Hioki SS7012 Instruction Manual

Page 1
SS7012
Instruction Manual
DC SIGNAL SOURCE
Visit us at www.TestEquipmentDepot.com
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Mar. 2015 Revised edition 8 SS7012A981-08 15-03H
EN
* 6 0 0 2 8 7 5 1 8 *
Page 2
Contents
Introduction .......................................................................... 1
Confirming Package Contents ............................................. 2
Safety Information ................................................................ 3
Operating Precautions ......................................................... 7
Chapter 1 Overview 13
1.1 Product Overview .............................................. 13
1.2 Features ............................................................14
1.3 Names and Functions of Parts .......................... 15
1.4 Sourcing and Measurement Flowchart ............. 23
Chapter 2 Sourcing, Measurement and
Loop test Preparation 25
2.1 Supplying Power ............................................... 25
Installing or Replacing the Batteries .................. 25
Connecting the AC Adapter (Option) ................ 27
2.2 Connecting the Input Cord and the Test Lead .. 28
2.3 Turning the Power On and Off .......................... 33
2.4 Pre-Operation Inspection ..................................34
Chapter 3 Sourcing 37
3.1 Sourcing Example ............................................. 38
3.2 Sourcing Constant Voltage and
Constant Current ...............................................39
3.3 Sourcing Thermoelectromotive Force ............... 43
3.4 0°C Check of the
100 Ω Resistance Thermometer ....................... 50
3.5 Output Monitor Function .................................... 51
3.6 Memory Source Function .................................. 54
Saving setting value to memory ........................ 55
Recall sourcing .................................................. 59
SS7012A981-08
i
Page 3
ii
Scan sourcing ....................................................61
Initializing setting values ....................................63
3.7 When setting value is flashing ...........................65
Chapter 4 Measurement 67
4.1 Measurement Example ......................................67
4.2 Voltage Measurement and
Current Measurement ........................................68
4.3 Temperature Measurement ...............................71
Chapter 5 Loop Test 75
5.1 Loop Test Overview ...........................................75
5.2 Testing a Distributor (4-20 mA) .........................76
5.3 Testing a two-wire transmitter ...........................79
Chapter 6 Other Function 81
6.1 USB Communication Function ..........................81
Chapter 7 Specifications 83
7.1 General Specifications .......................................83
7.2 Accuracy ............................................................86
Chapter 8 Maintenance and Service 91
8.1 Troubleshooting .................................................91
8.2 Replacing the Circuit Protection Fuse ...............94
8.3 Cleaning ............................................................96
8.4 Error Indication ..................................................97
8.5 Other Indication .................................................97
8.6 When calibrating the instrument ........................98
Index Index 1
Page 4

Introduction

Introduction
Thank you for purchasing the HIOKI Model SS7012 DC Signal Source. To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future refer­ence.
1
Page 5

Confirming Package Contents

Confirm that these contents are provided.
SS7012 DC Signal Source (1)
Instruction Manual (1)
9168 Input Cord (1 set)
LR6 alkaline batteries (4)
L9170-10 Test lead (1 set)
Spare fuse (1)
2
Confirming Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Page 6

Safety Information

Options
9184 Temperature Probe (reference junction compensation) 9380 Carrying Case (Holds main unit only, soft case) 9782 Carrying Case (Holds options, hard case) 9445-02 AC Adapter (For Japan, US, and Canada) 9445-03 AC Adapter (For EU) SS9000 Communication Package
(Includes USB cable and USB driver software)
Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Using the instrument in a way not described in this manual may negate the provided safety features. Be certain that you understand the instructions and pre­cautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from instrument defects.
3
Page 7
Safety Information
4
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
Safety Symbols
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative impor­tance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in seri­ous injury or death to the user. Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user. Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument. Indicates advisory items related to perfor­mance or correct operation of the instru­ment.
Page 8
Symbols for Various Standards
WEEE marking: This symbol indicates that the electrical and electronic appliance is put on the EU market after August 13, 2005, and producers of the Member States are required to display it on the appliance under Article 11.2 of Directive 2002/ 96/EC (WEEE).
This is a recycle mark established under the Resource Recycling Promotion Law (only for
Ni-MH
Japan).
This symbol indicates that the product conforms to regulations set out by the EC Directive.
Other Symbols
Indicates a prohibited action.
Safety Information
5
(p. #)
[ ] The names of setting objects are indicated by
OUTPUT
(Bold char-
acters)
The screen of this instrument displays characters in the follow­ing manner.
Indicates the location of reference information.
square brackets [ ].
Bold alphanumeric characters in this manual indi­cate key labels.
Page 9
Safety Information
6
Measurement categories
To ensure safe operation of measurement instruments IEC 61010 establishes safety standards for various electrical envi­ronments, categorized as CAT II to CAT IV, and called measure­ment categories.
Primary electrical circuits in equipment connected to an AC
CAT II
CAT III
CAT IV
Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instru­ment is rated could result in a severe accident, and must be carefully avoided. Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement applications could result in a severe accident, and must be carefully avoided.
electrical outlet by a power cord (portable tools, household appliances, etc.) CAT II covers directly measuring electrical outlet recepta­cles.
Primary electrical circuits of heavy equipment (fixed instal­lations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Page 10

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Preliminary Checks
Before using the instrument for the first time, verify that it oper­ates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Instrument Installation
Operating temperature and humidity: 0 to 40°C at 80%RH or less (non-condensating) Temperature and humidity range for guaranteed accuracy: 23±5°C, 80%RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high tem­perature
Exposed to liquids Exposed to high hu­midity or condensa­tion
Exposed to high lev­els of particulate dust
In the presence of cor­rosive or explosive gases
Exposed to strong electromagnetic fields Near electromagnetic radiators
Near electromagnetic radiators (e.g., high­frequency induction heating systems and IH cooking utensils)
7
Subject to vibration
Correct measurement may be impossible in the pres­ence of strong magnetic fields, such as near transform­ers and high-current conductors, or in the presence of strong electromagnetic fields such as near radio trans­mitters.
Page 11
Operating Precautions
8
Handling the Instrument
The maximum rated voltage between input terminals and ground is 30 VAC/ 60 VDC. Attempting to measure voltages exceeding 30 VAC/ 60 VDC with respect to ground could damage the instrument and result in personal injury.
• This instrument is designed for safe operation only at low voltage. Ensure that no more than 60 V is present between any terminal and ground, and that the potential difference between any two terminals does not exceed 60 V. The instrument cannot be guaranteed in an overrange situation, such as due to an inadvertent electric shock or faulty insulation, and we cannot accept responsibility for any consequences that might occur from misuse.
• The output terminals and the voltage and current input terminals (terminals for the standard resistor) are all iso­lated. Before connecting the terminals, ensure that the target device will not provide excessive output to, or require excessive input from, the terminals. The instru­ment cannot be guaranteed in an overrange situation, such as due to an inadvertent electric shock or faulty insulation, and we cannot accept responsibility for any consequences that might occur from misuse.
• Ensure that the input does not exceed the maximum input voltage or current to avoid instrument damage, short-cir­cuiting and electric shock resulting from heat building.
• Do not allow the instrument to get wet, and do not take measurements with wet hands. This may cause an elec­tric shock.
• The case is not hermetically sealed for protection against explosion, so do not use in a flammable atmosphere.
Page 12
Operating Precautions
• When the power is turned off, do not apply voltage to the volt­age input terminals. Doing so may damage the instrument.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• If this instrument has been vibrated or impacted, restart it before use.
• Before sourcing or measurement, make sure of the function position. If current or voltage in excess of sourcing or mea­surement range is input, the instrument may be damaged.
• Note that the instrument may be damaged if the applied volt­age or current exceeds the measurement range.
• If output is impossible when is displayed which indicates outputting, set to with OUTPUT ON/OFF key and set output to ON again.
• To avoid corrosion and damage to this instrument from battery leakage, remove the batteries from the instrument if it is to be stored for a long time.
To avoid problems with battery operation, remove the batteries from the instrument if it is to be stored for a long time.
9
Page 13
Operating Precautions
10
Handling the Cords
• Before using the instrument, make sure that the insula­tion on the test lead and input cord is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki representative for repair.
• Use the 9168 Input Cord with DC 28 V or less. If this level is exceeded, electric shock may occur.
• Always turn both devices OFF when connecting and dis­connecting an interface connector. Otherwise, an electric shock accident may occur.
• To avoid breaking the test lead and input cord, do not bend or pull them.
• For safety reasons, when taking measurements, only use the L9170-10 Test lead provided with the instrument.
Page 14
Operating Precautions
11
Handling the AC Adapter and the Nickel Hydride Batteries
• Turn the instrument off before connecting the AC adapter to the instrument and to AC power.
• Use only the supplied Model 9445-02 AC Adapter or Model 9445-03 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
The nickel hydride battery is subject to self-discharge. Be sure to charge the nickel hydride batteries before ini­tial use. If the battery capacity remains very low after correct recharging, the useful battery life is at an end.
Page 15

1.1 Product Overview

Two-wire transmitter and distributor tests (4-20mA) can be easily performed. (Loop tests)
Source waveforms can be freely programmed through communica­tion with a computer.
13

Overview Chapter 1

1.1 Product Overview
This instrument is a compact DC signal source (calibrator). It can perform maintenance or loop tests of instrumentation sys­tems and can be operated from a computer via a USB cable. Furthermore, it can calibrate thermometers and measure DC voltage and DC current.
Page 16

