Hitachi Koki CS150NX Instruction Manual

Page 1
Important
2014.11
HITACHI
TABLETOP
MICRO ULTRACENTRIFUGE
CS150NX
INSTRUCTION MANUAL
Before using this CENT RIFUGE carefully read through this INSTRUCTION MANUAL t o ensure efficient, safe oper at i on. Keep this INSTRUCTIO N M ANUAL available as an important referenc e when using the CENTRIFUGE.
MFG.No. CAT.No. Manual code
The appearance or speci ficat ion of the products covered in this manual is subject to partial change f or i mprovement.
S99806110
Page 2
Page 3
Copyright © 2013 Hitachi Koki Co., Ltd.
The names of actual companies and products mentioned herein may be the trademarks of their respective owners.
All rights reserved. No part of this document may be reproduced or transmitted in any form or any means without permission from Hitachi Koki Co., Ltd.
Page 4
General description
The CS150NX is designed to separate liquid-suspended materials having different densities and particle size. The CS150NX is a product that pursues user-friendliness and reliability based on our many years of experience in developing centrifuges. This centrifuge offers many features that we are confident of satisfying your requirements. These features include the following.
1. The maximum speed is 150,000 rpm (1,050,000 x g).
2. Small and compact tabletop type
3. The touchscreen with easy-to-see color liquid crystal screen is incorporated.
4. The displayed language can be switched over between Japanese and English.
5. Various alarm indications can notify users of the causes and necessary actions of the troubles. It can realize troubleshooting easier and quicker.
6. The rotor can be set by simply mounting it on the drive shaft in the rotor chamber (quick setting type.)
7. Samples can be easily balanced visually (see Section 2-1-2).
8. This product spins very quietly, and is thus well suited for personal use.
9. A CFC-free thermomodule cooling system is employed featuring a powerful cooling capacity.
(Samples can be kept at 0 at maximum speed when the ambient temperature is 30℃. )
10. The real-time control (RTC) feature enables setting a start time or a finish time, thus letting you
run your machine at a desired date and time.
11. Centrifugal force (RCFmax and RCFavg) can be displayed and set (Note 1).
12. Twenty varieties of nine stepped modes can be programmed for a wide range of applications
such as step running.
13. In addition to a door lock and an imbalance detector, two independent microprocessors are
incorporated for overspeed detection (a dual CPU overspeed prevention mechanism) for even greater safety.
4
Note 1: RCF: Relative Centrifugal Force
Page 5
SAFETY NOTICES
r
Safety reminders
Carefully read and fully understand the following safety instructions.
Operate your instrument according to the instruction manual.
Be sure to observe the all safety precautions in the instruction manual and safety instructions on your instrument. If neglected, personal injury and/or instrument damage can be caused.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
The safety reminders are indicated as shown below. The signal words "DANGER", "WARNING" and "CAUTION" are indicated together with the hazard alert symbols in this manual.
DANGER
This note indicates an imminently hazardous situation, which if not
strictly observed, could result in personal severe injury or possible death.
WARNING
This note indicates a potentially hazardous situation, which if not
strictly observed, could result in personal severe injury o possible death.
CAUTION
This note indicates a potentially hazardous situation, which if not
strictly observed, could result in personal injury or severe damage to the instrument.
This hazard alert symbol indicated together with a signal word is a reminder to emphasize important safety instructions.
"NOTE" indicates a note which has no direct bearing on personal safety.
Do not perform any operation not specified in the instruction manual. If any problem is
found on your instrument, contact a Hitachi Koki authorized sales/service representative. Although the safety precautions in the instruction manual and safety instructions on your
instrument have been fully considered, an unexpected situation may arise. Observe the instructions in the instruction manual and always be careful yourself when operating this instrument.
i
Page 6
SAFETY NOTICES
Mechanical Safety
WARNING
Do not open the door while the rotor is spinning.
Do not attempt to slow or stop the spinning rotor by hand.
Do not incline or move the instrument while the rotor is spinning. Do not place any object on the instrument or lean on the instrument.
Do not attempt to unlock the door forcefully while the rotor is spinning.
For operator safety, maintain a 30-cm "clearance envelope" around the instrument while the rotor is spinning. Do not store dangerous substances capable of developing flammable or explosive vapors on nor near the centrifuge.
Repairs, disassembly, and other modifications to the centrifuge are strictly prohibited unless performed by a Hitachi Koki authorized sales/service representative.
Do not use the other’s manufacturer’s rotor without Hitachi Koki’s permission.
Always use a quick-setting rotor for this ultracentrifuge. The screw-type rotors are inapplicable.
Check the chemical resistance chart attached to the rotor, and do not use any sample inapplicable to the rotor (including the buckets). Using such a sample could corrode the rotor (including the buckets).
Do not exceed the maximum rated speed of the rotor or buckets in use.
Do not use corroded, scratched or cracked rotor and buckets. Check that the rotor and buckets are free of such abnormalities before operation.
When using a swing rotor, check that the buckets are properly engaged with the rotor pins before operation. Wrong setting can cause severe damage to the instrument. Make sure that all the rotor buckets are of the same type.
If abnormal sound or vibration occurs, stop the operation immediately and contact a Hitachi Koki authorized sales/service representative.
CAUTION
Position the centrifuge on a level surface, such a table or laboratory bench that can support the weight of the centrifuge and resist vibration.
Be sure to remove the rotor from the rotor chamber when centrifuge is not used for a long time or when the machine is moved. Otherwise the drive shaft (crown) may be damaged.
Before using a rotor, be sure to read through the rotor instruction manual.
Check the chemical resistance chart attached to the rotor, and do not use any sample inapplicable to the tubes, tube caps, bottles, or bottle caps, etc. Using such a sample could corrode or deteriorate such parts.
Use the rotor tubes and bottles within their actual capacities.
Mount the rotor onto the drive shaft gently and properly.
Do not drop the rotor or apply excessive force to the drive shaft to avoid damage to the drive shaft.
Maximum rotor speed depends on the tubes or adapters to be used. Follow the instructions on the rotor instruction manual.
This centrifuge might convey vibration to a table while the rotor is rotating. If a measuring device is positioned on a table and near this centrifuge, use a measuring device carefully.
Approximately even quantities of sample in the tubes are sufficient for balancing, and extremely different sample quantities must be avoided ( Refer to Section 2-1-2 for the sample balancing).
Clean the inside of the drive hole (crown hole) of the rotor and the surface of the drive shaft (crown) of the centrifuge once a month.
Storing the rotor on the shelf is permitted if the shelf is taken necessary countermeasures against earthquakes not to drop the rotor.
ii
Page 7
SAFETY NOTICES
CAUTION
Do not pour any solution such as water, detergent, or disinfectant directly into the rotor chamber and be careful not to spill the sample into the rotor chamber. If you do so, the bearings of the drive unit might corrode or deteriorate and it might cause vacuum failure.
Do not operate the display panel (Touchscreen) using a ball-point pen.
Safety during installation and maintenance
DANGER
WARNING
CAUTION
Before removing the cover, top deck, or other component for maintenance, always turn off the POWER switch of the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
For maintenance and repairing of the rotors, tubes, etc., see the rotors and tubes instruction manual.
After installation and before any test-run, this ultracentrifuge always needs the internal check by a Hitachi Koki authorized sales/service representative.
Repairs, disassembly, and other modifications to the centrifuge are strictly prohibited unless performed by a Hitachi Koki authorized sales/service representative.
If the centrifuge is exposed to ultraviolet rays for a long time, the color of the covers may be changed or the coating may be peeled off. After use, cover the centrifuge with a cloth to protect it from direct exposure.
Electrical Safety
WARNING
CAUTION
Your centrifuge must be grounded properly to avoid electrical shock hazards.
Do not place containers holding liquid in the rotor chamber or on or near the
instrument. If they spill, liquid may get into the instrument and damage electrical components.
If the machine will not be used for a long time, turn off the main circuit breaker.
iii
Page 8
SAFETY NOTICES
Safety against Risk of Fire
WARNING
Chemical and Biological Safety
WARNING
This centrifuge is not explosion-proof. Never use explosive or flammable samples or materials that chemically react vigorously. Do not centrifuge such materials in this instrument nor handle or store them near the instrument.
Take all necessary safety measures before using samples that are toxic or
radioactive, or blood samples that are pathogenic or infectious. You use such
samples at your own responsibility.
Take all necessary safety measures when Risk Group Ⅱ materials (as identified in the World Health Organization “Laboratory Biosafety Manual”) are handled, and that more than one level of protection shall be provided in the case of materials of a higher group.
If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
If there is a possibility that the centrifuge, rotor, or an accessory is contaminated by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor, or the accessory properly before requesting repairs from a Hitachi Koki authorized sales/service representative.
It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or parts properly before returning them to a Hitachi Koki authorized sales/service representative.
Notice for an Earthquake
Depending on the magnitude, an earthquake might damage centrifuge. If you observe some abnormality, stop using the centrifuge immediately and ask for inspection by the Hitachi Koki service representative.
iv
Page 9
Table of contents
1. Description ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-1
1-1 External view of ultracentrifuge ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-1 1-2 Structure ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-2
1-2-1 Touchscreen ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-2 1-2-2 Rotor chamber・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-5 1-2-3 Safety devices ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1-6
2. Operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-1
2-1 Run preparation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-2
2-1-1 Starting up this machine ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-2 2-1-2 Preparing tubes/bottles and rotor ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-3 2-1-3 Applicable rotors ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-4
2-2 Basic operation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-5
2-2-1 Setting run conditions ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-5 2-2-2 Operational procedure ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-10 2-2-3 Acceleration rates and deceleration rates・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-13
2-3 How to use the option function・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-14
2-3-1 Programmed operation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-15 2-3-2 Step-mode operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-30 2-3-3 Selecting the rotor・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-41 2-3-4 Displaying and setting RCF ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-43 2-3-5 User login・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-47
2-4
Features of the MENU screen ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-49
2-4-1 Displaying the Run History and loading the information about run parameters ・・・・・・ 2-50 2-4-2 Rotor Catalog ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-51 2-4-3 Spin-down operation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-52 2-4-4 RTC (real-time control) operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-53 2-4-5 Customizing the settings ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-57 2-4-6 Administrator (Admin) functions・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-60
2-5 Emergency recovery from power failure・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2-73
3. Maintenance ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-1
3-1 Rotor Chamber ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-2 3-2 Drive Shaft (Crown) ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-2 3-3 Cabinet・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-2 3-4 Chamber door seal ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-3
3-4-1 Removing the door seal O-ring ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-3
3-5 Vacuum pump・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-4 3-6 Others ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3-4
v
Page 10
Table of contents
4. Troubleshooting ·········································································································· 4-1
4-1 Alarm indicators ······································································································ 4-2
4-2 User-corrected Problems ························································································ 4-4
5. Installation ···················································································································· 5-1
6. Warranty ······················································································································· 6-1
7. Specifications ·············································································································· 7-1
8. Supply list
····················································································································· 8-1
APPENDIX
Decontamination sheet
WEEE Compliance Marking for Control of Pollution Caused by Electronic Information Products
(THE PEOPLE’S REPUBLIC OF CHINA)
vi
Page 11
1-1 External view of ultracentrifuge
1. Description
1-1 External view of ultracentrifuge
The CS150NX Ultracentrifuge is tabletop type and requires a small amount of space.
Touchscreen (display/operation)
Door
POWER switch
Top deck
Unit: (mm)
Front cover
NOTE
582
590
External connection
* This size is the minimum size and depends on the installation conditions.
Fig. 1-1 External view of CS150NX Tabletop Micro Ultracentrifuge
*
408
1-1
Page 12
1-2-1 Touchscreen
1-2 Structure
1-2-1 Touchscreen
The touchscreen with color liquid crystal screen is incorporated in the CS150NX centrifuge. You can set
the run conditions, perform the operation, and display Run History, Programmed Run, and User
Customizations Screens by pressing the screen.
Fig. 1-2-1 shows the touchscreen.
Tabs of each screen
(1) Field display
(8) USER button
The following screen appears by pressing the SPEED , TIME , or TEMP button.
(3) VACUUM   button/Indicator
(2) Rotor indicator
(4) START button
Display at the normal operation
(7) RTC button
(6) PROGRAM button
(5) STOP button
(9) Keypad display
Display when setting the run conditions such as the speed etc.
Fig. 1-2-1 Touchscreen
1-2
Page 13
Functions of the Run screen
1-2-1 Touchscreen
No.
Name and symbol
Field display
(1)
SPEED column (RCF column)
TIME column
TEMP column
button
Displays various fields. The SPEED(RCF), TIME, and TEMP fields give the current status indicator in the top row and the setting indicator in the bottom row. (For setting, see Section 2-2-1.)
SPEED (
(Top row) Displays speeds in increments of 10 rpm at lower than 5,000 rpm, and in steps
of 100 rpm at 5,000 rpm or more.
