HLD PhysioTouch A02, LymphaTouch F02 User Manual

Page 1
USER MANUAL
Ver. A02 SW rev 2.0x
The manual contains instructions for use, maintenance and handling of the device, as well as safety precautions and warnings. Read through carefully before using the device, to ensure safety and eciency for you and your patient. Keep for future reference.
Page 2
2
Page 3
1. Safety 5
2. Application specification 7
3. Parts description 9
4. Connectors 10
5. Symbols and product markings 11
6. Installation 12
7. Using the PhysioTouch® - LymphaTouch® device 16
8. Treatment settings 20
9. Maintenance and cleaning 27
10. Troubleshooting 28
11. Service 30
12. Software upgrade 31
13. Disposal 33
14. Technical data 34
15. EMC 36
3
Page 4
Copyright©
© HLD Healthy Life Devices Ltd. Helsinki, Finland. All material included in this manual is intended solely for use related to the use of the PhysioTouch® - LymphaTouch® is a trademark of HLD Healthy Life Devices Ltd. FDA device class I Quality ISO9001:2008 / 13485:2003
710359
Manufacturer
HLD Healthy Life Devices Ltd Kuortaneenkatu 2 00510 Helsinki Finland www.physiotouch.com www.lymphatouch.com
Safety and EMC certificate
HLD PhysioTouch® - LymphaTouch® A02 conforms with Medical Device Directive 93/42/EEC and medical electrical equipment electromagnetic compatibility IEC 60601-1-2.
0537
4
Page 5
1. Safety
In this User Manual, safety precautions and warnings are indicated with the following signal words. Make sure to read and understand all safety precautions and warnings before starting to use the device.
Caution – potentially hazardous situation which, if not avoided, may cause minor to moderate injuries, or damage to the device.
Warning - potentially hazardous situation which can cause injuries or damage the device, if instructions are not followed.
Danger – potentially hazardous situation which can result in serious injury or death, if instructions are not followed.
5
Page 6
1.1 Safety precautions
• Clean the device and accessories regularly following the instructions and the regulations of your institution, to ensure safe and hygienic use.
• Use and store the device only under the environmental conditions specified.
• Handle the device with care, and follow the instructions for maintenance.
• Use only accessories and mounting provided by or approved by HLD Healthy Life Devices Ltd.
• Use treatment cups and treatment setting that are best suited to the anatomy and skin type of the patient you are treating.
• Avoid situations where the unit might fall or collide with a heavy object.
• Inspect the device and the related accessories for cracks or other signs of overuse, and replace accessories as required.
• Make sure the filter in the treatment cup is in place before attaching it to the Treatment Head, to avoid dust and debris from being sucked into the device.
• Never use the unit for purposes other than those recommended by HLD Healthy Life Devices Ltd. HLD Healthy Life Devices Ltd will not be liable for any inappropriate use of the equipment.
• Do not treat directly over skin with open wounds.
• Avoid situations where the cable can be wrap around the neck.
• Do not leave the device in proximity of pets, pests or children.
• This device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in Chapter 14
6
Page 7
2. Application specification
Intended use:
PhysioTouch® - LymphaTouch® is a negative pressure-based therapy device and aid for physiotherapy and lymphatic drainage therapy. The device is meant to be used by persons who are knowledgeable about the treatment method or under supervision of those persons.
Intended conditions of use:
The environment, where this device is used typically physiotherapy setting including hospitals, physiotherapy centres.
Contraindications:
Do not use on a pregnant patient without prior consultation with the attending doctor. Do not use on a patient with cancer during active treatment period or without prior consultation with the attending doctor. Do not use on acute infection. Be very careful when applying on fragile skin and on the throat area. Do not use on a patient with a diagnosed current venous thrombosis. Do not use on a patient suering from protein poor oedema due to severe heart failure, cardiac oedema. If you are uncertain of the suitability of the device for treatment of your patient, seek medical advice prior to starting the treatment. In addition, all contraindications of physiotherapy and manual lymphatic drainage apply.
