Hoefer SG500 User Manual

user manual
Hoefer SG500
Gradient maker
umSG500-IM /Rev.E0/08-12
Page finder
Pouring gradients from the top ..........................4
Pouring gradients from the bottom ......................5
2. Generating convex and concave exponential
gradients ..........................................................6
To generate an exponential gradient ....................7
3. Care and maintenance .......................................9
4. Ordering information ..........................................9
pi
push-pull valve
reservoir (back) chamber
mixing (front) chamber
connector channel
outlet connector (4 mm ID tubing)
slide stop
Fig 1. Hoefer SG500 gradient maker.
Included but not shown:
Hose clamps (4)
SG500 instructions
The Hoefer® SG500 gradient maker is designed for producing linear gradients of aqueous solutions ranging in volume from 100–500 ml. The SG500 can be used to create convex and concave exponential gradients with the addition of a one-holed rubber stopper, a piece of rigid tubing and a piece of flexible tubing. The unit is fabricated from cast acrylic plastic sheet and tubing. It is suitable for casting acrylamide pore gradient gels and delivering salt gradients for low pressure chromatography systems.
p1

1. Generating linear gradients

open
closed
To generate a linear gradient between two concentrations, equal volumes of solutions of the two concentrations are measured into the two chambers of the gradient maker. As solution is delivered out of the mixing chamber, an equal volume flows in from the reservoir chamber where it is rapidly diluted and mixed to unifor­mity by a magnetic stir bar. The initial concen­tration delivered will be that of the solution in the mixing chamber, the final concentration will be that of the reservoir chamber. For the most consistent delivery of gradients, a peristaltic pump is recommended.
1
Make sure all parts are clean and liquid flows freely through all the connecting channel of the slide valve and pump and delivery tubing.
2
Add a magnetic stir bar of the appropriate dimensions (20–30 mm long) to the mixing chamber and place the unit on a magnetic stirrer. If volumes will be less than half the capacity of the unit, an identical stir bar should be placed in the reservoir chamber as well to balance the displacement and prevent backflow into the reservoir when the chambers are first connected. Connect tubing to the outlet connector and pump, and adjust pump speed, if used. Position or connect the tubing to the receiving vessel (gel casting unit, column, etc.).
Fig 2. Open and closed positions of slide valve.
p2
3
Close the slide valve (out on side of white slide stop button, Fig 2). If the outlet tubing is not controlled by a pump, clamp it off near the gradient former. Add the required volume of the final solution to the reservoir (back) chamber.
Loading...
+ 8 hidden pages