Hioki RM9006 Instruction Manual

Introduction
HIOKI RM9006A961-00
Safety Information
Parts Names
RM9006
MAINTENANCE TOOL
Instruction Manual
June 2022 Edition 1
RM9006A961-00 22-06H
EN
[600633700]
Thank you for choosing the Hioki RM9006 Maintenance Tool. To ensure your ability to get the most out of this instrument over the long term, please read this manual carefully and keep it available for future reference. Please review the instruction manual of the RM2610 Electrode Resistance Measurement System and “Operating Precautions” that came with the RM2610 before using the device.
The latest edition of the instruction manual
The contents of this manual are subject to change, for example as a result of product
improvements or changes to specications.
The latest edition can be downloaded from Hioki’s website.
https://www.hioki.com/global/support/download
Target audience
This manual has been written for use by individuals who use the product or provide information about how to use the product. In explaining how to use the product, it assumes electrical knowledge (equivalent of the knowledge possessed by a graduate of an electrical program at a technical high school).
Trademark
Microsoft is either a registered trademark or a trademark of Microsoft Corporation in the United States and other countries.
CAUTION
Please review the safety information below before
using the device.
Failure to do so could cause damage to the device.
Usage Notes
Be sure to follow the precautions listed below in order to use the device safely and in a manner that allows it to function
e󰀨ectively.
CAUTION
Do not place the device on an unstable stand or
angled surface.
Failure to do so could cause improper use of the device, resulting in serious bodily injury or damage to the device.
Turn o󰀨 the RM2611 Electrode Resistance Meter
before connecting or disconnecting the RM9005
Connection Cable.
Failure to do so could cause damage to the device.
Do not unplug the USB cable while
communications are ongoing.
Failure to do so could cause damage to the device and your computer.
The device requires a current of at least 100 mA to operate. Check
the specications of your computer or USB hub before use.
2
1
3
78
Holder
3 3
65
Side
4
Overview
The instrument is used to maintain the RM9004 Test Fixture.
You can clean a test xture with air while observing its probes
with the camera (microscope). Eight-direction air blasts are available.
Notations
Safety notations
This manual classies seriousness of risks and hazard levels
as described below.
Indicates a potentially hazardous situation
that, if not avoided, could result in minor
CAUTION
or moderate injury or potential risks of damage to the supported product (or to other property).
Indicates a mandatory action.
Indicates a prohibited action.
Inspection
Inspect the device for damage due to storage or transport and
verify operation before use. If you nd any damage, contact your authorized Hioki distributor or reseller.
Troubleshooting
If the device seems to be malfunctioning, contact your
authorized Hioki distributor or reseller.
Cleaning
CAUTION
If the device becomes dirty, wipe the device clean
with a soft cloth moistened with water or a neutral detergent.
Never use solvents such as benzene, alcohol, acetone,
ether, ketone, thinners or gasoline. Doing so could deform and discolor the instrument.
Focus magnication adjustment dial
11
Digital microscope*
22
Air supply ports (×8)
33
USB cable
44
Vent
55
Transparent protective cover
66
The serial number consists of nine digits. The rst two
77
digits indicate the year of manufacture, while the second two digits indicate the month of manufacture. Do not remove this sticker as the number is important.
Polarizing dial (adjustment unnecessary for this application)
88
Clip adjusting screw (adjustment unnecessary for this
99
application)
9
Symbol for a standard
Indicates that the product complies with standards imposed by EU directives.
* You can download specialized software from the digital
microscope. The specialized software has a built-in image zoom
function. For more information, visit the following website.
https://www.dino-lite.com/download.php
Specications
HIOKI RM9006A961-00
Operating environment
Operating temperature and humidity range
Storage temperature and humidity range
Model on which the
device is mountable
Dimensions Approx. 110W × 91.5D × 135H mm
Weight Approx. 480 g (16.9 oz.)
Product warranty duration
Accessories Instruction Manual
Dimensions of air nozzles that can be inserted into the air supply ports
Consumption current 0.5 A (at a voltage of 5 V)
Interface USB 2.0
Operating environment (computer)
Indoor use, pollution degree 2, altitude up to 2000 m (6562 ft.)
