Change detailed settings
(measurement conditions and system-related
settings)
Setting judgment criteria
(p. 36)
Transmitting data (p. 87)Printing (p. 89)
Computer communications
(p. 109)
External Control (p. 95)
Enable statistical
calculation (p. 84)
Instrument interface settings must
be configured before printing, using
communications or external control.
Turning Power Off (p. 25)
Be sure to read the "Operating Precautions" (p. 8) before use.
Installing, Connecting and Turning On
3
Measurement Flow
Settings of RM3542A
Calculation, Printing, Communication, and External Control Settings
When Finished
4
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it
fails to operate according to the specifications, contact your authorized Hioki distributor or reseller.
Thank you for purchasing the Hioki Model RM3542A (RM3542-50/RM3542-51) Resistance
Meter.
To obtain maximum performance from the instrument, please read this manual first, and keep
it handy for future reference.
Model RM3542-51 includes the GP-IB interface.
Registered trademarks
• Windows and Visual Basic are either registered trademarks of Microsoft Corporation in the
United States and other countries.
• Teflon is a registered trademark of E. I. du Pont de Nemours and Company.
Verifying Package Contents
Inspection
Content confirmation
Options
Measurement Probes and Fixtures
(connect to measurement jacks)
Model 9140-10 4-terminal Probe
Model 9262 Test Fixture
Model 9263 SMD Test Fixture
Model IM9100 SMD Test Fixture
Alligator-clip-type measurement probes.
These general-purpose dual-electrode clips fit
a wide range of conductor thicknesses.
Measurable terminal diameter:
0.3 mm to 5 mm
This fixture is for measuring
lead components.
Measurable sample dimension:
Lead diameter:
φ0.3 mm to φ2 mm
Lead pitch: 5 mm or more
(less than 20 mΩ residual resistance after
zero adjustment)
This fixture is for measuring
chip components.
Measurable sample dimension:
Sample width: 1 mm to 10 mm
(less than 20 mΩ residual resistance after
zero adjustment)
This fixture is for measuring ultra-small
SMD components.
Measurable sample dimension:
JIS(EIA): Approx. L mm x W mm
0402 (01005): 0.4 mm x 0.2 mm
0603 (0201): 0.6 mm x 0.3 mm
1005 (0402): 1.0 mm x 0.5 mm
(less than 100 mΩ residual resistance after zero
adjustment)
Interface Cables
Model 9637 RS-232C Cable
(9-pin to 9-pin/crossover cable/1.8
m)
Model 9638 RS-232C Cable
(9-pin to 25-pin/crossover cable/1.8
m)
Model 9151-02 GP-IB Connector
Cable (2 m)
Contact your authorized Hioki distributor or reseller for details.
5
Verifying Package Contents
Precautions during shipping
During shipment of the instrument, handle it carefully so that it is not damaged due to a vibration or shock.
6
Safety Information
Safety Information
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not
described in this manual may negate the provided safety features.
Before using the instrument, be certain to carefully read the following safety notes:
If persons unfamiliar with electricity measuring instrument are to use the instrument, another
person familiar with such instruments must supervise operations.
Mishandling during use could damage to the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
Marks on This Instrument
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to a corresponding topic in the Instruction Manual.
Indicates AC (Alternating Current).
Alarm Symbols
In this document, the risk seriousness and the hazard levels are classified as follows.
Indicates an imminently hazardous situation that will result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator
or damage to the instrument or malfunction.
Indicates advisory items related to performance or correct operation of the instrument.
Symbols for Various Standards
This symbol indicates that the product conforms to regulations set out by the EC Directive.
This symbol indicates laws and regulations regarding the disposal of electrical and electronic appliances in the Member States of EU (WEEE directive).
Other Symbols
Indicates the prohibited action.
7
Safety Information
(p. )
*
[ ]
SET
(Bold characters)
Unless otherwise specified, Windows represents Windows 95, 98, Me, Windows NT4.0, Windows 2000, Windows XP, or
Windows Vista.
Click: Press and quickly release the left button of the mouse.
Double click: Quickly click the left button of the mouse twice.
Indicates the location of reference information.
Indicates that descriptive information is provided below.
Bold characters within the text indicate operating key labels.
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the
following meanings:
f.s. (maximum display value)
The maximum displayable value. This is usually the name of the currently selected range.
rdg.(reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
dgt.(resolution)
The smallest displayable unit on a digital measuring instrument, i.e. a "1" as the least-significant digit.
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
• Using a measuring instrument in an environment designated with a higher-num-
bered category than that for which the instrument is rated could result in a severe
accident, and must be carefully avoided.
• Never use a measuring instrument that lacks category labeling in a CAT II to CAT IV
measurement environment. Doing so could result in a serious accident.
When directly measuring the electrical outlet receptacles of the primary electrical circuits
CAT II:
CAT III:
CAT IV:
in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
When measuring the primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets
When measuring the circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution panel)
8
The instrument can be used with the stand (p. 16) in the upright position.
It can also be rack-mounted (p. A8).
50 mm or more
10 mm or more
Rear Panel
50 mm or more
Unplugging the power cord kills power to the instrument. Be sure to provide enough
unobstructed space to unplug the power cord immediately in an emergency.
Operating Precautions
Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various
functions.
Preliminary Checks
Before using the instrument for the first time, verify that it operates normally to ensure that no
damage occurred during storage or shipping. If you find any damage, contact your authorized
Hioki distributor or reseller.
Instrument Installation
Installing the instrument in inappropriate locations may cause a malfunction of instrument or may give rise to an accident. Avoid the following locations:
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic field or electrostatic charge
• Near induction heating systems (such as high-frequency induction heating systems
and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
Installing
• To prevent overheating, be sure to leave the specified clearances around the instrument.
• Install the instrument with the bottom facing down.
Do not place the instrument on an unstable table or an inclined place. Dropping or knocking
down the instrument can cause injury or damage to the instrument.
Handling the Instrument
To avoid damage to the instrument, protect it from physical shock when transporting and
handling. Be especially careful to avoid physical shock from dropping.
This instrument may cause interference if used in residential areas. Such use must be
avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
Handling the Fixture
Before using a test fixture, read the instructions provided with it.
Before Turning Power On
• Before turning the instrument on, make sure the supply voltage matches the voltage
indicated on its power connector. Connection to an improper supply voltage may
damage the instrument and present an electrical hazard.
• To avoid electrical accidents and to maintain the safety specifications of this instru-
ment, connect the power cord provided only to an outlet.
9
Operating Precautions
Avoid using an uninterruptible power supply (UPS) or DC/AC inverter with rectangular wave
or pseudo-sine-wave output to power the instrument. Doing so may damage the instrument.
Before Connecting EXT. I/O Connector
To avoid electric shock or damage to the equipment, always observe the following
precautions when connecting to the EXT. I/O connectors:
• Always turn off the power to the instrument and to any devices to be connected
before making connections.
• During operation, a wire becoming dislocated and contacting another conductive
object can be serious hazard. Use screws to secure the EXT. I/O connectors.
To avoid damage to the instrument, always observe the following precautions when
connecting to the EXT. I/O connector.
• Do not apply voltage or current to the EXT. I/O terminals that exceeds their ratings
(p. 103).
• Ensure that devices and systems to be connected to the EXT. I/O terminals are
properly isolated.
• When driving relays, be sure to install diodes to absorb counter-electromotive
force.
• The IISO_5 V pin of the EXT. I/O connector is a 5 V power output. Do not apply exter-
nal power to this pin. Be careful not to short-circuit ISO_5 V to ISO_COM.
• The IISO_12 V pin of the EXT. I/O connector is a 12 V power output. Do not apply
external power to this pin. Be careful not to short-circuit ISO_12 V to ISO_COM.
See "Connector Type and Signal Pinouts" (p. 96).
10
Operating Precautions
Before Connecting to the RS-232C Connector or SET MONITOR Connector
• Use a common ground for both the instrument and connected device.
Using different ground circuits will result in a potential difference between the instrument's
ground and the connected device.
If the communications cable is connected while such a potential difference exists, it may
result in equipment malfunction or failure.
• Before connecting or disconnecting any the communications cable, always turn off the
instrument and the connected device. Failure to do so could result in equipment malfunction or damage.
• After connecting the communications cable, tighten the screws on the connector securely.
Failure to secure the connector could result in equipment malfunction or damage.
Before Measuring
To avoid electrical hazards and damage to the instrument, do not apply voltage
exceeding the rated maximum to the EXT. I/O connectors.
• Do not apply any voltage to the measurement jacks. Doing so may damage the unit.
• Never attempt to measure at a point where voltage is present. In particular, do not measure a transformer or motor immediately after a temperature increase test or withstandvoltage test, as the instrument could be damaged by induced voltage or residual charge.
• Battery internal resistance cannot be measured with this instrument. It will sustain damage.
To measure the battery internal resistance, we recommend the Hioki 3554, 3555, BT3562,
BT3563 and 3561 Battery HiTesters.
• To obtain the guaranteed measurement accuracy, allow at least 30 minutes warm-up.
• The instrument internally retains all settings (but not measurement values), such as measurement range and comparator settings. However, measurement settings made through
the RS-232C or GP-IB interface are not memorized.
• In the 100
occur due to the influence of thermo electromotive force.
• The DC resistance of a power transformer cannot be measured. When measuring objects
with a large L, such as choke coils and other inductors, measurement values may be
unstable. In such cases, contact your authorized Hioki distributor or reseller.
• Carefully insulate all H
cannot be performed and an error will occur if the core and shield wires touch.
Ω or higher ranges (LOW POWER: OFF setting), a measurement error may
, H
, L
CUR
POT
POT
and L
wiring. Proper 4-terminal measurements
CUR
1.1 Product Overview and Features
This instrument employs the 4-terminal method to quickly and accurately measure the DC resistance of
components, such as resistors and ferrite beads.
It includes advanced contact-check, comparator, and data output functions. The intuitive user interface
and high noise immunity are ideal for use with taping machines and separators.
Resistance Measurement
The factory defaults (initial settings)
are optimized for chip-component resistance measurements. It can also
measure devices that are otherwise
difficult to measure with a high current, such as ferrite bead and small
multilayer inductors (low-power resistance measurement) (p. 28).
It is also suitable for measuring imperial 008004 sized components with
small rated voltage (Applied Voltage
Limit Function) (p. 64)
Interface Communications
Connect the instrument to a controller via the RS-232C or GP-IB
interface to control this instrument or acquire the measurement data
(p. 109)
Send the measurement value
and calculation results to the
printer.
Use a commercially available printer with a serial interface to print the
measurement values and calculation results. (p. 89)
Save and Output the Measurement Values
Measurement values can be stored
in the internal memory (p. 79).
Statistical calculations can be performed on the stored data, which
can be transferred to a computer in
batch form (however, stored data
cannot be confirmed internally).
Judge the Measurement Values
Measurement values are compared
with a pre-specified reference value
or range, and the result is outputted
externally and indicated by the
COMP indicators (comparator function) (p. 36)
Connect a PLC or I/O Board
By connecting to the EXT. I/O connector, it is possible to control the
instrument from a PLC. In addition
to the comparator results, various
measurement anomaly signals
can be outputted. (p. 95)
Upper limit
Lower limit
When using two instruments, a difference in settings disables measurement and causes a warning
notification (Setting Monitor Function)
(p. 59).
Optional Hioki probes and fixtures are
available to connect to the measurement jacks
(BNC jacks
(p. 5)).
Alternatively,
commercially
available cables,
such as the 1.5D2 V coax, can be
used (p. 24)
OverviewChapter 1
1.1Product Overview and Features
11
1
12
Ultra Fast and Accurate Measurements
Increase Productivity
The factory default settings are optimized for chip-component resistance measurements. Enhanced contactto-measurement and contact check-to-determination,
within 1 ms.
When using the low-power resistance measurement
and the 100 mΩ to 10 Ω ranges, the offset-voltage compensation (OVC) function minimizes the effects of thermal emf (p. 65).
Because measurement results are judged as pass/fail
with a 10 ppm resolution, it is ideal for high-speed Class
B resistor testing.
High-Speed Data Output and Ample Memory
The Data Output function transfers measured data at
5 ms/sample, even via RS-232C.
