Appendix A. Laemmli System Gels ........................29
Gel recipes..........................................................36
Appendix B. Bibliography .....................................37
Ordering information ............................................39
pi
•
Important Information – English
• If this equipment is used in a manner not specified by Hoefer, Inc. the protection provided by the
equipment may be impaired.
• This instrument is designed for indoor laboratory
use only.
• Only accessories and parts approved or supplied
by Hoefer, Inc. may be used for operating, maintaining, and servicing this product.
• Only use a power supply that is CE marked or
safety certified by a nationally recognized testing
laboratory.
• The safety lid must be in place before connecting
the power supply leads to a power supply.
• Turn all power supply controls off and disconnect
the power leads before removing the safety lid.
• Circulate only water or 50/50 water/ethylene glycol
through the heat exchanger if so equipped. Do
not connect the heat exchanger to a water tap or
any coolant source where the water pressure is
unregulated.
• Never introduce antifreeze or any organic solvent
into any part of the instrument. Organic solvents
will cause irreparable damage to the unit!
• Do not operate with buffer temperatures above
the maximum specified technical specifications.
Overheating will cause irreparable damage to the
unit!
Duležité Informace – Czech
• Pokud by toto zařízení je použito způsobem, který
není podle Hoefer, Inc. ochrana poskytovaná na
základě zařízení může být narušena.
• Tento nástroj je určen pro vnitřní použití v
laboratoři pouze.
• Pouze příslušenství a části schválen, nebo poskytnutých Hoefer, Inc. mohou být použity pro provoz,
údržbu, a údržbě tohoto výrobku.
• zdroj napájení používají jen že je opatřen
označením CE osvědčena nebo bezpečnost
vnitrostátně uznanými zkušebními laboratoř.
• Bezpečnosti lid musí být zavedena před připojením
napájecí zdroj napájení vede k.
• Turn veškeré napájení kontroly vypnuto a odpojit
před odběrem energie vede bezpečnostní víko.
• Rozeslat pouze voda nebo 50/50 voda/ethylenglykolu prostřednictvím výměník tepla je li to vybavena. Nemají připojení výměník tepla s vodními
setřepná nebo jakékoli chladicí kapaliny zdroje, kde
tlak vody je neregulo.
• Nikdy zavést prostředek proti zamrznutí nebo
jakákoli organická rozpouštědla do jakékoli části z
tohoto nástroje. Rozpustidlům způsobí nenapravitelné poškození jednotka!
• Nejsou provozována s pufru teplotách nad
maximální stanovenou technickými specifikacemi. Přehřátí způsobí nenapravitelné poškození
jednotka!
Vigtig Information – Danish
• Hvis dette udstyr bruges i en måde ikke specificeret ved Hoefer, Inc. den beskyttelse, som er blevet
forsynet af udstyret kan måske svækkes.
• Dette instrument er designet for indendørs laboratoriumbrug bare.
• Bare tilbehør og del godkendede eller forsynede
ved Hoefer, Inc. kan måske bruges for drive, funktionsfejl, og betjening dette produkt.
• bruger Bare en strømforsyning, der er CE
markerede eller sikkerhed, som er blevet attesteret
af en, som nationalt er blevet anerkendt prøve
laboratorium.
• Sikkerhedlåget må være på plads før forbinding
strømforsyningsblyet til en strømforsyning.
• Drejer alle strømforsyningskontroller af og afbryder
kraftblyet før fjerning sikkerhedlåget.
• Cirkulerer bare vand eller 50/50 vand/ethylene
glykol gennem varmeveksleren i så fald udrustet.
Forbind ikke varmeveksleren til en vandhane
eller nogen kølemiddelkilde hvor vandtrykket er
unregulated.
• Introducerer Aldrig antifreeze eller noget organisk
opløsningsmiddel ind i nogen del af instrumentet.
Organiske opløsningsmidler vil forårsage uboelig
skade til enheden!