(Bottom row) Sets speeds from 5,000 to maximum speed in increments of 1000 rpm.
For RCF , see Section 2-3-4.
TIME
(Top row) Displays the remaining operation time. If the running time is set to
(Bottom row) Specifies a setting in the range from 1 minute to 99 hours 59 minutes in
TEMP
(Top row) Displays in steps of 0.1 °C. (Bottom row) Sets a setting in the range from 0 °C to 40 °C in increments of 1 °C
The column is shifted from SPEED (RCF) column to RCF (SPEED)column by pressing this button.
The Acceleration rate and Deceleration rate are set by pressing this button.
Speed indicator)
The lower three digits (one, ten, and one-hundred positions) display zeros.
Maximum speed: 150,000 rpm
(running time indicator)
HOLD
steps of minutes and hours.
(temperature indicator)
When the pressure in the rotor chamber equals the atmospheric pressure, the
temperature of the inside of the rotor chamber is kept at 25 to prevent condensation.
Functions and actions
, this field displays time elapsed.
ACCEL (acceleration mode indicator)
button
Displays acceleration modes 1 through 9.
DECEL (deceleration mode indicator)
Displays deceleration modes 1 through 9, along with free coast (F).
(2) Rotor indicator
(3) VACUUM button
ROTOR CATALOG is displayed by pressing this button and you can select the desired rotor.
Press this button to turn the vacuum pump on or off. When the vacuum pump
is turned off, the pressure in the rotor chamber will change to equal the atmospheric pressure. (While the rotor is rotating, you can not turn off the
vacuum pump.) Temperature control starts as soon as the vacuum pump is turned on. Displays the following four stages according to the vacuum of the rotor chamber. 1 Atmospheric state. The vacuum pump is not activated.
2 Low vacuum. The rotor waits at 5,000 rpm until the vacuum
reaches an intermediate level.
3 Intermediate vacuum.
4 High vacuum.
NOTE
If the sample is sensitive to a
temperature rise, do not press the START button until the chamber is at a high vacuum level.
1-3
Page 14
1-2-1 Touchscreen
Functions of the Run screen
No. Name and symbol
(4)
button
(5)
button
(6)
(7)
(8)
button
button
button Displays the user name (see Section 2-4-6).
Functions of the keypad display
No. Name and symbol
(9) Keypad display
When entering the deceleration rate, is displayed in the position of .
FREE
Hold
External connection
Functions and actions
Starts rotor rotation. If VACUUM is off, pressing this button activates the vacuum pump and starts temperature control.
Stops rotor rotation.
Displays the program No. if program operation has been selected. Press this button to specify program settings (see Section 2-3-1).
Displays the start time or finish time for centrifugation. Press this button to set a start time or finish time (see Section 2-4-4).
Use the keypad display to enter numeric values for run parameters.
When entering a time: Switches from hours to minutes
Hold
FREE
Press this to save the entered setting.
Enter
Press this to close the keypad display.
When entering the operation time: Sets continuous running.
When entering deceleration conditions: Sets free coasting.
Press this to cancel input (for example, if you enter the
CE
wrong number or the wrong value for a run parameter).
X
Functions and actions
No. Name and symbol Functions and actions
(10) USB (host side)
(11) USB (device side)
Use the USB connection to output the operation history of the centrifuge to a USB flash drive.
Use for connecting optional items.
1-4
Page 15
1-2-2 Rotor chamber
r
The structure of the rotor chamber (vacuum chamber) is shown in Fig. 1-2-2.
Cooling bowl (Rotor Chamber)
Cools the rotor.
Chamber door seal
1-2 Structure
Drive shaft (crown) Sets the rotor and rotates.
Protector (Protection steel tube)
Over-speed sensor Stops the rotor when it rotates faster than the maximum allowable speed of the rotor. (Two sensors provided)
Imbalance detection device Detects abnormal balance and stops the rotor.
Fig. 1-2-2 Rotor chamber
Temperature senso Measures the temperature in the rotor chamber.
1-5
Page 16
1-2 Structure
1-2-3 Safety devices
(1) Protection of rotor chamber
Should the rotor spinning at high speed fails (or comes off the drive shaft), the safety of the operator is
ensured by the thick protector (Protection steel tube) enclosing the cooling bowl (Fig. 1-2-2).
(2) Imbalance detector
If during operation the vibration of the rotor becomes excessive due to serious imbalance or improper
bucket setting, the imbalance detector detects the situation and decelerates the rotor immediately.
However, the ultracentrifuge is designed to tolerate imbalance associated with visual balancing-it is
equipped with an imbalance tolerant drive. (For more information on the balancing of rotors, see
Section 2-1-2 "Preparing tubes/bottles and rotor".)
(3) Door lock system
The chamber door automatically locks for safety while the rotor is spinning. When the power supply is
off, the door remains locked. The door can only be opened and closed when the rotor is at rest and the
rotor chamber is vented. To open the door in the event of a power failure, see Section 2-5 "Emergency
recovery from power failure".
(4) Overspeed detector
This ultracentrifuge incorporates a detector designed to prevent the rotor from spinning at a speed
exceeding the maximum allowable speed.
Two independent microprocessors (CPUs) check the rotor for overspeed, thus increasing safety further
(dual CPU overspeed preventive mechanism).
The first CPU detects overspeed and performs control and display. Should the rotor be set to a speed
exceeding the permissible speed, this CPU detects an alarm massage of “SPEED” from the low speed
range (about 2,000 rpm), and stops the rotor.
(But this second CPU does not display an alarm message, because it is not connected to the display-
performing CPU. Should the alarm device be activated, pressing START button will not run the
instrument. Turn off the POWER switch, wait for several minutes, turn the POWER switch on again,
then pressing START button.)
1-6
Page 17
Other features
A
2. Operation
2. Operation
The CS150NX Ultracentrifuge is capable of operation in more than one mode to meet a wide range of
applications. The outline of each available mode is given below:
Brief description Reference
Section 2-2,
Normal operation
Programmed operation
Speed
Time
You can store set run conditions in memory for later use in repeated operation.
Store
Program No.
Call
Basic operation
Section 2-3-1, Programmed operation
フ  
Step-mode operation
ァ  
ン  
ク  
シ  
RCF (centrifugal force) value display setting
ョ  
Speed
Time
Speed
Time
More than one normal operation can be combined into a sequence of operations or step for successive centrifugation.
Speed
Time
This feature calculates centrifugal force (RCF) values from set speed. It can also calculate reversely, i.e., finding speed from such values.
RCF value
setting
Speed
Time
Section 2-3-2, Step-mode operation
Section 2-3-4, Displaying and setting RCF
Spin-down operation
RTC (real-time control) operation
Spin-down operation is useful to remove the adhered samples on the interior walls of the tubes.
Run starts or completes at a required date and time.
Press START button.
Speed
finish date and
time required
Time
2-1
Section 2-4-3, Spin-down operation
Section 2-4-4, RTC (real-time control) operation
Page 18
2-1 Run preparation
2-1 Run preparation
WARNING1. This centrifuge is not explosion-proof. Never use explosive or flammable samples, or
materials that chemically react vigorously. Do not centrifuge such materials in this
instrument nor handle or store them near the instrument.
2. Take all necessary safety measures before using samples that are toxic or radioactive,
or blood samples that are pathogenic or infectious. You use such samples at your
own responsibility.
CAUTION :Do not place containers holding liquid in the rotor chamber, on the centrifuge, or near the
centrifuge. If spilt, liquid might get into the instrument and damage electrical and
mechanical components.
2-1-1 Starting up this machine
Before setting run conditions, display the Run screen (Screen for setting run conditions)
(1) Displaying the Run screen (Screen for setting run conditions)
1. Turn on the POWER switch.
2. The initial screen appears.
Initial screen
3. The Run screen appears.
2-2
Run screen
Fig.2-1-1 Initial screen and Run screen
Page 19
2-1 Run preparation
A
2-1-2 Preparing tubes/bottles and rotor
The CS150NX allow you to balance, by eye, tubes or bottles containing a sample solution and then centrifuge
them. Make sure that the approximate difference between meniscus levels of sample solution in tubes or
bottles is within 5 mm (See fig. 2-1-2).
Within 5mm
For an S140AT rotor, an S100AT5 rotor, and an S50
rotor, set the approximate meniscus level difference to
within 3 mm, and for an S110AT rotor and an S80AT3
rotor, set it to within 4 mm.
Tubes
Fig. 2-1-2 Balancing tubes/bottles containing a sample solution
CAUTION Do not run the ultracentrifuge in an extremely imbalance condition. This might cause a
mechanical failure. When balancing tubes or bottles by eye, alarm message
“IMBALANCE” might appear. Balance tubes or bottles more accurately again if alarm
message “IMBALANCE” appears.
Depending on the type of tube or rotor combined with this machine, an excessively low liquid level may limit
the speed or crush the tube.
• When sealed tubes are used, fill the tubes with liquid to the full level.
CAUTION : 1. Before using a rotor, read the rotor instruction manual carefully.
2. Do not use any corroded, scratched, cracked, or otherwise damaged rotor or bucket.
Before operation, always check if there is no corrosion and damage on the rotor
surface.
3. Before running a swing rotor, make sure that each bucket is hooked on a pin securely.
Poor setting might seriously damage the instrument. Even though all of them are not
used, make sure to install all buckets at any time. Make sure that all the rotor buckets
are of the same type.
4. Some tubes and adapters are inapplicable at the maximum speed of the rotor.
Refer to the rotor instruction manual carefully.
5. The S58A rotor, S55A rotor, and S50A rotor can handle sample whose mean density is
1.2 g/ml or less.
Do not rotate this rotor at its maximum speed when using a sample whose mean
density is over 1.2 g/ml. Refer to the rotor instruction manual (Part No. S999276) for
how to limit the maximum speed when using a sample whose mean density is
over 1.2 g/ml.
2-3
Page 20
2-1 Run preparation
2-1-3 Applicable rotors
Rotors applicable to the CS150NX are limited to the quick setting type illustrated below. A quick setting rotor
can be installed by merely placing it on the drive shaft (crown) in the rotor chamber. Screw-type rotors used in
older models (CP120H/CP100H, CS120/CS100, CS120EX/CS100EX, and CS120FX/CS100FX) are
inapplicable to the CS150NX.
Rotor set handle
Cap (blue)
(without a screw hole)
Cover (black)
Quick setting type
(1) Applicable rotor
CAUTION : 1. Always use a quick setting rotor for this ultracentrifuge. The screw-type rotors are
inapplicable.
Quick-setting rotors are not applicable in older models (CP120H/CP100H,
CS120/CS100, CS120EX/CS100EX, and CS120FX/CS100FX).
2. Install the rotor carefully and securely on the drive shaft in the rotor chamber. Forcefully
moving the shaft sideways or dropping the rotor onto the drive shaft may damage the
shaft.
3. The S52ST swinging bucket rotor and S50ST swinging bucket rotor are specifically
designed for the CS150NX, CS-GXseries, and CS-GXL series ultracentrifuges.
These rotors cannot be used in any other centrifuge.
Screw – type for older models
(2) Inapplicable
Screw cap (with a screw hole)
Cover (blue)
2-4
Page 21
2-2 Basic operation
2-2 Basic operation
WARNING: Do not incline or move the instrument while the rotor is spinning.
Do not place any object on the instrument or lean on the instrument.
CAUTION: 1. Do not press the touchscreen with a sharp-pointed object such as a ballpoint pen.
2. If abnormal sound is heard during the operation, stop the operation immediately
and contact a Hitachi Koki authorized sales/service representative.
2-2-1 Setting run conditions
This section will first describe the screen for basic operation (the Run screen). For the display at the normal
operation and the display when setting the run conditions, refer to Section 1-2-1.
[Run Screen]
The screen for displaying run conditions and operational status is called the Run screen.
Speed, time, and temperature are displayed in two rows: the top row displays the current actual run
conditions, while the bottom row displays the set run conditions.
The acceleration (ACCEL) and the deceleration (DECEL) fields display set conditions.
Operation state display field
Setting value display field
Fig. 2-2-1 Run screen
2-5
Page 22
2-2 Basic operation
y
[Display and operations when entering the run parameters]
The keypad display appears by pressing the SPEED column, TIME column, button.
ACCEL: DECEL:
(1) Press the column of the
desired item to turn the first digit blue.
Color of the first digit:Blue
(2) Enter the desired numeric value
with the keypad display. (e.g.) SPEED:150,000rpm
Press the .
1 5 0
(3) If you do not want to change
any other setting, press the
button on the
Enter
keypad display.
If you do want to change
other setting, press the
column of the desired item,
and the pressed column will
become the input wait state.
The new setting is displayed
on the setting value displa
field.
Keypad display]
[
Setting value display field
Fig. 2-2-2 Setting value display field
2-6
Page 23
2-2 Basic operation
The next page describes how to set run condition by citing some examples.