Up to 30,000 measurements can be stored, and all
data can be exported at the end of measuring each
reel.
This function is ideal for system setup, debugging and
process management.
Multiple Interfaces
EXT. I/O is a noise proofed structure isolated from the
measurement and control circuits (p. 95).
All data can be acquired in real time using the built-in
38.4-kbps high-speed RS-232C interface.
Connect the commercially available printer with a serial interface to print the measurement values and statistical calculation results (p. 89)
The GP-IB interface can also be used for Model
RM3542-51 (specified when shipping (p. 109)).
Low-Power Function (p. 28)
For ranges from 1000 mΩ to 1000 Ω, the low-power
resistance measurement is provided to minimize the
measurement current. Safely measure devices that
are otherwise difficult to measure with high current,
such as ferrite-bead and multilayer inductors.
Clearly Visible Display and Intuitive Operation
The high-contrast LCD provides clear visibility, helping to avoid setting mistakes. The optimum range is
selected automatically when comparator thresholds
are entered.
The Auto Memory Function Is Convenient
for Sampling Tests(p. 81)
The auto memory function is convenient for sampling
tests after screen-printing.
When the measurement values become stable, the
measurement value is automatically acquired and statistical calculations are performed at the same time.
The beeper gives a notification when the specified
number of values are stored.
Selecting [PRINT] (screen display) prints the measurement values and statistical calculation results(p.
93).
Fixtures for Component Measurements (p. 5)
The BNC-type measurement jacks exhibit good noise
immunity.
Ready availability and easy assembly ensure a
smooth system setup.
Various test fixtures are available for Hioki LCR HiTesters.
Features
1.1 Product Overview and Features
Reliable Contact Checks (p. 50)
Contact checking (that was previously performed before and after measuring) is now performed during
measurement, so probe bounce and contact resistance
fluctuations can be detected. Contact checking time
can be shortened, improving tact times.
Features
Measurement Circuit Strongly Immune to
Contact Resistance Fluctuations
The effects of contact resistance fluctuations are reduced even when scattering occurs near the end of
probe life. Such effects are minimized by the fast response of the measurement circuit.
Contact Improvement Function
(Contact Improver) (p. 52)
The Contact Improver function improves bad contacts
between probes and test samples. Contacts errors are
reduced by penetrating oxidation and impurities between probes and samples.
Reducing contact errors can increase productivity and
quality. The intensity of the contact improver function
can be adjusted according to probe type.
Reject Faulty Data - Voltage Level Monitor
Function (p. 55)
When the contact resistance of the H
CUR
and L
CUR
leads fluctuates, the measurement current changes
momentarily. Such momentary changes are not detectable by typical contact checking.
The Voltage Level Monitor Function indicates a contact
error if the detection voltage changes significantly. It
can enhance the reliability of the measurement value.
Minimize Human Error and Risk
- Settings Monitor Function (p. 59)
If the settings are found to be different after comparing
the setting conditions of two instruments, an alarm is
sounded to prohibit the TRIG input. Helps to prevent
human errors by avoiding setting mistakes.
Reliable Four-Terminal Measurement Probe Short-Circuit Detection Function (p.
57)
If a conductive foreign object is present between the
POT and CUR probe tips, the reliable four-terminal
measurements cannot be maintained. When not
measuring, resistance between the POT and CUR
probe tips is measured and short-circuit probe anomalies are detected.
Strong Electrical Noise Immunity
The specified measurement accuracy is achieved
even with a ±1.5 kV mixed pulse noise. The floating
measurement section design is highly impervious to
electrical noise, minimizing the effect on measurement values even when turning large-induction motors on and off.
The free-range power supply input (90 to 264 V) is
essentially unaffected by voltage fluctuations, so stable measurements are possible even under poor
power conditions.
DUT
Foreign Object
POT
CUR
DUT electrode
Previously
RM3542A
Contact Condition
Contact
Check
Measuring
Probe Bounce
Contact
Condition
Detection
Voltage
Error
Good Contact
Poor Contact
An error occurs
because the detection
voltage changes significantly.
Good Contact
ERROR
Measurement
Contact Check
Contact Condition
Contact Improvement
Function (Contact
Improver)
Contact Improvement
Measuring
Contact Check
ONON
13
1.1 Product Overview and Features
1
14
AB
C
D
E
F
G
H
I
E
D
1.1 Product Overview and Features
Block Diagram
• A constant current (determined by the measurement range) is applied between the H
voltage is measured between the H
voltage (B) by the constant current flow (A). (
POT
and L
terminals. The resistance value is obtained by dividing the measured
POT
A, B)
CUR
and L
terminals while
CUR
• The effects of a large offset voltage, such as from thermal emf, are reduced by reversing the current and measuring
twice in positive and negative directions (A). (A)
• The constant current source (A) and voltmeter (B) circuit designs are largely unaffected by contact resistance. (A, B)
• Faulty measurement values caused by unstable or chattering contact conditions can be eliminated by monitoring (C)
the detection voltage (B) waveform (voltage level monitor function). (B, C)
• The voltmeter is provided with sufficient time for integration (the default setting is 0.3 ms) to achieve stable measure-
ments. (The integration time can be reduced to 0.1 ms to support higher speeds.) (B)
• Before measuring, the Contact Improver circuit (D) optimizes the contact when the probes touch the DUT. (D)
• Also, performing contact checking (E) before measuring can detect short circuits between the CUR and POT terminals
caused by a clogged probe tip (probe short-circuit detection function). (
E)
• When measurement starts, the contact check circuit (E) and constant current monitor (F) are activated to monitor for
faulty conditions while measuring. (
E, F)
• The dual CPU (C and G) design provides ultra-high-speed measurements and a fast system response. (C, G)
• Protection from electrical noise is provided by the isolation between the Measurement and Control blocks (H). (H)
• The 90 V to 264 V wide range switching power supply (I) can provide stable measurements even in poor power quality
environments. (I)
15
Display Screen
Monochrome graphical LCD
There are three screen types:
Measurement, Basic Settings and
Detailed Settings.
Screen Configurations (p. 17)
F1 - F4 key
(F Key: General)
Selects the correspondingitem on the right sideof thedisplay.
Judging the Measurement Values
(p. 36)(Comparator Function)
Set a reference value and range for judging
the measurements, which can be confirmed
by the COMP indicators.
Select this to judge the measurement values relative to a reference
value and tolerance (%).The REF%
setting display appears.
Press this button to judge the measurement values relative to the setting range (absolute value).The
ABS setting display appears.
Setting
POWER Button
Turns the instrument on and
off. (p. 25)
• Off: power is off
(When power is not supplied)
• Red light: power is off
(when power is supplied)
• Green light: power on
Turning the Power
On and Off
Entering numerical values
Enter a numerical value.
(these are called the numeric keys)
Switches the sign of a numerical value.
Selects the unit of measure.
Sets the tolerance values.
Deletes the value in the selected field.
Accepts the displayed comparator threshold values.
Aborts comparator setting and
returns to the previous display.
Selecting the setting contents
(we call these "F keys")
Selects the corresponding
item on the right side of the
display.
Cursor keys
Move among the displayed
setting items.
The cursor location is indicated by reverse characters.
Measurement jacks
Connect measurement probes and fixtures (p. 23).
• H
CUR
jack: Current source terminal
• H
POT
jack: High voltage detection
terminal
• L
POT
jack: Low voltage detection
terminal
• L
CUR
jack: Current detection termi-
nal
• GUARD jack: Shield terminal
(for the measurement of GND)
See "Before Measuring" (p. 10)
Connecting Probes
COMP indicator
Displays Comparator
Results (p. 36).
Hi The measurement
value is above the
upper limit.
IN Pass (meets crite-
ria)
Lo The measurement
value is below the
lower limit.
Viewing the
Comparator
Results
Viewing the
Measurement Values
and Settings
Buzzer
beeps.
1.2 Names and Functions of the Parts
1.2Names and Functions of the Parts
Front Panel
1
16
Serial Number
Shows the serial number.Do not remove this
label, as it is required for product support.
Connect the supplied power cord (p. 22).
See "Before Turning Power On" (p. 9)
Connecting the Power Cord
RS-232C Connector
The RS-232C interface can be used to connect to
a PLC or computer (p. 109).
It is also used to connect to a commercially available printer with a serial interface for printing (p.
89).
RS-232C Communications
Printer Output
GP-IB Connector
(RM3542-51 only)
The GP-IB interface can be used to connect to a computer
(p. 109).
GP-IB Communications
SET MONITOR Connector
Connect another instrument to compare the settings of two units (p. 59).
See "Before Connecting to the RS-232C Con-
nector or SET MONITOR Connector" (p. 10)
Compare Two Instruments
EXT. I/O Connector
Connect to a PLC or I/O board to control
the measurement start time, and to acquire the comparator results (p. 95).
Extend the stand until it clicks into place.
Make sure to extend both legs of the stand.
Collapsing the stand
Fold in the stand until it clicks into place.
This instrument can be rack mounted.
See: Rack Mounting (p. A8)
Please retain the parts removed from this instrument to be used again.
1.2 Names and Functions of the Parts
Rear Panel
17
Detailed Settings Screen (p. 19)
Measurement Settings screens
[MEAS SETTINGS]
Data Memory Setting screen[DATA SETTINGS]
System screens[SYSTEM]
Normal display (measurement screen) (p. 18)
Basic Settings screen (p. 18)
Returns to previous screen.
Indicates a Continued Screen.
Returns to previous screen.
1.3 Screen Organization
1.3Screen Organization
The instrument has three general display screen types: Measurement, Basic Settings and Detailed
Settings.
See "11.3 Error Displays and Solutions" "Error Displays and Solutions" (p. 197) for error displays.
The screen examples in this manual are reversed (black on white) for the best visibility on the printed page. However, the
instrument screens can actually be displayed only with white characters on a black background.
1
18
Normal display (measurement screen)
This screen normally appears while measuring. View
currently measurement values and measurement
conditions.
Some parts of the display depend on the comparator
mode and other settings.
Measurement
value
Settings Menu (corresponding to the F keys)
Displayed contents depend on the current function settings.
Parentheses ( ) indicate the corresponding F key.
MENU (F1)Displays the Basic Settings screen.
PRINT (F2)Printing (p. 91)
Appears only when the interface is set
for the printer.
STAT (F3)Statistical calculation results (p. 84)
Appears only when statistical calculation is enabled
NUMBER (F4) Sets the Auto-Memory number (p. 81).
Appears only when the auto-memory
function is enabled
Displays the memory number, the number of passed and failed products, in
the lower left of the screen.
UNDO (F3)Deletes the previously stored measure-
ment and calculation result
(executes only once) (p. 86)
Appears only when the calculation result screen is displayed.
ALLCLR (F4)Clears all memory and calculations (p.
86).
Appears only when calculation results
are displayed.
LOCAL (F1)Cancels the remote status (p. 116).
UNLOCK (F1) Cancels the key-lock status (Hold for
one second) (p. 68).
Criteria Setting Values (p. 36)
Displayed contents depend on the
selected comparator mode.
(Ex.: REF% mode)
To display the Basic Settings screen
Measurement Conditions
Shows current setting contents. Displayed contents depend on
the current settings.
Appears only when the zero-adjustment is enabled (p. 34).
OVC/
OFF (no display)
(OVC: offset voltage compensation)
Appears only when the offset voltage
compensation is enabled (p. 65).
LP/
OFF (no display)
Appears only when the low-power
resistance measurement function
is enabled (p. 28).
S/ OFF (no display) Appears when scaling function is
enabled (p. 46)
NUMAppears only when numeric input is
enabled.
RMTRemote status(p. 116)
M.LOCKDisables all operations except the
comparator settings (p. 67).
F. LOC KDisables all operations including
the comparator settings (p. 67)
Basic Settings screen
Make basic measurement condition settings on this screen.
Measurement speed and range can be changed while
viewing the measurement values (when the Trigger
source is set to the internal trigger [INT]).
TRG
Changing the trigger source (start control
method) (p. 31).
When selecting [TRG: EXT], [MANU] appears by
pressing the F4 key (Measures manually once).
RANGEChange the range (p. 32).
SPEEDChange the measurement speed (p. 29).