1.2 Features

14
1.2 Features
Source function (p.39) (p.43)
Sourcing DC voltage from -25.000 V to +25.000 V (minimum resolution 100 μV), and DC current from -25.000 mA to
25.000 mA (minimum resolution 1 μA). Sourcing thermoelec­tromotive force by setting temperature (thermocouples: K, E, J, T, R, S, B, N; JIS C1602-1995, IEC 584).
Measure function (p.68) (p.71)
Measurement of DC voltage from -28.000 V to +28.000 V (minimum resolution 100 μV), and of DC current from -
28.000 mA to 28.000 mA (minimum resolution 1 μA). Tem­perature measurement from -25°C to 80°C with the optional Model 9184 Temperature Probe.
Loop test (p.75)
The output of this instrument is bipolar and it can sink cur­rent. Therefore, loop test in the instrumentation system is possible.
Output monitor function (p.51)
This function monitors the applied current of the constant voltage source. This function monitors the applied voltage of the constant current source. This function monitors the reference junction temperature of the thermoelectromotive force source.
Memory source function (p.54)
Up to 20 units of source data (settings) for each function can be saved to memory. Settings saved to memory can be sourced in recall source mode or scan source mode.
AC power is not required for operation
Common LR6 alkaline batteries or nickel hydride batteries support use in situations where AC power is not available.
Continuous long-term operation
The optional Model 9445-02 AC Adapter or 9445-03 AC Adapter allows continuous long-term operation.
Page 17

1.3 Names and Functions of Parts

Front panel
Display (p.19)
Operation Keys (p.21)
Ter m ina ls (p.18)
1.3 Names and Functions of Parts
15
Page 18
1.3 Names and Functions of Parts
Right side
RJ sensor terminal
Connect the optional Model 9184 Temperature Probe (reference junction compensation). (p.30) (p.71)
USB port
The SS7012 can communicate with a computer via a USB cable. (p.81)
AC adapter terminal
Connect the optional Model 9445-02 AC Adapter or 9445-03 AC Adapter.The input voltage range is 9 V DC±10% and the adapter has a positive center pin. (p.27)
Power switch
Turns the power for the instru­ment on or off. (p.33)
: Power switch OFF
: Power switch ON
16
Page 19
Rear panel
Battery cover
Open this battery cover to install the LR6 alkaline batteries or the nickel hydride bat­teries. (p.25)
1.3 Names and Functions of Parts
17
Page 20
18
Termi nals
Voltage input terminal
Inputs voltage for voltage measurement. Input resis­tance is about 1 MΩ.
Current input terminal/ Terminal for the standard resistor
Current input
Inputs current for current measurement function. The SS7012 has built-in input protection fuse. The input resistance of current detection and input protection fuse is about 30 Ω.
Standard resistor
When current measurement function is off, a 100 Ω standard resistor is connected for the 0°C check of the 100 Ω standard resistance thermometer.
Output terminal
When output is on, the setting value will be output. This termi­nal is shorted with about 50 kΩ resistance when output is off. The SS7012 has built-in output protection fuse.
1.3 Names and Functions of Parts
Page 21
1.3 Names and Functions of Parts
Display
Various values are displayed, including functions, source values, mea­surement values, monitor values, thermocouples, polarity, units, output status and warnings.
Upper values
Lower values
Upper triangles
Lower triangles
Right side triangles
19
SOURCE ON OFF
MEASURE
K
EJTRSBN
MONITOR ADJ RMT
C mA
C mA
V
V
Indicates values and settings related to source func­tions.
Indicates that the setting for the source function is being output.
Indicates that the setting for the source function is not being output.
K/ E/ J/ T/ R/
S/ B/ N
Indicates the specified thermocouple type.
Indicates values and settings related to measure functions.
Indicates that the output monitor function has been set. (p.51)
This is the battery indicator (Battery low warning). When this indicator starts flashing, either exchange the batteries or connect the AC adapter. When this indicator is flashing, output cannot be set to ON. (If the SS7012 is outputting when the indica-
Flashing of
setting value
tor starts flashing, output continues until the output is set to OFF. Furthermore, the accuracy of measurement values cannot be guaranteed.
The setting value flash when it is not being output. (p.65) If a setting value is flashing, press the
OUTPUT key to set the output to OFF.
Page 22
1.3 Names and Functions of Parts
Upper triangles
Lower triangles
Right side triangles
20
Indicates that the SS7012 is in remote control mode. The instrument cannot be operated by the keys. Use the LOCAL key to enable key operation. If the USB cable is removed when the is dis­played, the indicator does not turn off.
Indicates the specified source function.
Upper values Displays setting value for the source functions.
Lower values
Displays the measurement values of the measure function and the address and sourcing time interval of the memory source function.
Indicates the specified measure function.
Indicates the specified source mode.
Off : Normal source mode RECALL : Recall source mode SCAN : Scan source mode
Page 23
1.3 Names and Functions of Parts
Operation Keys
21
+/- key
0 ADJ/ LOCAL key
MONITOR/ SKIP key
MEMORY/ ENTER key
MEASURE FUNCTION
key
Selects the polarity of the output setting value.
1. Possible to return the display to "00000" in voltage measurement func­tion ([V:2.5V], [V:25V]), or current mea- surement function ([A: 25mA]).
2. If pushed while is displayed, the
indicator goes out and local con-
trol is possible.
1. Toggles the measure function and out­put monitor function. (p.51)
2. When writing to memory in the memory source function, this key sets a skip setting. (p.55)
3. When pushed simultaneously with the
MEMORY/ENTER key, toggles the
source function and the memory write mode of the source function. (p.55)
1. Toggles between the normal source, recall source, and scan source. (p.59) (p.61)
2. When writing to memory in the memory source function, this key confirms the setting value. (p.55)
3. When pushed simultaneously with the
MONITOR/SKIP key, toggles the
source function and the memory write mode of the source function. (p.55)
Sets the measure function for the instru­ment.
Page 24
1.3 Names and Functions of Parts
22
TC TYPE key
SOURCE FUNCTION
key
Sets the thermocouple type in the thermo­electromotive force source function ([TC:0°C], [TC:RJ]).
Sets the source function for the instru­ment.
OUTPUT ON/OFF key
/ key
ADDRESS
key
INTERVAL
key
Switches the output on or off in the source function. This key can be used as the start/stop for memory source function.
The digit increases or decreases by one count each time the key is pressed. Press and hold the key to change the digit contin­uously.
These are the two / keys on the left of the arrow key group. In the memory source function, these keys specify the address.
These are the two / keys on the right of the arrow key group. In the memory source function, these keys specify the sourcing time interval.
Page 25

1.4 Sourcing and Measurement Flowchart

1. Inspections before switching on the power and using the instrument
Connecting the AC Adapter (p.27) Installing the Batteries (p.25) Turning the Power On (p.33) Pre-Operation Inspection (p.34)
2. Sourcing
Constant voltage (p.39) Constant current (p.39) Thermoelectromotive force (p.43) Standard register (100 Ω) (p.50) Output monitor function (p.51) Memory source function (p.54)
2. Measurement
Voltage (p.68) Current (p.68) Temperature (p.71)
3. Turning the Power Off (p.33)
2. Loop test (p.75)
or
or
1.4 Sourcing and Measurement Flowchart
23
Page 26

2.1 Supplying Power

25

Sourcing, Measurement and Loop test Preparation Chapter 2

2.1 Supplying Power

Installing or Replacing the Batteries

• Use the common LR6 alkaline batteries or the common nickel hydride batteries.
• To avoid electric shock, turn off the power switch and dis­connect the test leads or input cords from the target device before replacing the batteries.
• After replacing the batteries, replace the cover before using the instrument.
• Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity dur­ing installation. Otherwise, poor performance or damage from battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit, disassemble or dispose of in fire. Do not recharge alka­line batteries. Handle and dispose of batteries in accor­dance with local regulations.
Page 27
2.1 Supplying Power
26
• Do not use the LR6 alkaline batteries together with the nickel hydride batteries.
• To avoid corrosion and damage to this instrument from battery leakage, remove the batteries from the instrument if it is to be stored for a long time.
• The indicator flashes when the remaining battery capacity is low. In this case, the instrument's reliability is not guaranteed. Replace the battery immediately.
• When the nickel hydride batteries have not been used for a long time and are not repeat charge and dis­charge, the device may not operate normally. (This may also occur immediately after purchase.)
• Please use only LR6 alkaline batteries or nickel hydride (HR6) batteries. Please do not use manga­nese, oxyride batteries, etc.
Before installing or exchanging batteries, turn the
1.
power switch off ( ).
Open the battery cover and install four batteries with
2.
the correct polarity.
Be sure to close the battery cover.
3.
Page 28
2.1 Supplying Power
27

Connecting the AC Adapter (Option)

• Turn the instrument off before connecting the AC adapter to the instrument and to AC power.
• Use only the supplied Model 9445-02 AC Adapter or Model 9445-03 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
Turn the power switch off ( ).
1.
Connect the 9445-02 AC Adapter or the 9445-03 AC
2.
Adapter output plug to the AC adapter terminal for the instrument. (The input voltage range is 9 VDC±10% and the adapter has a positive center pin)
After making sure that the voltage of the power supply
3.
being used matches the supply voltage of the AC adapter, put the plug in the outlet.
Page 29