NOTE (1) If you enter the wrong value, press the CE button and then enter the correct value.
If you have pressed the Enter button, repeat step (1) of the previous page, and then enter
the correct value.
(2) When setting the two or more of three run conditions (SPEED, TIME, and TEMP), you do not
have to press the Enter button after each setting. You can store the setting by pressing the
desired item.
(3) If the system is running in (HOLD) and you want to set it to shut down at a future time, enter a
new time setting while the instrument is in operation; enter the sum of the time elapsed plus the
time remaining. If, for example, this machine has run continuously for five hours and you want
to stop it one and a half hours later, press the TIME column, then enter
.
6
:
3
0
2-7
Page 24
2-2 Basic operation
e
y
s
f
• How to set speed, running time, temperature, and other parameters
Here are some examples and descriptions:
Setting item
Typical setting 150,000rpm 2 hours 30 minutes 4°C
1 Press the SPEED,TIME,
or TEMP column to show the keypad display. The last entered value
2
is displayed on each column. Press the column of the desired item again if the color of the first digit is white. Go to procedure 3 if the color of the first digit is blue.
3 Enter the desired numeric
value with the keypad display.
7 8 9
4 5 6
1 2 3
0
Operation procedure
Entered numbers ar moved to the left ever time a new number i entered.
Hold
Enter
RPM
(SPEED)
Color:Blue
1 5 0 2
The last three digits
are fixed.
CE
×
Running time
TIME
Color:Blue
3 0
Press the
button to turn the number of minutes blue.
Temperature
(TEMP)
Color:Blue
4
For a continuous run, press the
Hold
button.
Make a check, then if
4
you want to change other setting, press
Set it to 150,000rpm.
Set it to 2:30.
(2 hours 30 minutes)
Set it to 4
°C
the column of the desired item. If you do not want to change any other setting, press the
Enter
button. Use the
CE
button to cancel input.
Can be set to any value in the range from 0 to 40 °C in increments o 1 °C.
Setting range and units
Can be set to any value in the range from 5,000 rpm to maximum speed in increments of 1,000 rpm.
Can be set to any value up to 99 hours 59 minutes in increments of 1 minute.
2-8
Page 25
2-2 Basic operation
Setting item
Acceleration
(ACCEL)
Typical setting 9 7
1
Press the button
ACCEL:5 DECEL:5
to show the keypad display.
Press the column of the
2
desired item.
Enter the desired numeric
3
value with the keypad display.
9 7
Deceleration
(DECEL)
Operation procedure
5
7 8 9
4 5 6
1 2 3
0
Make a check, then if you want to change other setting, press the column of the desired item. If you do not want to change any other setting, press the button.
Enter Use the button to cancel input.
Setting range and units
CE
CE
FREE
Enter
×
For free coast,
press the button.
FREE
Set it to 9. Set it to 7.
7
1-9 1-9
+
free coast(FREE)
2-9
Page 26
2-2-2 Operational procedure
2-2-2 Operational procedure
Given below is a description of the operational procedure for a normal run.
Before starting up this machine, carefully read the operation manual for your rotor and make sure
NOTE
that you have selected the correct type of tubes and entered the correct amount of sample.
Process Touch panel operation Screen display and notice
1 Turn on the POWER switch on
this machine.
2 Install the rotor. (the quick
setting rotor.)
3 Set run conditions.
Press the VACUUM button.
4
(You can omit this step.)
The touchscreen is displayed.The door is unlocked.
Install the rotor securely on the crown. (It does not have to be
screwed into the crown.)
Always use a quick setting rotor (see Section 2-1-3
“Applicable rotors”).
See Section 2-2-1 “Setting run conditions” and set run
conditions.
The machine starts evacuating the rotor chamber.Temperature control starts.The degree of vacuum in the rotor chamber is displayed on
the vacuum indicator on the VACUUM button.
(1) In a low vacuum
(1 indicator)
(2) In an
intermediate vacuum
(2 indicator)
(3) In a high vacuum
(3 indicator)
Press the START button
5
If the rotor compartment has moisture or frost on it, it takes a
long time to reach an intermediate high vacuum. In that case, wipe it off with a clean, dry cloth or sponge.
If the sample is sensitive to a temperature rise, do not press
the START button until the chamber is at high vacuum level.
The START button blinks and the rotor starts turning.
.
The timer begins operating. (When the actual run timer is
activated, the timer begins operating after the speed reaches the set speed.)
When the speed reaches the set speed, the START
button lights up.
This ultracentrifuge waits at 5,000 rpm until an intermediate
vacuum is reached.
2-10
Page 27
Process Touch panel operation Screen display and notice
The specified centrifugation
6
time elapses (time-out). Or press the STOP button.
The rotor stops.
7
8
Press the VACUUM button.
The STOP button blinks and the rotor starts decelerating.
The STOP button lights up.
The stop signal sounds to indicate that the rotor has stopped.The vacuum stops, the air leak valve activates, and the rotor
chamber reaches atmospheric pressure.
The door unlocks, and is able to be opened and closed.
2-2-2 Operational procedure
Take out the rotor.
9
Stop the rotor completely before taking it out.
NOTE : When the rotor chamber is vacuumed insufficiently before starting operation or the ambient
temperature is low (10°C or below), the vacuum waiting time at 5,000rpm may become longer. Also,
during acceleration up to the set speed, the instrument may become a vacuum waiting state.
Therefore, before starting operation, vacuum the rotor chamber sufficiently (approx. 15 minutes)
and then press the VACUUM button and check that the displayed vacuum is at a high vacuum
level (3 indicator).
2-11
Page 28
2-2-2 Operational procedure
r
r
A
r
Fig. 2-2-3 summarizes the operational procedure.
START
Turn on the POWER switch.
Open the chamber doo
and install the rotor.
Close the door and set
the run conditions.
Press the VACUUM
*
button.
Press the START button.
For information on how to set the run conditions, see Section 2-2-1.
(1) Vacuum pump is ON.
Temperature control starts. (2) If the sample in the rotor is sensitive to a temperature rise, wait until the rotor chamber is
at high vacuum level.
The rotor is rotating at set speed.
Reaches the set time o press the STOP button.
Rotor stops.
Press the VACUUM button.
Both of the vacuum pump and temperature control turn off.
ir vent opens and the doo
lock is released.
Open the chamber door, remove the rotor, then turn ultracentrifuge power off.
End
*This step may be omitted, in which case pressing the START
button later in the procedure turns on the vacuum pump and the rotor stays at 5000 rpm to wait for chamber vacuum to reach its intermediate level.
Fig. 2-2-3 Operational procedure
2-12
Page 29
2-2-3 Acceleration rate and deceleration rate
r
f
2-2-3 Acceleration rate and deceleration rate
In order to meet various experimental protocols, the acceleration and deceleration rates can be adjusted. The figure and table below show the relationship between ACCEL/DECEL code numbers selected and resulting approximate acceleration/deceleration times.
Speed (rpm)
1 - 9 : ACCEL/DECEL code numbers
(No Brake)
Time
Code
No.
Acceleration time
(minutes)
from rest to 5000 rpm
Minimum time*
0.5 1
1.5 2
2.5 3
3.5 4
-
*: The minimum time is the one that occurs
when the rotor is being accelerated o decelerated with the maximum torque o the driving motor. This time varies with the type of rotor in use.
Deceleration time
(minutes)
from 5000 rpm to rest
Minimum time*
Coasting
deceleration
NOTE: The times for acceleration and deceleration may be longer than the values listed above,
depending upon the type of rotor in use.
Typical examples of application of acceleration and deceleration rates
Code No.
Type of centrifugation
Density gradient centrifugation using a vertical rotor
DNA separation by CsCl isopycnic separation (self-forming gradients)
Pelleting using a fixed angle rotor
Density gradient centrifugation using a swinging bucket rotor
ACCEL DECEL
57
97
99
5 to 8 5 to 8
The sample and gradient in tubes reorient during acceleration and deceleration. Therefore, the sample and gradient can become mixed, especially in wide tubes, if you use rapid acceleration or deceleration. You can operate at maximum acceleration because the density gradient is not formed during the run. As for the deceleration, it is better to decelerate slowly to obtain sharp bands. Rapid pelleting of samples is possible (the run time decreases). The sample and gradient do not reorient. Therefore, mixing of the layers is less than that in the case of using a vertical rotor. But it is safe not to accelerate or decelerate the rotor by selecting minimum time.
Characteristic of separation
2-13
Page 30
2-3 How to use the OPTIONS function
2-3 How to use the option function
This ultracentrifuge incorporates a number of features, such as step-mode and other programmed running, display and setting of centrifugal force, and RTC (real-time control) that can run the centrifuge at a required date and time. Buttons for these features are displayed and specified on the Run screen.
RTC button
SPEED/RCF button
PROGRAM
:
button
SPEED/RCF
:
button
Fig. 2-3-1 Option buttons
Programs, stores, and recalls run conditions. This feature also offers a step-mode operation: a continuous run of multiple run conditions.
This feature is used to cause the system to automatically calculate and display an RCF value. It also sets an RCF value and calculates the speed. RCFmax indicates the maximum centrifugal force for the maximum radius Rmax of the rotor which is used. RCFavg indicates the average centrifugal force for the average radius
Ravg of the rotor which is used.
PROGRAM button
Ravg
Rmax
RTC button
:
Sets a start time or a finish time and runs
Fig. 2-3-2 Rotor radius
the ultracentrifuge at a desired date and time.
The above features can be used in combination.
NOTE
To perform a combination of PROGRAM and RTC , first set PROGRAM and then set the
RTC .
Once RTC is activated, you cannot change the run time. You therefore cannot activate
PROGRAM .
2-14
Page 31
2-3-1 Programmed operation
2-3-1 Programmed operation
When a centrifugal condition is to be used frequently, entering the same condition every time you want to
perform centrifugation is inconvenient.
This ultracentrifuge has a programmed operation feature that stores run conditions. Storing run conditions
which you often use allows you to call those conditions however often you may wish, thus saving time in
setting. (Even while the POWER switch is OFF, this centrifuge retains the conditions entered.)
This centrifuge incorporates the program areas indicated below. It has twenty memory areas and nine steps
in each memory unit.
Running this machine with each memory unit retaining multiple steps will allow you to change the speed, run
time, temperature, and others while in operation.
(Step-mode operation)
Memory 1 Step 1 Step 2 ······ Step 9
Memory 2 Step 1 Step 2 ······ Step 9
Memory 3 Step 1 Step 2 ······ Step 9
·
·
·
·
·
Memory 20 Step 1 Step 2 ······ Step 9
·
·
·
·
·
Fig. 2-3-3 Program areas
·
·
·
·
·
······
······
······
······
······
·
·
·
·
·
2-15
Page 32
2-3-1 Programmed operation
p
r
(Basic operation of the programmed operation feature)
Press the PROGRAM
button.
Program No. which stores the
arameters is displayed.
run
If you create, change, o
delete a program, see the
below.
If you create or change a program,
If you delete a program,
If you load the run parameters,
into the centrifuge, see the
below.
Programmed operation possible
Press the Run button
to go back to the Run
screen. The loaded run
parameters are displayed on
the Run screen.
Press the Select button.
Press the Delete button.
*
Press the column of the
desired Program No
.
NOTE You cannot create, change, or delete a program while in running. Perform these operations while not in
running. However, you can search the PROGRAM screen every time.
*If you do not know Program No. of the program which you need, see Section 2-3-1 (2) (b).
2-16
Page 33
(1) Programming procedure for run conditions (creating or changing)
Shown below is the procedure for storing (creating) or changing an run condition.
If you want to store the user name in the program, you have to load the user name into the centrifuge
in advance (see Section 2-4-6 (1)).
Process Touch panel operation Screen display and notice
Press the PROGRAM
1
button on the Run screen.
2-3-1 Programmed operation
PROGRAM button
PROGRAM screen
appears.
If Program name and Parameter are displayed in the
column of the Program No., it indicates that the run parameters has been already stored in this program.
If you want to create the program, press the column where
Program name and Parameter have not been displayed.
2-17
Page 34
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Press the column of the
2
program No. that you want to create (change) the program. Check that the pressed column turns blue and press the
Select button.
If you want to store the run parameters as Program No.4., press the column of the Program No.4.
Select button
If you want to change the contents of the program, change the desired item (see the below).
Procedures for entering the
3
program name are described as follows.
a) Press the column of the
program name in the PROGRAM screen.
b) Enter the program name in
the Program name screen.
c) Press the Enter button.
The PROGRAM
screen appears.
Program name column
User name column
Rotor column
a) The Program name
screen appears.
b) If you enter the
“DNA”,
the “DNA” is
displayed on the
column of the
program name.
2-18
c) The PROGRAM
screen returns to view.
Page 35
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Procedures for selecting the
4
user name are described as follows.
a)The PROGRAM
User name screen appears.
aPress the column of the
user name in the PROGRAM screen.
bSelect the user name and
press the column of the selected user name.