7
Page 8
Intended medical indication:
Improvement of lymphatic circulation in the treated area, improvement of secondary lymphedema and reduction of secondary lymphedema of the Arm (SLA) Post Mastectomy. Device is used on any patient deemed to be appropriate for a physical treatment. The patient has no control of device. Device has no specific age or weight limits on patients.
Operating principle
The mechanism of action is based on the eects of negative pressure in tissues. Negative pressure created by the treatment device stretches the skin and the tissue underneath, pulling on anchoring filaments to dilate the endothelial openings of lymph vessels. Vertical stretching of the fascial (connective tissue) structures is accomplished at the same time, expanding the space for circulation of blood and lymph. Lymph and the metabolic waste products that impede the healing process can then flow more easily from the interstitial space into lymph vessels, and excess fluid is carried away by the body’s own lymph transport mechanisms to re-join the venous circulation.
8
Page 9
3. Parts description
1. Main unit
2. Handle
3. On / O switch
4. Touch screen
5. Treatment Head
6. Cable
7. Dierent sizes of treatment cups for the Treatment Head
8. Soft covers for the treatment cups (detachable on the larger treatment cups)
9. Treatment cup filter
10. DC-power, country adapter included
9
Page 10
4. Connectors
1. Device’s serial number
1
2
2. Integrated air hose and cable connector
3. Mains power
4. LAN for future use
5. USB
3
Warning – Do not touch USB port while treating patient or let the patient touch the USB while being treated. Do not use USB or LAN connectors while treatment in progress.
10
4
5
Page 11
5. Symbols and product markings
In addition to the connector markers described above, you may find the following symbols and markings in the device:
Symbol Description
0537 The product is CE-marked and the number indicates the notified body.
Read the User Manual before using the device.
The device is classified as Class BF, according to the standard IEC 60601.
The device has double insulation; connection to mains does not require a grounding pin.
Special instructions are needed to dispose device properly.
Lithium-ion battery must be disposed properly.
The carton package of the device can be recycled.
This side up.
Weights less than 5.51 lbs.
Do not use sharp instrument to open the package
Fragile. Handle with care.
Humidity limitation. See more from the table page 34.
Keep dry.
Temperature limitation. See more from table page 34.
11
Page 12
6. Installation
6.1 Unpacking
Carefully unpack all parts of the device and confirm that all parts are included and intact.
Caution - We recommend you to keep the box in case you need to return the device or send the device for service – it is most safely transported in the box it was originally shipped in.
6.2 Internal Battery
PhysioTouch® - LymphaTouch® has internal battery which is advised to charge full before first usage.
During battery recharge, amber light blinks in the power button. When battery is fully charged amber light is on constantly. When device is used via battery and the device is on, the light is green.
12
Page 13
6.3 Connecting and disconnecting the cable
Attach the integrated air hose and cable to the round connector on the left side of the unit. Make sure the cable is tightly connected, but do not use excessive force or rotate connector. To disconnect, pull the connector straight.
6.4 Connecting and disconnecting the Treatment Head
Attach the integrated air hose and cable to the round connector on the end of the Treatment Head. Make sure the cable is tightly connected, but do not use excessive force or rotate connector. To disconnect, pull the connector straight.
13
Page 14
6.5 Connecting a treatment cup to the Treatment Head
Select the appropriate treatment cup for the patient and body area in question. Please refer to the Chapter 7.4 ”Selecting a treatment cup”.
Connect the treatment cup to the Treatment Head as indicated in the figure.
Caution - Make sure the treatment cup is connected in right orientation, as indicated by the guiding rails.
6.6 Connecting removable soft part to the treatment cup
60mm and 80mm treatment cups have separate soft parts. Connect the treatment cup soft part to treatment cup as indicated in the figure.
14
Page 15
6.7 Placing the PhysioTouch® - LymphaTouch® device for treatment use
We recommend mounting the PhysioTouch® - LymphaTouch® to the Roll Stand (see figure) when the device is used primarily in one location. The Roll Stand is designed for safe, ecient and ergonomic use.
