4.1 Adjusting the White Balance of
the Light Source (White Balance
Adjustment Assistance Function)
4.2 Precaution
Measurement status
When a single color light enters or when
a light with large radiometric quantity of
only one color enters
.............................................48
.................................48
.................................49
.......46
5 External Control51
5.1 External Input Terminals and Signals
5.2 InternalCircuitConguration
5.3 Frequently Asked Questions about
External Input
........................................54
..............53
51
6 Application Software
(RGBLaserUtility)
6.1 Startup Procedure
Screenconguration(Maindialog)
6.2 Setting the LAN
6.3 Measured Value Capture Settings
6.4 Measurement Settings
6.5 Chromaticity xy Chart Display
6.6 Modulation Frequency Measurement
6.7 Dark Measurement
6.8 Normal Measurement
Executing the normal measurement
Measurement range optimization
Capturing the measured values
Measured values of each color
Measured values of RGB mixed light
Overall judgment display
Modulation frequency display during
measurement
10.1 Relationship between Radiometric
Quantity and Photometric Quantity
10.2 Prevention of External Noise Entry
Prevention measures against noise
entry from power supply line
10.3 LAN Settings on the Computer
....................100
....99
..100
.........101
IndexIndex1
Warranty
ii
Introduction
Thank you for purchasing the Hioki TM6102 RGB Laser Meter, TM6103 RGB Laser Luminance
Meter, or TM6104 Optical Power Meter.
To obtain maximum performance from the instrument over the long term, be sure to read this
manual carefully and keep it handy for future reference.
1
Trademarks
• Adobe and Adobe Reader are trademarks of Adobe Systems Incorporated.
• Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation
in the United States and other countries.
• CORE i5 is a registered trademark of Intel Corporation.
License agreement
• The “RGBLaserUtility” application software is included with the instrument. This software requires
a license agreement. Please use it only after reading and accepting the license agreement inside
the CD.
Precautions during shipment
Store the packaging in which the instrument was delivered, as you will need it when transporting
the instrument. (p. 93)
2
3
4
5
6
7
8
9
10
Appx.Ind.
1
Conrming Package Contents
Conrming Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during
shipping. In particular, check the accessories and connectors. If damage is evident, or if it fails to
operate according to the specications, contact your authorized Hioki distributor or reseller.
Main unit and accessories
Conrm that these contents are provided.
Model TM6102 RGB Laser Meter
Model TM6103 RGB Laser Luminance Meter
Model TM6104 Optical Power Meter
TM6102
TM6103
TM6104
Instruction Manual (this document)Light shielding cap (Expressed as “cap” in this document)
Instrument is shipped with the cap attached.
Be sure to attach the cap when the instrument is not in use.
For the TM6102 and the TM6104
Power cordLAN cable
Category 6A (CAT 6A)
Straight cable 3 m
Model Z1008 AC AdapterApplication disc (CD)
• Application software
RGBLaserUtility
• Communication Command Instruction Manual
(PDF version)
(This manual describes the communication
commands.)
• Sample program
• Software license agreement
• The latest version can be downloaded from our
website.
For the TM6103
Option
The following options are available for the instrument. Contact your authorized Hioki istributor or
reseller when ordering.
Model Z1008 AC Adapter
2
Measurement Flowchart
Measurement Flowchart
Installation
Check that no contaminant or
1
dust is adhered to the detector
window of the instrument.
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly
tested for safety prior to shipment. However, using the instrument in a way not described in this
manual may negate the provided safety features.
Before using the instrument, be certain to carefully read the following safety notes.
WARNING
• Protective gears
Wear appropriate protective gear before measuring a strong laser light.
•
With regard to the electricity supply, there are risks of an electric shock, a heat
generation, a re, and an arc ash due to a short-circuit. Individuals using an
electrical measuring instrument for the rst time should be supervised by a
technician who has experience in electrical measurement.
CAUTION
Mishandling during use could damage to the instrument. Be certain that you understand
the instructions and precautions in the manual before use.
Notations
In this document, the risk seriousness and the hazard levels are classied as follows.
DANGER
WARNING
CAUTION
IMPORTANT
Indicates an imminently hazardous situation that will result in death or serious injury
to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to
the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury
to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks
with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
*
p.
[ ]
Unless otherwise specied, “Windows” represents Windows 7, Windows 8, or Windows 10.
•
Additional information is presented below.
Reference
Menus, dialogs, buttons in a dialog, and other names on the screen and the keys are
indicated in brackets.
4
Safety Notes
Symbols afxed to the instrument
Indicates cautions and hazards.
When the symbol is printed
on the instrument, refer to a
corresponding topic in the
Instruction Manual.
Indicates DC (Direct Current).
Accuracy
We dene measurement tolerances in terms of f.s. (full scale) and rdg. (reading) values, with the
following meanings:
f.s.
rdg.
f.s. (Maximum display value)
The maximum display value. Generally, this value indicates the range currently being used.
(Reading or displayed value)
The value currently being measured and displayed on the measuring instrument.
Symbols for various standards
Indicates the Waste Electrical and Electronic
Equipment Directive (WEEE Directive) in EU
member states.
Indicates that the product conforms to
regulations set out by the EU Directive.
1
2
3
4
5
6
7
8
9
10
Appx.Ind.
5
Usage Notes
Usage Notes
• Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions.
• Ensure that your use of the product falls within the specications not only of the instrument itself,
but also of any accessories, options and other equipment being used.
• Before using the instrument, verify that it operates normally to ensure that no damage occurred
during storage or shipping. If you nd any damage, contact your authorized Hioki distributor or
reseller.
• This instrument may cause interference if used in residential areas. Such use must be avoided
unless the user takes special measures to reduce electromagnetic emissions to prevent
interference to the reception of radio and television broadcasts.
• The instrument itself does not radiate any laser light. However, when a strong
laser light is irradiated to the detector window of the sensor, a strong reected
light is generated. In this case, do not look into the detector window directly.
Doing so may adversely affect your eyes or cause visual disturbance.
The instrument can measure lasers classied into classes I to IIIB. As required,
refer to the risks during measurement of the relevant class laser and cautions
on use stated in IEC60825-1 and FDA21CFR1040.10. When handling a laser
product, always follow the caution and warning labels adhered to the laser
product, and the contents described in the instruction manual.
• Refer to the description of safe use of the laser and laser system stated in
ANSI Z136.1. Only authorized operators who have trained in operation of the
laser and laser system are allowed to perform the measurement.
• Do not input a light exceeding the maximum input level. Otherwise, accurate
measurements cannot be performed or the sensor may be damaged by
excessive energy
• When an extremely focused beam is measured, the energy density on the
sensor is excessive, causing inaccurate measurements. In addition, the sensor
may deteriorate.
• To avoid the risk of re, do not irradiate a strong laser light to the instrument
or cap, nor focus a strong laser light on the instrument or cap. In particular, do
not place any combustible materials near the instrument in the unmanned state
such as automatic measurement.
• Use only the supplied Model Z1008
range is 100 V to 240 V AC at a frequency of 50 Hz/60 Hz. To avoid electrical
hazards and damage to the instrument, do not apply voltage outside of this
range.
• To prevent an electrical shock and to maintain the safety specications of this
instrument, connect only the power cord provided to an outlet.
WARNING
.
AC Adapter. The AC adapter input voltage
6
CAUTION
• Avoid using an uninterruptible power supply (UPS) or DC/AC inverter with rectangular
wave or pseudo-sine-wave output to power the supplied AC adapter. Doing so may
damage the instrument.
• The instrument consists of precision optical components. Dropping the instrument
or subjecting it to mechanical shock may damage it. Optical components inside the
instrument may fall out of alignment if the instrument is dropped or subjected to
mechanical shock, affecting measured values.
• To avoid damage to the instrument, protect it from physical shock when transporting
and handling it. Be especially careful to avoid physical shock due to dropping it.
• Do not connect the power supply improperly. Doing so may damage the instrument’s
internal circuitry.
• If a light outside the measurement wavelength range enters, this may cause the
sensor to deteriorate. Attach the cap when the instrument is not in use.
• Attach the cap when the instrument is not used for a long time.
Installing the instrument
Installing the instrument in inappropriate locations may cause a malfunction of instrument or may
give rise to an accident. Avoid the following locations.
Usage Notes
1
2
3
4
WARNING
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to a strong electromagnetic eld or electrostatic charge
• Exposed to high quantities of dust particles
• Near induction heating systems (such as high-frequency induction heating
systems and IH cooking equipment)
• Susceptible to vibration
CAUTION
• Do not place the instrument on an unstable table or an inclined place. Dropping or
knocking down the instrument can cause injury or damage to the instrument.
• The instrument consists of precision optical components. The instrument should be
securely mounted on a jig using the screw hole in its base. Dropping the instrument or
applying an impact to it can cause the accuracy to deviate from its specication. If an
impact is applied to the instrument, it needs to be inspected.
• When orienting the instrument so that a part other than its base is facing down, x it
in place so that it cannot fall. Failure to do so may cause a re or other malfunction in
the instrument.
• The instrument is housed in a metal case and emits heat. Be sure to leave adequate
space around the instrument. Failure to do so may cause the ambient temperature to
rise, affecting measured values and potentially damaging the instrument.
• Install the instrument so that no load is applied to the detector window.
5
6
7
8
9
10
Appx.Ind.
In an emergency, unplug the power cord to kill power to the instrument. Be sure to provide enough
unobstructed space to unplug the power cord immediately.
7
Usage Notes
Securing the instrument
Securing the instrument using the tripod mounting screw holes in the rear
• Install the instrument using the tripod mounting screw holes in
the rear.
• Sufciently strengthen the surface on which the instrument is
installed so that the instrument is not unstable.
• For the TM6103, securing only the rear may cause an unstable
installation of the instrument. Therefore, be sure to secure the
instrument using the tripod mounting screws on the bottom.
Usable screw: 1/4-20UNC, hole depth 7.5 mm
Securing the instrument using the tripod mounting screw holes in the bottom
• Install the instrument using the tripod mounting screw holes in
the bottom.
Usable screw: 1/4-20UNC hole depth 7.5 mm
• A stand and others necessary for the installation work should be
prepared by the customer.
Detector window (face detecting the light)
CAUTION
• Do not touch the detector window. The instrument may fail to operate to its full level of
performance if the detector window is dirty.
• Avoid contacting the detector window with sharp objects (the tip of a pair of
tweezers, etc.) or hard surfaces. The instrument may fail to operate to its full level of
performance if the detector window is damaged.
• Never use solvents that contain benzene, alcohol, acetone, ether, ketones, thinners or
gasoline. They can deform and discolor the detector window. (p. 93)
Light shielding cylinder (TM6103)
CAUTION
When handling the TM6103, observe the following cautions to perform correct
measurement.
• Do not put any foreign object in the light shielding cylinder.
• Do not apply a large load to the light shielding cylinder.
• Do not touch the inside of the light shielding cylinder with bare hands.
• Do not damage the inside of the light shielding cylinder using a sharp object.
• Do not apply an impact to the light shielding cylinder.
• Never attempt to modify, disassemble, or repair the light shielding cylinder.
8
AC Adapter
Usage Notes
WARNING
• Turn the instrument off before connecting the AC adapter to the instrument and
to AC power.
• Use only the supplied Model Z1008 AC Adapter.
1
Before starting the external control
To avoid damage to the instrument, do not apply a voltage exceeding the rated
maximum to the external input terminals.
• Always turn both devices OFF when connecting and disconnecting an interface
connector. Otherwise, an electric shock may occur.
• To avoid an electric shock or damage to the equipment, always observe the
following precautions when connecting to the external input terminals:
• Always turn off the power to the instrument and to any devices to be
connected before making connections.
• Be careful to avoid exceeding the ratings of the external input terminals.
• Connect cables securely to the external connector. During operation, a
wire becoming dislocated and contacting another conductive object can be
serious hazard.
• Use a common ground for both the instrument and the computer. Using different
ground circuits will result in a potential difference between the instrument’s ground
and the computer’s ground. If the communications cable is connected while such a
potential difference exists, it may result in equipment malfunction or failure.
