitor will display your systolic pressure, diastolic pressure, and pulse readings.
The monitor automatically finds where your measurement results fall on the American Heart Association (AHA 2017) table and provides a cue if your reading falls into one of the stages that could potentially indicate increased risk. Please refer to page 19 for more information on this feature.
The appearance of the icon indicates that a pulse irregularly consistent with an irregular heartbeat was detected during measurement. Refer to page 20 for more information on the Irregular Heartbeat Detector.
NAME/FUNCTION OF EACH PART
Arm Cuffs
Air Tube and Connector
LCD Display
Battery Cover (Located on bottom of unit)
Date/Time Set Buttons
Other Accessories:
MEMORY Recall Button
1 6V DC Output AC Adapter (not included) 4 AA size, 1.5V alkaline batteries
Note: Please unload the batteries when operating with the AC adapter for an extended period of time.
User Select Button
START/STOP Button
DISPLAY EXPLANATIONS
Date/Time Indicator
Irregular Heartbeat Detector
PM
Systolic Pressure
Diastolic Pressure
Heart Rate
Display Symbols:
User 1: Appears when the monitor is operated by User 1.
User 2: Appears when the monitor is operated by User 2.
Weak Battery Symbol: Appears when batteries should
be replaced.
Pulse Symbol: Shows the heart rate per minute.
Risk Category Index: See page 19 for more information.
Irregular Heartbeat Detector: See page 20 for
more information.
Memory Average: Displays average of last 3 readings
If
and any of the following letters and numbers appear in the area that systolic pressure should be displayed, an error has occurred with your reading. See the Troubleshooting section of this manual for more information.
Excessive Body Motion Detected/Measurement Error: Wrap the cuff correctly and keep arm steady during measurement. Measure again.
Air Circuit Abnormality: Check cuff connection. Measure again.
Pressure Exceeding 300 mmHg: Turn the unit off to clear, then measure again.
Error Determining Measurement Data: Measure again.
System Error: Measure again.
10
INSTALLING BATTERIES
1. Press down on latch and lift the battery cover to open the
battery compartment.
2. Install or replace 4 AA sized alkaline batteries in the
battery compartment according to the indications inside the compartment.
3. Close the battery cover by pushing in the top end of the
battery door.
Battery Cover
Replace the batteries if:
• The weak battery symbol appears on the display.
• Nothing appears on the display when the power is switched on.
Note:
• Date and time will need to be reset if batteries are removed
or replaced.
• Replace all batteries at one time (as simultaneous set). Use
only 1.5V AA alkaline batteries. Use only the size and type of batteries specified.
• When installing batteries, observe proper +/- polarities. Incorrect
installation of batteries may cause damage to the unit.
• When the batteries are removed, the measurement values
stored in the memory are retained. However, the date and time must be reset.
• Remove batteries when unit is not in use for extended periods of
time to prevent damage due to possible leakage.
• Clean contacts on battery and in battery compartment with a
soft, dry cloth each time you install batteries.
11
• Batteries are hazardous waste. DO NOT dispose of them
together with household garbage.
• DO NOT dispose of batteries in fire. Batteries may explode
or leak.
• Recycle or dispose of properly in accordance with local, state,
province, and country regulations.
USING THE AC ADAPTER (NOT INCLUDED)
This monitor is designed for operation with batteries or an optional AC adapter. Please use only with compatible AC adapter with required voltage and current as indicated in this manual. Contact HoMedics Consumer Relations at 1-800-466-3342 to purchase a compatible AC adapter.
Note:
• When the AC adapter is your main power supply, make sure the
adapter plug can be easily removed from the unit.
• Please unload the batteries when operating with an AC adapter
for an extended period of time. Leaving the batteries in the compartment for a long time may cause leakage, which may lead to damage to the unit.
• No batteries are needed when operating with an AC adapter.
• Date and time will need to be reset if AC adapter is unplugged
and unit is without batteries.
Connect the AC adapter with the AC adapter jack in the back of the unit as shown below.
