Hioki RM3542, RM3542-01 Operation Guide

RM3542, RM3542-01
RESISTANCE HiTESTER
Operation Guide
November 2009 Edition 1 RM3542A983-00 09-11H Printed in Japan
Thank you for purchasi ng the HIOKI RM3542/3 542 -01 Resistance Hi Tester. This guide describes bas ic scr een oper ation s and comparator setting procedures. Before using the instrument the first time, please read the operating manual.
ower On/Off
Unlit: power off (while power is not supplied)
Red light: power off (while power is supplied)
Green light: power on
omparator Results
COMP indicator LEDs)
i Measured Value is
above Upper limit
N Pass (meets criteria)
o Measured Value is
below Lower limit
Measurement Screen
Measurement screen ope rations are shown on the back of this sheet.
Decision criteria
Measurement cable jacks
Screen Organization
(Comparator function)
REF% Decide by reference value and tolerance (%) ABS Decide by specified threshold values
Numerical value entry and related settings
Select and change display screens and settings
Basic Settings Screen
TRG Change trigger source (INT/EXT) RANGE Change range (100mΩ to 100MΩ)
SPEED 0ADJ Turn zero-adjust on/off
COMP Turn comparator on/off LOCK Enable/disable key lock (OFF/FULL/MENU)
Change measurement speed (FAST/MED/ SLOW)
Detailed Settings
Screens
(MISC MEAS MEAS SETTINGS) Measurement Settings
DELAY1 Adjust the delay from probing to trigger input DELAY2 Adjust measurement object electrical response INT(FAST/ MED/ SLOW) Make fine adjustment to integration time CONT CHECK Contact check threshold settings CONT IMP Contact Improver function settings VOLT MONITOR Voltage level monitor function settings CURRENT MODE Current mode setting (CONT/ PULSE)
(MISC DATA DATA SETTINGS) Data Settings
AUTO MEMORY Turn Auto-Memory on/off STATISTICS Statistical calculation on/off DATA OUT Automatically output measured value
(MISC SYSTEM SYSTEM) System Settings
SET MONITOR Turn two-instrument setting comparison on/off PROBE CHECK Turn probe short-circuit detection on/off RETRY Retry function on/off TRIG EDGE (External I/O) Set trigger rising/falling edge EOM (External I/O) Set EOM (end-of-measurement) signal INTERFACE Communications interface settings LOW POWER Low-power resistance component measurement on/off JUDGE BEEP Comparator beeper settings KEY BEEP Key beeper on/off CLOCK(Y-M-D) Set internal clock LINE FREQ Set power source frequency CONTRAST Adjust screen contrast BACKLIGHT Adjust screen backlight RESET Initialize ADJUST Instrument Adjustment
1
Setting Comparator Decision Criteria for Measured Values
Decide According to
Reference Value and Tolerance
12
REF% Mode
Setting screen
Decide According to
Upper/Lower Thresholds
Select item
12
3
Change numerical value:
3
Change numerical value:
Change sign:
Set reference value
(example 10.5Ω)
Set upper tolerance
(example +4.5%)
Set lower tolerance
(example -4.5%)
When setting the upper tolerance, the lower tolerance is automatically set to its negative equivalent. Change the setting as needed.
Set upper thresholds
(example 150mΩ)
Cancel:
4
Return to Measurement screen
Cancel:
4
ABS Mode
Setting screen
Measurement Screen Indicators
Zero-adjust (ON)
Appears only when OVC is enabled
Trigger Source
INT: Internal trigger EXT: External trigger
Measurement range
Tenkey Input Enabled
Key-Lock State
F.LOCK: Changes to settings are disabled M.LOCK: All except comparator settings are disabled
Remote control
(OVC: Offset Voltage Compensation)
Appears only when the Low-Power Resistance function is ena ble d
Measurement speed
Select item
Change Settings
Cancel key-lock (hold 1s)
Set lower thresholds
(example 50mΩ)
Return to Measurement screen
Measurement Fault (Error) Indicators
+OvrRng
-OvrRng
C.E. Hi
C.E. Lo
C.E. Volt
- - - - - - -
Out of Range (over- or under-range)
The measured value is above or below the measurement range, or open device.
Contact Check Fault
Indicates possible contact fau lt or broken cable.
Voltage Level Monitor Fault
Probe contact may be unstable , or electri ­cal noise ingress may be present.
No Measurement Data
No measurement value is present imme­diately after power on or after changing range. Errors with other causes can pre­vent display of measured values.
PRINT
STAT
NUMBER
Enable local control
Prints measured value
(Enabled when PRINT is selected as the INTERFACE setting on the SYSTEM screen)
Displays statistical calculation results
(Enabled when ON is selected as the STATISTICS set­ting on the DATA screen)
Sets the number of measurements for Auto-Memo­ry mode
(Enabled when ON is selected as the AUTO MEMORY setting on the DATA screen)
Prints statistical data
PRINT
UNDO
ALLCLR Deletes all memories and calculations
(Enabled when PRINT is selected as the INTER­FACE setting on the SYSTEM screen)
Deletes the previously stored measurement and calculation result (only the most recent measurement can be deleted)
2