0ADJZero-adjustment function ON/OFF (p. 34)
COMPComparator function ON/OFF(p. 36)
LOCKKey-lock function ON/OFF (p. 67)
Measurement condition
settings
Move with cursor keys.
Measurement condition
setting conditions
Return to previous screen
To the Measurement Settings screen [MEAS SETTINGS]
To the Data Memory Setting screen [DATA SETTINGS]
To the System screen [SYSTEM]
MISCTo display the Detailed Settings screen.
1.3 Screen Organization
Measurement Settings screens
[MEAS SETTINGS]
Set instrument system-related settings on this screen.
SET MONITORTwo units measurement condition comparison
ON/OFF(p. 59)
PROBE CHECK Probe short-circuit detection function ON/OFF
(p. 57)
RETRYRetry function setting (p. 62)
VOLT LIMITSets voltage limit function to ON/OFF (p. 64)
TRIG EDGE(EXT. I/O) Set the trigger rising/falling edge
Shows detailed settings for measurements.
Use when adjusting the measurement speed, stability and
measurement fault detection functions.
DELAY1Adjust the delay between the probing and
trigger input (p. 44).
DELAY2Adjust the target electrical response (p. 44).
SCALING
(A*R+B)
Compensate the measured value using scaling
function (p. 46)
INT(FAST/ MED/
SLOW)
Make fine adjustment to the integration time
(p. 48).
CONT CHECKContact check threshold setting (p. 50)
CONT IMPContact improver function setting (p. 52)
VOLT MONITOR Voltage level monitor function setting (p. 55)
CURRENT
MODE
Current mode setting (p. 56)
Data Memory Setting screen
(Save, analysis, output)
[DATA SETTINGS]
These are settings for memory and statistical calculation
functions.
AUTO MEMORY Auto-memory function ON/OFF (p. 81)
STATISTICSStatistical calculation ON/OFF (p. 84)
DATA OUTAuto-export the measurement values
(communication) ON/OFF (p. 87)
Detailed Settings
19
1.3 Screen Organization
1
20
1.3 Screen Organization
21
Installing this instrument (p. 8)
1
Rear Panel
Front Panel
Connecting Measurement Probes
(p. 23)
Turning Power On (p. 25)
5
Connect to the test sample.
After measurements are completed,
turn the power off. (p. 25)
Connect the external interface
(as needed)
4
• Using the printer (p. 89)
• Using RS-232C or GP-IB interface
(p. 109)
• Connecting to the PLC or I/O Board
(p. 95)
• Automatically comparing the settings of
two instruments
(Settings Monitor function) (p. 59)
2
3
2
3
4
5
Setting the Instrument (p. 27)
6
Connecting the Power Cord (p. 22)
Measurement
PreparationsChapter 2
Be sure to read the "Operating Precautions" (p.8) before installing and connecting this instrument.
See "Appendix 4 Rack Mounting" (p. A8) for rack mounting.
2
22
Rear Panel
1 Confirm that the supply voltage matches the instrument,
and connect the power cord to the power inlet on the instrument.
2 Connect the plug of the power cord to the outlet.
The POWER button on the front panel lights up in red.
In event of a power outage, operation resumes with the
same settings when power is restored (breaker reset, etc.).
Power inlet
2.1 Connecting the Power Cord
2.1Connecting the Power Cord
• Before turning the instrument on, make sure the supply voltage matches the voltage
indicated on its power connector. Connection to an improper supply voltage may
damage the instrument and present an electrical hazard.
• To avoid electrical accidents and to maintain the safety specifications of this instru-
ment, connect the power cord only to a 3-contact (two-conductor + ground) outlet.
• Before using the instrument, make sure that the insulation on the power cord is
undamaged and that no bare conductors are exposed. Any damage can cause electric shock, contact your dealer or Hioki representative.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the
power outlet.
2.2 Connecting Measurement Probes and Test Fixtures
Example of defeated guard measurement
A
Connecting measurement probes
Connecting a fixture
Connect it directly to the measurement
jacks with the label side facing up, and
fix it with the levers on the left and right
sides.
Connect the red plugs to the
H
CUR
and H
POT
jacks, and
the black plugs to the L
CUR
and L
POT
jacks.
Black plugs
Red plugs
Measurement jack
connector guide
Lock
Cable BNC
connector groove
2
1
Align the slots in the BNC plug
with the guide pins on the jack on
the instrument, then push and
twist the plug clockwise until it
locks.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-wise, and pull it out.
Making and extending your own probes (p. 24)
2.2Connecting Measurement Probes and Test
Fixtures
23
Connect your measurement probes, optional Hioki probes, or test fixtures to the measurement
jacks.
Refer to "Options" (p.5) for details of the Hioki option.
See the instructions provided with the fixture for operating details.
• Do not apply any voltage to the measurement jacks. Doing so may damage the unit.
• When disconnecting the BNC connector, be sure to release the lock before pulling off the
connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable,
can damage the connector.
• We recommend using optional Hioki fixtures.
• Use the GUARD jack only for the HighResistance Measurements shield, and
avoid more than a 10 mA current flow.
This jack is not for guarding the network
resistance measurements.
Connection Methods
2
24
Wiring Diagram
Before Wiring
• Twist the H
POT
and L
POT
wires, and the H
CUR
and L
CUR
wires together.
If not twisted together, measurement values may be unstable and errors may occur when measuring with
low-power resistance, or low resistance values.
• See the block diagram (p. 14) for the internal circuit details.
• Probes and measuring objects should be shielded at BNC or GUARD jack potential.
• Measurement probe length: keep it within 2 m (with a conductor resistance of 500 mΩ/m or less).
Long cables are more susceptible to noise, and the measurement values may be unstable.
• Extensions should maintain the four-terminal structure. If the wiring is converted to a two-terminal
structure in wiring, the correct measurement may not be possible due to the effects of the wiring and
contact resistance.
• Cables and measuring objects should be shielded.
• After extending the measurement probes, verify that the operation and accuracy conform to the
"Measurement Specifications" (p.192).
• When cutting off the ends of the optional measurement probes, make sure that the H
CUR
, H
POT
,
L
POT
, and L
CUR
shield wires and core wires do not come into contact. Such contact will made accu-
rate measurement impossible.
When Extending the Measurement Probes
Observe the following when extending the measurement probes.
2.2 Connecting Measurement Probes and Test Fixtures
Making Your Own Measurement Probes
Recommended measurement probe specifications
Conductor resistance500 mΩ/m or less
Capacitance150 pF/m or less
Cable dielectric material
Connector insulating material
Length2 m or less
Example: JIS standard 3C-2V, 1.5D-2V MIL standard RG-58A/U
*. Teflon is a registered trademark of E. I. du Pont de Nemours and Company.
Polyethylene (PE), Teflon* (TFE), Polyethylene foam (PEF)
Insulation resistance 10 GΩ or more
Te fl o n* (TFE), Polybutylene terephthalate (PBT)
Insulation resistance10 GΩ or more
2.3 Turning the Power On and Off
Press the POWER button (it lights up in green).
Self-test
Indicates an error (p. 197).
After Power-On
A self-test (instrument diagnostic routine) is performed.
During the self-test, the following information is displayed while the hardware is verified.
Error
No Errors
Normal display (measurement screen)
The following information is displayed during selftesting:
• Manufacturer and model name
• Firmware versions (main, and measurement)
• Communication setting
• Power Line Frequency Setting
2.3Turning the Power On and Off
Turning Power On
25
2
When the power is turned on, the same setting as when the power was last turned off
appears (backup function).
Turning Power Off
When powered up for the first time, the default settings appear.
See: "Default Settings" (p.76)
Before Starting Measurement
To obtain precise measurements, provide about 30 minutes for warm-up after turning the
power on.
Measurement settings are recalled from when the power was last turned off (settings
backup).
However, measurement settings made through the RS-232C or GP-IB interfaces are not
retained, although they can be stored using the
:SYSTem:BACKup command (p. 149).
Press the POWER button (it lights up in red, power OFF).
Disconnect the power cord from the power inlet to extinguish the POWER button light.
When the power is turned on again, operation resumes with the same settings as when last
turned off.
If a power outage (e.g., breaker trip) occurs when the instrument is turned on, it will automatically turn on again when the power is restored without pressing the POWER button.
26
2.3 Turning the Power On and Off
27
Do not use the instrument if damage is
found, as electric shock or short-circuit
accidents could result.
Contact your dealer or Hioki representative.
Metal Exposed
Is the power cord insulation torn, or is any
metal exposed?
1
No Metal Exposed
Before using the instrument for the first time, verify that it operates normally to ensure that
no damage occurred during storage or shipping. If you find any damage, contact your dealer
or Hioki representative.
Peripheral Device Inspection
Is the insulation on a measurement probe
torn, or is any metal exposed?
Metal Exposed
If there is any damage, measurement
values may be unstable and measurement errors may occur.
Replace the cable with an undamaged
one.
No Metal Exposed
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the self-test screen appear (model
name, version no.) (p. 25)?
Not displayed
2
The power cord may be damaged, or
the instrument may be damaged internally. Request repairs.
Does the Measurement screen appear after
self-test?
No
Displayed
The instrument may be damaged internally. Request repairs.
See: "11.1 Troubleshooting" (p. 195)
"11.3 Error Displays and Solutions" (p. 197)
An error is
displayed (Err)
Displayed
Inspection complete
Please read the "Operating Precautions" (p. 8) before use.
3.1 Pre-Operation Inspection
Measurement Settings
(Basic Measurements)
Chapter 3
Refer to "Measurement Flow" (p. 3) for an outline of the measurement process from preparation to
end-of-measurement.
3.1Pre-Operation Inspection
3
28
Measurement range: 0.0000 mΩ (100 mΩ range) to
120.0000 MΩ (16 ranges)
Normal Resistance Measurement
General-purpose resistors
Measurement range: 0.000 mΩ (1000 mΩ range) to
1200.000 Ω (6 ranges)
Low-power resistance measurement
Hard-to-measure components
such as ferrite bead or layered
inductors, or other elements
sensitive to measurement
current
LP
LP appears at the top of the screen.
The Basic Settings screen appears.
1
The System screen appears.
[SYSTEM]
Selection
2
Selection
1
2
Normal Resistance Measurement (default)
Low-power resistance measurement
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
3.2 Measuring Object Types
3.2Measuring Object Types
The instrument provides two measurement methods: resistance measurement, and low-power
resistance measurement. Select the appropriate measurement method for the type of component to
be measured. For general-purpose resistor measurements, use the factory defaults. The power
applied to the measuring object = Resistance Value x (Measurement Current)
See: "(6) The Sample Becomes Warm" (p. A6)
(Example)If the resistance to be measured is 100 Ω
= 100 µWLow-Power Resistance Measurement, 100 Ω Range
2
.
Ranges 1000 Ω range or higher (LOW POWER: OFF) cannot be used for inductor measurements.
Open the Basic Settings screen.
1
Open the System screen.
2
Select the low-power mode, as needed.
3
Return to the Measurement screen.
4
29
The Basic Settings screen appears.
1
Selection
2
Refers to table below
(default)
Press the up/down
cursor keys to change
the setting.
3.3 Setting the Measurement Speed
3.3Setting the Measurement Speed
The measurement speed can be set to FAST, MED (MEDIUM), or SLOW. A slower measurement
speed provides greater measurement precision.
A faster measurement speed results in greater susceptibility to environmental noise. Ensure that
measurement probes and the sample are sufficiently shielded.
Open the Basic Settings screen.
1
Select the measurement speed.
2
3
3
Return to the Measurement screen.