2.2 Connecting the Input Cord and the Test Lead

28
2.2 Connecting the Input Cord and the Test Lead
The SS7012 has voltage input terminal, current input terminal (terminal for the standard resistor) and output terminal. In addition, RJ sensor terminal is provided for temperature mea­surement.
To prevent an electric shock accident, confirm that the white or red portion (insulation layer) inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable.
Removable sleeves are attached to the metal pins at the ends of the test leads. The test leads can also be used with the sleeves removed.
Removing and attaching the sleeves
The tips of the metal pins are sharp, so take care not to injure yourself.
Removing the sleeves
Gently hold the bottom of the sleeves and pull the sleeves off. Safely store the removed sleeves so as not to lose them.
Attaching the sleeves
Insert the metal pins of the test leads into the holes of the sleeves, and firmly push them all the way in.
Page 30
2.2 Connecting the Input Cord and the Test Lead
Voltage measurement
Connect the L9170-10 Test lead to the voltage input ter­minal.
Connect the red and black connectors to the + and - input terminals, respectively. Input only voltage to the voltage input terminal.
L9170-10 Test lead
Current measurement
Connect the L9170-10 Test lead to the current input ter­minal.
Connect the red and black connectors to the + and - input terminals, respectively. Input only current to the current input terminal.
L9170-10 Test lead
(red)
(black)
(red)
(black)
Connecting Method
29
Page 31
2.2 Connecting the Input Cord and the Test Lead
[TC:RJ] sourcing, Temperature measurement
Connect the optional 9184 Temperature Probe to the RJ sensor terminal.
9184 Temperature Probe
Sourcing
With the output off, connect the 9168 Input Cord to the output terminal.
Connect the red and black connectors to the + and - output terminals, respectively.
9168 Input Cord
(red)
(black)
0°C check for the 100 Ω resistance thermometers
Connect the 100 Ω standard resistance thermometer to the terminal for the standard resistor.
(red)
(black)
30
Page 32
2.2 Connecting the Input Cord and the Test Lead
• This instrument is designed for safe operation only at low voltage. Ensure that no more than 60 V is present between any terminal and ground, and that the potential difference between any two terminals does not exceed 60 V. The instrument cannot be guaranteed in an overrange situation, such as due to an inadvertent electric shock or faulty insulation, and we cannot accept responsibility for any consequences that might occur from misuse.
• The output terminals and the voltage and current input terminals (terminals for the standard resistor) are all iso­lated. Before connecting the terminals, ensure that the target device will not provide excessive output to, or require excessive input from, the terminals. The instru­ment cannot be guaranteed in an overrange situation, such as due to an inadvertent electric shock or faulty insulation, and we cannot accept responsibility for any consequences that might occur from misuse. (For the maximum value of the output voltage/current and voltage/ current input, refer to the specifications)
• To avoid electrical accidents, confirm that all connections are secure. The increased resistance of loose connec­tions can lead to overheating and fire.
• Be sure to connect the voltage input and current input ter­minals correctly. An incorrect connection could damage or short circuit this instrument.
31
Page 33
2.2 Connecting the Input Cord and the Test Lead
32
• When switching the power on and off, remove the target device from all terminals. Furthermore, always set the output to OFF when connecting or disconnecting the output terminal and target device. In either of these cases, failure to heed this caution may result in damage to the instrument or to the target device.
• Do not input current more than 25 mA DC to the terminal for the standard resistor. Doing so may damage the resistor of the instrument.
• Output terminal and current input terminal have circuit protec­tion fuses. If the fuse is blown, output or current measurement is impossible. Exchange the fuses before operation.
See: "8.2 Replacing the Circuit Protection Fuse" (p.94)
• When output is off, the output terminals are shorted
with a resistance of about 50 kΩ.
• Input resistance of the voltage input terminal is
approx. 1 MΩ and that of the current input terminal is approx. 30 Ω.
Page 34

2.3 Turning the Power On and Off

33
2.3 Turning the Power On and Off
Before turning the instrument on, make sure the supply voltage matches that indicated on the AC adapter. Connec­tion to an improper supply voltage may damage the instru­ment or AC adapter and present an electrical hazard.
When switching the power on and off, remove the target device from all terminals. Furthermore, always set the output to OFF when connecting or disconnecting the output terminal and tar­get device. In either of these cases, failure to heed this caution may result in damage to the instrument or to the target device.
• If power switch is set to on ( ) for the instrument, even when sourcing or measurement is not executed current is still consumed. The lifetime of battery there­fore shortens. The power switch should be turned off ( ) when unused.
• Because turning the power on or off places a load on the instrument, wait at least 5 seconds before turning the power off or on again.
• If the power doesn’t turn on when turning on the power again, turn the power off and leave for more than 5 seconds.
• The warm-up time of this instrument is 5 minutes. Source values and measurement values can be unstable within 5 minutes after turning the power on.
Page 35

2.4 Pre-Operation Inspection

Set the power switch to the " " position. When the power switch is on, all settings except memory are initialized to appear initial screen after LCD display indications are all lit.
(ON)
Set the power switch to the " " position. Turn off the power switch to go off the LCD display indications.
(OFF)
Is the insulation of the test lead or input cord to be used damaged, or is bare metal ex­posed?
Is damage to the AC adapter evident?
Is damage to the instrument evident?
Do not use the instrument if even one point of dam­age is found, as it can cause electric shock. If damage is evident, re­quest repairs.
Yes
Yes
Yes
34
Turning Power On
Turning Power Off
2.4 Pre-Operation Inspection
Before using the instrument, verify that it operates normally to ensure that no damage occurred during storage or shipping.
Instrument inspection
1.
Check the following points.
Page 36
Inspection at power on
The batteries may have run out. Replace the bat­teries with new ones and then check again.
No
Inspection complete
Yes
The instrument may be damaged internally. Request repairs.
No An error is displayed
When the power is switched on, does the screen come on?
An error is displayed
Yes
Use a powered outlet and then check again.
Yes
The power cord may be disconnected or the AC adapter or instrument in­ternals may be malfunc­tioning. Request repairs.
Is a power failure or another cause preventing power from reaching the outlet?
No
Yes
Inspection complete
The instrument may be damaged internally. Request repairs.
No
When the power is switched on, does the screen come on?
An error is displayed
2.
< When using batteries >
< When using AC adapter >
2.4 Pre-Operation Inspection
35
Page 37
37

Sourcing Chapter 3

Always inspect the instrument before sourcing.
See: "2.4 Pre-Operation Inspection" (p.34)
Always set the output to OFF when connecting or disconnecting the output terminal and target device.
• If the display shows all zeros or is out of the setting range while using the disabled.
• Settings can be changed consecutively using the
/ keys. However, this operation cannot be used in
the following cases.
• When the changed setting exceeds the permitted range
• When the changed setting exceeds “00000” and the
polarity changes (For example, when the setting is “00001”, all keys and the right-most key can be used.)
+/- keys, polarity changing is
Page 38

3.1 Sourcing Example

38
3.1 Sourcing Example
The SS7012 can source constant voltage, constant current, and standard register. Instead of the thermocouple, a voltage corresponding to a spec­ified temperature can be sourced.
Constant voltage (CV) (p.39)
In [CV:2.5V], constant voltage can be sourced from -2.5 to +2.5 V with the resolution of 100 µV. In [CV:25V], constant voltage can be sourced from -25 to +25 V with the resolution of 1 mV.
Constant current (CC) (p.39)
In [CC:25mA], constant current can be sourced from -25 to +25 mA with the resolution of 1 µA.
Thermocouple (TC) (p.43)
In [TC], voltage can be sourced according to the setting temperature instead of thermocouples. Eight types of thermocouples are supported: K, E, J, T, R, S, B, and N. Thermometers using thermocouples can be calibrated.
Standard register (Rs) (p.50)
100 Ω standard resistor is connected to the terminal for the standard resistor (current input terminal). The resister can be used for the 0°C check of the 100 Ω standard re­sistance thermometer.
Output monitor function (p.51)
When a constant voltage is sourced, the current flowing in the target device can be monitored. (this function is used when testing the two­wire transmitter in a loop test.) When a constant current is sourced, the voltage between the output terminals can be monitored. When a thermoelectromotive force is sourced, the reference junction temperature can be monitored. ("0°C" is displayed in [TC:0°C], the temperature measured with the optional 9184 Temperature Probe is displayed in [TC:RJ])
Memory source function (p.54)
Values to be sourced can be recorded ahead of time. The recorded value can be switched using a single key operation or automatically. (When testing a distributor in the loop test, 4-20 mA can be easily sourced.) For each source function, up to 20 values can be recorded.
Page 39

3.2 Sourcing Constant Voltage and Constant Current

SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
3.2 Sourcing Constant Voltage and Constant Current
Source DC voltage and DC current.
Turn the power switch on.
1.
Press the SOURCE FUNCTION key
2.
the required number of times to set the constant voltage source function ([CV:2.5V], [CV:25V]) or constant cur­rent source function ([CC:25mA]).
At power on, [CV:2.5V] is set. Pressing the
SOURCE FUNCTION key changes the setting
in the lower shown sequence.
[CV:25V] [CC:25mA] [TC:0°C]
[CV:2.5V] [TC:RJ]
In the upper part of the screen, the moves to the specified function.
39
Page 40
40
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
3.2 Sourcing Constant Voltage and Constant Current
Use the / keys and +/- key to set
3.
the value to be sourced.
The five / keys correspond to the five columns of the setting.
The ranges within which the values can be set are as follows:
[CV:2.5V] : [CV:25V] : -25.000 V to 25.000 V [CC:25mA] : -25.000 mA to 25.000 mA
If the target device has a power
4.
source, switch it off and then con-
-2.5000 V to 2.5000 V
nect the target device to the output terminal of the SS7012. If necessary, switch the power of the target device back on. If the target device doesn’t have a power source, simply connect the target device to the output terminal of the instrument.
Page 41
3.2 Sourcing Constant Voltage and Constant Current
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
Press the OUTPUT ON/OFF key.
5.
is displayed on the screen and the set­ting value is output. However, if the load gets excessive and the setting value cannot be output, the setting val­ue flashes. If the setting value starts flashing, press the OUTPUT ON/OFF key to set the output to OFF.
See: "3.7 When setting value is flashing" (p.65)
41
If you want to change the value to be
6.
sourced during sourcing, use the
/ keys and +/- key.
The new set value is output.
Page 42
3.2 Sourcing Constant Voltage and Constant Current
42
Press the OUTPUT ON/OFF key.
7.
is displayed on the screen and the out-
put changes to OFF.
Never connect this instrument to any voltage sources when sourcing constant voltage. Furthermore, never connect this instrument to an electric current source when sourcing constant current. Doing so may damage the instrument and the target device.
• In constant voltage source function ([CV:2.5V], [CV:25V]), do not set the output to exceed ±25 mA. Furthermore, in constant current source function ([CC:25mA]), do not set the output to exceed ±25 V. Doing so may damage the instrument and the target device.
• When using constant current source function, if the target device is connected incorrectly, overvoltage can be applied to the target device, which may result in damage to this instru­ment or to the target device.
If the indicator is flashing, output cannot be set to ON. (If it flashes during output, output continues until output is set to OFF.) Replace the batteries or connect the AC adapter.
Page 43