• Before connecting or disconnecting any communications cables, always turn off the
instrument and the computer. Failure to do so could result in equipment malfunction
or damage.
• To prevent equipment failure, use the recommended wire type to connect to the
external input terminals.
Recommended wire
Single strand: φ0.65 mm (AWG #22)
Multi-strand: 0.32 mm
2
2
DANGER
3
WARNING
4
5
6
CAUTION
7
8
9
(AWG #22)
Acceptable limits
Single strand: φ0.32 mm to φ0.65 mm (AWG #28 to #22)
Multi-strand: 0.08 mm
Strand diameter: minimum φ0.12 mm (per strand)
Standard insulation stripping length: 9 mm to 10 mm
Button pressing tool: Blade screwdriver (shaft diameter: 3 mm, tip width 2.6 mm)
2
to 0.32 mm2 (AWG #28 to #22)
10
Appx.Ind.
9
Usage Notes
CD precautions
• Exercise care to keep the recorded side of discs free of dirt and scratches. When writing text on a
disc’s label, use a pen or marker with a soft tip.
• Keep discs inside a protective case and do not expose to direct sunlight, high temperature, or
high humidity.
• Hioki is not liable for any issues your computer system experiences in the course of using this
disc.
10
1
Overview
1.1 Overview and Features
The instrument measures the centroid wavelength and radiometric quantity of laser light sources
and calculates the chromaticity and photometric quantity.
In addition, the target value of the radiometric quantity (each of the red, green, and blue radiometric
quantities) and the tolerance of the radiometric quantity necessary to adjust the chromaticity and
photometric quantity to specied values are presented.
All of the control and display of the measurement are performed on the computer.
Highly accurate measurement dedicated for RGB lasers
The centroid wavelengths and radiometric quantities of the red, green, and blue lasers are
measured simultaneously by means of the discrete centroid wavelength method*.
Additionally, the highly accurate photometry and colorimetry are achieved.
*: A method to measure the red, green, and blue centroid wavelengths and radiometric quantities
of the RGB laser and calculate the chromaticity and photometric quantity from the color-matching
function.
The industry’s rst traceability to national standards in laser illuminance
The industry’s rst traceability to national standards in laser illuminance was achieved using a
monochromatic laser source. (As of May, 2017)
It is Hioki’s original compliant, which renovates the conventional traceability to the standard lamp.
1
Overview
Shortening of adjustment process
When the white balance is adjusted, the target value of the radiometric quantity and the tolerance
of the radiometric quantity are calculated from the measured result to contribute to shortening of
the adjustment process.
Stable high-speed measurement
Various modulation frequencies (screen refresh rates) are supported to ensure stable
measurement.
Measurement of centroid wavelength
In addition to the chromaticity and photometric quantity, the centroid wavelength, which is used to
control the RGB laser module product itself, can be measured.
In addition, the measured centroid wavelength is utilized for the quality control or production control
of the RGB laser module.
Sensors suitable for the measuring object
Three types of sensors, illuminance (TM6102), luminance (TM6103), and optical power (TM6104),
suitable for various measuring objects such as HMD, HUD, and projector or various measurement
scenes in the production process, are prepared.
11
Model TM6102 RGB Laser Meter
1.2 Model TM6102 RGB Laser Meter
The TM6102 RGB Laser Meter measures the light that is irradiated uniformly to the entire detector
window from the RGB laser module of the projector or HUD (head up display).
The reference surface of the illuminance measurement is REF.LEVEL shown in the drawings of the
dimensions. (p. 15)
The measurement settings and measurement items except for the following points are the same as
the TM6103 and TM6104.
• Radiometric quantity → Irradiance
• Photometric quantity → Illuminance
Measurement example
12
Example: Laser projector
Parts Names and Functions
Model TM6102 RGB Laser Meter
Front
Right side
Rear
2
3
6
7
(p. 9)
1
Overview
8
(p. 6)
1
11
4
5
Bottom
9
10
10
No.DescriptionReference
Detector window
1
(φ11.3 mm ±0.1 mm)
Power LEDp. 27
2
LAN connectorp. 29
3
AC adapter connectorp. 26
4
General purpose screw hole
5
(Used to secure the power cord.)
External input terminalsp. 51
6
Communication mode switchp. 26
7
MAC address
8
Power switchp. 27
9
Tripod mounting screw holesp. 8
10
Serial No.
11
(Do not peel off the label because it is necessary for production control.)
p. 8
–
–
–
13
Model TM6102 RGB Laser Meter
C
L
±1
65
83±1
±0.5
27.3
11.3
±0.1
126
±1
Dimensions
65±1
83 ±1
27.3 ±0.5
11.3 ±0.1
φ
(Unit : mm)
14
Model TM6102 RGB Laser Meter
M3 Effective Depth 5
1.5 ±0.3
1
Overview
18.5
±0.5
C
L
1/4-20UNC Effective Depth 7.5
27.3
±0.5
C
L
122.85
±1
±0.5 82.35
37.35
±0.5
REF.LEVEL
(Measurement reference surface)
(45) 5.7 htpeD evitceffE CNU02-4/1 ×2 )5.04(
(Unit : mm)
15
TM6103 RGB Laser Luminance Meter
1.3 TM6103 RGB Laser Luminance Meter
The TM6103 RGB Laser Luminance Meter is a light shielding cylinder type luminance meter.
The TM6103 measures the luminance at a close range of the display (example: the distance from
the top end of the light shielding cylinder to the display is 10 mm).
Before starting the measurement, install the TM6103 so that it is perpendicular to the display. (To
measure the luminance more accurately, align the light axis of the light emitted from the display and
the light axis of the TM6103.)
The light source that becomes the measuring object needs to be larger than the measurement eld
diameter of the TM6103.
To perform accurate measurements, it is recommended to measure the light source that is
sufciently larger than the measurement eld diameter.
The TM6103 measures the average luminance within the measurement eld diameter in the same
way as the conventional luminance meter.
When a directional light source (light source with an inconstant luminance) is measured using
luminance meters with different angle-of-visibilities, the measured value that differs among them.
This phenomenon occurs because the calibration light source of the luminance meter has a
constant luminance area, but the measuring object does not have one.
• When the light source of the measuring object has a constant luminance area, differences in the
angle-of-visibility do not affect the measured luminance value.
• The measured luminance is a value for which the angle dependency of the light source luminance
is averaged by the angle-of-visibility. Therefore, when the light source of the measuring object
does not have a constant luminance area, the angle-of-visibility of the luminance meter affects
the measured luminance value, in principle.
To perform the measurement with excellent reproducibility, it is necessary to align the light axis
of the light emitted from the display and the light axis of the luminance meter. In this case, it is
recommended to use an optical bench.
The measurement settings and measurement items except for the following points are the same as
the TM6102 and TM6104. (p. 12)
• Radiometric quantity → Radiance
• Photometric quantity → Luminance
CAUTION
Removing or disassembling the light shielding cylinder may cause inaccurate measurements. (p. 8)
Never attempt to remove or disassemble the light shielding cylinder.
Measurement example
16
Light source (Target)
Parts Names and Functions
FrontRear
TM6103 RGB Laser Luminance Meter
Right side
(p. 6)
10
1
2
3
4
Bottom
5
(p. 9)
1
Overview
6
7
8
9
9
No.DescriptionReference
Power LEDp. 27
1
LAN connectorp. 29
2
AC adapter connectorp. 26
3
General purpose screw hole
4
(Used to secure the power cord.)
External input terminalsp. 51
5
Communication mode switchp. 26
6
MAC address–
7
Power switchp. 27
8
Tripod mounting screw holesp. 8
9
Serial No.
10
(Do not peel off the label because it is necessary for production control.)
The TM6104 Optical Power Meter measures the power and centroid wavelength of the laser beam
irradiated at the center of the detector window.
When performing the measurement, you have to consider the beam diameter.
The beam diameter is dened as a maximum distance between the points in the cross section
of the beam where the optical power density is e
(JIS C6182).
Although IEC1040 adopts e
−1
as the denition of a beam diameter instead of e−2, this instrument
uses the denition of JIS C6182.
To perform accurate measurements, it is necessary to make all powers of the beam enter the
detector window.
In addition, it is recommended to use a laser with a beam diameter of 3.1 mm or less so as to
perform accurate measurements.
This makes the diameter that is 3 times larger than the beam diameter smaller than the diameter of
the detector window (= φ11.3 mm) as a guide since the beam usually has a spread.
Example: When the beam diameter is 1 mm, make the beam enter an area within 6 mm dia. from
the center of the detector window.
−2
(= 0.1353) to the maximum value in the beam
However, this does not apply when the beam has no spread.
The measurement settings and measurement items except for the following points are the same as
Before using the instrument, verify that it operates normally to ensure that no damage occurred
during storage or shipping. If you nd any damage, contact your authorized Hioki distributor or
reseller.
Inspecting peripheral devices
Check itemAction
Using a damaged connection cable may cause an electrical
Is the sheath of any connection cable damaged or
any metal part exposed?
Inspecting the instrument
Check itemAction
shock or a short circuit accident. Do not use any damaged
connection cable.
Replace the damaged connection cable with a new one.
2
Preparations
Is the instrument damaged?If the instrument is damaged, send the instrument for repair.
Is any dust or contaminant adhered to the detector
window?
Does the power LED light up?
Execute the self-test to check if an error occurs.
(Use the
When no error occurs,
When an error occurs,
Input the modulation frequency signal to the SYNC
terminal and execute the modulation frequency
measurement. Is the measured result of the
modulation frequency within the assumed range?
TST?
command to execute the self-test.
*
PASS
is returned.
FAIL
is returned.)
If the detector window is contaminated, clean it. (p. 93)
• If the power LED does not light up, the power cord contains
a broken wire or the inside of the instrument malfunctions. If
the instrument malfunctions, send it for repair. (p. 27)
• If the power LED blinks in red, an error occurred inside the
instrument. After connecting the instrument via TCP/IP, use
ESR?,:SYSTem:ERRor?
the
*
of the error, and then take appropriate corrective actions
corresponding to the error.
Details of error: Communication Command Instruction
Manual (CD)
• When an error occurs, use the
to check the details of the error, and then take appropriate
corrective actions corresponding to the error.
• For details about the error, see the Communication Command
Instruction Manual (CD).
• When executing the self-test using the application software,
see “Self-test” (p. 79).
If the measured result of the modulation frequency is not
within the assumed range, the inside of the instrument may
malfunction. Send the instrument for repair.
command to check the details
:SYSTem:ERRor?
command
Is the measured value of the photometric quantity
changed by changing the brightness of the
irradiation light?
Is the measured result of the centroid wavelength
within the assumed range when the laser light
source with the known wavelength is measured?
If the measured value of the photometric quantity is not
changed, the inside of the instrument may malfunction. Send
the instrument for repair.
If the measured result of the centroid wavelength is not
within the assumed range, the inside of the instrument may
malfunction. Send the instrument for repair.
25
Connecting the AC Adapter and Power Cord
2.2 Connecting the AC Adapter and Power Cord
Check the following points before starting the connection work.
• Check that the cap is attached to the detector window.
• Thoroughly read “AC Adapter” (p. 9).
Rear
Connect the AC adapter to
1
the AC adapter connection
terminal of the instrument.
Connect the power cord to
2
the AC adapter.
Connect the power cord
3
to a commercial power
supply.
2.3 Setting the Communication Setting Mode
Change the LAN communication setting mode using the communication setting mode switch. (p. 31)
The LAN communication setting mode cannot be changed when the power is ON.
Rear
26
Switch
DFLT
USER
Communication
setting mode
Fixed setting
mode
User setting
mode
Settings
• The LAN settings are as follows.
IP address: 192.168.0.254
Subnet mask: 255.255.255.0
Default gateway: 0.0.0.0 (None)
Communication command port: 1024
• The xed setting mode is used to make the LAN settings.
• The xed setting mode can also be used when one-to-one communication
with the computer is performed.
Use the LAN settings that are set using the following communication
GreenLitRunning when the communication setting mode is set to the user setting mode.
OrangeLitRunning when the communication setting mode is set to the xed setting mode.
2
Preparations
An internal error is occuring.
RedBlink
OrangeBlink
‒OffThe power is off.
