HIOS HP-10, HP-1, HP-100 Operation Manual

Digital Torque Meters
HP-100 / HP-10 / HP-1
(Option)
Operation Manual
(May 2019)
ET-C006 19A
Contents
Changing the Fidaptor shock absorber spring
Applications and features of
HIOS Digital Torque Meters ............................2
Safety Precautions ............................................ 3
HIOS Digital Torque Meter controls .................. 4
List of items included with torque meter and
explanations of Fidaptor
and threaded shaft ......................................... 5
Fidaptor .............................................................6
Operating instructions ........................................7
How to measure screwdriver torque
using the Fidaptor ...........................................7
Explanations of functions ............................ 8
DIP switch .......................................................8
Torque settings ...............................................8
Memory ..........................................................9
Average torque measurement ......................10
However high the quality of your screwdriver, you cannot make full use of its potential unless you know the right amount of torque for each job and make sure that is the torque being delivered. With a HIOS digi­tal torque meter you can make torque settings on screwdrivers, check torque wrenches or make torque measurements of various kinds of rotation tools. With an optional attachment, it is also possible to transfer measurement data to a computer for use in integrated production management systems.
-
Care of the Fidaptor .........................................10
Fidaptor parts ...................................................11
Using the adapter socket for torque
measurement of instruments
other than screwdrivers ................................12
Battery charger.................................................12
Types of data output.........................................13
Customer service .............................................13
Inspection and calibration ................................13
Troubleshooting ...............................................14
Specifi cations ...................................................14
10
External features diagram ................................15
Analog and serial data
output specifi cations .....................................15
China RoHS2 Table ..........................................16
........10
Easy to use, accurate and low in cost, HIOS digital torque meters are used by manufacturers of electronic goods, cameras, watches and other kinds of precision instruments, as well as by makers of automobiles and aircraft.
Torque value display can be automatically reset
Applications
to zero.
Torque measurement of power screwdrivers,
torque drivers and torque wrenches
Certifi cation of performance of torque tools
Torque measurement of other kinds of rotation
tools (may require adapter)
Examples of uses of torque measurement data
Reference for preparation of tap screw coun-
terbore diameters
Sorting screws by required torque values
Classifying jars by screw cap tightness
Destructive testing of small parts
Quality control of plastic screw threads and
plastic parts
Features
HP-1 is on request product. It can be measured
below 10N•cm.
Up to 100 data measurements can be stored in
memory at a time.
Checks maximum, minimum and average
Torque value display are automatically recali-
brated.
Values are displayed digitally for ease and ac-
curacy in reading out data.
Peak torque values are frozen until the display
is reset to insure accuracy when setting torque of instruments.
Can be set for rotation in either direction, for
easy measurement of both tightening and loos­ening torque.
Counterclockwise torque measurement is indi-
cated by a minus sign in front of displayed values.
The strain meter is directly attached to the
torque sensor, as part of the instrument's simple, rugged design.
Can be integrated into a statistical process con-
trol system.
Small, light and operated by rechargeable bat-
tery, this instrument is highly portable.
torque measurements.
-2-
Data can be output analogically, as waveforms
or numerical printouts, for a wide range of uses. (Requires optional analog cord.)
An input tool is available to make possible
transfer of measurement data to a computer. (See page 13 for details.)
Comes with an AC adapter that can be used
with 100− 240 volt power sources. Runs on an environment-friendly nickel-hydrogen battery.
Safety Precautions
Please read the following safety precautions to ensure safe and correct operation.
Before using
Before you begin using the instrument, read the
operating instructions and also the precautions given on the seal affi xed to the instrument.
HIOS cannot assume responsibility for ac-
cidents or diffi culties when the instrument has been remodeled or disassembled, or when it has been used in a way other than that explained in the operating instructions.
During operation
Keep the working area orderly to ensure that
extraneous objects do not become entangled when the screwdriver or other instrument being measured is operated.
Never point the driver bit in any direction other
than into the torque meter.
If the torque meter shows any functioning ab-
normality, cease operation immediately and contact HIOS Corporation or your HIOS dealer.
Operators of power tools should always button
cuffs and shirt fronts and fasten zippers on ap­parel when conducting operations.
Operators of power tools should not wear gloves,
as this interferes dangerously with having a proper grip on the tools.
Precautions in operating a digital
torque meter
Never exceed the maximum torque levels indi-
cated in the specifi cations on page 14. Exceed- ing capacity load levels by 120% will result in internal damage to the meter.
The battery charger is for NiMH batteries only. Nev-
er use any other kind of battery, even from HIOS.
Operating the Fidaptor
See pages 10 and 11 concerning operation of
the Fidaptor.
