Information in this Technical Reference Manual is subject to change without notice due to
correction or enhancement. The information described in this manual is solely the property of
GSE. No part of this manual may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopying and recording and sold for any monetary figure
without the express written permission of GSE.
GSE LOCATIONS
GSE Scale Systems
42860 Nine Mile Road
Novi, MI 48375
U.S.A.
Phone: (800) 755-7875
www.gse-inc.com
GSE Canada, Inc.
617 East Lake Road
Airdrie, Alberta Canada T4B 2B8
Phone:(403) 948-9921
Fax: (403) 948-1449
Model 370............................................................................................................................................... 2
MODEL 375........................................................................................................................................... 3
Both ...................................................................................................................................................... 45
First Zero.............................................................................................................................................. 62
Last Zero............................................................................................................................................... 63
Only Zero.............................................................................................................................................. 66
This chapter describes the components of the Model 370/375 such as keypad, e nclosure and
display. Also included is wiring for the load cell and communication port.
EEnncclloossuurree
The enclosure is made of die cast aluminum with rib enforcement that includes built in carrying
handles and a line cord wrap for ease of portability. This design provides extra strength and
durability. The enclosure is painted with powder coat and the counting surface is stainless steel.
LLooaadd CCeellll RReeppllaacceemmeenntt
Figure 1-1: Model 370 and Model 375 Enclosure
The load cell can be exchanged by removing two M6 1x 20mm (size) bolts from the bottom plate
and two M6 1 x 16mm (size) bolts from the top spider assembly.
Install the new load cell by placing it in the load cell cavity. Fasten the load cell by installing the
bolts on the bottom plate. Reinstall the spider assembly. NOTE: When exchanging a load cell
with another, the overload stop and corner overloads must be reset. Refer to Table 1-1 for
setting values.
Table 1-1: Load Cell Overload Settings
Load Cell Stop Set to
Load Cell Center Overload 120 % of full scale
Corner Overloads 60 % of full scale
LOAD CELL CONNECTIONS
The load cell cable connects to J10 of the main board. The J10 connector is accessible from the
load cell cavity of the Model 370/375 enclosure.
Table 1-2: Load Cell Connection to Main Board
Pin Designation Function
1 Shield
2 - Sense
3 + Sense
4 - Signal
5 + Signal
6 - Excitation
7 + Excitation
KKeeyyppaadd
All of the keys perform different functions. Some keys have more than one function.
MODEL 370
The Model 370 offers a 6-button durable and versatile Polymeric switch plate with large keys for
ease of use. The keypad is easily cleaned with a damp cloth or non-abrasive cleaner. Each key is
assigned two distinct functions. Various key combinations are also used. Each key has secondary
functions; allowing an operator to enter target values, perform piece samples, access setup
parameters, etc.
Figure 1-4: Model 370 Keypad
Functions
The Model 370 keypad performs different functions in the Weigh Mode, the Setup Mode, and the
Calibration Mode. Secondary functions for each key allow you to perform additional tasks.
2
Key Press Weigh Mode Count Mode Setup Mode
Performs a gross zero
<
=
:
>
9
<>
:>
<:
<=
function and/or clears an
entry in progress. Hold
this key on power-up to
turn on the display
regardless of P420.
Performs a print function
and/or ‘scrolls’ through
digits during setpoint
entry.
Toggles between ‘lb’, ‘kg’
third unit (if enabled)
and/or advances cursor to
next entry position.
Performs an auto-tare
function (if enabled)
and/or accepts an entry in
progress.
Toggles between display
modes and/or restores
power to the Model
370/375 (if auto-shutoff
enabled).
Goes to the count mode
Access Setup Mode.
Break out of remote
display mode.
No function. No function.
Absolute clear – clears an
entry in progress and/or
clears the value of a
parameter.
Backspace – erases the
right-most digit during
data entry.
Performs a quantity zero
function and/or clears an
entry in progress.
Performs a print function
and/or ‘scrolls’ through
digits during setpoint
entry.
Toggles through standard
sample sizes during a
sample and/or begins a
new sample entry.
Performs an auto-tare
function and/or accepts an
entry in progress.
