Hioki RM9004 Instruction Manual

RM9004
TEST FIXTURE
Instruction Manual
July 2019 Edition 1 RM9004A961-00 19-07H
EN
Introduction
Thank you for choosing the Hioki RM9004 Test Fixture. Preserve this manual carefully and keep it handy to make full use of this device for a long time. Familiarize yourself with the RM2610 Electrode Resistance Measurement System Instruction Manual and the separate document entitled “Operating Precautions” before using the device.
Target audience
This manual has been written for use by individuals who use the product in question or who teach others to do so. It is assumed that the reader possesses basic electrical knowledge (equivalent to that of someone who graduated from the electrical program at a technical high school).
Safety notations
This manual classies seriousness of risks and hazard levels
as described below.
Indicates a potentially hazardous situation that
CAUTION
IMPORTANT
may result in minor or moderate injury to the operator or damage to the device or malfunction.
Indicates information or content that is particularly important from the standpoint of operating or maintaining the device.
Indicates an action that must not be performed.
Indicates an action that must be performed.
Operating Precautions
Follow these precautions to ensure safe operation and to
obtain the full benets of the various functions.
CAUTION
Do not place the device on an unstable or uneven surface. Doing so could cause the device to fall or turn over, causing bodily injury or damage to the device.
Do not touch the tips of the probes. When an object other than the measurement target comes into contact with probes, they may be damaged.
IMPORTANT
• The RM9004 Test Fixture’s probes are consumable parts. They require periodic replacement but are not customer­replaceable. Please contact your authorized Hioki distributor or reseller for more information.
• Always use the test xture case when storing or
transporting the test xture.
Overview
The RM9004 Test Fixture is designed exclusively for use with the Electrode Resistance Measurement System. It should be
afxed to the RM9003 Press Unit prior to use.
Specications
Operating environment
Operating temperature and humidity
Storage temperature and humidity
Connector Test xture terminal (Connect to Model
Dimensions Approx. 130W × 100D × 890H mm
Mass Approx. 770 g (27.2 oz.)
Number of probes 46
Probe (buckling probe)
Probe distance 120 µm
Product warranty period
Accessories Instruction manual
Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
23°C±5°C (73°F±9°F), 80% RH or less (no condensation)
0°C to 50°C (32°F to 122°F), 80% RH or less (no condensation)
RM2611 by using the RM9005 Connection Cable)
(5.12″W × 3.94″D × 35.04″H)
Tip protrusion: Approx. 250 µm
Probe diameter (reference value) : φ75 µm
Tip shape: needle, tip SR
Probe load (reference value) : 0.044 N/per
probe Material: tool steel, gold coating
3 years Probes are considered consumables and as such are not covered by the product warranty.
Test xture case
Mounting screws × 4 Probe inspection board
Inspection
Check if there is any damage to the device occurred during storage or shipping and verify that device operates normally before using it. See the RM2610 Electrode Resistance Measurement System Instruction Manual for how to inspect using the included probe inspection board.
Troubleshooting
If damage is suspected, read the “Before having your product repaired” section of the RM2610 Electrode Resistance Measurement System Instruction Manual before contacting your authorized Hioki distributor or reseller.
Precautions when transporting the device
Be sure to follow these precautions.
• To avoid damage to the device, remove the RM2611 Electrode Resistance Meter and RM9003 Press Unit from the device. Additionally, use the packaging in which the device was packed when you purchased it, and be sure to doublebox it. Hioki cannot guarantee that the device will not be damaged during transport.
• Attach a description of the issue when sending out your device for repair.
See the RM2610 Electrode Resistance Measurement System Instruction Manual for connector pin and measurement probe assignments.
1 2 3 4
Assembly (RM9003, RM9004, RM9005)
Part Names
Cleaning
CAUTION
To avoid equipment damage, turn off the RM2611 Electrode Resistance Meter before connecting or disconnecting the RM9005 Connection Cable.
Preparations
Remove the four mounting screws from the top of the
RM9004 Test Fixture and remove the test xture from the test xture case. After removing the test xture from the case, do not place it
on a surface with the probe surface facing down.
Assembly
Place the RM9003 Press Unit’s up/down lever in the
1
raised position.
Slide the RM9004 Test Fixture into position along
2
the RM9003 Press Unit’s test xture insertion rails.
Push the test xture toward the rear until it won’t go
any further.
Pull the test xture lock lever toward you and
3
downward and then let go to lock.
The RM9004 Test Fixture will lock in place.
Connect the grounding cable on the rear of the
4
RM9003 Press Unit to the ground terminal on the RM9004 Test Fixture.
Connect the connector on the RM9004 (the female
5
side) and connector on the RM2611 (the male side) using the RM9005 Connection Cable and screws that hold it in place.
Secure the RM9005 Connection Cable in place with
6
the cable clamp.
There are screws for attaching the cable clamp on the left and right sides of the RM9003 Press Unit. Choose the appropriate side based on the position of the device.
IMPORTANT
When connecting the RM9005 Connection Cable, insert the
cable into the connector rmly and tighten it in place with screws.
If the screws loosen, the resulting poor contact could cause a measurement error.
• Exercise care not to lose the test xture case’s mounting screws,
which you will need in order to store the RM9004 Test Fixture during transport.
Up/down lever
1
RM9003 Press Unit
Test xture
insertion rail
2
RM9004 Test Fixture
6
Clamp
Test xture lock lever
Test xture insertion rail
3
4
Ground terminal
Front and top of the RM9004
(Top)
Mounting screws × 4
Test xture
Case
(Front)
Remove the mounting screw found at each of the four corners on top
of the xture and remove the test xture from its case.
Bottom and rear of the RM9004
Connector
(Rear)
Ground terminal
Probe guard
Probe
(Bottom)
CAUTION
• Use a compressed air pressure of 207 kPa (2.1 kgf/ cm2, 30 psi) or less.
• Wear a dustproof mask and dustproof goggles while cleaning the device.
IMPORTANT
The probes are extremely thin. Do not clean them with anything
other than air. Tighten the test xture clamp knob securely so that
it is not loose.
Cleaning the probe
Verify that the test xture lock lever is in the
1
lowered position.
Rotate the test xture clamp knob counterclockwise
2
to disengage the lock.
While doing so, rotate the test xture clamp knob
toward the UNLOCK position until the screw is fully removed while holding the RM9004 Test Fixture still.
The base (bottom) of the RM9004 Test Fixture will gradually move so that it faces toward the front of the unit.
Once the base of the test xture is fully facing
3
toward the front, pull out the maintenance plunger
to lock the RM9004 Test Fixture in place.
Remove dust from the probes using the
4
compressed air gun or blower.
Repeat the above procedure in reverse after
5
cleaning to return the test xture to its original
state.
Test xture clamp knob
2
1
Grounding cable
Test xture lock lever
5
Screw
RM9005 Connection Cable
4
1 2 3 4
Probe
3
Maintenance plunger