AC Adapter
Input: 100-240V~0.18A 50/60Hz Output: 6V DC 1A 6W Max
Connect AC Adapter
12
DATE & TIME SET PROCEDURE
1. To adjust the date and time, press the Date/Time set button.
2. The display will show a blinking number showing the MONTH.
Change the MONTH by pressing the + button. Each press will increase the number by one in a cycling manner. Press the Date/Time set button again to confirm the entry, and the screen will show a blinking number representing the DAY.
3. Change the DAY, HOUR AND MINUTE as described in Step
2 above, using the + button to change the numbers and the
Date/Time set button to confirm the entries.
Press the button
blinking MONTH
increasing MONTH
confirm the entry
repeat process to set
DAY, HOUR AND MINUTE
13
USING THE ARM CUFF
Very Important: Proper cuff size is critical to accurate measurement.
This monitor comes with 2 sizes of arm cuffs: Standard: Fits 9"–13" (23 cm – 33 cm) Large: Fits 13"–17" (33 cm – 43 cm)
The appropriate cuff is suitable for your use if the colored arrow falls within the solid color line as shown below.
Colored arrow
Position cuff edge 0.6˝
1.0˝
(1.5
ARROW MUST FALL
WITHIN YELLOW BAND
ARROW MUST FALL
WITHIN YELLOW BAND
Center tube over middle of arm.
Center tube over middle of arm
2.5cm) above elbow.
Fit cuff securely. Allow room for two fingers to fit between the cuff and your arm.
Solid colored line
14
1. If the cuff is not assembled, pass the end of the cuff furthest
Position cuff edge
0.8
1.2 inches (2
3cm)
above elbow.
Fit cuff
Allow
to fi
and
Cen mi
away from the tubing through the metal D-ring in order to form a loop. The smooth side without the felt material should be on the inside of the cuff loop.
ARROW MUST FALL
End of cuff
WITHIN YELLOW BAND
D-ring
2. Plug the cuff tube into the left side of the unit.
Insert plug
(0.8”~1.2”)
2~3 cm
3. Open the arm cuff so that your arm may be placed through it.
4. Remove tight fitting clothing from your upper left arm so that
the cuff can be wrapped around your exposed arm.
5. Position cuff edge 0.8-1.2
inches (2-3 cm) above elbow.
15
6. Center tube over middle of arm.
7. Pull the end of the cuff so that it tightens
evenly around your arm. Press the hook and loop material together to secure. Allow room for 2 fingers to fit between the cuff and your arm.
8. Lay your arm on a table (palm upward)
so the cuff is at the same height as your heart. Make sure the tube is not kinked.
Note:
• The device should not be used when your arm is wounded
or injured.
• If it is not possible to fit the cuff to your left arm, it can also be
placed on your right arm. However, all measurements should be made using the same arm.
• To use on the right arm, you must
position the artery symbol over
Main Artery
the main artery. Locate the main artery by pressing with 2 fingers approximately 1" (2 cm) above the bend of your elbow on the inside of your right arm. Identify where the pulse can be felt the strongest. This is your main artery.
16
MEASUREMENT PROCEDURE
Note:
• Blood pressure changes with every heartbeat and is in
constant fluctuation throughout the day.
• Blood pressure measurement can be affected by the position
of the user, his or her physiologic condition, and other factors.
Before measurement:
To help ensure an accurate reading, follow these instructions BEFORE taking a measurement:
• Wait 1 hour after exercising, bathing, eating, drinking
beverages with alcohol or caffeine, or smoking.
• Sit quietly and rest for 15 minutes.
• Stress raises blood pressure. Avoid taking measurements
during stressful times.
• Take your reading in a comfortable environment as
measurements can be affected by hot or cold temperatures. Take your blood pressure at a normal body temperature.
During measurement:
• DO NOT talk or move your arm or hand muscles.
• DO NOT cross your legs. Sit with feet flat on the floor.
• DO NOT touch cuff or monitor during measurement.
If you are using this blood pressure monitor for the first time, please remove the protective film from the screen.
17
1. Press the User Select button to choose User 1 or User 2.
OR
2. With the cuff wrapped around your arm, press the START/
STOP button. Do not inflate the cuff unless it is wrapped around your arm. All digits will light up to check the display functions. The checking procedure will be completed after about 3 seconds.
3. After all symbols disappear, the display will show “00”. The
monitor is ready to measure and will automatically inflate the cuff to begin measurement. The cuff will then start deflating as measurement continues.