Do not use this instrument for measurement of
continuous impact tools like air drivers or im­pact wrenches.*
Do not disassemble the instrument or loosen the
screws on its body, as this could cause inaccu­racy in measurement.
Use of the battery charger
See page 12 for instructions on battery charging.
Use only the products specifi ed in these operat-
ing instructions for connection to the data out­put ports.
When disconnecting cables, grasp the cable plug
and take care not to damage connector pins.
Always turn the power switch off after complet-
ing measurement operations.
Do not strike the acrylic display panel or place
heavy objects on it.
Avoid excessive internal adjustments for cali-
brating or other purposes.
Handle the instrument with care and take care
not to drop it.
This instrument is sensitive to ambient condi-
tions. Avoid using it in the following kinds of places:
Places where water, oil or other liquids could
fall onto or get inside the instrument.
Places subject to vibration, dust or hot air.
Outdoors, or places where electrical discharg-
es may occur.
Places of high temperature or humidity. (Operat-
ing humidity and temperature ranges for digital torque meters are 25%−65% and 15
Any other place where there is a danger of the
35℃).
instrument being damaged or its functioning impaired.
Please do not store the instrument in places of
high temperature or high humidity. This could result in condensation forming inside, which could endanger functioning.
* HIOS offers other instruments for measurement
of air drivers, impact wrenches, etc. Please con­tact HIOS Corporation or your HIOS dealer for information.
-3-
HIOS Digital Torque Meter controls
Functions
Power switch【POWER】 Button switch. The LED lights when the power is on. * The power is automatically turned off when the instrument is not operated (or torque
Reset switch【RESET】 Used to return the display to zero or print out data. Display Measuring unit display
Unit selector【UNIT】 Switches between one of the following combinations of torque measurement units,
Measuring mode selector PEAK: Freezes display at peak torque value. (Press reset to cancel)
MODE】 TRACK: Displays current torque value.
F.PEAK: Freezes display at fi rst peak torque value. AC connector For battery charger. Analog data output Port for analog voltage data output.
ANALOG OUT】
Digital data output Press reset to output measurement data. This functions makes it possible to use the
DATA OUT】 instrument in statistical process control systems.*
DIP switch For setting auto reset timer and turning data output on ("P.ON" LED indicates "on" setting). Torque set switch A【SET】 Sets maximum torque load at which warning tone will sound Memory function toggle switch The LED lights when the memory function is on.
M_SET
Torque set switch B Max load warning tone The LED lights when this function is on. on/off switch【BZ Memory clear switch【M_CR】 Deletes data in memory. Average measurement When average measurement mode is on, the LED lights and the torque measure-
mode switch【AVE】 ment average values are displayed. Adapter socket For connecting of torque tool, using a HIOS Fidaptor or other type of adapter.
*1 Please push the power switch and detach it immediately until a buzzersounds. *2 Statistical Process Control (SPC) uses statistical techniques to monitor variations in manufacturing processes for use in quality control.
Control Function
load is less than 10) for a period of 10 minutes.
depending on which combination the purchaser ordered: Nmlbfin, NmNcm, Nmkgfcm
Sets the torque level at which the warning tone sounds.
【▲】
Model Nm Ncm Nmm HP-1 0.00 0.0 HP-10 0.000 0.0 — HP-100 0.00 0
1
2
-4-
Items included with torque meter
100
Market
Areas
Korea
Taiwan
Hong Kong
Singapore
Europe
U.S.A.
Canada
Australia
Korea
Taiwan
Hong Kong
Singapore
Fidaptor with Safety Guard
HIOS shank 5
Measuring range:
0.5 3N•m (Black Spring) Part No. TF5G-Z 1 pc.
Hex. shank
(Convertible shaft) 5 mm sub­tense (Upper)
6.35 mm subtense (Lower)
Measuring range:
0.5 3N•m (Black spring) Part No. TF6XG-Z 1 pc.
Hex. shank (convertible shaft)
Part No.
TF6XG-0010
HIOS shank 4
Measuring range:
0.15 0.6N•m (Yellow Spring) Part No. TF4G-Z 1 pc.
Model
HP-
Supplied Accessories
Driver Shaft
HIOS shank 4
5 mm subtense (Upper)
6.35 mm subtense (Lower) 1 pc.
TF5G-0010-4
Part No.
Accessories for exclusive use of HP-100.
Grease(*)
1 pc.
Part No. TF-G3 1 pc.
NOTE:
“Grease” and “Fidaptor Buffer Spring” are attached
to HP-100 with H5 shank model only.
* Grease is to be applied on the Fidaptor with
safety guard and Driver shaft appropriately.