3.3 Sourcing Thermoelectromotive Force

Thermocouple
Copper conductor
Reference junction (0°C)
Container filled with ice water, etc.
Ther­mometer
+
-
43
3.3 Sourcing Thermoelectromotive Force
Outputs voltage according to the setting temperature instead of thermocouples. Thermometers using thermocouples can be calibrated. There are 2 types of thermoelectromotive force source func­tions. The temperature of voltage input terminal (reference junction) of the thermometer can be set to either 0°C or room temperature.
[TC:0°C] : Use when sourcing thermoelectromotive force with
the temperature of voltage input terminal (reference junction) of the thermometer at 0°C. As shown in the diagram below, use the [TC:0°C] setting when the calibrating the thermometer measuring the thermo­electromotive force using 0°C as a reference. When using this function, create 0°C by putting ice water in a thermos flask or other means and make the refer­ence junction to 0°C.
.
Page 44
3.3 Sourcing Thermoelectromotive Force
Compensation thermometer
-
Thermocouple
Thermocouple or compensated conductors
+
Ther­mometer
44
[TC:RJ] : Use when calibrating a thermometer that can com-
pensate the temperature of voltage input terminal (reference junction) of the thermometer. As shown in the diagram below, use the [TC:RJ] setting when calibrating a thermometer that compensates the thermoelectromotive force of thermocouples at the reference junction temperature. However, the optional 9184 Temperature Probe is required to measure the temperature to compensate when using this function.
.
1.
Turn the power switch on.
Press the SOURCE FUNCTION key
2.
the required number of times to set the thermoelectromotive force source function ([TC:0°C], [TC:RJ]).
At power on, [CV:2.5V] is set. Pressing the
SOURCE FUNCTION key changes the setting
in the lower shown sequence.
[CV:25V] [CC:25mA] [TC:0°C]
[CV:2.5V] [TC:RJ]
In the upper part of the screen, the moves to the specified function.
Page 45
3.3 Sourcing Thermoelectromotive Force
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
Press the TC TYPE key to set ther-
3.
mocouple ([K], [E], [J], [T], [R], [S], [B] or [N]).
Use the / keys and +/- key to set
4.
the value.
The ranges within which the values can be set are as follows:
[K] :
-174.0°C to 1372.0°C
[E] : -220.0°C to 839.0 °C [J] : -208.0°C to 1108.0 °C [T] : -169.0°C to 400.0°C [R] : -50 °C to 1768 °C [S] : -50 °C to 1768 °C [B] : 300 °C to 1820 °C [N] : -113 °C to 1300.0 °C
45
Page 46
3.3 Sourcing Thermoelectromotive Force
< Figure 1 >
46
Connect the thermometer to the out-
5.
put terminal.
The connection methods for [TC:0°C] and
[TC:RJ] differ.
[TC:0°C]: Connect it as indicated in Figure 1.
The optional 9184 Temperature Probe may be connected to the RJ sensor terminal or not, without ef­fect. Create 0°C by putting ice water in a thermos flask or other means and make the reference junction to 0°C.
Page 47
< Figure 2: When using a copper conductor >
The temperature probe close to the thermometer.
Copper conductors
< Figure 3: when using a thermocouple the same type as the thermometer or compensated conduc­tors >
The temperature probe close to the output terminal of the SS7012.
Thermocouple or compensated conductors
3.3 Sourcing Thermoelectromotive Force
[TC:RJ]: Connect the optional 9184 Tempera-
ture Probe to the RJ sensor termi­nal. As indicated in Figure 2, connect the temperature probe close to the thermometer when us­ing a copper conductor, or close to the output terminal of the SS7012 as indicated Figure 3, when using a thermocouple the same type as the thermometer or compensated con­ductors.
47
Page 48
3.3 Sourcing Thermoelectromotive Force
48
Press the OUTPUT ON/OFF key.
6.
is displayed on the screen and the set­ting value is output. However, if the load gets excessive and the setting value cannot be output, the setting val­ue flashes. If the setting value starts flashing, press the OUTPUT ON/OFF key to set the output to OFF.
See: "3.7 When setting value is flashing" (p.65)
The output value will change if the
7.
setting value is changed with the
/ keys and
Press the OUTPUT ON/OFF key.
8.
is displayed on the screen and the out-
put changes to OFF.
To avoid damage to the instrument and the target device, in the thermoelectromotive force source function ([TC:0°C], [TC:RJ]), never apply an external voltage or current to the output terminal.
• The reference junction compensation temperature ranges are as follows:
[K], [E], [J], [T], [R], [S], [N]: -25.0 to 80.0°C [B]: 0 to 80°C
• In the thermoelectromotive force source function ([TC:0°C], [TC:RJ]), use a thermocouples thermometer with an input resistance of at least 100 kΩ. If the input resistance is under 100 kΩ, the following error will be added to the specification precision: 100 × (1-Rin/ (2.2+Rin))% (where Rin is the thermocouples ther­mometer input resistance.)
+/- key.
Page 49
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
3.3 Sourcing Thermoelectromotive Force
• When output in the thermoelectromotive force source function ([TC:RJ]) is on, reference junction compensa­tion will be performed about every 5 seconds, and the output value compensated.
• In the thermoelectromotive force source function ([TC:RJ]), when the detected temperature of the optional 9184 Temperature Probe is outside the -25°C to 80°C range, or if connection is incorrect, the mes­sages "rJErr" will be displayed even when the output on switch is pressed, and output will remain off.
• To perform reference junction compensation at a tem­perature out of the operating temperature range of the main instrument, be careful that the ambient tempera­ture of the main instrument.
• If the instrument is moved to a new location, or the thermometer being compensated for is changed, errors may result due to changes in the timetempera­ture constants. Do not attempt to perform measure­ments until the instrument has had sufficient time to reach thermal stability.
• If the indicator is flashing, output cannot be set to ON. (If it flashes during output, output continues until output is set to OFF.) Replace the batteries or connect the AC adapter.
49
Page 50
3.4 0°C Check of the 100 Ω Resistance Thermometer
50
3.4 0°C Check of the 100 Ω Resistance
Thermometer
100 Ω reference resistor is connected between reference resis­tor terminals, which are the same terminals as the current input terminals. This resistor is for 0°C check of the 100 Ω standard resistance thermometer. Usable except for current measurement function ([A:25mA]). Make sure measurement function is never set to current measurement function before connecting the instrument to the 100 Ω standard resistance thermometer.
Do not input current more than 25 mA DC to the terminal for the standard resistor. Doing so may damage the resistor of the instrument.
Page 51

3.5 Output Monitor Function

51
3.5 Output Monitor Function
• In the constant voltage source function ([CV:2.5V], [CV:25V]), the current that flows into the target device is displayed on the LCD display.
• In the constant current source function ([CC:25mA]), the volt­age between the output terminals is displayed on the LCD dis­play.
• In the thermoelectromotive force source function, the reference junction temperature is displayed on the LCD display ("0°C" is displayed in [TC:0°C], the temperature measured with the optional 9184 Temperature Probe is displayed in [TC:RJ]).
Turn the power switch on.
1.
Press the SOURCE FUNCTION key
2.
the required number of times.
Set the source function.
Use the / keys and +/- key to set
3.
the value to be sourced.
If the target device has a power
4.
source, switch it off and then con­nect the target device to the output terminal of the SS7012. If necessary, switch the power of the target device back on.
Page 52
52
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
3.5 Output Monitor Function
Press the OUTPUT ON/OFF key.
5.
is displayed on the screen and the set­ting value is output. However, if the load gets excessive and the setting value cannot be output, the setting val­ue flashes. If the setting value starts flashing, press the OUTPUT ON/OFF key to set the output to OFF.
See: "3.7 When setting value is flashing" (p.65)
Press the MONITOR key.
6.
The instrument switches from the measure function to the output monitor function.
Read the monitor value at the lower
7.
portion of the display.
Press the MONITOR key again.
8.
The instrument switches from the output mon­itor function to the measure function.
Press the OUTPUT ON/OFF key.
9.
put changes to OFF.
is displayed.
is displayed.
is displayed on the screen and the out-
Page 53
3.5 Output Monitor Function
Monitor ranges are shown below:
[CV:2.5V] Load current : -28.00 mA to 28.00 mA [CV:25V] Load current : -28.00 mA to 28.00 mA [CC:25mA] Load voltage : -28.00 V to 28.00 V [TC:0°C] Reference junction temperature: 0°C [TC:RJ] Reference junction temperature
: -25.0°C to 80.0°C
If the output is outside this range, the display will indicate "oF" in the constant voltage source function ([CV:2.5V], [CV:25V]) and constant current source function ([CC:25mA]), and "rJErr" in the thermocouple power function ([TC:RJ]). In this case, turn off the output and confirm the target device. If output is continued, the unit and the target device may be damaged.
• If the output is OFF, the display will indicate " - - - - - ".
• If the indicator is flashing, output cannot be set to ON. (If it flashes during output, output continues until output is set to OFF.) Replace the batteries or connect the AC adapter.
53
Page 54