Toggles between display
modes and/or restores
power to the Model
370/375 (if auto-shutoff
enabled) and/or toggles
through standard sample
sizes during a sample.
Performs a sample for an
accurate part count and
calculates an average
piece weight.
Access Setup Mode.
Break out of remote
display mode.
No function. Clears any entry in progress.
Backspace – erases the
right-most digit during
sample entry.
Exits the Setup Mode and/or
answers “NO” to query
prompts and/or clears an entry
in progress.
‘Scrolls’ through digits during
data entry.
Advances cursor to next entry
position and/or cycles prompts.
Accepts an entry in progress
and/or ‘scrolls’ through
parameter sub-set selections
and/or answers ‘YES’ to query
prompts.
Advances to the next setup
parameter.
Accepts an entry in progress
and/or ‘scrolls’ through
parameter sub-set selections
and/or answers ‘YES’ to query
prompts.
No function.
Return to the previous setup
parameter.
Backspace – erases right-most
digit during data entry.
The Model 375 offers an 18-button durable and versatile Polymeric switch plate with large keys
for ease of use. The keypad is easily cleaned with a damp cloth or non-abrasive cleaner. The
number keys make entering a tare value or average piece weight easier. Each key is assigned at
least one distinct function.
3
Figure 1-5: Model 375 Keypad
Functions
All of the keys perform different functions. Some keys have more than one function.
Key Press Weigh Mode Count Mode Setup Mode
<
=
:
>
J-H,
I
K
9
<>
Performs a gross zero
function and/or clears an
entry in progress. Hold
this key on power-up to
turn on the display
regardless of P420.
Performs a print function
and/or ‘scrolls’ through
digits during setpoint or
time/date entry.
Toggles between ‘lb’, ‘kg’
or third unit (if enabled).
Performs an auto-tare
function (if enabled).
Toggles between display
modes and/or restores
power to the Model
370/375 (if auto-shutoff
enabled).
Numeric keys used to
manually enter a value for
tare weight or target
value.
Absolute clear – clears an
entry in progress and/or
clears the value of a
parameter.
Enters the counting mode
and/or accepts an entry in
progress.
Access Setup Mode.
Break out of remote
display mode.
Performs a quantity zero
function and/or clears an entry
in progress.
Performs a print function.
Toggles through standard
sample sizes.
Performs an auto-tare function
and requests a piece sample
and/or accepts an entry in
progress.
Toggles between display
modes and/or restores power
to the Model 370/375 (if autoshutoff enabled) and/or toggles
through standard sample sizes
during a sample.
Numeric keys used to
manually enter a value for tare
weight, average piece weight,
sample size etc.
Clears any entry in progress or
abort a sample.
Performs a sample for an
accurate part count and
calculates an average piece
weight and/or accepts an entry
in progress.
Access Setup Mode. Break
out of remote display mode.
Exits the Setup Mode
and/or answers “NO” to
query prompts and/or
clears an entry in
progress.
Return to the previous
setup parameter.
Shows the current
parameter number
Accepts an entry in
progress and/or ‘scrolls’
through parameter sub-set
selections and/or answers
‘YES’ to query prompts.
Advances to the next
setup parameter.
Numeric keys used to
manually enter a value.
Clears any entry in
progress.
Accepts an entry in
progress and/or ‘scrolls’
through parameter sub-set
selections and/or answers
‘YES’ to query prompts.
No function.
4
DDiissppllaayy
The Model 370 and Model 375 come standard with a six digit, 7-segment bright green LED
display. The Model 370 and Model 375 will display alphanumeric data, but due to the nature of 7segment LED and the limitation of six digits, some information is abbreviated.
The display also has 11 annunciators to show weight and status information. The SP1, SP2, and
SP3 annunciators are red, green, and yellow. Also there is an annunciator for a third unit under
kg. Place the third unit label above the third Annunciator (refer to Figure 1-6). See page 24 for
third unit setup.
All segments and annunciators are illuminated for a brief display test upon power up. The current
gross weight is then displayed in default units.
