1. | Features | 1 |
---|---|---|
2. | Accessories. | 2 |
3. | Precautions | 2 |
4. | Controls and connectors | 6 |
5. | How to produce the bright line | 18 |
6. | Method for connecting signals | 20 |
7. | Measuring procedure | 20 |
8. | Adjustments | 36 |
9. | Maintenance | 36 |
10. | Specifications | 36 |
11. | External view | 11 |
12. | Schematic diagrams | 41 |
44 |
The Hitachi V-522/V-523/V-525 is a portable-type, advanced-class oscilloscope with a bandwidth of DC to 50 MHz designed with the emphasis on operability and portability and has a following features.
The instrument has a bandwidth from DC to 50 MHz.
Sensitivity is 1 mV/div
Employment of a large square CRT makes waveforms easier to observe.
Employment of an internal graticule CRT permits waveforms observation to be made without parallax error.
Allows a proportionate observation of any part of an input waveform even when it is in a large amplitude.
Allows simultaneous observation of X1 and X10 sweeping waveforms.
Even an observation of two waveforms of different frequencies, the waveform of the each cannel is stably triggered.
Employment of a new TV sync separator circuit allows the instrument to observe TV signals stably.
Focusing shift is automatically corrected.
Enlarges any given portion of signal for easy observation.
Set values can be displayed on the CRT for photographing. a waveform with the data.
The voltage difference and the time difference are measured by the two electronic cursors and displayed on the CRT.
automatically converted and displayed on the CRT.
This instrument is shipped with the following standard accessories.
Precautions to be observed to lengthen the service life of this instrument.
The operating ambient temperature is 0°C or more.
The operating ambient humidity is 35–85%.
Since an accidental intrusion of water may also cause troubles, do not place a water-filled containers such as a vase on the oscilloscope.
do not use the instrument near an equipment generating strong magnetic force.
* Do not leave a hot soldering iron on the cabinet or the screen.
* Do not try to turn the instrument upside down. Otherwise, knobs may be broken.
* Do not use the instrument upright, leaving BNC cable connected to EXT BLANKING terminal on the rear panel. Otherwise, the cable may be damaged.
The handle of the V-522/V-523/V-525 can be positioned for carrying or as a tilt-stand for the instrument. To position the handle, press in at both pivot points and turn the handle to the desired position. 13 positions are provided for convenient carrying or viewing.
When not in use, put the dust-proof cover on the instrument and store it with care.
Recheck the operating procedure and if problem persists, contact a nearly service station or agent.
* Never use strongly volatile agent such as benzine and thinner.
When cleaning the inside, insure beforehand that no electricity remains in the condensers of the power supply circuit. "NOT FOR AN OPERATOR"
* Cleaning of CRT
Dirty surface of CRT screen tends to cause measuring errors. The screen surface becames visible when the bezel is removed.
Remove the stains on CRT and filter by using a clean and soft cloth, paying attention not to impair them.
When the stain is extremely heavy wash them with neutral washing agent and then leave them stand until the moisture is removed naturally
• If the screen is installed while it is moistened water rings may be formed and the waveform may be blurred to become hard to observe Pay attention not to leave finger prints on it.
* Check the line voltage.
The operating voltage range of this oscilloscope is as shown below. Check the line voltage without fail before turning on the power switch.
Rating | Line Voltage (50/60Hz) |
---|---|
AC100V | AC 90V - 110V |
AC120V | AC108V- 132V |
AC220V | AC198V- 242V |
AC240V | AC216V- 264V |
Nominal volts ±5% at 400Hz.
Prior to shipment the voltage selector is set properly. When the oscilloscope is intended to be used on a different voltage, relocate the voltage selector in the procedures which follow
4) Connect the power connector
In order to protect the circuit against overcurrent, a 1A (make use of AC100V or AC120V) or 0.5A (make use of AC220V or AC240V) is used on the primary side of the power supply. When this fuse is below out, check thoroughly the cause, repair any faulty point present, and then replace with a specified fuse. Do not try to use the fuse other than the specified ones. Otherwise, fault may be caused or danger may be invited.
(Particularly, do not use a fuse different from the specified one in current capacity and in length.) The standards if the fuses are as follows
Shape
(Diameter x length)mm |
JIS type name | |
---|---|---|
1A | 5.2φ x 20 | 250V-MQ4 1A |
0.5A | 5.2¢ x 20 | 250V-MQ4 0.5A |
Do not increase the brightness of the spot and trace too much. Your eyes may be strained and the fluorescent surface of CRT may be burnt.