3.6 Memory Source Function

54
3.6 Memory Source Function
This function records sourcing values and then later recalls them for sourcing. For each source function, up to 20 values can be recorded. In
[TC:0°C] and [TC:RJ], the thermocouple type can also be
recorded. There are 2 sourcing methods.
Recall sourcing (p.59): Single key operation for changing the setting.
Scan sourcing (p.61) : Automatically changes the setting value at
In Chapter 5, a loop test method is introduced as a practical example of using the memory source function.
See: "Chapter 5 Loop Test" (p.75)
In [CV:25V] and [CC:25mA], setting values are already saved as a default.
Address 1 2 3 4 5
CV:25V
CC:25mA
a set time interval.
1 V2 V3 V4 V5 V
-4 mA -8 mA -12 mA -16 mA -20 mA
Page 55
3.6 Memory Source Function
Confirm that a does not appear near [MEMORY] at the upper right of the screen (normal source mode). If a is displayed, press the MEMORY key the required number of times until it is no longer displayed.
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN

Saving setting value to memory

Turn the power switch on.
1.
Press the SOURCE FUNCTION key to
2.
set the source function.
Press the MONITOR key and the
3.
MEMORY key together.
The system will shift to the memory write mode. "01" is displayed in the memory address and output setting value of the address is dis­played. At this time, if the address "01" has been skipped before, "- - - - -" is displayed and numerical value is not displayed.
55
Page 56
3.6 Memory Source Function
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
MEASURE MONITOR ADJ RMT
V
C
mA
V
C
mA
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
56
Use the / keys and +/- key to set
4.
the value to be sourced.
The five / keys correspond to the five columns of the setting.
Press the ENTER key.
5.
After flashing, the setting value is set. Next, the instrument advances automatically to the next address.
Page 57
3.6 Memory Source Function
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
57
If you do not want to save a setting value to the address, press the SKIP key. (Skip setting)
"- - - - -" flashes and then the instrument ad­vances automatically to the next address.
Repeat steps 4. and 5. to save setting values to memory. A maximum of 20 items can be saved.
Press the MONITOR key and the
6.
MEMORY key together.
Exit memory write mode.
Page 58
3.6 Memory Source Function
58
Never switch off the power in memory write mode. It may dam­age the SS7012.
• Addresses with skip settings cannot be called by the memory source function.
• If the setting value or " value or skip setting is being written to memory.
• The sourcing time interval is set using scan source mode ([SCAN]).
• In memory write mode, measurements and output monitoring cannot be performed.
• Memory data is shared in the thermoelectromotive force source functions ([TC:0°C], [TC:RJ]).
- - - - -" is flashing, the setting
Page 59
3.6 Memory Source Function
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
12
3

Recall sourcing

Press the SOURCE FUNCTION key to
1.
set the memory sourcing function.
Press the MEMORY key once to set
2.
to the recall source mode ([RECALL]).
Use the ADDRESS keys (the 2 left-
3.
side / keys) to set the address to be sourced.
The output setting value in memory is dis­played at the sourcing screen.
Connect the target device.
4.
Press the OUTPUT ON/OFF key.
5.
59
is displayed on the screen and the set-
ting value is output.
Page 60
3.6 Memory Source Function
Normal source mode
Recall source mode
Scan source mode
60
Press the OUTPUT ON/OFF key.
6.
is displayed on the screen and the out-
put changes to OFF.
Repeat steps 3. to 6.
7.
Press the MEMORY key twice.
8.
Exit recall source mode.
• In recall source mode, the output setting value cannot be changed. The / keys only set addresses. To change the setting value, see "Saving setting value to memory" (p.55).
• When output is ON, press the MONITOR key to enable the output monitor function and monitor output. When output is OFF, measurement and monitoring cannot be performed.
• Every time the MEMORY key is pressed, the source mode changes in the following sequence.
• Addresses that can be set for recall sourcing are only those set between 1 and 20. (Those with no skip set­ting.)
• If the indicator is flashing, output cannot be set to ON. (If it flashes during output, output continues until output is set to OFF.) Replace the batteries or connect the AC adapter.
Page 61

Scan sourcing

SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
1
2
3
Start address for sourcing
4 Sourcing time
interval
1.
2.
3.
4.
3.6 Memory Source Function
61
Press the SOURCE FUNCTION key to set the memory sourcing function.
Press the MEMORY key twice to set to the scan source mode ([SCAN]).
Use the ADDRESS keys (the 2 left­side / keys) to set the address to be sourced.
The output setting value in memory is dis­played at the sourcing screen.
Use the INTERVAL keys (the 2 right­side / keys) to set the sourcing time interval.
The sourcing time interval is from 1 to 99 sec­onds. A sourcing time interval cannot be set for each address.
Connect the target device.
5.
Page 62
3.6 Memory Source Function
Normal source mode
Recall source mode
Scan source mode
62
Press the OUTPUT ON/OFF key.
6.
is displayed on the screen and output starts. The setting switches by the specified time in­terval.
Press the OUTPUT ON/OFF key.
7.
is displayed on the screen and the out­put changes to OFF. Scan sourcing also stops.
Press the MEMORY key once.
8.
Exit scan source mode.
• In scan source mode, the output setting value cannot be changed. The / keys only set addresses. To change the setting value, see "Saving setting value to memory" (p.55).
• When output is ON, press the MONITOR key to enable the output monitor function and monitor output. When output is OFF, measurement and monitoring cannot be performed.
• Every time the MEMORY key is pressed, the source mode changes in the following sequence.
• Addresses that can be set for scan sourcing are only those set between 1 and 20. (Those with no skip set­ting.)
• If the indicator is flashing, output cannot be set to ON. (If it flashes during output, output continues until output is set to OFF.) Replace the batteries or connect the AC adapter.
Page 63
3.6 Memory Source Function
63

Initializing setting values

• Initialize setting values saved to memory. Initialization by source function is possible. Furthermore, all functions can be initialized at same time.
• This feature is useful when re-saving settings to memory or when batch initializing the contents of functions that are no longer used.
• After initialization, the memory contents are as follows.
Address 1 2 3 4 5
CV:25V
CC:25mA
Those not listed above are as follows.
Setting count : 00000 Source setting : Skip (Not used) Sourcing time interval : 1 s Thermocouple : [K] ([TC:0°C], [TC:RJ] source function)
1 V2 V3 V4 V5 V
-4 mA -8 mA -12 mA -16 mA -20 mA
Turn the power switch on while hold-
1.
ing both the 0 ADJ and the MEMORY keys.
The display shows "CLr" when the Memory Clear mode is activated. If it does not, repeat step 1.
Page 64
3.6 Memory Source Function
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
64
Press the SOURCE FUNCTION key to
2.
select the function for which settings are to be initialized.
After selecting the thermoelectromotive force source function, all functions will be selected: ([CV:2.5V] [CV:25V] [CC:25mA]
[TC:0°C], [TC:RJ] → All functions)
Press the MONITOR key.
3.
The settings of selected functions are initial­ized and “CLr” flashes on the screen.
Turn the power off.
4.
To avoid damaging the instrument, do not turn the power off in the middle of the Memory Clear procedure. The "CLr" display should be blinking before turning the power off.
Page 65

3.7 When setting value is flashing

65
3.7 When setting value is flashing
When the OUTPUT key is pressed and output is set to ON, set­ting value may flash. When setting value is flashing, the value being output may be lower than the setting value. (Overload) Press the OUTPUT key to set the output to OFF. Check the connection of the SS7012 and the target device. If there are unexpected shorts or contact failures, eliminate them. If the setting value is still flashing even after setting output to ON, the performance level of the SS7012 is being exceeded. (See below.)
In the constant voltage source function of this instrument, the possible electrical current that can flow is ±25 mA. (Current must flow to source the specified constant voltage.) If a current exceeding ±25 mA flows, the source voltage may become lower than the setting value. In other words, the set voltage cannot be sourced in this condition. Furthermore, in the constant current source function, the applied voltage is ±25 V. (Voltage must be applied for the specified con­stant current to flow.) If a voltage exceeding ±25 V is applied, the source current may become lower than the setting value. In other words, the set current cannot be sourced in this condition.
Page 66

4.1 Measurement Example

67

Measurement Chapter 4

Always inspect the instrument before measurement.
See: "2.4 Pre-Operation Inspection" (p.34)
4.1 Measurement Example
The following are the voltages, currents, and temperatures that the SS7012 can measure.
Voltage (V) (p.68)
In [V:2.5V], measurement is possible from -2.8 to +2.8 V with the res­olution of 100 µV. In [V:25V], measurement is possible from -28 to +28 V with the reso­lution of 1 mV. In all cases, DC voltage can be measured.
Current (A) (p.68)
In [A:25mA], measurement is possible from -28 to +28 mA with the res­olution of 1 µA. DC current can be measured.
Temperature (TEMP) (p.71)
In [TEMP], measurement is possible from -25 to +80°C with the resolu­tion of 0.1°C. The optional 9184 Temperature Probe is required.
Page 67