Use the communication command (
See: Communication Command Instruction Manual (CD)
Running in the version up mode (boot mode).
The communication setting mode becomes the xed setting mode regardless of the
communication setting mode switch state.
ESR?, :SYSTem:ERRor?
*
) to check the details.
27
Installing the Application Software
2.5 Installing the Application Software
Recommended computer operating environment
CPUCore i5, 2.7 GHz or faster
OSWindows 7/ Windows 8/ Windows 10
Memory8 GB or more
Screen displayResolution 1024
Hard diskFree capacity 100 MB or more
InterfaceLAN 100BASE-TX (TCP/IPv4 connection)
CD-ROM driveFor software installation
The application software supplied with the instrument may run slowly depending on the computer
operating environment and the type of other application software that is used at the same time. It is
recommended to run the application software in the recommended operating environment.
768 dots, 64,000 colors or more
×
Installation
Before starting the installation, exit all applications running on the computer.
Log into the computer using an account with administrative privileges such as
1
“administrator”.
Execute X:\RGBLaserUtility\Setup.msi contained in the TM6102/TM6103/TM6104
2
Application Disc (CD).
(X: is a CD-ROM drive.)
After executing setup.msi, follow the instructions that appear on the screen to proceed the installation.
After the installation has been completed and all connections have been performed,
3
start the program in one of the following ways.
• Double-click the [RGBLaserUtility] icon on the desktop.
• From the [Start] menu of Windows, click [All Programs] > [HIOKI] > [RGBLaserUtility].
Uninstallation
When the application software is no longer needed, delete it in one of the following ways.
28
• Click
[Control Panel] > [Uninstall a program], and then delete [HIOKI RGBLaserUtility].
• Click [Control Panel] > [Programs and Features], and then delete [HIOKI RGBLaserUtility].
Using a LAN
2.6 Using a LAN
The instrument is equipped with an Ethernet 100BASE-TX interface as a standard.
The instrument is connected to the network using the recommended LAN cable (p. 30), and then
the instrument can be controlled using the computer.
When a program is created and the instrument is connected to the communication command port
via TCP/IP, the instrument can be controlled using communication commands. (For details, see the
Communication Command Instruction Manual (CD).)
Network multiple instruments and computer.
Hub
Flow of preparations
Connect the LAN cable (p. 30).
1
Set the communication conditions of the instrument (p. 31).
2
Perform one-to-one connection
between the instrument and computer.
Constructing the network environment
For LAN settings on the computer, see “LAN Settings on the Computer” (p. 101).
2
Preparations
• Do not connect the instrument to the existing network.
When the instrument is started in the xed setting mode or version up mode, it uses
the xed IP address (192.168.0.254). Therefore, when there is the same IP address in
the existing network, the IP addresses are overlapped.
• Do not send more commands than necessary. Doing so may lower the communication
speed when multiple instruments are controlled.
• Accessing using a wireless LAN or router is not recommended.
Description of setting items
An address to identify individual instruments that are connected on the network.
IP address
Subnet mask
Default gateway
Communication
command port
number
Set a unique IP address that does not overlap with other instrument.
The instrument uses IP version 4. The IP address is expressed by four decimal numbers separated
by a dot (.) like “192.168.0.1”.
This setting separates the IP address into the address part showing the network and the address
part showing the instrument.
Normally, the subnet mask is expressed by four decimal numbers separated by a dot (.) like
“255.255.255.0”.
When the communicating computer and the instrument are in different networks, specify the IP
address of the device that becomes a gateway.
When performing one-to-one connection or when using no gateway, set “0.0.0.0” for the instrument.
Specify the connected TCP/IP port number for communication commands.
CAUTION
29
Using a LAN
Example of network environment construction
Example 1: Connect one computer and multiple instruments using a hub.
When a local network that is not connected externally is constructed, it is recommended to use a
private IP address shown in an example for the IP address.
Example 2: Perform one-to-one connection between the computer and instrument using the
LAN cable.
When performing one-to-one connection between the computer and instrument, you can set a
desired IP address. However, it is recommended to use a private IP address.
Computer: 192.168.0.1
IP address
Subnet mask255.255.255.0
Default gateway
Instrument: 192.168.0.2 (Set an IP address different from the computer
setting.)
Computer: No entry
Instrument: 0.0.0.0
Connecting the LAN cable
Connect the instrument and computer using the LAN cable.
If the green LED on the LAN connector is not lit even after the instrument has been connected to
the LAN, the instrument or connection device may malfunction or the LAN cable may have a broken
wire.
Rear
Recommended cable
Category 6A (CAT 6A) cable (Maximum 100 m)
• Since the instrument supports only 100BASE-TX,
the communication can be performed using at least
category 5 (CAT 5) cable. However, to ensure the high
quality communication, it is recommended to use the
Green LED
Lit: Linking
Blink: Communicating
Orange LED
Lit: 100BASE-TX
category 6A cable.
• The instrument incorporates a cable automatic
recognition function. Either the straight cable or cross
cable can be used.
30
Setting the LAN
To set the LAN of the instrument, follow the settings and steps below.
• Be sure to perform one-to-one connection with the computer.
• Set the communication setting mode to the xed setting mode.
Using a LAN
Perform one-to-one connection
between the instrument and computer.
Turn OFF the instrument.
1
Disconnect the LAN cable from the instrument and computer.
2
Start the instrument in the xed setting mode.
3
Rear
Set the communication mode switch to the xed setting mode
An example of communication commands is described in the following document.
For details about the communication commands, see the Communication
Command Instruction Manual (CD).
3.1 Various Settings
Precaution
• If you turn OFF the power during measurement or setting change, the settings might not be
saved. In this case, the backup error occurs when the instrument is started next time, and then
the settings are reset.
Before turning OFF the power, execute the following items.
1. Check that the response of
2. After checking, turn OFF the power without triggering or changing the settings.
• When using the application software supplied with the instrument, exit the application software,
and then turn OFF the instrument. If you turn OFF the power during measurement or setting
change, the backup error may occur when the instrument is started next time.
OPC?
*
is 1.
3
Measurement Condition Settings
Settings related to trigger
Trigger source (p. 60)
Bus trigger: Triggering is enabled by the communication command (
External trigger: Triggering is enabled from the TRIG terminal.
Communication command
:TRIGger:SOURce <BUS/EXTernal>
Trigger edge (p. 60)
Set to detect the rising edge of the external trigger signal, or to detect the falling edge of the
external trigger signal.
Communication command
:TRIGger:EDGE <RISE/FALL>
Trigger delay (p. 60)
Set the delay time from the trigger signal to the measurement start.
The trigger delay function does not apply to the modulation frequency measurement.
Communication command
:TRIGger:DELay <
Delay time
[sec]>
*TRG
).
35
Various Settings
Measurement mode settings
Normal measurement mode (p. 68) to (p. 78)
The RGB laser light is measured in this mode.
Dark measurement mode (p. 69)
The dark is measured in this mode.
Modulation frequency measurement mode (p. 68)
The modulation frequency is measured in this mode.
Input the modulation frequency signal to the external input terminal (SYNC terminal) to measure its
period.
Set the result of the modulation frequency measurement to the modulation frequency.
Communication command
:MODE <NORMal/DARK/PULSe>
Settings common to normal measurement and dark measurement
Modulation frequency settings (p. 61)
CAUTION
If the modulation frequency is set incorrectly, accurate measurements cannot be
performed. Set the modulation frequency that matches the blinking period of the
measuring object.
For the light source that is lit continuously, set the modulated light function to OFF.
The set value of the modulation frequency is not reected in the parameter settings inside the
instrument while the modulated light function is OFF.
When setting the modulation frequency, set the modulated light function to ON, and then set the
modulation frequency result or known modulation frequency.
The parameters such as the measurement time based on the set modulation frequency are set
inside the instrument.
Communication command
:PULSe <1/0/ON/OFF>
:PULSe:FREQuency <
Frequency
[Hz]>
36
Auto range settings (p. 71)
Set the auto range to ON or OFF for each of three sensors that are the R, G, and B sensors.
The auto range may not operate properly depending on input light conditions.
In that case, turn off the auto range and set the range properly.
Set an arbitrary range (1 to 16) for each of three sensors that are the R, G, and B sensors.
When the range is set in the auto range ON state, the auto range setting is OFF.
A larger range number increases the measurement sensitivity and a weaker light can be measured.
Various Settings
3
Communication command
:RANGe:R <
:RANGe:G <
:RANGe:B <
Range number (1 to
Range number (1 to
Range number (1 to
16)>
16)>
16)>
Normal measurement settings
Color-matching function (p. 61)
Select the color-matching function from 2° or 10°.
The chromaticity is calculated using the set color-matching function.
However, the photometric quantity is always calculated using the color-matching function with a
eld-of-view of 2° regardless of the color-matching function settings.
In addition, when a eld-of-view of 10° is set, the following measurement items are not measured.
• Correlated color temperature
• Delta uv
• NTSC ratio
Communication command
:ANGLe <2/10>
Measurement Condition Settings
Averaging (p. 63)
The averaging is performed the specied number of times.
When the number of averaging is increased, the measurement may take some time.
Number of averaging = 1 to 100 times
Communication command
:AVERaging <1
to
100>
37
Correction Functions
3.2 Correction Functions
Centroid wavelength input
mode (p. 39)
Centroid wavelength
offset correction (p. 39)
Radiometric quantity gain
correction (p. 40)
Chromaticity xy offset
correction (p. 40)
Photometric quantity gain
correction (p. 40)
The photometry is performed using the centroid wavelength set for each of R, G,
and B.
The centroid wavelength is corrected using the centroid wavelength offset value
set for each of R, G, and B.
The radiometric quantity is corrected using the radiometric quantity gain value set
for each of R, G, and B.
xy of the RGB mixed light (lights with R, G, and B wavelengths are mixed) are
corrected.
The photometric quantity of the RGB mixed light (lights with R, G, and B
wavelengths are mixed) is corrected.
The corrections of the centroid wavelength and radiometric quantity may affect the calculations of the
chromaticity and photometric quantity.
Therefore, when the chromaticity or photometric quantity is corrected in the state where the centroid
wavelength or radiometric quantity is corrected, this is double-correction.
It is recommended to apply the correction in accordance with the combinations shown in the table below.
: Recommended. – : Not recommended.
Centroid
wavelength
input mode
Centroid wavelength input
mode
Centroid wavelength offset
correction
Radiometric quantity gain
correction
Chromaticity xy offset
correction
Photometric quantity gain
correction
*: When the double-correction of the centroid wavelength input mode and centroid wavelength offset is set, the centroid
When the detection level is less than 10%, the measurement accuracy of the centroid wavelength
becomes worse. This affects the measured value of the radiometric quantity. In this case, set the
centroid wavelength beforehand when a strong laser light is measured, and then measure the
radiometric quantity.
• The settable centroid wavelength value must be only in the measurable wavelength range.
• When the centroid wavelength input mode is used, the centroid wavelength offset correction does
not reect to the measurement result.
• When the centroid wavelength input mode is set to ON, the auto range operation is not
performed. After setting the range, execute the measurement.
• Be sure to specify an appropriate range, and then perform the measurement.
3
Measured value
Measured value of centroid wavelength = Set centroid wavelength value
Set the centroid wavelength offset correction when you want to correct the measured value of the
centroid wavelength.
When the centroid wavelength offset correction is enabled, the measurement may not be performed
near the upper and lower limits of the centroid wavelength measurement range.
In this case, the centroid wavelength measurement is treated as an underow, and other measured
values are also treated as underows.
When the centroid wavelength input mode is used, the centroid wavelength offset correction does
not reect to the measurement result.
Measurement Condition Settings
Measured value
Measured value of centroid wavelength = Centroid wavelength obtained from measurement +
Offset value
Input the modulation frequency signal to the external input terminal (SYNC terminal) to measure its
period.
Set the result of the modulation frequency measurement to the modulation frequency.
Modulation frequency measurement settings
Averaging (p. 64)
The averaging is performed the specied number of times.