The input withstand voltage of each input connector and probe input is as follows. Never apply a voltage higher than specified.
INPUT direct,300V (DC + AC peak at 1 kHz)When x10 probe is used400V (DC + AC peak at 1 kHz)When x 1 probe is used300V (DC + AC peak at 1 kHz)EXT TRIG INPUT300V (DC + AC peak)EXT BLANKING30V (DC + AC peak)
To maintain instrument accuracy, perform the calibration
of the V-522/V-523/V-525 at least every 1000 hours of operation, or every six months if used infrequently.
This model can be operated from either a 100-volt, a 120-volt, a 220-volt or a 240-volt nominal line voltage source.
The Line Voltage Selector assembly on the rear panel converts the instrument from one operating range to the other. In addition, this assembly changes the primary connections of the power transformer to allow selection of one of four regulating ranges. The assembly also includes the line fuse. Use the following procedure to convert this instrument between nominal line voltage or regulating ranges.
Fig. 4-1 V-525 Front View
Fig. 4-3 V-522 Front View
The POWER is set on at the pushed-in position, and set off at the released position.
This lamp goes on in red when the power supply is in ON state.
After obtaining an appropriate brightness by operating INTENsity, adjust FOCUS until the bright line is clearest. Although the focus is also corrected automatically when INTEN is rotated, the focus is sometimes slightly shifted.
Controls graticule illumination. Useful to illuminate the graticule when viewing in a dark area, photographing.
Used to aline the trace of CRT with the horizontal graticule.
This knob also works as the brightness adjust variable resistor. Brightness is increased by rotating INTENsity
clockwise.
This is inlet for detachable AC power cord.
BNC connector for vertical axis input. The signal input to this terminal becomes the X-axis signal when the instrument is used as an X-Y oscilloscope.
The same as CH1, but when the instrument is used as an X-Y oscilloscope, the signal input to this terminal becomes the Y-axis signal.
The switches used to select the coupling system between the input signal and vertical axis amplifier.
AC At this setting the signal is connected through a condenser. The DC component of the input signal is cut off and only the AC component is displayed.
A step attenuator which selects vertical deflection factor. Set it to an easily observable range corresponding to the amplitude de of the input signal.
Multiply the reading by 10 when the 10:1 probe is used in combination with the instrument.
Fine tuning device used to vary the vertical deflection sensitivity continuously. Attenuation of less than 1/2.5 is obtained when this device is rotated in the reverse direction of the arrow to the full.
This control is used when comparing waveforms or when measuring the rise time of a square wave in 2-channel observation. Normally this control is left rotated in the direction of the arrow to the full. When the knob is at PULL position (pulled up state) the gain of the vertical axis is magnified 5 times and the maximum sensitivity becomes 1mV/DIV.
Light when VAR is out of CAL detent position.
This knob used to adjusting the position of the vertical axis.
The image rises with the clockwise rotation of this knob and falls with the counterclockwise rotation.
When the knob is pulled up, the adjustment range of the trace position of the vertical axis can be magnified by the DC OFFSET function. Therefore, the peak value of an input waveform with large amplitude can be measured.
(Keep pushed-in for the normal operation.)
The same as CH1, but when the knob is at PULL position (pulled up state), this is used to inverse the polarity of the input signal applied to CH2.
This control is conveniently used in the comparison of two waveforms having different polarity or in the observation of the waveform of the difference signal (CH1) - (CH2) between CH1 and CH2 using ADD.
This switch is used to select the operation mode of the vertical deflection system.
Output connector providing a sample of the signal applied to the CH1 connector.
DC OFFSET
24 25 DC BAL
connector
There are used for the ATT balance adjustment. See 8. ADJUSTMENTS . for the details.
Sweep time ranges are 19 steps from 0.2 µs/div to 0.2s/ div.
X-Y This position is used when using the instrument as an X-Y oscilloscope.
In this position the X (horizontal) signal is connected to the input of CH1; the Y (vertical) signal is applied to the input of CH2 and has a deflection range from less than one millivolt to 5 volts/div at a reduced band-wide of 500 kHz.
This control works as CAL and the sweep time is calibrated to the value indicated by TIME/DIV.
TIME/DIV of sweep can be varied continuously when shaft is out of CAL position.