Instructions for connecting the PhysioTouch® - LymphaTouch® main unit to the Roll Stand are included in the Roll Stand package.
Caution – Use only mounting elements provided or recommended by HLD Healthy Life Devices Ltd to mount the device. If the device is not attached properly, it may fall and cause injury or damage to the equipment.
Caution – HLD0901 Roll stand is designed solely to be used with PhysioTouch®
- LymphaTouch® device. Do not connect other devices or equipment to the Roll stand. Basket of the roll stand is designed to withstand the weight of 4 treatment cups and the treatment head. Do not exceed this limit.
Caution – If used without Roll stand place the device on a flat surface and be sure that the device is securely on top of the surface and device can’t falll o.
15
Page 16
7. Using the PhysioTouch® - LymphaTouch® device
7.1 Turning the device on and o
Turn the device on by pressing the on/o switch in the lower right hand corner. Device starts in 35 seconds. To turn o the device, press the on/o switch. The device will also shut down automatically after an hour of inactivity, while on battery power.
7.2 Screen saver
The display will shut down and the device will enter a power-saving mode when the device has not been used for 30 minutes. Reactivate the unit by pressing the on/o switch or by touching the screen.
16
Page 17
7.3 Patient preparation
PhysioTouch® - LymphaTouch®–treatment does not require special preparation of the patient. The treatment is applied to bare skin.
Moderate application of oil or massage cream may allow the treatment cup to glide more smoothly on the skin. Avoid excessive use of oil or massage cream.
If the patient’s skin is very dry and producing dandru, the filter in the treatment cup may get clogged faster than normal. This will result in diculty achieving the set pressure. Replace the treatment cup in order to continue an ecient treatment. The error message below is displayed if filter is clogged. Do not use talcum powder or other powdery top agents on the skin.
17
Page 18
7.4 Selecting a treatment cup
Select the proper size treatment cup for each treatment situation. Please refer to the table below for general guidance, while using your own clinical judgment as to what is an appropriate size for each situation.
Treatment cup Size Removable rubber ring Body areas
• Local scars
• Small joints
35 mm
HLD0335
50 mm
HLD0350-2
60 mm X
HLD0360
• Face
• Head and neck
• Local scars
• Joint areas
• Face
• Head and neck
• Palm area
• Whole body
18
HLD0380
80 mm X
• Torso
• Upper and lower limbs
Page 19
7.5 Accessories
Code Accessory
HLD1300 Treatment head
HLD0901 Roll Stand See page 15.
HLD1020-35 Treatment cup 35mm, 20pcs
HLD1020-50 Treatment cup 50mm, 20pcs
HLD1020-60 Treatment cup 60mm, 20pcs
HLD1020-80 Treatment cup 80mm, 20pcs
HLD0122 Treatment Cable
HLD1601 PhysioTouch® - LymphaTouch®
carrying bag
Use these codes to order accessories from the nearest distributor.
19
Page 20
8. Treatment settings
Turn on the device. PhysioTouch® - LymphaTouch® main screen will appear (see picture). The default setting is 80 mmHg with 2 s. pulsations. The unit will start operation automatically when the treatment cup is put close to the treated target.
1
2
1. The setting is indicated on the main screen on the top left, as well as graphically in the form of a wedge. During treatment a blue wedge appears which indicates the current vacuum in the treatment cup.
2. The vacuum is adjusted with the touch screen by moving the slider at the desired location.
If the desired vacuum is not achieved make sure that the treatment cup selected is suitable to the aected area of the body. If necessary replace the treatment cup with a more appropriate size, or reposition the treatment cup on the skin carefully. Also, the patient’s body hair may cause air leakage from the treatment cup. Refer to the section ”Troubleshooting.”
20
Page 21
1
2
1. The bar on the right hand side of the screen indicates the distension of the skin in the middle of the treatment cup, as a result of the applied negative pressure.
2. The battery icon in the lower left hand corner indicates the battery charge level. When the device is connected to the mains, the top of the battery icon is a power plug icon and when the battery is charging, a flash icon.