4.2 Voltage Measurement and Current Measurement

SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
68
4.2 Voltage Measurement and Current Measurement
Measure DC voltage or DC current as following procedure.
Turn the power switch on.
1.
Press the MEASURE FUNCTION key
2.
the required number of times to set the voltage measure function ([V:2.5V], [V:25V]) or current measure function ([A:25mA]).
At power on, [FUNCTION OFF] is set. Press­ing the MEASURE FUNCTION key changes the setting in the lower shown sequence.
[V:2.5V] [V:25V] [A:25mA]
[FUNCTION OFF] [TEMP]
In the lower part of the screen, the moves to the specified function.
Page 68
4.2 Voltage Measurement and Current Measurement
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
3.
<For voltage measurement>
Connect the test leads to the voltage input terminals for the instrument and short out the ends of the test leads.
<For current measurement>
Connect the test leads to the current input terminals for the instrument and open the ends of the test leads.
Press the 0 ADJ key to execute the
4.
zero adjustment.
Connect the end of the test lead to
5.
the target device and read the dis­played value.
69
In order to prevent damage to the instrument or the target device, do not input voltage in excess of ±28 V to the voltage input terminal and do not input current in excess of ±28 mA to the current input terminal. Do not input voltage to the current input terminal.
Page 69
4.2 Voltage Measurement and Current Measurement
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
70
If the input is outside the measurement range, the dis-
play will indicate "oF".
• The zero adjustment function works only when the measurement value is within ±100 counts. When the measurement value is within ±100 counts, press the
0 ADJ key to revert the display to “00000”.
• Every time pressing the 0 ADJ key, the input value is memorized. The LCD shows the difference between the memorized value and the input value.
• In the following cases, the zero adjustment function does not work and “AdJEr” is displayed, even if the
0 ADJ key is pressed.
• In [V:2.5V], the input value exceeds ±0.0100 V
• In [V:25V], the input value exceeds ±0.100 V
• In [A:25mA], the input value exceeds ±0.100 mA
Page 70

4.3 Temperature Measurement

9184 Temperature Probe
71
4.3 Temperature Measurement
Temperature measurement is possible with the optional 9184 Temperature Probe.
Connect the optional 9184 Tempera-
1.
ture Probe to the instrument.
Turn the power switch on.
2.
Page 71
4.3 Temperature Measurement
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
72
3.
4.
Press the MEASURE FUNCTION key 4 times to set the temperature mea­sure function ([TEMP]).
At power on, [FUNCTION OFF] is set. Press­ing the MEASURE FUNCTION key changes the setting in the lower shown sequence.
[V:2.5V] [V:25V] [A:25mA]
[FUNCTION OFF] [TEMP]
In the lower part of the screen, the moves to the specified function.
Read the displayed value.
Page 72
4.3 Temperature Measurement
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
73
• If the temperature measurement function is selected without the optional 9184 Temperature Probe being connected, the message "rJErr" will be displayed. When a Temperature Probe is connected, the error disappears and temperature measurement starts.
• The optional 9184 Temperature Probe cannot be used out of the temperature measurement range between
-25 and 80°C. If measurement exceeds this tempera­ture range, the lead insulation of the temperature probe being used is damaged. In addition, if the input is outside measurement range, the display will indicate "rJErr".
Page 73

Loop Test Chapter 5

The connections for the two-wire transmitter, etc.
Sensor
Two-wire transmitter
Distributor
Output signal
4-20 mA
24 V

5.1 Loop Test Overview

Loop tests can be performed with the SS7012. The loop tests are for systems that include two-wire transmitters and distributors. (See the diagram below). Two-wire transmitters and distributors can be tested using the SS7012. The two-wire transmitter receives the sensor output and con­verts it to a 4-20 mA current signal. In addition to supplying the power to the two-wire transmitter, the distributor outputs a signal corresponding to the current signal to a later stage.
5.1 Loop Test Overview
75
Page 74

5.2 Testing a Distributor (4-20 mA)

Distributor test
Current source
SS7012
Distributor
4-20 mA
Output signal
76
5.2 Testing a Distributor (4-20 mA)
The 4-20 mA current flow of the distributor can be tested by using the SS7012 to sink the 4-20 mA current (a two-wire trans­mitter simulation). For the distributor test, setting values of the source current are minus because the SS7012 sinks current. It is assumed that the memory source function setting of the SS7012 is default. When the setting is not at the default, initial­ize the memory settings or set -4 mA/ -8 mA/ -12 mA/
-16 mA/ -20 mA.
See: "Initializing setting values" (p.63)
"Saving setting value to memory" (p.55) Note that when you make the setting yourself, they may not match with the [ADRS] values of the following procedure.
Page 75
5.2 Testing a Distributor (4-20 mA)
77
Test procedure
1. Check which distributor input terminal has the higher potential.
Power on the SS7012 and perform the pre-operation
1.
inspection.
See: "2.4 Pre-Operation Inspection" (p.34)
Press the MEASURE FUNCTION key to set the mea-
2.
sure function to [V:25V].
Connect the voltage input terminal of the SS7012 and
3.
the input terminal of the instrumentation.
Read the voltage of the distributor input terminals
4.
and check which terminal has the higher potential.
Release the connection with the distributor of the
5.
SS7012.
2. Checking the memory source function setting
Press the SOURCE FUNCTION key to set the source
1.
function to [CC:25mA].
Press the MEMORY key to switch from normal source
2.
to recall source. (The selected memory source func­tion is displayed at the upper right display area.)
Check that [ADRS] is “01” and the setting is -4.000
3.
mA.
Press the right ADDRESS key (the ADDRESS keys
4.
are the 2 / keys on the left) and switch [ADRS] from “01” to “02”.
Check that the setting is -8.000 mA.
5.
Press the ADDRESS key. Check that, as [ADRS]
6.
increases from “03” to “04” to “05”, the setting changes to -12.000 mA to -16.000 mA to -20.000 mA.
Press the ADDRESS key to set [ADRS] to “01”.
7.
Page 76
5.2 Testing a Distributor (4-20 mA)
78
3. Connect the SS7012 and the distributor and then test.
Connect the output terminal of the SS7012 and the
1.
input terminal of the distributor as shown in the fig­ure. Be careful of the connection direction. Connect the positive side of the output terminal of the SS7012 with the input terminal of the distributor with the higher potential. Connect the negative side of the output terminal of the SS7012 with the input terminal of the distributor with the lower potential.
Press the OUTPUT ON/OFF key to set the output to
2.
ON. (During output, is displayed in the display area.) A current of 4.000 mA is now flowing. (The set­ting is -4.000 mA.)
Check the indicated value of the meter.
3.
In this state, press the ADDRESS key to set [ADRS] to
4.
02”. A current of 8.000 mA is now flowing. (The set­ting is -8.000 mA.)
Check the indicated value of the meter.
5.
Similarly to below, increase [ADRS] to “03”, “04”, and
6.
05". (The current flow increases to 12.000 mA,
16.000 mA, and 20.000 mA). Check that the indicated value of the meter accordingly.
Press the OUTPUT ON/OFF key to set the output to
7.
OFF.
Release the connection with the distributor of the
8.
SS7012.
Turn off the power of the SS7012.
9.
Page 77

5.3 Testing a two-wire transmitter

Two-wire transmitter test
4-20 mA
Voltage source 24 V (Using the monitor function)
SS7012
Sensor
Two-wire transmitter
5.3 Testing a two-wire transmitter
The two-wire transmitter can be tested by sourcing voltage (volt­age of the power for the transmitter of the distributor) from the SS7012 and monitoring the output current. In this case, it is assumed that the power voltage of the two-wire transmitter is 24 V. (24 V loop power)
Power on the SS7012 and perform the pre-operation
1.
inspection.
See: "2.4 Pre-Operation Inspection" (p.34)
Press the SOURCE FUNCTION key to set the source
2.
function to [CV:25V].
Press the / keys to set the source setting to
3.
24.000 V (the voltage of the distributor).
Connect the two-wire transmitter and the SS7012 as
4.
shown in diagram “Two-wire transmitter
Press the OUTPUT ON/OFF key to set the output to
5.
6.
ON. (During output, is displayed in the display area.)
Press the MONITOR key to set the output monitor function to ON.
79
test”.
Page 78
5.3 Testing a two-wire transmitter
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C
mA
V
C
mA
EJTRSBN
80
Insert a suitable input into the two-wire transmitter
7.
(pressure, temperature, etc.). Read and check the current value from the SS7012 at this time. (Do this for all the inputs that required for testing. The display example is as follows.)
Press the MONITOR key to set the output monitor
8.
function to OFF.
Press the OUTPUT ON/OFF key to set the output to
9.
OFF.
Release the connection with the distributor of the
10.
SS7012.
Turn off the power of the SS7012.
11.
Page 79

6.1 USB Communication Function

81

Other Function Chapter 6

6.1 USB Communication Function
The SS7012 can communicate with a computer. (Remote con­trol) The communication function can set the SS7012, as well as record and save settings and measurements on the computer.
To use this function, the optional SS9000 Communica­tion Package is required. For more details about using the communication func­tion, see the Remote operation manual included with the SS9000 Communication Package.
Page 80