Then the control is rotated in the direction of arrow to the full, the CAL state is produced and the sweep time is calibrated to the value indicated by TIME/DIV. Counterclockwise rotation to the full delays the sweep by 2.5 times or more.
Light when SWP VAR is out of CAL detent position.
This knob is used to move the bright line in horizontal directions. It is indespensable in the measurement of the time of waveform.
Bright line is moved toward right when the knob is rotated clockwise and toward left with counterclockwise rotation.
Sweep is magnified 10 times by pulling out knob of POSITION. In this case the sweep times is 1/10 of the value indicated by TIME/DIV. Bring the position of the waveform desired to be magnified observed to the outer of the scale by operating — POSITION of the horizontal axis. Next, switch x10 MAG switch to PULL (pulled out state). Then the waveform placed at the center is magnified in right and left directions. The sweep
time in this case is 10 times the sweep speed obtained by TIME/DIV, in other words, the reading is 1/10 of the sweep time indicated.
CH1 input signal is displayed alternately by each single sweep of X1 (NORM) and X10(MAG).
This switch is used to select the triggering signal source sweep.
(32)
This switch is used to select the internal triggering signal source sweep.
MODE For observing two waveforms, the sync signal changes alternately corresponding to the signals on CH1 and CH2 to trigger the signal.
Input terminal for use for external triggering signal of sweep
This knob is used to decide at which portion of the waveform should the sweep be started by setting trigger level. This knob is also enabled to switch SLOPE.
Depressed position (normal state) is for SLOPE and PULL position (state in which the knob is protruding) is for SLOPE.
AUTO The instrument is brought into automatically triggering sweep in which sweep is always conducted.
In the presence of triggered signal, normal triggered sweep is obtained and the waveform stands still. In the case of no signal or out of triggering, sweep line will appear automatically. This setting is convenient in usual cases.
NORM Triggered sweep is obtained and sweep is conducted only when triggering is effected. No
Input terminal for brightness modulation. It is of the DC coupling. The brightness is reduced with a positive signal and increases with a negative signal.
Output terminal of calibration square wave of about 1kHz and 0.5V. It has a tip terminal. It is used to calibrate the probe combination.
Earth terminal of the oscilloscope.
(6) Readout function switches (V-525 only)
(41)
Selects the REF (X) cursor. Press this switch to make the intensity of the REF (X) cursor brighter. By pressing the REF switch (39) and the \Delta switch (40), bath the REF (X) and the \Delta cursors are displayed.
Selects the $\Delta$ (+) cursor. Press this switch to make the intensity of the $\Delta$ (+) cursor brighter.
The cursor selected by the REF switch 39 or the $\Delta$ switch 40 can be shifted in the directions of arrow marked on the switches.
NOTE: With the REF and the ∆ switches pressed, the two cursors can be shifted simultaneously in the directions of arrows with the CURSOR switches.
Controls the brightness of the cursors and the readout displays. A clockwise rotation increases the intensity. At this time the focus of characters may not match with that of a waveform. Adjust the FOCUS control to
obtain an optimum focus.
This switch is used to select the operation mode of the single time base with delayed sweep.
Main sweep appears on the screen. This is used in normal operation.
Although the sweep on the screen is main sweep it indicates the single time base with delayed sweep by intensity modulation.
• DELAY
The intensity modurated portion is magnified.
This control is used to set the starting point of the single time base with delayed sweep. Five ranges (1 µs-10 µs, 10 µs-100 µs, 100 µs-1 ms, 1 ms-10 ms, 10 ms-100 ms) of delayed time can be set continuously with DELAY VAR.
Delayed time control combined with DELAY TIME.
Before turning ON the POWER switch, insure the power supply voltage is within the range of 108–132V for AC 120V set, 198–242V for AC 220V set, and 216–264V for AC 240V set.
Insert the plug of the power cord on the rear panel into the power supply wall socket and set the controls as follows.
POWER | OFF | ||
---|---|---|---|
INTEN | Counterclockwise to the full | ||
FOCUS | Midrange | ||
AC-GND-DC | GND | ||
1, POSITION |
Midrange (the knob is in the de-
pressed) |
||
V. MODE | CH1 | ||
TRIG | AUTO | ||
TRIG SOURCE | INT | ||
INT TRIG | CH1 | ||
TIME/DIV | 0.5 ms/div | ||
Midrange |
Set all the levers of the switches to the upper side.