• Driver ikke med stødpudetemperaturer over
maksimummet specificerede tekniske specifications. Overheding vil forårsage uboelig skade til
enheden!
pii
•
Belangrijke Informatie – Dutch
• Indien deze uitrusting in een manier wordt
gebruikt die niet door Hoefer, Inc. is gespecificeerd
de bescherming die door de uitrusting is verzorgd
kan worden geschaad.
• Dit instrument is voor binnenlaboratoriumgebruik
enkel ontworpen.
• Enkel onderdelen en delen keurden goed of
leverden door Hoefer, Inc. kan voor het bedienen
worden gebruikt, handhavend en onderhouden
van dit product.
• gebruik Enkel een netvoeding die CE is markeerde
of veiligheid die door een is gecertificeerd die
nationaal is herkend testene laboratorium.
• Het veiligheidsdeksel moet in plaats voor het
verbinden van de netvoeding leidt tot een
netvoeding zijn.
• Doe alle netvoedingscontroles Uit en koppel los
de machtleiding voor het verwijderen van het
veiligheidsdeksel.
• Circuleer enkel water of 50/50 water/ethyleenglycol door de hitte exchanger zo ja uitrust.
Verbind de hitte exchanger naar een waterkraan
of koelmiddelbron niet waar de waterdruk niet
geregulariseerd is.
• Stel Nooit antivriesmiddel of organische oplosmiddelen in deel van het instrument voor. Organische
oplosmiddelen zullen onherstelbare schade aan de
eenheid veroorzaken!
• Bedien niet met buffertemperaturen boven het
maximum specificeerde technische specificaties.
Oververhittend zal onherstelbare schade aan de
eenheid veroorzaken!
Tärkeää Tietoa – Finnish
• Jos tätä varusteita käytetään tavassa ei määritetty
Hoefer, Inc. suojelu ehkäisty varusteille saattaa olla
avuton.
• Tämä väline suunnitellaan sisälaboratoriokäytölle
vain.
• Vain lisävarusteet ja osat hyväksyivät tai toimitti
Hoefer, Inc. oheen ää voi käyttää käyttämiselle,
valvoalle, ja servicing tämä tuote.
• Vain käyttää käyttöjännitettä joka on CE merkitsi
tai turvallisuus joka on todistanut aidoksi ohi joka
on kansallisesti tunnustettnut testaaminen laboratoriota.
• Turvallisuuskansi täytyy olla paikallaan ennen
yhdistäminen käyttöjännitelyijyjä käyttöjännitteeseen.
• Kiertää kaikki käyttöjännitevalvonnat ja irrottaa
valtalyijyt ennen poistaminen turvallisuuskantta.
• Kiertää vain vesi tai 50/50 vesi/ethyleneä glycol
siinä tapauksessa varustetun lämmönvaihtimen
läpi. Älä yhdistä lämmönvaihdinta vesinapautukseen eikä jäähdytysnestelähteeseen, missä vesipaine on unregulated.
• Pakkasneste eikä orgaaninen liuotin välineen
osassa ei esitele Koskaan. Orgaaniset liuottimet
aiheuttavat korvaamattoman vahingon yksikköön!
• Ei käytä puskuria yllä olevia lämpötiloja enintään
määritetyillä teknisillä täsmennyksillä. Ylikuumeneminen aiheuttaa korvaamattoman vahingon
yksikköön!
Information Importante – French
• Si cet équipement est utilisé dans une manière pas
spécifié par Hoefer, Inc. la protection fourni par
l’équipement pourrait être diminuée.
• Cet instrument est conçu pour l’usage de laboratoire intérieur seulement.
• Seulement les accessoires et les parties ont
approuvé ou ont fourni par Hoefer, Inc. pourrait
être utilisé pour fonctionner, maintenir, et entretenir ce produit.
• utilise Seulement une alimentation qui est CET a
marqué ou la sécurité certifié par un nationalement reconnu essayant le laboratoire.