3. During treatment, it is possible to choose either pulsating or continuous treatment. The pulsation button at the bottom right is green with bristles when pulsation is activated. Button is green and continuous when continuous treatment is selected. To modify the pulsation setting, touch the pulse button on the main screen, and the pulsation setting screen will appear.
3
21
Page 22
8.1 Pulsation screen
1
2 3 4
1. Set work/rest ratio by sliding the arrow at the top of the screen between 30 – 90%.
2. Set pulse length by sliding the arrow, or by pressing the desired location on the axis on the lower part of the screen.
3. Constant negative pressure. Return to main screen and deactivate pulsation.
4. Return to main screen and keep the pulsation on.
22
Page 23
8.2 Treatment timer
PhysioTouch® - LymphaTouch® has pre-set treatment times that can be selected from the timer display. Press the treatment button to open a display where you can select the desired duration of treatment.
Remaining duration of treatment is shown on the main display. See the picture below. When treatment time ends PhysioTouch® - LymphaTouch® will notify with treatment head valve ending.
23
Page 24
8.3 High-frequency vibration treatment feature
High-frequency vibration treatment feature can be activated from the High-frequency screen by pressing the high-frequency button on the main screen.
On the High-frequency screen there are two buttons; the left one is for turning the high frequency treatment o and the right one is for high frequency treatment. Initial settings are: the high frequency treatment mode o and the pulsation treatment mode activated.
24
Page 25
Constant high-frequency vibration without pulsation can be set on by deactivating pulsation treatment mode and activating the high frequency treatment mode.
By activating both the high-frequency vibration and the pulsation treatment mode, combination treatment can be started. Treatment mode is now vibrating pulsation treatment.
25
Page 26
8.4 Numeric keypad
Numeric keypad button can be found from timer setting screen. Numeric keypad –screen is used mainly in service purposes.
26
Page 27
9. Maintenance and cleaning
Check the device for visible damage or malfunctions regularly, and replace accessories if required.
9.1 Treatment cup maintenance and cleaning
Wipe the treatment cups carefully using a mild disinfectant or mild detergent after each use. Disconnect the soft covers from the larger treatment cups (60 mm and 80 mm) during the cleaning process. The treatment cups are not autoclavable.
Inside the treatment cup is a small, white filter that prevents dust and debris from entering the device. When cleaning the treatment cup, be careful not to touch the filter, as the detergent may also cause clogging of the filter. Manufacturer recommends that mouthpieces are replaced with new ones when there is visible wear and tear or when the cups have been used longer than three months.
9.2 Cleaning main unit and other parts
Clean the PhysioTouch® - LymphaTouch® main unit, cable and Treatment Head by wiping them with a cloth containing mild detergent. Be careful not to scratch the display. Do not autoclave any parts of the device.
27
Page 28
10. Troubleshooting
The section below provides assistance in certain problem situations. If the suggested corrections do not fix the situation, contact device’s retailer for more instructions, or return the device for service.
Problem: Device will not turn on
• Battery may be empty. Connecting to mains will start charging the battery, and allow using the device with mains power immediately.
• There may be an internal error which has caused the system to halt. Press the on/o switch for 10 seconds to reset the device.
Problem: Treatment cable disconnected
• Verify that the cable is properly connected both to the device and to the Treatment Head. Contact service if necessary.
Problem: Suction will not start when the Treatment Head is brought to the patient’s skin.
• If the treatment cup is not properly connected, the error message will be displayed. Verify that the treatment cup is properly connected to the Treatment Head. Replace treatment cup if necessary.
28
Page 29
Problem: Device is overheated
• Device is used with maximum performance long period of time in hot environment.
• Wait until the unit cools down and the error message disappears.
Problem: Suction initiates, but the set negative pressure is not achieved.
• Ensure that the treatment cup fits tightly on the area you are attempting to treat. Try to find a location where you can achieve an airtight connection or replace treatment cup with the size that fits better.
• The filter in the treatment cup may be clogged. Replace the treatment cup.