After ending all the setting mentioned above, turn ON the POWER and, 15 second later, rotate the INTEN knob clockwise. Then the sweep bright line will appear. If the observation
is to be started immediately, set the FOCUS control at a point where the bright line is sharpest.
If the instrument is not used with the power supply turned on rotate the INTENsity counterclockwise to reduce the brightness and also blur the FOCUS.
For usual observation, leave the following non-calibrating function section set to "CAL" position.
VARIABLE |
Rotate in the direction of arrow.
In this case the VOLTS/DIV is calibrated to its indicating value. |
|
---|---|---|
SWP VAR |
Leave the knob in depressed state.
In this case the TIME/DIV is calibrated to its indicating value. |
Align the bright line with the horizontal scale line at the center of the screen by operating CH1 POSITION. In some cases the bright line may be oblique to the scale slightly by the effect of earth magnetism. In this case, bring the bright line until it lies on the horizontal scale line at the center of the screen by properly adjusting the semi-fixed variable resistor TRACE ROTATION on the front panel.
Use CH1 or CH2 when not observing the phase difference between two waveforms or when engaging in a operation other than X-Y operation. Make the following settings when using CH1.
MODE Switch of Vertical defection system | CH1 |
---|---|
MODE Switch of TRIG | AUTO |
TRIG SOURCE | INT |
INT TRIG | CH1 |
Under these settings almost all the repetitive signals of about 25 Hz or more applied to CH1 can be synchronized and observed by adjusting TRIG LEVEL. Since the MODE of horizontal axis is at AUTO position, the bright line appears even when no signal is present or when input coupling switch is at GND position. This means that the measurement of DC voltage can be measured. The following switching is needed when observing low frequency signals of about 25 Hz or less.
NORM
When using only CH2, use the instrument after making the following settings.
MODE Switch of Vertical Axis | CH2 |
---|---|
TRIG SOURCE | INT |
INT TRIG | CH2 |
Observation of two waveforms can be made easily by setting the MODE switch of vertical axis to ALT or CHOP. When observing two waveforms of high repetition frequencies set the MODE switch to ALT and, in the case of low frequencies, set it to CHOP.
When measuring the phase difference, measure after effecting synchronization with leading phase signal.
The first step of measurement is introduce the signal desired to measure to the oscilloscope properly. Do it with utmost care.
Use the attached probe, AT-10 AK 1.5, when measuring a high frequency wave with high accuracy.
It should be noted, however, that since the input signal is attenuated by this probe to 1/10 before it is input to the oscilloscope the use of the probe is disadvantageous for low signals, and that at the same time the measuring range is extended by that amount for high signals.
(a) A good example
(b) A bad example
For better measurement it is required to use an earth attachment available at option.
Multiply the reading of VOLTS/DIV by 10. For example, if the VOLTS/DIV is 50mV/DIV, then read the waveform as
500 mV/div x 10 = 500 mV/div
• To avoid measurement error, put the probe in the following correction state and check it before measurement without fail.
Connect the tip of the probe to the output terminal CAL 0.5V of 1kHz calibration square wave voltage. When this correction capacity value is at optimum the waveform takes the shape as shown in Fig.(a) as follows.
If the waveform is as shown in Fig. (b) or Fig. (c), rotate the semifixed adjusting screw on the matching box of the probe by using a screwdriver until the optimum state is obtained.
(2) At time of direct connection
When connecting a signal directly to the oscilloscope not using the attached probe AT-10 AK 1.5 (10:1), pay attention to the following points in order to minimize the measurement error.
• When performing observation using a bare lead wire, no trouble occurs of the circuit to be measured is of low impedance and high level.
However, note that, in most cases, measurement error may be caused by static stray coupling with other circuit and power line.
This measurement error cannot be ignored even in low frequency region.
In general, it is safe to avoid measuring with nonshielded connecting wire. When using a shielding wire connect one end of the shield to the earth terminal of the oscilloscope and the other end to the grounding of the circuit to be measured. It is desirable to use a coaxial cable with BNC type connector.
• The following cautions must be observed when performing a wide band measurement. It is necessary to terminate with the characteristic impedance of the cable when measuring a rapid rising waveform or a high frequency wave.
Especially when using a long cable, the absence of a terminating resistor will necessarily lead to a measurement error derived from ringing phenomenon. Some measuring circuits require a terminating resistor equal to the characteristic impedance of the cable also on the measurement terminal side.