Third Unit
Figure 1-6: Model 370 and Model 375 LED Display
ANNUNCIATORS
Annunciators provide mode and status information. When illuminated, they indicate the following
conditions:
MOTION
GROSS Displayed value represents the current gross weight
NET Displayed value represents the current net weight
QTY Displayed value represents the current piece quantity (Count)
lb The displayed value is represented in pounds
kg The displayed value is represented in kilograms
Oz, lb oz, g The displayed value is represented in either ounces, pound ounces or grams
RReeaarr PPaanneell CCoonnnneeccttiioonnss
AC Power Com 1 Options Future Option
Displayed weight is at center-of-zero (± ¼ display graduation)
Scale is in motion. Motion inhibited transmits and motion inhibited setpoint activation will be delayed
until motion ceases
COM 1
PLACE 370/375 BARCODE HERE
POWER (15W MAX)
50-60 Hz 12-20 VAC
DC 12-36V
1 NONE
6 GND
2 RXD
7 RTS
3 TXD
8 CTS
4 *
9 +5V
5 GND
5 GND
MOVE JUMBER E6
FOR +5V ON PIN 4
BATTERY
ON/OFF
www.gse-inc.com Novi, MI USA
Figure 1-7: Rear Panel
AUX 2
AUX 3
ETHERNET
5
COMMUNICATION PORT 1
Comm1 is standard with the Model 370/375. Connector pin outs for Comm 1 (DB9) are provided
in Table 1-3 below.
It is possible to connect a Model 370/375 to another 300 Series or 60 Series indicator and ha ve
the remote indicator display a copy of the master indicator or customize the display of the remote
(slave) indicator. The remote (slave) indicator’s keypad will operate the master indicator. Refer to
page 55 to configure the master and remote (slave) displays. There are several different
configurations for Master to remote (slave) indicator connection and setup. Refer to Table 1-3 for
comm port connections.
Supply Voltage Jumper (E6)
This allows for a choice in scanner power supply pins and to minimize modifications to a cable or
connector.
Comm port 1 has two pins on the DB 9 connector which provide + 5 volts to supply power to a
scanner. The jumpers located on E6 of the main board will reroute the power (+ 5 volt) to either
pin 9 or pin 4 depending on jumper orientation.
Pin 9 (Default) Pin 4
The power (+ 5 volt) and remote key pins are switched. Refer to Figure 1-8 for the jumper
location on the main board. Refer to the Barcode Scanner manual for the power pin location.
Figure 1-8: Comm Port 1 Supply Voltage Jumper Location
6
Chapter 2: O
This chapter will provide detailed instructions on installing each option. Additional hardware may
be needed for the Model 370/375 depending on which options you will be installing.
Also this chapter provides instructions on how to configure all Model 370 / 375 options. This
includes setup mode parameters, connections, calibration and testing.
PTION INSTALLATION
CCoommmmuunniiccaattiioonn
RS-485 NETWORKING
The module converts the standard RS-232 communication on comm port 1 to RS-485. However
the advantage of using the RS-485 module, aside from the ability to transmit over long distances,
is the ability to network multiple indicators or parts counters using the same communication wires.
When networking indicators or parts counter, it is necessary to set up a network address for each
scale. The network module itself does not require addressing, rather each indicator or pa rts
counter must be enabled for network addressing in the setup mode. Refer to page 26 for details
on the RS-485 enable parameter (P250) and the network address parameter (P251).
Installation Instructions
1. DISCONNECT POWER! UNPLUG THE MODEL 370/375 TO INSURE DAMAGE WILL NOT
OCCUR DURING OPTION INSTALLATION.
2. Remove the six 8 mm screws (size) from the bottom plate. Separate the top enclosure from
the bottom plate.
3. Remove the IC chip and white jumper from the U4 socket on the main board.
4. Snap in the plastic spacers into the three mounting holes surrounding the U4 socket. Refer
to Figure 2-1.
5. Gently press the option board into the socket and make sure the board snaps onto the
standoffs.
6. Reinstall the enclosure bottom plate.
Figure 2-1: RS-485 / 20 mA Option Installation
RS-485 Connections (Comm Port 1)
The Model 370/375 will be connected to a peripheral via the DB 9 connector on comm port 1.