30
3.3 Setting the Measurement Speed
Relationship Between Measurement Range and Speed
(factory defaults)
Measurement
range
Ω
100 m
Ω
1000 m
3 Ω
10 Ω
100 Ω
300 Ω
1000 Ω
10 kΩ
30 kΩ
100 kΩ
300 kΩ
1000 kΩ
3 MΩ
10 MΩ
30 MΩ
100 MΩ
LOW POWER: OFFLOW POWER: ON
FASTMEDSLOWFASTMEDSLOW
3.8 ms13 ms
2.0 ms6.4 ms
1.6 ms6.0 ms
1.6 ms6.0 ms
0.9 ms3.6 ms
0.9 ms3.6 ms
0.9 ms3.6 ms
1.0 ms3.6 ms
0.9 ms3.6 ms
1.3 ms3.8 ms
1.3 ms3.8 ms
2.5 ms6.0 ms
2.5 ms6.0 ms
5.3 ms
5.8 ms
26 ms
22 ms
23 ms
20 ms
23 ms
20 ms
46 ms
39 ms
43 ms
36 ms
41 ms
35 ms
41 ms
34 ms
41 ms
34 ms
21 ms
17 ms
21 ms
17 ms
21 ms
17 ms
21 ms
17 ms
21 ms
17 ms
21 ms
18 ms
21 ms
18 ms
21 ms
18 ms
21 ms
18 ms
23 ms
20 ms
86 ms
72 ms
86 ms
72 ms
‑‑‑
2.3 ms12 ms
2.3 ms12 ms
2.3 ms12 ms
1.7 ms6.1 ms
3.2 ms7.6 ms
7.2 ms12 ms
‑‑‑
‑‑‑
‑‑‑
‑‑‑
‑‑‑
‑‑‑
‑‑‑
‑‑‑
‑‑‑
42 ms
35 ms
42 ms
35 ms
42 ms
35 ms
41 ms
34 ms
43 ms
36 ms
47 ms
40 ms
Integration time can be optionally
set for each range (p. 48).
• When internal triggering is selected, the EXT. I/O TRIG signal and the ∗TRG command are ignored
(except for memory storage and statistical calculations).
• To measure samples such as inductors that require time to settle, adjust delay time (DELAY2). Start with
a long delay, and gradually shorten it while watching for the measurement value to settle.
See: "4.2 Setting Pre-Measurement Delay" (p. 44)
• When external triggering [EXT] is enabled, the Auto-Memory function is forcibly disabled (OFF).
Trigger signals are automatically generated internally for continuous measurement.
Measure with internal [INT] triggering
To measure automatically
Measurements are triggered by an external signal. Manual measurement triggering is also available.
• Apply a trigger signal at the EXT. I/O connector (p. 95)
• Send the ∗TRG command by communications interface
(p. 138)
• Press F4 [MANU] (only appears when EXT is selected)
Measure with external [EXT] triggering
To measure at specific times
To retain measurement values
The Basic Settings screen appears.
1
Selection
2
Internal trigger
External trigger
(default)
Press F3 [EXT] to display the F4 [MANU] indicator.
Press to trigger a measurement manually.
Press the up/down
cursor keys to change
the setting.
Continuous measurement (:INITIATE:CONTINUOUS ON) is the normal trigger state when operating from
the front panel. Selecting the internal [INT] trigger source activates continuous triggering ("free-run"). When
external [EXT] triggering is selected, each external trigger event initiates one measurement.
Continuous measurement can be disabled by sending the
RS-232C or GP-IB. When continuous measurement is disabled, trigger acceptance is controlled only by the
host (computer or PLC).
See: Refer to "Trigger" (p. 152) and "9.8 Data exporting methods" (p. 163) for trigger commands.
:INITIATE:CONTINUOUS OFF command via
32
3.5 Selecting the Measurement Range
3.5Selecting the Measurement Range
The measurement range can be set as follows.
When the threshold values of the comparator are set with the panel keys, the measurement range is
selected automatically according to the settings (reference values or upper/lower thresholds, refer
to the following table).
When the comparator settings are made by remote control commands, the measurement range is
not affected.
Changing the Range
If the resistance value of the measuring object is very small compared to the measurement range, the measurement error increases.
Ranges 1000 Ω range or higher (LOW POWER: OFF) cannot be used for inductor measurements.
Auto-Range (when making comparator settings)
LOW POWER: OFF (p. 28)
VOLTAGE LIMIT: OFF (p. 64)
Reference value (REF%)
and
upper limit (ABS) ranges
0.00 m
100.1 mΩ to 1000.9 mΩ1000 mΩ100.1 mΩ to 1000.9 mΩ 1000 mΩ0.0 mΩ to 1000.9 mΩ1000 mΩ
1.001 MΩ to 3.009 MΩ3 MΩ0.501 MΩ to 1.509 MΩ3 MΩ−−
3.010 MΩ to 10.009 MΩ10 MΩ1.510 MΩ to 5.009 MΩ10 MΩ−−
10.01 MΩ to 30.09 MΩ30 MΩ5.01 MΩ to 15.09 MΩ30 MΩ−−
30.10 MΩ to 120.00 MΩ100 MΩ15.10 MΩ to 120.00 MΩ100 MΩ−−
Ω to 100.09 mΩ100 mΩ0.00 mΩ to 100.09 mΩ100 mΩ−−
1.001 Ω to 3.009 Ω3 Ω1.001 Ω to 3.009 Ω3 Ω1.001 Ω to 3.009 Ω3 Ω
3.010 Ω to 10.009 Ω10 Ω3.010 Ω to 10.009 Ω10 Ω3.010 Ω to 10.009 Ω10 Ω
10.01 Ω to 100.09 Ω100 Ω10.01 Ω to 100.09 Ω100 Ω10.01 Ω to 100.09 Ω100 Ω
100 .1 Ω to 300.9 Ω300 Ω100.1 Ω to 300.9 Ω300 Ω100.1 Ω to 300.9 Ω300 Ω
301.0 Ω to 1000.9 Ω1000 Ω301.0 Ω to 1000.9 Ω1000 Ω301.0 Ω to 1200.0 Ω1 000 Ω
1.001 kΩ to 10.009 kΩ10 kΩ1.001 kΩ to 5.009 kΩ10 kΩ−−
10 .01 kΩ to 30.09 kΩ30 kΩ5.01 kΩ to 15.09 kΩ30 kΩ−−
30.10 kΩ to 100.09 kΩ100 kΩ15.10 kΩ to 50.09 kΩ100 kΩ−−
100.1 kΩ to 300.9 kΩ300 kΩ50.1 kΩ to 150.9 kΩ300 kΩ−−
301.0 kΩ to 1000.9 kΩ1000 kΩ151.0 kΩ to 500.9 kΩ1000 kΩ−−
Selected
range
LOW POWER: OFF (p. 28)
VOLTAGE LIMIT: ON (p. 64)
Reference value (REF%)
and
upper limit (ABS) ranges
Selected
range
LOW POWER: ON (p. 28)
Reference value (REF%)
and
upper limit (ABS) ranges
Selected
range
Manual Range Selection
The Basic Settings screen appears.
1
Selection
2
Press the up/down
cursor keys to change
the setting.
Increments
the range.
Decrements
the range.
Open the Basic Settings screen.
1
Select the range.
2
Selectable ranges depend on the low-power resistance setting (p. 28).
• When low-power resistance measurement is set to OFF: 100 mΩ, 1000 mΩ, 3 Ω, 10 Ω, 100 Ω, 300 Ω, 1000
• When low-power resistance measurement is set to ON: 1000 mΩ, 3 Ω, 10 Ω, 100 Ω, 300 Ω, 1000 Ω
Return to the Measurement screen.
3
33
3.5 Selecting the Measurement Range
3
34
The Basic Settings screen appears.
1
Selection
2
Internal trigger
Press the up/down
cursor keys to change
the setting.
1
Selection
2
Disables zero adjustment (cancel).
Executes zero adjustment.
3.6 Zero Adjustment
3.6Zero Adjustment
When four-terminal measurement (Kelvin connection) is not practical such as when measuring very
small samples, the additional inherent resistance of the two-terminal wiring should be canceled out.
The zero-adjustment function can cancel out up to 10
Before Zero Adjustment
• The guaranteed accuracy of the instrument applies to four-terminal connections without zero adjustment.When using
four-terminal connections, do not execute zero adjustment.
Executing zero adjustment with incorrect wiring may amplify measurement error. However, zero adjustment
may be needed even with four-terminal measurements if they are affected by a large offset voltage, such as
due to thermal emf (LOW POWER: OFF, in 100 Ω to 100 MΩ range).
• Execute zero adjustment when the ambient temperature has changed, or when a probe is replaced.
Execute zero adjustment after the warm-up period following power on.
Ω additional resistance.
1
2
3
4
5
Open the Basic Settings screen.
Select the internal [INT] trigger mode.
Short the probes together.
Confirm that the measurement value does not exceed 10 Ω.
If no measurement value is displayed, increment the measurement range (p. 32).
Select whether to enable or disable zero adjustment.
6
After confirming that the measurement value does not exceed 10 Ω, execute zero adjustment.
Return to the Measurement screen.
When Zero Adjustment Fails
If zero adjustment fails, the following error message appears.
35
3.6 Zero Adjustment
Before attempting zero adjustment again, confirm the following:
• When measuring with the 10 Ω range, confirm that the displayed value does not exceed 10
Ω.
• Confirm that the probe connections are correct.
3
36
Measurement value>upper limit
Upper limit≥Measurement value≥
Lower limit
Measurement value<lower limit
Select the REF% (relative values) judgment
Decide whether a
measurement value is within
specified tolerance limits
relative to a specified reference
value (p. 37)
Select the ABS (absolute values) judgment
Decide whether a measurement
value is between specified
upper and lower threshold
Comparator results can be output to an external device (via
EXT. I/O connector) when the comparator reference/tolerance or upper/lower threshold values have been set.
See: "Chapter 8 External Control" (p. 95)
Comparator results are also indicated by the COMP Hi/IN/Lo
panel lamps, and by audible beeper (disabled by default).
See: "Setting the Comparator Judgment Beeper" (p. 70)
The comparator judment mode can be set as one of the following:
Before Using the Comparator Function
• When the measurement value is out of the selected measurement range, comparator judgment indicators
appear as follows. In the event of a measurement fault, no judgment is made.
See:"3.8 Confirming Faulty Measurements" (p. 40)
Out-of-Range Display
• If power is turned off during comparator setting, changes to settings are lost as they revert to their previous
values. To accept the settings, press the ENTER key.
• When setting comparator criteria, the appropriate range is selected automatically.
Refer to "Auto-Range (when making comparator settings)" (p. 32) for range settings.
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled).
Press the units key to accept the setting and move the cursor to the upper threshold.
Set the positive tolerance.
3
Press the % key to accept the setting and move the cursor to the negative tolerance value.
The negative tolerance is initially set to the same absolute value as the positive tolerance (change as needed).
Set the negative tolerance in the same way (as needed)
4
Accept the settings and return to the Measurement screen.
5
39
Upper threshold
Lower threshold
1_15_150_150.0 mΩ
(Example: 150
Selection
To Reset Numerical Values
Deletes entered digits.
This key is enabled only when entering
numerical values.
To change the value after selecting the units, use
the cursor keys to select the item to
change, then enter the new value with the numeric keys.
1
2
1
2
Selection
(Example: 50
5_50_50.0 mΩ
• Internal calculations are performed on floating-point values, and judgments round up any fraction of the
least-significant digit.
• Displayed values of the reference and tolerances are rounded according to the selected range. Internal
calculations use data not rounded off, so judgments are based on the entered (setting) values.
• An error message appears if you press ENTER with the positive tolerance < the negative tolerance.
See:"11.3 Error Displays and Solutions" (p. 197) (ERR:001)
To abort the setting process, press . Settings are abandoned and the display returns to the
previous screen.
Open the absolute value threshold setting screen.
1
Set the positive tolerance.
2
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled). Error message is not displayed.
3
Press the units key to accept the setting and move the cursor to the lower threshold value.
Set the negative tolerance in the same way.
3
Press the units key to accept the setting and move the cursor to the upper threshold.
Accept the settings and return to the Measurement screen.
4
40
Appears when the measurement value is outside of the measurement or display range.
Check for a broken sample component.
The comparator result is Hi when +OvrRng is displayed, and Lo when -OvrRng is displayed.
No external measurement fault signal (ERR
) is output.
Measurement out of range
The resistance between the H
POT
and H
CUR
probe contacts, and between the L
POT
and L
CUR
probe contacts, are measured and compared with specified contact fault values.
An error message appears when the measurement value reaches or exceeds the specified contact
fault values.
If this error persists, probe wear or cable failure may be the cause.
If the error is not cleared by shorting the tips of a known-good measurement probe, the instrument
requires repair.
Contact Check Fault
This method monitors the stability of the voltage between H
POT
and L
POT
probe contacts.
