PH-200 1/06
TAKING MEASUREMENTS
The PH-200 can take measurements of pH levels and temperature.
pH Measurements
1. Remove the cap.
2. Press the ‘ON/OFF’ button. The display will become active.
3. Dip the meter into the water sample, liquid or solution to be tested.
4. Lightly swirl
the meter to ensure the removal of trapped air bubbles or electric
charges. Do not tap it against the glass.
5. The meter will display a reading almost immediately. Keep
the meter in the liquid until the reading stabilizes (approx. 30
seconds) for an accurate reading. NOTE – It is a normal
function of pH for the reading to fluctuate and not fully
stabilize.
6. To view the reading out of the liquid, quickly press the
‘HOLD/MODE’ button while the meter is in the liquid. This
will hold the reading on the screen. Quickly pressing the
‘HOLD/MODE’ button again will release it.
7. Press the ‘ON/OFF’ button to turn the meter off.
8. Shake any excess water off the meter and rinse with distilled
or de-ionized water. Put the cap back on.
Temperature Measurements
The temperature reading is always displayed on the LCD panel during measurement mode, and
is shown simultaneously for pH readings. It is not shown when the meter is in calibration mode.
The default temperature reading for the meter is in Celsius.
1. Remove the cap.
2. Press the ‘ON/OFF’ button. The display will become active.
3. The temperature reading is always displayed on the LCD panel (except in calibration
mode), and is shown simultaneously for pH readings.
4. The default mode for temperature is Celsius. To change the temp mode, quickly
press the ‘TEMP/CAL’ button to switch from Celsius to Fahrenheit or vice-versa.
5. Dip the meter into the water sample or solution to be measured.
6. The temperature reading will change immediately (unless the solution is at room
temp). For very hot or cold liquids, the reading may take slightly longer to stabilize.
7. Press the ‘ON/OFF’ button to turn the meter off.
8.
Shake any excess water off the meter and rinse with distilled or de-ionized water. Put
the cap back on.
NOTE – Because of the sensitivity of the pH sensor and reference tube, it is not recommended
to use your meter as a thermometer in very hot or very cold liquids.
Switching Temperature Modes:
The default temperature reading for the meter is in Celsius. To change the temperature mode,
quickly press the ‘TEMP/CAL’ button to switch from Celsius to Fahrenheit or from Fahrenheit to
Celsius.
CALIBRATION
Your PH-200 meter has been factory-calibrated to a 7.00 pH buffer solution. While this is
suitable for most applications, it may be necessary to recalibrate the meter for more accurate
results. Additionally, it should be noted that pH buffer solutions may vary amongst
manufacturers. For best results, the PH-200 should be recalibrated to a pH buffer solution that
suits your specific needs.
The PH-200 features 3-point auto-ranging calibration with digital fine tuning. The meter can be
calibrated to pH buffer solutions of 4, 7 or 10 pH. To recalibrate the meter:
1. Turn the meter on by pressing the ‘ON/OFF’ button. Note that the reading will
fluctuate when the meter is not submersed in a liquid. It will begin to stabilize once it
is in a pH buffer solution.
2. Dip the meter into a pH buffer solution of 4, 7 or 10 pH. Lightly stir the meter for
approximately 30 seconds. The reading will change during this time and may continue
to change slightly after 30 seconds. This is normal.
3. Press and hold the ‘TEMP/CAL’ button. The temperature display will change to a
‘CAL’ image.
4. The meter will automatically adjust to a reading
within a range of the solution the meter is in.
5. Change the reading so that it matches the
calibration solution. For example, if your
calibration solution is 4 pH, decrease or
increase the reading until it reads ‘4.00’. To
increase the reading, press the ‘UP’ button
(TEMP/CAL). To decrease the reading, press
the ‘DOWN’ button (HOLD).
- If the calibration reading reaches the
minimum or maximum level within the
range (approx. +/- .5 pH), the screen will
display the ‘minimum calibration reached’
icon or ‘maximum calibration reached’
icon, respectively. Note that this occurs
only within the range of the sample the
meter is currently in. When ‘CAL’ does
not flash, it means calibration is in the
middle of the range.
6. To set the calibration, press and hold the
‘TEMP/CAL’ button until the screen reverts back to the measurement mode.
7. Your meter is now re-calibrated.
NOTE – The sensor is very sensitive. It is normal for the reading to change by .01 or .02.