Refer to Table 2-1 for wiring connections.
Installing jumpers 1, 2 and 4 on the RS-485 option board electrically connects pin RX B(+) to pin
TX B(+), and pin RX A(-) to pin TX A(-) on the option board. This effectively provides two + and
two - pin connections, enabling easy connection of network lines in parallel from device to d evice
without having to position two wires into the same lever socket. A B(+) line from each device on
the network should be connected in parallel to the next device on the network. This is also true for
all A(-) lines.
The units inside the two end-points of the network loop will utilize both A(-) pin connections and
both B(+) pin connections. The units at the end-points of the network will utilize only one A(-) pin
connection and one B (+) pin connection.
Full Duplex (4-wire)
Removing jumpers 1, 2 and 4 on the RS-485 option board requires that the transmit and receive
lines be wired independently of one another. The RX B(+) and RX A(-) receive lines must be
wired in parallel to the next device's RX B(+) and RX A(-) receive lines and the TX B(+) and TX A
(-) transmit lines must be wired in parallel to the next device's TX B(+) and TX A(-) transmit lines.
8
In order to connect network lines in parallel from device to device it is necessary to position two
wires into the same lever socket. This requires that the wire used to build the network be 24 AWG
or smaller to allow both wires to fit into the same lever socket.
Both Half Duplex and Full Duplex
The network boards on both end-points should install jumper 3 on the RS-485 option board to
engage the 120 W termination resistor (R8). The boards between the two en d-points should
remove jumper 3 on the RS-485 option board.
The isolated ground (ISO GND) should be connected in parallel from unit to unit. A shielded
twisted two pair cable is recommended throughout the network.
20 MA CURRENT LOOP OPTION
This option will convert the comm port 1 to a 20 mA current loop instead of an RS-232. This is a
digital communication signal and should not be confused with a 4 to 20-mA (or 0-20 mA), which
are analog output signals. The intended use is primarily printers and scoreboard displays.
This option will mount into the socket for the comm port RS-232 chip on the main board. This
option will exclude the use of the RS-485 option board. (I.e. only the RS-485 OR the 20 mA
option can be installed into a Model 370/375).
Installation Instructions
1. DISCONNECT POWER! UNPLUG THE MODEL 370/375 TO INSURE DAMAGE WILL
NOT OCCUR DURING OPTION INSTALLATION.
2. Remove the six 8 mm screws (size) from the bottom plate. Separate the top enclosure
from the bottom plate.
9
3. Remove U4 and the white jumper from the socket of the main board.
4. Snap in the plastic spacers into the three mounting holes surrounding the U4 socket.
Refer to Figure 2-1.
5. Gently press the option board into the socket and make sure the board snaps onto the
standoffs.
6. Reinstall the enclosure bottom plate.
BI-Directional
Both the transmit output and the receive input of the Model 370/375 are available as 20 mA
signals. None of the handshake signals are supported for the 20-mA current loop operation.
Baud
Only baud rates of 9600 baud and less are supported.
Active/Passive
The Tx output may be used as an active or passive output from the Model 370/375. Either active
or passive is chosen depending upon which terminals are used for the connections. In active
mode the Model 370/375 supplies the current. In passive mode, the external device supplies the
current. The Rx input is available in passive mode only.
Isolation
The input and output are electrically isolated from the main board as well as earth ground and
each other, for both passive and active modes of operation. Isolation is a minimum of 1000v.
Max Voltage
Active mode Tx current loop provides a driving voltage of 12v. This will allow 20 mA current flow
with up to a 600 ohm load. Passive mode will work with an external driving voltage of up to 50v.
Connections
The field connections to the 20 mA circuitry will be made at the 9 pin 'D' connector on the rear of
the Model 370/375.
Transmit Current Input Active = TA
Transmit Current Input Passive = TP
Transmit Output = TXO
Receive Current Input = RXI
Receive Current Output = RX
Comm port 1 pin out Connection
1 No connection
2 RX IN
3 RX Pass
4 VCC
5 TX Active
6 Ground
7 TX OUT
8 TX Pass
9 VCC
Cable
The length for the current loop is 1000 ft. maximum. This is for the entire loop, not from device to
device. Example: 1 transmitter and 1 receiver can have a maximum of 500 ft. of 2 conductor
cable between them.