Then press the Enter button.
bThe PROGRAM
screen returns
to view.
If you selected the
“Hitachi-Koki”,
the “Hitachi-Koki”
is displayed on
the column of the
user name.
2-19
Page 36
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Procedures for selecting the
5
rotor are described as follows.
aPress the column of the rotor
in the PROGRAM screen.
bPress the Rotor Catalog
button in the PROGRAM Rotor Management screen.
You can select your rotor in the PROGRAM Rotor
Management screen, if you registered your rotor in advance (see Section 2-4-6 (5) “Rotor Management”).
a The PROGRAM
Rotor Management screen appears.
cSelect the rotor in the
PROGRAM ROTOR CATALOG screen and press the column of the selected rotor.
Then press the Enter button.
bThe PROGRAM
ROTOR CATALOG
screen appears.
cThe PROGRAM
screen returns
to view.
If you selected the
“S150AT”,
the “S150AT” is
displayed on
the column of the
Rotor.
2-20
Page 37
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Procedures for selecting the
6
run parameters are described as follows.
Entering the SPEED value If you want to enter the SPEED value as the parameter, press the Edit button after checking that the color of the SPEED column is blue.
SPEED column
Edit button
The Enter
PROGRAM
screen appears.
Entering the RCF value If you want to enter the RCF value as the parameter, press the RCFmax (RCFavg) column. Press the Edit button after checking that the color of the RCFmax (RCFavg) column turns blue.
RCFmax column RCFavg column
Edit button
2-21
Page 38
A
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Procedures for entering the run
7
parameters are described as follows.
aPress the Edit button
after checking that the color of the STEP1 column is blue. Enter the run parameters on the keypad display (see Section 2-1-2). If you want to enter the acceleration and deceleration rate, press the
ACCEL/DECEL button
OK button
ACCEL/DECEL button. .
Example: SPEED:150,000 rpm TIME:2hr.30min. TEMP:20°C
CCEL:9
DECEL:9
b) Press the OK button
after entering the run parameters. (If you want to change the run parameters, press the column of the
Run button
Edit button desired STEP. Press the Edit button after checking that the pressed column is blue and enter the run parameters.)
To enter the run parameters in multiple steps in a step-
mode operation, continue the operation from the above (see Section 2-3-2 “Step-mode operation”).
You can not skip a step. (You can not enter anything in
STEP 2 with nothing entered into STEP1.)
a) The keypad
display is shown.
b)The entered run
parameters are displayed in the Enter PROGRAM screen.
8
Press the Run button in the PROGRAM screen.
The Run screen
appears.
Program name
NOTE (1) If you make and store changes in a column that already stores run parameters, the previous
parameters are replaced by the new parameters.
(2) You cannot store a run parameter while in running (while the rotor is rotating). Always perform
this function while not in running.
2-22
Page 39
2-3-1 Programmed operation
(2) How to perform a programmed operation
Shown below is how to perform a “programmed operation”, that is, how to call a stored set of run parameters and run this centrifuge accordingly.
(a) If you know the program name you need
Process Touch panel operation Screen display and notice
Turn on the POWER switch on
1
the centrifuge.
Install a rotor. Install the rotor securely on the shaft. (The rotor does not
2
Touchscreen appears. The door unlocks.
have to be screwed into the shaft.)
Always use a quick setting rotor (see Section 2-1-3
“Applicable rotors.”)
3
Press the PROGRAM button on the Run screen.
PROGRAM button
The PROGRAM
screen appears.
2-23
Page 40
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Press the column of the
4
desired program name and press the Run button after checking that the pressed column turns blue.
Run button
STEP1-3
In this case, this program contains three steps.
In the Run screen, the run parameters for the first step in this program are displayed.
Run the centrifuge under
5
normal operation without making changes to the run parameters.
Program name
The program name is displayed on the PROGRAM
button indicator.
If the “STEP:1-3” is displayed on the PROGRAM
button indicator, this means that the program you have just called stores multiple run parameters (step-mode feature). For details, see Section 2-3-2 “Step-mode operation”.
Run this centrifuge according to Section 2-2-2 “Operation
procedure”.
If you make changes to the run parameters (such as
SPEED and TIME) after calling a program, the program you have just called is canceled. You must call it again to use it.
2-24
Page 41
(b) If you do not know the program name you need
Process Touch panel operation Screen display and notice
2-3-1 Programmed operation
Turn on the POWER switch on
1
the centrifuge.
Install a rotor. Install the rotor securely on the shaft. (The rotor does not
2
Touchscreen appears. The door unlocks.
have to be screwed into the shaft.)
Always use a quick setting rotor (see Section 2-1-3
“Applicable rotors.”)
3
Press the PROGRAM button on the Run screen.
PROGRAM button
The PROGRAM
screen appears.
2-25
Page 42
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Press the column of the
4
desired program name whose contents you want to check. Then press the Select
button after checking that the pressed column turns blue.
Check the user name and rotor name in the Program screen and decide whether this selected program is the program that you need or not.
Select button
If you think that the selected program is the program you
need, press the Edit button to check the run
parameters. If you do not think that the
selected program is the program you need, press the
x button to switch to the previous screen.
Edit button
After checking the run parameters, press the
Run button if the selected program is the program you need.
STEP1-3
In this case, this program has three steps.
In the Run screen, the run parameters for the first step in this program are displayed.
Run button
Program name
The program
name is displayed on the PROGRAM
button indicator.
If the “STEP:1-3” is displayed on the
RROGRAM
button indicator, this means that the program you have just called stores multiple run parameters (step­mode feature). For details, see Section 2-3-2 “Step-mode operation”.
2-26
Page 43
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Run the centrifuge under
5
normal operation without making changes to the run parameters.
Run this centrifuge according to Section 2-2-2 “Operation
procedure”.
If you make changes to the run parameters (such as
SPEED and TIME) after calling a program, the program you have just called is canceled. You must call it again to use it.
NOTE (1) To perform a combination of a programmed run with RTC (real-time control) (see Section 2-4-
4 “RTC operation”), call a programmed memory unit, then set RTC. The system will then
calculate the total of the running times of all steps of the programmed run and calculate the
start time for RTC. Therefore, cannot call the program memory after setting RTC.
2-27
Page 44
x
2-3-1 Programmed operation
(3) Deleting a program
This section explains how to delete a program. To delete a program, delete all steps in that program.
You cannot delete a program while in running (while the rotor is rotating). Always perform this
NOTE
function while not in running.
Process Touch panel operation Screen display and notice
1 Press the PROGRAM button
on the Run screen.
PROGRAM button
Press the column of the
2
program name which you want to delete. Then press the
Delete button after checking that the pressed column turns blue.
The PROGRAM
screen appears.
The deletion
confirmation bo
is displayed.
2-28
Page 45
2-3-1 Programmed operation
Process Touch panel operation Screen display and notice
Press the Yes button on
3
The program is
the deletion confirmation box.
deleted.
2-29
Page 46
2-3-2 Step-mode operation
2-3-2 Step-mode operation
This ultracentrifuge incorporates a step-mode operation feature, which stores two or more run conditions in
one program memory area and switches between different values of speed, running time, temperature, and
other parameters while in operation. This centrifuge can store up to nine steps.
This section explains how to make settings by citing some examples.
(1) How to activate a step-mode operation
[Typical settings]
Shown below is the example of a three-step run and how to activate a step-mode operation.
Step 1 Step 2 Step 3
Speed 150,000 rpm 120,000 rpm 100,000 rpm
Run time 1 h 30 min 1 h 30 min
Temperature 20°C 20°C 20°C
Acceleration mode 9 9 9
Deceleration mode 9 9 7
Step 1
150,000
Step 2
120,000
Step 3
100,000
Speed (rpm)
1 h 30 min
1 h
30 min
Run time
Fig. 2-3-2 A typical step-mode run
2-30
Page 47
2-3-2 Step-mode operation
Process Touch panel operation Screen display and notice
Press the PROGRAM
1
button on the Run screen.
PROGRAM button
The PROGRAM
screen appears.
If a Program name and Parameter are displayed in the
column of a Program No., run parameters have already been stored for that program.
If you want to create a new program, press a column that
does not display a Program name or Parameter.
2-31
Page 48
2-3-2 Step-mode operation
Process Touch panel operation Screen display and notice
Press the column of the
2
program No. for which you want to create (or change) a program, and make sure that the pressed column turns blue.
Then press the Select button.
If you want to store the run parameters as Program No.4., press the column of the Program No.4.
Select button
The PROGRAM
screen appears.
Enter a program name as
3
follows:
a) Press the column of the
program name in the PROGRAM screen.
b) Enter the program name in
the Program name screen.
c) Press the Enter button.
Program name column
User name column
Rotor column
a)The Program name
screen appears.
b)If you enter the
“DNA-STEP”,
these characters
are displayed on
the column of the
program name.
c)The PROGRAM
screen reappears.
2-32
Page 49
2-3-2 Step-mode operation
Process Touch panel operation Screen display and notice
Select a user name as follows:
4
a) The PROGRAM
User name
aPress the column of
screen appears.
User name in the PROGRAM screen.
bPress the column of the user
name you want to select.
Then press the Enter button.
bThe PROGRAM
screen reappears.
If you selected the
“Hitachi-Koki”,
these characters
are displayed on
the column of the
user name.
2-33
Page 50
2-3-2 Step-mode operation
Process Touch panel operation Screen display and notice
Select a rotor as follows:
5
aPress the column of the rotor
in the PROGRAM screen.
bPress the Rotor Catalog
button in the PROGRAM Rotor Management screen.
If you registered your rotor in advance (see Section 2-4-6
(5) “Rotor Management”), you can select it in this PROGRAM Rotor Management screen,
a The Program
Rotor Management screen appears.
cIn the PROGRAM ROTOR
CATALOG screen, press the column of the rotor you want to select.
Then press the Enter button.
bThe PROGRAM
ROTOR CATALOG
screen appears.
cThe PROGRAM
screen reappears.
If you selected the
“S150AT”,
these characters
are displayed on
the column of the
Rotor.
2-34
Page 51
2-3-2 Step-mode operation
Process Touch panel operation Screen display and notice
Select run parameters as
6
follows:
To enter a SPEED value If you want to enter a SPEED value as a parameter, press the SPEED column and make sure that it turns blue. Then press the Edit button.
SPEED column
Edit button
The Enter
PROGRAM
screen appears.
To enter a RCF value If you want to enter a RCF value as a parameter, press the RCFmax (or RCFavg) column and make sure that it turns blue. Then press the Edit button.
RCFmax column RCFavg column
Edit button
2-35
Page 52
A
2-3-2 Step-mode operation
A
A
Process Touch panel operation Screen display and notice
Enter run parameters as
7
follows: aMake sure that the STEP 1
column is blue, and then press the Edit button. Enter the run parameters for the STEP 1 on the keypad display (see Section 2-1-2). If you want to enter an acceleration or deceleration rate, press the ACCEL/DECEL button.
Run parameters of the STEP 1: SPEED: 150,000 rpm TIME: 1hr.30min. TEMP: 20°C
CCEL: 9
DECEL: 9
ACCEL/DECEL button
OK button
b) After entering the run
parameters, press the OK button (If you want to change run parameters, press the column of the desired STEP, and make sure that it turns blue. Then press the Edit button.
c) Enter the run parameters for
STEP 2 on the keypad
Run button
You can not skip a step. (If you enter nothing in STEP 1,
you cannot enter anything in STEP 2.)
Edit button
display in a similar way.
Run parameters of the STEP 2: SPEED: 120,000 rpm TIME: 1hr. TEMP: 20°C
CCEL: 9
DECEL: 9
d) Enter the run parameters for
the STEP 3 on the keypad display in a similar way.
Run parameters of the STEP 3: SPEED: 100,000 rpm TIME: 30min. TEMP: 20°C
CCEL: 9
DECEL: 9
a) The keypad
display appears.
b)The run
parameters
entered for
STEP 1 are
displayed in the
Enter PROGRAM
screen.
c)The run
parameters entered for STEP 2 are displayed in the Enter PROGRAM screen.
d)The run
parameters entered for STEP 3 are displayed in the Enter PROGRAM screen.
2-36
Page 53
Process Touch panel operation Screen display and notice
8
Press the Run button in the
The Run screen
PROGRAM screen.
Run button
2-3-2 Step-mode operation
appears.
Run the centrifuge under
9
normal operation without changing to the run parameters.
Program name
Run the centrifuge according to Section 2-2-2 “Operation
procedure”.
If you change to a run parameters (such as SPEED or
TIME) after calling a program, the program you have just called is canceled. You must call it again to use it.
Pressing the START button rotates the rotor under the run
parameters of STEP 1.
When STEP 1 ends, the system moves to STEP 2 and
STEP 3 automatically.