BNC type terminating resistor (50 Ω) is conveniently used for this purpose.
a capacity of about 100 pF per meter, its effect on the circuit to be measured cannot be ignored. Use a probe to minimize the effection the circuit.
When the length of the shield wire used or when the length of the non-terminated cable reaches 1/4 wave length or its multiples within the band of V-422 [V-222] type (1/4 wavelength is about 1.5 meter [about 3 meter]) when using a coaxial cable at 40MHz [20 MHz], oscillation may be caused near 5 mV/DIV range.
This is caused by the resonance between the externally connected high-Q inductance and the input capacity and can be avoided by reducing the Q.
Connect the cable or shield wire to the input connector by way of a serially connected 100Ω to 1 kΩ resistor, or perform measurement at, other VOLTS/DIV range.
Set the TIME/DIV switch to X-Y. Then the instrument works as an X-Y oscilloscope.
Each input is applied to the instrument as follows.
X-axis signal (horizontal axis signal) | CH1 | INPUT |
---|---|---|
Y-axis signal (vertical axis signal) | CH2 | INPUT |
In this case leave the horizontal axis mangnification switch (PULL-MAG x10 knob) at depressed position.
The first things to do are as follows.
Set input coupling to GND and decide the zero level properly.
Set VOLTS/DIV appropriately and set AC-GND-DC to DC. Since the bright line shifts here by the amount of DC voltage, the DC voltage of the signal can be obtained by multiplying the shift width by the indicated value of VOLTS/DIV. When VOLTS/DIV is 50 mV/DIV, then 50 mV/div x4.2 = 210 mV/However, if the probe AT-10 AK 1.5 is in use, the true value of the signal becomes 10 times the value, or 50 mV/div x 4.2 x 10 = 2.1V.)
The same as paragraph 7 (1), "DC voltage measurement", but here those is no need of matching the zero level with the scale line. Move the zero level at will to a position easy to observe.
In the drawing is follows, VOLTS/DIV is 1V/DIV, 1V/ div x5 = 5 Vp-p (50 Vp-p at time using the prove AT-10AK 1.5). When magnifying and observing a smallamplitude signal, superimposing on a high DC voltage, set input coupling of AC. The DC voltage is cut off and AC voltage can be observed by increasing sensitivity.
This will be explained taking the drawing at follows as an example.
One period covers the time A and time B, which are separated from each other by 2.0 div on the screen.
When the sweep time is 1 ms/DIV, the period is given by 1 ms/div x 2.0 = 2.0 ms
= 2.0 × 10-3 s
Accordingly, the frequency is
1/(2.0 × 10-3) = 500 Hz
(However, when the knob MAG x 10 is at pulled out position, TIME/DIV must be converted to 1/10 since the sweep is magnified.)
Triggering signal source "SOURCE" is selected as offering reference signal when measuring the time difference between two signals. Assume that pulse trains as shown in (a). Then (b) shows the case when CH1 is taken as the triggering signal source and (c) the case where CH2 is taken.
This means that CH1 is used as the triggering signal when investigating the length of time by which the signal of CH2 is delayed from the signal of CH1. CH2 is used in the reversed case. In other words, the signal leading in phase is selected as the triggering signal source.
If this process is reversed, the portion to be measured may sometimes not appear on the screen. Thereafter, equalize the amplitudes of the two signals appearing on the screen or superimpose one on another.
Read the time difference by the interval between 50% amplitude points of the two signals.
Sometimes the superimposing method is more convenient from the point of view of procedure.
Since the pulsed wave contains many high-frequency wave components (higher harmonics) depending on its width or
period, pay the same attention as given to high frequency signals when handling it. Accordingly, use a probe or coaxial cable and shorten the earth lead wire as much as possible.
To measure the rise time pay attention not only to the abovementioned items but also to measurement error. The following relationship exists between the rise time Trx of the waveform to be measured, the rise time Trs of oscilloscope, and the rise time Tro displayed on the screen.
Tro =
When the rise time of the pulse going to be measured is sufficiently longer than the rise time of the oscilloscope (7ns in our case), the effect of the rise time of the oscilloscope on the measurement can be neglected. However, if both are close to each other, measurement error may be caused.
The true rise time is given by
Trx =
Moreover, in general, in a circuit free from waveform distortion such as overshoot and sag, the following relationship is established between frequency band and rise time.
The rise time and fall time are determined by the time elapsed between the 10% to 90% values of pulse width. This oscilloscope is provided with graduations for 0%, 10%, 90%, and 100% on the screen, which facilitate measurement.