Number of averaging = 1 to 10 times
Communication command
:PULSe:AVERaging <1
Measurement edge (p. 64)
Set to count the rising edge of the modulation frequency, or to detect the falling edge of the
modulation frequency.
to
10>
3
Measurement Condition Settings
Communication command
:PULSe:EDGE <RISE/FALL>
Modulation frequency measurement execution and modulation
frequency settings
See: “Modulation Frequency Measurement” (p. 68)
Set the measurement mode to the modulation frequency measurement.
1
:MODE PULSe
Make the modulation frequency measurement settings.
2
:PULSe:AVERaging <1
:PULSe:EDGE <RISE/FALL>
Input the modulation frequency signal between the SYNC terminal and the ground
3
terminal.
Execute the modulation frequency measurement (apply triggering).
4
:READ?
TRG
*
Set the dark estimation.
5
:DARK:ESTimate <ON/OFF>
Set the modulation frequency.
6
:PULSe <ON/OFF>
:PULSe:FREQuency <
When the dark estimation is ON, acquire the estimation result.
7
:DARK:ESTimate:RESult?
Set the measurement mode to the dark measurement mode or normal measurement
8
mode.
:MODE <NORMal/DARK>
10>
to
Modulation frequency
[Hz]>
41
Dark Measurement
3.4 Dark Measurement
Before starting the measurement (after warming up for 30 minutes or longer), attach the cap, and
then be sure to execute the dark measurement.
When the dark measurement is performed, the sensor offset can be canceled and the correct
measured value can be acquired.
Additionally, it is recommended to perform the dark measurement every time the set value of the
modulation frequency is changed.
• After the dark measurement has been executed, the value that is corrected using the acquired
dark value is reected in the measured result. This dark value is valid until the power is turned off.
• To perform more highly accurate measurements, if the atmospheric temperature changes after
the dark measurement has been executed, execute the dark measurement again.
• When the dark measurement is not performed after the power has been turned on, the dark
measurement value at shipment (default dark value) is reected in the measurement result.
To measure a weak light, the following dark measurements are recommended.
• Increase the number of averaging of the dark measurement.
• Execute the dark measurement frequently.
Dark measurement settings
Range in which the dark measurement is performed (p. 64)
Set the measurement to be performed in all ranges or only in the xed range.
When all ranges are selected, the dark measurement is performed in all ranges.
When the xed range is selected, the dark measurement is performed only in the range selected at
this time.
However, when the auto range setting is ON, the dark measurement is performed forcibly in all
ranges regardless of the settings.
Communication command
:DARK:TYPE <ALL/FIX>
Averaging (p. 64)
The averaging is performed the specied number of times.
To measure a weak light, it is recommended to increase the number of averaging.
When the number of averaging is increased, the measurement may take some time.
Number of averaging = 1 to 100 times
Communication command
:DARK:AVERaging <1
to
100>
Dark measurement result judgment (p. 64)
Whether the dark value is correct can be judged.
When the dark measurement is performed with the judgment set to OFF, the dark measurement
succeeds in any dark measurement state. However, when the dark measurement is executed in the
light incidence state, subsequent normal measurement may not be performed correctly.
Therefore, it is recommended to set the dark measurement judgment to ON and check the
judgment after the dark measurement has been executed.
Communication command
:DARK:JUDGment <1/0/ON/OFF>
42
Dark Measurement
Dark estimation
The dark estimation is used when the modulation frequency setting is changed.
When the modulation frequency setting is changed in the dark estimation ON state, the dark value
at the new modulation frequency is calculated.
• When the modulation frequency is changed in the dark estimation ON state, be sure to check
that the dark estimation is performed correctly.
• To execute the dark estimation correctly, set the modulated light function to ON beforehand.
After that, perform the dark measurement in all ranges of all colors. (p. 64)
It is recommended to set the number of averaging of the dark measurement to 10 or more.
• The dark estimation is performed in all ranges regardless of the setting of the range in which
the dark measurement is performed.
• When the modulation frequency is changed in the modulated light function OFF state, the dark
estimation is not executed.
• The dark estimation is used only in a range of the modulation frequency ±5 Hz or less after
the dark measurement has been performed. The dark value obtained by the dark estimation
is an estimate value. Basically, it is recommended to execute the dark measurement when the
Acquire the measurement result of the dark measurement, and then check that the dark
6
value is appropriate.
to
to
100>
300.0000>
Set the measurement mode to the normal measurement mode.
7
43
Dark Measurement
44
4
The centroid wavelength, radiometric quantity, tristimulus value, chromaticity, photometric quantity,
correlated color temperature, NTSC ratio, and dominant wavelength of the laser light source that is
a measuring object are measured.
The measurement can be performed in two ways shown below.
Normal Measurement
Measurement using the
application software
supplied with the
instrument.
Measurement using
communication
commands.
Measurement itemCommand
Centroid Wavelength
1. Make preparations for measurement.
See: “2 Preparations” (p. 25)
2. Execute the measurement in accordance with the application software
RGBLaserUtility.
The instrument can be controlled using commands.
An example of communication commands is described in the following.
For details about the communication commands, see the Communication
Adjusting the White Balance of the Light Source (White Balance Adjustment Assistance Function)
4.1 Adjusting the White Balance of the Light
Source (White Balance Adjustment Assistance
Function)
The instrument provides a function that assists the work to adjust the white balance of the RGB
laser light source, which is a measuring object, to the target chromaticity and photometric quantity.
Acquiring the target value of the radiometric quantity
When inputting the target chromaticity and photometric quantity for the white balance before
starting the measurement, you can acquire not only the measured values, but also the target values
of the R, G, and B radiometric quantities. (The target value of the radiometric quantity is expressed
in the unit of the radiometric quantity.)
You can adjust the light source to the target chromaticity and photometric quantity for the white
balance only by adjusting the radiometric quantity of the RGB laser to the target value of the
radiometric quantity.
• The ne adjustment of the radiometric quantity of the RGB laser, which operated by sense while
checking the measured values of the chromaticity and photometric quantity, is no longer needed.
• Since the target value of the radiometric quantity can be acquired in the RGB laser simultaneous
incidence state, the processes such as turning on the light sources in order are not needed.
Acquiring the tolerance of the radiometric quantity
To make the white balance adjustment assistance function easier to use, the tolerance of the
radiometric quantity can be acquired in accordance with the tolerances of the chromaticity and
photometric quantity for the white balance you have input.
When inputting the tolerances of the chromaticity and photometric quantity for the white balance
before starting the measurement, you can acquire the tolerances of the R, G, and B radiometric
quantities. (The tolerance of the radiometric quantity is expressed in the unit of the radiometric
quantity.)
You can adjust the chromaticity and radiometric quantity of the light source to their tolerances
you have specied by adjusting the radiometric quantity of the RGB laser to the tolerance of the
radiometric quantity you have acquired.
You can adjust the white balance more easily by acquiring not only the target value of the
radiometric quantity, but also the tolerance of the radiometric quantity.
Execution example
Set the measurement contents. (p. 35)
1
Set the measurement range and the number of averaging.
Set the target values of the chromaticity and photometric quantity.
2
46
Adjusting the White Balance of the Light Source (White Balance Adjustment Assistance Function)
Set the tolerances of the values you have input in step 2.
Target Value of Chromaticity
Target Value of Chromaticity
Target Value of Photometric Quantity
x>,<
Tolerance of Chromaticity
y>,<
Tolerance of Chromaticity
>,<
Tolerance of
x>
y>
4
Normal Measurement
Adjust the radiometric quantity of the RGB laser light source so that it enters the
7
tolerance of the radiometric quantity.
Execute the measurement, and then check that the chromaticity xy and photometric
8
quantity are in a range of the target value ± the tolerance.
:READ?
TRG
*
:TARGet:RESult:RGB?
Even when the radiometric quantity of the RGB laser light source is adjusted to the tolerance of the
radiometric quantity, the chromaticity and photometric quantity of the light source may not enter their
tolerances. In this case, adjust the radiometric quantity of the RGB laser light source so that it gets close to
the target value of the radiometric quantity.
47
Precaution
4.2 Precaution
Measurement status
Be sure to check the measurement status during measurement.
Measurement
status
Error10
Overow8
Underow7
Excessive
input
Unbalance
Low input5
No dark4
Centroid
wavelength
input mode
Measurement
stop status
Nonmeasurement
Normal0
Measurement
status No. of
Command
An error occurred that affects the measured
value.
The light with an unmeasurable radiometric
quantity was measured.
• The light whose centroid wavelength
cannot be calculated was measured.
• The light whose radiometric quantity
cannot be calculated was measured.
9
6
3
2
1
A radiometric quantity larger than the
maximum input entered.
There was a color whose measured
radiometric quantity is less than 1/20 of the
maximum radiometric quantity measured
among three colors, R, G, and B. The
measured value related to the color with
less than 1/20 of the maximum radiometric
quantity may not be accurate.
The detection level in the range used for
the measurement was less than 10%.
The measurement can be performed, but
the measurement cannot be performed
accurately.
The following occurred in a range used for
the normal measurement.
• No dark measurement was performed.
• The modulation frequency used for the
dark measurement did not match the
modulation frequency used for the normal
measurement.
The wavelength set by the user was
calculated as the centroid wavelength.
The measurement was stopped during
measurement.
The measured value was acquired in the
non-measurement state.
The measured value was acquired in the
appropriate state.
DetailsCorrective action
The repair is required. Contact your
authorized Hioki distributor or reseller.
Change the measurement range to set
an appropriate measurement range. (If
an overow occurs in the auto range,
this light is out of the measurement
range.)
Change the measurement range to set
an appropriate measurement range. (If
an underow occurs in the auto range,
this light is out of the measurement
range.)
If the excessive input state continues
for an extended period of time, this may
cause a malfunction. Do not allow a light
larger than the maximum input to enter.
Do not use the measured value related
to the color with less than 1/20 of the
maximum radiometric quantity.
Increase the measurement range of the
target color to select a measurement
range in which the detection level is 10%
or more.
• Execute the dark measurement.
• Set the modulation frequency in the
dark estimation ON state.
‒
‒
‒
‒
48
Precaution
Measured value showing
abnormal value
1.0E+99Error status
1.0E+90Non-measurement status
1.0E+80Overow status
1.0E+70Underow status
Details
When a single color light enters or when a light with large radiometric
quantity of only one color enters
When a weak light with less than 1/20 of the strongest radiometric quantity among three colors, R, G,
and B, is measured, a weak light may not be measured properly due to the effect of the strong light
color.
At this time, the measurement status shows unbalance.
When the measurement status shows unbalance, it is recommended not to use the measured
value.
4
Normal Measurement
49
Precaution
50
5
The following items can be executed using the external input terminals on the rear of the
instrument.
After checking that external control devices are turned off, connect the instrument and external
control devices.
All signals are insulated by digital isolators.
Check the input rating and internal circuit conguration, and then understand the safety precautions.
After that, connect the control system and use it in a correct manner.
External Control
• The modulation frequency is measured.
• The instrument is controlled by inputting the measurement trigger signal from the external device.
5.1 External Input Terminals and Signals
Thoroughly read “Before starting the external control” (p. 9”) beforehand.
Connectors used
Rear
Signal layout
Signal
name
GND−GND−−
SYNCINModulation frequency signal
TRIGINExternal trigger
1
2
3
I/OFunctionLogic
Rising
edge
Rising
edge
Push in the terminal button using a tool such as a slotted
screwdriver.
With the button pushed in, insert an electric wire into the
electric wire connection hole.
Release the button.
The electric wire is then locked.
5
External Control
Falling
edge
Falling
edge
51
External Input Terminals and Signals
Input signal
When the trigger source is external, the measurement is performed once on the rising
edge or falling edge of the TRIG signal.
The trigger source can be set by the
The direction of the detection edge can be set by the
TRIG
(p. 35)
The delay from the trigger input to the measurement start can be set by the
:TRIGger:DELay
When the trigger signal is input during measurement, it is ignored.
When the trigger source is set to the communication bus, the trigger measurement is
not performed.
The period of the SYNC signal is measured during measurement of the modulation
SYNC
frequency.
Input a signal synchronized with the modulation frequency of the laser light.
Timing chart
• The level of the TRIG signal shows the voltage level.
• The following gure shows the operation when the edge of the external trigger is set to the rising
edge.
