- 2 -
Notes for use
●
Never apply the torque exceeding the maximum value displayed on the instrument. If you do so, the
instrument will break inside.
●
For correct handling of Fidaptor, refer to page 4.
●
Do not use attachments except Fidaptor and other specied attachments.
●
Do not use the instrument with a device that gives repeated shocks such as an air screwdriver or an
impact wrench. (*2)
●
When you attach Fidaptor to the socket, always x it with screws in four directions. (*3)
●
Do not loosen the screws xing the socket. (It may cause a precision error.)
●
For correct handling of the charger, refer to page 11.
●
Do not connect to the data output connector anything but the devices specied in the operation
manual.
●
When you insert or remove a connector to/from a cord, hold the connector head and check the pin
arrangement.
●
Always turn o the power after nish using the instrument.
●
Do not hit or apply a load to the display plate (acrylic plate).
●
Do not change the internal volume for calibration, etc.
●
Do not handle the detector roughly or drop it.
●
Do not use the detector in inappropriate places described below:
・
A place where water, oil and other uids may be scattered
・
A place where vibrations, dust or heat may exist
・
Outdoors and a place where electric noises may exist
・
A place with high temperatures and high humidity (preferable temperature: 15°C-35°C, preferable
humidity: 25%-65%)
・
Other places where malfunctions and functional damages may be caused
●
Do not store the instrument where the temperatures or humidity may change signicantly. If you do
so, condensation may occur inside the detector resulting in functional damage.
◎
Important
*1. Before performing zero adjustment, set the unit to the TRACK mode and check if “zero” is
displayed.
*2. If you want to measure the torque of an air screwdriver or an impact wrench, we have the HIT
series for that purpose. Please consult with us.
*3. When you attach Fidaptor to the socket, x it rmly at four points. (Refer to the external views of
the detector on pages 5 and 15.)
*4. If you turn the zero adjustment knob in the PEAK mode, the reset function will be disabled.
In that case, turn the measurement mode switch to TRACK and perform “Peak” measurement
after conrming zero is displayed. (Refer to Section 7.8 on page 6.)