10
Connected Devices
While 20 mA current loops can allow for more than one transmitter and/or receiver, the Model
370/375 and/or option board do not include any address recognition or collision avoidance and/or
detection to promote this usage. If the 20-mA loop is intended to be used in this manner, proper
planning for these issues is required.
Table 2-2: Connecting to External Devices
Typical Installations
Model 370/375 External Device
Passive 20 mA Output
Passive 20 mA Input
Active 20 mA Output
TP RX+
TXO RXRXI RX+
RX RXTXO RX+
TA RX-
Active 20 mA Input
Active 20 mA Output
Passive 20 mA Input
AAnnaalloogg OOuuttppuutt OOppttiioonn
This section provides procedures for installing and configuring the analog output module.
The analog output modules enable the Model 370/375 to generate a 0-10VDC, active 0-20mA or
active 4-20mA output signal corresponding to the value of most operating parameters. For
parameter setup see page 35. For details on testing and troubleshooting see page 77. For
analog output calibration see page 71. Requires an option mounting bracket kit (24370B-300A0)
for internal installation.
Installation Instructions
1. DISCONNECT POWER! UNPLUG THE MODEL 370/375 TO INSURE DAMAGE WILL NOT
OCCUR DURING OPTION INSTALLATION
2. Remove the (6) 38-31-8710 M5 x 0.8 x 10 mm screws from the enclosure bottom plate and
set it aside.
3. Discard the mounting hardware provided with the analog output kit and use the hardware
provided with the option-mounting bracket kit. Refer to page 13 for parts included with this
kit.
4. Snap in the (4) nylon spacers in the 4 holes either to the right of the notch or the left of the
notch in the option mounting bracket.
5. Position the Analog Output Option Board on the spacers so the ribbon cable points away
from the notch on the option-mounting bracket. Carefully press the analog output option onto
the spacers.
6. Install the (4) nylon hex nuts to secure the option board. Refer to Figure 2-3 for details.
7. If a setpoint option is also being used, connect the 6" option ribbon cable from J1 of the
analog output option board to J1 of the setpoint option board.
8. Connect the 22-30-25520 6.5" ribbon cable (optional) to from J2 of the analog output board to
J3 of the main board.
9. Install the option-mounting bracket in the Model 370/375 enclosure. Refer to page 13 for
option mounting bracket installation instructions.
10. Connect all necessary wiring to the Analog Output Option Board.
11. Reinstall the enclosure bottom plate.
11
Figure 2-2: Model 370/375 Analog Output and Setpoint Option Installation
Analog Output Connections
Pin Connection Name Description
1 Earth Ground Non-isolated earth ground (future use).
2 + 5 VDC Isolated 5 volt source (future use).
3 Isolated Ground Provides an isolated ground connection.
4 Vout Used for the 0-10 VDC analog signal output.
5 Isolated Ground Provides an isolated ground connection.
6 Iout Used for 4-20 or 0-20 mA analog signal output.
SSeettppooiinntt OOppttiioonn
The setpoint option provides the ability to running a process, turning on stack lights or anything
else that may require an output signal. Eight different configurations are offered. Requires an
option mounting bracket kit (24370B-300A0) for internal installation.
Table 2-3: Setpoint Option Board Part Numbers
Quantity Part Number Description
1 420925-36594 DC 4 output
1 420926-36589 DC 4 input
1 420924-36584 CD 2/in - 2/out
1 420923-36574 Low voltage AC, 4 input
1 420918-36536 Low voltage AC, 2/in - 2/out
1 420922-36579 AC 4 output
1 420923-37093 High voltage AC, 4 input
1 420918-37092 High voltage AC, 2/in - 2/out
Installation Instructions
1. DISCONNECT POWER! UNPLUG THE MODEL 370/375 TO INSURE DAMAGE WILL NOT
OCCUR DURING OPTION INSTALLATION
2. Remove the (6) 38-31-8710 M5 x 0.8 x 10 mm screws from the enclosure bottom plate and
set it aside.