2-37
Page 54
2-3-2 Step-mode operation
NOTE (1) Step display
The PROGRAM button indicator displays steps as follows:
In this case, this program contains three steps. In the Run screen, the run parameters for the first step in this program are displayed.
When the system finishes separating step 1 and moves to step 2,
STEP:1-3
STEP:2-3
When the system moves from step 2 to the final step 3,
STEP:3-3
Thus, you can see at a glance how many steps are stored in the specified memory and which step the centrifuge is following.
(2) You cannot store a run parameter while in running (while the rotor is rotating). Always perform
this function while not in running.
(3) To perform a combination of a step mode run with an RTC (real-time control) run (see Section 2-
4-4 “RTC (real-time control) operation”), call a program memory unit, then set RTC.
The system then calculates the total of running times of all steps of the programmed run and
calculates the startup time for RTC. You therefore cannot call a program memory unit after setting
RTC.
2-38
Page 55
(3) Deleting a program
x
This section explains how to delete a program. To delete a program, delete all steps in that program.
You cannot delete a program while in running (while the rotor is rotating). Always perform this
NOTE
function while not in running.
Process Touch panel operation Screen display and notice
1 Press the PROGRAM button
on the Run screen.
2-3-2 Step-mode operation
PROGRAM button
Press the column of the
2
program name which you want to delete. Then press the
Delete button after checking that the pressed column turns blue.
Press the Yes button on
3
the deletion confirmation box.
The PROGRAM
screen appears.
The deletion
confirmation bo
is displayed.
The program is
deleted.
2-39
Page 56
2-3-2 Step-mode operation
(3) Other procedures
1) Making changes to the run parameters Call the memory unit you want to make changes to and make the changes. Alternatively, delete the memory unit, then store a new set of run parameters. To make a change that will result in fewer steps, delete the memory unit and then enter a new set of run conditions and store them.
2) Running the centrifuge starting from an intermediate step You cannot run the centrifuge starting from an intermediate step in a memory unit that contains multiple steps. Store (register) run parameters for the intermediate step and later steps in another memory unit. Then call the memory unit and run it.
3) What if a SPEED alarm goes on? If a step stores a speed exceeding the maximum allowable speed of your rotor, the system will detect it in the STEP1 run and display the SPEED alarm. Double-check the speed of all steps and correct any wrong ones.
4) Stopping the centrifuge in operation Press STOP button. The rotor stops and the system does not move to the next step.
2-40
Page 57
2-3-3 Selecting the roto
r
r
r
2-3-3 Selecting the rotor
This ultracentrifuge stores the maximum and average radii of each rotor in the internal memory. Setting a speed causes this centrifuge to automatically calculate and display the RCF (relative centrifugal force) value, while setting an RCF value causes the centrifuge to automatically calculate and display the speed. You can control the rotor data such as total operation hours and number of runs by registering your rotors in the centrifuge.
(1) Procedure for selecting the rotor
Process Touch panel operation Screen display and notice
Check that the desired rotor is
1
displayed in the ROTOR indicator of the RUN SCREEN. When the desired rotor is not displayed, pess the ROTOR indicator.
ROTOR indicato
Check whether the desired
2
rotor is displayed in the Rotor Management screen or not. When the desired rotor is displayed, press the column of the desired rotor and press the OK button. When the desired rotor is not displayed, press the Rotor Catarog
button.
Press the column of the
3
desired rotor on the ROTOR CATALOG screen. Check that the pressed column turns blue and press the Enter
button.
The Rotor
Management screen appears.
Nothing is displayed in the ROTOR indicate by pressing the Cancel button.
You can select your rotor in the Rotor Management
screen, if you registered your rotor in advance (see Section 2-4-6 (5) “Rotor Management”).
The ROTOR
CATALOG screen appears.
The pressed column turns from white to blue.
2-41
Page 58
r
2-3-3 Selecting the rotor
Process Touch panel operation Screen display and notice
The desired rotor is displayed
4
in the ROTOR indicator of the Run screen.
The Run screen
returns to view.
ROTOR indicato
NOTE (1) Selecting the rotor is required to display and set an RCF value (see Section 2-3-4 “Displaying
and setting RCF”).
(2) You can select your rotor in the Rotor Management screen, if you registered your rotor in
advance (see Section 2-4-6 (5) “Rotor Management”).
(3) Selecting the rotor is required when the rotor lockout function is enabled (see Section 2-4-6 (3)
“Rotor Lockout”).
2-42
Page 59
r
2-3-4 Displaying and seting RCF
2-3-4 Displaying and setting RCF
This ultracentrifuge stores the maximum and average radii of each rotor in the internal memory. Setting a speed causes this centrifuge to automatically calculate and display the RCF (relative centrifugal force) value, while setting an RCF value causes the centrifuge to automatically calculate and display the speed. Given below is a description of how to display and set RCF.
(1) How to display an RCF value
Process Touch panel operation Screen display and notice
Check that the desired rotor is
1
displayed in the ROTOR indicator of the Run screen. Go to step 3 when the desired rotor is displayed. When the desired rotor is not displayed, pess the ROTOR indicator to show the ROTOR CATALOG screen.
ROTOR indicato
Press the column of the rotor
2
you want to see the RCF value. The column turns from white to blue. Check that the pressed column turns blue and press the Enter button.
The pressed column turns from white to blue.
2-43
Page 60
2-3-4 Displaying and seting RCF
Process Touch panel operation Screen display and notice
3
When RCFmax or RCFavg is displayed in the column, touch the    button to
SPEED
RCF
show SPEED in the SPEED column. Press the SPEED column and enter the SPEED value you want to convert into the RCF value with the keypad display.
(e.g.) Rotor: S150AT
SPEED: 100,000 rpm Press , and press the
0
1
Enter button to enter the
setting.
See Section 2-2-1 “Setting run conditions.”
SPEED column
0
4 Press the button to
SPEED
RCF
turn the SPEED column to the RCFmax column or the RCFavg column.
The RCF value
of the rotor displayed in the ROTOR indicator is displayed on the setting value display field of the RCF column.
This is the RCF value for the SPEED value entered in step 3.
RCFmax column: The maximum RCF value for the rotor set
speed is displayed.
RCFavg column: The average RCF value for the rotor set
speed is displayed.
2-44
Page 61
r
2-3-4 Displaying and seting RCF
(2) How to set an RCF value
Select the desired rotor and enter an RCF value, and the centrifuge will calculate, set, and display the speed. Shown below is the procedure.
Process Touch panel operation Screen display and notice
Check that the desired rotor is
1
displayed in the ROTOR indicator of the Run screen. Go to step 3 when the desired rotor is displayed. When the desired rotor is not displayed, press the ROTOR column to
ROTOR
indicato show the ROTOR CATALOG screen.
Press the column of the rotor
2
you want to see the RCF value. The column turns from white to blue. Check that the pressed column turns blue and press the Enter button.
The pressed column turns from white to blue.
3
When “SPEED” is displayed in the column, press the
SPEED
   button to show
RCF
“RCFmax” (“RCFavg”) in the RCFmax (RCFavg) column. Press the RCFmax (RCFavg) column and enter the SPEED value you want to convert into the SPEED value with the keypad display.
(e.g.) Rotor: S150AT
RCFmax: 850x1,000 xg
Press
0
, and press the
8
Enter button to enter the
setting.
See Section 2-2-1 “Setting run conditions”.
RCFmax column
5
2-45
Page 62
2-3-4 Displaying and seting RCF
Process Touch panel operation Screen display and notice
4
Press the button to
turn the RCFmax (RCFavg) column to the SPEED column.
SPEED
RCF
The SPEED value
of the rotor
displayed in
the ROTOR
indicator is
displayed on
the setting value
display field of
the SPEED
column.
This is the SPEED value for the RCF value entered in step 3.
NOTE  (1) To shift from the SPEED column to the RCF column, press the button.
SPEED
RCF
(2) This centrifuge is not capable of checking rotors. Select the correct rotor.
Especially when you wish to determine a speed based on an RCF value, selecting the wrong
rotor and thus setting an RCF value exceeding the capacity of your rotor will cause the system to
calculate and set a speed exceeding the maximum allowable speed. (However, this
ultracentrifuge has an overspeed detector, which prevents an overspeed run.)
(3) When you run this centrifuge at a speed determined based on an RCF value, there may occur a
slight error (of up to 2%) between the set RCF value and the actual value, because the speed
setting is set in increments of 1,000 rpm.
2-46
Page 63
2-3-5 User login
After logging in, individual users can control the operation history of the centrifuge.
(1) Procedure for user login
Process Touch panel operation Screen display and notice
Press the USER column on the
1
Run screen.
2-3-5 User Login
USER column
Press the column of the
2
desired user name on the User Management screen. Check that the pressed column turns blue and press the LOGIN button. (“Admin” was registered as a
user name for this centrifuge before shipment.)
Enter the desired PIN on the
3
keypad display and press the
Enter button.
PIN is the number (4 digits) which you entered on the PIN registration screen (see Section 2-4-6 (1)).
(“1111” is the PIN of “Admin”
registered as a user name for
this centrifuge before shipment. If necessary, change the PIN.)
The User
Management screen appears.
The pressed column turns from white to blue.
LOGIN button
The Login screen
appears.
PIN box
The masked PIN
is displayed.
Enter button
2-47
Page 64
2-3-5 User Login
Process Touch panel operation Screen display and notice
The user name is displayed on
4
the user column of the Run
Run screen
returns to view.
screen.
USER column
NOTE  (1) User login is required to start operation when the user lockout function has been enabled as
described in Section 2-4-6 (2) “User Lockout.” You cannot operate the centrifuge without logging
in. Perform registration according to Section 2-4-6 (1) “User Management” before attempting to
log in.
2-48
Page 65
y
(
2-4 Features of the MENU screen
2-4 Features of the MENU screen
The Menu screen appears by pressing the MENU screen tab on the Touchscreen. These features are
designed to allow you to use the CS150NX with additional handy options (see figure 2-4-1).
MENU tab
Fig. 2-4-1 MENU screen
The functions of each icon on the MENU screen are described in the table below.
Function Icon Description
The information about the run parameters of a previous normal
Run History
Rotor Catalog
Spin-down
operation
RTC
(real-time control)
operation
Customize
operation can be stored automaticall load it for a new run loading the information about run parameters”).
You can view the applicable rotor names and their specifications (see Section 2-4-2 “Rotor catalog”).
While you press the PULSE button on the Run screen and are holding down it, the centrifuge accelerates up to the set speed at the maximum acceleration rate. When you release this button, it starts deceleration. This feature is useful to remove the adhered samples on the interior walls of the tubes (see Section 2-4-3 “Spin-down operation”).
Run starts or completes at a required date and time (see Section 2-4-4 “RTC operation”).
You can select rotor stop signal etc. which enable you to easily run the centrifuge (see Section 2-4-5 “Customizing the settings”).
see Section 2-4-1 “Displaying the Run History and
in the centrifuge and you can
Manager
(Admin)
Press the desired icon. Then the corresponding item is displayed.
You can set the items which the administrator should set such as language (see Section 2-4-6 “Administrator (Admin) functions”).
2-49
Page 66
2-4-1 Displaying the Run History and loading the information about run parameters
2-4-1 Displaying the Run History and loading the information about run parameters
The information about the run parameters of a previous normal operation can be
stored automatically in the centrifuge and you can load it for a new run.
Process
1
2
Touch panel operation Screen display and notice
Press the Run History icon on the MENU screen.
If you want to load the information about the run parameters for a new run, press the column of the information which you want to load for a new run. The column turns from white to blue. Check that the pressed column turns blue and press the Load for a new run button.
The Run History screen
appears.
Select the column.
If you want to check the details of the selected information,
press the View Details button.
3 Press the Run screen tab to
return to the Run screen.
The Run History Details
screen appears.
The Run screen appears.
2-50
Page 67
2-4-2 Rotor Catalog
You can view the applicable rotor names and their specifications.
2-4-2 Rotor Catalog
Process
1
Touch panel operation Screen display and notice
Press the Rotor Catalog icon on the MENU screen.
< : Previous Page button
> : Next Page button
2 If you want to view the MENU
screen, press the x button or MENU screen tab.
If you want to view the Run screen, press the Run screen tab.
The Rotor Catalog screen
appears.
Number of tubes x ml
The MENU screen appears.
The Run screen appears.
2-51
Page 68
f
2-4-3 Spin-down operation
2-4-3 Spin-down operation
While you press this button and are holding down it, the centrifuge accelerates up to the set speed at the maximum acceleration rate. When you release this button, it starts deceleration. This feature is useful to remove the adhered samples on the interior walls o the tubes.
Process Touch panel operation Screen display and notice
1 Press the PULSE Operation
icon on the MENU screen.
The PULSE button
is displayed on the
Run screen.
2 While you press the
PULSE button and are holding down it, the rotor
When the rotor stops, the PULSE button disappears and the
START button and the STOP button appears.