7.1 General Specifications

Specifications Chapter 7

7.1 General Specifications
83
Operating environment
Storage temperature and humidity
Operating temperature and humidity
Dielectric strength
Applicable Standards
Effect of radiated radio-frequen­cy electromag­netic field
Effect of conducted ra­dio-frequency electromagnet­ic fiel d
Indoors, Pollution degree 2, up to 2000 m (6562-ft.) ASL
-20 to 50°C (-4 to 122°F) 80%RH or less (non-condensating)
0 to 40°C (32 to 104°F) 80%RH or less (non-condensating)
720cccc VrmsAC (Current sensitivity: 1 mA) for 1 minute Between output terminals and input terminals
Safety EN61010 EMC EN61326
EN61000-3-2 EN61000-3-3
5%f.s. at 3 V/m
3%f.s. at 3 V
Page 81
7.1 General Specifications
84
9445-02 AC Adapter 9445-03 AC Adapter 9 V, 1 A
Rated supply voltage: 100 to 240 VAC (Voltage fluctuations of ±10% from the rated supply
voltage are taken into account.)
Rated supply frequency: 50/60 Hz
Power supply
Continuous operating time
Dimensions
Mass Approx. 570 g (20.1 oz.) (without batteries)
Product warranty period
Output resistance
Input resistance
Maximum input voltage
Maximum input current
Maximum rated voltage to earth
Temperature and humidity for guaranteed accuracy
Anticipated transient overvoltage: 2500 V Maximum rated power: 6 VA
LR6 alkaline battery
Rated supply voltage: 1.5 DC × 4 Maximum rated power: 6 VA
Nickel hydride batteries (HR6 × 4)
Rated supply voltage: 1.2 DC × 4 Maximum rated power: 6 VA
When using nickel hydride batteries (min. 2450 mAh) (23°C reference value) Approx. 200 minutes (after full recharge, [CC:25mA], 20 mA source, 250 Ω load) Approx. 170 minutes (after full recharge, [CV:25V], 25 V source, 1 kΩ load) Approx. 280 minutes (after full recharge, [V:2.5V], when input terminal is shorted)
Approx. 104W × 180H × 58D mm (4.09"W × 7.09"H × 2.28"D) (without protrusions)
1 year
CV: 300 mΩ or less TC: 5 Ω or less
DCV: 1 MΩ±1% DCA: 30 Ω or less (Including approx. 15 Ω for the fuse)
28 VDC
28 mADC
28 VDC Anticipated transient overvoltage: 330 V
23±5°C, 80%RH or less
Page 82
Warm-up time 5 minutes
Battery voltage
Period of guaranteed accuracy
Source output method
Source functions
Measure functions
Monitor measurement
Added functions
Accessories
Options
(Battery low warning indicator)
When flashing, accuracy cannot be guaranteed.
1 year
Bipolar sink-source system
DC constant voltage [CV:2.5V] [CV:25V] DC constant current [CC:25mA] Thermoelectromotive force [TC:0°C] [TC:RJ] Standard resistance [Rs]
DC voltage [V:2.5V] [V:25V] DC current [A:25mA] Temperature [TEMP]
Load current [CV:2.5V] [CV:25V] Load voltage [CC:25mA] Reference junction temperature [TEMP]
Memory source function
(Recall sourcing [RECALL], Scan sourcing [SCAN])
Zero adjustment function
([V:2.5V] [V:25V] [A:25mA] Within ±100)
Overload display function Communication function
(Virtual COM port by USB interface 1.1)
Battery low warning function
(When the power voltage is 4.6 V or less, flashes.)
9168 Input Cord 1set L9170-10 Test Lead 1set Fuse 250VF50mAH (φ5.2 × 20 mm) 1 LR6 alkaline batteries 4 Instruction Manual 1
9184 Temperature Probe
(reference junction compensation) 9380 Carrying Case (Holds main unit only, soft case) 9782 Carrying Case (Holds options, hard case) 9445-02 AC Adapter (For Japan, US, and Canada) 9445-03 AC Adapter (For EU) SS9000 Communication Package
(Includes USB cable and USB driver software)
7.1 General Specifications
85
Page 83
86
Sourcing

7.2 Accuracy

Sourcing
Constant voltage
[CV:2.5V]
0 to ±2.5000 V
Constant voltage
[CV:25V]
0 to ±25.000 V
Constant current
[CC:25mA]
0 to ±25.000 mA
Thermoelectromo­tive force [TC:0°C]
[K]
-174.0 to 1372.0°C
[E]
-220.0 to 839.0°C
[J]
-208.0 to 1108.0°C
[T]
-169.0 to 400.0°C
[R]
-50 to 100°C
101 to 1768°C 1°C
[S]
-50 to 100°C
101 to 1768°C 1°C
[B]
300 to 600°C
601 to 1820°C 1°C
[N]
-113.0 to 1300.0°C
7.2 Accuracy
Function
Reso-
lution
100 μV
1 mV
1 μA
0.1°C
0.1°C
0.1°C
0.1°C
1°C
1°C
1°C
0.1°C
Accuracy Remarks
±0.03% of setting ±300 μV
±0.03% of setting ±3 mV
±0.03% of setting ±3 μA
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±2.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±0.5 °C
Maximum output ±25 mA or more (sink-source) When shorted, 50 mA or less. Load regulation
0.3 mV/mA Maximum output
±25 V or more (sink-source) When open, 40 V or less.
Page 84
7.2 Accuracy
Sourcing
Memory sourcing
Function
Thermoelectromo­tive force [TC:RJ]
[K]
-174.0 to 1372.0°C
[E]
-220.0 to 839.0°C
[J]
-208.0 to 1108.0°C
[T]
-169.0 to 400.0°C
[R]
-50 to 100°C
101 to 1768°C 1°C
[S]
-50 to 100°C
101 to 1768°C 1°C
[B]
300 to 600°C
601 to 1820°C 1°C
[N]
-113.0 to 1300.0°C
Standard resistor [Rs] 100 Ω
Memory
[RECALL], [SCAN]
One of the following 4 functions: [CV:2.5V] [CV:25V] [CC:25mA] [TC:0°C]/[TC:RJ] Maximum number of steps: 20 Sourcing time interval: 1 to 99 s (1 s step)
Reso-
lution
0.1°C
0.1°C
0.1°C
0.1°C
1°C
1°C
1°C
0.1°C
Accuracy of sourced value depends on accuracy of selected function. The timing accuracy of the sourcing time interval is not rated.
Accuracy Remarks
±0.05% of setting ±1. 0°C
±0.05% of setting ±1. 0°C
±0.05% of setting ±1. 0°C
±0.05% of setting ±1. 0°C
±0.05% of setting ±2. 0°C
±0.05% of setting ±1. 5°C
±0.05% of setting ±2. 0°C
±0.05% of setting ±1. 5°C
±0.05% of setting ±3. 0°C
±0.05% of setting ±2. 0°C
±0.05% of setting ±1. 0°C
±0. 2 Ω
Using the 9184 Temperature Probe The accuracies list­ed to the left are when the monitor measurement value is 23±5°C
At other tempera­tures, 9184 temper­ature measurement accuracy portions are added. ±0.5°C added (0 to 50°C) ±1.5°C added (-25 to 80°C)
When current mea­sure function is off, standard resistance input terminal
Memory data is shared in [TC:0°C] and [TC:RJ] func­tions.
87
Page 85
7.2 Accuracy
Measurement
Monitor
88
Measurement
% rdg. is the % of the reading Sampling rate: Approx. 1.67 times/s
Function
Voltage [V:2.5V] 0 to ±2.8000 V
Voltage [V:25V] 0 to ±28.000 V
Current [A:25mA] 0 to ±28.000 mA
Temperature
[TEMP]
-25.0 to 80.0°C Load current
[CV:2.5V]
0 to ±28.00 mA Load current
[CV:25V]
0 to ±28.00 mA Load voltage
[CC:25mA]
0 to ±28.00 V Reference junction
temperature
[TC:0°C]
Reference junction temperature
[TC:RJ]
-25.0 to 80.0°C
Reso-
lution
100 μV ±0.03% rdg.±300 μV
1 mV ±0.03% rdg.±3 mV
1 μA ±0.03% rdg. ±3 μA
±0.5°C (23±5°C)
0.1°C
±1.0°C (0 to 50°C) ±2.0°C (-25 to 80°C)
10 μA ±0.1% rdg. ±20 μA
10 μA ±0.1% rdg. ±20 μA
10 mV ±0.1% rdg. ±20mV
0.0°C 0°C reference
±0.5°C (23±5°C)
0.1°C
±1.0°C (0 to 50°C) ±2.0°C (-25 to 80°C)
Accuracy Remarks
After executing a zero adjustment
Using the 9184 Temperature Probe
Using the 9184 Temperature Probe
Page 86
7.2 Accuracy
Temperature coefficient
For temperatures other than 23°C±5°C, the following is added.
Sourcing
[CV:2.5V] : ±0.005% [CV:25V] : ±0.005% [CC:25mA] : ±0.005%
of setting ±30 mV/°C of setting ±0.2 mV/°C of setting ±0.2 mA/°C
[TC:0°C] ([K], [E], [J], [T], [N])
[TC:0°C] ([R], [S], [B]): ±0.005%
: ±0.005%
of setting ±0.05°C/°C of setting ±0.10°C/°C
[TC:RJ] ([K], [E], [J], [T], [N], [R], [S], [B])
: ±0.01%
of setting ±0.10°C/°C
[Rs] : ±0.005 Ω/°C
Measurement
% rdg. is the % of the reading
[V:2.5V] : ±0.005% rdg. ±10 μV/°C [V:25V] : ±0.005% [A:25mA] : ±0.005%
rdg. ±0.1 mV/°C rdg. ±0.1 μA/°C
[TEMP] : ±0.05°C/°C [MONITOR] ([CV:2.5V], [CV:25V])
[MONITOR] ([CC:25mA]): ±0.01%
: ±0.01%
rdg. ±1 μA/°C rdg. +1 mV/°C
[MONITOR] ([TC:RJ]) : ±0.05°C/°C
89
Page 87