Fidaptor
Buffer Spring
To use for measuring Lower range:
0.15 0.6N•m (Yellow Spring) 1 pc.
Exclusive Battery Charger
TCH-MULTI
Part No.: (for AC100-240V, A plug) PSE, cUL
JAPAN, U.S.A, Canada, Mexico, Taiwan, Thailand, Philippines
TCH-MULTI-CN
Part No.: (for AC100-240V, A plug) CCC
China
TCH-MULTI-CE
Part No.: (for AC100-240V, A plug) CE
Vietnam, Indonesia, India, Singapore, Malaysia, Korea, Hong Kong, U.K., Ireland, South America, The Middle East, Africa, Australia, New Zealand
TCH-EU
Part No. (for AC100-240V, C plug) CE
Europe (except U.K., Ireland)
1 pc
1 pc
1 pc
1 pc
Calibration Sheet
HP-10
Europe
U.S.A.
Canada
Australia
Korea
Taiwan
Hong Kong
HP-1
Old models of the Fidaptor and the intershank are not compatible with model HP-100.
The main unit and accessories come all together in a dedicated aluminum attache case.
No Fidaptor is available for HP-1. Please make test piece reffering to page 12 and measure with your screw.
Singapore
Europe
U.S.A.
Canada
Australia
Hex. shank
(convertible shaft) 5 mm subtense (upper)
6.35 mm subtense (Lower)
Measuring range:
0.15 0.6N•m (Yellow Spring) Part No. TF6SG-Z 1 pc.
(quantity: 3)
Note : Other battery chargers are not available.
-5-
Fidaptor
Fidaptor with safety cover
The Fidaptor is included with the torque meter. Connects clutch-type power screwdrivers to the torque meter socket, for measurement of torque
Limit Line on the Fidaptor TFG series
How to Use The Fidaptor TFG series
Loosen the spring until the top of
the bearing housing reaches to the
desired area.
The desired area to loosen the spring up to
at the point where the clutch disengages. Features a safety cover and a mechanism that has almost the same effect as that obtained in actual tapping screw fastening work.
Parts name: Bearing housing
Note (to keep accuracy of torque
measurement)
• Loosen the spring up to the desired area for storing the Fidaptor properly.
• Scratch or wipe with chemicals such as benzin or alcohol could cause the serigraphed “Limited line” peel-off.
• We recommend to follow the guide written in the “ maintenance and inspection of of the Fidaptor “ in the operation manual and replace the Fidaptor driver shaft with a new one after 2,500 strokes.
Threaded shaft
(comes with Model HP-100)
Use as adapter for bits that are not the 5 mm HIOS standard diameter. (Refer to diagram on page 10
when changing shock absorber spring.)
-6-
Operating instructions
First check the state of the battery charge. When the power is on and the battery has an insuffi cient charge, the display reads "LOBAT." When this happens, use the battery charger to recharge the battery for at least 3 hours and no more than 6 hours.
1. Make sure the digital torque meter is stable.
2. Select the unit of measurement desired
UNIT】 switch). (N
(N•mkgf•cm) (N•cm←→N•mm, only for HP-1)
3. Conduct zero adjustment. Set【MODE】 switch to "TRACK", then press both the 【SET】and
switches at the same time for less than 2
【▲】
seconds. This adjusts the meter to zero. (Note: If you press the switches for than 2 seconds, the meter will go into another mode. If this hap­pens, press【RESET】, then repeat the zero adjustment process.)
4. Set【MODE】switch to "PEAK" or "F.PEAK".
 In peak mode the peak torque value reached
during measurement is displayed. Thus, the dis­play does not change. Model HP-100 measures up to 10 N•m of torque, which corresponds to a display value of 15. Please conduct measure­ments within this range.
In "F.PEAK" mode the first peak value is
displayed and does not change even if higher torque is applied.
5. Press the【RESET】switch to cancel the dis­play.
6. Minus signs appear in front of measurements of counterclockwise torque.
7. To output data using a Mitutoyo Input Tool, set the No. 3 connector of the DIP switch (page 8) to the on position (so the "P.ON" LED lights), set the【MODE】switch to "PEAK" and press the【RESET】switch (which functions the same as the Enter key on a computer) . The Input Tool will convert the measurement data for direct input to a spreadsheet application. In order to output data, please conduct measure­ment within the instrument's peak measurement range, as indicated in the specifi cations.
 Minus signs denoting counterclockwise torque
are transferred with the measurement data.
8. Use the Fidaptor to measure torque of power screwdrivers, as explained below.
9. It is possible to fashion adapters for torque mea­surement of other types of tools. (See "Using
mN•cm) (N•mlbf•in)
the adapter socket for torque measurement of instruments other than screwdrivers.")