To cancel the spin-down operation, press the tab of the MENU
rotates.
screen tab and press the PULSE Operation icon again.
NOTE  (1) While performing the spin-down operation, you cannot change run parameters until the rotor
stops.
2-52
Page 69
2-4-4 RTC (real-time control) operation
2-4-4 RTC (real-time control) operation
The CS150NX contains an internal clock, allowing you to run the machine at a
specified start or finish time for centrifugation. This feature for running the machine at a specified time is called the RTC (real-time control) feature. The feature saves you the trouble of calculating the delay time for “delayed-start operation”.
Explained below is how to perform an RTC operation, with an example.
Example:If you wish to install your rotor on the centrifuge under the run conditions listed below on the evening of January 10 and to take out the samples around 8:00 the next morning;
(1) Rotor: S150AT (2) RPM: 150,000 rpm (3) Separation time: 2 hours (4) Control temperature: 4 (5) Acceleration mode: 9 (6) Deceleration mode: 7
150,000
4:30
2h
January 10 January 11
6:00 8:04
8:00 am
Fig. 2-4-4 A typical RTC operation
In this example, you set the above run conditions (2) through (6), set the start time for RTC operation to 6:00, January 11 and start the centrifuge. (You can make an identical setting by setting the finish time to 8:00 instead of setting the start time to 6:00.)
2-53
Page 70
2-4-4 RTC (real-time control) operation
t
(1) How to perform RTC an operation
Process
1
Touch panel operation Screen display and notice
Enter the run parameters on the Run screen.
Press the RTC button.
To specify the run time (centrifugation time), do not set
If you want to set a run to start
HOLD. Enter the desired numeric value. on a particular date and time, press the Start Time button.
If you want to set a run to stop
Start time button
Stop time button
on a particular date and time, press the Stop Time button.
Enter the start time or stop time on the keypad display.
RTC  button
The Set RTC screen
appears.
You can enter the star
time or stop time on the keypad display.
If you want to change any of the Month, Day, Hour, or Minuites
setting, press the column of the desired item and enter the numeric value.
To set the hours, use a number between 0 and 23 (in the 24-
hours system).
Enter a time later than the current time. When setting the “stop
time”, allow for centrifugation time and set the centrifugation start time to a time later than the current time.
You cannot set the centrifugation start time to a date more than
20 days in the future.
2-54
Page 71
s
2-4-4 RTC (real-time control) operation
t
Proces
Touch panel operation Screen display and notice
2 Press the Enter button.
Press the OK button and
3
the START button.
The RTC run does not ge
activated unless you press
the START button.
The START
confirmation box
appears.
The START time and STOP time for centrifugation are
displayed on the RTC field by pressing the START button. This centrifuge will turn the rotor automatically at the specified time.
Pressing the START button enters the system into a
condition causing the centrifuge to wait until the set time comes. This centrifuge will turn the rotor automatically at the specified time and run it for the specified time.
Once the RTC setting is made, you cannot change the
run time (centrifugation time). If you wish to make a change to the run time, cancel RTC first. If you want to cancel the RTC, press the RTC button on the Run screen and press the Cancel button on the SET RTC screen.
If you cancel RTC when the user lockout function (see
Section 2-4-6 (2) on this function) is enabled, the following screen appears. Inputting the “login user PIN” or the “administrator PIN” is required in this screen.
2-55
Page 72
2-4-4 RTC (real-time control) operation
NOTE
1. You cannot make an RTC setting in any of the following cases: (1) When the Run screen is set to HOLD (continuous run)
Set the run time (centrifugation time) not to HOLD but to a numerical value.
(2) When it is past the start time
Set the start time to a time later than the current time.
(3) When the start time is more than 20 days after the current time
Set the time to a time no more than 20 days afterwards.
2. To change the run time (centrifugation time) after making an RTC setting, cancel RTC and then set a new
run time.
3.To perform a combination of a programmed operation (including a step-mode operation) with an RTC run,
call a program memory unit, then set RTC. The system calculates the total run times of all steps of the programmed operation and calculates the start time for RTC. Therefore, cannot call the program memory unit after setting RTC.
4. To stop this ultracentrifuge in RTC operation, press the STOP button. The system then stops RTC and
stops the rotor.
2-56
Page 73
2-4-5 Customizing the settings
f
2-4-5 Customizing the settings
You can customize the settings on the Run screen and the settings on the CUSTOM screen such as the stop signal, sound volume and backlight.
Press the Customize icon on the MENU screen. Then the CUSTOM screen showing the five functions is displayed as shown in Fig. 2-4-5.
Tab of each screen
Fig. 2-4-5 CUSTOM screen
The functions of each icon on the CUSTOM screen are described in the table below.
Function Icon Description
Zoom
Stop signal
Volume
Brightness
Vacuum Standby
Setting
Press the desired icon. Then the corresponding item is displayed. After setting, press the tab o the desired screen.
The display on the Run screen can be zoomed in (see Section 2-4-5 (1)).
The rotor stop signal can be selected from six kinds of sound including five tunes and electric beep (see Section 2-4-5 (2) ).
The volume of the stop signal can be adjusted (see Section 2-4-5 (3)).
The brightness of the screen can be adjusted (see Section 2­4-5 (4)).
The conditions of the waiting for the vacuum of the rotor acceleration state can be specified (see Section 2-4-5 (5)).
2-57
Page 74
x
c
2-4-5 Customizing the settings
(1) Zoom
The display on the Run screen can be zoomed in.
1. Normal:Displays the ordinary Run screen
2. Zoom:The speed and time display is zoomed when 20 seconds have passed after reaching the set speed.
Fig. 2-4-5 (1) Zoom Setting screen
Press either the NORMAL or ZOOM button and make sure that the selected button is surrounded with a green frame. Then press the x button or the CUSTOM tab to store the setting.
To return from the zoom screen to the normal screen during operation, press anywhere except the
STOP button. Then the screen returns to the normal screen. Press the Customize icon of the MENU
screen and select the NORMAL according to the above procedure.
(2) Stop signal
The rotor stop signal can be selected from si kinds of sound including five tunes and electri beep.
Press the column of the desired stop signal. Then the stop signal sounds and the column of the selected stop signal turns blue.
Press the button to store the setting.
Enter To set another function on the CUSTOM screen, press the x button or the CUSTOM tab.
(3) Volume Adjustment
The volume of the stop signal can be adjusted.
The sound volume is turned up as the green range in the volume setting indicator increases. The stop signal does not sound if there is only the black range. Adjust the sound volume by pressing the following buttons.
Fig. 2-4-5 (2) Stop signal Setting screen
< The volume is lowered by pressing this button.    > The volume is turned up by pressing this button.
The stop signal does not sound by pressing this button.
    The stop signal sounds at full volume by pressing this button.
Press the x button or the CUSTOM screen tab to store the setting.
2-58
Fig. 2-4-5 (3) Volume Setting screen
Page 75
(4) Backlight setting
The backlight brightens up as the green range in the backlight setting indicator increases. The backlight level is darkest if there is only the black range. Press the following buttons to adjust the brightness of the screen. Press the x button or the CUSTOM screen tab to store the setting.
The brightness is decreased by pressing
<
>
The brightness of the screen can be adjusted.
this button.
The brightness is increased by pressing
this button.
2-4-5 Customizing the settings
Fig. 2-4-5 (4) Backlight Setting screen
(5) Vacuum standby setting
The conditions of the waiting for the vacuum of the rotor acceleration state can be specified.
1. Normal: When the vacuum level returns from medium to low during acceleration at 5,000 rpm or higher speed, the centrifuge keeps on running at that speed until the vacuum level becomes medium again.
2. No Waiting Acceleration: The centrifuge keeps on accelerating when it is
accelerating at 5, 000 rpm or higher speed even after the vacuum level returns from medium to low.
Fig. 2-4-5 (5) Vacuum Standby Setting screen
Press either the Normal or No Waiting Acceleration button and make sure that the selected button is surrounded with a green frame. Then press the x button or the CUSTOM screen tab to store the setting.
NOTE (1) Select Normal mode when using the sample that is sensitive to a temperature rise. If you select
No Waiting Acceleration , the rotor does not wait until the vacuum reaches an intermediate level and it might cause the rise in the rotor retention temperature by the friction heat.
(2) If the rotor chamber has frost on it by the repeated operation, it takes a long time to reach an
intermediate vacuum. In that case, wipe it off with a cloth, a sponge, etc.
(3) Once you select the condition of the waiting for the vacuum ( Normal or
No Waiting Acceleration ), your centrifuge memorizes this condition of the waiting for the
vacuum. If necessary, change the condition of the waiting for the vacuum.
(4)If the low vacuum condition persists for more than one minute in the No Waiting Acceleration
mode while the rotor rotates at the set speed, the alarm message “VACUUM ERROR” is displayed and the rotor is stopped.
2-59
Page 76
2-4-6 Administrator (Admin) function
2-4-6 Administrator (Admin) functions
You can set the admistrative items such as language mainly.
Press the Manager(Admin) icon on the MENU screen. Then the ADMIN screen showing the eight functions is displayed as shown in Fig. 2-4-6.
Tab of each screen
Fig. 2-4-6 ADMIN screen
2-60
Page 77
2-4-6 Administrator (Admin) function
The functions of each icon on the ADMIN screen are described in the table below.
Function Icon Description
User Management
User Lockout
Rotor Lockout
Japanese
Rotor Management
The user name can be registered and deleted (see Section 2-4-6 (1)).
User login can be required to start operation (see Section 2-4-6 (2)).
Run history of each rotor can be controlled (see Section 2-4-6 (3)).
English and Japanese can be toggled in display (see Section 2-4-6 (4)).
You can use and control the rotor data such as total operation hours and number of runs (see Section 2-4-6 (5)).
Actual Run Timer
Date and time display
Centrifuge ID
Service contact
Actual run timer can be selected (see Section 2-4-6 (6)).
The date and the time can be set (see Section 2-4-6 (7)).
Centrifuge ID can be set for identification (see Section 2-4-6 (8)).
The operation history data of the centrifuge can be exported to a
USB/LAN
communication
USB flash drive. If your centrifuge has a LAN communication option, LAN communication start/stoppage can be specified.
(see Section 2-4-6 (9)).
Press the desired icon. Then the corresponding item is displayed. After setting, press the tab of the desired screen.
2-61
Page 78
s
2-4-6 Administrator (Admin) function
(1) User Management
You can register up to 40 users with the system. How to store (register) and how to change a user name are described below.
Proces
1
Touch panel operation Screen display and notice
Press the User Management icon. Then press the Add User button.
(“Admin” whose PIN is “1111”
has been registered in this centrifuge before
shipment. If necessary,
change the PIN.)
Add User button
User name box
The User
Management
appears.
x button
A screen to
register a user
name is
displayed.
Clear button
Shift button
Space button
Delete button
Enter button
Press the Shift button to toggle case.
Press the Clear button to delete all the characters in
the user name box.
Press the Space button to put one space after the last
character in the user name box.
Press the Delete button to delete the last character in
the user name box.
Press the Enter button to turn the screen to the PIN
registration screen.
Press the x button to return to the user administration
screen.
2-62
Page 79
s
2-4-6 Administrator (Admin) function
Proces
2
Touch panel operation Screen display and notice
Register the user name and touch the Enter button.
Enter button
PIN box
The PIN registration
screen is displayed.
3
Enter the desired PIN number (4 digits) in the PIN box and touch the Enter button.
The masked PIN
number is displayed.
The PIN confirmation
screen is displayed.
2-63
Page 80
s
A
2-4-6 Administrator (Admin) function
Proces
4
Touch panel operation Screen display and notice
Enter the same PIN number (4 digits) as the one entered in the
PIN (confirm) box
above step 3 in the PIN (confirm) box. Press the Enter button.
Difference of access level Admin: The administrator can
do every operation of this centrifuge.
Enter button
User: The user cannot use the
Admin functions when the user lockout function is enabled.
The masked PIN
number is displayed.
The access level
selection box is
displayed.
ccess level
selection box
To place the registered user on
5
the same access level with the administrator, press Administrator in the access level selection box. If you do not want to place the registered user on the same access level with the administrator, press User ..
(The screen display shows the
case when placing the registered user on the same access level with the administrator.)
For the difference of access level, refer to Section 2-4-6 (2).
The screen returns
to the User Management screen. ”Hitachi-Koki” is added to the user name and “Administrator” is displayed as the
Access level button
PIN button
To change the access level or the PIN number after
access level.
registration of the user name, press the column of the desired user name and check that the column turns blue. Press the Access level button or the PIN button to make a change.
To delete the user name after registration, press the
column of the desired user name and check that the column turns blue. Press the Delete User button to delete it.
2-64
Page 81
(2)User Lockout
r
Enable : User login is required to start operation.
You can not set run conditions and start
operation without logging in. Prior user
registration is required.