An error message appears when voltage instability is detected due to chattering of the probe contacts.
If this error is displayed frequently, the probes may be degraded due to wear.
C.E. Volt may also be displayed when external noise is strong.
Voltage Monitor fault
See: "Out-of-Range Detection Function" (p. 41)
See: "4.5 Checking for Poor or Improper Contact
(Contact Check Function)" (p. 50)
See: "4.7 Detecting Measurement Voltage Faults
(Voltage Level Monitor Function)" (p. 55)
This display appears after changing measurement settings and before the next measurement is
performed.
- - - - - - -
Display Examples: Display Measurement State and Appearance with Open-Circuit Probe
Display Measurement State
Current Monitor Results
Normal (PASS)Fault (FAIL)
Contact Check Results
Voltage Level Monitor
Results
Normal
(PASS)
Display: Measurement Value
COMP indicator: According to the
measurement value
Display: +OvrRng/ -OvrRng
COMP indicator: Hi/ Lo
(when connection to the measuring
object is broken)
Fault
(FAIL)
Display: C.E. Hi/ C.E. Lo/ C.E. Volt
COMP indicator: No judgment
EXT.I/O: ERR signal output
Display: C.E. Hi/ C.E. Lo/ C.E. Volt
COMP indicator: No judgment
EXT.I/O: ERR signal output
+OvrRng
-OvrRng
C.E. Hi
C.E. Lo
C.E. Volt
Display
Display
Display
This method monitors the regulated measurement current for normal flow through the measuring object.
An error is detected mainly when a measurement fault occurs due to an open measuring object or poor contact between the H
CUR
and L
CUR
probes.
The error display depends on the contact check and voltage level monitor states (refer to the table below).
Current Monitor Fault
See: "Current Monitor Function" (p. 41)
The measurement fault display differs according to detection order and settings.
3.8 Confirming Faulty Measurements
3.8Confirming Faulty Measurements
When a measurement is not performed correctly, a measurement fault indicator appears and a
measurement fault signal is output at the ERR
detection).
The instrument detects measurement faults by the following four methods.
pin of the EXT. I/O connector (except for out-of-range
Measurement Fault Detection Order
NoNoNoNoNo
No
No
Measurement Fault DetectionDisplayEXT.I/O
41
3.8 Confirming Faulty Measurements
Probe short-circuit
Hi Wiring Contact Error
Lo Wiring Contact Error
Voltage Level Monitor Error
Constant-Current Error
Below Lower Limit
Above Upper Limit
No Measurement Data
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
ERR: 021
C.E. Hi
C.E. Lo
C.E. Volt
+OvrRng
-OvrRng
+OvrRng
- - - - - - - -
PRB_SHORT output,
ERR output
ERR output, CE_HI
output
ERR output, CE_LO
output
ERR output
HI output
LO output
HI output
Measurement fault detection
proceeds in the order shown
at the left, ending with display
of the first detected error.
Corresponding measurement fault signals are also
output at the EXT. I/O connector.
3
Out-of-Range Detection Function
Examples of Out-of-Range Faults
Out-of-Range DetectionMeasurement Example
The measurement value is outside
of the measurement range.
The relative tolerance (%) display
of the measurement value exceeds
the display range (999.999%).
The zero-adjusted value is outside
of the display range.
While measuring, input voltage exceed the A/D converter input range.
10 kΩ range is used to measure 13 kΩ.
Measuring 500 Ω (+2400%) with a reference value of 20 Ω
1 In the 1 Ω range with 0.5 Ω zero-adjustment in effect, measuring 0.1 Ω provides a zero-adjusted value of -0.4 Ω, which is outside of the display range.
Measuring a large resistance value in an electrically noisy environment
Current Monitor Function
The instrument supplies constant measurement current through the measuring object via the H
probes. A current monitor fault occurs if constant current cannot be attained. If the contact check and voltage
level monitor results are normal, the out-of-range and comparator result displays indicate "Hi".
Example of Current Monitor Fault
• Broken DUT (open work)
•H
•H
CUR
CUR
or L
or L
probe contact fault
CUR
cable break
CUR
CUR
and L
CUR
42
3.8 Confirming Faulty Measurements
4.1 Making Range-Specific Measurement Settings
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Selects the range to use.
Customizing Measurement
SettingsChapter 4
(set as needed)
Change measurement settings as appropriate for your application.
Refer to "Detailed Settings Screen" (p. 19) for the available settings.
43
4.1Making Range-Specific Measurement
Settings
The setting affects only the selected range (except for the DELAY1 setting).
Open the Basic Settings screen.
1
Open the Measurement Settings Screen.
2
Select the range to change the setting.
3
4
Set the items as needed.
4
44
The DELAY1 setting is common to all ranges.
The default setting is 0 ms (corresponding to trigger signal input
at the same time as probe contacts become stable).
Setting DELAY1 affects measurements in all ranges.
Set DELAY1
Adjust this setting to allow for
probe contact mechanical
stabilization.
Set DELAY2 to the time needed for stabilization after measurement current is applied, such as may be required for inductive
components.
The setting affects only the selected range.
The default setting is 0 ms (corresponding to resistance measurement of non-inductive components).
Set DELAY2
Adjust this setting to allow for
stabilization of the measurement
sample.
DELAY1 and DELAY2 Timing Chart
* Internal delay is provided to suit purely resistive (non-reactive) measuring objects,
and is different for each measurement range.
Acquisition
Stable Contact
Internal
Probe Contact
Condition
Start
TRIG
Measurement
Current
Acquisition of
Measurement
value
End of
Measurement
EOM
4.2 Setting Pre-Measurement Delay
4.2Setting Pre-Measurement Delay
This setting specifies the delay between trigger signal input and the start of measurement.
Adjust this setting to delay measurement until the measurement value has time to stabilize, so that
even if the sample is connected after triggering, measurement starts only after the specified delay.
The delay can be set by two methods, as follows.
45
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Setting range: 0.0 ms (default) to 100.0 ms
[DELAY1] is common to all ranges, while
[DELAY2] can be set for each range independently (p. 43).
Numerci keypad
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4.2 Setting Pre-Measurement Delay
Estimated Delay time
Set the delay so that inductance does not affect measurements.
To fine tune the delay, begin with a longer delay than necessary, then gradually shorten it while watching the
measurement value.
Open the Basic Settings screen.
1
Open the Measurement Settings Screen.
2
4
Set [DELAY1] or [DELAY2].
3
Return to the Measurement screen.
4
46
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function(default)
(go to step 6)
Enables the function
When measuring the resistance of current sensing resistors, there may be a discrepancy between
the resistance value at the time of mounting on the board being used and the resistance value
obtained when measuring the component alone (due to the effects of the probing location and other
factors). The scaling function corrects the measured resistance value obtained from the component
alone to yield the resistance value during actual use. Scaling is performed by mean of the following
equations:
R
= A × R + B
S
R :Measured value before compensation
R
Resistance value after compensation
S:
A :Compensation coefficient setting range:
B :Offset resistance setting range
: ±0.0000 mΩ to ±99.9999 MΩ (default: 0.0000 mΩ)
0.50000 to 2.00000 (default:: 1.00000)
• Scaling calculation will be performed on a measured value after zero-adjustment. Consequently, measure
value may not be equal to zero after zero-adjustment.
• Changing the comparator settings automatically sets the scaling function to OFF.
1
2
3
Open the Basic Settings screen.
Open the Measurement Settings Screen.
Enable or disable the scaling function.
Set the compensation coefficient.
Selection
1
2
Setting range: 0.50000 to 2.00000 (default: 1.00000)
Numeric keys
3
Selection
1
2
Setting range: ±0.0000 mΩ to ±99.9999 MΩ (default: +0.0000 mΩ)
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
(The settings for the current
measurement range are displayed.)
1
Selection
2
4.4 Setting the Measurement Integration Time Option
4.4Setting the Measurement Integration Time
Option
The integration time can be optionally set for each range by selecting FAST, MED, or SLOW measurement speed.
Integration time can be set in ms or PLC
*
units.
* PLC = Power Line Cycle, where one PLC is the time equivalent to one period of the power line waveform.
At 50 Hz, one PLC = 1/50
PLC setting units are useful where measurements may be affected by power line noise (high- or low-resistance measurements)
Default setting
LOW POWER
Range
100 mΩ
1000 mΩ
3 Ω
10 Ω
100 Ω
300 Ω
1000 Ω
10 kΩ 0.3 ms 3.0 ms 1 PLC OFF−−−−
30 kΩ 0.3 ms 3.0 ms 1 PLC OFF
100 kΩ 0.5 ms 3.0 ms 1 PLC OFF
300 kΩ 0.5 ms 3.0 ms 1 PLC OFF
1000 kΩ 1.5 ms 5.0 ms 1 PLC OFF----
3 MΩ 1.5 ms 5.0 ms 1 PLC OFF---10 MΩ 2.5 ms 1 PLC1 PLC OFF---30 MΩ 2.5 ms 1 PLC1 PLC OFF----
100 MΩ 1 PLC2 PLC4 PLC OFF----
*1. Low-power resistance measurement (p. 28)
*2. Two measurements are made within the above integration time.
Integration time
FASTMEDSLOWFASTMEDSLOW
0.5 ms 5.0 ms 1 PLC
0.3 ms 2.5 ms 1 PLC
0.3 ms 2.5 ms 1 PLC
0.3 ms 2.5 ms 1 PLC
0.3 ms 3.0 ms 1 PLCOFF 0.3 ms 2.5 ms 1 PLC
0.3 ms 3.0 ms 1 PLCOFF 0.3 ms 2.5 ms 1 PLC
0.3 ms 3.0 ms 1 PLCOFF 0.3 ms 2.5 ms 1 PLC
th
of a second, and at 60 Hz, one PLC = 1/60th of a second.
*1
: OFFLOW POWER*1: ON
OVC
ON
ON
ON
ON
*2
*2
*2
*2
Integration time
−−−−
0.5 ms 5.0 ms 1 PLC
0.5 ms 5.0 ms 1 PLC
0.5 ms 5.0 ms 1 PLC
−−−−
−−−−
−−−−
OVC
ON
ON
ON
ON
ON
ON
*2
*2
*2
*2
*2
*2
1
2
Open the Basic Settings screen.
Open the Measurement Settings Screen.
Select the integration setting units.
Selection
1
Sets in units of time.
Sets in units of power line cycles.
2
The setting is specific to the selected range (p. 43).
Selection
1
Setting range:
• When setting in ms units:
0.1 ms to 100.0 ms
• When setting power-line-cycle units:
1 to 6 PLC (60 Hz), 1 to 5 PLC (50 Hz)
2
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
3
Select the integration time.
4
49
4.4 Setting the Measurement Integration Time Option
4
Return to the Measurement screen.
5
• The instruments accuracy specifications are applicable only with the default integration times.
Investigate your measurement requirements carefully before changing the integration time.
• When the effects of power line noise can be ignored, the integration time can be set longer than the default to reduce
scattering of measurement values. On the other hand, if the integration time is too short, scattering increases.
For high-or low-resistance and low-power resistance measurements that are easily affected by power line noise, we
suggest setting according to the power line period (PLC units).
50
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function (go to step 5).
Enables the function (default).
The setting is specific to the selected range
(p. 43)
A contact fault occurs when a measured
value exceeds the threshold setting.
4.5 Checking for Poor or Improper Contact (Contact Check Function)
4.5Checking for Poor or Improper Contact
(Contact Check Function)
This function detects poor contact between the probes and measuring object, and broken measurement probes.
The instrument continually monitors the resistance between the H
L
and L
CUR
When the resistance is outside of the specified value, a contact check fault occurs and the C.E. Hi
or C.E. Lo error message appears. No comparator judgment is applied to the measurement value.
When these error messages appear, check the probe contacts, and check for broken measurement
probes.
(If the error is not cleared by shorting the tips of a known-good measurement probe, the instrument
requires repair.)
probes from the start of integration (including response time) and while measuring.
POT
CUR
and H
probes and the
POT
• During low-resistance measurement, poor contact of the H
range measurement.
• When contact checking is disabled, measurement values may be displayed even when a probe is not contacting the measuring object.
1
2
3
Open the Basic Settings screen.
Open the Measurement Settings Screen.