In the case shown in the Fig. (a) below where two waveforms greatly different in amplitude alternate, the waveform is doubled if the trigger level is not set properly. In the case where the trigger level is selected as Y line two waveforms, one starting with A and advancing to B, C, D, E, F, . and the other starting with E and advancing to F, G, H, I . , will appear alternately on the screen. They will be doubled as shown in Fig. (b), for which no synchronization can be taken.
In such a case, rotate LEVEL clockwise until the trigger level comes to Y' line. Then the waveform on the screen becomes the one is shown in Fig. (c) above which start with B and advances to C, E, F, . and which allows synchronization.
Table 1.
SOURCE | INT | LINE | EXT | |||
---|---|---|---|---|---|---|
INT | TRIG | CH1 | CH2 |
VERT
MODE |
||
v | CH1 | CH1 | CH2 | CH1 | ||
CH2 | CH1 | CH2 | CH2 | Line | External | |
M
O |
ALT | CH1 | CH2 |
CH1 CH
2(ALT) |
||
D | СНОР | CH1 | CH2 | ADD | ||
E | ADD | CH1 | CH2 | ADD |
When the SOURCE switch to INT, INT TRIG switch to VERT MODE, and MODE switch to ALT, the input signals applied to CH1 and CH2 become trigger source alternatively at each sweep. Consequently, even for an observation of two waveforms of different frequencies, the waveform of the each channel is stably triggered.
In this case, the signal should be applied to both CH1 and CH2, and the two signals have the same level portion in excess of the rated amplitude each other. There should be a common portion of levels available that is above the rated amplitude of CH1 and CH2.
When a sinewave is applied to CH1, and a square wave is applied to CH2, "A"s in Fig. 1 are the levels possible for synchronization.
In order to expand the synchronization range, AC coupling is applied to the CH2 side.
When either the CH1 or CH2 input signal is smaller as shown in Fig.2, adjust the VOLTS/DIV switches (13) and (14) to obtain sufficient amplitudes.
The VERT MODE triggering requires 1.5 div more than the amplitude required for an observation of CH1 or CH2.
Fig. 2
The VERT MODE triggering is not possible when the signal is applied to only one channel as illustrated on Fig.3.
Caution: Do not use the INT TRIG to the VERT MODE VAR when PULL x5 GAIN (15) and/or (16) are in the pulled out position (x5 GAIN mode).
Jittering wave as shown below may appear on the screen when a gently-sloping signal is displayed by approximately 10 cycles or less, with the VERT MODE for the INT TRIG switch and the ALT position for the MODE select switch. For detailed and clear observation of each signal, set the MODE select switch to the CH1 or CH2.
In the work concerned with TV, complexed signals containing video signal, blanking pedestal sitgnal, and synchronizing signal are often measured. However, since the waveform is complexed, a special circuit is needed to effect a stable synchronization with vertical waveform.
Exclusive circuit for con |
Exclusive circuit for this
instrument (Principle drawing) |
||
---|---|---|---|
General circuit |
Simple synchronizing
circuit |
TV exclusive synchronizing
separator circuit |
|
Circuits |
Video signal
To trigger circuit →> |
To trigger circuit |
Vc
To trigger circuit |
Hard to synchronize, because
video signal is applied directly as trigger signal. |
Synchronization is more easily
effected than in the circuit shown at left, because the signal is integrated to remove high frequency components. |
Stable synchronization is on-
tained since SYNC pulse is picked up, amplified, and then integrated to remove high fre- quency components. |
(NOTE) This oscilloscope synchronizes with only (-) synchronizing signal.
The oscilloscope provides the DC offset voltage display of ±1 to ±100V according to the range. An output terminal for voltage reading. (Except : x5 GAIN, out of CAL)
Fig. 7-1 Diagram of DC OFFSET
See Fig. 7–1 for the function of the DC OFFSET. There is following relation among the differential amplifier output voltage e0, the input voltages ei and E2.
(Where is a DC input and is an AC input)
Set the controls so that E2 = EDC is attained. Then e0 = A X eAC is obtained.
The DC component can be removed for an observation.
VOLTS/DIV | DC OFFSET voltage |
---|---|
5mV/div-50mV/div | more than ±1V (x1) |
0.1V/div-0.5V/div | more than ±10V (x10) |
1V/div – 5V/div | more than ±100V (x100) |
When measured with a DMM connection, multiply the DMM reading by the multiplier written above in indentaion (_____).