10.Be sure to turn off the power and remove the tool from the socket when you are fi nished mea- suring.
How to measure screwdriver
torque using the Fidaptor
1. Insert the Fidaptor into the digital torque me­ter's socket, then fi t the bit of the screwdriver to be measured into the head of the Fidaptor. The Fidaptor is for bits of 5 mm. Model HP-100 comes with two threaded shafts in addition to the Fidaptor that can be used for three differ­ent bit diameters. (Refer to diagram on page 10 when changing shock absorber spring.)
2. Take some tension off the Fidaptor's spring by running the screwdriver in reverse "REV", so that spring can be turned slightly by hand.
3. Press the reset switch to return the display to zero.
4. Run the driver in the forward direction "FOR" until it stops automatically. The Fidaptor spring will tighten.
5. When the driver stops the display will show the output torque of the driver.
6. Release the tension on the spring by running the driver in the reverse "REV" direction, then press the reset switch to cancel the display.
7. Repeat the above procedure to confirm the output torque value. Refer to the output torque guide table in the operating instructions for your HIOS power screwdriver to set the torque adjustment nut to an appropriate torque value.
-7-
Explanations of functions
DIP switch
Always turn off the power before changing DIP switch settings.
Auto-reset
DIP switch settings
A.RESET TIME P.ON/OFF
Switch number
  
Time to auto-
1
reset
  
Data output
Notes
1
The reset switch does not function when auto-reset is on.
2
The error display (see below) appears when a printer
1 sec ON ON
2 sec ON OFF
5 sec OFF ON
OFF OFF OFF
Printer ON
Printer OFF
1 2 3 4
−−
− −
−−
2
−−ON−
− −
− −
OFF
−
or the Mitutoyo Input Tool is connected for data out­put and the LED lights.
Error display
Data output
1. Set the mode switch to "PEAK" or "F.PEAK".
2. Set data output DIP switch number 3 to the on position.
3. Connect a printer or the Mitutoyo Input Tool to the Digital data output port. (Pressing the reset switch in this state will result in the error dis­play. An error will also result if you try to print out data with a printer that cannot print out nu­merical data containing decimal points.
4. Conduct measurement, then press the reset switch to output the data. Average【AV E】values can­not be output.
Output of data from the memory (memory func-
tion on)
Note: Please cancel auto-reset.
1. Press the reset button for 2 seconds.
2. When the display shows the let-
ters "Prn," press reset.
3. All the torque data, from memory number
0001, will be output. The display fl ickers dur- ing data transfer.
Setting torque
The digital torque meter is shipped with the warn­ing tone set to sound when maximum torque load is reached. This tone also sounds when the torque set-
ting has been changed. The torque setting can be changed as desired. The warning tone signals that the set torque value has been measured.
Example (torque set at 100) Data is output only for measurements up 100.
Setting
Warning tone
Track Peak
Reset
ON ON
Switch functions for setting torque Switches used
① ②
SET
SET】switch
① 【
RESET
1. Press for 2 seconds to enter torque setting mode. Press for 2 seconds once again to exit torque set­ting mode.
2. When the meter is in
Setting torque
torque setting mode, the 4-digit display
0 - 90 or 1 0 - 9 0 - 9
can be adjusted, one digit at a time. Use the SET switch to select the digit for adjustment. The selected digit blinks on and off. Each time the SET switch is pressed, the next digit to the left is selected (or, if pressed while the leftmost digit is selected, the rightmost digit will be selected.)
② 【▲】
switch
When the meter is in torque setting mode, this
switch increases the selected number by one each time it is pressed. Note that the leftmost digit can only be 0 or 1.
RESET】 switch
③
Pressing this
switch readies
Setting torque
(Example: Changing a torque setting from "0000" to "0100")
Ready for
measurement
①【SETswitch
The last 0 flickers
Press for 2 seconds to enter torque setting mode.
The last 0 is now selected.
the meter to begin measure­ment. Pressing it while in torque setting mode cancels the pro­cess and takes the meter out of torque setting
①【SETswitch
②【▲】switch
①【SETswitch
Press twice to select the third 0.
Press once to increase the selected digit from to 1.
Press for 2 seconds to exit torque setting mode.
mode.
-8-
Memory
Objective
Measurement data can be stored in the memory for management of everyday operations.
Storing and reading out measurement data
1. The memory can store 100 measurements. The measurements are numbered automatically, from 0001 to 0100. If data from previous measurements remains in the memory, the new measurements begin from the next available number. (Fig. 1)
2. Fig. 2 shows how to display data stored in the memory.
3. When the 100th result has been stored in the memory, delete some of the data (Fig. 3) or all of the data (Fig. 4) to make room for the next series of measurements.