Enable or disable the function.
CUR
or L
probe may be detected as an out-of-
CUR
4
Select the contact check fault threshold resistance.
4.5 Checking for Poor or Improper Contact (Contact Check Function)
Probe contacts can be improved by applying current from the POT to the CUR probes before measuring.
The Contact Improver function applies voltage to the sample. Be careful when measuring
samples with characteristics that may be affected.
The current used for the Contact Improver functions can be selected as follows.
17 mA, 25 mA, 35 mA (default), 50 mA
Higher current provides more effective contact improvement, but at the cost of faster probe deterioration.
Contact Improver current can be set to be disabled (OFF), enabled (ON), or PULSE.
The PULSE setting applies the contact improvement current for approx. 100 µs to 300 µs immediately before
measurement. The PULSE setting is useful to decrease Joule heating of the DUT based on contact improvement current if the measuring object is susceptible to change in characteristics.
100 mΩ range to 100 kΩ range300 kΩ range to 100 MΩ range
DUT current
DUT voltage20 V max.15 V max.
*
2 mA max.60 mA max.
*. It takes several microseconds for the DUT current to reach the steady-state value. Until the steady-state
value is reached, a transient current that is approximately equal to the contact improvement current setting
(default setting: 35 mA) will flow.
SettingTiming chart (Contact Improver Function)
OFF
ON
*Internal delay is different for each range.
SettingTiming chart (Contact Improver Function)
Measurement
Start
TRIG
Contact
Improver
current
Probe
Contact
Condition
Measuring
Internal
delay *
Stable Contact
Approx.
100
μs
Measurement
Start
TRIG
Contact
Improver
current
Probe
Contact
Condition
Measuring
Internal
delay *
Stable Contact
Approx.
300
μs
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
PULSE
When both applied voltage
limiter and low-power resistance measurement function are disabled.
PULSE
When the applied voltage
limit function or low-power
resistance measurement
function is enabled.
For ranges between 300 kΩ and 100 MΩ, the [PULSE] setting is enabled by default.
Before measuring in the ranges from 300 kΩ to 100 MΩ with the Contact Improver function set to [ON], verify
that measurements are not biased.
Open the Basic Settings screen.
1
Open the Measurement Settings Screen.
2
54
Selection
2
Disables probe contact improvement
(go to step 4).
Enables probe contact improvement.
Contacts improvement current is ap-
plied immediately before measuring only.
1
The setting is specific to the selected range (p. 43)
Set the Contact Improver current timing to disabled (OFF), enabled (ON), or PULSE.
(When selecting ON or PULSE)
Set the current limit value.
Return to the Measurement screen.
55
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function (default)
(go to step 5).
Enables the function.
Disables the function of all ranges
(go to step 5).
The setting is specific to the selected range (p. 43)
Selection
1
2
LOOSE
*
NORMAL
*
SEVERE
*. Default setting: 10LOOSE is the default for the 0 MΩ
range, and NORMAL for ranges other than 100 MΩ.
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4.7 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)
4.7Detecting Measurement Voltage Faults
(Voltage Level Monitor Function)
When a measurement voltage fault occurs due to probe chattering, the C.E. Volt error message
appears on the measurement screen and an
appear when external noise is strong.
• Set the voltage level monitor to Loose, or OFF (disable).
Open the Basic Settings screen.
1
ERR signal is output. The C.E. Volt error may also
4
Open the Measurement Settings Screen.
2
Enable or disable the function.
3
Select the voltage level monitor threshold.
4
Return to the Measurement screen.
5
56
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Measurement current is applied while
awaiting trigger.
Measurement current is applied only
while measuring (default).
The setting is specific to the selected range (p. 43)
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4.8 Applying Current Only When Measuring (Current Mode Setting)
4.8Applying Current Only When Measuring
(Current Mode Setting)
When the Contact Improver function is set to Pulse or disabled (CONT IMP: PULSE or OFF) and
measurement current is set for pulse output, open-circuit voltage when not measuring does not
exceed 20 mV.
When the Contact Improver function is enabled (CONT IMP: PULSE or ON setting), the current mode setting
is ignored even if set to continuous (CURRENT MODE: CONT setting). (The Contact Improver function forces
pulse operation with measurement current applied only during measurement.)
1
2
3
Open the Basic Settings screen.
Open the Measurement Settings Screen.
Select whether to apply current when not measuring.
4
To apply measurement current continuously (CONT setting) even when waiting for a trigger, confirm that
the Contact Improver function is disabled (CONT IMP: OFF, (p. 52)).
Return to the Measurement screen.
4.9 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
DUT
Foreign Object
POT
CUR
DUT
electrode
Timing chart
(Probe Short-circuit Detection)
Probe short-circuit
PRB_SHORT
Comparator
Hi
,IN,Lo
End of
measurement
EOM
Measurement
Start
TRIG
Measurement
status
Transport
Measurement
stage
DUT1 contact
DUT2 contactTransport
Probe short-circuit
Transport
DUT1
measurement
DUT2
measurement
Short-circuit
detection
Detection timing setting
Available
Short-circuit
No judgment
4.9Test for Short-Circuited Probe
(Probe Short-Circuit Detection Function)
Four-terminal measurements are not possible when a conductive foreign object is present between the POT and CUR probe tips. To detect
short-circuited probes, this function measures the resistance between
the CUR and POT terminals after a specific time (initially 5 ms) following the end of measurement.
Probe short-circuit detection is disabled by default.
When a probe short-circuit is detected, an error message appears on
the measurement screen, and the
output.
(
ERR:021 Probe short error)
PRB_SHORT and ERR signals are
57
4
Short-circuit detection can also be controlled by asserting the active-low PRB_CHECK of the EXT. I/O signal.
Asserting the PRB_CHECK
end of measurement (p. 95).
About Probe Short-Circuit Detection
• If probes are connected to the measuring object during probe short-circuit detection, it is determined as
probe short circuit. Ensure that the probes have sufficient time to separate from the measuring object.
• Probe short-circuit detection occurs within about 1 ms.
• The threshold for probe short-circuit detection is fixed at 500 Ω, so if the resistance between CUR and POT
probes is larger, detection is not possible.
signal while measuring causes short-circuit detection to be performed after the
58
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Probe short-circuit detection is not used
(default) (go to step 5).
Probe short-circuit detection is used.
Selection
1
2
Setting range: 1 to 100 ms, 5 ms (default)
Short-circuit detection is delayed for the specified time following the end of measurement.
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4.9 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
• Even while the probe short-circuit detection function is set to be disabled, short-circuit detection is performed when the
EXT. I/O PRB_CHECK
• When the internal trigger [TRG: INT] source is selected, short-circuit detection is not performed after the end of mea-
surement. However, short-circuit detection can still be executed by asserting the PRB_CHECK signal is executed.
Enables/Disables Probe Short-Circuit Detection Function
signal is asserted.
1
2
3
4
Open the Basic Settings screen.
Open the System screen.
Enable or disable the function.
Set the probe short-circuit detection timing.
5
Return to the Measurement screen.
59
Automatic
Comparison
1st Stage
2nd Stage
Transport
Direction
ABCDE
The SET MONITOR connectors are identical
to RS-232C connectors.
Be careful to avoid connecting the wrong
connectors.
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
The setting is not monitored (default) (go
to step 7).
The setting is monitored.
4.10 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)
4.10 Comparing the Measurement Settings of
Two Instruments (Settings Monitor Function)
This function automatically compares the settings of two
instruments to determine whether they are the same.
Only comparator settings and measurement speed are
compared.
When the settings differ, an alarm notification is displayed
and subsequent.
TRIG signal input is prevented from starting measure-
ment.
When the settings of two instruments match, TRIG input is
accepted and measurement starts.
However, if the range defined by the upper and lower thresholds of the second stage is broader than that of the first stage,
measurement still starts despite the different threshold settings.
4
Connect the SET MONITOR connectors of the two instruments using a Hioki 9637 RS-
1
232C cable.
Open the Basic Settings screen.
2
Open the System screen.
3
Enable or disable the function. Enable the function on both instruments.
4
60
Selection
1
2
Selects this instrument as the 1st stage.
Selects this instrument as the 2nd stage.
3
Enter the difference in tolerance (%) to be allowed at the 2nd stage from the tolerance range set for the 1st stage.
Setting range: 0.000 to 9.999%
4
Example:
If the 1st stage is set to measure 12 Ω
±0.800%, and the 2nd is to measure 12 Ω
±1.000%, the tolerance will be 0.300%
Selection
1
2
Selects this instrument as the 1st stage.
Selects this instrument as the 2nd stage.
4.10 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)
5
6
Select the instrument to serve as the 1st stage, and set its tolerance range.
Set the instrument to serve as the 2nd stage.
7
Return to the Measurement screen.
Tolerance Range Setting Conditions
61
1st Stage
TRIG Input Accepted
TRIG Input Inhibited
TRIG
Input Accepted
When changing the reference value
Change the reference value to match
the 1st stage
The error message appears when
the settings do not match.
2nd Stage
1st Stage
2nd Stage
1st Stage
2nd Stage
SET MONITOR: ON 1st 0.300% (on the SET MONITOR: ON 2nd (on the System screen)
When an error is displayed
ERR:003
Setting monitor error. (COMP)Comparator settings do not match. Please check.
ERR:004
Setting monitor error. (SPEED)Measurement speed settings do not match. Please check.
4.10 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)
Practical Example
4
62
Retry
Chatter
Probe
Contact
Condition
Start
TRIG
Contact
Improver
Measurement
Current
Contact
Check
End of Measurement
EOM
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Retries disabled (go to step 5).
Retries enabled (default).
4.11 Retrying Measurement After a Fault (Retry Function)
4.11 Retrying Measurement After a Fault (Retry
Function)
The Retry function enables measurement to be
automatically retried when a measurement fault
occurs due to probe chatter.
During Retry, all measurement operations including Contact Improvement and DELAY2 (but excluding DELAY1)
are restarted.
If a measurement fault persists after the specified continuous retry interval (e.g., if the measured object is not
connected), retrying is aborted and the EOM
output.
When Retry is enabled, the maximum time to end-ofmeasurement occurs when recovering from a measurement fault immediately before the retry interval expires,
which approaches the sum of the retry interval setting
plus normal measurement time.
Decreasing test throughput may indicate probe maintenance is required.
signal is
1
2
3
Open the Basic Settings screen.
Open the System screen.
Enable or disable the function.
Set a continuous retry interval.
Selection
1
2
Setting range: 1 to 50 ms (default: 2 ms)
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4
Return to the Measurement screen.
5
63
4.11 Retrying Measurement After a Fault (Retry Function)
4
64
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
1
2
Applied voltage is not limited(default).
Applied voltage is limited.
Selection
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
4.12 Limiting Measurement Voltage (Applied Voltage Limiter Function)
4.12 Limiting Measurement Voltage (Applied
Voltage Limiter Function)
Applied voltage used for measurements can be limited to 5 V or lower.
When [ON] is selected, an adequate measurement range in which the "measured current
ence value or
selected.
upper comparator threshold" (voltage) does not exceed 5 V will be automatically
× refer-
1
2
3
4
Open the Basic Settings screen.
Open the System screen.
Enable or disable the function.
Return to the Measurement screen.
The instrument will display OvrRng (over range) when measuring a sample with applied voltage of 5 V or greater.
To maintain measurement precision, the instrument self-calibrates every ten minutes to compensate for internal circuit offset voltage and gain drift. This function cannot be disabled.
During self-calibration, the subsequent measurement is delayed for about 6 PLC + 10 ms (PLC =
Power Line Cycles) for internal circuit compensation.
Self-Calibration Timing
Supplied power 50 Hz: 130 ms, 60 Hz: 110 ms
• When the timing of self-calibration overlaps with a measurement, self-calibration is postponed until the end
of measurement.
• When a trigger signal (measurement start signal) is applied during self-calibration, the start of the triggered
measurement is postponed until self-calibration is finished.
• Self-calibration executes automatically after changing comparator or measurement speed settings.
• During self-calibration, measurement current and the Contact Improver current are inhibited.