Read next paragraph for the detailed explanation of measurement with a DMM connection.
In order to readout the voltage level digitally, connect a DMM (digital multimeter) to the DC OFFSET output terminal and set the oscilloscope to the DC OFFSET mode. See Fig. 7–2.
1) Measuring DC component
Align level (b) with the center graticule and read the DMM digital value. (+2V should be displayed.)
When the above is performed with an input sensitivity of 50mV, then 4 div of amplitude must be present on the oscilloscope screen.
The AC component is observable on the oscilloscope screen where as the DC component is measurable on a DMM, and there is no need for any complicated switching as such needed for conventional oscilloscopes.
More over, the DC OFFSET function provides readings of peak to peak value (p-p) of eo on a DMM.
Align level (a) with the center graticule line on the screen, read the DMM value and name it V(a). Then align level (c) with the graticule line and readout V(c). The p-p
the difference between V(a) and V(c), and the value can be read digitally on the DMM.
As explained above, the DC OFFSET function with DC OFFSET output terminal offers an improved operational convenience and a highly accurate measurement in detailed waveform portions.
(11) Operation procedure of signal time base with delayed sweep. (V-523 only)
This instrument has a convenient function that enlarges any given portion of signal for easy observation.
<Figure b)>
4) Set the DISPLAY switch to DELAY position and expanded the display using TIME/DIV. <Figures c) & d)>
The following set values are displayed when the vertical made is not X-Y, the MODE select switch is not set to ADD, or the CH1 ALT MAG switch is OFF.
Set values of CH1 signal are displayed when the vertical mode is CH1, ALT, or CHOP. These values are not displayed in the CH2 mode.
The sweep time is displayed when the TIME/DIV select switch is set to a position other than X-Y.
x10 MAG TIME/DIV UNCAL : > Normal state · Blank TIME DIV : 20 ns to 0.2 s ΔV cursor measurement value display
Polarity display: +, -
àb c
Note: When the MODE select switch is set to CH2, the measurement value is displayed with divisions (0.00 to 8.00 div).
When the CH1 VOLTS/DIV select switch
is set to the uncalibrated position, "UNCAL" is displayed.
Measurement value display: 0.0 ns to 2.000 s
Note: When the SWP VAR control is set to the uncalibrated position, "UNCAL" is displayed.
b
Set input coupling to GND and decide the zero level properly.
Set VOLTS/DIV appropriately and set AC-GND-DC to DC. Since the bright line shifts here by the amount of DC voltage, the DC voltage of the signal can be obtained by multiplying the shift width by the indicated value of VOLTS/DIV. When VOLTS/DIV is 50 mV/DIV, then 50 mV/div x4.2 = 210 mV/However, if the probe AT-
10 AK 1.5 is in use, the true value of the signal becomes 10 times the value, or 50 mV/div x 4.2 x 10 = 2.1V.) the difference between V(a) and V(c), and the value can be read digitally on the DMM.
As explained above, the DC OFFSET function with DC OFFSET output terminal offers an improved operational convenience and a highly accurate measurement in detailed waveform portions.
The voltage difference (ΔV) and the time difference (ΔT) are simultaneously measured by the REF (X) cursor and the Δ (+) cursor, and the measurement values are displayed on the screen.
(a) Cursor selection
The REF (X) cursor and the \Delta (+) cursor is selected by their respective select switches. The intensity of the selected cursor is made brighter.
With the REF (X) and the cursors selected, the two cursors can be shifted simultaneously.
(b) Cursor shifting
Bring the REF (X) cursor to the desired position since the difference between the REF (X) cursor and the \Delta (+) cursor is measured, using the REF (X) cursor as a reference cursor.
Combination use of the CURSOR shift switches makes the cursor shift at angle of 45°.
Every pressing of the CURSOR shift switch shifts the cursor by one resolution and continuous pressing makes the cursor shift continuously.
(c) Measurement value display
The voltage difference (ΔV) and the time difference (ΔT) are simultaneously displayed on the upper part of the screen.
In case that the cursor is shifted at angle of 45°, the ΔV and the ΔT displays are simultaneously changed. The measurement values are the values automatically converted by the set values of the CH1 VOLTS/DIV switch and the TIME/DIV switch
In the x5 GAIN and the x10 MAG modes, the automatic conversion is performed.
