Non-contact infrared thermometer
981C
Instruction manual
1Introduction
Compact, rugged and easy to use. Just aim and push the button, read current surface temperatures in less than a second. Safely measure surface temperatures of hot, hazardous or hard-to-reach objects without contact.
How it works Infrared thermometer measures the surface temperature of an object. The units optics sense emitted, reflected, and transmitted energy which is collect and focused onto a detector. The units electronics transmitted energy which is display on the unit. For increased ease and accuracy the laser pointer makes aiming even more precise.
Cautions
Infrared thermometer should be protected for the following:
--EMF(electro-magnetic fields) from arc welders, induction heaters.
--Thermal shock(cause by large or abrupt ambient temperature changes allow 1 hours for unit to stabilize before use).
--Do not leave the unit on or near objects of high temperature.
Warning
Do not point laser at eye or indirectly off reflective surfaces.
1. When take measurement, point thermometer toward
the object to be measured and hold the yellow trigger. The object under test should be large than the spot size calculated by the field of view diagram.
2. Distance spot size: As the distance from the object increase, the spot size of measuring area becomes large.
3. Field of view: Make sure the target is larger than the unit’s spot size. The smaller the target the close measure distance. When accuracy is critical, make sure the target is at least twice as large as the spot size.
4. Emissivity: Most organic materials and painted or oxidized surfaces have an emissivity of 0.95. Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cover the surface to be measured with masking tape or flat black paint. Measure the tape or painted surface when the tape or painted reach the same temperature as the material underneath.
2Quick start instruction
1. Press battery door clip, install battery correctly. Pull the trigger, LCD display reading battery icon. Release the trigger and the reading will hold for 7 seconds
LCD display: A LCD display B MAX C Low temperature warning setting D Average E High temperature warning setting F Emissivity G MIN H Storage date NO. I Low battery indication J Laser Light K High temperature warning L Measure data M Low temperature warning N Temperature Unit O Date HOLD
2. Locating a hot spot aim the thermometer outside the
(Figure 3)
(Figure 1)
(Figure 2)
area of interest, then scan across with up and down motions until you locate the hot spot.(please turn on the laser to for accurate measuring)
3. Diagram description (1) TriggerPress once, Power on. Press one more time ,
Reading shows on the screen and Turn to Date HOLD model while release the trigger, auto data hold will keep for 10 seconds. 10 sencond without any operation, will auto power off.
(2) Laser Light&Up bottomPress once, turn on laser
light, press one more time, turn off laser light. While setting on warning temperature setting , emissivity setting and Date storage Reading this bottom work as UP bottom.
(3) Back Light/Down bottomPress once, turn on back
light , press one more time, turn off backlight. While setting on warning temperature setting , emissivity setting and Date storage Reading this bottom work as DOWN bottom.
(4) MODE BottomPress this bottom to chose mode
E-MAX-AVG-MIN-LAL-HAL circularly AMAXMeasure current max reading BMINMeasure current min reading CAVGAverage of all the reading data DHALHigh Temperature Warning , Chose mode HAL to set the warning temperature(HIGH), press UP/DN to select temperature. while measure date is higher than setting temperature, H will display on the screen ELALLow Temperature Warning , Chose mode HAL to set the warning temperature(LOW), press UP/DN to select temperature. while measure date is lower than setting temperature, L will display on the screen
(5) MEM BottomFor reading the storage dataPress
once to read the date。Press“UP/DN”to select date It can store only 12 data, while over 12 date, the new one will cover the old oneOnce take out battery, all the date will be gone.
(6) ℃ /
0
F/0F Change, If you want to change the
temperature unitPlease press this bottom
(7) LED Screen (8) Battery door: When replace battery door, please using
the finger indents to pull open the battery door.
3Maintenance
1) Lens cleaning: Blow off lose particles using clean compressed air. Gently brush remaining debris away with a moist cotton cloth.
2) Case cleaning: Clean the case with a damp sponge/cloth and mild soap.
3) Please take out the battery when not using for a long time.
Note:
1) Do not use solvent to clean lens.
2) Do not submerge the unit in water.
4specifications
Temperature
range
-30 to 550(-22to10220F)
Accuracy
±3℃ or ±3%of rdg,-30 to 0(-22 to 320F) ±2℃ or±2% of rdg,,0 to 100(32 to 2120F) ±3℃ or ±3% of rdg,,≥100(2120F)
Repeatability
1%of reading or 1
Response time
500msec, 95%response
Spectral response
8-14um
Emissivity
0.11.0 adjustable
Ambient operating range
0 to ~60(32 to 1400F)
Relative humidity
10-95% RH noncondensing
Storage temperature
-20~60(-4~1400F) without battery
Ambient temp range of guarantee for accuracy
23℃~28
Weight/dimensions
155g;165×72×41mm
Power
9V battery ,6F22 or NEDA 1604
Battery life
Laser models:12hrs
distance spot ratio
8:1
Attached list :Applicable Emissivity for Different Material ( For reference only)
Material
Emissivity
Material
Emissivity
Asphaltum
0.90 to 0.98
TextileBlack
0.98
Beton
0.94
Human Skin
0.98
Cement
0.96
Soap bubble
0.75 to 0.80
Sand
0.90
Charcoal
powder
0.96 Soil
0.92 to 0.96
Lacquer
0.80-0.95
Water
0.92 to 0.96
Lacquer
reluster
0.97
Ice
0.96 to 0.98
Rubber Black
0.94
Snow
0.83
Plastic
0.85-0.95
Glass
0.90 to 0.95
Timber
0.90
Ceramic
0.90 to 0.94
Paper
0.70-0.94
Marble
0.94
Chromic oxide
0.81
Gypsum
0.80 to 0.90
Copper Oxide
0.78
Compo
0.89 to 0.91
Iron Oxide
0.78 to 0.82
Brick
0.93 to 0.96
Stainless steel and
Aluminum
0.2-0.3
(Figure 5)
(Figure 4)