3. Discard the mounting hardware provided with the setpoint option kit and use the hardware
provided with the option-mounting bracket kit. Refer to page 13 for parts included with this
kit.
4. Snap in the (4) nylon spacers in the 4 holes either to the right of the notch or the left of the
notch in the option mounting bracket.
12
5. Position the setpoint option board on the spacers so the ribbon cable points away from the
notch on the option-mounting bracket. Carefully press the analog output option onto the
spacers.
6. Install the (4) nylon hex nuts to secure the option board. Refer to Figure 2-3 for details.
7. If an analog output option is also being used, connect the 6" option ribbon cable from J1 of
the setpoint option board to J1 of the analog output option board.
8. Connect the 22-30-25520 6.5" ribbon cable (optional) to from J2 of the setpoint board to J3
of the main board.
9. Install the option-mounting bracket in the Model 370/375 enclosure. Refer to page 13 for
10. Connect all necessary wiring to the setpoint option board.
11. Reinstall the enclosure bottom plate.
SETPOINT CARD CONNECTIONS
Using one of the software setpoint configurations (see General Setpoint Setup on page 36) in
conjunction with the setpoint option board gives the Model 370/375 the ability to directly control
external devices such as valves, relays, actuators, etc.
There are up to three setpoint outputs available. The activation and deactivation is controlled by
the setpoint configuration. The outputs are capable of driving up to one Amp at 20-280VAC & 2
Amp at 3-60VDC. The solid state relays are normally open (NO) contacts.
OOppttiioonn MMoouunnttiinngg BBrraacckkeett
The option mounting bracket kit (24370B-300A0) is required for option installation. The bracket
will accommodate up to four options. The options installed on the option-mounting bracket are the
Analog Output and Setpoint. The options mount on the bracket with the hardware provided with
each option kit.
1. DISCONNECT POWER! UNPLUG THE MODEL 370/375 TO INSURE DAMAGE WILL NOT
OCCUR DURING OPTION INSTALLATION.
2. Install all options on the mounting bracket. Refer to the specific option section for installation
instructions.
3. Remove the (6) 38-31-8710 M5 x 0.8 x 10 mm screws from the enclosure bottom plate and
set it aside.
4. Use the (2) M5 self-tapping screws to fasten the option-mounting bracket to the Model
370/375 enclosure. For the installation drawing refer to Figure 2-3.
5. Reinstall the enclosure bottom plate.
13
Figure 2-3: Option Bracket Installation
14
Chapter 3: S
This chapter will cover the configuration of the scale parameters and how to configure options
such as analog output, setpoint and communication modules.
To prevent accidental changes to the Model 370/375 Setup, a sequence of keystrokes is used to
gain access to the Setup Mode. If the Model 370 is set for remote display, unplug the unit and
press [ZERO] + [SELECT] upon power up. Follow the instructions below to a ccess the setu p
mode.
To prevent accidental changes to the Model 370/375 Setup, a sequence of keystrokes is used to
gain access to the Setup Mode. If the Model 375 is set for remote display, unplug the unit and
press [ZERO] + [SELECT] upon power up. Follow the instructions below to a ccess the setu p
mode.
100 [SELECT]
23640 [SAMPLE/Enter]
These keystrokes must be made within five seconds, or the Model 375 will return to the Weigh
Mode.
To
access the Setup Mode:
1. From the Weigh Mode, key in 100 [SELECT] DISPLAY READSSetup ~ Enter Code
2. Key in 23640 [SAMPLE/Enter]
DISPLAY READS Chgs ~ Poss!
P110.- - ~ F.S.= ~ 100
To
access Setup in a view-only mode:
1. From the Weigh Mode, key in 100 [SELECT] DISPLAY READSSetup ~ Enter Code
2. Press [SAMPLE/Enter]
DISPLAY READS No ~ Chgs
P110.- - ~ F.S.= ~ 100
SSeelleeccttiinngg aa PPaarraammeetteerr
To advance to the next parameter (Model 370 and Model 375):
Selection parameters have a pre-defined list of choices to pick from. Each choice is numbered
and corresponds to a certain value. The choice number is shown to the right of the decimal point
within the parameter number. Repeatedly pressing [TARE] (Model 370) or [SAMPLE/Enter]
(Model 375) while viewing a selection parameter cycles through the available choices, or you can
key in the choice number.