4
4.14 Compensating for Thermal EMF Offset
(Offset Voltage Compensation - OVC)
This function automatically compensates for offset voltage resulting from thermal emf or internal
instrument bias. (OVC: Offset Voltage Compensation)
See: "Appendix 2 Effect of Thermal emf" (p. A2)
The following value is known to be a true resistance value from RP (>0), the value measured with
current flowing in the positive direction, and R
negative direction.
Offset voltage compensation is automatically enabled in the following conditions, and cannot be modified or
disabled:
• When a range from 100 mΩ to 10 Ω is selected.
• Low Power Resistance Measurement (LOW POWER ON)
When the test object is inductive, some delay [DELAY2] is required (p. 44) to allow adequate current flow
before starting measurement.
(<0), the value measured with current flowing in the
N
66
4.14 Compensating for Thermal EMF Offset (Offset Voltage Compensation - OVC)
5.1 Disabling and Enabling Key Operations
Key operations other than comparator settings (REF%, ABS, units and numeric
keys) and F1 [UNLOCK] keys are disabled.
To disable key operations: [M.LOCK] is displayed when returning to the [MENU]
measurement screen.
Disabling All Except Comparator Settings
Only comparator
settings are enabled.
All key operations except F1 [UNLOCK] are disabled.
To disable key operations: [F.LOCK] is displayed when returning to the [FULL]
measurement screen.
Disabling All Key Operations Including Comparator
Key operations to change
settings are disabled.
(although key-lock can
be canceled)
Asserting (Low) the EXT. I/O KEY_LOCK signal disables all panel keys, including
F1 [UNLOCK] and F1 [LOCAL] (disables remote control) (p. 95).
To disable the key-lock function and re-enable the keys, de-assert (High) the
KEY_LOCK
signal.
Disabling All Panel Keys
All key operations
are disabled.
The Basic Settings screen appears.
1
Selection
2
Key operations enabled (default).
Disables all except key-lock cancel.
Disables all except key-lock cancel and
comparator setting change.
[UNLOCK] is displayed only
when key-lock is enabled by
front panel key operations.
System SettingsChapter 5
5.1Disabling and Enabling Key Operations
Disabling Key Operations (Key-Lock Function)
Activate the key-lock function to disable the instrument's front panel key operations.
Three key-lock levels are available to suit specific purposes.
67
5
Open the Basic Settings screen.
1
Enable or disable key operations.
2
Return to the Measurement screen.
3
68
5.1 Disabling and Enabling Key Operations
Re-Enabling Key Operations (Key-Lock Cancel)
Key-lock can be canceled only when [UNLOCK] is displayed.
Press and hold F1 [UNLOCK] for one second.
If key operations are disabled by the
keys.
KEY_LOCK signal, de-assert (High) the signal to unlock the
5.2 Setting the Comparator Judgment and Key Beepers
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Disables the beeper.
Enables the beeper (default).
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
5.2Setting the Comparator Judgment and Key
Beepers
Enabling or Disabling the Key Beeper
The key beeper sound can be enabled and disabled.
The key beeper is enabled (ON) by default.
Open the Basic Settings screen.
1
69
5
Open the System screen.
2
Select whether to enable or disable the key beeper.
3
Return to the Measurement screen.
4
70
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Disables the beeper (default)
(go to step 5).
Enables the beeper.
Selection
1
2
IN (beep when within range)
HI/LO (beep when out of range)
LOW (beep when below lower threshold)
HIGH (beep when above upper threshold)
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
5.2 Setting the Comparator Judgment and Key Beepers
Setting the Comparator Judgment Beeper
The comparator judgment beeper can be enabled and disabled.
The judgment beeper is disabled (OFF) by default.
1
2
3
Open the Basic Settings screen.
Open the System screen.
Select whether to enable or disable the judgment beeper.
4
5
Select the judgment beep conditions.
Return to the Measurement screen.
71
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Automatically detect local line frequency
(default).
When the line frequency is 50 Hz.
When the line frequency is 60 Hz.
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
5.3 Power Line Frequency Manual Setting
5.3Power Line Frequency Manual Setting
For proper electrical noise suppression, the instrument needs to be set to match the power line frequency.
With the default setting (AUTO), the instrument attempts to automatically detect the line frequency,
but manual setting is also available.
Unless the line frequency is set correctly, measurement values may be unstable.
An error message appears if line noise is high enough to prevent correct frequency detection
(ERR:041 (p. 197)). In that case, set the instrument's line frequency manually.
When the [AUTO] setting is selected, the line frequency is automatically set to 50 or 60 Hz when the instrument is turned on or reset.
However, automatic detection is not available when the line frequency changes after turning power on or
resetting.
If the actual line frequency deviates from 50 or 60 Hz, it is set to one of the closest frequencies.
5
Examples: If the actual line frequency is 50.8 Hz, select the 50 Hz setting.
If the actual line frequency is 59.3 Hz, select the 60 Hz setting.
Open the Basic Settings screen.
1
Open the System screen.
2
Select the line frequency to be used.
3
Return to the Measurement screen.
4
72
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Enter the last two digits of the year, and the month, day, hour, minutes and
seconds in that order.
(The cursor will move automatically.)
Clock settings cannot be canceled.
5.4 Setting the Clock
5.4Setting the Clock
To record and print the correct time when using statistical calculations (p. 84), the clock needs to be
set correctly.
The time of printing is also output when printing statistical calculation results.
1
2
3
Open the Basic Settings screen.
Open the System screen.
Set the date and time.
4
Return to the Measurement screen.
73
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
0~100%, 5% step
50% (default)
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
5.5 Adjusting Screen Contrast
5.5Adjusting Screen Contrast
The screen may become hard to see when ambient temperature changes. In this case, adjust the
contrast.
Open the Basic Settings screen.
1
Open the System screen.
2
5
Adjust the contrast.
3
Return to the Measurement screen.
4
74
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
0~100%, 5% step
80% (default)
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
5.6 Adjusting the Backlight
5.6Adjusting the Backlight
Adjust backlight brightness to suit ambient illumination.
• When external triggering [TRG: EXT] is selected, backlight brightness is automatically reduced after nonoperation for one minute.
• Be aware that the display may be hard to see when brightness is set too low (near 0%).
1
2
3
Open the Basic Settings screen.
Open the System screen.
Adjust the backlight.
4
Return to the Measurement screen.
75
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Returns all settings to their factory
defaults.
Cancels the operation.
Executes.
5.7 Initializing (Reset)
5.7Initializing (Reset)
The instrument can be reset by following three methods.
• System reset from the System screen: Returns all settings (except the clock) to factory defaults.
• Turn the instrument on while simultaneously holding the REF% and ABS keys: Returns all settings (except
the clock) to factory defaults.
• Reset by remote control command: returns all settings (except communication and clock settings) to their
factory defaults.
∗
RST command (non-backup) (p. 135)
:SYSTem:RESet command (p. 151)
This procedure describes system reset from the System screen.
Open the Basic Settings screen.
1
5
Open the System screen.
2
Select RESET.
3
Select whether to cancel or proceed to execute system reset.
4
The Measurement screen is displayed when system reset finishes.
0.1 ms to 100 ms
1PLC to 6PLC (60 Hz)
1PLC to 5PLC (50 Hz)
OFF/ ON
50 Ω/ 100 Ω/ 150 Ω/ 200 Ω/ 300 Ω/
400 Ω/ 500 Ω
OFF/ ON/ PULSE
17 mA/ 25 mA/ 35 mA/ 50 mA
OFF/ ON/ ALLOFF
LOOSE/ NORMAL/ SEVERE
EXT
100 MΩ
FAST
OFF
ON
OFF
0 ms
0 ms
OFF
A:1
B:0
Depends on measurement range
ON, 200 Ω
ON, 35 mA (ranges from
100 mΩ to 100 kΩ)
PULSE, 35 mA (ranges
from 300 kΩ to 100 MΩ)
ON, NORMAL (LOOSE
when 100 MΩ range)
PULSE
OFF
OFF
OFF
Trigger setting function
(p. 31)
Range switching function
(p. 32)
Measurement speed (p. 29)
Zero adjustment function (p. 34)
Comparator function (p. 36)
Key-Lock function (p. 67)
Probe delay setting (p. 44)
DUT response setting (p. 44)
Scaling function (p. 46)
Integration time (p. 48)
Contact check (p. 50)
Contact Improvement (p. 52)
Voltage Monitor Function (p. 55)
Current mode setting function (p. 56)
Auto-Memory function (p. 81)
Statistical calculation function (p. 84)
Data output function (p. 87)
77
5.7 Initializing (Reset)
DisplaySetting value
SYSTEM
SET MONITOR
PROBE CHECKOFF/ ON, 0 to 100 ms
RETRYOFF/ ON, 1 to 50 ms
VOLT LIMITOFF/ ON
TRIG EDGEOFF EDGE/ ON EDGE
EOM
INTERFACEGP-IB/ RS232C/ PRINT
PRINT MODE
LOW POWEROFF/ ON
JUDGE BEEP
KEY BEEPOFF/ ON
CLOCKClock setting (p. 72)
LINE FREQAUTO/ 50 Hz/ 60 Hz
CONTRAST0 ~100
BACK LIGHT0 ~100
RESET-
ADJUST-
OFF/ ON,
1st/ 2nd, 0.000% to 9.999%
PULSE/ HOLD
1 to 100 ms
NORMAL/ SAMPLE, 1 to 100,
ALL/ IN, 1/L/ 3/L
OFF/ ON
OFF/ IN/ HI/LO/ LOW/ HIGH
Default valueSettings
OFF, 1st, 0.000%
OFF, 5 ms
ON, 2 ms
OFF
ON EDGE
PULSE, 5 ms
RS232C, 9600bps
GP-IB, ADR01, LF
NORMAL, 100, ALL, 3/L
OFF
OFF, HI/LO
ON
AUTO
50
80
-
-
Settings Monitor function (p. 59)
Probe short-circuit detection function
(p. 57)
Retry function (p. 62)
Applied Voltage Limiter Function
(p. 64)
Start Logic Setting (p. 106)
End-of-measurement pulse width
(p. 105)
Interface setting (p. 113)
Set printing method (p. 91)
Low-Power Resistance Measurement (p. 28)
Comparator judgment beeper
(p. 70)
Key beeper (p. 69)
Line frequency (detection) setting
(p. 71)
Screen contrast adjustment
(p. 73)
Screen backlight adjustment
(p. 74)
Reset (p. 75)
Adjustment screen (p. A13)
5
78
5.7 Initializing (Reset)
79
Stores up to 30,000 measurement data using the EXT. I/O TRIG
signal or by pressing the F4 [MANU] key on the Basic Settings
screen.
Data Memory Function (p. 80)
Store measurement values
at specified timing.
This is convenient for batch exporting
data to a controller while switching
reels.
Measurement values are automatically stored as they become
stable.
When the specified number of data points (up to 99) is acquired,
the beeper sounds and auto-storing halts.
Auto-Memory Function (p. 81)
Store data after measurement
value has stabilized.
This is convenient for sample inspection after printing (vapor deposition)
resistors on a board.
Minimizes transfer time by eliminating the need for transmit requests from the remote controller.
(RS-232C interface only)
Data Output Function (p. 87)
Automatically output (export)
measurements at the end of
measurement.
Stored measurement data are lost when the instrument is turned off.
Therefore, be sure to print out or export important data to a PC.
Storing and
Exporting DataChapter 6
Measurement values can be stored or automatically exported, according to application. Stored data
can be output to a printer, RS-232C or GP-IB. Also, statistical calculations can be applied to internally stored data.
6
80
6.1 Storing Data at Specified Timing (Data Memory Function)
6.1Storing Data at Specified Timing
(Data Memory Function)
Measurement values are stored in the instrument's internal memory at the following timings (up to
30,000 points).
• All measurements performed by external (EXT) triggering
• When a trigger is applied during internally (INT) triggered measurement
The following three storage methods are available:
• Stores when an EXT. I/O TRIG
• Stores when a ∗
• Stores when the F4 [MANU] key on the [MENU] - [TRG] setting screen is pressed.
• This function can only be enabled by remote control. The data memory function should be enabled by
remote command beforehand. This setting is not available on the front panel.
• Stored memory data cannot be viewed on the instrument's screen. Use remote control commands to export
stored data.