Memory full display
The fl ickering display "FUL" ap- pears when the memory is full. To clear this display, press【RESET】.
Fig. 1 Recording readout
Example of memory readout
Memory mode on (LED on)
Ready for
measurement
Measurements of driver
torque
③【RESETswitch
Fit the power screwdriver into the socket, using the Fidaptor or other adapter
When the memory function is on, pressing this switch returns the instrument to a state of readiness to begin recording data again. Press to end recording, to display or to delete data.
Fig. 2 Displaying recorded data
Example of data display
(Memory function on, LED on, measurement data in memory)
Ready for
measurement
①②【SETand【▲】
switches
①【SETswitch
②【▲】switch
③【RESETswitch
SET】advances the display to the next measure-
ment result.
【▲】
urement result.
Press once to advance the display to the next data result. The data number (e.g. 0002) is displayed below. The measurement data and data number flicker alternately.
Press once to advance the display to the previous data result.
Data readout mode terminated when this switch is pressed, when there is no action for 30 seconds or a new torque signal is received.
goes back to the previous meas-
Switch functions in memory mode
Switches used
① ② ④
SET
SET】switch
①
RESET M_SET M_CR
When the memory function is on, pressing this
switch displays the last measurement in the memory.
② 【▲】
switch
When the memory function is on, pressing this
switch displays the first measurement in the memory.
RESET】switch
③ 【
When the memory function is on, pressing this
switch returns the instrument to a state of readi­ness to begin recording data again. Press to end recording, to display or to delete data.
M_SET】 switch
④ 【
Press for 2 seconds to enter or exit measuring
mode. The LED lights when measuring mode is on.
M_CR】 switch
⑤ 【
Pressing this switch deletes memory contents.
Note that data can be output (see "Data Output"
on page 8) whether or not the LED is on.
Fig. 3 Partial deletion of measurement data
Example of partial deletion of torque value data
(Memory function on, LED on, measurement data in memory)
Ready for measurement
Memory readout mode
on
⑤【M_CRswitch
Exit partial deletion mode RESET】pressed or no action for 30 seconds
Data deleted
⑤【M_CRswitch
Display measurement result for deletion
Partial data deletion mode display
Press once
Each time a measurement result is delet­ed, each of the measurement results shift one place backwards to occupy the va­cated places in the memory.*
Display flickers
* Partial deletion of measurement results
Data shifts to previous position
DATA1 DATA2 DATA3 DATA4
0001 0002 0003 0004
DATA1 DATA3 DATA4
0001 0002 0003 0004
Fig. 4 Full deletion of measurement data
Example of full deletion of measurement data
(Memory function on, LED on, measurement data in memory)
Ready for measurement
⑤【M_CRswitch
Exit full deletion mode RESET】pressed or no action for 30 seconds
Data deleted
⑤【M_CR】switch
Full data deletion mode display
Press for 2 seconds
Press once; data deletion takes 5 seconds
Display flickers
-9-
Measurement of average torque
Objective
Simple monitoring of torque for normal operations
Measurement procedure
Makes possible measurement of the minimum, average and maximum values among sets of mea­surements, each set consisting of up to ten measure­ments.
1. The measurement results are automatically num­bered, from 0001 to 0010. Negative displays are not recorded. Measurements in average mode are automatically deleted when the meter exits aver­age mode.
2. If less than 10 measurements are made, the dis­play shows only those measurements.
0001 0002 0003 0010
1st measurement deleted
0001 0002 0003 0009 0010
0001 0002 0003 0009 0010
11th measurement recorded
Switch functions in average measurement mode
Switches used
③
Notes
1. Be sure to set【MODE】switch to "PEAK" or "F.PEAK" before entering average measurement mode.
2. The【UNIT】switch does not function in average measurement mode.
Changing the Fidaptor's shock-
absorbing spring (Model HP-100)
Two springs are supplied with Model HP-100. The black spring is high-tension and the yellow spring is low-tension. The 5 mm Fidaptor is shipped with the black spring attached. For low-torque measure­ment, replace with the yellow spring.
Fit the driver bit into the head of the Fidaptor.

Turn the driver on in the reverse direction to loosen the threaded shaft and change the springs.
RESET
RESET】 switch
③ 【
AVE
Puts the instrument into a state of readiness to
begin measurement when in average measure­ment mode. Press to end measurement or dis­play of measurement sets.
AV E】 switch
⑥ 【
Press for 2 seconds to enter or exit average
measurement mode. The LED lights when in average measurement mode. The figure show the average measurement and display cycles.