• Le couvercle de sécurité doit être à sa place avant
connecter l’alimentation mene à une alimentation.
• Tourner tous contrôles d’alimentation de et
débrancher les avances de pouvoir avant enlever le
couvercle de sécurité.
• Circuler seulement de l’eau ou 50/50 glycol d’eau/
éthylène par l’exchanger de chaleur si si équipé. Ne
pas connecter l’exchanger de chaleur à un robinet
d’eau ou à la source d’agent de refroidissement où
la pression d’eau est non régulée.
• Ne Jamais introduire d’antigel ou du dissolvant
organique dans n’importe quelle partie de
piii
•
l’instrument. Les dissolvants organiques causeront
des dommages irréparables à l’unité!
• Ne pas fonctionner avec les températures de
tampon au-dessus du maximum a spécifié des
spécifications techniques. La surchauffe causera
des dommages irréparables à l’unité !
Wichtige Informationen – German
• Wenn diese Ausrüstung gewissermaßen nicht
angegeben durch Hoefer, Inc. verwendet wird,
kann der durch die Ausrüstung zur Verfügung
gestellte Schutz verschlechtert werden.
• Dieses Instrument wird für den Innenlaborgebrauch nur dafür entworfen.
• Nur Zusätze und Teile genehmigten oder lieferten
durch Hoefer, Inc. kann für das Funktionieren, das
Aufrechterhalten, und die Wartung dieses Produktes verwendet werden.
• Verwenden Sie nur eine Energieversorgung,
die CE gekennzeichnet oder durch ein national
anerkanntes Probelaboratorium bescheinigte
Sicherheit ist.
• Der Sicherheitsdeckel muss im Platz vor dem
Anschließen der Energieversorgung sein führt zu
einer Energieversorgung.
• Alle Energieversorgungssteuerungen abdrehen
und die Macht trennen führt vor dem Entfernen
des Sicherheitsdeckels.
• Nur Wasser oder 50/50 Glykol des Wassers/
Äthylens durch den Wärmeaustauscher, wenn so
ausgestattet, in Umlauf setzen. Verbinden Sie den
Wärmeaustauscher mit einem Wasserklaps oder
jeder Kühlmittel-Quelle nicht, wo der Wasserdruck
ungeregelt wird.
• Führen Sie nie Frostschutzmittel oder jedes
organische Lösungsmittel in jeden Teil des Instrumentes ein. Organische Lösungsmittel werden
nicht wiedergutzumachenden Schaden der Einheit
verursachen!
• Mit Puffertemperaturen über angegebenen
technischen Spezifizierungen des Maximums
nicht funktionieren. Die Überhitzung wird nicht
wiedergutzumachenden Schaden der Einheit
verursachen!
Informazioni Importanti – Italian
• Se quest’apparecchiatura è usata in un modo
specificato da Hoefer, Inc. la protezione fornito
dall’apparecchiatura potrebbe essere indebolita.
• Questo strumento è disegnato per l’uso di laboratorio interno solo.
• Solo gli accessori e le parti hanno approvato o
hanno fornito da Hoefer, Inc. potrebbe essere
usato per operare, per mantenere, e per revisionare
questo prodotto.
• usa Solo un alimentatore che è CE ha marcato o la
sicurezza certificato da un nazionalmente riconosciuto testando il laboratorio.
• Il coperchio di sicurezza deve essere nel luogo
prima di collegare i piombi di alimentatore a un
alimentatore.
• Spegne tutto i controlli di alimentatore e disinserisce i piombi di potere prima di togliere il coperchio di sicurezza.
• Circola solo l’acqua o 50/50 glicole di acqua/etilene
attraverso lo scambiatore di calore se così equipaggiato. Non collegare lo scambiatore di calore a un
rubinetto di acqua o qualunque fonte di refrigerante dove la pressione di acqua è sregolata.
• Non introduce mai l’antigelo o qualunque solvente
organico in qualunque parte dello strumento. I
solventi organici causeranno il danno irreparabile
all’unità!