• Do not input any TRIG signal after the external trigger measurement has been started. (The input
TRIG signal is disabled.)
command. (p. 35)
:TRIGger:SOURce
:TRIGger:EDGE
command. (p. 35)
command.
Trigger monitoring startTrigger monitoring
TRIG
Internal
state
The measurement completion can be checked using communication commands.
t2t3
Each time of timing chart
ItemDescriptionTimeRemarks
t0Trigger pulse OFF time0.3 ms or longer
t1Trigger pulse ON time0.3 ms or longerRise or fall can be selected.
ONOFF
t1t0
measurement
start
t0
Measurement standbyDuring
52
Trigger judgment time +
t2
Measurement start wait time
t3Delay time0 to 1000 ms
0.3 ms to 0.9 msNo command send/receive state
In accordance with the setting of the
trigger delay. (p. 35)
5.2 Internal Circuit Conguration
Input circuit
Internal insulation 5 V
内部絶縁5V
2 k
Ω
2kΩ
10 k
Ω
10kΩ
TRIG
TRIG
4700pF
4700 pF
Internal insulation 5 V
内部絶縁5V
2 k
2kΩ
Ω
Internal insulation
内部絶縁コモン
common
Internal Circuit Conguration
Digital isolator
デジタルアイソレータ
GND
GND
Electrical specications
InputInsulationDigital isolator
Input HI2.4 V to 5 V
Input LO0 V to 0.4 V
Response pulse width300 µs or longer
Maximum applied voltage6 V
10 k
Ω
10kΩ
4700pF
4700 pF
Internal insulation common
内部絶縁コモン
5
External Control
53
Frequently Asked Questions about External Input
5.3 Frequently Asked Questions about External
Input
Frequently asked questionAnswer/Corrective action
What kind of connection method is
used to input the trigger?
What kind of connection method
is used to input the modulation
frequency synchronizing signal?
How is the input (control) checked?
Can the PLC be connected directly?
Input the HIGH level (2.4 V to 5.0 V) or LOW level (0 V to 0.4 V)
pulse signal, or the rectangular wave signal between the TRIG
terminal and the ground.
Input the HIGH level (2.4 V to 5.0 V) or LOW level (0 V to 0.4 V)
pulse signal, or the rectangular wave signal between the SYNC
terminal and the ground.
If the operation does not start even when the TRIG signal is input,
short-circuit the TRIG signal and the ground terminal instead of the
pulse signal.
(Be careful not to short-circuit the power supply.)
When the output is a relay or an open collector, the direct
connection can be performed.
(Before connecting, check that the voltage level or owing current
does not exceed the rating.)
54
Application Software
6
(RGBLaserUtility)
RGBLaserUtility is application software dedicated for the TM6102 RGB Laser Meter, TM6103 RGB
Laser Luminance Meter, and TM6104 Optical Power Meter.
RGBLaserUtility controls the TM6102, TM6103, and TM6104 to check the measurement results.
When performing more advanced control or when creating an arbitrary program, perform the control
using commands while referring to the Communication Command Instruction Manual (CD).
Notations
• Windows dialog box is expressed as a “dialog”.
• Names and keys on screens, such as menu names, dialog names, and buttons in dialogs are
enclosed by square brackets ([]).
Mouse operations
Click:Press and quickly release the left button of the mouse.
Right-click:Press and quickly release the right button of the mouse.
Double click:Quickly click the left button of the mouse twice.
Drag:
While holding down the left button of the mouse, move the mouse and then release
the left button to deposit the chosen item in the desired position.
Font sizes on the screen
RGBLaserUtility does not support high DPI. Setting to high DPI may disturb the layout. Set the DPI
as follows.
Windows versionSetting procedure
Windows 7,
Windows 8,
Windows 8.1
Windows 10
Select [Control Panel] > [Appearance and Personalization] > [Display] >
[Set custom text size (DPI)], and then change the setting to “Smaller - 100%
(default)”.
Select [Settings] > [System] > [Display] > [Change the size of text, apps, and
other items], and then change the setting to “100%(Recommended)”.
6
Application Software (RGBLaserUtility)
55
Startup Procedure
6.1 Startup Procedure
Start up the application software in one of the following ways.
See: "Installing the Application Software" (p. 28)
• Double-click the [RGBLaserUtility] icon on the desktop.
• From the [Start] menu of Windows, click [All Programs] > [HIOKI] > [RGBLaserUtility].
Screen conguration (Main dialog)
Menu bar
Operation button
Modulation
frequency value
Serial No. of
connected instrument
Measurement status
Detection level and
range settings
Overall judgment
1
4
23
5
Red radiometric quantity
target value list
Red measured
value list
Measurement counter
Measured values of
RGB mixed light
Green radiometric quantity
target value list
Green measured
value list
Blue radiometric quantity
target value list
Blue measured
value list
Check boxes and radio buttons
SettingsStatusCheck boxRadio button
Set to ON (enabled).Check on.
Set to OFF (disabled).Check off.
Operation buttons
Start Normal Measurement
1
Start Dark Measurement
2
Capture Measured Value
3
Start Modulation Frequency Measurement
4
Stop Measurement
5
56
6.2 Setting the LAN
[Communication Settings] on the menu bar > Communication Settings dialog
LAN settings of the instrument
connected from the computer.
Setting the LAN
Making the LAN settings of the instrument connected from the computer
RGBLaserUtility uses “LAN settings of the instrument connected from the computer” to perform the
TCP/IP connection to the instrument.
See: "Setting the LAN" (p. 31)
Communication test
Enter a command in the Send Data eld and click [Send]. This command is sent to the port number
of the IP address you have set above.
When a query command is sent, the received data is displayed in the Received Data eld.
Connection start
1
Send command entry eld
2
Received data display eld
3
Send
4
1
2
4
6
Application Software(RGBLaserUtility)
3
57
Measured Value Capture Settings
6.3 Measured Value Capture Settings
[Capture Settings] on the menu bar > Capture Settings dialog
Set the measured value
1
capture method.
Set the additional save
2
measurement items.
Set the save destination.
3
Set the format of the
4
save le.
Set the measured value capture method.
1
1
2
3
Time interval:
0.0 to 3600.0 sec.
* :
• The time intervals operate as follows.
“Measurement + Measured value acquisition” →
“Interval time wait” → “Measurement + Measured
value acquisition”
• Even when the time interval is set to 2.0 sec.,
the measured value may not be captured to a le
at intervals of 2.0 sec. This is because the time
interval includes the period of time necessary for
“Measurement + Measured value acquisition”.
Example:
For measurement settings for which “Measurement
+ Measured value acquisition” takes 1.0 sec., the
measured value is captured to a le at intervals of
3.0 sec.
Number of measurement
times:
1 to 1,000,000,000 times
1. Click the [Start Meas] operation button to start the
measurement.
2. Press [Capture] on the measurement screen or the
Enter key on the keyboard.
• The measured values that are displayed at this time can
1
be captured to a text le or an Excel
• The trigger source is recommended to be set to BUS. For
EXT trigger, measurement will not be executed until TRIG
terminal is input to the trigger signal.
All measured values from the measurement start to the
measurement stop are captured to a text le or an Excel
2
le.
Click the [Start Meas] operation button to start the
measurement.
• The values that have been measured the specied
number of times at specied time intervals are captured
to a text le or an Excel
• The
trigger source is recommended to be set to BUS. For
EXT trigger, measurement will not be executed until TRIG
3*
terminal is input to the trigger signal.
• The measurement counter on the main screen increments
every time the measurement is performed until the
measurement counter is reset. (p. 75) To perform
the interval measurement from count number 1 of the
measurement counter, reset the measurement counter
before starting the measurement.
®
le.
®
le.
®
58
Measured Value Capture Settings
Set the additional save measurement items.
2
All measured data displayed on the measurement screen are captured to a text le or an Excel
le. When there are additional save items, check on the relevant check boxes ().
Measurement counter
1
1
2
Date when the measured value was saved.
2
Time when the measured value was saved.
3
3
Modulation frequency measurement settings when the
4
5
6
Set the save destination.
3
Create a save le in a desired place.
1
2
4
measurement was performed.
Measured value of dominant wavelength
When the dominant wavelength saving is enabled, the measured value
5
acquisition time becomes longer than that when it is disabled. (Up to
approximately 300 ms)
Measured values of tristimulus values XYZ
6
Save le.
1
(Save to a text le.)
Specify a save le.
(Clicking
2
a desired le.)
[Browse] allows you to select
®
Save to an Excel® file.
3
3
The le is initialized to a le name using the date and time when the
software is started. However, the folder at the save destination is started in
the previously set place.
Example:
Time when the software
was started.
File name20170102083015.csv
Set the format of the save le.
4
Set the types of the decimal point and separator when setting the save destination in [Save to
File].
However, note that this format setting does not affect the numerical value displayed on the
software. The decimal point displayed on the software is xed to a period (.) despite the locale
setting.
1
2
8:30 and 15 seconds on January 2,
2017
Decimal point: Period (.), comma (,)
1
Separator: Comma (,), semicolon (;), tab, space
2
6
Application Software(RGBLaserUtility)
59
Measurement Settings
6.4 Measurement Settings
[Measurement Settings] on the menu bar > Measurement Settings dialog
Make the basic settings.
1
(1) Select the model of the instrument to be connected.
The units of the radiometric quantity and photometric quantity, and the unit of the save data are
changed in accordance with the selected model.
It is necessary that the selected model matches the model of the instrument to be connected.
If the selected model does not match the model of the instrument to be connected, the
measurement cannot be started.
Unit of RGBLaserUtility
Settings
TM6102W/m
TM6103W/srm
TM6104WlmWlm
*1: Shows the same unit as the display unit.
Radiometric
Quantity
2
2
Photometric
Quantity
lxW/m2 *
2
cd/m
Unit of the header (save item name) to be
saved to a text le or an Excel
Radiometric Quantity
1
W/sr*m2 *
1
(2) Set the trigger.
®
le
Photometric
Quantity
lx
cd/m2 *
1
1
2
3
60
Trigger source
1
Trigger edge
2
Trigger delay
3
[EXT]: External trigger, [BUS]: Communication trigger
[RISE]: Rising edge, [FALL]: Falling edge
Make the settings common to the normal measurement and dark measurement.
2
(1) Make the detailed settings.
Click the tab.
Measurement Settings
1
Setting for color-matching functions
1
Modulated light function
2
Modulation frequency
3
Sensor range settings for red laser *
4
Sensor range settings for green laser *
5
Sensor range settings for blue laser *
6
Used to check the maximum measurable radiometric quantity and measurement time in the
measurement range.
7
See: "Acquire the upper limit of measurable radiometric quantity." (p. 61)
(2) Acquire the upper limit of measurable radiometric quantity.
2
3
4
5
*: When the Auto range is selected and the settings are
6
transmitted to the instrument using RGBLaserUtility,
the initial value of the range is set to Range 1. To set
the centroid wavelength input mode to ON, select
the xed range. (When the centroid wavelength input
mode is ON, the instrument does not perform the
auto range operation.)
7
1. Click.
2. Enter the centroid wavelength
of each color.
Enter the centroid wavelength
of the laser that is a measuring
object.
3. Click.
4. Check the sending of the modulated light function and modulation frequency setting.
When [Yes] is clicked and the upper limit of measurable radiometric quantity is acquired, the [Cancel] of
(
measurement setting dialog will be invalid.)
[Yes]: Transmits the modulated light function and modulation frequency setting to acquire the upper limit of
measurable radiometric quantity and measurement time. The modulated light function and modulation
frequency setting of the instrument will change.
[No]: Stops acquiring the upper limit of measurable radiometric quantity and measurement time.
6
Application Software(RGBLaserUtility)
61
Measurement Settings
When the acquisition is completed, the upper limit of measurable radiometric quantity and
measurement time of each color are displayed.
However, when the TCP/IP connection to the instrument cannot be performed, the upper limit of
measurable radiometric quantity and measurement time cannot be checked.
The upper limit of measurable
red radiometric quantity
The upper limit of measurable radiometric quantity may exceed the maximum input. However, the radiometric
quantity to be input to the instrument should not exceed the maximum input.