Example of average measurement
(Average measurement mode, LED on.)
Ready for measurement
Torque measurement
③【RESETswitch
Display of average measurement data
(Data is displayed while in average measurement mode and is deleted when the mode is terminated.)
⑥ Each time theAVEswitch is pressed minimum, average and maxi- mum values are displayed flickering.
Fig. 1
Use Fidaptor or other adapter to fit torque tool into the torque meter socket.
Press once. "0001" flickers.
Repeat measurements until sufficient data is obtained.
Min Ave
(Same for auto-reset.)
Max
Care of the Fidaptor
Please read carefully the following points about handling the Fidaptor.
Using the Fidaptor
1. Never place a load on the Fidaptor that exceeds
the indicated measurement capacity.
2. Use the Fidaptor that is designed for the torque
level of the tool to be measured.
3. Keep the Fidaptor loose, and after measurement
remove it from the socket.
4. Keep the driver and Fidaptor upright during mea-
surement. Keep torque from the upper surface to within 5 kg, or 2 kg, in the case of a low-torque Fidaptor.
5. When conducting a series of measurements, ap-
ply grease* to parts ①, ② and ⑤ of the Fidap­tor, as shown in the fi gure on page 11.
-10-
6. See that each measuring cycle lasts 5 seconds or more to minimize wearing of parts.
7. Never attach the Fidaptor unless you intend to use it immediately. When it is not in use, always remove it from the torque tool and remove the tension from the spring.
8. Always be sure that the Fidaptor is properly at­tached to the torsion tool.
9. Do not use the Fidaptor if it has been bent or in some way modifi ed.
* Use the grease supplied with the torque meter to
apply to parts as necessary. This grease can be purchased separately (part number TF-G3).
Fidaptor parts
Name of part Type Part number
Threaded shaft
①
Bearing set
②
Bearing housing
③
Spring
④
Plate nut
⑤
Maintenance and inspection of the Fidaptor
1. Apply the grease periodically to parts ①, ② and in the fi gure below.
2. Always inspect the parts of the Fidaptor before
use.
(1) Is there grease on parts ①, ② and ⑤?
(2) Have the threads of the threaded shaft be-
come worn?
(3) Is there dirt or foreign material on the
threaded shaft?
3. The Fidaptor must eventually be replaced. It
should be periodically inspected and replaced when necessary.
HP-100 Hex shank
HP-10 HIOS 4 mm TF4G-0010 HP-100 TF5-0020 HP-10 CL65-0250 HP-100 TF5G-0030 HP-10 TF4-0030 Black spring (0.5 - 3N•m) TF5-0051 Yellow spring (0.15 - 0.6N•m) CL65-0040 HP-100 TF5G-0050 HP-10 TF4G-0050
HIOS 4 mm TF5G-0010-4
HIOS 5 mm TF5G-0010
(fi ts 5 mm and 6.35 mm sockets)
TF6XG-0010
Guide to when it's time to replace Fidaptor parts
:
① ,② ①
④ ⑤
Note
The safety cover provides protection against the possibility of shaft breakage due to wearing or metal fatigue, but for an added measure of safety it is recommended that the user follow the above guide to parts replacement.
Other parts (available for separate purchase)
Item
Fidaptor Hex. shank
Fidaptor Hex. shank
Standard accessory set for HP-100
Driver
Shaft
6.35 mm subtense
6.35 mm subtense
Measuring
Range
3 9N•m TF6U-Z
3 9N•m TF6UP-Z
Threaded shaft and bearing set: After 2,500 strokes
:
Threaded shaft only: When it becomes bent or its threads are worn
:
Spring: Replace together with threaded shaft after 2,500 strokes
Plate nut
:
Part No. Note
TF5G-SET
: After 5,000 strokes
*1
For HIOS torque drivers with lever­start.
*2
For HIOS CL-9000 or high torque derives with both lever-start and push-to-start. See the Manual p.4 of Standard accessories (without battery charger).
-11-
*1 TF6UP-Z
35mm
*2 TF6U-Z
15mm
Use of the socket
In addition to the Fidaptor or threaded shaft, it is possible to make one's own adapter to fi t torsion tool bits of different diameters into the socket for measurement. Keep the following points in mind when fashioning such an adapter.
Details of the socket
+0.04
+0.060
9.5 +0.035
20.0 -0.02
6.0
6.0
HP-10 • P-100
HP-1
+0.04
-0.02
20.0
6.0
45.0
Cautions in making a socket adapter
Make sure that the adapter fi ts snugly into the
socket.
Make sure the adapter is strong enough that it
won't break during measurement.