• Verify that the treatment cup is properly connected to the Treatment Head.
• Verify that the soft treatment cup cover is properly connected to the treatment cup and intact.
• Wear and tear of the treatment cup may result in leaks appearing. Verify and replace the treatment cup if necessary.
29
Page 30
11. Service
If the device malfunctions or the error message below appears, first try to reset the device by pressing the on/o switch for 10 seconds. If the error message reappears, contact the reseller and inform the error code displayed on the lower right corner.
Warning, Danger - Only persons certified by HLD Healthy Life Devices Ltd. are authorized to perform servicing of the device. Opening the device or servicing the device without authorization will result in voiding the warranty and may lead to serious injury or damage to the equipment.
30
Page 31
12. Software upgrade
HLD Healthy Life Devices Ltd releases new software versions to the PhysioTouch® - LymphaTouch® unit from time to time. Software can be easily upgraded by using a USB memory stick.
The software upgrade is provided either via email, internet, or directly on a USB stick. More detailed instructions on performing the software upgrade and detailed description of the new software features will be included with each software upgrade.
If you have received a software zip file upgrade via email or downloaded it from the www.physiotouch. com Web site, unzip the file to the USB memory stick root folder. There should be the following two files on the memory stick when the upgrade process is initiated: hld_update_2.xx.img and upgrade.spec. Other files on the memory stick will not hamper the upgrade process.
31
Page 32
Proceed with the software upgrade according to the following steps:
• Keep the device connected to mains during the upgrading.
• Turn o the device.
• Connect the memory stick to the USB connector on the left side of the device.
• Turn on the device, and follow the instructions on the screen.
• Do not disconnect the memory stick, until instructed to do so.
• The device will restart and update the device’s main board and Treatment Head if necessary. If the Treatment Head is not connected at the time, it will be updated the next time it is connected.
• Upgrading is complete and the device is ready for use.
If you have several PhysioTouch® - LymphaTouch® treatment devices in use, it is recommended that all units are upgraded to the same software level.
32
Page 33
13. Disposal
This is an electronic device containing a lithium-ion battery as an internal power source. Do not throw away the device with normal household waste at the end of its life. Consider the environment: deliver the device for recycling at an ocial collection point for electronic devices. Check with the appropriate recycling organization for local disposal information.
33
Page 34
14. Technical data
Dimensions (w x h x d) 19 x 10 x 9,5 cm
Weight Main unit 1.2 kg
Power requirements Adapter 100-240 VAC, 50/60 Hz
Type HLD-2200 Lithium Ion 10,8V,
5.2Ah, 56Wh
Replacement interval ~3 years
Duration time in use 8 hours
Battery
(*) Estimated time. Time can vary due to the ambient and / or device temperature. Note: Designed lifetime for the adapter is two years.
34
Duration time in stand by 18 hours (*)
Charging while in use 8 hours (*)
Charging while in stand-by 5 hours (*)
Page 35
Environmental conditions
Use 10°C - 40°C,
humidity 15 - 93% non-condensing, altitude 0 - 3000 m
Storage and transport 0° - 40° C,
humidity 15 - 95% non-condensing, altitude 0 - 3000 m
Classification
Conformity to standards
Performance specifications
Designed lifetime
Medical device (CE-MDD), class 2a
ISO 13485:2003 IEC 60601-1 IEC 60601-1-11
Negative pressure adjustment
Pulsation adjustment Range 0.5-5 s
Vibration adjustment Range 20-90 Hz
Range 0-250 mmHg Accuracy ± 10% + ± 10 mmHg
Accuracy ± 10% + ± 0.05s
Accuracy ± 10%
5 years
35
Page 36
15. EMC
15.1 Guidance and manufacturer’s declaration – electromagnetic emissions
The PhysioTouch® - LymphaTouch® is intended for use in the electromagnetic environment specified below. The customer or the user of the PhysioTouch® - LymphaTouch® should assure that it is used in such an environment.
Warning: Use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by HLD Healthy Life Devices as replacement parts for internal components, may result in increased electromagnetic emissions or decreased immunity of the device to such emissions.
