Hioki RM9003 Instruction Manual

RM9003
PRESS UNIT
Instruction Manual
July 2019 Edition 1 RM9003A961-00 19-07H
EN
Introduction
Thank you for choosing the Hioki RM9003 Press Unit. Preserve this manual carefully and keep it handy to make full use of this device for a long time. Familiarize yourself with the RM2610 Electrode Resistance Measurement System Instruction Manual and the separate document entitled “Operating Precautions” before using the device.
Target audience
This manual has been written for use by individuals who use the product in question or who teach others to do so. It is assumed that the reader possesses basic electrical knowledge (equivalent to that of someone who graduated from the electrical program at a technical high school).
Safety notations
This manual classies seriousness of risks and hazard levels
as described below.
Indicates a potentially hazardous situation that
CAUTION
IMPORTANT
may result in minor or moderate injury to the operator or damage to the device or malfunction.
Indicates information or content that is particularly important from the standpoint of operating or maintaining the device.
Indicates an action that must be performed.
Inspection
Check if there is any damage to the device occurred during storage or shipping and verify that device operates normally before using it.
Troubleshooting
If damage is suspected, read the “Before having your product repaired” section of the RM2610 Electrode Resistance Measurement System Instruction Manual before contacting your authorized Hioki distributor or reseller.
Cleaning
To clean the device, wipe it gently with a soft cloth. Do not wipe moving parts such as the shaft or up/down lever as they are coated with oil.
Operating Precautions
Follow these precautions to ensure safe operation and to
obtain the full benets of the various functions.
CAUTION
Do not place the device on an unstable or uneven surface. Doing so could cause the device to fall or turn over, causing bodily injury or damage to the device.
Specications
Operating environment
Operating temperature and humidity
Storage temperature and humidity
Dimensions Approx. 230W × 325D × 370H mm (9.06″W ×
Measuring object size
Up/down stroke Approx. 27.5 mm
Measurement stage Resin
Mass Approx. 8.5 kg (299.8 oz.)
Operation method The test xture is raised and lowered
Lowering mechanism
Tilt mechanism This mechanism orients the tips of the test
Connecting and disconnecting the
test xture
Locking the test
xture
Up/down lock function
Product warranty period
Accessories Instruction manual
Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
0°C to 50°C (32°F to 122°F), 80% RH or less (no condensation)
12.80″D × 14.57″H)
(excluding protrusions) Distance from the probe contact position to
the press unit support: approx. 144 mm (5.67″)
148 mm × 210 mm (A5 size) Thickness: up to 10 mm
graduated plate (10 mm, 5 mm)
Center line in the middle
manually. The test xture descends under its
own weight until it comes into contact with the electrode sheet.
The downward speed is cushioned by a damper. Time to lowered position: 1 s to 4 s (within the operating temperature range)
xture’s probes so that they face the operator
(in order to improve probe maintainability). The mechanism can be locked in the
measurement orientation with the test xture
clamp knob.
The test xture can be slid in and out. No set
screws are used.
Test xture lock lever
• Raising the test xture engages the lock.
• A screw holds the lock mechanism in place (for use when transporting the system).
• A slide disengages the lock function (to improve convenience during use).
3 years
Quick manual
Overview
The RM9003 Press Unit, which is designed exclusively for use with the Electrode Resistance Measurement System,
is used after being afxed to the RM9004 Test Fixture. It
serves to place the probes in contact with the measurement surface on the object under test (the electrode sheet). A tilt mechanism improves maintainability, while an up/down lock mechanism improves safety.
Part Names
1
2
3
4
5
6
Up/down lever
1
Test xture clamp knob
2
Test xture insertion rail
3
Test xture lock lever
4
RM9004 Test Fixture
5
Measurement stage
6
Lock release lever
7
Lock release lever clamp knob
8
Lock disable pin
9
Clamp
10
Serial number
11
(The serial number consists of 9 digits. The rst two [from
the left] indicate the year of manufacture, and the next two indicate the month of manufacture. Required for product control. Do not peel off the label.)
Shaft
12
(Do not touch as the shaft is coated with lubricating oil.)
Maintenance plunger
13
Graduated plate
14
7
8
9
10
11
12
13
14
1 2 3 4
Assembly (RM9003, RM9004, RM9005)
CAUTION
To avoid equipment damage, turn off the RM2611 Electrode Resistance Meter before connecting or disconnecting the RM9005 Connection Cable.
IMPORTANT
The device ships with the up/down lever locked with the lock release lever clamp knob. Loosen the knob before using the
device for the rst time.
Place the RM9003 Press Unit’s up/down lever in the
1
raised position.
Slide the RM9004 Test Fixture into position along
2
the RM9003 Press Unit’s test xture insertion rails.
Push the test xture toward the rear until it won’t go
any further.
Pull the test xture lock lever toward you and
3
downward and then let go to lock.
The RM9004 Test Fixture will lock in place.
Connect the grounding cable on the rear of the
4
RM9003 Press Unit to the ground terminal on the RM9004 Test Fixture.
Connect the connector on the RM9004 (the female
5
side) and connector on the RM2611 (the male side) using the RM9005 Connection Cable and screws that hold it in place.
Secure the RM9005 Connection Cable in place with
6
the cable clamp.
There are screws for attaching the cable clamp on the left and right sides of the RM9003 Press Unit. Choose the appropriate side based on the position of the device.
IMPORTANT
When connecting the RM9005 Connection Cable, insert the
cable into the connector rmly and tighten it in place with screws.
If the screws loosen, the resulting poor contact could cause a measurement error.
• Exercise care not to lose the test xture case’s mounting screws, which you will need in order to store the RM9004 Test Fixture during transport.
Test xture
insertion rail
RM9004 Test Fixture
Up/down lever
1
RM9003 Press Unit
2
3
Test xture lock lever
Test xture insertion rail
Basic Operation
IMPORTANT
• Exercise caution not to pinch your ngers or other body parts when lowering the RM9004 Test Fixture.
• Do not touch the tips of the probes when handling the electrode sheet.
Place an electrode sheet on the measurement stage.
1
The probes will make contact with the bold graduated marks.
Pull the up/down lever toward you and down while
2
pulling the lock release lever toward you.
The RM9004 Test Fixture will move downward under its own weight.
Once the RM9004 Test Fixture is fully lowered, start
3
measurement from the main screen of RM2612 Resistance Calculation Software.
See the RM2610 Electrode Resistance Measurement System Instruction Manual
Once measurement completes, raise the up/down
4
lever.
Remove the electrode sheet from the measurement
5
stage after verifying that the RM9004 Test Fixture is fully raised.
4
2
Up/down lever
Lock release lever
Precautions when transporting the device
Be sure to follow these precautions.
• To avoid damage to the device, remove the RM9004 Test Fixture and RM9005 Connection Cable from the device. Additionally, use the packaging in which the device was packed when you purchased it, and be sure to double­box it. Hioki cannot guarantee that the device will not be damaged during transport.
• Attach a description of the issue when sending out your device for repair.
• Raise the up/down lever and then tighten the lock release lever clamp knob to ensure that the press unit does not move up or down during transport.
Lock release lever clamp knob
6
Clamp
5
Screw
RM9005 Connection Cable
4
Ground terminal
Grounding cable
3
5
Measure
PC
1
RM2611 Electrode Resistance Meter
1 2 3 4
Loading...