4. When the measurement is completed, the
cuff will deflate entirely. Systolic pressure, diastolic pressure, and pulse will be shown simultaneously on the LCD screen. The
measurement is then automatically stored into memory.
18
PM
Note:
(No Symbol)
• If the monitor detects Excessive Body movement during the
measurement, EE “error” will display on the screen. Keep arm steady and measure again.
• This monitor will reinflate automatically if the system detects
that your body requires more pressure for measurement.
• This monitor automatically turns off approximately 1 minute
after last operation. You may also press the START/STOP button to turn the unit off.
• To interrupt the measurement, you may press the START/
STOP (recommended), MEMORY, User Select or Date/ Time SET button. The cuff will deflate immediately after a
button is pressed.
RISK CATEGORY INDEX
This monitor comes equipped with a Risk Category Index that automatically compares each reading to defined levels established by the American Heart Association (AHA 2017) as described earlier in this manual, and provides a helpful cue if your reading falls into one of the stages that could potentially indicate increased risk. No cue is given if the reading falls in the normal range as defined by AHA. Please note that cues provided by this monitor are only intended to assist you in using this table. The table and cues are only provided for convenience to help you understand your noninvasive blood pressure reading as it relates to the AHA information. They are not a substitute for a medical examination by your physician. It is important for you to consult with your physician regularly. Your physician will tell you your normal blood pressure range as well as the point at which you may actually be considered to be at risk.
Stage 2 Hypertension Elevated Normal (No Symbol)
Stage 1 Hypertension Hypertension Crisis
19
IRREGULAR HEARTBEAT DETECTOR (IHB)
The appearance of the icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement. Usually this is not a cause for concern. However, if the symbol
Irregular Heartbeat Detector
appears often, we recommend that you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
Movement, shaking, or talking during the measurement can result in pulse irregularities that may cause the appearance of this
icon. Therefore, it is of great importance to not move or talk
during measurement.
To determine the presence of an irregular heartbeat, the average of the heartbeat intervals is calculated with the first 3 normal effective heartbeat values. It is important to note that the average is not a strict mathematical averaging of all recorded intervals. At least 3 beats with 25% or greater difference from the average heartbeat interval will generate the icon on the screen.
IMPORTANT INFORMATION:
This blood pressure monitor is not designed for use by people with arrhythmias nor for diagnosing or treating an arrhythmia problem. As a safeguard, we recommend that if you have arrhythmias such as atrial or ventricular premature beats and atrial fibrillation or any other special conditions, you should check with your physician before using your blood pressure monitor.
PM
20
RECALLING VALUES FROM MEMORY
This monitor can be used by 2 individuals. Each user can store
up to 60 measurements.
1. Press the User Select button to select User 1 or User 2.
2. Press the MEMORY button to access the memory.
3. The monitor will first display the calculated average applied to
the last 3 memories.
21
4. Every new press of the MEMORY button will recall a previous
reading. The latest reading will be recalled first.
CLEARING VALUES FROM MEMORY
1. Press the User Select button to select User 1 or User 2.
2. Press and hold the DATE/TIME Set buttons + at the same
time while in memory recall mode, and the data for the selected user will be erased automatically.
and
Note: Once deleted, your readings cannot be restored.
22
IMPORTANT NOTES REGARDING YOUR BLOOD PRESSURE MEASUREMENT
• Take your reading in a comfortable environment as
measurements can be affected by hot or cold temperatures. Take your blood pressure at normal body temperature.
• DO NOT move or talk during measurement as this can elevate
readings.
• DO NOT move or cross legs during measurement. Keep feet
flat on floor.
• DO NOT touch cuff or monitor during measurement procedure.
• It is suggested that you take your measurements at the same
time each day and use the same arm for consistency.
• Users should wait a minimum of 5 minutes before taking
additional measurements. More time may be necessary depending upon your physiology.
• The measurement results that users receive are for
reference only. If users have any blood pressure concerns, please consult a physician.
• Once inflation reaches 300 mmHg, the unit will deflate
automatically for safety reasons.
• This device may have difficulty determining the proper blood
pressure for pregnant women and for users with irregular heartbeat, diabetes, poor circulation of blood, kidney problems, or for users who have suffered a stroke.