Warning: PhysioTouch® - LymphaTouch® should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the PhysioTouch® - LymphaTouch® should be observed to verify normal operation in the configuration in which it will be used.
36
Page 37
Emission tests according to the test specification IEC 60601-1-2.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Group 1 The PhysioTouch® - LymphaTouch® uses RF energy
only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
Not Applicable (low input power)
Complies
The PhysioTouch® - LymphaTouch® is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
37
Page 38
15.2 Guidance and manufacturer’s declaration – electromagnetic immunity
The PhysioTouch® - LymphaTouch® is intended for use in the electromagnetic environment specified below. The customer or the user of the PhysioTouch® - LymphaTouch® should assure that it is used in such an environment. Portable and mobile RF communications equipment can aect medical electrical equipment. Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can aect this equipment and should be kept at least a distance d (see tables on pages 40-41) away from the any part of the PhysioTouch® - LymphaTouch® including cables.
Immunity test IEC 60601
test level
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
38
± 6 kV contact ± 8 kV air
± 2 kV for power supply lines ± 1 kV for input/output Lines
Compliance level
± 6 kV contact ± 8 kV air
± 2 kV for power supply lines ± 1 kV for input/ output Lines
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Page 39
Immunity test IEC 60601
test level
Surge IEC 61000-4-5
± 1 kV line(s) to line(s) ± 2 kV line(s) to earth
Compliance level Electromagnetic
environment – guidance
± 1 kV line(s) to line(s) ± 2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency
<5 % U for 0,5 cycle; 40 % U for 5 cycles; 70 % U U <5 % U for 5 s
(>95 % dip in UT)
T
(60 % dip in UT)
T
(30 % dip in
T
) for 25 cycles;
T
(>95 % dip in UT)
T
<5 % U for 0,5 cycle; 40 % U for 5 cycles; 70 % U U <5 % U for 5 s
3 A/m 3 A/m Power frequency
(>95 % dip in UT)
T
(60 % dip in UT)
T
(30 % dip in
T
) for 25 cycles;
T
(>95 % dip in UT)
T
(50/60 Hz) magnetic field IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the PhysioTouch® ­LymphaTouch® requires continued operation during power mains interruptions, it is recommended that device is installed with battery option.
magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
39
Page 40
Immunity test IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz – 80 MHz
3 Vrms Wireless communications equipment such as wireless
home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can aect this equipment and should be kept at least a distance d away from the any part of the PhysioTouch® - LymphaTouch® including cables. The recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d=1.2*√(P) MHz to 800 MHz d=2.3*√(P)800 MHz to 2,5 GHz
Radiated RF IEC 61000-4-3
3 V/m 80 MHz – 2.5 GHz
3 V/m
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined
a
by an electromagnetic site survey,
should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the PhysioTouch is used exceeds the applicable RF compliance level above, the PhysioTouch should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the PhysioTouch
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
40
Page 41
Recommended separation distances between portable and mobile RF communications equipment and the PhysioTouch® - LymphaTouch®
The PhysioTouch® - LymphaTouch® is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PhysioTouch® - LymphaTouch® can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the PhysioTouch® - LymphaTouch® as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter (m) 150kHz – 80MHz
d=1.2*√P
80MHz – 800MHz d=1.2*√P
800MHz – 2.5GHz d=2.3*√P
0.01 0.1 2 0.1 2 0.23
0.1 0.37 0.37 0.74 1 1.2 1.2 2.3 10 3.7 3.7 7.4 100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
41
Page 42
424344
Page 43
Page 44
HLD Healthy Life Devices Ltd. Kuortaneenkatu 2 FI - 00510 Helsinki Tel. +358 (0)10 423 7300 Fax +358 (0)10 423 7309 [email protected], [email protected] www.physiotouch.com www.lymphatouch.com ISO 9001:2008 / 13485:2003 Certified company © Copyright 10/2016 v8.0 HLDID-30-1238 EN
Directive 93/42/EEC
0537
Loading...