• Non opera con le temperature di tampone al di
sopra del massimo ha specificato le descrizioni
tecniche. Il surriscaldamento causerà il danno
irreparabile all’unità!
Viktig Informasjon – Norwegian
• Hvis dette utstyret blir brukt i en måte ikke spesifisert ved Hoefer, Inc. beskyttelsen som ha blitt git
av utstyret kan bli svekket.
• Dette instrumentet er utformet for innendørs laboratoriumbruk bare.
• Bare tilbehør og deler godkjente eller forsynte ved
Hoefer, Inc. kan bli brukt for drive, vedlikeholde, og
betjene dette produktet.
• bruker Bare en kraftforsyning som er CE merket
eller sikkerhet som ha blitt sertifisert av et som
nasjonalt ha blitt anerkjent prøver laboratorium.
piv
•
• Sikkerheten lokket må være på plass før forbinding
kraftforsyningene blyene til en kraftforsyning.
• Vender all kraftforsyningsstyring av og frakopler
kreftene blyene før fjerning sikkerheten lokket.
• Sirkulerer bare vann eller 50/50 vann/ethylene
glykol gjennom oppvarmingen veksleren i så fall
utstyrer. Ikke forbind oppvarmingen veksleren
til en vanntapp eller noe kjølemiddelkilde hvor
vannet trykket er unregulated.
• Introduserer Aldri antifreeze eller noe organisk
løsemiddel inn i noe del av instrumentet. Organiske løsemiddler vil forårsake irreparabel skade på
enheten !
• Driver med buffertemperaturer over maksimum
ikke spesifiserte teknisk spesifikasjoner. Å overoppheting vil forårsake irreparabel skade på enheten !
Wazne Informacje – Polish
• Jeżeli ten sprzęt jest wykorz ystywany w sposób nie
określone przez Hoefer, Inc. do ochrony przewidzianej przez urządzenie może zostać obniżony.
• Instrument ten jest przeznaczony do użytku w
laboratoriach kryty tylko.
• Tylko akcesoriów i części zatwierdzone lub dostarczone przez Hoefer, Inc. mogą być wykorzystane do
eksploatacji, utrzymania i obsługi tego produktu.
• korzystać jedynie zasilacza że jest noszące oznakowanie CE lub bezpieczeństwa uwierzytelnione
przez uznane na poziomie krajowym laboratorium
badawcze.
• Bezpieczeństwo lid musi być w miejsce przed
podłączeniem zasilania prowadzi do zasilania.
• Zaś wszystkie źródła zasilania urządzenia sterujące
off i odłączyć moc prowadzi przed odbiorem
bezpieczeństwa lid.
• Krążą tylko wody lub wody 50/50/ethylene glycol
wymiennik ciepła poprzez jeśli tak wyposażone.
Nie należy połączyć wymiennik ciepła woda z
kranu lub jakimkolwiek chłodziwo źródła, jeżeli
ciśnienie wody jest nieuregulowanych.
• Nigdy nie wprowadzać rozpuszczalnika organicznego przeciw zamarzaniu lub jakichkolwiek
na dowolną część dokumentu. Rozpuszczalniki
organiczne spowoduje nieodwracalne szkody dla
jednostki!
• Nie działają w buforze temperatury powyżej
maksymalnego określone specyfikacje techniczne.
Przegrzania spowoduje nieodwracalne szkody dla
jednostki!
Informações Importantes –
Portuguese
• Se este equipamento é usado numa maneira não
especificada por Hoefer, Inc. que a protecção fornecida pelo equipamento pode ser comprometida.
• Este instrumento é projectado para uso de interior
de laboratório só.
• Só acessórios e partes aprovaram ou forneceu por
Hoefer, Inc. pode ser usada para operar, manter, e
servicing este produto.
• Só usa um estoque de poder que é CE marcou ou
segurança registrada por um nacionalmente reconhecido testando laboratório.