The upper limit of measurable
green radiometric quantity
The upper limit of measurable
blue radiometric quantity
Measurement time (when the
number of averaging is 1) (from
sampling start to calculation end)
62
Make the normal measurement settings.
3
1
2
4
5
6
7
Measurement Settings
1. Click the tab.
2. Click the tab.
3
Average times
1
White balance adjustment assistance function settings (p. 46)
The following shows the PASS judgment
conditions of the instrument.
Each Target Value − Each Tolerance ≤ Each
Measured Value ≤ Each Target Value + Each
Tolerance
For the gure shown on the left, the PASS
judgment conditions are as follows.
900 ≤ Photometric Quantity ≤ 1100
0.331 ≤ Chromaticity x ≤ 0.335
0.330 ≤ Chromaticity y ≤ 0.336
When the centroid wavelength input mode is set
to ON, the auto range operation is not performed.
Be sure to specify an appropriate range, and then
perform the measurement.
6
Application Software(RGBLaserUtility)
Radiometric quantity gain correction settings (p. 40)
5
ON/OFF
Photometric quantity gain correction settings (p. 40)
Make the modulation frequency measurement settings.
5
[ALL]: Executes the dark measurement in all ranges.
[FIX]: Executes the dark measurement only in the currently set
range.
When this judgment is set to ON, the dark measurement failure
message appears if the dark measurement result is the error
judgment.
Click the tab.
1
2
Average times
1
[RISE]: Measures the period between the rising edge and the next rising edge of
Measurement
2
edge
the SYNC signal.
[FALL]: Measures the period between the falling edge and the next falling edge
of the SYNC signal.
64
Read the settings of the instrument.
6
Save the settings to a le.
7
Measurement Settings
When you click [Load from Instrument], the settings are read from the
instruction at the connection destination set in the [Communication
Settings] dialog, and then they are reected in the [Measurement
Settings] dialog.
If the connection to the instrument cannot be performed or if the
command send or receive has failed, the settings in the dialog are not
updated.
Read the settings from the le.
8
When you click [Save to File], the settings displayed in the
[Measurement Settings] dialog are saved to a specied le (extension
.dcm).
6
Application Software(RGBLaserUtility)
Clicking [Load to File] displays the load le selection screen.
When you select a setting le (extension .dcm), the settings saved in
this le are reected in the [Measurement Settings] dialog.
65
Measurement Settings
Reect the settings in the instrument.
9
RGBLaserUtility is started with these settings at the next startup.
1. Click [OK].2. Select whether the settings are transmitted to the instrument or
not.
(Select [Yes] or [No].)
Cancel to change the settings.
10
12
[Yes]:Transmits the settings to the instrument and closes the
[Measurement Settings] dialog.
[No]:Does not transmit the settings to the instrument and
closes the [Measurement Settings] dialog.
Clicking [Cancel] closes the [Measurement Settings] dialog.
66
Chromaticity xy Chart Display
6.5 Chromaticity xy Chart Display
Select [Tool] on the menu bar, and then check on the [Chroma Chart] check box ().
When the measurement is started, the chromaticity xy chart is displayed.
Coordinates of green chromaticity
Hiding the chromaticity xy chart
Stop the measurement.
1
Select the menu bar > [Tool], and then check off the [Chroma Chart] check box ().
2
Click [×] of the chromaticity xy chart.
3
Coordinates of RGB mixed
Coordinates of blue chromaticity
Coordinates of red chromaticity
6
Application Software(RGBLaserUtility)
light chromaticity
CAUTION
The measured value is not updated while the display size of the chromaticity xy chart is changed.
Changing the display size repeatedly may cause slow updating of the measured value.
67
Modulation Frequency Measurement
6.6 Modulation Frequency Measurement
12
3
This display is shown when the
measurement cannot be performed
correctly. (p. 78)
Start Modulation Frequency Measurement
Measures the modulation frequency to be input between the SYNC terminal and the ground.
1
When the modulated light function of the instrument is ON, RGBLaserUtility sets the measured
modulation frequency for the instrument.
Stop Measurement
2
Stops the measurement during measurement of the modulation frequency.
Modulation Frequency Measurement Result
Displays the modulation frequency value.
3
[--------] is displayed during measurement.
If correct measurements cannot be performed, [ERROR] is displayed.
68
6.7 Dark Measurement
Start Dark Measurement
1
Stop Measurement
2
Stops the measurement during dark measurement.
Dark Measurement
1
2
When the dark measurement is completed, the following is displayed.
Dark measurement succeeded:Dark measurement failed:
Click [OK] to close the dialog.The dark level is too large.
Take corrective actions such as attaching the cap,
and then perform the dark measurement again.
6
Application Software(RGBLaserUtility)
69
Normal Measurement
6.8 Normal Measurement
Executing the normal measurement
Start Normal Measurement
1
1
2
Stop Measurement
2
If the measurement start cannot be executed, check the following items.
SymptomCheck item or causeCorrective action and reference page
The communication with
the instrument could not be
performed.
The model of the instrument
set in RGBLaserUtility did not
match the instrument at the
connection destination.
The le at the save
destination could not be
opened.
The LAN settings may be incorrect.Check the LAN settings. (p. 57)
Set the model of the instrument to be
connected.
Saving to a text le.
Saving to an Excel
®
text.
"Select the model of the instrument
to be connected." (p. 60)
The le at the save destination may
already be opened.
The Excel
for some reason.
®
le could not be opened
70
Measurement range optimization
When the manual range is selected, select the measurement range of each color in which the
detection level becomes maximum.
1
2
Normal Measurement
3
4
5
Detection Level
1
Auto range settings*
Check on: Auto range is ON.
Check off: Auto range is OFF (manual range).
2
When the auto range is changed from ON to OFF in a weak light or unstable light measurement
status, the subsequent measurement may be performed in an inappropriate range. When the auto
range is set to OFF, select a range in which the detection level becomes maximum.
Measurement range that is used when the displayed measured value has been measured.*
3
Increase the range. (Increase the measurement sensitivity.)
4
Decrease the range. (Decrease the measurement sensitivity.)
5
*: The settings are sent to the instrument when the normal measurement or dark measurement is started.
6
Application Software(RGBLaserUtility)
71
Normal Measurement
Capturing the measured values
When the measurement is started, the measured values are captured and saved in accordance
with the contents of the measured value capture settings.
See: "Measured Value Capture Settings" (p. 58)
When [With Capture button or ENTER key] is selected for the measured value capture method,
the measured values displayed at this time are captured to a text le or an Excel
®
le.
There are two kinds of capture methods as described below.
• Click [Capture] (capture the measured values) during measurement.
• Press the [Enter] key on the keyboard.
Save item nameColorDetailsRemarks
Count–Measurement counter
Save or non-save can be set.Date–Measurement completion date
Time–Measurement completion time
RGB:Photometry[lx]
RGB:Delta Photometry[lx]
RGB:x
RGB:Delta-x
RGB:y
RGB:Delta-y
Target_Photometry[lx]Target value of photometric quantity
Target_xTarget value of chromaticity x
Target_yTarget value of chromaticity y
Tolerance Photometry[lx]Tolerance of photometric quantity
Measured value of photometric quantity
(RGB mixed light)
Measured value of photometric quantity
(RGB mixed light) − Target value of
photometric quantity
Measured value of chromaticity x (RGB
mixed light)
Measured value of chromaticity x (RGB
mixed light) − Target value of chromaticity
x
Measured value of chromaticity y (RGB
mixed light)
Measured value of chromaticity y (RGB
mixed light) − Target value of chromaticity
y
RGB
mixed
light
The unit varies depending on the
connected instrument.
–
The unit varies depending on the
connected instrument.
–
The unit varies depending on the
connected instrument.
–Tolerance yTolerance of chromaticity y
The unit varies depending on the
connected instrument.
Save or non-save can be set.RGB:YTristimulus value Y (RGB mixed light)
–
72
Normal Measurement
Save item nameColorDetailsRemarks
R:Status
R:Centroid Wavelength[nm]Measured value of centroid wavelength–
R:Dominant Wavelength[nm]Dominant wavelengthSave or non-save can be set.
R:Radiometry[W/m2]Radiometric quantity
R:Photometry[lx]Photometric quantity
R:xChromaticity x
R:yChromaticity y
R:XTristimulus value X
R (Red)
R:ZTristimulus value Z
R:Target(P)[W/m2]Target value of radiometric quantity
R:Target-Meas(P)[%]
R:Target Min(P)[W/m2]
R:Target Max(P)[W/m2]
R:RangeRange
R:Detection Level[%]Detection level
G:Status
G:Centroid Wavelength[nm]Measured value of centroid wavelength–
G:Dominant Wavelength[nm]Dominant wavelengthSave or non-save can be set.
G:Radiometry[W/m2]Radiometric quantity
G:Photometry[lx]Photometric quantity
G:xChromaticity x
G:yChromaticity y
G:XTristimulus value X
(Green)
G:ZTristimulus value Z
G:Target(P)[W/m2]Target value of radiometric quantity
G:Target-Meas(P)[%]
G:Target Min(P)[W/m2]
G:Target Max(P)[W/m2]
G:RangeRange
G:Detection Level[%]Detection level
Measurement status
(Measured value of radiometric quantity −
Target value of radiometric quantity)/Target
value of radiometric quantity × 100
Target lower limit value of radiometric
quantity
Target upper limit value of radiometric
quantity
Measurement status
G
(Measured value of radiometric quantity −
Target value of radiometric quantity)/Target
value of radiometric quantity × 100
Target lower limit value of radiometric
quantity
Target upper limit value of radiometric
quantity
The displayed measurement status is
saved.
The unit varies depending on the
connected instrument. (p. 61)
The unit varies depending on the
connected instrument. (p. 61)
–
Save or non-save can be set.R:YTristimulus value Y
The unit varies depending on the
connected instrument. (p. 61)
–
The unit varies depending on the
connected instrument. (p. 61)
The unit varies depending on the
connected instrument. (p. 61)
–
The displayed measurement status is
saved.
The unit varies depending on the
connected instrument. (p. 61)
The unit varies depending on the
connected instrument. (p. 61)
–
Save or non-save can be set.G:YTristimulus value Y
The unit varies depending on the
connected instrument. (p. 61)
–
The unit varies depending on the
connected instrument. (p. 61)
–
6
Application Software(RGBLaserUtility)
73
Normal Measurement
Save item nameColorDetailsRemarks
B:Status
B:Centroid Wavelength[nm]Measured value of centroid wavelength–
B:Dominant Wavelength[nm]Dominant wavelengthSave or non-save can be set.
B:Radiometry[W/m2]Radiometric quantity
B:Photometry[lx]Photometric quantity
B:xChromaticity x
B:yChromaticity y
B:XTristimulus value X
B (Blue)
B:ZTristimulus value Z
B:Target(P)[W/m2]Target value of radiometric quantity
B:Target-Meas(P)[%]
B:Target Min(P)[W/m2]
B:Target Max(P)[W/m2]
B:RangeRange
B:Detection Level[%]Detection level
Modulation Frequency[Hz]–Modulation frequency setting
Measurement status
(Measured value of radiometric quantity −
Target value of radiometric quantity)/Target
value of radiometric quantity × 100
Target lower limit value of radiometric
quantity
Target upper limit value of radiometric
quantity
The displayed measurement status is
saved.
The unit varies depending on the
connected instrument. (p. 61)
The unit varies depending on the
connected instrument. (p. 61)
–
Save or non-save can be set.B:YTristimulus value Y
The unit varies depending on the
connected instrument. (p. 61)
–
The unit varies depending on the
connected instrument. (p. 61)
The unit varies depending on the
connected instrument. (p. 61)
–
Save or non-save can be set.
When the modulated light function is OFF,
the setting is saved as “CW”.
74
Normal Measurement
Saving to a text le (p. 59)
Every time the measured value is captured, a line feed is inserted.
After the measurement has been started, a header (save item name) is added to the save le only
when the measurement counter is 0.
Before starting the measurement, be sure to check that a le to be saved using RGBLaserUtility is
not opened by other software.
If the le you want to save is opened by other software, the following symptoms may occur.
• An error occurs and the measurement cannot be started.
• The measured value cannot be saved correctly.