(See Section 2-4-6 (1) “User control” and 2-3-5 “User login.”)
In addition, login of a user having the administrator access level
2-4-6 Administrator (Admin) function
is required to call up the Admin screen.)
Fig. 2-4-6 (1) User Lockout Setting screen
Disable : User login is not required to start operation.
Press either the Enable or Disable button and make sure that the selected button is surrounded with a red frame. Then press the x button or the ADMIN screen tab to store the setting.
3 Rotor Lockout
Enable : Select a rotor from the registered rotors
to start operation. It is impossible to
select from the rotor catalog. (See
Section 2-4-6 (5) “Rotor Management.”)
Disable : The centrifuge is operable without selecting the
rotor. It is possible to select a rotor from the roto
catalog.
Fig. 2-4-6 (2) Rotor Lockout Setting screen
Press either the Enable or Disable button and make sure that the selected button is surrounded with a red frame.
Then press the x button or the ADMIN screen tab to store the setting.
4Japanese
You can select whether to give displays in English or Japanese by pressing this button.
2-65
Page 82
2-4-6 Administrator (Admin) function
f
5 Rotor Management
You can use and control the rotor data such as total operation hours and number o runs by registering your rotors in the centrifuge in advance. (It is recommended to use the rotor lockout function (see 2-4-6 (3)) together in order to control the rotor data such as total operation hours and number of runs.)
(1) Registering the rotor
Process Touch panel operation Screen display and notice
1 Procedures for registering the
rotor are described as follows.
a Press the
Rotor Management icon on the ADMIN screen.
bPress the Add rotor
button.
a The Rotor
Management screen appears.
bThe ROTOR
CATALOG
screen appears.
c) Press the column of the
desired rotor name and
press the Enter
button.
d) Enter the S/N (Serial
number ) and press the
Enter button.
c The Enter S/N
screen appears.
Enter a 4-digit
number: In the
case of “123,”
prefix zero to the
number (“0123”).
d) Registration is
completed if it is
a new rotor
having no
operation history.
2-66
Page 83
2-4-6 Administrator (Admin) function
Process Touch panel operation Screen display and notice
2 Procedures for entering the
total operation hours of the rotor are described as follows.
a Press the column of the
desired rotor on the Rotor Management screen. Then check that the pressed column turns blue
and press the Edit Hours button.
b Enter the total operation
hours of the rotor and press the Enter button.
The total operation
hours of a rotor
having operation
history can be
registered. This
operation is not
required for a new
rotor.
2-67
Page 84
2-4-6 Administrator (Admin) function
Process Touch panel operation Screen display and notice
3 Procedures for entering the
total number of runs of the rotor are described as follows.
a Press the column of the
desired rotor on the Rotor
Management screen. Then check that the pressed column turns blue
and press the Edit Runs button.
b Enter the total number of
runs of the rotor and press the Enter button.
The total number
of runs of a rotor
having operation
history can be
registered. This
operation is not
required for a new
rotor.
4 Press the Run screen tab. The Run screen appears.
2-68
Page 85
r
2-4-6 Administrator (Admin) function
(2) Deleting a registered rotor
Process Touch panel operation Screen display and notice
1 Press the Rotor Management
icon on the ADMIN screen.
2 Press the column of the
rotor which you want to delete on the Rotor Management screen. Then check that the pressed column turns blue and press
the Delete rotor button.
The Rotor
Management
screen appears.
The deletion
confirmation box
is displayed.
3 Press the Yes button on
the deletion confirmation box.
The registerd roto
is deleted.
2-69
Page 86
t
r
2-4-6 Administrator (Admin) function
6 Actual Run Timer setting
<Actual Run Timer is enabled.>
When you press the Enable button,
the timer starts counting after reaching the
set speed. <Actual Run Timer is not enabled.>
When you press the Disable button,
Fig. 2-4-6 (3) Actual Run Timer Setting screen
the timer starts counting immediately afte
the start of operation.
Press either the Enable or Disable button and make sure that the selected button is surrounded with a red frame.
Then press the x button or the ADMIN screen tab to store the setting.
7 Date and time
Use this function to precisely set the curren time setting of the internal clock. Set an exact date and time for RTC operation.
Fig. 2-4-6 (4) Setting at time screen
Press the column of the desired item (Year, Month, Day, Hour, or Minute). Check that the pressed
column turns to red and enter the date and time on the keypad display.
If you want to change the other items, press the column of the other items.
You can also turn the desired column to red by pressing the > button and < button. If you do not want to change the other items, press the Enter button on the keypad display.
Example: Year:2008 Month:December Day:11th Time:8:58am
Fig. 2-4-6 (5) Setting at time screen
If you want to change the setting of the other items on the ADMIN screen, press the ADMIN screen tab.
2-70
Page 87
(8) ID /Contact
Press the Machine ID column.
2-4-6 Administrator (Admin) function
When you have two or more centrifuges, you can set the centrifuge ID for each centrifuge for identification.
Service contact:
Set the centrifuge ID (4-digit number).
Make a memo of your service person’s contact address.
Fig. 2-4-6 (6) Centrifuge ID and service contact setting screen
Press the Service contact column.
Fig. 2-4-6 (8) Enter Service contact screen Fig. 2-4-6 (7) Enter Centrifuge ID screen
9 USB/LAN communication
[USB MEMORY]
The operation history of the centrifuge can be output in CSV format* to a USB flash drive on the market. *CSV format: This is a versatile file format that is compatible with spreadsheet softwares such as EXCEL.
USB (host side)
Insert the USB flash drive on the market into the USB for external connection (host side) at the right side of the
Press the DATA EXPORT button. (Writing in the USB flash drive
is automatically started.)
centrifuge.
Fig. 2-4-6 (9) USB/LAN communication setting screen
2-71
Page 88
2-4-6 Administrator (Admin) function
NOTE (1) Be After writing in the USB flash drive is finished, the following screen appears.
If you press the YES button, the run history data stored in the centrifuge is deleted. If you press the NO button, the run history data stored in the centrifuge is not deleted. However, if the information about 100 sets of the run parameters of the previous normal operations has been stored in the centrifuge, the oldest information of the run parameters is deleted when you perform the centrifuge run.
(2) Be sure to back up your data before using the USB flash drive. Hitachi Koki will not be liable for
any data loss.
(3) This centrifuge cannot use the USB flash drive with security function. (4) When powered on with the USB flash drive, the centrifuge may not recognize the USB flash
drive. In such a case, pull out the USB flash drive from the centrifuge and then insert it again into the centrifuge.
[LAN Communication]
If your centrifuge has a LAN communication option, LAN communication start/stoppage can be specified. Insert the LAN cable into the LAN port at the right side of the centrifuge.
The following message is displayed. “Transmission was completed. Do
you want to delete run history data stored in centrifuge?”
LAN port(option)
NOTE
Connect :
Disconnect : LAN communication stops.
Even if your centrifuge does not have a LAN communication option, the Connect button is surrounded with a red frame as you press the Connect button. In that case, this icon is always displayed on the Run screen, but there is no problem.
LAN communication starts. Press this button to establish the communication between “himac LogManager” (option) on your computer and your centrifuge. It is possible to manage operation results of this centrifuge by using “himac LogManager” (option). If you press this button, either of the below icons is displayed on the upper side of the VACUUM button on the Run screen.
There is no problem in communication between “himac LogManager” (option) on
your computer and your centrifuge.
There might be problems in communication between “himac LogManager” (option)
on your computer and your centrifuge. Follow the below instructions.
Check whether the LAN cable is connected to the centrifuge properly. Check whether the “himac LogManager” on your computer is started. Check whether the LAN works properly.
2-72
Page 89
2-5 Emergency recovery from power failure
2-5 Emergency recovery from power failure
DANGER: Before removing the front cover for troubleshooting, always turn off the POWER switch of
the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
WARNING:
CAUTION:
1. Rotation of the rotor
The rotor stops as a result of free coast. However, if the rotor did not decelerate to 300 rpm or lower when
a power failure occurred, it will, or lower when a power failure occurred, after power returns, automatically
accelerate again and return to the set speed before the power failure. If the speed goes below 300 rpm, the
rotor will decelerate and stop after power returns.
If the rotor is stopping when the power returns, this centrifuge does not accept the VACUUM button etc. (see Section 4-1 “Alarm indicators POWER FAILURE : ***** rpm”).
2. Touchscreen display
All displays are off during a power failure. After power returns, the instrument starts control again at the
settings specified before the power failure. (The settings have battery-backed back-ups.)
The instrument then displays an alarm message, indicating that it has suffered a power failure.
3. Removing the rotor from the ultracentrifuge
If a power failure may last a long time, open the door and take out the rotor as described below.
(1) Never open the door during rotation.
(2) Never touch the rotor during rotation.
Never conduct operations in a manner other than as described in this operation manual.
Contact with your nearest service representative if any problem arises.
WARNING:
The rotor chamber is in a vacuum with low air resistance. The rotor may therefore go on turning for
approximately 120 minutes before stopping. Allow enough time before opening the door.
Before following this procedure, listen carefully for sound from within the rotor chamber
and make sure that the rotor is not turning.
While the rotor is rotating, never unlock the chamber door.
2-73
Page 90
2-5 Emergency recovery from power failure
(1) Make sure that the rotor has stopped.
(2) Unplug the power cord of the centrifuge from the wall outlet. (3) Remove the four screws that fasten the both sides of the front cover by Hex key, remove the front
cover while pulling it towards you, and remove the cover.
Top deck
Vacuum
Front cover
release knob
Front cover fastening screw
Front cover hook
Fig. 2-5-1. Removing the front cover
(4) Remove the vacuum release knob at the left of the vacuum chamber (by turning its tip
counterclockwise) to allow air to enter the rotor chamber. When the rotor chamber reaches
atmospheric pressure, do not forget to return the vacuum release knob to its previous position.
2-74
Page 91
2-5 Emergency recovery from power failure
r
(5) Pull the release wire of the door lock in the left side of the vacuum chamber toward you, and open the door.
After opening the door, make sure that the rotor is not rotating. If it is rotating, close the door immediately.
WARNING: Never touch the rotor while it is rotating.
Vacuum chambe
Vacuum release knob
Release wire of the door lock
Square hole of the frame
Fig. 2-5-2 Door lock and vacuum release knob
(6) After taking out the rotor, insert the hooks (which are on both sides of the front cover) into the frame, and
insert the front cover into the gap between the top deck and the front cover supporting plate while holding
up the front cover a little. Then fasten them with the fastening screws.
Top deck
Front cover
Front cover supporting plate
Hook
Fig. 2-5-3 Inserting the front cover
(7) Plug the power cord into the wall outlet.
(8) After the power returns, turn on the power switch of the centrifuge. Press the VACUUM button to turn
on the vacuum pump for approximately 15 minutes, and then press the VACUUM button to turn off the
vacuum pump. Then turn off the power switch on the centrifuge. (These procedures are effective ways of
keeping the vacuum pump in good condition.)
2-75
Page 92
3. Maintenance
3. Maintenance
Be sure to read and keep in mind the following cautionary information before maintenance.
DANGER: Before removing the cover, top deck, or other component for maintenance, always turn
off the POWER switch of the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
WARNING: (1) If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic
or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
(2) If there is a possibility that the centrifuge, rotor, or an accessory is contaminated
by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor, or the accessory properly before requesting repairs from a Hitachi Koki authorized sales/service representative. Note that Hitachi Koki cannot repair the centrifuge, rotor, or the accessory unless sterilization or decontamination is completed
.
(3) It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or
parts properly before returning them to a Hitachi Koki authorized sales/service representative. In such cases, copy the decontamination sheet at the end of this manual and fill out the copied sheet, then attach it to the item to be returned. Hitachi Koki may ask you about the treatment for the centrifuge, rotor or the part if the decontamination is checked and judged as insufficient by Hitachi Koki. It is your responsibility to bear the cost of sterilization or decontamination.
If you have any question, please send e-mail tohimac@hitachi-koki.co.jp.
Note that Hitachi Koki cannot repair or inspect the centrifuge, the rotor or the accessory unless sterilization or decontamination is completed.
CAUTION: Do not perform any operation not specified in this manual. If any problem is found on
your centrifuge, contact a Hitachi Koki authorized sales/service representative.
3-1
Page 93
3-1 Rotor chamber 3-2 Drive shaft (Crown) 3-3 Cabinet
This centrifuge does not require complicated maintenance and inspection. For longer and safe use of this centrifuge without trouble, observe the following instructions.
CAUTION: Using a cleaning or sterilization method other than the ones recommended in this
instruction manual might cause corrosion or deterioration of the centrifuge. Refer to the chemical resistance chart provided with the rotor, or contact Hitachi Koki.
CAUTION: Disconnect the power cord from the outlet before cleaning or sterilizing the centrifuge.
For information on the maintenance of rotors and tubes, see the rotor instruction manual provided with the rotor.