• A tampa de segurança deve estar em lugar antes
de ligar o estoque de poder leva a um estoque de
poder.
• Desliga todos controlos de estoque de poder e
desconecta os chumbos de poder antes de retirar a
tampa de segurança.
• Circulam só água ou 50/50 glicol de água/ethylene
pelo exchanger de calor se for assim equiparam.
Não ligue o exchanger de calor a uma torneira de
água nem qualquer fonte de refrigerante onde a
pressão de água é não regulado.
• Nunca introduz anticongelante nem qualquer
orgânico solvente em qualquer parte do instrumento. Orgânico solvente causará agressão
irreparável à unidade!
• Não opera com temperaturas de buffer acima do
máximo especificou especificações técnicas. Superaquecer causará agressão irreparável à unidade!
Información Importante – Spanish
• Si este equipo es utilizado en una manera no especificado por Hoefer, Inc. la protección proporcionado por el equipo puede ser dañada.
• Este instrumento es diseñado para el uso interior
del laboratorio sólo.
• Sólo accesorios y partes aprobaron o suministraron
pv
•
por Hoefer, Inc. puede ser utilizado para operar,
para mantener, y para atender a este producto.
• Sólo utiliza una alimentación que es CE marcó o
la seguridad certificada por un nacionalmente
reconocido probando el laboratorio.
• La tapa de la seguridad debe estar en el lugar
antes de conectar la alimentación lleva a una
alimentación.
• Apaga todos controles de alimentación y desconecta los plomos del poder antes de quitar la tapa
de la seguridad.
• Circula sólo agua o 50/50 glicol de agua/etileno
por el intercambiador de calor si ése es el caso
equiparon. No conecte el intercambiador de calor a
un toque de la agua ni cualquier fuente del líquido
refrigerante donde la presión del agua está libre.
• Nunca introduce anticongelante ni algún solvente
orgánico en cualquier parte del instrumento. Los
solventes orgánicos causarán daño irreparable a
la unidad!
• No opera con temperaturas de búfer encima del
máximo especificó especificaciones técnicas. Recalentar causará daño irreparable a la unidad!
Viktig Information – Swedish
• om denna utrustning används i ett sätt som inte
har specificeras av Hoefer, Inc. skyddet tillhandahöll vid utrustningen kan skadas.
• Detta instrument formges för inomhuslaboratorium användning bara.
• Bara medhjälpare och delar godkände eller levererade vid Hoefer, Inc. kan användas för fungera,
underhålla, och servicing denna produkt.
• använder bara en kraft tillgång som är CE
markerade eller säkerhet intygade vid en nationellt
erkänd testande laboratorium.
• Säkerheten locket måste vara på platsen före
koppla kraften tillgången blyen till en kraft tillgång.
• Vänder sig alla kraft tillgång kontroller av och
kopplar bort kraften blyen före flytta säkerheten
locket.
• Cirkulerar bara vatten eller 50/50 vatten/ethylene
glycol genom värmen exchanger i så utrustad fall.
Inte kopplar värmen exchanger till en vatten kran
eller något kylmedel källa där vattnet trycket är
unregulated.
• Inför aldrig kylvätska eller något organiska
lösningsmedel in i någon del av instrumentet.
Organiskt lösningsmedel ska orsaka irreparable
skada till enheten!
• Använd inte med buffert temperaturer över
det högsta angivna tekniska specifikationerna.
Överhettning skulle orsaka irreparabla skador på
enheten!
pvi
•
Waste Electrical and
Electronic Equipment (WEEE)
English
French
German
Italian
Spanish
This symbol indicates that the waste of electrical and
electronic equipment must not be disposed as unsorted
municipal waste and must be collected separately. Please
contact an authorized representative of the manufacturer
for information concerning the decommissioning of your
equipment.