Do not attempt to use an adapter that delivers
more than 10 kg of vertical torque.
Do not use this torque meter to measure impact
torque tools.
Take care to observe all other principles of
torque measurement.
9.0
10.0
6.0
4-M3.0 tap
29.0
0.5
24.0
2.7
5.0
Attention
The dimensions indication of the socket part is not to scale.
ڧ
Battery charger
To charge batteries for the HIOS torque meter use only the battery charger provided with the meter. Always turn the power off when charging. The fi rst time will require 6 hours for a full charge.
Caution
Do not recharge the battery for longer than 6 hours.
Use only NiMH batteries, never any other kind,
even if obtained from HIOS Corporation.
Do not use the meter while charging the battery.
When the letters "LOBAT" appear in the display,
stop measurement and recharge the battery.
±0.5
14mm in depth
+0.05
6.35
-0
Do not use the battery charger for any purpose
other than charging batteries for this instrument.
Do not place objects on top of the battery char-
ger's cord. Do not subject the cord to extreme bending or tie it in knots.
Be sure the power to the instrument is off when
attaching or detaching the battery charger.
Do not attempt to remove the battery from the
instrument.
Danger!
Leaving the battery charger on for the maximum
recharge time of 6 hours could lead to overheat­ing, damage to the instrument or fi res.
-12-
How to recharge
1. Turn off the power to the torque meter and insert the battery charger cord into a power source.
2. When recharging is complete, unplug the cord, turn the torque meter power switch on and check that the display goes on.
3. Wait for the battery charger to cool, and store it in an appropriate location.
Types of data output
Use the Mitutoyo Input Tool (available for pur­chase from many dealers) to transfer measurement data to a computer.
Model: IT-005D 264-005 (For any standard Win-
dows computer)
Connection cable NO.937244 2m (sold sepa-
rately)
(Please contact Mitutoyo Corporation or a Mitutoyo dealer about purchasing these items.)
Types of analog data output
An analog cord is available for separate purchase
1.5 m, item HP-8060
( connect a pen recorder, oscilloscope or voltmeter to the HIOS torque meter. When using these instruments, read their operation manuals to insure correct operation.
Customer service
Repair
1. Service charges will be made for repairs under the following circumstances:
(1) Malfunctioning or damage caused by incor-
rect use of the instrument, the instrument has been disassembled or it malfunctions due to attempted repair.
(2) Oil has been added to the socket, switches
or inside the instrument.
(3) The instrument has been damaged during
shipping, by dropping, etc.
(4) Damage due to fi re, exposure to gas, earth-
quake, water, irregular power supply or other type of disaster.
(5) Service charges will also be made for cali-
bration, inspection or parts replacement for the Fidaptor, etc.
2. No charge will be made for service in the case of inspection and/ or calibration of the same part that becomes necessary within three months af­ter inspection or calibration has been performed.
). This cable can be used to
(This does not apply under circumstances (1) (4) above.)
Shipping and handling charges incurred for re-
pair service must be paid by the customer. Please direct questions about customer services to HIOS Corporation or your HIOS dealer.
Attention
The product that you have purchased contains a rechargeable battery. The battery is recyclable. At the end of it's useful life, under various state and local laws, it may be illegal to dispose of this battery into the municipal waste stream. Check with your local solid waste offi cials for details in your area for recycling options or proper disposal.
Inspections and calibration
HIOS torque meters employ a unique sensor mech­anism that maintains accuracy for long periods. A tool for checking its accuracy (the HIOS calibration check bar) is provided, but in addition to this, we recommend that the torque meter be sent in to HIOS Corporation for inspection at least once a year. (Note that a fee is charged for this service.)
HIOS torque meters are shipped calibrated to meet our accuracy standards. Depending on the condi­tions under which the meter is used, accuracy may be affected over the course of months or years. Therefore, HIOS offers calibration and, when nec­essary, overhaul services. We also recommend use of the periodic calibration documentation included with the torque meter. This documentation shows the relation of the HIOS standards to Japan's stan­dards.
Cautions to be observed when
sending the torque meter in for calibration or repair
1. Pack the meter in the attache case in which it was originally shipped to protect it from violent impacts during shipping.
2. Do not put anything into the attache case other than the items originally shipped inside. Never include items such as homemade mounting plates, tools or extra bits. HIOS cannot assume responsibility for these.
3. Please explain the nature of the diffi culty with the instrument in suffi cient detail.
-13-
Troubleshooting
Before you conclude the torque meter is broken, go through the following troubleshooting guide. If, by fol­lowing the instructions, you are unable to resolve the diffi culty, please contact HIOS Corporation or your HIOS dealer.