TRG command is received (p. 138)
signal is received (p. 95)
Data Memory Function Operating Procedure
1
2
3
4
Moreover, stored data is automatically erased at the following timings:
• When the memory function setting (including auto-mem-
• When the range is changed (p. 32)
• When changing comparator settings (p. 36)
Enable data memory mode.
Send this remote command to enable the data memory function:
:MEMory:MODE MEMory (p. 156)
Store measurement values.
Execute external trigger measurement, or apply a trigger during internally triggered measurement.
Export the stored data.
Send this remote command to export the measurement values stored in the instrument:
:MEMory:DATA? (p. 157)
Clear measurement data from instrument memory.
Send this remote command to erase the data from instrument memory:
:MEMory:CLEar (p. 156)
• When printing the statistical calculations (p. 93)
ory) is changed (p. 156)
• When the measuring object is changed (p. 28)
• Upon system reset (p. 75)
81
Printing (p. 93)
Prepare the printer (p. 89).
Enable Auto-Memory and set
the number of values to store.
Setting judgment
criteria. (p. 36)
Measure
Beeper notifies when the specified number of values are stored.
The Basic Settings screen appears.
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
6.2Store as soon as Measurement is Stable
(Auto-Memory Function)
This function automatically stores the measurement value each time the probes contact the sample
with internal triggering.
When the specified number of values have been acquired, auto-storage operation stops.
Statistical calculations are applied to the stored data, with results output to the screen or printer
(RS-232C).
See: "6.3 Performing Statistical Calculations on Measurement Values" (p. 84)
"Chapter 7 Printing" (p. 89)
Data storage and printing can be automatically controlled by the Auto-Memory function.
The Auto-Memory function is disabled by default.
Enable the Auto-Memory function before setting the number of values to store.
Enabling the Auto-Memory function affects other functions as follows:
• Statistical calculation is forcibly turned on.
• The voltage level monitor function is forcibly turned off (the setting itself is not set to OFF, but the function is
disabled).
• The trigger source setting is forcibly turned to internal [TRG: INT].
When the trigger source is set to external [TRG: EXT], the Auto-Memory function is forcibly disabled.
Deleting Stored Data
Stored data is automatically erased at the following timings:
• When the memory function setting (including data-mem-
ory) is changed (p. 156)
• When the range is changed (p. 32)
• When changing comparator settings (p. 36)
• When the power is turned off
• When printing (p. 93)
• When the measuring object is changed (p. 28)
• Upon system reset (p. 75)
• Upon setting the auto-memory number of values to store
(p. 82)
Enabling the Auto-Memory Function
Open the Basic Settings screen.
1
6
82
The Data Settings screen appears.
[DATA SETTINGS]
1
Selection
2
Selection
1
2
Auto-Memory function is not used (default).
Auto-Memory function is used.
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
When the function is enabled
Displays the Auto-Memory setting
screen.
COUNT (IN judgment) (Hi/Lo judgment)
Setting range: 1 to 99
To Reset Numerical Values
Deletes entered digits.
This key is enabled only when entering
numerical values.
(Example: 20 values are set to be stored.)
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
2
3
4
Open the Data Memory Settings screen.
Enable or disable the function.
Return to the Measurement screen.
Setting the Number of Values to Store
1
2
3
Open the Auto-Memory Settings screen.
Enter the number of values to store.
To abort the setting process, press . Settings are abandoned and the display returns to the previous screen.
Accept the settings and return to the Measurement screen.
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
Acquiring Measurement Values Automatically
Momentarily disconnect (open-circuit) the probes.
1
Connect the probes to the measuring object.
2
When the measurement is stable, the value is automatically stored and the count is incremented.
When the count reaches the specified number of values, a long beep sounds, and subsequent measurements are not stored.
The (one) last acquired value can be deleted. (Undo function (p. 86)).
83
6
84
n
x
x
=
16−
−
=
n
LoHi
Cp
σ
1
6
2
−
−+−−
=
n
xLoHiLoHi
CpK
σ
6.3 Performing Statistical Calculations on Measurement Values
6.3Performing Statistical Calculations on
Measurement Values
Statistical calculations can be performed and results displayed for up to 30,000 measurement values.
Printing is also available (p. 93).
Calculation types: average, maximum and minimum values, population standard deviation, sample
standard deviation, and process compatibility indices
Maximum
Minimum
Mean
Population standard
deviation
Standard deviation
of sample
Process capability
index*
(variation)
Process capability
index*
(bias)
• When only one valid data sample exists, standard deviation of sample and process capability indices are
not displayed.
• When σ
• The upper limit of Cp and CpK is 99.99. If Cp and CpK>are 99.99, the value 99.99 is displayed.
• Negative values of CpK are considered as CpK = 0.
• If statistical calculation is turned off and then back on without first clearing the calculation results, calculation resumes from the point when it was turned off.
• Measurement speed is reduced when statistical calculation is enabled.
• When Auto-Memory is enabled (ON), statistical calculation is forcibly enabled (ON).
• When statistical calculation is disabled (OFF), Auto-Memory is forcibly disabled (OFF).
= 0, Cp and Cpk are 99.99.
n-1
X
= MAX (x1, ....., xn)
max
X
= MIN (x1, ....., xn)
min
In these formulas, n represents the number of valid
data samples.
Hi and Lo are the upper and lower thresholds of the
comparator.
* The process capability index represents the quality
achievement capability created by a process, which
is the "Quality variation and width bias possessed
by the process".
Generally, depending on the values of Cp and CpK,
process capability is evaluated as follows:
When statistical calculation is enabled, F3 [STAT]
appears on the Measurement screen.
Confirm calculation results (p. 86)
6.3 Performing Statistical Calculations on Measurement Values
Using Statistical Calculations
When statistical calculation is set to ON and an EXT. I/O trigger signal is applied, operation will be
performed as follows depending on the trigger source setting:
• With external [EXT] triggering: One measurement is performed and subject to statistical calculation.
• With internal [INT] triggering: The next measurement value after the trigger signal is subjected to statistical
calculation.
Operation is the same in the following cases:
(Key Operations)(Remote Control)
• When the F4[MANU] key on the [MENU] - [TRG] selection
screen is pressed.
• When the F2[PRINT] key on the Measurement screen is
pressed (with internal triggering and Auto-Memory disabled.
Appears only when the interface is set for the printer.).
• When acquiring measurement values with the Auto-Memory
function (p. 81).
• ∗When a
received.
• When an EXT. I/O PRINT
on the Measurement screen (with internal
triggering and Auto-Memory disabled).
TRG remote control command is
signal is applied
Open the Basic Settings screen.
1
Open the Data Memory Settings screen.
2
Enable or disable the function.
3
Return to the Measurement screen.
4
6
86
Calculation Results screen is displayed
(only if statistical calculation is
enabled).
Num
Total data count Va lNumber of valid measurement values (error-free data) (Valid)
Ave
Mean
Sn
Population standard deviation
Max
Maximum
Sn1
Standard deviation of sample
Min
Minimum
Cp
Process capability index (variation)
Cpk
Process capability index (bias)
Output to the printer.
"Example Printouts (PRINT MODE:
NORMAL)" (p. 94)
Statistical calculation results and stored data are erased when printing finishes.
Erases the last measurement and calculation result (executes only once).
Erases all measurement values and statisti-
cal calculation results.
After selecting, a confirmation screen appears.
6.3 Performing Statistical Calculations on Measurement Values
Confirming, Printing, and Erasing Statistical Calculation Results
Statistical calculation results are displayed on the screen.
Printing is available using commercially available printers having a serial interface. Statistical calculation results are automatically erased after printing.
Before printing, select the [PRINT] interface setting.
See: "7.2 Setting the Instrument" (p. 91)
The number of valid samples can be confirmed on the Statistical Calculation Results screen.
• When the number of valid samples (Val) is zero, calculation results are not displayed.
• When only one valid data sample exists, no standard deviation or process capability indices are displayed.
1
2
Display the Calculation Results screen.
To print
To print, select the printer as the interface setting on the System screen (p. 91).
To erase
87
The Basic Settings screen appears.
The Data Settings screen appears.
• Set the interface to [RS232C] beforehand. This function is not applicable to the GP-IB Interface.
See: "9.4 Configuring the Communications Protocol" (p. 113)
• When internal [TRG: INT] triggering is selected, data is exported only when a TRIG signal is applied.
• Executing a :READ? query command exports duplicate measurement values.
• Even when using other queries, ensure that the query response and auto-exporting of measured values do
not overlap.
• The data format for measurement values can be selected as ASCII (default) or BINARY. Transfer time is
minimized when BINARY is selected.
See: ":SYSTem:FORMat <ASCii/ BINary>" (p. 150)
Open the Basic Settings screen.
1
Open the Data Memory Settings screen.
2
Enable or disable the function.
3
6
Return to the Measurement screen.
4
88
6.4 Auto-Exporting Measurement Values (at End of Measurement) (Data Output Function)
7.1 Connecting the Printer
Printing (p. 93)
• Measurement values and
comparator judgments
• Statistical calculation
results
Connecting the printer
to the instrument
Setting the
Instrument (p. 91)
Make printer
settings
PrintingChapter 7
7.1Connecting the Printer
Before connecting the printer
89
Because electric shock and instrument damage hazards are present, always follow
the steps below when connecting the printer.
• Always turn off the instrument and the printer before connecting.
• A serious hazard can occur if a wire becomes dislocated and contacts another con-
ductor during operation. Make certain connections are secure.
• As much as possible, avoid printing in hot and humid environments. Otherwise, printer life
may be severely shortened.
• Use only compatible recording paper in the printer. Using non-specified paper may not only
result in faulty printing, but printing may become impossible.
• The recording paper may jam if it move at angle relative to the roller.
Compatible printer
The requirements for a printer to be connected to the instrument are as follows.
Confirm compatibility and make the appropriate settings on the printer before connecting it to
the instrument.
Normal printing method (default)
(go to step 8).
Sampling type
Selection
1
Selection
2
Setting range: 1 to 999 (default: 100)
Numeric keypad
3
1
2
Prints all independent of comparator judgment (default).
Prints only if the comparator judgment is
Selection
Open the Basic Settings screen.
1
Open the System screen.
2
91
7.2 Setting the Instrument
Select print as the interface type.
3
Select the printing method.
4
Select the number of samples.
5
7
Select the printing conditions.
6
92
1
2
Prints one data set per line (default).
Prints three data sets per line.
Selection
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previous screen.
The confirmation screen appears.
7.2 Setting the Instrument
7
8
Select the number of data sets per line.
Return to the Measurement screen.
93
(When statistical calculation is enabled)
7.3 Printing
7.3Printing
Before Printing
Verify that the instrument and printer settings (p. 91) are correct.
Printing Measurement Values and Comparator Judgments
Printing by key operation
Press the PRINT key to print the measurement value currently displayed on the Measurement
screen.
Printing by external control
Measurement values and comparator judgments can be printed when the PRINT signal (the instrument's EXT. I/O connector) is set to Low (ON) (shorted with ISO_GND
* ISO_GND is one of the pins of the instrument's EXT. I/O connector.
When statistical calculation is enabled [STATISTICS: ON] and triggering source is set to internal triggering
[TRG: INT], statistical calculations are performed and measurement values are printed. When external [TRG:
EXT] triggering is selected, only measurement values are printed. Use the TRIG
calculations with external triggering.
*
).
signal to perform statistical
Printing Statistical Calculation Results
Statistical calculation results can be printed when auto-memory or statistical calculation is enabled
[ON]. To print, select PRINT on the screen or set the PRINT
nector) to Low (ON) (shorted with ISO_GND
To enable auto-memory:
See: "6.2 Store as soon as Measurement is Stable (Auto-Memory Function)" (p. 81)
To enable statistical calculation:
See: "6.3 Performing Statistical Calculations on Measurement Values" (p. 84)
*
).
signal (the instrument's EXT. I/O con-
7
If no valid data exists, only the data set is printed. When only one valid data set exists, standard deviation of
sample and process capability index cannot be printed.
94
Resistance measurements
1 0.8725mOhm Lo
2 0.484mOhm Lo
3 10.99998 Ohm IN
4 -10.0026 Ohm Lo
27 9.9986 Ohm Hi
28 9.996 Ohm Hi
29 0.01003kOhm Hi
30 0.00012MOhm Hi