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensing)
0°C to 50°C (32°F to 122°F), 80% RH or less (non-condensing)
RM9004 only
(4.33″W × 3.60″D × 5.31″H)
1 year
Blower
Cleaning lm
Nozzle outer diameter: 1 mm to 9 mm
Nozzle length: 40 mm or more
Windows 10, Windows 8, Windows 7
Basic Operation
CAUTION
Do not clean probes with liquid.
If the probe tips are cleaned with a liquid such as alcohol or water, the electrode material that has been attached to probes could melt, penetrate the gaps and solidify, preventing probes from moving.
Do not use any tool not specied by Hioki that
makes direct contact with the probe.
Doing so could cause damage to probes.
Wear a dust mask and protective goggles during
operation.
Failure to do so could cause splash of air to get into the eye or mouth.
Limit the projecting pressure of the air to 207 kPa
(2.1 kgf/cm2 or 30 psi).
Otherwise, the operator could su󰀨er from air embolism.
Check that the locking lever of the RM9004 Test
1
Fixture is lowered.
Disconnect the RM9005 Connection Cable from the
2
RM9004.
Disconnect the grounding cable from the RM9004’s
3
ground terminal.
Turn the test xture xing knob counterclockwise
4
to disengage the lock.
At this time, while pressing down the RM9004, turn the xing
knob on the Test Fixture to the UNLOCK side until the screw is completely removed.
The bottom of the RM9004 faces slowly.
Pull the RM9004’s maintenance plunger outward to
5
secure the RM9004 in place.
Test-xture xing knob
Test-xture
locking lever
Probe guard
Probes
11
66
44
55
Plunger
RM9004 Test Fixture
2, 32, 3
Ground terminal
RM9005 Connection Cable
Grounding cable
Hang the device on the probe guard.
6
Connect the device’s USB cable to your computer.
7
Use the computer’s camera function to display the
8
probe tips of the RM9004 on the display.
Rotate the focus magnication adjustment dial to focus on
and observe the tips of the probes. If there are multiple cameras connected to the computer, switch the cameras so that the image from the device’s camera is displayed.
Turn the digital microscope so that the TP6 probe is visible in the position shown below.
TP6
Try to eliminate deposits on the tips of the probes
9
by injecting air while checking the camera image.
Insert the blower’s nozzle all the way into the vent before
supplying air.
While observing deposits, spray air from multiple supply ports.
Operating precautions
• To ensure proper operation of the RM9004 over a long period of time, remove any probe deposits after you have
nished working.
• Remove any deposits as soon as possible. Repeated measurements while there are deposits on probes will cause the deposits to settle between the probes, making
it di󰀩cult to remove with air. Moreover, an error may be
displayed during the open and short inspections of the RM2610.
If deposits cannot be removed, have the text xture repaired.
• If any objects adhere to the transparent protective cover of the device, remove them with air or a soft, dry cloth.
Cleaning lm
Removing insulation at the tips of the probes using the
included cleaning lm may cause the electrical contact to resume. Use the cleaning lm only when the short test of
the RM2610 yields fail judgments frequently. The abrasive
contained in the cleaning lm could scratch the plating on the surface of the probes, causing it to oxidize, resulting in more
liable to yield fail judgment during the short inspection.
How to use
Place the cleaning lm on the measurement stage
1
with the non-glossy side facing up.
Move the test xture up or down several times and
2
allow the probes to make contact with the cleaning
lm.
The tips of the probes are polished, resulting in better contact.
Replace the lm with a new one when it comes ine󰀨ective.
• If the short inspection of the RM2610 yields fail judgment
frequently, even when a cleaning lm is used, have the
RM9004 repaired.
You will need:
• Air
(e.g., air duster spray, air gun, blower included accessories) Keep in mind that if the air contains oil or other foreign objects, it could settle on the RM9004.
• Nozzle outer diameter: 1 mm to 9 mm
• Nozzle length: 40 mm or more
• Computer
Prepare a computer that matches the operating environment of the device.
Check out the video on basic operation here.
https://www.hioki.com/global/products/ resistance-meters/resistance/id_6740
The customer is responsible for carrier charges.
Cleaning lm