23
CARE, MAINTENANCE, & CLEANING
• Clean the blood pressure monitor body and cuff carefully with
a slightly damp, soft cloth. Do not press. Do not wash cuff or use chemical cleaner on it. Never use thinner, alcohol, or petrol (gasoline) as cleaner.
• Leaky batteries can damage the unit. Remove the batteries
when the unit will not be used for a long time.
• Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components, including batteries.
• If the unit is stored near freezing, allow it to acclimate to room
temperature before use.
• This blood pressure monitor is not field serviceable. You should
not use any tool to open the device nor should you attempt to adjust anything inside the device. If you have any problems with this device, please contact HoMedics Consumer Relations (contact information can be found in the Warranty section).
• DO NOT immerse the unit in water as this will result in damage
to the unit.
• DO NOT subject the monitor or cuff to extreme temperatures,
humidity, moisture, or direct sunlight. Protect from dust.
• DO NOT fold the cuff and tubing tightly.
• DO NOT disassemble the monitor or cuff. If in need of repair,
refer to the Warranty section of this manual.
• DO NOT subject the monitor to extreme shocks (do not drop
on floor).
• DO NOT inflate the cuff unless wrapped around arm.
• DO NOT wrap the cuff around body parts other than your arm.
• DO NOT drop or insert any object into any opening or hose.
24
• To avoid accidental strangulation, keep this product away
from children. Do not drape tube around neck.
• This monitor may not meet its performance specifications if
stored or used outside of these temperature and humidity ranges:
Storage/Transportation • Operating
Environment Environment
Temperature:
-13°F~158°F (-25°C~70°C)
Temperature: 41°F~104°F (5°C ~40°C)
Humidity: less than 93% RH
Humidity: 15% ~ 93% RH
25
POTENTIAL FOR ELECTROMAGNETIC INTERFERENCE
To avoid inaccurate results caused by electromagnetic interference between electrical and electronic equipment, do not use the device near a cell phone or microwave oven. For most wireless communication devices, it is recommended to maintain a distance of 10.8 feet (3.3m) in order to avoid electromagnetic interference.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
FEDERAL COMMUNICATIONS COMMISSION COMPLIANCE STATEMENT
Changes or modifications to this equipment not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
26
ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions IEC 61000-3-3
• Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Immunity test IEC 60601 test level Compliance level Electromagnetic environment–
Electrostatic discharge (ESD) IEC 61000-4-2
Power frequency (50 or 60 Hz) magnetic field IEC 61000-4-8
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Interruptions and voltage variations on power supply input lines
IEC 61000-4-11
± 8 kV contact discharge
± 15 kV air discharge
30 A/m 50 or 60 Hz
± 2 kV for power supply
± 1 kV for input/output lines
± 2 kV Power lines ± 2 kV Power lines Mains power quality should be that
0% UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT; 1 cycles 70% UT; 25/30 cycles 0% UT; 250/300 cycles
Group 1 RF energy is used only to maintain device’s operation. Therefore,
Class B
Class A
Complies
its RF emissions are so low that it’s not likely to cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments, including domestic establishments, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
guidance
lines
± 8 kV contact discharge
± 15 kV air discharge
3 A/m 50 or 60Hz
± 2 kV for power supply
lines
± 1 kV for input/output
lines
0% UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT; 1 cycles 70% UT; 25 cycles 0% UT; 250 cycles
In air discharge testing, the climatic conditions shall bw within the following rages: Temperature:15˚~35˚, Humidity:30%~60%.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
of a typical commercial or hospital environment..
Mains power quality should be that of a typical commercial or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.
27
• Recommended separation distances between portable and mobile RF communication equipment
and the device.