8.1 Troubleshooting

91

Maintenance and Service Chapter 8

8.1 Troubleshooting
Do not attempt to modify, disassemble or repair the instru­ment; as fire, electric shock and injury could result.
• If damage is suspected, check the "Before returning for repair" (p.93) before contacting your dealer or Hioki representative.
• Getting the instrument wet or letting oil or dust enter inside its casing will certainly damage it, and is quite likely to cause an electric shock accident or a danger­ous conflagration. If the instrument has gotten seri­ously wet, oily, or dusty, stop using it and send it for service at an approved Hioki calibration facility.
Transporting
When sending the instrument for repair, remove the batteries and pack carefully to prevent damage in transit. Include cushion­ing material so the instrument cannot move within the package. Be sure to include details of the problem. Hioki cannot be responsible for damage that occurs during shipment.
Page 88
8.1 Troubleshooting
92
Replaceable Parts and Operating Lifetimes
Useful life depends on the operating environment and frequency of use. Operation cannot be guaranteed beyond the following periods. For replacement parts, contact your dealer or Hioki rep­resentative.
Part Life
Electrolytic Capacitors Approx. 9 years
There is a fuse inside in the power unit of the instrument. If the power does not turn on, the fuse may be blown. If this occurs, a replacement or repair cannot be performed by customers. Please contact your dealer or Hioki representative.
Page 89
8.1 Troubleshooting
Before returning for repair
Symptom Cause Remedy
Although you have turned on the power switch, the screen display does not ap­pear.
Constant volt­age, constant current, ther­moelectromo­tive force sourcing or cur­rent measure­ment are impossible.
The value is un­usual when sourcing ther­moelectromo­tive force.
Sourcing is not possible.
• Are the batteries cor­rectly inserted?
• Is the useful battery life at an end?
• Is the optional 9445­02 AC Adapter or the 9445-03 AC Adapter connected?
• Is the circuit protec­tion fuse fitted?
• Has the circuit pro­tection fuse blown?
• Are you attempting to calibrate the ther­mometer for internal reference junction compensation using
[TC:0°C]?
• Is the indicator flashing?
Insert the new batteries.
See: "Installing or Replac-
Connect the AC Adapter. (Check that power is reach­ing the outlet.)
See: "Connecting the AC
Change the circuit protec­tion fuse.
See: "8.2 Replacing the
Use the 9184 Temperature Probe and calibrate using
[TC:RJ].
• Insert the new batteries.
See: "Installing or Replac-
• Connect the AC Adapter. (Check that power is reaching the outlet.)
See: "Connecting the AC
93
ing the Batteries" (p.25)
Adapter (Option)" (p.27)
Circuit Protection Fuse" (p.94)
ing the Batteries" (p.25)
Adapter (Option)" (p.27)
Page 90

8.2 Replacing the Circuit Protection Fuse

94
8.2 Replacing the Circuit Protection Fuse
• To avoid electric shock, turn off the power switch and dis­connect the test leads or input cords from the target device before replacing the fuse.
• After replacing the batteries, replace the cover before using the instrument.
• Replace the fuse only with one of the specified character­istics and voltage and current ratings. Never use unspeci­fied fuses and never use the instrument after the fuse holder has shorted. This will damage the instrument and cause injury. Fuse type: 250VF50mAH (φ5.2 ×
The fuses are fitted on the PCB where the lower case is removed.
• The SS7012 has three fuses: two of them are for source func­tion, one for measure function. Fuses are common for all the function.
• The fuse for the constant voltage ([CV:2.5V], [CV:25V]) and constant current ([CC:25mA]) source functions is labeled "F201" on the circuit board.
• The fuse for the thermoelectromotive force source functions ([TC:0°C] and [TC:RJ]) is labeled "F202" on the circuit board.
• The fuse for the current measure function ([A:25mA]) is labeled "F401" on the circuit board.
20 mm)
Page 91
3
4,9
(2)
5
6
(1)
(3)
(4)
8.2 Replacing the Circuit Protection Fuse
Turn off the power switch.
1.
Remove the AC adapter, if
2.
the AC adapter or some­thing is connected to the AC adapter terminal.
Remove the battery cover
3.
and then remove all LR6 alkaline batteries or nickel hydride batteries.
Remove the two screws
4.
from the lower case, and then remove the lower case.
Release the connector cable
5.
connected to (1).
Change the blowing fuses.
6.
Confirm the following func­tions: (2): F202
Thermoelectromotive force source functions ([TC:0°C] and [TC:RJ])
(3): F401
Current measure function ([A:25mA])
(4): F201
Constant voltage source functions ([CV:2.5V] and
[CV:25V])
Constant current source function ([CC:25mA])
95
Page 92

8.3 Cleaning

Fuse Minus screwdriver
96
7. Connect the connector cable to (1).
8. Hook the back of the case into the tracks in the
front of the case. Do not allow the connector cable to be pinched by the case.
9. Retighten the two screws removed in step 4.
10. Install the batteries and mount the battery
cover.
When removing fuses, use a minus screwdriver and do not damage the parts on the circuit board.
8.3 Cleaning
To clean the instrument, wipe it gently with a soft cloth moist­ened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
Wipe the LCD gently with a soft, dry cloth.
Page 93

8.4 Error Indication

SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C m
EJTRSB
SOURCE ON OFF
K
V
C m
EJTRSB
SOURCE ON OFF
MEASURE
MONITOR ADJ RMT
K
V
C m
V
C m
EJTRSB
SOUC O O
MEASURE
MONITOR ADJ RMT
K
V
C m
V
C m
EJTRSB
MEASURE MONITOR ADJ RMT
V
C m
8.4 Error Indication
Error indication Description Remedial Action
Poor temperature probe connection
The measurement temperature at ther­mocouple B for
[TC:RJ] is a minus
value.
The input tempera­ture is outside the measurement range (of -25°C to 80°C).
Zero adjustment is not functioning
Connect the temperature probe correctly.
Set the measurement tem­perature between 0°C and 80°C.
Set the input temperature within the measurement range.
Set the measurement val­ue to within ±100 counts.
97

8.5 Other Indication

Internal error.
Abnormal setting val­ues saved to memo­ry.
Contact your dealer or Hio­ki representative.
Initialize the setting values saved to memory (p.63).
Indication Description Remedial Action
The input is out of the measurement range.
Set the input within the measurement range. (Change either the input value or range.)
Page 94

8.6 When calibrating the instrument

98
8.6 When calibrating the instrument
When calibrating the instrument, please ensure that the follow­ing conditions are met.
Environment Temperature 23±5°C
SS7012 Warm-up time 10 minutes or more
To calibrate the instrument, the following devices are necessary.
Voltmeter, ammeter, voltage/current generator, insulator and resistor with a sufficient accuracy to meet the accuracy specifications for the instrument
Temperature probe resistance and temperature conversion
0°C 918.57 Ω 23°C 993.51 Ω 40°C 1048.52 Ω
Ensure that the input does not exceed the maximum input voltage or current to avoid instrument damage, short-cir­cuiting and electric shock resulting from heat building.
To avoid damage to the instrument and the target device, never apply voltage to the current input terminal.
• Please refer to the respective calibrating devices' instruction manuals for the handling instructions.
• Please refer to JIS C1602-1995 and IEC 584 for the corresponding table for temperature and thermoelec­tromotive force.
Page 95
Index 1
Index
Numerics
0 ADJ key .........................21
0°C Check ........................ 50
24 V Loop Power .............. 79
A
A:25mA .............................68
AC Adapter Terminal ........16
ADDRESS keys ..........59, 61
AdJEr ..........................70, 97
C
CC:25mA .......................... 39
Cleaning ...........................96
Communication Function .. 81
Current Input Terminal .....18
CV:2.5V ............................39
CV:25V .............................39
D
Default Memory Values .... 63
Distributor Test .................76
E
ENTER key .......................21
Error Indication .................97
F
Flashing Setting Value .....65
I
Initializing Memory ........... 63
INTERVAL keys ............... 61
L
Load Current .................... 53
Load Voltage .................... 53
LOCAL key ....................... 21
Loop Test ......................... 75
M
MEASURE .......................19
MEASURE FUNCTION key
.........................................21
Measurement (Current) .... 68
Measurement (Temperature)
.........................................71
Measurement (Voltage) .... 68
Measurement Range ........ 67
MEMORY key ......21, 59, 61
Memory Source Function .54
MEMORY WRITE ............55
Memory Write Mode ......... 55
MONITOR ........................ 19
Monitor .............................51
MONITOR key ..................21
O
oF ...............................70, 97
Output Monitor ................. 51
Output Terminal ............... 18
Page 96
Index 2
R
Recall Sourcing ................ 59
Reference Junction .......... 43
Remote Control ................ 81
Replacing the Batteries .... 25
Replacing the Fuse .......... 94
Resistance Thermometer 50
RJ Sensor Terminal ......... 16
rJErr ........................... 73, 97
RMT ................................. 20
S
Saving to Memory ............ 55
Scan Sourcing ................. 61
SKIP key .......................... 21
Skip Setting ...................... 57
SOURCE FUNCTION key 22 Sourcing (Constant Current)
......................................... 39
Sourcing (Constant Voltage)
......................................... 39
Sourcing (Thermoelectromotive Force)
.......................................... 43
Sourcing Range ............... 38
Standard Resistance ....... 50
T
TC:0°C ............................. 43
TC:RJ ............................... 44
TC TYPE key ................... 22
TEMP ............................... 71
Temperature Coefficient .. 89 Terminal for the
Standard Resistor ........... 18
Testing a
Two-wire Transmitter ....... 79
Thermocouple ...................43
Thermometer Calibration ..43 Two-wire Transmitter
Simulation ........................76
U
Unusual value in sourcing thermoelectromotive force
..........................................93
USB Port ...........................16
V
V:2.5V ...............................68
V:25V ................................68
Voltage Input Terminal .....18
Z
Zero Adjustment .........69, 70
Page 97
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Loading...