In this case, press any one of the six switches to reset to the initial set values.
The ATT balance of the vertical axis can be made easily.
Large 6" screen with internal graticule. Approximate 12kV acceleration potential.
P31 standard
8 x 10 div (div = 10 mm) Internal graticule
Possible (with automatic focus correction circuit)
Trace rotation | Present |
---|---|
Brightness adjustment | Possible |
Scale illumination | Variable (V-522, V-523 only) |
DC-coupled, positive-going signal decreases intensity: 5Vp-p signal causes noticeable modulation at normal
intensity: DC to 2MHz
Input impedance 33k ohm(typ.)
Maximum input voltage 30V (DC + peak AC)
DC to at least 50 MHz and rise time 7ns or less. DC to at least 7 MHz and rise time 50ns or less at magnifier extends. The AC coupled lower-3dB point is 10 Hz or less.
5mV/div to 5V/div in 10 calibrated steps in a 1-2-5 sequence. Uncalibrated continuous control extends deflection factor to at least 12.5 Volts per division in the 5 Volts/div position. x5 magnifier increases sensitivity of each deflection factor setting to 1mV/div.
±3% Additional error for magnifier ±2%.
CH1, CH2 (normal or invert), Alternate, Chopped (approximate 250kHz), Added
Approximately 1 MΩ in parallel with 25 pF
300V (DC + peak AC) or 500 Vp-p AC at 1kHz or less
AC, GND, DC
DC OFFSET Possible
DC OFFSET VOLT OUT Present
Dynamic range 6 division or mode
0.2µs/div to 0.2s/div in 19 calibrated steps in a 1-2-5 sequence. Uncalibrated continuous control extends deflection factor to at least 0.5 seconds per division in the 0.2 sec/div position. x10 mag extends maximum sweep rate to 20 ns/div
±3%
Additional error for magnifier ±2%
Automatic, Normal, TV (TV-H or TV-V)
Internal (Ch1, Ch2 V-MODE), Line, External
+, -
TV (-)
Frequency | Internal CV-MODE | External |
---|---|---|
20 Hz – 5 MHz | 0.5 div (2.0 div) | 200 mV |
5 MHz – 40 MHz | 1.5 div (3.0 div) | 800 mV |
40 MHz – 50 MHz | 2.0 div (3.5 div) | 1 V |
TV-V sensitivity: SYNC section less 1 div or 1V AUTO low band: Approximately 25 Hz
Trigger coupling
AC : 20 Hz to full bandwidth
Approximately 1 MΩ in parallel with 30 pF
Maximum input voltage 300V (DC + AC peak)
ALT MAG Present
Same as vertical deflection
X-bandwidth
DC to at least 500kHz
3°C or less from DC to 50kHz
An approximate 1kHz frequency 0.5V (±3%) square wave.
Output voltage is at least 20 mV/div into a 50 ohm load. Bandwidth is 50 Hz to at least 5 MHz.
Delay time 1 µs - 100 ms, 5 ranges Jitter 5,000: 1 5,000 : 1
(applicable to the modes other than X-Y) Vertical axis (CH1 only): V/DIV, UNCAL MAG (converted value)
Voltage difference
Time difference ΔT : Δ - REF
Note: Displayed when the vertical mode is the modes other than ADD.
Vertical | : | Within ±3 DIV |
---|---|---|
Horizontal | : | Within ±4 DIV |
VOLTAGE (50/60Hz) | FUSE |
---|---|
100 V ( 90 – 110 V) | 1A |
120 V (108 - 132 V) | 1A |
220 V (199– 242 V) | 0.5A |
240 V (216– 264 V) | 0.5A |
Power supply frequency: 50, 60, 400 Hz
Power | consumption: | |
---|---|---|
Approx. 30W Max. 40W at 120V 60Hz
Limit of operation temperature | 0 −+ 40°C |
---|---|
Limit of operation humidity | 35-85% |
Rated range of use temperature | +10 - +35°C |
Rated range of use humidity | 45-85% |
---|---|
Storage and transport temperature | -20-+70°C |
Approx. 310(W) x 130(H) x 370(D) mm (12.4(W) x 5.2(H) x 14.8(D) inch) Approx. 6.5kg (14.6 lbs)
V-525
V-523 Block Diagram
V-522/V-523/V-525 TRIG GEN 4
V-522/V-523 POWER SUPPL Y 🛞
V-522/V-523 RV/LAMP (9)-1
/31 .9