Saving to an Excel® le (p. 59)
The measured data is saved to the [RGB_LASER] sheet of Excel®. Every time the measured value
is captured, a line feed is inserted.
If Excel® is not started or there is no [RGB_LASER] sheet of Excel®.
Create a new book and save the data from cell A1 of the [RGB_LASER] sheet.
If there is the [RGB_LASER] sheet.
The measurement counter is 0: The data is saved beginning from the cell that is selected at the
start of the measurement.
The measurement counter is not 0: The data is saved in the cell one line below the one that
contains the previously measured data.
The measurement counter increments every time the measured value is saved to the le.
After the measurement has been started, a header (save item name) is added to the save le when
the measurement counter is 0 and Excel
CAUTION
• Do not perform any operation on Excel
operation is performed during saving, the save operation will fail and the measurement will be
canceled.
• When saving the data to an Excel
data to be saved incorrectly.
®
is not started.
®
while the data is saved to an Excel® le. If any
®
le, do not save it to merged cells. Doing so may cause the
Resetting the measurement counter
In the following cases, the measurement counter is reset to 0.
• When RGBLaserUtility is started.
• When [OK] is clicked in the Capture Settings dialog.
• When [OK] is clicked in the Measurement Settings dialog.
• When [Tool] is selected on the menu bar > [Reset Counter].
6
Application Software(RGBLaserUtility)
75
Normal Measurement
Measured values of each color
Even when all measured values of the radiometric quantities of three colors, R, G, and B, are within
the adjustment target values of the radiometric quantities, in rare cases the overall judgment may not
become the PASS judgment. In this case, perform the adjustment so that the radiometric quantity of
each color gets close to the target value until the overall judgment becomes the PASS judgment.
Before starting the white balance adjustment, check on the white balance adjustment assistance
function ([White Balance Adjustment Assistance Function]) check box (). (p. 63)
If each measured value is not a valid value, [--------] is displayed.
1
2
3
[Radiometric Quantity]: Target value of radiometric quantity
[Measurement-Target]: Difference between radiometric quantity and target value of radiometric
quantity
The calculation formula is as follows.
(Measured Value of Radiometric Quantity − Adjustment Target Value of Radiometric Quantity)/Measured
Value of Radiometric Quantity × 100
1
Adjust the output of each laser so that it gets close to 0%.
Color of [Measurement-Target]
Green: The measured value of the radiometric quantity is within the tolerance of the radiometric
quantity.
Red: The measured value of the radiometric quantity is out of the tolerance of the radiometric quantity.
White: The white balance adjustment assistance function is OFF.
Overow statusCentroid wavelength input mode status
Underow statusMeasurement stop status
Excessive input statusNormal status
Unbalance statusNon-measurement status
Low input status
Normal Measurement
Measured values of RGB mixed light
Before starting the white balance adjustment, check on the white balance adjustment assistance
function ([White Balance Adjustment Assistance Function]) check box (). (p. 63)
If each measured value is not a valid value, [--------] is displayed.
1
2
3
6
Application Software(RGBLaserUtility)
[Target]: Each target value
[Tolerance]: Each tolerance
1
[Measurement]: Each measured value
[Meas-Target]: Each measured value - each target value
[Photometric Quantity]: Photometric quantity
[x]: Chromaticity x
[y]: Chromaticity y
2*
Cell color
Green: Each measured value is within each target value ± each tolerance.
Red: Each measured value is out of each target value ± each tolerance.
White: The white balance adjustment assistance function is OFF.
[Radiometric Quantity]: Radiometric quantity
[Tcp]: Correlated color temperature
[Delta uv]: Delta uv
3
[u’]: Chromaticity u’
[v’]: Chromaticity v’
[NTSC Ratio]: NTSC ratio
*: The overall judgment is performed by the instrument. The photometric quantity and chromaticity
xy are judged by the application software and the display color is changed to the relevant color.
Therefore, in rare cases the overall judgment may not match the judgments of the photometric
quantity and chromaticity xy due to effects of the rounding error. In this case, use the overall
judgment as the judgment result.
77
Exiting the Application Software
Overall judgment display
DisplayDetails
Non-judgment status.
This status is displayed in the following cases.
• Immediately after RGBLaserUtility is started.
• The white balance adjustment assistance function is OFF.
• The measurement is started.
The photometric quantity and chromaticity xy are within their tolerances.
The photometric quantity and chromaticity xy are out of their tolerances.
• The measurement could not be started.
• The correct measurement could not be performed.
Modulation frequency display during measurement
The modulated light function is ON:The modulated light function is OFF:
The modulation frequency set on the
[CW] (Continuous Wave) is displayed.
instrument is displayed.
6.9 Exiting the Application Software
78
Click [×] at the upper right
corner of the main screen.
When the application software is started next time, it is started
with the settings (communication settings, capture settings, and
measurement settings) that are set when the software is terminated.
When the application software is terminated, the following les may
be created in the folder in which [RGBLaserUtility.exe] is located.
Do not delete these les as they are backup les of the settings.
• MeasSetData.ini
• RGBLaserUtility.ini
6.10 Others
Self-test
Select [Tool] on the menu bar > [Self Test].
When the self-test is completed, the following is displayed.
Others
The self-test is the PASS
judgment:
ErrorDetailsCorrective action
[ROM error]An operation error occurred in the internal ROM.
[Adjustment error]The adjustment value showed an abnormal value.
The self-test is the FAIL
judgment:
Errors that occurred are displayed.
Check the table shown below and take
corrective actions.
When the instrument needs to be
repaired, contact your authorized Hioki
distributor or reseller.
6
Application Software(RGBLaserUtility)
The MAC address showed an abnormal value.
[MAC address error]
[Product No. error]The written serial No. is inappropriate.
[Save memory error]
[Backup error]
[Conguration error] The conguration at startup failed.
[Measurement error] The sampling failed during measurement.
[Mode error]An inappropriate process was performed.
[RAM error]An operation error occurred in the internal RAM.The repair is required.
This may cause the communication failure to
occur.
The operation of the memory for storage of
settings was abnormal.
The instrument was started with the default
settings as the stored settings showed an
abnormal value.
The repair is required.
When the error is cleared by
restarting the instrument, there is
no problem.
The repair is required.[AD error]The AD value showed an abnormal value.
Please inform your authorized
Hioki distributor or reseller of the
details of the process you have
performed.
79
Others
Starting RGBLaserUtility by specifying the measurement setting le
(.dcm le)
Start RGBLaserUtility by dragging and dropping the measurement setting le.
Drag and drop the measurement
1
setting le (.dcm le) to
[RGBLaserUtility.exe].
sample.dcm
• This function can not be used when the measurement setting le is dragged and dropped to the shortcut.
• RGBLaserUtility is not started when the measurement setting le is dragged and dropped to the shortcut.
Start RGBLaserUtility using the shortcut.
Create a shortcut of RGBLaserUtility anywhere in your computer.
1
RGBLaserUtility.exe - shortcut
Start RGBLaserUtility.
2
Transmit the measurement setting le to the
instrument, to start RGBLaserUtility.
Right-click the shortcut you have created, and then click [Properties].
2
Enter a space and the full path to the le in the [Target:] eld, and then click [OK].
3
1
80
2
Double-click the shortcut specied in the [Target:] eld.
4
Transmit the measurement setting le to the instrument, to start RGBLaserUtility.
Adjusting the width of the measured value display list
Dragging the right end of each column can adjust the width of the relevant column in the measured
value display list on the measurement screen.
To return the width of the column to the default setting, click [Tool] on the menu bar >
[Reset Column Width].
Controlling multiple instruments
Create folders for the number of connected instruments in a folder other than [ProgramFile]
1
or [ProgramFile(x86)].
Others
Copy [RGBLaserUtility.exe] in the installation folder into each folder you have created.
2
Power on the instrument for which you set the communication settings, start
3
[RGBLaserUtility.exe] corresponding to this instrument, and then make the LAN settings.
Set the instruments one by one while referring to "Using a LAN" (p. 29).
Make the measured value capture settings and perform the measurement using
4
RGBLaserUtility.
• When controlling multiple instruments, select the text format to save the measured values.
The measured values from the multiple instruments cannot be saved to the same Excel
• The operation may become slow depending on the computer specications. If the operation is
slow, upgrade the computer specications.
®
le.
6
Application Software(RGBLaserUtility)
81
Menu List
6.11 Menu List
MenuDescription
Communication SettingsOpens the Communication Settings dialog.
Capture SettingsOpens the Measured Value Capture Settings dialog.
Measurement SettingsOpens the Measurement Settings dialog.
Chroma ChartSets to show or hide the Chromaticity xy Chart dialog.
Tool
Reset Column Width
Reset CounterResets the measurement counter to zero (0).
Self TestPerforms the self-test.
HelpDisplays the version information of RGBLaserUtility.
Returns the width of the measured value display list to the default
value.
6.12 Message List
MessageDescriptionCorrective action
Restart RGBLaserUtility.
Initializing LAN
communications failed.
The setting le contains an
error.
A connection failed.
Initializing the LAN communication failed.
The read measurement setting le contains
an inappropriate format, causing a read
failure.
The TCP/IP connection to the instrument
failed.
If the same message appears even after
RGBLaserUtility has been restarted,
RGBLaserUtility cannot be used in this
environment.
The measurement setting le in which an
error occurred cannot be used.
Check the TCP/IP settings on the computer
or instrument.
Succeeded in a
connection.
Sending a command failed.
Receiving a response
failed.
Conguring settings failed.
Succeeded in the
conguring settings.
The connected instrument
is not Model TM6102,
TM6103, or TM6104.
The connected instrument
is different from the
instrument type setting.
The TCP/IP connection to the instrument
succeeded.
Sending commands to the instrument
failed.
Receiving responses from the instrument
failed.
Setting the instrument failed.
The value set on the instrument did not
match the read value.
Setting the instrument succeeded.–
The connected instrument was not the
TM6102, TM6103, or TM6104.
The model of the instrument set in
RGBLaserUtility did not match the model of
the connected instrument.
There may be an obstruction such as a
noise in the communication environment.
There may be an obstruction such as a
noise in the communication environment.
Make the settings again.
Set the model of the instrument set in
RGBLaserUtility to the model of the
connected instrument.
–
–
82
MessageDescriptionCorrective action
The settings of
RGBLaserUtility is different
from the instrument
settings.
The measurement settings specied in the
instrument did not match the settings in
RGBLaserUtility. (Except for the auto range
settings and range settings.)
Message List
• To use the measurement settings set in
RGBLaserUtility, transmit the settings.
(p. 66)
• To use the measurement settings set
in the instrument, load the settings.
(p. 65)
Starting measurement
failed.
Opening the le failed.
Opening the Excel le
failed.
Saving the settings failed. Saving the measurement setting le failed.
Loading the settings failed. Reading the measurement setting le failed. The read le cannot be used.
Saving the measurement
data failed.
Measuring a dark value.The dark measurement is being performed.–
Succeeded in the dark
measurement.
Dark measurement failed. The dark measurement failed.
Starting the measurement failed.–
• Close the text le for storage of the
measured values if it is opened.
Opening the text le for storage of the
measured values failed.
Opening Excel
values failed.
Saving the measured data failed.
The dark measurement succeeded.–
®
for storage of the measured
• If the le name is inappropriate, enter
an appropriate le name in the “le
name.extension” format.
• If the folder name is inappropriate,
select an appropriate folder.
Install Excel
The save le may be in use. Stop using
the save le.
• Do not perform the save operation
when there is no measured data.
• Do not perform any operation on Excel
during the measurement while the data
is saved to an Excel
Perform the dark measurement with the
cap attached.
®
.
®
le.
®
6
Application Software(RGBLaserUtility)
An attempt was made to set the modulated
Are you sending the
modulation settings to the
instrument?
The le to be loaded is not
a DCM le.
The program has started
with all the default values
set.
Transferring the settings.The settings are being sent to the instrument.–
The connection was reset.
Are you sending the setting
to the instrument?
Setting the modulation
frequency failed.
Estimating a dark value
failed.
light function and modulation frequency
settings for the instrument.
This massage is displayed when the
measurable radiometric quantity is acquired.