3-1 Rotor chamber
CAUTION: Do not pour any solution such as water, detergent, or disinfectant directly into the
rotor chamber and be careful not to spill the sample into the rotor chamber. If you do so, the bearings of the drive unit might corrode or deteriorate and it might cause vacuum failure.
To maintain the rotor chamber, follow the instructions given below: (1) When the ultracentrifuge is not in use, keep the rotor chamber ventilated. (2) If the bowl is moist, wipe it with a clean, dry cloth or sponge. (3) If the rotor chamber is dirty, wipe it with a clean cloth wrung out in a diluted solution of a mild, non-
alkaline detergent. For sterilization, wipe it with a cloth etc. dampened with 70% ethanol.
3-2 Drive shaft (Crown)
CAUTION: Clean the inside of the drive hole (crown hole) of the rotor and the surface of the drive
shaft (crown) of the centrifuge once a month. If the drive hole or the drive shaft is stained or any foreign matter is adhered, the rotor may be improperly installed and come off during operation
This part is very important because the rotor is mounted on it and the crown transmits driving force to the rotor. Before mounting a rotor, wipe the outer surface of the crown with a soft cloth dampened with water sufficiently.
3-3 Cabinet
Always keep the top deck and the cabinet of the centrifuge clean to prevent dust and other materials from falling into the rotor chamber. Wipe the top deck and the cabinet with a cloth or sponge dampened with a diluted solution of neutral detergent. If any solution that is toxic, radioactive, or pathogenic is spilled inside or outside the centrifuge, take necessary action according to your proper laboratory procedures and methods.
3-2
Page 94
3-4 Chamber door seal
3-4 Chamber door seal
If the door seal O-ring is dusty or scratched, high vacuum level will not be obtainable. Always keep the door seal O-ring clean. When the ultracentrifuge is used frequently, take out the door seal O-ring and wipe it with a clean, soft cloth and then put a thin coat of vacuum grease on it every three to four months (ordinarily, once a year). If the door seal O-ring is damaged, replace it. Wipe the groove for the door seal O-ring with a clean, soft cloth dampened with alcohol or a similar solvent.
3-4-1 Removing the door seal O-ring
1. While the door is open, turn off the POWER switch and unplug the power cord from the wall outlet.
2. Open the door completely. Then hold the handle of the door and pull up the door.
Handle
Door seal O-ring
Fig. 3-4-1Removing the door seal O-ring
3. Use a toothpick etc. to take the door packing out of the groove. At that time, be careful not to damage the door seal O-ring or groove.
4. Clean the door seal O-ring and groove. If the door seal O-ring is damaged, replace it. {Part code of replacement door seal O-ring (not furnished with the accessories): 84520135}
5. Apply a thin coat of vacuum grease to the door seal O-ring, insert it into the groove, and close the
door.
3-3
Page 95
3-5 Vacuum pump 3-6 Others
3-5 Vacuum pump
If displaying the VACUUM alarm, poor evacuation, or other trouble occurs frequently, the vacuum pump oil may deteriorate or the oil mist trap may be clogged or the vacuum pump may have a problem. The vacuum pump oil should be changed once a year to prevent deterioration of the vacuum pump and clogging of the oil mist trap. Note that the oil change frequency may differ depending on the use conditions and the environment. Call a service representative when an oil change is required or any trouble occurs.
Oil for the vacuum pump is furnished with this centrifuge. Store it in a safe place and give it to the service representative if requested.
3-6 Others
(1) Storage period of service parts
Service parts are kept in stock ten years after the discontinuation of production. The term “service parts” means the parts that are necessary to ensure the correct functioning of the centrifuge.
3-4
Page 96
4. Troubleshooting
4. Troubleshooting
Be sure to read and keep in mind the following cautionary information before troubleshooting.
DANGER: Before removing the cover, top deck, or other component for troubleshooting, always
turn off the POWER switch of the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
WARNING: (1) If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic
or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
(2) If there is a possibility that the centrifuge, rotor, or an accessory is contaminated
by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor, or the accessory properly before requesting repairs from a Hitachi Koki authorized sales/service representative. Note that Hitachi Koki cannot repair the centrifuge, rotor, or the accessory unless sterilization or decontamination is completed
.
(3) It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or
parts properly before returning them to a Hitachi Koki authorized sales/service representative. In such cases, copy the decontamination sheet at the end of this manual and fill out the copied sheet, then attach it to the item to be returned. Hitachi Koki may ask you about the treatment for the centrifuge, rotor or the part if the decontamination is checked and judged as insufficient by Hitachi Koki. It is your responsibility to bear the cost of sterilization or decontamination.
If you have any question, please send e-mail tohimac@hitachi-koki.co.jp.
Note that Hitachi Koki cannot repair or inspect the centrifuge, the rotor or the accessory unless sterilization or decontamination is completed.
CAUTION: Do not perform any operation not specified in this manual. If any problem is found on
your centrifuge, contact a Hitachi Koki authorized sales/service representative.
Before removing the cover, top deck, or other component for troubleshooting, always turn off the POWER switch of the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
4-1
Page 97
4-1 Alarm indicators
This ultracentrifuge incorporates a “self-diagnosis feature” that diagnoses the cause of any problem which may occur when you start the centrifuge or while in operation.
4-1 Alarm indicators
If any trouble occurs, this machine gives a buzzer sound and displays an alarm message in the message field of the Run screen to warn of the trouble.
(1) Responding to an alarm signal
If an alarm message appears, remove the cause of the trouble as described below and press the CE button. You will then be able to resume your centrifugation.
WARNING:
Unspecified repairs, remodeling or disassembly of the centrifuge that is not listed
below is strictly prohibited by any person other than a Hitachi Koki authorized services representative.
If the alarm message persists even after you have done what is specified below, contact service representative to order a repair.
Alarm Cause Action
CLOSE DOOR
Close the door completely.
Press the VACUUM or START button.
VACUUM ERROR
Wipe off moisture inside the rotor chamber, or check a sample leakage.
1. The VACUUM or START button has been pressed with the chamber door left open.
1. Required level of vacuum cannot be reached.
2. After a satisfactorily high level of
vacuum was reached, it lowered (due to, for instance, sample leakage).
Close the door completely and press the
VACUUM or START button.
Wipe off the moisture from inside the rotor
chamber.
Clean the door packing, then apply a thin coat
of vacuum grease. (Refer to Section 3-4 “Chamber door seal.”)
Check if the sample is leaking from the rotor
and/or tubes. If so, the tubes may be overfilled or may be cracked or broken. In these cases, reduce the amount of sample in the tubes or change the tubes.
SET ROTOR
Set the rotor. Wait up to 20 min.
If the alarm is not cleared.
IMBALANCE
Check balance of samples.
Tighten rotor cover or cap securely.
1. The rotor is not installed.
1. Rotor is not properly balanced, and abnormal vibration has occurred in the rotor.
2. Rotor cover or cap is not properly tightened.
Install the rotor.
If a “SET ROTOR” alarm message appears,
this ultracentrifuge does not accept the CE or VACUUM button for 20 minutes to ensure safety. Wait at least 20 minutes, then press the CE button.
If you want to clear this alarm when 20 minutes have not passed since this alarm message appeared, see the ACTION of the alarm message “POWER FAILURE:***** rpm”.
Check if the sample tubes exceed allowable
imbalance level.
Check if any one of the tubes is deformed, and
if there is any sign of sample leakage.
Tighten rotor cover or cap securely.
INVALID SPEED
Check max. speed of the rotor.
Set its max. or lower speed again.
POWER FAILURE
Running re-started automatically or is suspended. Check status.
1. Rotor speed is set higher than the maximum allowable speed.
1. A power failure occurred while the rotor was rotating.
Refer also to Section 2-5 “Emergency recovery from power failure.”
Set the speed within their permitted limits.
Unless the set run time has elapsed, restart
the run.
If the instrument was automatically restored
and the rotor is rotating at set speed, then let the run continue.
4-2
Page 98
4-2 User-corrected Problem
Alarm Cause Action
POWER FAILURE:***** rpm
Door is closed. Refer to manual for further detail.
E11-E64
1. A power failure occurred while the rotor was rotating and the speed could not be detected after power returned.
(This alarm code appears when the
detector of this ultracentrifuge could not identify which status (status 1 or status occurred.
Status 1: The rotor stopped normally. Status
2: Although the rotor continued to
rotate, the speed of the rotor was not detected due to a
If this alarm code appears, this ultracentrifuge does not accept the CE or VACUUM button for 120 minutes to ensure safety (The same thing can be said of “E13”).
Refer also to Section 2-5 “Emergency recovery from power failure.”
Refer to NOTE.
failure of the speed detector.
This ultracentrifuge does not accept the CE or
VACUUM button for 120 minutes from power failure. If you are convinced that the rotor has come to a complete stop, press the SPEED column,
2)
HOLD , 9 , 0 , 1 , HOLD and Enter button in order when the Run screen is displayed. Then press the VACUUM button. This ultracentrifuge ac cepts the VACUUM button. First open the door a little to confirm that rotor has come to a complete stop. Then open the door properly.
Call a service representative.
If any alarm message between E11 and E64 appears, this ultracentrifuge requires maintenance by a service representative. When ordering a repair, inform us of the alarm code you have received.
E13 is an alarm code that indicates a failure of the speed detector.
NOTE
If this alarm code appears, this machine does not accept the CE or VACUUM button for 120 minutes to ensure safety. These 120 minutes are required for the rotor to stop. Wait at least 120 minutes before pressing the CE button. (This “120 minutes” is the time which passed after a power failure occurs.)
4-3
Page 99
4-2 User-corrected Problem
4-2 User-corrected Problem
Some problems are not identified and reported by the self-diagnostic capability of the ultracentrifuge. To correct these problems, take the action(s) described in the table below.
Symptom Possible cause Action
Ultracentrifuge will not start up when its power is on.
Rotor does not cool,
or rotor temperature is rising.
You press the START button, but the rotor will not turn.
The backlight level of the screen is too dark or too bright.
You can not open the door because the STOP button continues to blink.
1. Building power circuit breaker is open.
2.The power cord is unplugged from the outlet.
1. Rotor was accelerated at low or
intermediate vacuum level.
2. High vacuum level cannot be reached.
1. If the “SET ROTOR” alarm message appears, the overspeed detector may issue a signal that prevents the rotor from rotating even though you pressed the CE button. (This is not a fault.)
1. The brightness of the screen
adjusted.
1. You pressed the START button, and
then you pressed the STOP button
before the rotor started to rotate.
is not
Close the circuit breaker.
Plug the cord into the outlet.
Start accelerating the rotor when the rotor
chamber is at high vacuum level (after the vacuum indicator is displayed as the following).
In a high vacuum
Inspect the door seal O-ring (see Section
3-4).
Turn off the POWER switch, wait several
minutes, turn on the POWER switch again, and press the START button.
If this procedure still does not activate this ultracentrifuge, call a service representative.
Make the adjustment while referring to the
clause “Backlight setting” in Section 2-4-5 “Customizing the settings”
Because the detector of this ultracentrifuge
could not identify which status (status 1 or status 2) occurred, the safety devices function. The centrifuge maintains deceleration status at that time. Status 1: The rotor stops normally. Status 2: The speed detector is faulty. Open the door of the centrifuge after the safety devices are released (after from 10 seconds to approximately four minutes).
4-4
Page 100
5. Installation
5. Installation
This chapter describes the electrical power requirements, location and environment that you must provide for your ultracentrifuge prior to its installation by an authorized Hitachi Koki representative.
DANGER: Before removing the cover or other component, always turn off the POWER switch of
the instrument, unplug the power cord from the wall outlet, and wait at least three minutes to avoid the risk of electrical shock.
Note: The installation and leveling of your ultracentrifuge must be done by an authorized Hitachi Koki
representative. If they are done by anyone else, the ultracentrifuge warranty will be null and void.
1.Place of installation
(1) Position the centrifuge on a level surface, such a table or laboratory bench that can support the
weight of the centrifuge and resist vibration. (Place the centrifuge at least 5 cm away from the edge of the laboratory bench.)
(2) Ambient temperature for operation is 5 to 35°C. If the room temperature rises above 35°C, the
temperature of the rotor may become too high. Avoid installing the ultracentrifuge in direct sunlight.
(3) Keep the back of the instrument at least 8 cm away from the wall. We recommend you install the
sides of the instrument about 13cm away from the walls. Make sure that the air can circulate adequately around this ultracentrifuge. Avoid installing this ultracentrifuge close to a heat-generating device, which might reduce this centrifuge’s cooling capacity.
WARNING: For operator safety, maintain a 30-cm “clearance envelope around the instrument
and keep out that area while the rotor is spinning. Do not store dangerous substances capable of developing flammable or explosive vapors on nor near the centrifuge.
8cm or more
13cm
or more
13cm
or more
Fig.5-1 Clearance for ultracentrifuge
5-1