Problem Causes Resolving the problem
The display says "
There is no display
The displayed value does not return to zero
The displayed values do not stop changing
Unable to recharge the battery
LOBAT
The " reappears after battery recharging is complete
The display shows a numerical value when it shouldn't
LOBAT
"


" display







Insuffi cient battery charge
This may be due to long-term use of the meter or natural battery discharge over time.
Unable to conduct zero adjustment.
MODE
The
The battery charger plug may not be en­gaged all the way into its socket.
The connection may be to the wrong terminal.
The battery may have reached the end of its operational life.
The battery charge may be insuffi cient.
This may be due to noise occurring from the power source, etc.
switch is set to "
TRACK

Recharge the battery as explained on page 12. If this does not resolve the problem, contact HIOS or your HIOS dealer.

Turn the power switch off, conduct a full battery re­charge (6 hours, no more), then turn the power switch on and see if there is a display. If there is not, contact HIOS or your HIOS dealer.
Conduct zero adjustment.


"






PEAK
Set to "
Check that the plug is connected properly.
Connect to the battery charger socket.
If the above does not remedy the problem, contact HIOS or your HIOS dealer.
Send to HIOS for repair.
Recharge the battery for no more than 6 hours.
Press
."
RESET
switch to clear the display.
Specifi cations
Model HP-1 HP-10 HP-100
Peak range N•m 0.015 - 1.000 0.15 - 10.00
N•cm 0.15 - 10.00 1.5 - 100.0 15 - 1000
N•mm 1.5 - 100.0
Ibf•in 0.015 - 0.90 0.15 - 9.00 1.5 - 90.0
(kgf•cm) (0.15 - 10.00) (1.5 - 100.0)
Accuracy* within ±0.5% (at full scale)
Power fi ve 1.2 V NiMH cells 1,000 mA
Recharge time up to 6 hours
Running time on full charge 8 hours
(continuous operation)
Battery life About 300 chargings
Dimensions See fi gure
Weight 1.8 kg
Battery charger Runs on 100-240 V AC power (50 or 60 Hz) Output 7.3 - 8 V DC (0.5 A)
• Never exceed the maximum torque tolerance level for the instrument.
• Because battery life depends on operating conditions, this cannot be guaranteed.
-14-
External features diagram
230
30
Digital Torque Meter
PEAK
ON
TRAK
POWER
F.PEAK
MODEUNITRESET
35
Analog and data output specifi cations
MaxBZAveMinP.ON
SET
AVEM_CRM_SET
47.562.5
124
35
50
25
A display of 1000 means approximately 1 volt.
Serial data output specifi cations
1. Pin assignments
Mitutoyo MQ65-5P
65.2
12.75
95
12.2
12.7
Unit: mm
As for the dimensions, HP-1, 10 and 100 are
the same.
Note that speci cations or the external ap-
pearance of HIOS torque meters are sub­ject to change without notice.
GND: Ground
DATA: Outputs data in the for-
mat indicated below.
CK: Clock
RD: Demands data reception
④
REQ: Demands data output from exter-
nal source
- ④: Open drain
-0.3  +7 V (400 A max.) : Pull up to VDD (1.55 V)
2. Data output format
Following is an explanation of the meanings of the 13-digit data output from the torque meter.
d1 d2 d3 d4 d5 d6 d7 d8 d9 d10 d11 d12 d
Sign
Entry number
Data type
Each digit is output in four-bit binary notation from the least signifi cant bit in the sequence 2
13
Decimal point position
Measured value
0
3. Timing chart
MIN MAX UNIT
T
2  sec
0
T1 0.2 0.4 sec T2 0.2 0.4 mS T3 0.5 1 mS T4 0.2 0.4 mS
1
2
2
2
23.
-15-
China RoHS2 Table
If you are asked by China Customs, please show this table to them.
有害物质名称及含量标识格式
产品中有害物质的名称及含量
有害物質
部件名
铅(pb)汞(Hg) 镉(Cd)
电路板总成 ×
壳○○○ ○ ○ ○
充电池 ×
电源适配器 ×
-
-
-
-
本表格依据 SJ/T 11364 的规定编制。 ○: 表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下。 ×: 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
六价铬
(CR(VI))
多溴联苯
(PBB)
多溴二苯醚
(PBDE)
In addition, the China RoHS marks also is required at the product and product box. At the product, you can fi nd it at the bottom and it is marked on the product box. If you cannot fi nd the mark, please ask your distributor. In case of emergency, please cut the mark below and stick at the bottom of product and on the product box.
China RoHS mark
Note: minimum size of China RoHS mark is 5mm at outer diameter
-16-