The extension of the measurement setting
le to be read was different or the le does
not exist.
The program was started with the default
measurement settings as the read
measurement setting le, or backup
measurement settings were inappropriate.
The TCP/IP connection to the instrument was
disconnected during measurement.
Check whether the measurement settings are
transmitted to the instrument. This message
is displayed when you click [OK] in the
Measurement Settings dialog.
Setting the modulation frequency failed.Make the settings again.
The dark estimation failed.
A noise problem is assumed.
Perform the dark measurement with the
cap attached.
–
–
–
–
83
Message List
MessageDescriptionCorrective action
Cancellation is invalid.
Please click OK button.
The measurement upper limit radiometric
quantity is acquired, and the modulated light
function and modulation frequency setting is
changed. [Cancel] is invalid.
Click [OK]
.
Getting command-string
failed.
Internal error!!
An internal error occurred.
Contact your authorized Hioki distributor
or reseller.
84
7
Specications
7.1 General Specications
Operating environmentIndoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature
and humidity
Temperature 5°C to 35°C (41°F to 95°F)
Humidity80% RH or less (no condensation)
Storage temperature
and humidity
Standards
Power supply
InterfaceLAN
Dimensions
Mass
Product warranty period3 year
AccessoriesSee: “Main unit and accessories” (p. 2)
OptionSee: “Option” (p. 2)
Temperature −5°C to 45°C (23°F to 113°F)
HumidityLess than 35°C (95°F), 80% RH or less (with no condensation)
35°C (95°F) or higher and less than 40°C (104°F), 55% RH or less
(no condensation)
40°C (104°F) or higher and less than 45°C (113°F), 45% RH or less
(no condensation)
SafetyEN 61010
EMCEN61326 Class A
Model Z1008 AC Adapter (12 V, 1.25 A)
Rated supply voltage: 100 V AC to 240 V AC (Voltage uctuations of ±10% from the
rated supply voltage are taken into account.)
Rated supply frequency: 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
Maximum rated power: 9.5 VA (including the AC adapter), 1.9 VA (instrument only)
Range: 16-range conguration (For each color of RGB)
DefaultAuto range: ON (All colors)
Range: Minimum sensitivity range
The calculated maximum measurable power of each range (16 ranges) is
available using communication commands.
frequency measurement
DefaultNormal measurement
88
Functional specications
Normal
measurement
AveragingSettings1 to 100
Default1
Setting for color-matching
functions
CorrectionCentroid
wavelength
input mode
Centroid
wavelength
offset
Radiometric
quantity gain
Chromaticity
xy offset
Photometric
quantity gain
White balance adjustment
assistance function
Measured value • Centroid wavelength (R, G, B)
Settings2°/10°
Default2°
SettingsUser setting enabled: ON/OFF (Settable for each color of
RGB)
User setting centroid wavelength: Measurement centroid
wavelength range (Settable for each color of RGB)
DefaultEnabled: OFF (All colors)
Set centroid wavelength: 473 nm, 532 nm, 632.8 nm
SettingsOffset enabled: ON/OFF
Offset: −2.00 nm to 2.00 nm (Settable for each color of RGB)
DefaultOffset enabled: OFF
Offset: 0.00 nm (All colors)
SettingsGain enabled: ON/OFF
Gain: 1.00000e-3 to 1.00000e+3 (Settable for each color of
RGB)
DefaultGain enabled: OFF
Gain value: 1.00 (All colors)
SettingsOffset enabled: ON/OFF
Offset: −1.0000 to 1.0000 (Settable for xy of RGB mixed light)
DefaultOffset enabled: OFF
Offset: 0.0000 (xy)
SettingsGain enabled: ON/OFF
Gain: 1.00000e-3 to 1.00000e+3 (Settable only for RGB
mixed light)
DefaultGain enabled: OFF
Gain: 1.00
SettingsFunction enabled: ON/OFF
Target value of photometric quantity: 0.00000 to 3.00000e+8
Tolerance of photometric quantity: 0.00000 to 3.00000e+8
Target value of chromaticity xy: 0.00000 to 1.00000 (Set for
each of x and y.)
Tolerance of chromaticity xy: 0.00000 to 1.00000 (Set for
each of x and y.)
DefaultFunction enabled: OFF
Target value of photometric quantity: 0.00000
Tolerance of photometric quantity: 0.00000
Target value of chromaticity xy: 0.00000 (Set for each of x
and y.)
Tolerance of chromaticity xy: 0.00000 (Set for each of x and y.)
• Radiometric quantity (R, G, B, RGB mixed light)
The radiometric quantity differs from the irradiance/radiance/radiant ux
(optical power) depending on the type.
• Photometric quantity (R, G, B, RGB mixed light)
The photometric quantity differs from the irradiance/radiance/radiant ux
Measurement rangeAll-color range 1 (Minimum sensitivity range)
Measurement modeNormal measurement
Normal
measurement
EdgeRise
Delay time0.0 s
Averaging1
Centroid wavelength
input mode
Centroid wavelength
offset
Radiometric quantity
gain
Chromaticity xy offset
EnabledAll colors OFF
Set centroid wavelength
EnabledOFF
OffsetAll colors 0.0
EnabledOFF
GainAll colors 1.0
EnabledOFF
Offset0.0
Red: 632.8 nm, Green: 532 nm, Blue: 473
nm
Photometric quantity
gain
White balance
adjustment assistance
function
Measurement result judgmentON
Dark measurement
Modulation
frequency
measurement
Setting for color-matching functions2°
LANUser setting mode
Averaging1
Dark measurement rangeAll range measurement
Dark estimationOFF
Averaging1
Count edgeRise
EnabledOFF
Gain1.0
EnabledOFF
Target value
(Photometric quantity,
chromaticity xy)
Tolerance range
(Photometric quantity,
chromaticity xy)
IP address0.0.0.0 (Disabled)
Subnet mask255.255.255.0
Default gateway
All parameters 0.0
All parameters 0.0
0.0.0.0 (None)
92
Communication
command port
1024
8
Maintenance and Service
WARNING
Customers are not allowed to modify, disassemble, or repair the instrument.
Doing so may cause incorrect measurement results.
Calibrations
• The calibration period varies depending on the status of the instrument or installation
environment. We recommend that the calibration period be determined in accordance with the
status of the instrument or installation environment. Please contact your Hioki distributor to have
your instrument periodically calibrated.
• We do not support calibration of the TM6103 only. If you request calibration of the TM6103, the
calibration is performed with adjustment. Please understand this point in advance.
Backing up the data
The instrument may be initialized (returned to the factory default settings) when it is repaired or
calibrated.
Before you ask for repair or calibration, it is recommended to back up (save or record) the
measurement conditions and measured data.
Precautions during shipment
• To ensure safe handling, when transporting the instrument, please use the original box and
packing materials, but do not use if the box is damaged or warped, or if the packing materials
are in poor condition or incomplete. If the original box or packing materials provided with the
instrument cannot be used, contact your authorized Hioki distributor or reseller. We will send
appropriate box or packing materials.
• When packing the instrument, make sure to disconnect any cords including the power supply
cord from the main device.
• When transporting, avoid dropping or other excessive impacts.
• When sending the instrument for repair, include details of the problem.
• To prevent dew condensation during transportation, pack the instrument after putting it in a bag
that can be sealed. When moisture absorbents such as silica gel can be prepared, pack them
together with the instrument.
• Note that if the instrument is damaged due to transportation without using the original packing
boxes and cushioning materials, the repair cost will be charged even within the warranty period.
Cleaning
• Never use solvents that contain benzene, alcohol, acetone, ether, ketones, thinners or gasoline.
They can deform and discolor the instrument.
• To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent.
• If the detector window is contaminated, wipe it gently with a lint-free cloth such as lens cleaning
paper.
• If any ber remains on the detector window, blow it out with a blower for optical lenses.
8
Maintenance and Service
Disposal
Handle and dispose of the instrument in accordance with local regulations.
93
Troubleshooting
8.1 Troubleshooting
If damage is suspected, check the following before contacting your authorized Hioki distributor or
reseller.
Before sending the instrument for repair
SymptomCheck item or causeCorrective action and reference page
The power LED does not
light up even when the
power switch is turned on.
The power LED blinks in red
when the communication is
performed.
The measured value of
the centroid wavelength is
always the same value.
Even when the auto range
setting is set to ON, the auto
range does not operate.
• Is the power cord disconnected?
• Is the power cord connected
correctly?
An error occurs as the sent
commands include an improper
command.
Is the centroid wavelength input
mode set to ON?
Is the modulated light function set to
ON?
Is the detector window
contaminated?
Check that the power cord is connected
correctly
Perform one of the following.
• Send the commands one by one to
• After sending the commands, send
Details about
Command Instruction Manual (CD)
Check the setting of the centroid
wavelength input mode. (p. 39)
Set the modulated light function to ON,
and then set the modulation frequency
(p. 41)
Clean the detector window. (p. 93)
. (p. 26)
check for the command that causes the
power LED to blink in red.
ESR?
to check the error occurrence
*
location.
ESR?
: Communication
*
.
The measured value varies.
The measurement cannot
be performed correctly.
Is the instrument used in a noisy
environment?
Was the dark measurement
performed appropriately?
Is an unintentional correction
function enabled?
If the instrument is used in a noisy
environment, investigate the corrective
actions shown below.
• Take the noise prevention measures.
(p. 100)
• Keep the instrument away from noise
generation sources (motor
electromagnetic switch, electric power
line, or device in which sparks occur),
or perform the measurement in another
room.
• Get the power supply from an outlet that
is installed securely
• Get the power supply from a power
supply line dif
generates noises.
Enable the judgment function of the dark
measurement, and then perform the dark
measurement. (p. 42)
Check the settings of each correction
function. (p. 38)
ferent from a device that
, inverter,
.
94
Troubleshooting
SymptomCheck item or causeCorrective action and reference page
The communication cannot
be performed via LAN.
The TCP/IP connection of
the LAN is disconnected.
The LAN communication
may become slow.
Does the computer support
100BASE?
Are the communication settings
correct?
• Are the computer settings correct?
• Are the instrument settings
correct?
Is the TCP/IP connection process
performed before the instrument is
started?
Is there a broken wire in the LAN
cable or a connection failure of the
connector?
Is the instrument used in a noisy
environment?
Are the TcpNoDelay settings
correct?
Is the instrument used in a noisy
environment?
Check that the auto negotiation is enabled
in the network adapter settings of the
computer and that the 100 M link is set.
Check the communication setting mode
switch. (p. 26)
Check the network environment and the
LAN settings between the computer and
instrument. (p. 29 to p. 34)
Perform the TCP/IP connection process
approximately 5 sec. after the instrument
has been started.
Prepare a new LAN cable equivalent to
the cable supplied with the product, and
check the TCP/IP connection again.
See the corrective actions stated in “The
measured value varies.” (p. 94) shown
above.
See “Cautions on Communication
Program Creation” stated in the
Communication Command Instruction
Manual (CD).
See the corrective actions stated in “The
measured value varies.” (p. 94) shown
above.
Error display
See “Self-test” (p. 79) or
Instruction Manual (CD).
ESR?,:SYSTem:ERRor?
*
stated in the Communication Command
8
Maintenance and Service
95
Troubleshooting
96
9
The instrument uses a lwIP open source.
lwIP’s License
lwIP is licensed under the BSD license:
Copyright (c) 2001-2004 Swedish Institute of Computer Science.
All rights reserved.
Redistribution and use in source and binary forms, with or without modication, are
permitted provided that the following conditions are met:
1. Redistributions of source code must retain the above copyright notice, this list of conditions and
2. 2. Redistributions in binary form must reproduce the above copyright notice, this list of
3. 3. The name of the author may not be used to endorse or promote products derived from this
THIS SOFTWARE IS PROVIDED BY THE AUTHOR ``AS IS AND ANY EXPRESS
OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE
ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT,
INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES
(INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR
SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION)
HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT,
STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING
IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE
POSSIBILITY OF SUCH DAMAGE.
License Information
the following disclaimer.
conditions and the following disclaimer in the documentation and/or other materials provided
with the distribution.
software without specic prior written permission.
License Information
97
9
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.