The device is intended for use in an electromagnetic environment where radiated RF disturbances are
under control. User can help prevent electromagnetic interference by keeping the device at a minimum distance from portable and mobile RF communications equipment (transmitters). The table below details the maximum output power of transmit ter:
Rated maximum output power of
transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For tran smitters ra ted at a maximu m output power n ot listed above , the recommen ded separati on distance d i n metres (m) can be estim ated using the e quation appl icable to the fr equency of the t ransmitte r, where P is the max imum output po wer rating of th e transm itter in wat ts (W) acc ording to the tra nsmitter ma nufacture r. NOTE 1 At 80 M Hz and 800 MHz, t he separatio n distance fo r the higher fre quency range a pplies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
• Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments:
150 kHz to 80 MHz
Immunity test IEC 60601 test level Compliance
Conducted RF IEC 61000-4-6
Radiat ed RF IEC 61000-4-3 (Proxim ity fields from RF wireless communications equipment IEC 61000- 4-3)
NOTE 1 At 80 M Hz and 800 MHz, th e higher freque ncy range appl ies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec­tion from structures, objects and people.
a. Field s trengths fro m fixed transmi tters, such a s base statio ns for radio (cellu lar/cordless) te lephones and la nd mobile radios , amateur radio, A M and FM radio broa dcast and TV br oadcast can not be predicte d theoretica lly with accur acy. To asses s the electrom agnetic envir onment due to fix ed RF transmit ters, an elec tromagnetic si te survey shou ld be considere d. If the mea sured field st rength in the loc ation in which t he device is use d exceeds the appl icable RF comp liance level ab ove, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device. b. Over the f requency ran ge 150 kHz to 80 MHz, fie ld strengths s hould be less tha n 3 V/m.
3V rms At 0.15-80 MH z
6V rms At ISM & Rad io Amateur Freq.
10 V/m at 80-2 700 MHz AM Modul ation And 9 -28V/ m at 385 -6000 MHz,Pulse Mode and other Modula­tion. The system shall be teste d as specifie d in IEC60 601-1-2 Table 9 for proximi ty fields fr om RF wireless communications equipm ent using the te st methods specified in IEC 61000-4-3
Separation distance according to frequency of transmitter m
d = 1.2
3V rms At 0.15-80 MH z
6V rms At ISM & Rad io Amateur Freq.
10 V/m at 80-2 700 MHz AM Modul ation And 9 -28V/ m at 385 -6000 MHz,Pulse Mode and other Modula­tion. The system shall be teste d as specifie d in IEC60 601-1-2 Table 9 for proximi ty fields fr om RF wireless communications equipm ent using the te st methods specified in IEC 61000-4-3
80 MHz to 800 MHz
d = 1.2
level
Electromagnetic environment –
Portable and mobile RF communications equipm ent should be us ed no closer to any part of the device, including cables, than the recommended separation distance calculated from the e quation appl icable to the fr equency of the transmitter.
Recommended separation distance
d = 1.2 d = 1.2 8 0 MHz to 800 MHz d = 2 .3 800 MHz t o 2.7 GHz
where P is t he maximum ou tput power rat ing of the tra nsmitter in w atts (W) a ccording to the tran smitter man ufacturer a nd d is the recommended separation distance in metres (m). Field st rengths fro m fixed RF trans mitters, a s determ ined by an elect romagnetic s ite survey, a should b e less than the co mpliance lev el in each frequency range.b Inter ference may oc cur in the vicin ity of equipm ent marked wit h the following s ymbol:
800 MHz to 2.7 GHz
d = 2.3
guidance
28
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
SYMPTOMS
Unit does not turn on when the START/STOP button is pushed.
POSSIBLE
CAUSES
Batteries have run down. Replace them with
Battery polarities have been positioned incorrectly.
CORRECTION
four new AA alkaline batteries.
Reinsert the batteries in the correct positions.
EE measurement error symbol shown on display or the blood pressure value is displayed excessively low (or high).
E1 error symbol shown on display.
E2 error symbol shown on display.
E3 error symbol shown on display.
EP error symbol shown on display.
Cuff has been placed incorrectly.
Did you talk or move during measurement?
Shaking of the arm with the cuff on.
Air circuit abnormality. Cuff tube may not be plugged into monitor correctly.
Inflation pressure exceeding 300 mmHg.
Error determining measurement data.
System error. Measure again.
Wrap the cuff properly so that it is positioned correctly.
Keep arm steady during measurement. Measure again.
Check cuff connection. Measure again.
Switch the unit off, then measure again.
Rewrap the cuff properly so that it is positioned correctly. Take new measurement.
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found in the Warranty section.
29
Loading...
+ 67 hidden pages