Ce symbole indique que les déchets relatifs à l’équipement
électrique et électronique ne doivent pas être jetés comme
les ordures ménagères non-triées et doivent être collectés
séparément. Contactez un représentant agréé du fabricant
pour obtenir des informations sur la mise au rebut de votre
équipement.
Dieses Symbol kennzeichnet elektrische und elektronische
Geräte, die nicht mit dem gewöhnlichen, unsortierten
Hausmüll entsorgt werden dürfen, sondern separat
behandelt werden müssen. Bitte nehmen Sie Kontakt mit
einem autorisierten Beauftragten des Herstellers auf, um
Informationen hinsichtlich der Entsorgung Ihres Gerätes zu
erhalten.
Questo simbolo indica che i rifiuti derivanti da
apparecchiature elettriche ed elettroniche non devono essere
smaltiti come rifiuti municipali indifferenziati e devono invece
essere raccolti separatamente. Per informazioni relative alle
modalità di smantellamento delle apparecchiature fuori uso,
contattare un rappresentante autorizzato del fabbricante.
Este símbolo indica que el equipo eléctrico y electrónico no
debe tirarse con los desechos domésticos y debe tratarse por
separado. Contacte con el representante local del fabricante
para obtener más información sobre la forma de desechar el
equipo.
Swedish
Denna symbol anger att elektriska och elektroniska
utrustningar inte får avyttras som osorterat hushållsavfall och
måste samlas in separat. Var god kontakta en auktoriserad
tillverkarrepresentant för information angående avyttring av
utrustningen.
pvii
•
1. Unit function and description
The Hoefer® SE400 and SE410 Sturdier™
Vertical Slab Gel Electrophoresis Units are
intended for electrophoretic separation of
proteins and nucleic acids under both denaturing
and native conditions. Up to 28 samples can be
compared on a single slab gel. One gel (or two
gels if using the divider plate, ordered separately)
is cast in the casting stand side of the unit. The
size of the gel is 14 × 15 cm if using the SE400,
and 14 × 23 cm if using the SE410. After casting, the sandwich is transferred into the lower
buffer chamber for electrophoresis.
The basic unit includes one set of glass plates
(18 × 16 cm for the SE400, and 18 × 24 cm for
the SE410), two clamp assemblies (SE400: two
16 cm clamps; SE410: two 16 cm clamps and
two 8 cm clamps), and two cams. The complete
unit includes one 15-well comb and two spacers,
1.5 mm thick, in addition to the basic unit.
p1
•
Unpacking and disassembly
Unwrap all packages carefully and compare
contents with the packing list, making sure all
items arrived. If any part is missing, contact
your local sales office. Inspect all components
for damage that may have occurred while the
unit was in transit. If any part appears damaged,
contact the carrier immediately. Be sure to keep
all packing material for damage claims or to use
should it become necessary to return the unit.
This unit is partially assembled to protect
components during shipping. To disassemble:
1
Position the unit so that the electrical connectors
face you.
2
Note the holes at each side on the upper buffer
chamber. Rest your thumbs in these holes and use
your index fingers to lift the sides of the safety lid
gently until the electrode connectors unplug. First lift
the lid straight up so that the upper electrode shield
clears the upper chamber and then lift the lid out
(toward you) to remove it completely.
3
Lift out the upper buffer chamber and then the glass
plate assembly.
4
Remove the clamps by loosening the thumb screws.
p2
•
Fig 1. SE400 series
main components.
Included but not shown:
Cams
GelSeal grease, ¼ oz.
Spacer-Mate
Wonder Wedge
Level
Required but not included:
Approved power supply
safety lid
color-coded connectors (2)
upper buffer chamber
glass plates (2)
clamps
casting cradle
lower buffer chamber
color-coded leads (2)
leveling feet (4)
•
p3
Annotated inventory
Buffer chambers. Both buffer chambers are
chemically resistant to common electrophoretic
buffers but not to organic solvents or strong
acids and alkalis.
Safety lid. The lid contains both electrodes and
both electrode connectors. The electrode connectors plug into the lead connectors on the lower
buffer chamber. The color-coded leads plug into
color-coded jacks on the power supply.