P 200.00
Indicates Parameter Parameter Number Choice Number
For example, parameter 200 (P200) is a selection parameter that holds the baud rate for the
serial port. This is a selection parameter because a choice number between 00 and 12 must be
used. Each choice number corresponds to a different baud rate. To change the baud rate from
the default value of 9600 to 4800, perform the following steps from the Setup Mode.
To
change the baud rate from the default value of 9600 to 4800:
Key-In Parameters are not limited to a list of choices, although there may be upper and lower
value limits. A Key-In Parameter requires that a numeric value be entered using the front panel
keys. Key-In Parameters are shown with two hyphens after a decimal point within the parameter
number.
P 110.--
Indicates Parameter Parameter Number Denotes “Key In”
To
enter a Key-In Parameter (370):
1. Press [PRINT]. A decimal point is used to represent the entry position.
2. Press [PRINT] until the desired character appears.
3. Press [UNITS]. Another decimal point indicates the next entry position.
4. Repeat steps 1 and 2 until your desired entry value is shown.
5. Press [TARE] to enter your numerical value.
To
enter a Key-In Parameter (375):
1. Key in the number by using the numeric keys.
2. Press [SELECT] to accept the entry.
To
setup a full scale value of 250 lbs (370):
1. Access the setup mode.
DISPLAY READSP11 0.- - ~ F.S.= ~ 100
2. Press [PRINT] four times to select the first digit.
DISPLAY READS2
3. Press [UNITS] to advance to the next digit.
DISPLAY READS 2.
4. Press [PRINT] six times to select the next digit.
DISPLAY READS25
5. Press [UNITS] to advance to the next digit.
DISPLAY READS25.
6. Press [PRINT] once to select the next digit.
DISPLAY READS250
7. Press [TARE] to enter the value.
DISPLAY READSP11 0.- - ~ F.S.= ~ 250
To
setup a full scale value of 250 lbs (375):
1. Access the setup mode.
DISPLAY READSP110.- - ~ F.S.= ~ 100
2. Key in 250 [SELECT] to accept the entry.
DISPLAY READSP11 0.- - ~ F.S.= ~ 250
18
SSaavviinngg PPaarraammeetteerrss
To exit the Setup Mode and save changes (370):
1. Press [ZERO] to begin exiting Setup Mode. DISPLAY READS Enter ~ =CAL!
5. Press [SAMPLE/Enter] to complete exit.
DISPLAY READS 0.00
FFaaccttoorryy DDeeffaauulltt
Parameter 65001 and 65002 are available to return the Model 370/375 to factory settings.
Parameter 65001 will reset parameters including the calibration, while parameter 65002 resets all
parameters except the calibration.
After a factory default is performed and saved the Model 370/375 will be in the Quantity mode.
The display will show ‘do APS’ which means an average piece weight needs to be established.
Either perform a sample (refer to page 50) or press [SELECT] to chose another mode.
Return
1. Access the setup mode. See page 15.
2. Key in 65001 or 65002 [SELECT]
3. Press [TARE] once.
4. Press [TARE] once.
5. Press [ZERO] [ZERO] [TARE] [TARE]
Return
1. Access the setup mode. See page 15.
2. Key in 65001 or 65002 [SELECT]
to factory default (370):
DISPLAY READS P65002~default-Cal
DISPLAY READS Are U sure?~default=Enter
DISPLAY READS Setup done~P65002~default-Cal
DISPLAY READS do APS
to factory default (375):
DISPLAY READS P65002~default-Cal
3. Press [SAMPLE/Enter] once.
DISPLAY READS Are U sure?~default=Enter
The Model 370 and Model 375 have several parameters that can be configured to your sp ecific
application. Table 3-1 is a complete list of the available parameters. Also refer to page 23for
explanations of each parameter. The parameters shaded in gray do not apply to the Model
370/375.