Glass plates. Two 18-cm wide glass plates are
included. Plates for the SE400 are 16 cm long,
and plates for the SE410 are 24 cm long.
(A notched divider plate, ordered separately,
can be used to run two gels at the same time.)
Clamps. Two 16-cm clamps are required to
secure the 16 cm long sandwich. These and an
additional pair of 8-cm clamps are required to
secure a 24-cm long sandwich.
Casting stand. The caster can be leveled with
the adjustable leveling feet on the bottom of the
unit. A laminated gasket seals the bottom of
the glass plate assembly when it is locked into
the stand.
Cams. Cams are used twice; first to secure the
assembled sandwich in the casting stand and
again to lock the sandwich and the upper buffer
chamber together.
Rubber gaskets. There are two gaskets. The
laminated gasket fits into the bottom of the casting stand and provides the seal for the bottom
of the gel sandwich. The slotted gasket fits
under the upper chamber and provides the seal
between the sandwich and the upper chamber.
Two ridges help position this gasket.
p4
•
Spacer-Mate assembly template. Aligns spacers
for sandwich assembly.
Wonder Wedge Plate Separator Tool. Use to
disassemble gel sandwiches and to gauge spacer
and comb thickness.
Spacers. (May be ordered separately.) Spacers
determine the thickness of the gel. They are 2 cm
wide and are available in three thicknesses: 0.75,
1.0, and 1.5 mm.
Combs. (May be ordered separately.) Combs are
available in sizes that form 10, 12, 15, 20, or 28
wells. Preparative combs include 1 or 2 reference
wells in addition to a preparative well. Most
combs are available in all three thicknesses:
0.75, 1.0 and 1.5 mm.
All preparative combs, and combs with fewer
than 28 wells form wells that are 25 mm deep.
The 28-well comb forms wells that are only
15 mm deep so that wells do not collapse when
the comb is removed. The sample volume held
by each well depends on the gel thickness, well
depth and the number of wells per comb. Table
2 on page 15 lists volume per 1 mm depth for
wells created by each comb size. See the ordering
information for additional comb specifications.
p5
•
Specifications
Glass plate size SE400: 18 × 16 cm
SE410: 18 × 24 cm
Approximate gel size SE400: 14 × 15 cm
SE410: 14 × 23 cm
Max. wattage 20 W
Max. voltage 500 V at 40 mA
Max. amperage 30 mA/gel
(60 mA total at 325 V)
Max. temperature 45 °C
Environmental Indoor use: 4– 40 °C
operating conditions Humidity up to 80%
Altitude up to 2000 m
Installation category: II
Pollution degree: II
Dimensions (w × h × d) SE400: 24 × 28 × 15 cm
This declaration of conformity is only valid for the
instrument when it is:
• used in laboratory locations,
• used as delivered from Hoefer, Inc. except for alter-
ations described in the user manual, and
• connected to other CE-labeled instruments or prod-
ucts recommended or approved by Hoefer, Inc.
p6
•
2. Operating instructions
Procedures for casting gels and electrophoretic
separation follow. Included are instructions for
both single percentage (homogeneous) and
gradient polyacrylamide gels. Appendix A lists
recipes and Appendix B gives a bibliography.
2.1 Gel casting preparation
2.1.1 Options—precast gels and self-cast gels
The SE400 unit accepts standard precast gels
purchased from commercial suppliers as well
as self-cast gels, which can be prepared using
the built-in casting stand. (To cast multiple
14 × 16 cm gels, the Multiple Gel Caster Kit,
which holds up to 10 sandwiches, and the Gel
Caster Kit, which holds up to four sandwiches,
can be ordered separately.) Gels for the SE410
must be self-cast.
Glass plates, spacers, and clamp sets are sized
so that the assembled sandwich can be easily
aligned to create the required seal. When assembling sandwiches, take extra care to align all components for best results.